The Array Scanner as Microdensitometer Surrogate: A Deal with the Devil or... a Great Deal?



Similar documents
Diagnostics for Digital Capture using MTF

Assessment of Camera Phone Distortion and Implications for Watermarking

Video Camera Image Quality in Physical Electronic Security Systems

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Scanners and How to Use Them

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.

Digital Camera Imaging Evaluation

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Characterizing Digital Cameras with the Photon Transfer Curve

Bernice E. Rogowitz and Holly E. Rushmeier IBM TJ Watson Research Center, P.O. Box 704, Yorktown Heights, NY USA

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL

Module 13 : Measurements on Fiber Optic Systems

Using visible SNR (vsnr) to compare image quality of pixel binning and digital resizing

Why use ColorGauge Micro Analyzer with the Micro and Nano Targets?

Lecture 14. Point Spread Function (PSF)

Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

Measuring Line Edge Roughness: Fluctuations in Uncertainty

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

Overview. Proven Image Quality and Easy to Use Without a Frame Grabber. Your benefits include:

RGB Color Managed Workflow Example

The Effect of Network Cabling on Bit Error Rate Performance. By Paul Kish NORDX/CDT

A Comprehensive Set of Image Quality Metrics

DOLBY SR-D DIGITAL. by JOHN F ALLEN

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

KODAK PLUS-X AEROGRAPHIC Film 2402

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Basler. Line Scan Cameras

Dispelling the Myths of Color Scanning

Uses of Derivative Spectroscopy

Digital Halftoning Techniques for Printing

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

Experimental Evaluation of Museum Case Study Digital Camera Systems

Introduction to Fourier Transform Infrared Spectrometry

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

SR2000 FREQUENCY MONITOR

Coating Thickness and Composition Analysis by Micro-EDXRF

Application Note Noise Frequently Asked Questions

SENSITOMETRIC CHARACTERISTICS OF THE EOS C300 DIGITAL CINE CAMERA

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

PCM Encoding and Decoding:

Determination of source parameters from seismic spectra

Optimizing IP3 and ACPR Measurements

Digitization of Old Maps Using Deskan Express 5.0

The effect of temperature on the bandwidth of telecommunications cabling in commercial buildings. Focus on the IBDN System 4800LX from NORDX/CDT

MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE

Physics testing of image detectors

RF Measurements Using a Modular Digitizer

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

Film Scanner The term Film Scanner can refer to a dedicated slide and negative film scanner or to a capture type scanner.

White paper. In the best of light The challenges of minimum illumination

GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2

Author: Dr. Society of Electrophysio. Reference: Electrodes. should include: electrode shape size use. direction.

Using the Normalized Image Log-Slope

Physics Lab Report Guidelines

AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis

Color Management Terms

Encoders for Linear Motors in the Electronics Industry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

QUALITY CONTROL OF THE IUE FINAL ARCHIVE

Glencoe. correlated to SOUTH CAROLINA MATH CURRICULUM STANDARDS GRADE 6 3-3, , , 4-9

CS171 Visualization. The Visualization Alphabet: Marks and Channels. Alexander Lex [xkcd]

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

HP Smart Document Scan Software compression schemes and file sizes

Plastic Film Texture Measurement With 3D Profilometry

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices

AVR127: Understanding ADC Parameters. Introduction. Features. Atmel 8-bit and 32-bit Microcontrollers APPLICATION NOTE

Outline. Quantizing Intensities. Achromatic Light. Optical Illusion. Quantizing Intensities. CS 430/585 Computer Graphics I

Synthetic Sensing: Proximity / Distance Sensors

Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014

Basler scout AREA SCAN CAMERAS

Log-Likelihood Ratio-based Relay Selection Algorithm in Wireless Network

Basler beat AREA SCAN CAMERAS. High-resolution 12 MP cameras with global shutter

Overview. Raster Graphics and Color. Overview. Display Hardware. Liquid Crystal Display (LCD) Cathode Ray Tube (CRT)

Basler pilot AREA SCAN CAMERAS

Color holographic 3D display unit with aperture field division

Centrifugal Fans and Pumps are sized to meet the maximum

Untangling the megapixel lens myth! Which is the best lens to buy? And how to make that decision!

Digital Image Basics. Introduction. Pixels and Bitmaps. Written by Jonathan Sachs Copyright Digital Light & Color

RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA

Avalanche Photodiodes: A User's Guide

SHEET 1 COMMUNICATION NETWORKS COMPUTER AND TELEPHONE SYSTEMS

Big Ideas in Mathematics

The Optical Design of the High Resolution Imaging Channel for the SIMBIO-SYS experiment on the BepiColombo Mission to Mercury

Manual Analysis Software AFD 1201

Quick start guide! Terri Meyer Boake

TDS5000B, TDS6000B, TDS/CSA7000B Series Acquisition Modes

Gran Via Corts Catalanes, BARCELONA Tel Fax

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff

Four Wave Mixing in Closely Spaced DWDM Optical Channels

On the Operational Quality of Fingerprint Scanners

Fraunhofer Diffraction

KIP 3000 Series MULTIFUNCTION SIMPLICITY

Determining the Resolution of Scanned Document Images

PIXEL-LEVEL IMAGE FUSION USING BROVEY TRANSFORME AND WAVELET TRANSFORM

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS

Digital Imaging for Photographic Collections Foundations for Technical Standards

Transcription:

The Array Scanner as Microdensitometer Surrogate: A Deal with the Devil or... a Great Deal? Don Williams Eastman Kodak Company, Rochester, NY, USA, 14650-1925 ABSTRACT Inexpensive and easy-to-use linear and area-array scanners have frequently substituted as densitometers for lowfrequency (i.e., large-area) hard copy image metrology. A number of workers and standard s committees 1 have in fact advocated such use. Increasingly, scanners are also being tasked for high spatial frequency, image microstructure metrology, which is usually reserved for high-performance microdensitometers that use microscope optics, photomultiplier tubes (PMT), and log amps. It is hard to resist their adoption for such use given the convenience level. The high speed, large scan areas, auto-focus, discomfiting low cost, and low operator skill requirements of commercial off-the-shelf (COTS) scanners makes one question if their use for such purpose is somehow too good to be true though. To confidently judge their limitations requires a comprehensive signal and noise spatial frequency performance evaluation with respect to available driver options. This paper will outline and demonstrate evaluation techniques that use existing ISO metrology standards for modulation transfer function (MTF), noise, and dynamic range with a comparison to the performance of a Photometric Data Systems (PDS) microdensitometer. Keywords: MTF, dynamic range, microdensitometer, metrology, standards 1. INTRODUCTION Despite their half century legacy, microdensitometers continue to serve as de-facto standardized instruments for high spatial frequency photometric image metrology of both transmissive and reflective media. Microdensitometers however continue to excel in their flexibility and imaging performance. Independent aperture and sampling settings, calibration freedom, ease of operator intervention, and PMT/log amp detection make them invaluable as high-end research and development tools. For these reasons, they are unlikely to ever be completely replaced. But, costly or unattainable maintenance, high skill requirements, and slow speeds have forced many metrology centers to abandon them altogether in favor of their speedier and less-expensive brethren, the array scanner. Is this a deal with the devil though? Will the immediate convenience of rapid and low cost data collection allowed by array scanners haunt adopters with mid-grade imaging performance, cross-device inconsistency, and non-standardized driver settings? Or, is it really a great deal? Are the high dynamic range and resolution claims (summarily and incompletely specified in bits and sampling frequency ) of COTS scanners truly achievable in a scientific signal-to-noise ratio sense at such low cost? Using established scientific practices and standardized ISO protocols, a scanner imaging performance accreditation for metrology purposes will be demonstrated. While the methods described here can be applied to any device intended for quantitative metrology, the candidate COTS scanner chosen for comparison here is an Imacon 848 scanner in transmission mode. The single greatest reason for considering the Imacon 848 as a microdensitometer surrogate had been the claim of true 8000 dpi performance over a 6 frame 35mm film chop. A single frame RGB scan at this resolution will take about five minutes. On the microdensitometer the same resolution scans would take about 8 hours. Clearly, this almost 100X decrease was a factor. Three primary imaging performance metrics were measured on the Imacon 848. They are 1) Opto-Electronic Conversion Function (OECF), 2) noise and signal-to-noise ratio (SNR), and 3) modulation transfer function (MTF). All were measured according to protocols defined in ISO 14524, ISO 21550, and ISO 16067-1 respectively. The OECF and SNR metrics can be generalized to any scan resolution mode. The MTF results are format dependent however because the effective sampling resolution is dictated by a reduction optics design. Fortunately, the 35mm size format provides the highest potential resolution along with being a popular format for existing microdensitometer requests. So, the 35mm mode was selected for MTF characterization.

2. OPTO-ELECTRONIC CONVERSION FUNCTION OECF Unlike microdensitometers which usually encode signals in one of two ways - linear with transmission or density - the finished file encoding for scanners like the Imacon 848 is a function of the various settings available in the software driver. Only the effect of the gamma settings was of interest for this characterization and is described here. The generic driver settings were as follows. Highlight = 255, Shadow= 0, Gamma= variable, Sharpening = disabled, 16 bit capture The OECF results for the labeled gamma settings are shown in Figure 1. Though captured in 16 bits, the ordinate is labeled in 8 bit equivalent counts for ease of interpretation. Those wishing for the 16 bit equivalent should multiply the ordinate values by 255. The curves of Fig. 1 apply to all color channels equivalently. Figure 1 Imacon OECF curves ; gamma 1.5 thru 3.5 ; color channel equivalence The behaviors of the OECFs as a function of gamma are typical. Low gammas are characterized with high low-density slopes that quickly diminish at the high densities. This would suggest better signal detection for low density values using low gamma settings. Similarly, greater high-density slopes at high gamma settings might suggest their choice for tasks requiring high density signal detection. Qualitatively, note the increasing plateau behavior of the OECF at high gamma settings. While this would suggest a decrease in effective dynamic range, noise levels also need to be considered for such a measure. So, the OECF is a necessary but insufficient tool for evaluating a device s dynamic. The noise and SNR metrology for the Imacon 848 and PDS microdensitometer are covered in the next section. 3. NOISE AND SIGNAL-TO-NOISE RATIO (SNR) The measured noise levels for the Imacon 848 and PDS densitometer are shown in Fig. 2. The PDS noise levels remain fairly constant at 0.003 rms density over its range of operation and is clearly superior to the Imacon at densities above 1.5. The Imacon 848 outperforms the PDS up to densities of 1.5 but quickly degrades beyond a 0.02 rms density at a mean density of 2.60.

Figure 2 RMS density for Imacon and PDS microdensitometer Note that no distinction was made for the Imacon 848 density noise levels as a function of gamma setting. Though the rms noise in counts does change as a function of gamma, in terms of equivalent density it does not. This is supported further through the incremental SNR plots of Figure 3. The SNR and thus signal detection abilities of the Imacon 848 remain essentially unchanged for gamma settings between 1.5 and 3.5. This, of course, is the beauty of using incremental SNR as outlined in ISO 21550 and 21550. It is independent of gamma setting. Figure 3 Incremental* SNR for Imacon and PDS microdensitometer * Incremental SNR uses the slope of the OECF as its signal metric and is used to measure a systems ability to distinguish incremental rather than absolute signals. For example, how well a density of X can be detected from X + δx.

3. MODULATION TRANSFER FUNCTION - MTF The single greatest reason for considering the Imacon as a microdensitometer surrogate had been the claim of true 8000 dpi performance over a 6 frame 35mm chop. This claim is reasonably well supported through the MTFs of Figures 4 and 5. The Imacon Colorflex driver allows resolution selection for native and non-native resolutions. Only the native resolutions were tested. Standard resolution MTFs are shown in Fig. 5 while high resolution mode MTFs are illustrated in Fig. 4. For a given sampling frequency, no directional or spectral MTF differences were found. Therefore, the illustrated responses are the same for all colors and scan directions. Figure 4 - High Resolution Imacon MTFs Figure 5 - Standard Resolution Imacon MTFs A number of the MTFs for each mode are not classically shaped optical MTFs. That is, not characterized by a monotonically decreasing shape. The unorthodox behaviors at 1575, 2100, and 3150 dpi ( Fig. 5) and 2000, 2667, and 4000 dpi (Fig. 4) suggest that some form of forced processing is occurring. For these reasons, it was not recommended that these sampling resolutions be used when performing spatial metrology grade scans with the Imacon 848. Either the 6300 dpi (low res. mode) or 8000 dpi (high res. mode) samplings are more suitable. For comparison purposes, the measured MTFs of the microdensitometer using a 10µm, 6.7 µm and 5µm diameter circular aperture are plotted in Figure 6 along with the 6300 dpi and 8000 dpi Imacon 848 MTFs. This illustrates that a 10 µm circular microdensitometer aperture response best matches the Imacon 848 at 8000 dpi. The 6300 dpi match is slightly less congruous but probably still suitable for microdensitometer tasks using a 10-13 µm circular apertures. From theoretical considerations, circular aperture MTF responses are roughly equivalent to square aperture responses 20% smaller in size. ( e.g. 10 µm circular aperture response is approximately equal to an 8 µm square aperture response). The above two sampling frequencies are equivalently efficient in terms of bandwidth per pixel. This can be deduced by taking the ratio of each limiting resolution frequency 2 (10% MTF response ) to the half-sampling frequency. The half sampling frequency for 6300 dpi is 124 cy/mm. For 8000 dpi it is 157 cy/mm. These are the maximum useful frequencies that these sampling rates can support as dictated by the Nyquist criterion. The ratio of their measured limiting resolution ( maximum measured frequency), 100 cy/mm and 120 cy/mm respectively ( see Fig. 6), to their halfsampling frequencies ( maximum theoretical frequency) are roughly 80%. The equivalence and relatively high value of this ratio is a good indication that optics or image processing functions are not unduly limiting the spatial frequency potential of this scanner.

Figure 6 MTF comparisons between Imacon and PDS microdensitometer Similar thinking can be applied to any scanner where a continuum of sampling rates is allowed. Only the naïve among us would believe that increasing sampling rates would be followed by true unbounded spatial frequency response improvements. Take the example of the suite of MTFs for a flatbed scanner in Fig.7. Each MTF curve corresponds to a different sampling rate, as labeled, allowed by the scanner driver. MTF sampling rate progression for flatbed scanner 1 0.9 0.8 Response 0.7 0.6 0.5 0.4 600 dpi 900 dpi 1200 dpi 1500 and 1800 dpi 0.3 0.2 300 dpi 0.1 0 0 5 10 15 20 25 Frequency (cy/mm) Figure 7 MTF sampling rate progression for candidate flatbed scanner Though the sampling rates of Fig. 7 are equally incremented by 300 dpi, the frequency response improvements across sampling rates are not proportional. For instance, there is a very large MTF improvement in the 300-600 dpi transition while there is barely any in the 600-900 dpi transition. Indeed, there is a decrease in MTF performance in the 1200-1500 dpi transition. A simple contour plot technique can be used to more easily evaluate these sampling rate MTF behaviors as illustrated in Fig. 8.

0 Equi-Response MTF Contours ( 0.1 increments) 1.0 0.8 0.6 0.4 0.2 0.1 1900 1700 1500 1300 1100 900 700 500 Sampling Frequency (dpi) 300 2 4 6 8 10 12 14 16 18 20 22 24 Frequency (cycles/mm) * Dashed line indicates half-sampling rate for corresponding dpi 26100 Figure 8 - MTF contour plots for flatbed scanner Using the 10% MTF contour for demonstration, one can use Fig. 8 to easily track which dpi settings offer the most efficient bandwidth sampling. For instance, the relatively large rightward outcroppings at 600 and 1200 dpi suggest a possible native resolution detent of the scanner. Beyond 1200 dpi it is clear that the MTF performance dramatically decreases, even at the 1300 dpi level. This type graphic allows visualization of overall MTF performance for better sampling rate selection depending on the source object being measured. 4. SUMMARY For most of today s hardcopy media types, array scanners can easily substitute for micro-densitometers. But OECF, MTF, and noise characterizations are required beforehand to accredit their absolute photometric performance. It is often said, the devil is in the details. Characterizing these details with the protocols outlined in ISO 14524, ISO 21550, and ISO 16067-1 can make the difference between the deal with the devil and the great deal. This paper has provided examples of how these protocols can be used to map and identify the micro signal detection capabilities of array scanners. REFERENCES 1. D. René Rasmussen, et al, INCITS W1.1 Standardization for Evaluation of Perceptual Macro-Uniformity for Printing Systems, Proc. IS&T 2003 PICS Conference, IS&T, 96-101, 2003.. 2. Don Williams and P.D. Burns, Diagnostics for Digital Capture using MTF, Proc. IS&T 2001 PICS Conference, IS&T, 227-232, 2001.