Fit-For-Purpose Method Validation For Successful Biomarker Measurement To Support MRTP Applications Rafiq Islam Senior Director, Global Bioanalytical Services
Presentation Outline Definition of Modified Risk Tobacco Product (MRTP) Challenges to establishing MRTPs Biomarker as a tool to support Modified Risk Tobacco Product Application (MRTPA) Key considerations during biomarker method validation Key considerations during sample analysis Summary 2
Modified Risk Tobacco Product (MRTP) MRTP means any tobacco product that is sold or distributed for use to reduce harm or the risk of tobaccorelated disease associated with commercially marketed tobacco products. MRTPs can be introduced under risk modification order or exposure modification order Source: FDA Draft Guidance: Modified Risk Tobacco Product Applications 3
Modified Risk Tobacco Product (MRTP) Risk modification requires Evidence that MRTP significantly reduces harm and the risk of tobacco related diseases to individual tobacco users Benefit the health of the population as a whole Exposure modification requires Scientific evidence is not available and cannot be made available without conducting long term epidemiological studies for an application under 911(g)(1) Scientific evidence of substantial overall reductions in exposure to the harmful substance(s) Benefit the health of the population as a whole 4 Source: FDA Draft Guidance: Modified Risk Tobacco Product Applications
Challenges to Establishing Reduced Harm The concept of harm reduction/modified risk products is complex and sensitive Reducing the content of the toxin reduce exposure of toxin reduced harm Harm reduction strategies must be measured by their effects on health outcome Most important health outcome associated with tobacco are delayed in onset Lack of understanding of biological events from chronic smoking to disease manifestation 5
Challenges to Establishing Reduced Harm Epidemiological studies are the gold standard Epidemiological studies require; Long times due to long latency period of smoking related diseases such as cardiovascular diseases, COPD, lung cancer etc. to observe clinical end point Large number of subjects with well-matched confounding factors such as age, sex, ethnicity, and other lifestyle factors Subject compliance to the study protocol over long period of time Low subject attrition rate 6
Challenges to Establishing Reduced Harm MRTP development requires faster and more controllable method to assess Biomarkers offer an alternative and cost-effective approach for evaluation of potential harm reduction from MRTPs during product development The profiles of biomarkers may be used to understand biological events from smoke inhalation to disease manifestation Endpoint is the biological effect in response to smoking as opposed to disease manifestation 7
Biomarker Definition Biomarker is a good example of a term whose dictionary definition is not keeping pace with the word s changing significance in the real world The US National Institutes of Health definition: a characteristic that is objectively measured and evaluated as an indicator of normal biological processes, pathogenic processes or pharmacological responses to a therapeutic intervention 8
Types of Biomarkers Biomarker of exposure - potential to interact with a biologic macromolecule Biomarker of biologically effective dose - binds to or alter a macromolecule Biomarker of biological event with potential to lead to harm - measurement of an effect attributable to exposure Biomarker of harm - generally diagnostic biomarkers used to identify the presence of disease 9
Biomarkers to Support MRTPA Types of biomarkers based on the intended purpose Decisive Supportive Explorative Biomarker validation processes are continuous and iterative, and driven by the intended purpose of the biomarker data Fig: Conceptual diagram of fit-for-purpose method validation. Source: Lee et. al (2006) 10
Method Validation Bioanalytical Guidance assay must be: 11 Selective Sensitive Accurate Precise Stable Sample collection and handling Freezer (-20 o C or -80 o C) Freeze/Thaw UV light sensitivity Benchtop Pre-extraction Post-extraction DOCUMENTED!
Sample Collection Procedure Sample collection vials must be tested for presence of analyte Example- Some sample collection vials may contain 1-4 ng of nicotine Non-specific binding of analytes to the collection vials should be tested and detected early and resolved prior to sample collection Proper additives need to be chosen and adsorption test should be performed Allocate enough time for bioanalytical scientists to troubleshoot if non-specific binding or leaching is observed 12
Method Validation - Selectivity The analytical method should be able to differentiate the analyte of interest and IS from endogenous components in the matrix and other components in the sample Metabolite analysis primarily a concern with nicotine and TSNAs. As pharmacokinetic evaluations are observed with nicotine in serum or plasma, it is expected that the assay validation data demonstrate sufficient selectivity in the presence of high concentration metabolites In the presence of low concentration target analytes, high concentration metabolite concentrations should be supplemented to ensure conversion during collection, sample extraction, and detection does not occur Alternatively, back-conversion of metabolites can be checked by applying incurred sample reanalysis (ISR) or incurred sample stability (ISS) 13
Method Validation - Selectivity Additional requirements regarding medications allowed during studies has been noted. Stability testing in the presence of con-meds should be evaluated. While this is clearly not possible for all tobacco constituents, it is common practice to include selectivity testing of an OTC cocktail when evaluating assay selectivity Evaluation of the impact of sample hemolysis and lipemia is now expected for assays using serum or plasma samples Acceptance Criteria: The response in individual matrices should be no less than 20% of the LLOQ for the analyte in at least 80% of the tested individual matrices 14
Method Validation - Carry-over and Contamination All sources of carry-over and contamination should be minimized Carry-over should be assessed by injecting blank samples after high concentration samples or calibration standard at the upper limit of quantification. All possible sources of contamination should be investigated. Nicotine is the most abundant organic compound emitted during smoking, deposits on indoor surfaces and lasts up to months. Example: For high sensitivity nicotine method it was necessary to rinse every surface (transfer tubes, pipette tips, injection vials, etc.) with methanol to minimize contamination. Acceptance Criteria: The response in the blank sample should be equal or less than 20% of the LLOQ for the analyte 15
Method Validation - Sensitivity Sensitivity (Lower Limit of Quantitation - LLOQ): LLOQ is the lowest concentration of analyte in a sample which can be quantified reliably, with an acceptable accuracy and precision The analyte signal of the LLOQ sample should be at least 5 times the signal of the blank sample For decisive (in some cases for supportive) biomarker, it is critical to have appropriate LLOQ. Just because it might be difficult or expensive, it does not provide a waiver Celerion has one of the most sensitive assays for the measurement of NNN It was found that >20% of the clinical samples were below LLOQ. It is difficult (if not impossible) to do comparative statistics when a lot of the results are <LLOQ New method is in development using state-of the-art MS technology 16
Method Validation - Precision The precision of the analytical method describes the closeness of the repeated individual measures of analyte Acceptance criteria: The coefficient of variations (CV) for with-in run imprecision and between-run imprecision should not exceed 15% (20% at LLOQ) For decisive biomarkers it may not be sufficient! Improving assay precision for biomarkers can have a significant impact on statistical results 17
Assay Variability's Impact on Number of Subjects 500 450 78 fewer subjects 400 Number of subjects 350 300 250 200 150 44 fewer subjects Full Variability 15% Less Variability 100 14 fewer subjects 50 0 1 2 3
Method Validation and Cross-Validation Minimum Requirements At Celerion the following tests are required per SOP for Bioanalytical Method Validation Validation Test Initial Validation Cross-validation Precision and Accuracy X X Multi-lot Matrix Effect / Matrix Factor X Blank Matrix Selectivity X X Hemolyzed Sample Matrix Effect X Turbid Sample Matrix Effect X Sensitivity Assessment X X Recovery Assessment X Long-term stability X Freeze/thaw and Short-term stability X X Post Preparative Stability X Stock Stability X Sample Collection and Handling Stability X Frozen Aliquot Storage Stability X Dilution Integrity X X Processed Sample Integrity X Stress Test X X Sample Shipping Stability X
Sample Analysis Standards & QCs Each batch must contain its own standard calibration curve Minimum of six different calibrator concentrations Range of the standard curve should reflect the expected range of the study sample concentrations 135 115 95 75 55 Clinical Sample Results Standard QC 35 15-5 -1 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 20
Sample Analysis Standards & QCs Each batch must contain its own standard calibration curve Minimum of six different calibrator concentrations Range of the standard curve should reflect the expected range of the study sample concentrations Each batch must contain a minimum of 3 QC concentrations in duplicate (low, middle and high) What s the difference between standards and QCs? Philosophically standards create the calibration curve and QCs demonstrate that non-standards can be accurately measured so they must be different Different weighings getting standards and QCs to match each other QCs are stored with study samples to verify sample integrity 21
"Caffeine" Mass(es): "195.169/138.200Da" PeakName: "none" Annotation: "" Comment: Index: 10 Sample Type: Standard Sample 20.0 ng/ml Concentration: Conc: 20.6 ng/ml Calculated Date: 8/3/2012 Acq. Time: 5:31:47 PM Acq. Modified: Yes Algorithm: Specify Parameters - MQ III Proc. Percentage: 80 Noise Factor: 2 Peak-Split. Largest Peak: No Report Width: 3 points Smoothing Window: 30.0 sec RT End Time: 2.81 min Intensity, cps 2400 2450 2300 2350 2200 2250 2100 2150 2000 2050 1900 1950 1800 1850 1700 1750 1600 1650 1500 1550 1400 1450 1300 1350 1200 1250 1100 1150 1000 1050 900 950 800 850 700 750 600 650 500 550 400 450 300 350 200 250 100 150 0 50 1.70 1.68 1.81 1.96 2.07 2.32 1.8 1.9 2.0 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 1.7 Time,min 2.67 2.83 2.93 3.30 3.37 3.46 "Caffeine" Mass(es): "195.169/138.200Da" PeakName: "none" Annotation: "" Comment: Index: 10 Sample Type: Standard Sample 20.0 ng/ml Concentration: Conc: 20.6 ng/ml Calculated Date: 8/3/2012 Acq. Time: 5:31:47 PM Acq. Modified: Yes Algorithm: Specify Parameters - MQ III Proc. Percentage: 80 Noise End Time: 2.80 min Intensity, cps 2450 2350 2400 2250 2300 1800 1850 1700 1750 1600 1650 1500 1550 1400 1450 1300 1350 1200 1250 1100 1150 1000 1050 900 950 800 850 700 750 600 650 500 550 400 450 300 350 200 250 100 150 0 50 1.70 1.68 1.81 1.96 2.07 2.32 1.8 1.9 2.0 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 1.7 Time,min 2.67 2.83 2.93 3.30 3.37 3.46 Sample Analysis: Chromatographic Integration Big Deal the vast majority of serious audit findings involve scientists modifying integration parameters for QCs to get them to pass acceptance criteria Base. Sub. Window: Peak Height: Min. Peak Width: Min. 1.00 min cps 500.00 sec 0.00 RT: 2.63 min Expected Relative RT: No Use Type: Int. Time: Retention Area: Height: Start Time: To Base Base min 2.67 counts 14099.428 cps 1.88e+003 2.50 min QC = +15.7% QC = +14.9% Sub. Window: 1.00 min Base. Factor: 2 2200 Peak-Split. Largest Peak: No Report Peak Height: 500.00 cps Min. Peak Width: 0.00 sec Min. Width: 3 points Smoothing Window: 30.0 sec RT RT: 2.63 min Expected Use Relative RT: Type: Int. Time: Retention Area: Height: Start Time: No Manual min 2.67 counts 12107.104 cps 1.77e+003 2.55 min 2150 2100 2050 2000 1950 1900 Consistent with other samples Original computer generated Inconsistent with other samples Manually drawn 22
Sample Analysis Chromatographic Integration Big Deal A significant number of serious audit findings involve scientists modifying integration parameters for QCs to get them to pass acceptance criteria. Best Practice Review all chromatographic integration and manually redraw baselines then lock-it-down prior to performing regression analysis. It should be very difficult to modify integration following regression analysis. 23
Sample Analysis - Acceptance Criteria Standard Curve Acceptance Must use the same regression model and weighting option that was chosen during method validation Must have an SOP in place that describes how to reject standard(s) Only sample concentrations that are within the range of your accepted standards can be reported QC Acceptance The predicted concentration of 2/3 of the QC samples must be within 15% of their nominal concentration (20% for ligand binding assays) At least 50% of the QCs must be accepted at each QC level 24
Sample Analysis: Batch Acceptance Criteria Blank (no internal standard) and standard zero samples Prepared with matrix that was free of interferences At least 50% of these samples must have a response that is <20% of the response of the lowest standard reject low standard Sample Dilution Can only report sample concentrations that fall within the range of the standard curve Sample concentrations that exceed the concentration of the highest standard must be diluted Must prepare dilution QCs and dilute them using the same respective dilution (for example, 1/10 dilution) Dilution QCs are used for acceptance of the diluted samples 50% of dilution QCs must be within 15% of their nominal concentration 25
Sample Analysis: Incurred Sample Reproducibility (ISR) Purpose is to demonstrate that re-assaying study samples is reproducible. Recommended by FDA and EMA Generally accepted practice: minimum 20 samples If n < 1000 samples then reanalyze 10% If n >1000 samples then reanalyze 100 samples + 5% of samples beyond 1000 Select more subjects with fewer samples per subject Acceptance Criteria: The difference between the repeat value and the original value < 20% (< 30% for ligand binding assays) for 2/3 of the samples Is this OK for biomarkers when precision is so important? 26
Sample Analysis: ISR Investigation of failed ISR testing has found the following common root causes Analyte stability Failure to bridge reference standard materials/stocks/ standards Metabolite conversion or selectivity concerns Analytical technique failure in the laboratory (i.e. dilution error or performance variability) While the original intent was to demonstrate consistent method performance, failure due to assay variability has infrequently been observed. 27
Sample Analysis Reporting Sample Concentrations Demonstrate the sample analysis plan was followed Must have an SOP to select samples for re-analysis Must show that the acceptance criteria for each batch was met Demonstrate inter-batch reproducibility: List all of the QCs by batch and perform statistical analysis (% mean deviation and %CV) on each QC level for the study study precision Report sample concentrations that were < the concentration of the lowest standard as < LLOQ Report sample concentrations that were > the concentration of the highest standard as >ULOQ Prepare a table by batch as either passed or failed 28
Summary Evidence that MRTP significantly reduces harm and the risk of tobacco related diseases to individual tobacco users and benefit the health of the population as a whole is REQUIRED. The profiles of number of biomarkers can be used to support reduced harm from MRTPs. Choosing appropriate panel of biomarkers is critical to the success of MRTP applications Fit-for-purpose Method Validation in a GLP-compliant lab is critical to the success of MRTP Applications Bioanalytical methods must be selective, accurate, precise, stable and documented! Principles of GLP along with most up to date industry practices should be utilized for sample processing and data reporting Incurred Sample Reanalysis (ISR) analysis a requirement Just because it might be difficult or expensive does not provide waiver for advanced method validation 29
Acknowledgment Dr. Ray Farmen Kirk Newland Curtis Sheldon Celerion Team 30
References Stratton K, Shetty P, Wallace R, Bondurant S, editors. Clearing the smoke. Assessing the science base for tobacco harm reduction. Washington DC : National Academy Press, 2001:636pp. Guidance for Industry Bioanalytical Method Validation, 2001 EMA Bioanalytical Guidelines (2011) Quarterly reports on 483 documents Pharmaceutical Outsource Solutions, Inc. 2009 White Paper on Recent Issues in Bioanalysis The Calibration and Validation Group, Bioanalysis (2010) 2(1), 53-68 Lee JW, Devanarayan V, Barrett YC, Weiner R, Allinson J, Fountain S, Keller S, Weinryb I, Green M, Duan L, Rogers JA, Millham R, O'Brien PJ, Sailstad J, Khan M, Ray C, Wagner JA. Fit-for-purpose method development and validation for successful biomarker measurement. Pharm Res. 2006 ;23(2):312-28.Critical topics in ensuring data quality in bioanalytical LC-MS method development, Bioanalysis (2010) 2(6), 1051-1072 31