ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards.



Similar documents
Micro Motion ELITE CMFS Series Meter

Chemical Measurement Solutions

ATEX Installation Instructions for Micro Motion ELITE CMFS Sensors

Oilfield Services Measurement Solutions

Best Practices for Reducing Transmitter Life Cycle Costs

CALIBRATION PRINCIPLES

Integrated Custom Systems

Measuring equipment in a laboratory.

Calibration Service Selection Guide

CMMs and PROFICIENCY TESTING

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards

A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

ISO and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi

Micro Motion 3098 Gas Specific Gravity Meter

AMS Suite: Intelligent Device Manager with the DeltaV system

MICRO MOTION 5700 TRANSMITTER SECURITY FEATURES AND BEST PRACTICES BY JASON LEAPLEY, MICRO MOTION, INC.

AN IN-SITU VERIFICATION TECHNOLOGY FOR CORIOLIS FLOWMETERS

High Level Importance: ISO Microbiology compliance

ISO/IEC QUALITY MANUAL

Smart Wireless Solutions. Intelligent Well Production Starts Here

RITTER Multiple Sonic Nozzle Calibration System

A.1 Sensor Calibration Considerations

This document was prepared in conjunction with work accomplished under Contract No. DE-AC09-96SR18500 with the U. S. Department of Energy.

Regulations for custody transfer measurements: meeting the needs? Focus on liquid flow measurement systems/applications

Building Commissioning

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist

Calibration Certificate

Custody Transfer Measurement. with the V-Cone Flowmeter

Fisher FIELDVUE Instrumentation Improving Safety Instrumented System Reliability

calibrate confidence with Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Abstract Introduction White Paper

Advances in Thermal Dispersion Mass Flow Meter Accuracy

Enhanced calibration High quality services from your global instrumentation partner

AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL

Case study Realistic and best uncertainties with humidity instruments

Micro Motion Net Oil Computer Software and NOC System

ISO17025 INTERPRETATION DOCUMENT FOR CEC TEST METHODS

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

Monitoring Water Quality For Today s Demanding World

AMS Suite: Intelligent Device Manager

unless the manufacturer upgrades the firmware, whereas the effort is repeated.

Operations Management and the Integrated Manufacturing Facility

ALACC Frequently Asked Questions (FAQs)

OVERVIEW OF STANDARD 7E & STANDARD 20

Gravimetric determination of pipette errors

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Micro Motion Model D and DL Coriolis Flow and Density Meters

EVALUATION AND ACCREDITATION OF SUPPLIER TEST FACILITIES GP 10

Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance

Agilent Electronic Calibration (ECal) Modules for Vector Network Analyzers

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors

OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Rosemount Orifice Plate in Wet Gas

Measuring instruments in industry.

Keysight Technologies Specifications Guidelines. Technical Overview

Software Verification and Validation

Grant Agreement No SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

Daniel Venturi Tubes and Flow Nozzles Brochure

Procedure for Equipment Calibration and Maintenance

FLUXUS Energy. Manage energy effectively using non-invasive measuring technology. Heat usage, chilled flows and compressed air flow rate

CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company

Baumann Little Scotty Bronze Control Valve

OpenEnterprise. Lifecycle Services

IFS ApplIcAtIonS For Document management

Application of the Orifice Meter for Accurate Gas Flow Measurement page 1. Application of the Orifice Meter for Accurate Gas Flow Measurement.

JOFRACAL Temperature and Pressure Calibration Software

ROC800L Liquid Application Software

American Association for Laboratory Accreditation

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES


Evaluating the Accuracy of Maxim Real-Time Clocks (RTCs)

MATRIX GEMINI LIMS AT WORK IN THE FOOD INDUSTRY

DeltaV SIS for Burner Management Systems

How to calibrate an RTD or Platinum Resistance Thermometer (PRT)

The Calibration Chain: Role of BIPM,

American Association for Laboratory Accreditation

Data Center.

Ultrasonic Meters: A Natural Choice To Measure Gas Flow

S-series DeviceNet Interface Card

Bristol Series 5453 Indicating Pneumatic Pressure Controllers/Transmitters

Uncompromising Customer Service. Baumann TM Sanitary and Aseptic Process Valves

System Integration. system integration at a glance. Manufacturing Capabilities. Control and Identity Systems

Achieving Functional Safety with Global Resources and Market Reach

Agilent Measuring Noninsertable Devices

ISO 9001:2008 QUALITY MANUAL. Revision B

1 Quality Assurance and Quality Control Project Plan

OPERATIONAL STANDARD

Weight Calibration ComparatorPac TM Simple Calibration Efficient Workflow Reliable Results

Quick Start Guide , Rev DA March Rosemount 8714D (Calibration Standard) Magnetic Flowtube Simulator

Models SFP33 & SFP32 Pressure Transmitters

Experis Gas Mixtures. When accuracy, stability and reliability are a must

Improving Chromatic Dispersion and PMD Measurement Accuracy

Maximize Your Up Time With A Preventative Maintenance Plan Designed For The Way you Work s s s s.

Cisco Nexus Planning and Design Service

POWER. Your Partners in Availability POWER

Agilent Automotive Power Window Regulator Testing. Application Note

ICES Guidelines for Profiling Float Data (Compiled January 2001; revised April 2006)

Process Measurement Assurance Program For U.S. State Metrology Laboratories 1

International Laboratory Accreditation Cooperation. Laboratory Accreditation or ISO 9001 Certification? global trust. Testing Calibration Inspection

Transcription:

MICRO MOTION WHITE PAPER BY M. LEE, D. STANDIFORD, MICRO MOTION, INC. A. PRUYSEN, EMERSON PROCESS MANAGEMENT ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards KNOWLEDGE.micromotion.com Abstract At FLOMEKO 2000 Micro Motion, Inc. (MMI) reviewed the concept of Transfer Standard Method (TSM) flow stands for the calibration of flow instruments, in particular the uncertainty analysis for calibration of Coriolis meters. This paper is an extension of the topic, discussing in depth the issues involved with ISO 17025 accreditation, maintaining calibrations, and traceability. The company has chosen to use Coriolis flowmeters as Global Reference Meters (GRM) to maintain the traceability on the TSM stands operating worldwide. To accomplish this ISO 17025 accreditation was successfully sought for the SSF Primary Flow Stand (PFS) that is used to calibrate the GRM's. This paper will discuss the experiences associated with accrediting the SSF Primary Flow Stand, the processes involved with creating a GRM, calibrating TSM stands worldwide and the implications to TSM stand uncertainty. This paper includes: Review of SSF Primary Flow Stand Uncertainty and Traceability Internal Proficiency Testing Methodology Practical experiences in attaining ISO 17025 accreditation Process for GRM calibrations Advantages to using the GRM process Keywords: Coriolis, traceability, ISO 17025, transfer standard, flow meter Introduction The three purposes of this paper are: Review the TSM stand calibration process used by the flowmeter manufacturer. Introduce the Global Reference Meter (GRM) process that is used to maintain traceability and calibration of the TSM stands. Discuss the ISO 17025 accreditation process the manufacturer went through in support of the traceability required for the TSM stands and the GRM process. TSM process Previously, at FLOMEKO 2000, the manufacturer introduced the use of TSM stands, using Coriolis mass flowmeters as primary reference standards, to calibrate its flowmeters. The review of the process will include: a) TSM operation b) Traceability, as discussed at FLOMEKO 2000 c) Uncertainty of PFS gravimetric Standing Start Finish (SSF) stand and TSM stand TSM operation Figure 1 is a simplified P & ID (process & instrument diagram) of the TSM design. Different TSM calibration stands with different flow rate capabilities are all of a similar design. All TSM stands use pneumatically powered clamping systems, which facilitate rapid installation of the UUT (unit under test refers to the meter(s) being calibrated and adjusted). The flow meters Tank Pump UUT Unit Under Test Temperature Pressure Control Valve QCM (Quality Check Meters) (4 or 5 meters) RM (Reference Meters) (4 or 5 meters) Figure 1. Simplified P & ID for a TSM Flowmeter Calibration Stand

MICRO MOTION WHITE PAPER Page 2 of 6 called RM (reference meters) and QCM (quality check meter) on each TSM stand are identical in size and calibration. For some meter technologies, the UUT can be more than one sensor installed in series. The UUT is calibrated and adjusted to the RM. When verifying a meter's performance, after determining its flow calibration factor, the TSM calibration stand with its automated programming checks the UUT's low flow performance by comparing it to that of one of the in-line reference meters smaller than itself. Each time a reading is obtained from the UUT, the RM and QCM meters are also read. This is one of the key features of the TSM stand design: The QCM provides a continuous, real time, first level check on the calibration process for the UUT and on the stability of the RM, as well as on other potentially confounding effects such as entrained air and leakage. PFS and TSM uncertainty As developed in Ref 1 the Calibration and Measurement Capability (CMC), formerly referred to as the Best Measurement Capability (BMC) for mass of the PFS was 0.012%, assuming a k value of 2 for a 95% confidence level. This result will be revisited when ISO 17025 accreditation of the PFS is discussed. Also developed in Ref 1 the CMC for mass of the TSM stands was 0.013%, assuming k=2. This result will not be developed any further in this paper. TSM traceability International mass standard - Paris USA mass standards at NIST Rice Lake (mass standards supplier) ISO 17025 Accredited Micro Motion mass standard NIST traceable Micro Motion Primary Stand scales Micro Motion Primary Stand, ISO 17025 Accredited Micro Motion Global Reference Meter Micro Motion RM meter for the TSM flow stand Micro Motion TSM flow calibration stand Figure 2. Original Calibration Traceability for TSM Stands Global reference meters: GRM process With the success of the TSM stands in MMI's USA and Asia Pacific facilities, the program was extended. TSM stands were installed and are operating at the company's Mexico manufacturing facility. Also TSM stands were installed and are in the process of being commissioned at the company's Netherlands facility. The company now has nine stands, at four facilities, on three continents to support, figure 3. Originally the concept was to maintain calibration and traceability through reference meter (RM) replacement (see figure 2 for traceability chain). Location of MMI TSM Calibration Stands by Summer of 2003 Figure 3. Calibration Stand Locations This methodology had several issues. First, the methodology for maintaining the calibration did not look at the "system". System integrity was maintained through secondary methods, a measurement assurance program (MAP) and on-board Quality Check Meters. Although adequate, the methods were secondary, not primary calibrations. Second, the company had originally decided to use its own Coriolis flow meters because of their long term stability, so un-installing and replacing ran counter to the underlying philosophy of the stands. Third, like all good meter installations, if the process is running as desired taking the line apart and installing a new piece of equipment is not efficient. To overcome these issues the company developed a process based on the transfer standard methodology, using what are called Global Reference Meters (GRM). The process involves using a set of reference meters, calibrated in precisely the same manner as the TSM RM's. These meters are then installed in the UUT position on the TSM stand and used to calibrate the corresponding flow path, i.e. 1 inch meter calibrates the 1 inch meter path, of the RM. These same meters are used to calibrate all of the TSM stands globally. The process has three advantages. First, the calibration process is performed with the GRM in the calibration position, providing a system view equivalent to a production calibration. Second, the calibration requires no alteration to the stand, therefore being less labor intensive and lowering the risk of secondary problems associated with meter replacement. Third, it provides a means of comparing the company's TSM

ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards Page 3 of 6 stands globally, an internal round robin. GRM traceability This process would then change the previously discussed traceability chain to the new scenario shown below in figure 4. The change in the traceability path are 1) the State of Colorado is replaced by the company's mass standard supplier, Rice Lake Weighing Systems, and 2) the TSM stands traceability is now maintained by the GRM, not the RM's original calibration. Initially the company plans on using the PFS stand in Boulder, Colorado, USA as a hub, bringing the meters back for verification calibrations between trips to the various calibration stands. Although there is no indication that this is required due to meter stability, it seems prudent initially to insure the process is operating correctly. The full primary calibration of the GRM's will be conducted annually. The final piece to ensuring traceability chain was to get ISO 17025 accreditation on the PFS calibration stand, located within the company's Measurement Technology and Test Lab (MTTL). This would then make the GRM's fully traceable through an accredited path all the way back to the international mass standard. 17025 Accreditation: quality system International mass standard - Paris USA USAmass massstandards at at NIST NIST Rice Lake (mass standards supplier) ISO 17025 Accredited Micro Motion mass standard NIST traceable Micro Motion Primary Stand scales Micro Motion Primary Stand, ISO 17025 Accredited Micro Motion Global Reference Meter Micro Motion RM meter for the TSM flow stand Micro Motion TSM flow calibration stand Figure 4. Revised Traceability for TSM Stands Prior to obtaining ISO 17025 accreditation the company had already attained ISO 9000:2000 certification. This certification covered the company's facilities in the USA, Mexico, The Netherlands, and China. All facilities are under the same certificate, with the same quality system. Initially, when the effort to obtain ISO 17025 was undertaken a stand-alone quality system was being developed. Given the rigorous nature of ISO 17025 the belief was that the way to assure compliance was a narrowly focused, dedicated quality system for the Measurement Technology and Test Laboratory (MTTL) Primary Flow Stand (PFS). Although very focused to the needs of the MTTL, it soon became apparent that a lot of work was being duplicated. (An example would be the creation of a Corrective/Preventative Action procedure for the MTTL when a company wide procedure already existed and was judged acceptable to ISO 9000 already.) Recognizing the fact that the needs of section 4 of the ISO 17025 standard were satisfied with the existing quality system the team was allowed to focus on the more technical aspects of accreditation, mainly section 5, technical requirements. The activities in the technical requirements concentrated on four areas: uncertainty, traceability, lab management and proficiency. 17025 Accreditation: uncertainty The full detail of the uncertainty analysis for the PFS was previously published at FLOMEKO 2000, reference 1. We are not going to revisit the entire analysis here, rather we will talk about the main component in maintaining the levels previously published, scale stability. The stand is comprised of three scales to provide the full span of desired flow rates, 0.2 kg/min to 3200 kg/min. Procedurally the scales are required to be calibrated annually, but to build up a history of performance, calibrations were performed approximately every six months. Figure 5 below, for the small scale is representative of these periodic calibrations. The mass standards the MMTL uses for the PFS have an uncertainty of 0.002%. This graph illustrates how the scales are essentially stable within the uncertainty of the mass standards used. Based on these results a 0.003% allowance for scale stability was figured into the uncertainty analysis for scale drift. 17025 Accreditation: traceability The reference standards that impact the uncertainty of the PFS include: mass standards, pressure transmitters, and temperature transmitters. As shown in figure 4, the company's mass standards are ISO 17025 traceable to the International mass standard in Paris. The pressure and temperature standards, used in real-time buoyancy corrections, are maintained through outside services. The pressure transmitters are calibrated on-site using an ISO 17025 accredited calibration service. The temperature calibration are a secondary calibration, where the secondary reference is sent out to the original supplier, ISO 17025 accredited, for calibration. 17025 Accreditation: lab management The operation of the 7000 lb/min Primary Flow Stand (PFS) within Micro Motion, Inc.'s (MMI's) Measurement Technology Test Lab (MTTL), brings with it a new set of requirements in

MICRO MOTION WHITE PAPER Page 4 of 6 0.006 0.005 0.004 6-Apr-01, Check With New Mass Standards 31-Aug-01, Periodic Check 8-Apr-02, Full Annual Calibration 10-Dec-02, 17025 Assessment Demonstration 0.003 % Error % Error 0.002 0.001 0.000-0.001 Mass Standard Uncertainty Uncertainty Budget -0.002-0.003-0.004 9 11 13 15 17 19 21 23 25 27 Nominal Mass, kg Figure 5. PFS Small Scale Calibration History (without adjustment) addition to those already in place for ISO 9000 requirements. Most notable of those, is that unlike the typical engineering flow stand that may be routinely modified to meet testing requirements, the PFS requires monitoring and formal revision control, for all software and hardware changes. For accreditation purposes, the PFS has both a major equipment list, and software revision documents, that are maintained by MTTL personnel. No equipment or software may be changed or modified without the approval of Quality management. The approval process for change requires that an assessment be done for the impact on the stands' performance, thus assuring that no effect adds to the stands uncertainty. Once the assessment is complete, before and after performance data is collected and analyzed for each change. If the analysis shows there is no effect, then the change is made permanent and the documentation is updated. The formal change process highlights one of the main challenges of the accreditation process, the accreditation on a single flow stand in a lab of many flow stands. Although the company demands thorough testing and justifications before modifying flow stands, the formal review process with the quality department was a departure from historical behavior. Issues with the PFS are more than "engineering" problems to be solved. Impact on calibrations, past and present, as well as performance implications must be taken into account. Training requirements for the PFS are different from the other flow stands as well. In addition to hands-on training, as done for all other engineering stands, test technicians that operate the PFS to perform primary calibrations must be able to competently explain the operating procedures to the MTTL manager. The MTTL manager will then quiz the technician for the proper understanding of what can affect calibration uncertainty. When satisfied, the MTTL manager will certify the test technician is competent to perform the PFS primary calibrations and place the appropriate training document in their personnel file. 17025 Accreditation: proficiency testing For proficiency testing there were several difficulties to overcome: The PFS is a Coriolis-specific stand MMI is a commercial enterprise whose most suitable test partners from the respect of technology are competitors The uncertainty attained, 0.012%, was so low as to greatly reduce the number of potential partners Prior to receiving accreditation, MMI was not as attractive as a test partner Currently the company is seeking suitable partners to perform proficiency testing. To monitor ongoing system quality the company performs internal proficiency testing. This is accomplished by using three control meters, one appropriately sized meter for each scale. Prior to accreditation these meters were calibrated periodically, but not on a scheduled basis. To satisfy the ongoing

ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards Page 5 of 6 needs of accreditation the control meters are now calibrated on a monthly basis. When analyzing the proficiency of independent labs or stands the proficiency testing formula shown below is used: the ISO 17025 accreditation of the PFS, to a very low uncertainty (0.012%), that is used to calibrate the GRM's. The process allows in situ calibration of the TSM stands, provides a global round robin of the TSM stands and has greatly simplified the logistics of maintaining the TSM stands. With: 1.0 U 95 LAB representing the CMC, with k=2, of the lab under test U 95 REF representing the CMC, with k=2, of the lab being used for reference LAB representing the calibration of the lab under test REF representing the calibration of the lab being used for reference For internal proficiency testing the LAB is the current calibration and the REF is the original meter calibration. The REF value is never changed, thus yielding a measure of the stands long term stability. Because the assumption of independence cannot be made for calibrations on the same stand, the standard formula was modified as shown: where: En (Internal) En (Internal) 0.0 8/1/2001 10/31/2001 1/30/2002 5/1/2002 7/31/2002 10/30/2002 1/29/2003 Figure 6. PFS Large Scale E n History References Test Date FLOMEKO 2000, Traceability and Uncertainty Analysis for a Calibration Process for Flowmeters, Using Coriolis Flowmeters as Reference by Robert DeBoom, Marc Buttler, Andrew Kolbeck and Aart Pruysen. Annex Certificate of Accreditation and Scope of Accreditation. U 95 (LAB,REF) indicates the potential uncertainty factor for correlation effects between calibrations on the same stand. r LAB,REF is the correlation coefficient, assumed to be 0.5 The process is then reviewed, with En<= 1 indicating a system where differences are not statistically significant. Figure 6 below is typical of the calibration results for the control meters. The control meter shown is for the largest scale. Assessment The company went through the assessment process on December 10 and 11, 2002. The accrediting body was A2LA. Following the successful assessment and review by the accrediting body, MMI received accreditation on February 8, 2003. The scope of accreditation is for calibration of Coriolis mass flow, at flow rates of 0.2 kg/min to 3200 kg/min, with a CMC of 0.012% (Annex 1). Conclusion Micro Motion was able to develop a process for maintaining traceability on the TSM calibration stands globally. This was accomplished by extending the use of its own Coriolis flowmeters as transfer reference standards. The process is enabled by

MICRO MOTION WHITE PAPER Page 6 of 6 ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards Presented at FLOMEKO 2000.micromotion.com Micro Motion supports PlantWeb field-based architecture, a scalable way to use open and interoperable devices and systems to build process solutions of the future. The contents of this publication are presented for informational purposes only and, while every effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, expressed or implied, regarding the products or services described herein or their use or applicability. We reserve the right to modify or improve the designs or specifications of our products at any time without notice. Micro Motion, Inc. USA Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado 80301 T (303) 530-8400 (800) 522-6277 F (303) 530-8459 www.micromotion.com Micro Motion Europe Emerson Process Management Wiltonstraat 30 3905 KW Veenendaal The Netherlands T +31 (0) 318 549 549 F +31 (0) 318 549 559 Micro Motion Japan Emerson Process Management Shinagawa NF Bldg. 5F 1-2-5, Higashi Shinagawa Shinagawa-ku Tokyo 140-0002 Japan T (81) 3 5769-6803 F (81) 3 5769-6843 Micro Motion Asia Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T (65) 6 777-8211 F (65) 6 770-8003 WP-00701 2004. Micro Motion, Inc. All rights reserved.