Routine Ultrasound Equipment Tests for Quality Assurance. Contents



Similar documents
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Diagnostic Ultrasound Performance Testing & Accreditation

Ultrasound Phantoms: B-Mode, Doppler and Others. Heather M. Pierce, B.S. Norfolk, VA

INTRODUCTION. A. Purpose

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

Quality Control and Maintenance Programs

Prepublication Requirements

Scanning Acoustic Microscopy Training

Breast MRI Quality Control

Physics testing of image detectors

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen.

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Quality Control of Full Field Digital Mammography Units

Ultrasonic Wave Propagation Review

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Checklist for the Qualification of Digital Detector Array Systems

Robot Perception Continued

WHERE IN THE WORLD JILL LIPOTI?

Pipeline External Corrosion Analysis Using a 3D Laser Scanner

Accreditation Is Coming

The promise of ultrasonic phased arrays and the role of modeling in specifying systems

Performance testing for Precision 500D Classical R/F System

MQSA Quality Control Manual for Monochrome Displays for Mammography

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions

Medical imaging monitors specification guidelines

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Automated Optical Inspection is one of many manufacturing test methods common in the assembly of printed circuit boards. This list includes:

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT

Synthetic Sensing: Proximity / Distance Sensors

A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality

Update on ACR Digital Mammography QC Manual

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy

Module 13 : Measurements on Fiber Optic Systems

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

Encoded Phased Array Bridge Pin Inspection

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany

BIOMEDICAL ULTRASOUND

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Staff Doses & Practical Radiation Protection in DEXA

Understanding Digital Modalities: System Integration and Use

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Digital Breast Tomosynthesis QC Requirements

Scanners and How to Use Them

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

Ultrasound - Vascular

Impact Echo Scanning Technology for Internal Grout Condition Evaluation in Post-Tensioned Bridge Ducts

Automotive Applications of 3D Laser Scanning Introduction

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Percutaneous Abscess Drainage

Preparation iagnostic Medical Sonographer Overview"

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications

MRI QA Technologist s Tests

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices

Coronis 5MP Mammo. The standard of care for digital mammography

How can I be sure it will be performed to a high standard?

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

How To Use An Ultrasound For Medical Research

NDT to Identify Concrete Bridge Deck Deterioration

Dental Radiography Core Subject. Digital Radiography

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

First floor, Main Hospital North Services provided 24/7 365 days per year

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Forensic Science: The Basics. Microscopy

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

SAR ASSOCIATED WITH THE USE OF HANDS-FREE MOBILE TELEPHONES

The Basics of Scanning Electron Microscopy

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Quantitative Imaging In Clinical Trials Using PET/CT: Update

Basics of Ultrasound Imaging

Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014

LIST OF CONTENTS CHAPTER CONTENT PAGE DECLARATION DEDICATION ACKNOWLEDGEMENTS ABSTRACT ABSTRAK

CURRICULUM VITAE. Kalpana M. Kanal, Ph.D., DABR

Technical data. General specifications. Indicators/operating means. 30 Hz Multiplex operation 30 Hz / n, n = number of sensors, n 5

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

Lecture 14. Point Spread Function (PSF)

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

1. General Description

Vivid S6. Cardiovascular ultrasound system. imagination at work. GE Healthcare

Transcription:

Routine Ultrasound Equipment Tests for Quality Assurance James A. Zagzebski, Ph.D. Nick Rupert, M.S. Ryan DeWall, M.S. Tomy Varghese, Ph.D. Dept. of Medical Physics University of Wisconsin, Madison, WI, USA Contents Elements of a Routine Test Program What system components are considered most important in routine testing? Concentrate on transducers! Steps to carry out tests on a regular basis Activity of Ultrasound Lab Accreditation Groups Basic QA methods Phantoms and tests 1

Routine QA Program Guidelines available from US facility accreditation bodies (ACR; AIUM) Emphasis is given to safety, cleanliness, and very basic performance tests. Issues related to image performance, such as spatial resolution, are not included (except ACR breast). Traditional parameters such as distance measurements are believed by many to be unnecessary. Includedi n some protocols, but only briefly. Routine QA (ACR General US Program) System sensitivity and/or penetration capability 1999: Do for 2 probes (?) Image Uniformity Look for element dropout Photography and other hard copy recording 2010: workstation monitor display Low contrast object detectability Assurance of electrical and mechanical safety and sterility Verify measurement accuracy during program initiation 2

American College of Radiology Requirements: General Sonography ACR physics committee developing this standard based it on measurements using a phantom. A phantom was selected (fall,1998) but turned out to be much more costly than originally promised. Physics Committee chose to leave the choice of phantoms to the user ACR staff published the following: Using a phantom will be helpful in responding to questions about low contrast detectability. However, the use of a phantom is optional at this time. Questions relating to characteristics associated with system sensitivity and image uniformity may be answered without the use of a phantom or test object. American Institute of Ultrasound in Medicine: General US QC 2008 Originally called QA in the Clinic Sonographers and clinicians helped draft Outlines what to do Sonographers Physicists/engineers Limited information on methodology Requires a phantom Phantom left to users See www.aium.org 3

American Institute of Ultrasound in Medicine: General US QC American Institute of Ultrasound in Medicine: General Sonography (physicist, engineer) Transducer Uniformity Phantom Electronic probe tester Sensitivity, Maximum depth of visualization into a phantom Distance measurement accuracy Target detection and imaging Focal targets such as simulated cysts or low contrast objects Choice of phantoms left to users Image display fidelity Cables OK, air filters clean, mechanical and electrical inspection Minimum frequency Annual for physicists/engineer tests Program includes daily, weekly checks by sonographers 4

Water-based gels Currently used materials: Advantages: Speed of sound = 1540 m/s Attenuation proportional to frequency Backscatter Disadvantages: Subject to desiccation Must be kept in containers Requires scanning window Currently used TM materials: Solid, non-water-based materials (urethane) Advantages: Not subject to desiccation No need for scanning window Produce tissue-like backscatter Disadvantages: C= 1430-1450 m/s Attenuation not proportional to frequency (~f 1.6 ) Surface easily damaged if not cleaned regularly to remove gels 5

Maximum Depth of Visualization Considered by many as a good overall check of the integrity of the system FOV at 18 cm (or set to match the phantom) Output power (MI) at max Transmit focus at deepest settings Gains, TGC for visualization to the maximum distance possible Maximum Depth of Visualization How far can you see the speckle pattern in the material? 6

Objective Maximum Depth of Visualization Gibson, Dudley, Griffith, A computerized Quality Control System for B-mode Ultrasound, Ultrasound Med Biol 27:1697 :1697-1711, 1711, 2001. Shi, Al-Sadah, Mackie, Zagzebski, Signal to Noise Ratio Estimates on Ultrasound Depth of Penetration (abstract only), Medical Physics 30: : 11367, 2003. Gorny, Tradup, Bernatz, Stekel, and Hangiandreou, Evaluation of an Automated Depth of Penetration Measurement ofr the Purpose of Ultrasonic Scanner Comparison, (abstract only), J. Ultrasound Med 23: S76, 2004. MPV Compute mean pixel value vs depth for phantom (signal) and for noise only (noise) Depth where signal/noise = 1.5 =DOP SNR -DOP tracks well Observer-DOP 16 Observers-DOP & SNR-DOP comparision (ATL C5-2) 15 14 13 Depth (cm) 12 11 10 (Al-Sadah, UW-Madison) 9 8 7 Observer's DOP DOP using SNR'= 1.5 0.0 0.2 0.4 0.6 0.8 1.0 1.2 Mechanical Index MI 7

Routine QA: Image Uniformity Considered to be the most impor- tant and useful test! Ideally: No evidence of element dropout No vertical shadows No loss of sensitivity near edges Uniformity Turn off spatial compounding, speckle reduction Use 1 tx focus Search for shadows emanating from transducer Common in new and old probes! 8

Coupling to a flat surface phantom scanning window with curvilinear transducers Coupling to a flat surface phantom scanning window with curvilinear transducers Solution 1: rock transducer from side to side 9

Coupling to a flat surface phantom scanning window with curvilinear transducers Solutside to side Coupling to a flat surface phantom scanning window with curvilinear transducers 10

Coupling to a flat surface phantom scanning window with curvilinear transducers Solution 2: Use a liquid TM material Coupling to a flat surface with curvilinear transducers 1 Sensitivity 0.8 Volts p-p 0.6 0.4 0.2 0 1 11 21 31 41 51 61 71 81 91 101 111 121 Elements Electronic Probe test Solution 2: Use a liquid TM material (pseudo liquid TM slurry; Madsen et al.) 11

Coupling to a flat surface phantom scanning window with curvilinear transducers Solution 3: Use a phantom having concave windows (Goodsitt et al) Visualizing 1-2 element dropouts Use persistence; translate transducer. 12

Visualizing 1-2 element dropouts Use persistence; translate transducer. Visualizing 1-2 element dropouts Record Image loops while translating the transducer; Compress a 100 frame loop into 1 image (averages speckle) 1 image Ultrasonix SonixTouch Enables masking off individual elements Elements 95 and 96 masked 13

Visualizing 1-2 element dropouts Record Image loops while translating the transducer; Compress a 100 frame loop into 1 image (averages speckle) Average image Ultrasonix SonixTouch Enables masking off individual elements Elements 95 and 96 masked Visualizing 1-2 element dropouts Use persistence; translate transducer Use Record Image loops while translating the transducer; compress a 100 frame loop into 1 image (averages speckle) Phased array transducer performance Need electronic probe tester Transducer test modes on the ultrasound instrument are possible; manufacturers need to be encouraged to provide these Future 2-D probes, probes that do part of the beam forming within the probe housing 14

Gray Scale Photography, Workstation Fidelity Important for monitor on machine to be set up properly to view all echo levels available and entire gray bar pattern. Set up during acceptance testing Take steps to avoid the meat hook effect (mark or inscribe contrast and brightness controls) Workstation monitors and/or hardcopy devices should be matched to the machine monitor. AAPM, July, 2010 Check hardcopy, workstation displays. Are all gray bar transitions visible? 15

SMPTE Test Pattern Available on most scanners 0% to 100% gray bar pattern Squares for detecting geometric distortion Are all gray transitions visible? Distance Measurement Accuracy: Vertical Actual 2.5 Measure 2.46 error 1.5% Acceptable 16

Distance Measurement Accuracy: Horizontal Actual 1.0 Measure 1.01 error <1% Acceptable Routine QA (ACR General US Program) Distance Measurement Accuracy tests as needed Necessary? ( Scanner is a transducer tied to a computer. ) May be important for specific uses Images reregistered from 3-D data sets Workstation measurements Radiation seed implants 17

ACR Recommended QC for Breast US Maximum Depth of visualization Vertical and horiz measurement accuracy Semiannually Semiannually Service eng./physicist Service eng./physicist Hardcopy Recording Semiannually Service eng./physicist Uniformity Semiannually Service eng./physicist Elec-mechanical condition Semiannually Service eng./physicist Anechoic void perception Semiannually Service eng./physicist Ring down Semiannually Service eng./physicist Lateral Resolution Semiannually Service eng./physicist QC Checklist Semiannually Service eng./physicist Adherence to infection control procedures After each biopsy Sonographer Clean transducers After each biopsy Sonographer Vertical and horiz measurement accuracy Quarterly Sonographer Grey-scale photography Quarterly Sonographer Dead Zone? No suitable targets for dead zone testing, even in phantoms that advertise deadzone tragets! 18

Anechoic Sphere Imaging, a better choice Anechoic Sphere Imaging 19

Image of a phantom is useful for qualitative comparisons! Conventional Spatial Compounding Images obtained during routine Breast QC testing, 3/2010 Image of a phantom is useful! Conventional Spatial Compounding Images obtained 1 month later, after a software change; 20

Routine QA: Cleanliness, safety Transducer Inspection Delaminations Frayed cables Proper cleaning Console Air filters Viewing monitor, keyboard clean Lights, indicators Wheels, wheel locks Proper cleaning Summary Setting up, maintaining an equipment QA program is straight forward The ACR listed procedures are a useful, basic QA program Directed by physicist or lab personnel Integrated effort including lab and technical staff Requires a Phantom Closely correlates with AIUM list of factors to test Transducer uniformity is a frequent fault in today s scanning machines Computational methods can be developed for objective tests 21

References Goodsitt et al, Real-time B-mode ultrasound quality control test procedures, Medical Physics 1998; 25:1385-1406. 1406. J Zagzebski, US quality assurance with phantoms. In Categorical Course in Diagnostic Radiology Physics: CT and US Cross-Sectional Imaging, Edited by L. Goldman and B. Fowlkes, 2000, Oak Brook, IL: Radiological Society of North America, pp. 159-170. 170. J Zagzebski and J Kofler, Ultrasound Equipment Quality Assurance, in Quality Management in the Imaging Sciences, ed. By J Papp, 2002, St. Louis, Mosby, pp. 207-215. 215. QA Manual for Gray Scale Ultrasound Scanners, 1995, American Institute of Ultrasound in Medicine, Laurel, MA. D Groth et al, Cathode ray tube quality control and acceptance program: initial results for clinical PACS displays, Radiographics 2001, 21: 719. 22