BORON IN GLASS DETERMINATION USING WDXRF



Similar documents
Coating Thickness and Composition Analysis by Micro-EDXRF

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system

X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING

X Ray Flourescence (XRF)

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

Spectral distribution from end window X-ray tubes

APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS

Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist

Elemental Analyses by ICP-AES

Electron Microprobe Analysis X-ray spectrometry:

OLIVÉR BÁNHIDI 1. Introduction

BRUKER ADVANCED X-RAY SOLUTIONS GUIDE TO XRF BASICS

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

Vacuum Evaporation Recap

EDXRF of Used Automotive Catalytic Converters

Gold Refining and Coin Manufacturing at the Royal Canadian Mint

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

FTIR Instrumentation

Application Note # EDS-10 Advanced light element and low energy X-ray analysis of a TiB 2 TiC SiC ceramic material using EDS spectrum imaging

XRF MAPPING: NEW TOOLS FOR DISTRIBUTION ANALYSIS

h e l p s y o u C O N T R O L

RoHS Test Methods. Presented by: Bruce Peterson

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE

Glancing XRD and XRF for the Study of Texture Development in SmCo Based Films Sputtered Onto Silicon Substrates

ANALYSIS OF LEAD IN CANDLE PARTICULATE EMISSIONS BY XRF USING UNIQUANT 4

Introduction to Energy Dispersive X-ray Spectrometry (EDS)

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

13C NMR Spectroscopy

Raman spectroscopy Lecture

X-Ray Study of Soft and Hard Magnetic Thin Films

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Surface Characterization - Marie-Geneviève Barthés-Labrousse

Lectures about XRF (X-Ray Fluorescence)

3 - Atomic Absorption Spectroscopy

Raman and AFM characterization of carbon nanotube polymer composites Illia Dobryden

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Physics 441/2: Transmission Electron Microscope

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

DATA MINING WITH DIFFERENT TYPES OF X-RAY DATA

How to measure absolute pressure using piezoresistive sensing elements

Introduction to Fourier Transform Infrared Spectrometry

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Field portable XRF analysis of environmental samples

Energy Dispersive Spectroscopy on the SEM: A Primer

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER , CAS NUMBER ] IRON ORE PELLETS

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff

Quantifying Amorphous Phases. Kern, A., Madsen, I.C. and Scarlett, N.V.Y.

THEORY OF XRF. Getting acquainted with the principles. Peter Brouwer

Analytical Testing Services Commercial Price List ManTech International Corporation January 2016

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

Copyright by Mark Brandt, Ph.D. 12

Experiment 5. Lasers and laser mode structure

Using the PDF for material identification using elemental data. from XRF and SEM EDS.

- particle with kinetic energy E strikes a barrier with height U 0 > E and width L. - classically the particle cannot overcome the barrier

Instrumentation. (Figure 2)

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

How To Understand Light And Color

Neuere Entwicklungen zur Herstellung optischer Schichten durch reaktive. Wolfgang Hentsch, Dr. Reinhard Fendler. FHR Anlagenbau GmbH

Solid State Detectors = Semi-Conductor based Detectors

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

Coating Technology: Evaporation Vs Sputtering

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF)

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

The Physics of Energy sources Renewable sources of energy. Solar Energy

Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

Radiation Detection and Measurement

Lecture 9, Thermal Notes, 3.054

The Basics of Scanning Electron Microscopy

WAVES AND ELECTROMAGNETIC RADIATION

Acousto-optic modulator

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF

Introduction to Elemental Analysis by ED-XRF Justin Masone Product Specialist 3 June 2015

2. Deposition process

X-ray Techniques: Overview. Ron Jenkins

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation?

Laser sintering of greens compacts of MoSi 2

Oxford Instruments Analytical technical briefing. Wavelength Dispersive X-ray Microanalysis

Preface Light Microscopy X-ray Diffraction Methods

Introduction to X-Ray Powder Diffraction Data Analysis

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

CHEM 1411 Chapter 5 Homework Answers

1. Photon Beam Damage and Charging at Solid Surfaces John H. Thomas III

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale

5.33 Lecture Notes: Introduction to Spectroscopy

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

The study of structural and optical properties of TiO 2 :Tb thin films

Experiment: Crystal Structure Analysis in Engineering Materials

Back to Basics Fundamentals of Polymer Analysis

New Portable X-Ray Diffraction/X-Ray Fluorescence Instrument (XRD/XRF)

For Touch Panel and LCD Sputtering/PECVD/ Wet Processing

DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY

Transcription:

269 ABSTRACT BORON IN GLASS DETERMINATION USING WDXRF Alexander Seyfarth Bruker AXS Inc., 5465 East Cheryl Parkway, Madison, Wisconsin 53711-5373 The application of wavelength-dispersive XRF to the analysis of boron in glass is reviewed with respect to physical, instrumental, calibration, and maintenance considerations. Advances in XRF instrumentation have enabled improvement in the analysis of boron. Reduced window thickness and high-current capabilities of X-ray tubes, as well as new analyzer crystals, produce a higher intensity boron signal. With correct determination of background and matrix interferences for successful quantification, only the sample itself limits the analysis. Observed boron migration in some glasses needs to be understood in order to establish stable, routine quantification in a process environment. INTRODUCTION Boron is an important component of glass. It exerts a strong influence on the viscosity of the melt, imparting a beneficial effect to the chemical resistance and lowering the expansion coefficient of the finished article. Boron is widely used in the production of E type glasses (7 wt%), neutral glass for pharmaceutical applications (7 to 11 wt%), and Pyrex -type borosilicate glasses (13 wt%). Because of the high volatility of boron in the melting furnace, a very accurate procedure is needed to meet the strict boron concentration specifications required for the finished product and to permit any necessary batch adjustment to be implemented almost in real time. XRF is well suited for glass analyses. Since the 1980s, XRF has been the method of choice because of its ease of sample preparation and rapid results. In the production control of soda-lime glass, the technique has already replaced wet chemical techniques because of its high precision and fast measurement time. Boron, being a very light element, poses challenges to the analyst as well as to the instrument. Demand for boron analysis in the glass industry prompted the development of new generation wavelength-dispersive X-ray fluorescence (WDXRF) spectrometers. ASPECTS OF THE ANALYSIS OF BORON To put the difficulty of the analysis of boron in glass into perspective, consider its physical and chemical properties [1 3]. Physical data for boron is given in Table I. Table I. Physical data for boron. Atomic number (Z) K shell absorption edge (ev) Kα radiation (ev) Mass absorption coefficient µλ (cm 2 /g) Fluorescence yield (Ω) 5 192 183 3,350 1.25 10-4 Boron s low atomic number and low K-shell energy cause various issues for XRF analysis which will be addressed further.

This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website www.dxcicdd.com ICDD Website - www.icdd.com

270 EXCITATION AND FLUORESCENCE YIELD The emitted intensity, I B-Kα, of the characteristic fluorescence radiation is governed by the overlap of the excitation spectrum and the absorption coefficient µ λ of boron [4]. Absorption by the tube window cuts off the low energy (long wavelength) part of the excitation tube spectrum. Because of this, boron is excited with the primary X-ray continuum and the rhodium L lines. For optimum primary continuum, the tube window must be as thin as possible. Currently window thicknesses range from 150 to 30 µm; 75 µm has become the industry standard since it balances price, performance, and durability. Fluorescence yield is another important factor in the analysis of boron. Fluorescence yield is proportional to Z 4 and is therefore close to zero for low atomic number elements. Using the highest available tube current at the lowest possible voltage, the emission of boron photons from the sample can be optimized. Starting in 1988, the possible settings for tube current have increased from 100 ma to today s maximum of 170 ma [Siemens SRS 303 (1988) to Bruker S8 TIGER 4 kw (2006)]. Figure 1. Comparison of boron intensity in 3 kw and 4 kw WDXRF systems. As can be seen in Figure 1, 40% greater boron intensity was obtained from a boronphosphosilicate glass (BPSG) sample (4 wt% B) analyzed on a 4 kw instrument in 1999 than a 3 kw instrument in 1988 [SRS 3000 with 75 µm end window X-ray tube and 3 and 4 kw generator (1999)].

271 ANALYZED LAYER, SAMPLE PREPARATION, AND X-RAY BEAM PATH Because the energy of boron Kα radiation is so low (183 ev), only a thin 0.6 µm (600 nm) surface layer of the sample can be analyzed. The density and composition of the glass can, in some cases, reduce this down to 200 nm. The glass samples to be analyzed must therefore be prepared with a good surface finish. Mellott compiled and investigated the surface roughnesses of commercial glasses using Atomic Force Microscopy (AFM) and found that many are in the range from 2 to 0.2 nm [5]. Quantitative boron analysis on non-polished glasses was much less accurate and repeatable than on the same polished material (private communication in 2004 with Brian Simpkins, Alcan Packaging, formerly Wheaton Glass, Millville, New Jersey). Reducing the path length of the X-ray beam (tube-to-sample and sample-to-detector) and using thin-window flow proportional detectors, combined with a controlled vacuum of better than 0.1 mbar, enables higher yield of detected boron radiation. Comparing the published sensitivities from a 1998 dual-turret spectrometer with those from a direct-loading S4-type system using the same X-ray tube, the type of generator shows a substantial effect, attributable to the optimized beam path and window design [6 8]. Changes in the vacuum during the measurement affect the count rate of the detected boron and needs to be taken into consideration for those systems that have large sample compartments (e.g., dual-turret systems). In contrast, direct-loading systems can usually achieve a stable vacuum without changeover during the whole measurement period. These direct-loading systems feature pressure-controlled sample introduction. Measurement is started only when the pressure in the sample compartment reaches a certain level, resulting in constant pressure and higher, more stable count rates for boron. ANALYZER CRYSTALS Figure 2. Comparison of boron intensity collected with a traditional crystal versus multilayer crystal. Ricardo et al. showed in 2001 that with a new type of patented multilayer detector crystal, the boron signal can be raised by 28% [9]. The XS-B crystal depicted in Figure 2 is a LaB 4 C multilayer and is available from Incoatec GmbH (Gesthacht, Germany). Schuster et al. performed a feasibility study for the analysis of boron in a coating of BPSG, a common semiconductor material, in 1987 [10]. Using a WDXRF system equipped with a 125 µm endwindow X-ray tube, 80 ma tube current, and early multilayer crystal (OVO 160), he obtained a sensitivity of 2.7 counts/s per wt% B. The detection limit was reported to be 0.04 wt% for 600 s measurement time.

272 Using a 200 Å Mo-B 4 C multilayer crystal (OVO-B) and an S4 Series WDXRF system optimized for beam path, Mauser achieved a detection limit of 0.02 wt % B in 100 s and a sensitivity of 330 counts/s per wt% B on the same type of material [7]. Using the XS-B multilayer crystal on the same S4 instrument further improved the sensitivity to 420 counts/s per wt% B as was shown by Behrens in 2004 [11]. CHOICE OF COLLIMATORS AND BACKGROUND POSITIONS Multilayer crystals produce wide peak profiles and should be used with coarse collimators. The rule of thumb in WDXRF is to use the coarsest collimator for the lightest element detection. Since achieving a high count rate is paramount for the analysis, peak resolution and interference corrections are not an issue. This holds true for BPSG, but in order to create a meaningful widerrange calibration for commercial borosilicate and E glass as well, count rate needs to be balanced against resolution [12]. K Ll, Ca Ll Cl Ll, B KA1 O (3) Zr Mz Si LB/ S Ll Ln Ca Ca and K L l lines, as well as Ca Ln and Zr M lines, should be separated as shown in Figure 3. The effect of higher order reflections of O cannot be reduced much by using pulse height discrimination and needs to be minimized by collimation. With Ca and Zr present in the analyzed material, the selection of background positions becomes difficult. The approach from Feather and Willis is particularly useful in this regard, as it can be used to perform interference correction on the background [13]. CALIBRATION APPROACH <21.778 Y : 0 to 2.260 KCps 61.978> Figure 3. Interference on and around the boron K signal. Obtaining a sufficient number of glass standards, as well as including enough samples to perform the interference corrections needed, pose the biggest challenge to the analyst. The quality of the boron certification by Manitol-based volumetric means is very much dependent on the experience of the analyst with that methodology. Few commercial labs now offer Manitol-based boron analysis. Inductively coupled plasma (ICP) methods are increasingly being offered instead, but these methods are time- and cost-prohibitive because of the requirement for sample digestion. Few glass samples of defined boron composition are commercially available. Many borosilicate glass samples, such as those available from Breitlaender GmbH (Hamm, Germany), are just premelt batch and not chemically certified. Borosilicate glass suppliers are the best source of sample material. They can produce a range of experimental matrix for the calibration based on past batch-melt samples. A comparison and discussion of the quality of certification techniques for boron in glass is unfortunately absent from the literature.

273 Using over 40 glass samples of standard 32 mm size and a small 28 mm aperture, a calibration for boron was established from 0.00 to 19.35 wt%. The calibration has a standard error of calibration (SEC) of 0.266 wt%. The sensitivity is 94 counts/s per wt% B 2 O 3. An empirical calculation using a concentration-based correction approach produced better results than any fundamental parameters (FP) model, even with variable influence coefficients. FP for boron, as taken from the NIST-compiled tables, are not well characterized and would need further improvements. Of greatest importance is interference correction of both the background and peak position to obtain a good net intensity for boron. MAINTENANCE OF CALIBRATION In order to maintain the calibration, drift correction is needed to account for changes in detector gas, foil, and tube signal that can influence the intensity and therefore the analysis. A very interesting study by Guadagnino, Sundberg, Michiels, and Brochot showed that some glass types exhibit a migration of boron when measured repeatedly [14]. Figure 4 shows the change in B 2 O 3 content in Pyrex-type and E-type glass as a function of the number of WDXRF measurements. Borosilicate glass should therefore be used as a drift correction and QC sample only after it has been conditioned. Figure 4. Long-term stability of boron in E-type and Pyrex-type glass

274 CONCLUSION The analysis of boron in glass using WDXRF has dramatically improved since its first application as a process control tool in the late 1980s. Nevertheless, it is still one of the most challenging applications, since the analyst must gather enough reference material and optimize conditions to fit the spectrometer capabilities. The choice of the right collimator, background and peak counting positions, correct pulse height analysis settings, and interference correction method greatly influences the range and durability of the resulting calibration. Maintenance of the calibration and a daily check of the system and calibration need to take the stability of the glass standards into account. ACKNOWLEDGMENTS The author thanks Brian Simpkins (Alcan Packaging, formerly Wheaton Glass, Millville, New Jersey) for his time performing the scans and tests on the borosilicate glass samples, as well as for evaluating the XS-B and OVO-B crystals side-by-side in a process environment. Many thanks as well to my colleagues at Bruker AXS Larry Arias, Kai Behrens, Karl Mauser, and Karen Roscoe for contributing to the presentation and article. REFERENCES [1] Bearden, J. A. Rev. Mod. Phys. 1967, 39, 78 124. [2] Krause, M. O. J. Phys. Chem. Ref. Data 1979, 8, 307 327. [3] White, E. W.; Johnson Jr., G. G. X-ray Emission and Absorption Wavelengths and Two- Theta Tables (ASTM Data Series DS 37A); American Society for Testing and Materials: Philadelphia, 1970. [4] Shiraiwa, T.; Fujino, N. Jpn. J. Appl. Phys. 1966, 5, 886 899. [5] Mellott, N. P.; Brantley, S. L.; Hamilton, J. P.; Pantano, C. G. Surf. Interface Anal. 2001, 31, 362 368. [6] Mauser, K. E.; Uhlig, S. SRS 3400 Boron in BPSG. Bruker AXS GmbH Application Note, 1999. [7] Mauser, K. E. S4 PIONEER Boron in BPSG. Bruker AXS GmbH Application Note, 2002. [8] Buehler, A.; Seyfarth, A.; Schlotz, R. Introduction to XRF. Bruker AXS GmbH Training Manual, 2000. [9] Ricardo, P.; Wiesmann, J.; Nowak, C.; Michaelsen, C.; Bormann, R. Appl. Opt. 2001, 40, 2747 2754. [10] Schuster, M.; Müller, L.; Mauser, K. E.; Straub, R. Thin Solid Films 1988, 157, 325 336. [11] Behrens, K. New Applications with the XS-B; S4 PIONEER. Bruker AXS GmbH Application Note, 2004. [12] Uhlig, S.; Müller, L. Boron Analysis in Ceramic and Glass Industries using Wavelength Dispersive X-ray Fluorescence Analysis. Siemens Analytical Application Notes 326, 1991. [13] Feather, C. E.; Willis, J. P. X-Ray Spectrom. 1976, 5, 41 48. [14] Guadagnino, E.; Sundberg, P.; Michiels, D.; Brochot, D. A collaborative study into the determination of boron in glass using X-ray fluorescence (XRF) spectroscopy. ICG/TC2 report, 2006.