Introduction to Mass Spectrometry. W.M. Keck Biomedical Mass Spectrometry Lab

Similar documents
Aiping Lu. Key Laboratory of System Biology Chinese Academic Society

Mass Spectrometry Based Proteomics

Electrospray Ion Trap Mass Spectrometry. Introduction

Tutorial for Proteomics Data Submission. Katalin F. Medzihradszky Robert J. Chalkley UCSF

Absolute quantification of low abundance proteins by shotgun proteomics

Mass Spectrometry Signal Calibration for Protein Quantitation

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

La Protéomique : Etat de l art et perspectives

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

Introduction to Proteomics 1.0

Master course KEMM03 Principles of Mass Spectrometric Protein Characterization. Exam

Pesticide Analysis by Mass Spectrometry

Research-grade Targeted Proteomics Assay Development: PRMs for PTM Studies with Skyline or, How I learned to ditch the triple quad and love the QE

Proteomics in Practice

Introduction to mass spectrometry (MS) based proteomics and metabolomics

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

Session 1. Course Presentation: Mass spectrometry-based proteomics for molecular and cellular biologists

Molecular Cell Biology. Prof. D. Karunagaran. Department of Biotechnology. Indian Institute of Technology Madras

Principles and Applications of Proteomics

using ms based proteomics

Laboration 1. Identifiering av proteiner med Mass Spektrometri. Klinisk Kemisk Diagnostik

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

High resolution mass spectrometry (HRMS*) in Graz

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

LC-MS/MS for Chromatographers

Mass spectrometry. Dr. Kevin R. Tucker

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

ProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry

Application Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008

Methods for Protein Analysis

Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit

Challenges in Computational Analysis of Mass Spectrometry Data for Proteomics

Proteomic Analysis using Accurate Mass Tags. Gordon Anderson PNNL January 4-5, 2005

AMD Analysis & Technology AG

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

Introduction to Proteomics

ProteinPilot Report for ProteinPilot Software

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

Global and Discovery Proteomics Lecture Agenda

ANALYSIS OF PROTEINS AND PROTEOMES BY MASS SPECTROMETRY

Analyzing Small Molecules by EI and GC-MS. July 2014

Appendix 5 Overview of requirements in English

Building innovative drug discovery alliances. Evotec Munich. Quantitative Proteomics to Support the Discovery & Development of Targeted Drugs

SpikeTides TM Peptides for relative and absolute quantification in SRM and MRM Assays

Waters Core Chromatography Training (2 Days)

13C NMR Spectroscopy

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Introduction to Proteomics

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics

How Mascot Integra helps run a Core Lab

Functional Data Analysis of MALDI TOF Protein Spectra

TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature

Standard Mixture. TOF/TOF Calibration Mixture. Calibration Mixture 1 (Cal Mix 1, 1:10) Calibration Mixture 1 (Cal Mix 1, 1:100)

ProSightPC 3.0 Quick Start Guide

Summary of the characteristics of different mass analyzers

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

PROTEIN SEQUENCING. First Sequence

amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Chapter 14. Modeling Experimental Design for Proteomics. Jan Eriksson and David Fenyö. Abstract. 1. Introduction

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics

Agilent 6000 Series LC/MS Solutions CONFIDENCE IN QUANTITATIVE AND QUALITATIVE ANALYSIS

MaxQuant User s Guide Version

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Sample Analysis Design Isotope Dilution

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

How To Use Gc-Ms

Quan%ta%ve proteomics. Maarten Altelaar, 2014

A Hybrid Microchip/Capillary Electrophoresis Mass Spectrometry Platform for Rapid and Ultrasensitive Bioanalysis

Error Tolerant Searching of Uninterpreted MS/MS Data

Pinpointing phosphorylation sites using Selected Reaction Monitoring and Skyline

SPECIAL FEATURE: TUTORIAL

Proteomic data analysis for Orbitrap datasets using Resources available at MSI. September 28 th 2011 Pratik Jagtap

Quantitative mass spectrometry in proteomics: a critical review

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Retrospective Analysis of a Host Cell Protein Perfect Storm: Identifying Immunogenic Proteins and Fixing the Problem

Quantification of Multiple Therapeutic mabs in Serum Using microlc-esi-q-tof Mass Spectrometry

Ionization of amino acids

Transcription:

Introduction to Mass Spectrometry W.M. Keck Biomedical Mass Spectrometry Lab Moore Health Sciences Library Rooms 1335 & 1337 May 18, 2010

The Keck Mass Spectrometry Lab of the Biomolecular Resource Facility Jordan Hall Rooms 1105 and 1034 434 924-0070 Nicholas E. Sherman, Ph.D., Director Jim Farmar, Ph.D., Assistant Director Carla Castro, Ph.D., Postdoctoral Fellow Rachel Reuther, M.S., Laboratory Technician John Shannon, Ph.D., Director, Shared Instrumentation Core Facility 2

Course Objectives How a Mass Spectrometer Works. What a Mass Spectrometer Can Tell You About Your Biomolecule. Types of Mass Spectrometers Available in the Keck MS Lab. Services Available in the Keck MS Lab. How to Work With the Keck MS Lab to Maximize Your Research Success. Examples 3

How Does a Mass Spectrometer Work? 1. Forms Ions from Molecules 2. Analyzes Ions by mass to charge (m/z) 3. Detects the Separated Ions 4. Collects the Data 4

Sample How Does a Mass Spectrometer Work? 760 Torr Ionization Source Analyzer Detector Data System 10-3 to 10-10 Torr 5

Mass Spectrometer Varieties 6 Aebersold W.M. Keck & Mann MS Nature Lab 2003, 198

Forming Ions: MALDI (Matrix-Assisted Laser Desorption Ionization) MALDI Laser energy strikes co-crystallized analyte/matrix under vacuum and sputtering produces a plume/explosion of analyte and matrix ions Singly Charged Ions (usually) Usually Paired with a Time of Flight Analyzer. Ions Formed from molecules: H + or H - or H ++ or higher Reactive, so keep under vacuum. Moved in the MS by electromagnetic fields. Ions can be made to react with other species in the MS. 7 National Magnet Lab www.magnet.fsu.edu

Forming Ions: ESI ElectroSpray Ionization ESI Molecules separated via HPLC. Jet of highly charged droplets emerge from capillary tip Solvent is evaporated. Multiply Charged Ions enter the analyzer. Paired with Quadrupoles, Ion Traps, FTs and Orbitraps Analyzers. Ions Formed from molecules: H + or H - or H ++ or higher Reactive, so keep under vacuum. Moved in the MS by electromagnetic fields. Ions can be made to react with other species in the MS. 8 National Magnet Lab www.magnet.fsu.edu

Analyzing Ions: Analyzer Types Deceleration Voltage Time of Flight Ions fly along a long tube Diagram shows a reflector Same m/z ions : Slow ions : Fast ions Detector Reflectron Voltage Smaller ions get to the detector 1 st Analogous to a gel Non-resonant ions Resonant ions Detector AC and DC voltages Quadrupole Voltages on 4 rods are scanned as ions enter from source. For a certain voltage, only a narrow range of m/z will pass. Other m/z s are rejected. Result: a Scanning Mass Filter Usually used in a series of 3 9 www.chemistry.adelaide.edu.au Angeletti, Einstein CoM 2007

Analyzing Ions: Analyzer Types Ion Trap Traps ions in a 2D field Both MS and MS/MS analyses can occur. Ions are sequentially ejected to the detector High sensitivity, Good range Low resolution, Low accuracy Orbitrap (FT) Ions are trapped between a central & an outer electrode Ions oscillate to produce a current Current is transformed into a m/z High sensitivity, good range Very High Resolution & Accuracy 10 Thermo Fisher

Detecting Ions: How? Plate and Ion Multiplier Type For ToF, Quadrupole and Ion Trap Conversion Dynode: ions produce 2⁰ particles Electron Multiplier: 2⁰ particles generate multiple electrons High sensitivity and High range ICR Type For FT and Orbitrap Measures the current induced by ions rotating in a magnetic field or around an electrode. High mass accuracy, high sensitivity, & good range Detectors can only count 11

Your Biomolecule(s): What Information Do You Want? Molecular Mass Structure Sequence Identity Quantity 12

Key Concept: Telling One Biomolecule from Another SDS PAGE Detection: Sensitivity vs. Specificity Stain: sensitive but non-specific; weak dynamic range Ab: very sensitive & specific for an epitope 13 kda? 10 kda? 7 kda? 5 kda? Separation by MW and Charge Resolution: Distinguish one biomolecule from another of similar properties 2000 Da +/- 500 Smaller MWs, Better Resolution Accuracy: Compare Standard Proteins to Sample Proteins. How accurate? What are the MWs of the sample Bands? 13 Bio-Rad

Key Concept: MS of Peptides not the Protein Problem: Proteins are too big; Small but important differences in the MS aren t seen and are difficult to locate Solution: Digest the protein to produce peptides Trypsin is the preferred enzyme: Very well-characterized Efficient Produces good-sized peptides (10 amino acid avg length) Cleaves between KX or RX except if X is Pro 14

Key Concept: Mass Accuracy Is Important Measured Value vs. True Value. Increase Mass Accuracy: Decrease Number of Matching Molecules from the Protein Databases. Modern MS Accuracy: within 10 ppm. Practical Limits to instrument accuracy. Effect of Mass Accuracy/Tolerance on Peptide Mass Fingerprinting Result Search m/z Mass Tolerance (Da) # Hits 1529 1 478 1529.7 0.1 164 1529.73 0.01 25 1529.734 0.001 4 1529.7348 0.0001 2 Liebler, 2002 15

Ion Intensity Ion Intensity Key Concept: Isotopes Natural Abundance: 12 C: 98.93% 13 C: 1.07% Use Monoisotopic Mass (all 12 C molecule) For High Mass Accuracy MI Mass: 1085.55 Avg Mass: 1086.28 Mass Mass 16

Key Concept: Resolution for Mass Spectrometers The ability of a MS to distinguish one ion from another of very similar m/z s. Glucagon at Different Resolutions 3481.62 MH + MS Resolution @ 1,000 MS Resolution @ 3,000 MS Resolution @ 10,000 MS Resolution @ 30,000 Isotopes Resolved! Matrix Science 17

Intensity Key Concept: Dynamic Range of Mass Spectrometers 1000:1 Dynamic Range Biomolecules in cells and tissues are present in different concentrations. MS can detect biomolecules that differ in concentration by up to 10^5. m/z 18

Key Concept: Advantages of a Mass Spectrometer Accuracy: >10 part per million Resolution: up to 10^6 Range: greater than 10^5 difference in concentration MS as a Universal Detector gives Sensitivity & Specificity. 19

Key Concept: Tandem MS or MS/MS To get more information about an ion: Fragment the ion into Product ions. Obtain the MS of the fragments. From the fragments, deduce how they fit together. Tandem MS uses 2 or more Analyzers in series. For Peptides: Sequence Information 20

Trp-Ser-Val-Lys 519.2926 (MH + ) y2: 246.18 y3: 333.21 y1: 147.11 b3: 373.19 21 b2: 274.12 m/z

Trp-Ser-Val-Lys 519.296 (MH + ) m/z b + Ions y+ Ions Trp-Ser-Val + Lys Trp-Ser + Val-Lys Trp + Ser-Val-Lys 22

Keck MS Lab: Normal Resolution MS Bruker MALDI-ToF MS Available for Training & Use. Confirmation of mass for a biomolecule (500-150,000 Da). Protein confirmation via Peptide Mass Fingerprinting (PMF). Isotopic Resolution Accuracy can be 50 ppm. Sensitive. Limits: Cannot determine the sequence of a peptide Not good for complicated mixtures No quantification. Bruker MALDI ToF 23

Keck MS Lab: Normal Resolution MS Three Quadrupoles in Series Quantification of proteins, peptides and metabolites Selected Reaction Monitoring (SRM, aka MRM) measures intensity of a specific transition for each peptide Parent ion to Product Ion High Sensitivity Resolution: 1000 Thermo TSQ Triple Quad 24

Our Lab: High Resolution MS Double Ion Trap & Orbitrap Resolution: 1:400,000 (max) Mass Accuracy: 1 ppm Very fast Tandem MS for: Sequence of peptides Locates the PTM to a specific amino acid in the peptide Quantitation by chemical labeling of tryptic digests. Global proteomics Thermo Orbitrap Velos 25

Our Lab: High Resolution MS Thermo LTQ ICR FT: Ion Trap (LTQ) Ion Cyclotron Resonance Fourier Transform MS Resolution: >1:400,000 Mass Accuracy: 1 ppm Tandem MS for: Sequence of peptides Global proteomics Thermo FT ICR 26

Keck MS Lab: MS Services 1. Confirmation of peptide identity by MALDI mass measurements. 2. Protein identification from gel or solution by Peptide Mass Fingerprinting (PMF). 3. Analysis of protein mixtures from tissue and media. 4. Absolute quantitation of proteins using labeled peptides and selective reaction monitoring. 5. ID of post-translational modifications (PO 4, Ac, Methylation, etc.,) & sites. 27

Keck MS Lab: MS Services 6. High resolution, High Mass Accuracy measurements of peptides and small proteins. 7. Protein identification and sequencing from gel or solution by ESI-LC/MS/MS. 8. Proteomics: comparison of proteins in samples. 9. De Novo (manual) sequence analysis of novel proteins to enable cloning. 10. Identification of binding partners. 28

Keck MS Lab: Sample Types Analyzed Gel Bands Coomassie stain Silver stain* Fluorescent stains Proteins in solution Immunoprecipitations Tissue samples Biofluids Media 29

Your Sample: Keep in Mind Important: Talk with us before you begin your experiment. Use our online sample submission process (we can show you how). One band or one spot may contain several proteins and in varying amounts. Minimize contamination. 30

In the Keck MS Lab Sample Modification: Cleanup Chemistry Digest Chromatography Mass Spectrometer Raw Data Analysis Results are reported to you as Protein Matches 31

Bruker MALDI TOF Experiment Was the correct version of Ghrelin synthesized? Wanted: GSS(oct)FLSP*EHQRVQQRKESKKP*P*AKLQP*R Octanoyl moiety adds 126.1 Da to Ghrelin (C 8 H 14 O). Each Heavy Proline (P*) adds 6 Da for a total of 24 Da. Are all Prolines made with heavy C and N atoms? Has the Ghrelin lost the octanoyl group to form des-ghrelin? 32

Bruker MALDI ToF in Linear Mode MH + Ghrelin 4P*: 3393.94 Da Ghrelin: 3369.8866 Da des-ghrelin 4P*: 3267.84 Da Ghrelin MS Analyses W.M. Keck Biomedical MS Lab

Bruker MALDI ToF in Reflector Mode MH + Theory Ghrelin 4P*: 3393.94 Da Ghrelin: 3369.8866 Da MH + Found 3394.36 Da +0.42 Da +125 ppm

Orbitrap and FT MS Experiment

βγ Isoforms Binding to Adenosine A1 Receptor: Gαi1 Complex Use Orbitrap Velos or FT MS Use SILAC: Stable Isotope Labeling with Amino Acids in Cell Culture For Relative Quantification of proteins from two cell cultures Compares the tryptic peptides of proteins From cells grown with Heavy Arg and/or Lys ( 13 C for +6 Da for each present). To cells grown with Light Arg/Lys ( 12 C).

βγ Isoforms Binding to Adenosine A1 Receptor: Gαi1 Complex? γ β α Receptor α β γ The receptor:gαβγ ternary complex is involved in starting signaling cascades. G protein consists of αβγ trimer bound to a receptor. The βγ dimer binds to both receptor and the Gα subunit. The complex regulates many different signaling proteins through the βγ dimer. 20+ Gα isoforms. 5 Gβ Isoforms 12 Gγ isoforms. Do specific receptors choose particular combinations of α, β, and γ subunits?

βγ Isoforms Binding to Adenosine A 1 Receptor: Gα i1 Complex Using SILAC Building a trap for βγ dimers Adenosine A 1 Receptor Fused with Gα i1 & His-Flag Tag βγ dimers binds to receptor:gα i1 fusion protein γ β α Non-specific Ab to β for total quantitation. Which β and γ isoforms bound to Adenosine A 1 :Gα i1?

FLAG M2 β γ Quantification of βγ Isoforms in an R:G Complex Recombinant Baculovirus in Sf9 Cells Purify & Quantify β 1 γ 2 β 2 γ 2 β 4 γ 2 β 1 γ 5 β 1 γ 7 β 1 γ 10 β 1 γ 11 β 1 γ 12 Spike in 10 β equiv Each βγ SDS PAGE β Trypsin Digest HPLC Adenosine A1:Gα Fusion Protein human HEK Cells Heavy Lys and Arg γ Lyse Ppt β α Total Amount of β Determined by a non-specific β Ab γ Wash Elute 1 β equiv Total for all βγ β γ β γ β γ β γ β γ β γ 1. Select γ 5 Specific Peptides 2. Compare Areas Heavy vs. Light peptides 3. Calculate Relative Amount of γ 5 1. Repeat Steps 1-3 for each γ Isoform 2. Calculate % of Each γ Isoform vs. Total γ Isoforms

Relative Abundance Relative Abundance A1STD - g5 - γkvvqqlr 5 100 95 Peptide: KVVQQLR RT: 0.28-32.05 MA: 5019923 100 90 Area Comparisons Are Valid for Isotopic Peptides 441.8001 90 85 80 75 80 70 60 50 40 30 20 m/z= 435.7700-435.7900 Area: 5.02 x10^6 γ 5 from Sf9 (light) 70 65 60 55 50 45 40 35 10 0 100 90 80 70 60 50 40 30 20 10 0 AA: 29847721 m/z= 441.7900-441.8100 5 10 15 20 25 30 Time (min) Area: 2.98 x10^7 γ 5 from HEK (Heavy) HEK : Sf9 = 6:1 30 25 20 435.7263 Resolution: 435.7796 vs. 435.7263 m/z 438.7249 442.3019 15 10 5 0 436.2818 436.7833 437.2821 438.2233 439.2919 439.7930 440.7971 441.2985 442.8032 436 437 438 439 440 441 442 443 m/z

Relative Abundance A1STD - g5 - γvkvsqaaadlk 5 100 Peptide: VKVSQAAADLK 381.2375 95 90 85 80 75 70 65 60 55 50 45 381.5719 40 35 30 25 20 15 10 5 0 377.2240 377.5586 377.8929 378.2241 378.7259 379.1908 379.5655 380.2213 380.7243 380.9030 381.9062 377.5 378.0 378.5 379.0 379.5 380.0 380.5 381.0 381.5 382.0 m/z 382.2212

Relative Abundance Relative Abundance γa1std 11 g11 - SGEDPLVK Peptide: SGEDPLVK 100 95 90 422.7238 100 95 90 422.7238 x100 85 85 80 75 70 65 60 80 75 70 65 60 55 50 Dynamic Range: Able to Quantify a peptide present at 0.25%! 55 45 423.2250 50 45 423.2250 40 35 30 40 25 425.7333 35 30 20 15 10 423.7262 426.2350 25 20 5 0 424.2278 424.7297 427.2210 427.7229 422.5 423.0 423.5 424.0 424.5 425.0 425.5 426.0 426.5 427.0 427.5 428.0 m/z 15 423.7262 10 5 0 424.2278 424.7297 425.7333 426.2350 427.2210 427.7229 422.5 423.0 423.5 424.0 424.5 425.0 425.5 426.0 426.5 427.0 427.5 428.0 m/z

βγ Isoforms Binding to Adenosine A 1 Receptor: Gα i1 Complex γ Isoform % Total γ +/- γ 2 6.3 0.9 γ 5 77.5 2.1 γ 7 2.3 0.1 γ 10 2.1 0.2 γ 11 0.03 0.01 γ 12 11.9 1.2

Thermo Triple Quad MS Experiment Quantitation by SRM Selected Reaction Monitoring (SRM) or Multiple Reaction Monitoring (MRM). Proteins, peptides, metabolites and PTMs of proteins. No Ab needed Variants easily detected. Biomolecule are separated by chromatography. Measures the fragment ion intensities of: A biomolecule. An isotopic (Heavy) version of the biomolecule which was added in a known amount. Compares the two intensities to arrive at a relative amount for the natural biomolecule. 44

Thermo Triple Quad MS: SRM or MRM HPLC SRM A. Schmidt, P. Picotti and R. Aebersold 45

FAK: Phosphorylation at Ser, Thr and Tyr Focal Adhesion Kinase I: 1052 Amino Acids Involved in cell motility, proliferation and apoptosis; Phosphorylation of Ser, Thr and Tyr controls interactions with other proteins. What Amount of Ser 722 is Phosphorylated? Experiment Design: 1. Digest FAK to produce natural or Light peptides. 2. Add a known amount of each Heavy tryptic peptide. (synthesized with 13 C/ 15 N Arg). 3. Run Spiked digest on LC-MSMS (triple quad). 4. For each Heavy/Light peptide, Compare area from a transition: MS ion (precursor)/msms ion (product ion).

SRM MS of +/- Phosphorylated PGYPSPR RT: 0.00-40.00 100 80 60 40 20 0 100 80 60 40 20 0 100 80 60 40 20 SM: 7B PGYPSPR*: 392.2/466.3; y4 ion Area: 4.84 x 10^7 20 fmole added 2.42 x 10^6 units/fmol PGYPSPR: 387.2/456.3; y4 ion Area: 1.69 x 10^6 0.70 fmole obtained from FAK PGYPsPR*: 432.2/546.2; y4 ion Area: 1.34 x 10^6 20 fmole added 0.67 x 10^5 units/fmol RT: 27.09 AA: 48354016 RT: 27.07 AA: 1686537 RT: 27.42 AA: 1338749 NL: 2.44E6 1. Phosphorylated Base Peak m/z= 465.76-466.76 PGYPsPR F: + c NSI SRM ms2 392.205 [318.144-318.146, & PGYPSPR 466.264-466.266, co-elute. 629.327-629.329] MS Genesis t100318t01 2. Different intensities for Phosphorylated vs. Non-Pi ions. NL: 8.16E4 Base Peak m/z= 455.76-456.76 F: + c NSI SRM ms2 387.201 [318.144-318.146, 456.256-456.258, 619.319-619.321] MS Genesis t100318t01 3. PGYPSPR: 0.7 fmol 4. PGYPsPR: 4.5 fmol 5. Phosphorylation at FAK NL: 7.58E4 Ser722: Base Peak m/z= 87% 545.73-546.73 or 1:6.4 F: + c NSI SRM ms2 432.188 [318.144-318.146, (Plain/Pi) 546.230-546.232, 709.293-709.295] MS Genesis t100318t01 0 100 80 60 40 20 PGYPsPR: 427.2/536.2; y4 ion Area: 3.02 x 10^5 4.5 fmole obtained from FAK RT: 27.29 AA: 301613 NL: 1.45E4 Base Peak m/z= 535.72-536.72 F: + c NSI SRM ms2 427.184 [318.144-318.146, 536.222-536.224, 699.285-699.287] MS Genesis t100318t01 0 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 Time (min)

Summary Points A Mass Spectrometer makes ions and measures them accurately. A Mass Spectrometer Can Give: Molecular Mass Sequence & Structure Identity Quantity Different Mass Spectrometers for Different Information. The Keck MS Lab Services include: Confirmation Discovery Quantification Talk to Us about Your Research Goals 48

Supplementary Slides to Introduction to Mass Spectrometry (May 18, 2010) 49

Mass Spectrometer Parameters Domon & Mann Science 2006 51

Comparison of Mass Analyzers TOF mass filter ion-trap sector FT-ICR mass resolving power 10 3-10 4 10 2-10 4 10 3-10 4 10 2-10 5 10 4-10 6 mass accuracy 5-50 ppm 100 ppm 50-100 ppm 1-5 ppm 1-5 ppm mass range >10 5 10 4 1.5 10 5 10 4 >10 4 linear dynamic range 10 2-10 6 10 7 10 2-10 5 10 9 10 2-10 5 precision 0.1-1% 0.1-5% 0.2-5% 0.01-1% 0.3-5% abundance sensitivity up to10 6 10 4-10 6 10 3 10 6-10 9 10 2-10 5 efficiency (transmission duty cycle) 1-100% <1-95% <1-95% <1% (scanning) <1-95% speed 10 1-10 4 Hz 1-20 Hz 1-30 Hz 0.1-20 Hz 0.001-10 Hz compatibility with ionizer pulsed & continuous continuous pulsed & continuous continuous pulsed & continuous cost moderate to high relatively low low to moderate moderate to high moderate to high size/weight/utility requirements benchtop benchtop benchtop lab instruments lab instrument Angeletti, Eistein CoM 2007

Schematic of a MALDI Tof MS 53 http://kenickbiochem09.

Mass Spectrometry References UVa: Search mass spectrometry AND Sherman Google: Search Siuzdak AND Scripps Google: Search ASMS Google: IonSource National Magnet Lab/FSU: www.magnet.fsu.edu/education/tutorials/tools/ Books: Introduction to Proteomics, Liebler Protein Sequencing and Identification Using Tandem MS, Kinter & Sherman Computational Methods for MS Proteomics, Eidhammer, et al., 54

Terms m/z : mass of an ion divided by the number of charges Dalton: MW: Molecular Weight Angiotensin I = 1296.48 g/mole MH + (mono): MonoIsotopic Angiotensin I = 1296.6848 m/z (z = 1) MH + (avg): Average Angiotensin I = 1297.5115 m/z MH +3 (mono): MonoIsotopic Angiotensin I = 432.8998 m/z (z = 3) 55