Instructions for Use. Device Description The AMPLATZER Vascular Plug II is a self-expandable nitinol mesh occlusion device (see Figure 1).



Similar documents
LifeStent XL Biliary Stent System

Bard LifeStar Biliary Stent System

ON-Q * Catheters and Introducers

When Procedural Support really matters. Navien TM A+ Intracranial Support Catheter

AERO. Tracheobronchial Stent Technology System

RELAY Plus Thoracic Stent-Graft System with Transport Delivery System

Contraindications: Malign or benign strictures in the upper part of esophagus close to the cricopharyngeal muscle.

Aspira* Peritoneal Drainage Catheter

PERIPHERALLY INSERTED CENTRAL CATHETERS (PICC) Fong So Kwan APN, Haematology unit Medical Department, QMH

Allium Ureteral Stent (URS)

Bard LifeStar Vascular Stent System

Wingspan Stent System with Gateway PTA Balloon Catheter

Jugular/Subclavian Vein Approach Instructions for Use

Aspira* Pleural Drainage Catheter

Policy and Procedure Flushing and / or Blood withdrawal Aspiration Procedure For PICC Line and Midline Catheters

GRAFTMASTER RX Coronary Stent Graft System PPL (2/1/13)

PERCUTANEOUS PD CATHETER IMPLANTATION SYSTEM

REVEAL LINQ LNQ11. Insertable Cardiac Monitor MRI procedural information. MRI Technical Manual

Subject: Magnetic Resonance Imaging (MRI) Information on Sorin Group Heart Valve Prostheses and Annuloplasty Devices

Hemostasis Solutions Boston Scientific is committed to improving patient care in the management of gastrointestinal bleeding.

OPERATING MANUAL CONTRACTOR STYLE AXIAL FANS

A4.7 Management of a totally occluded central catheter and persistent withdrawal occlusion (PWO)

WallFlex Biliary RX Stent. Fully, Partially and Uncovered Self-Expanding Metal Stents

Plasmax Plasma Concentration System BIOLOGICS

Tracheal Stent System

PATIENT INFORMATION BOOKLET

Overview of Newer Stent Devices for Aneurysm Treatment

Propane Conversion Kit PROPANE CONVERSION KIT INSTALLATION MANUAL

Morpheus SMART PICC CT Insertion Kit

Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery.

Bard Access Systems, Inc. How to Care For Your. Nursing Guide

NFC Communication Tray Model: HHX-IT3-Z

Guidance for implant removal. Straumann Dental Implant System

MyoSure Hysteroscopic Tissue Removal System MyoSure XL Tissue Removal Device Instructions for Use

BioFlo PICC. with ENDEXO Technology. Directions For Use A

I-140 Venipuncture for Blood Specimen Collection

Biliary and Endovascular Product Catalog

HUMULIN 70/30 KwikPen

Patient Information Guide Morpheus CT Peripherally Inserted Central Catheter

Instructions For Use CORDIS ENTERPRISE Vascular Reconstruction Device and Delivery System

Cook Celect and Günther Tulip Vena Cava Filter Patient Guide

ROTOR LOADER OWNER S MANUAL

Talent Thoracic Stent Graft with THE Xcelerant Delivery System. Expanding the Indications for TEVAR

The New Standard of Excellence, Redefining Deliverability, Versatility and Control

Plasmax Plasma Concentration System for Laparoscopic Roux-en-Y Gastric Bypass

Patient Information. PORT-A-CATH Implantable Venous Access Systems

Instructions For Use

How To Use An Ekosonic Endovascular System

Clinician Competency:

Vaxcel PICCs Valved and Non-Valved. A Patient s Guide

Percutaneous closure of paravalvular leaks EULOGIO GARCIA MD MADRID ~ SPAIN

DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual

Medications or therapeutic solutions may be injected directly into the bloodstream

User Manual THR840DUK Digital Thermostat

Achilles Tendon Repair, Operative Technique

EZ-PAGE Electrophoresis System USER MANUAL

USER MANUAL EN IN 7449 Power Tower insportline Power Tower PT500

Instructions for Use. Cordis S.M.A.R.T. CONTROL Vascular Stent System

Instructions and User Guide. Select Series: Vascular Access Ultrasound Training Models BPO100, BPBV110, BPP120, BPNB150. June 8, 2011 Rev 2.

Stainless Steel Single and Dual Circulation Kits

6900PTFX bioprosthesis Carpentier-Edwards PERIMOUNT Magna mitral pericardial

INSTALLATION INSTRUCTIONS. SharkBite. Connection System. The SharkBite System

Arrow NextStep and Arrow Cannon II Plus Chronic Hemodialysis Catheters

ELECTRIC WINE BOTTLE OPENER OWNER S MANUAL BO60. Visit our web site at

Global Water Instrumentation, Inc.

Instructions for Use HUMALOG KwikPen insulin lispro injection (rdna origin)

Kimberly-Clark Healthcare Catalogue Digestive Health. Digestive Health

PATIENT GUIDE. Understand and care for your peripherally inserted central venous catheter (PICC). MEDICAL

Replacement Heart Valve Product Description (Stented Tissue) Models Reference

149mm. Walk-Thru Assembly Gate. MODEL NO: 1161, 1167 Owner s Manual.

FINAL DOCUMENT. Global Harmonization Task Force. Title: Label and Instructions for Use for Medical Devices

Central Venous Catheters for Veterinary Technicians

Digital Fingerprint safe

To maintain a port of entry to venous flow when all available peripheral ports have failed.

Introduction... 3 Setting the Combination... 4 Securing the DEFCON CL... 5 Locking the DEFCON CL to your Notebook Computer...

Zero peel strength of acrylic pressure sensitive adhesive is at about 45 C (-50 F).

Deflectable & Steerable Catheter Handbook

Managing Your Non-Tunneled (Percutaneous) Catheter: PICC, SICC, and JCC. What is a PICC catheter?

UV100A Ultraviolet Air Treatment System

Section Two: Arterial Pressure Monitoring

Appendix L: Accessing/Deaccessing Implanted Central Venous Access Port

ANGIO-SEAL EVOLUTION VASCULAR CLOSURE DEVICE INSTRUCTIONS FOR USE

HUMULIN (HU-mu-lin) N

2009/JUL/20 at 9:18 a.m. Doc number: M710950B001 [Uni+] VALIANT THORACIC STENT GRAFT WITH THE CAPTIVIA DELIVERY SYSTEM Instructions for Use

Installation Instructions Hustler 6-BTV Trap Vertical

el-720 Carbon Dioxide Gas Leak Detector User s Guide

Instructions for Use

SUCTION, ASPIRATING TUBES

Retractable Screen. Replacement Screen Instructions

Installation and Operating Instructions Installation Instructions for SS EPE-316L Series

Peripherally Inserted Central Catheter (PICC) Patient Instructions

81000 Series Aluminum Gate Valve Installation and Repair Manual

CUESTA COLLEGE REGISTERED NURSING PROGRAM CRITICAL ELEMENTS

A PROCEDURAL GUIDE TO MIDLINE CATHETER INSERTION

RGC-IR Remote Gas Calibrator for IR400

Wireless Indoor/ Outdoor Thermometer

Extra Tall Wide Span Gate

A Guide for Patients Living with a Biliary Metal Stent

STATFlow. Direct-to-Drain System. For use with STATIM 2000 and STATIM 5000 Autoclaves. Installation Guide and Operator s Manual

Transcription:

Vascular Plug II Instructions for Use Device Description The AMPLATZER Vascular Plug II is a self-expandable nitinol mesh occlusion device (see Figure 1). A B Figure 1. AMPLATZER Vascular Plug II A. Nitinol mesh B. Screw attachment. Marker bands The device has a screw attachment for a delivery wire and radiopaque marker bands at both ends. The AMPLATZER Vascular Plug II is attached to a 135 cm delivery wire with a stainless steel screw. The AMPLATZER Vascular Plug II is packaged within a loader, pre-connected to the delivery wire in a hoop dispenser. A plastic vise is included with each device (Figure 2). A B D E A. Hoop dispenser B. Plastic vise. Delivery wire D. Loader E. Preloaded device Figure 2. AMPLATZER Vascular Plug II preloaded device. Plugs are available in the sizes referenced in Table 1. Does not contain natural rubber latex components *600198-002* 600198-002 09-2011 US 2009, 2011 AGA Medical orporation AMPLATZER is a registered trademark of AGA Medical orporation. AGA Medical products and technologies for which patents are granted and/or pending in the USA and/or other countries are listed at www.amplatzer.com/patents 1 Manufacturing facility: AGA Medical orporation 5050 Nathan Lane North Plymouth, MN 55442 USA +1.888.546.4407 Toll Free +1.763.513.9227 Phone +1.763.513.9226 Fax www.amplatzer.com

Table 1. Recommended delivery catheter or delivery sheath size Order Number a a. Some sizes not available in all areas. b. Refer to delivery catheter or delivery sheath manufacturer s product labeling for inner diameter (ID). Intended Use The AMPLATZER Vascular Plug II is indicated for arterial and venous embolizations in the peripheral vasculature. ontraindications None known. Vascular Plug II Diameter Fully Expanded (mm) Vascular Plug II Length Fully Expanded (mm) Minimum Delivery atheter or Delivery Sheath ID b (inch) Maximum Delivery atheter or Delivery Sheath ID b. (inch) Warnings The safety and effectiveness of this device for cardiac uses (for example, patent ductus arteriosus or paravalvular leak closures) and neurological uses have not been established. Potential Adverse Events Potential complications include, but are not limited to: death, migration of the device, stroke, or vessel perforation. Maximum Delivery atheter or Delivery Sheath Length (cm) 9-AVP2-003 3 6 0.056 0.067 100 9-AVP2-004 4 6 0.056 0.098 100 9-AVP2-006 6 6 0.056 0.098 100 9-AVP2-008 8 7 0.056 0.106 100 9-AVP2-010 10 7 0.070 0.106 100 9-AVP2-012 12 9 0.070 0.106 100 9-AVP2-014 14 10 0.086 0.106 100 9-AVP2-016 16 12 0.086 0.106 100 9-AVP2-018 18 14 0.098 0.106 100 9-AVP2-020 20 16 0.098 0.106 100 9-AVP2-022 22 18 0.098 0.106 100 Precautions This device was sterilized with ethylene oxide and is for single use only. Do not reuse or resterilize this device. Attempts to resterilize this device can cause a malfunction, insufficient sterilization, or harm to the patient. Use on or before the last day of the expiration month that is printed on the product packaging label. Patients with nickel allergy can experience an allergic reaction to this device. Do not use this device if the sterile package is open or damaged. This device should be used only by physicians who are trained in standard endovascular techniques. The physician should determine which patients are candidates for procedures that use this device. MR onditional 1 Through non-clinical testing, AMPLATZER devices have been shown to be MR onditional. A patient with an implanted AMPLATZER device can be scanned safely immediately after placement of the device under the following conditions: - Static magnetic field of 3 tesla or less - Spatial gradient magnetic field of 720 G/cm or less - Maximum MR system-reported, whole-body-averaged specific absorption rate (SAR) of 3 W/kg for 15 minutes of scanning During testing, the device produced a clinically non-significant temperature rise at a maximum MR system-reported, wholebody-averaged specific absorption rate (SAR) of 3 W/kg for 15 minutes of scanning in a 3-tesla MR system using a transmit/receive body coil. MR image quality may be compromised if the area of interest is in the exact same area or relatively close to the position of the device. Therefore, optimization of MR imaging parameters to compensate for the presence of this device may be necessary. 1. MR onditional as defined in ASTM F 2503-05. 2

Directions for Use 1. Access the vessel and perform an angiogram using standard technique to measure the diameter of the vessel at the desired occlusion site. 2. Select an AMPLATZER Vascular Plug II with a diameter approximately 30% 50% larger than the vessel diameter at the occlusion site. Ensure that the occlusion site has sufficient length to accommodate the deployed device length so that the device will not obstruct other vessels or anatomical structures. 3. Flush the hoop dispenser and loader with sterile saline until fluid exits the distal tip to purge air from the loader. 4. Remove the device (in the loader) and the delivery wire from the hoop dispenser. 5. Select a delivery catheter (see Table 1 for delivery catheter size). Note: If the inner diameter of the original access catheter is sufficient for the device size selected, that catheter may be used for delivery. Note: The delivery catheter length should be no more than 100 cm. 6. Advance the delivery catheter over the guidewire until the distal tip of the catheter is at the distal edge of the occlusion site. 7. Remove the guidewire. 8. Insert the loader into the delivery catheter through the Y-connector or hemostasis valve. 9. A single Y-onnector kit is recommended for use with the AMPLATZER Vascular Plug II. B. Braun Medical Inc., Order No. 610400. 10. Allow blood backflow or aspirate the system to ensure air is purged from the catheter and loader. AUTION: Do not overtighten Y-connector screw to avoid damaging loader. 11. Push on the delivery wire to advance the device into the delivery catheter. Remove the loader from the wire if desired. 12. Advance the delivery wire and device to the distal end of delivery catheter. Do not twist or rotate the delivery wire during advancement to ensure device does not prematurely detach. 13. Hold the delivery wire in place and slowly withdraw the delivery catheter to deploy the device at the occlusion site. 14. Verify position of the device. 15. If device position is unsatisfactory: - Stabilize the wire and re-advance the delivery catheter until the device is completely within the catheter. - Reposition and deploy, or remove the device from the patient. 16. If the device position is satisfactory: - Attach the plastic vise to the wire, and release the device, by rotating the delivery wire counterclockwise until it separates from the device. - Remove the delivery catheter and wire from the patient. Disposal The Instructions for Use are recyclable. Dispose of all packaging materials as appropriate. Devices can be returned to AGA Medical for disposal. ontact an AGA Medical representative or returns@amplatzer.com for instructions. Use solid biohazard waste procedures to discard devices. Warranty AGA Medical orporation warrants to buyer that, for a period equal to the validated shelf life of the product, this product shall meet the product specifications established by the manufacturer when used in accordance with the manufacturer's instructions for use and shall be free from defects in materials and workmanship. AGA Medical orporation's obligation under this warranty is limited to replacing or repairing at its option, at its factory, this product if returned within the warranty period to AGA Medical orporation and after confirmed to be defective by the manufacturer. EXEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY, AGA MEDIAL ORPORATION DISLAIMS ANY REPRESENTATION OR WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, INLUDING ANY WARRANTY AS TO MERHANTABILITY OR FITNESS FOR A PARTIULAR PURPOSE. See the Terms and onditions of Sale for further information. 3

Symbol Definitions The following symbols may appear on the device packaging: Symbol Definition Manufacturer ontains Phthalate EU authorized representative Reference number Product serial number Product lot number Use by date (Use on or before the last day of the expiration month noted on the product packaging.) Do not reuse Sterilized using ethylene oxide onsult operating instructions Keep dry Do not use if package is damaged Does not contain natural rubber latex components MR onditional Inner diameter Outer diameter Length 4

Usable length Hydrophilic coating Recommended delivery sheath/catheter dimensions Indication of conformity with the essential health and safety requirements set out in European Directives Federal law (USA) restricts this device to sale by or on the order of a physician (or properly licensed practitioner). 5