FACSCanto RUO Special Order QUICK REFERENCE GUIDE



Similar documents
GETTING STARTED. 6. Click the New Specimen icon.

Analyzer Experiment setup guide for LSRII and Canto s

Institut für Experimentelle Immunologie und Bildgebung

How to operate the BD FACSCanto flow cytometer

LSR II Users Guide Stanford Shared FACS Facility 06/10/13

UCHC Guide to FACS DiVa. LSR II instruments

Getting Started with BD FACSDiva Software

Navios Quick Reference

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit G June Beckman Coulter, Inc N. Harbor Blvd. Fullerton, CA 92835

BD FACSDiva TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY

Becton-Dickinson FACSCalibur Flow Cytometer Instructions

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract

BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin

1. Instrument Layout pg Summit Overview pg Data Storage and Data Files pgs Saving Data pg Start Up pg 8

BD FC Beads 8-Color Kit for BD OneFlow Assays

Attune Acoustic Focusing Cytometer

BD FACSComp Software Tutorial

BD FACSDiva Software Reference Manual

UCHC Guide to CellQuest Pro. FACS Calibur instruments

Science is hard. Flow cytometry should be easy.

Preparing BD Systems for Data Migration. Summary of the Upgrade Process on page 6. Copying User Data Files onto an Upgraded System on page 7

Company Setup 401k Tab

How to install and use the File Sharing Outlook Plugin

Establishing Optimum Baseline PMT Gains to Maximize Resolution on BD Biosciences Digital Flow Cytometers

BD FACSCanto II Instructions For Use

How To Create A Powerpoint Intelligence Report In A Pivot Table In A Powerpoints.Com

Hydras 3 LT Quick Start

OSF INTEGRATOR for. Integration Guide

BulkSMS Text Messenger Product Manual

Mail Merge Creating Mailing Labels 3/23/2011

BD CellQuest Pro Software Analysis Tutorial

Attune NxT Acoustic Focusing Cytometer

Cell Sorting. S3e Cell Sorter Quick Guide

SA-9600 Surface Area Software Manual

Microsoft Expression Web

Learn how to create web enabled (browser) forms in InfoPath 2013 and publish them in SharePoint InfoPath 2013 Web Enabled (Browser) forms

QIAGEN Rotor Gene Q: Software V2.0

owncloud Configuration and Usage Guide

Transitioning from TurningPoint 5 to TurningPoint Cloud - LMS 1

Creating a Participants Mailing and/or Contact List:

Mail Merge Microsoft Word and Excel Queries Scott Kern Senior Consultant

Quick Reference Guide Course Homepage Management (Faculty)

COMPENSATION MIT Flow Cytometry Core Facility

FACS-CALIBUR. USER-Guide

Hamline University Administrative Computing Page 1

WebSphere Business Monitor V7.0 Business space dashboards

BIGPOND ONLINE STORAGE USER GUIDE Issue August 2005

UF Health SharePoint 2010 Introduction to Content Administration

Customer admin guide. UC Management Centre

Creating and Issuing the Workstation Authentication Certificate Template on the Certification Authority

Adobe Dreamweaver CC 14 Tutorial

Integrating with IBM Tivoli TSOM

System Overview and Terms

CDOT Workflow ProjectWise Web Access Operations

Web Forms for Marketers 2.3 for Sitecore CMS 6.5 and

Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V.

Decision Support AITS University Administration. Web Intelligence Rich Client 4.1 User Guide

SQL Server 2005: Report Builder

Microsoft FrontPage 2003

ICP Data Entry Module Training document. HHC Data Entry Module Training Document

City of Dallas. WASP MOBILE ASSET Professional Edition Version 6.3 STEP-BY-STEP GUIDE

WebSphere Business Monitor V6.2 Business space dashboards

Migrating Groupwise Data

Use the Microsoft Office Word Add-In to Create a Source Document Template for Microsoft Dynamics AX 2012 WHITEPAPER

ORACLE BUSINESS INTELLIGENCE WORKSHOP

State of Nevada. Ektron Content Management System (CMS) Basic Training Guide

Crystal Reports Payroll Exercise

einstruction CPS (Clicker) Instructions

HHC Compensation Module Training Document

Senior Systems Cloud Services

Fairfield University Using Xythos for File Sharing

Kentico CMS 7.0 User s Guide. User s Guide. Kentico CMS

Working with H-drive, I-drive, and DropBox Files Online

Training Manual Version 1.0

Spherotech, Inc Irma Lee Circle, Unit 101, Lake Forest, Illinois

How Do I Upload Multiple Trucks?

SSL VPN Setup for Windows

Create a Simple Website. Intel Easy Steps Intel Corporation All rights reserved.

Sage Intelligence Report Designer Add-In

Network DK2 DESkey Installation Guide

How to Back Up and Restore an ACT! Database Answer ID 19211

Excel Pivot Tables. Blue Pecan Computer Training Ltd - Onsite Training Provider :: :: info@bluepecan.co.

emarketing Manual- Creating a New

User Manual Web DataLink for Sage Line 50. Version 1.0.1

Neoteris IVE Integration Guide

This exhibit describes how to upload project information from Estimator (PC) to Trns.port PES (server). Figure 1 summarizes this process.

Table of Contents. Speedi WIN Online Page 1 of 44 Speedi-WIN Online Install Instructions

LEGENDplex Data Analysis Software

Talend Open Studio for MDM. Getting Started Guide 6.0.0

Batch Scanning. 70 Royal Little Drive. Providence, RI Copyright Ingenix. All rights reserved.

Octet System Data Analysis User Guide

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

User Guide for TASKE Desktop

Mastering Mail Merge. 2 Parts to a Mail Merge. Mail Merge Mailings Ribbon. Mailings Create Envelopes or Labels

CyTOF2 Sample Acquisition: Super Sampler* (CyTOF software version )

Step by Step. Use the Cloud Login Website

HDAccess Administrators User Manual. Help Desk Authority 9.0

Budget Process using PeopleSoft Financial 9.1

Remote Storage Area (RSA) Basics

Transcription:

FACSCanto RUO Special Order QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW username and password) 3. Launch Tera Term, Click on Serial and make sure the Port is on COM1 and click OK First user of the day CST: 4. Turn on cytometer (Green button-left side), and launch FACSDiva Software. 5. Allow ~30 minutes warm-up for lasers. Continue while waiting. 6. Check sheath and waste levels. If emptying waste, add 1L household bleach and 500uL anti-foam reagent to the new waste tank. 7. Perform Fluidics Start-Up (~7 min.) (Cytometer>Fluidics Start Up) 1. Cytometer > CS&T. 2. Verify lot number. 3. Choose between load the tube manually or not 4. Run. Note: Failed CS&T is NOT acceptable. Note 2: 1 drop CS&T beads in 300uL FOR A NEW EXPERIMENT: USE COMP BEADS or SINGLE STAINED CELLS: 1. Open a new experiment. Select the experiment titled Blank Experiment. 2. Click cytometer settings. Delete all parameters then insert experiment specific fluorophores in the order you will run them for compensation purposes. 3. Select the H checkbox in the Inspector window to select height for each parameter. 4. In the Browser, right click Global Sheet1 and select Apply Analysis Template 5. In the Template dialog, select the Area Scaling worksheet and click OK 6. Create a new specimen by clicking the New Specimen button on the Browser toolbar 7. Expand the new specimen. 8. Adjust histograms and statistics according to fluorophores used (Use one fluorophores per laser). 9. Place the Acquisition Pointer on Tube_001.

Adjusting Area Scaling 10. Install the unstained positive sample onto the loading port and click Load in the acquisition dashboard. 11. Adjust the FSC and SSC voltages to place the particles on scale. Note: Keep record of FSC and SSC voltages. 12. Adjust the P1 gate around the population of interest in the FSC vs SSC plot 13. Click the Laser tab in the Cytometer window and adjust the FSC Area Scaling factor until the FSC-A signal matches the FSC-H signal. a. Increase the area scaling factor if the FSC-A signal is lower than the FSC-H b. Decrease the area scaling factor if the FSC-A signal is higher than the FSC-H 14. Unload the stained tube. 15. Load the comp beads tubes and in the parameter window adjust the FSC PMT voltage. Note: Adjust FSC PMT voltage in the parameter window if you are using compensation beads. 16. In the Laser tab adjust area scaling for each laser (being used) with appropriate fluorophores until the A signal matches the H signal. 17. In the Inspector, clear the checkbox for height for all parameters.

Optimizing PMT Voltages 18. Right-click Cytometer Settings > Application Settings > Create Worksheet. A sheet titled Global Sheet2 opens. 19. Load the unstained control sample onto the cytometer. Note: If beads have been used, enter FSC and SSC voltage recorded earlier under the parameters tab 20. Under the Parameters tab enter FSC and SSC voltages recorded earlier. 21. Adjust the FSC threshold. 22. Adjust the appropriate fluorescence voltage to center the negative population in the grey zone and around the + sign. Do not record. Note: However, lowering PMT voltage from the setting established by CS&T might affect the resolution of dim populations. 23. Load the fully stained sample to verify that all fluorescent parameters are on scale. If a positive population is off scale, lower the PMT voltage for that parameter until the positive population can be seen entirely on scale 24. Unload the fully stained sample from the cytometer. 25. Right-click Cytometer Settings > Application Settings > Save. a) Name the experiment including your initials and the date. b) Click Yes to apply the optimized threshold, if applicable. c) The newly saved Application Settings can be found under Cytometer > Catalogs to view or delete when no longer in use. d) The settings are now optimized for this new experiment and are relative to current CST. Note 1: Application settings are relevant until a new baseline is performed. Note 2: Even though the saved experiment template contains Application Settings, these Application Settings are not automatically synchronized with current CS&T every time a new experiment is created from the template Note 3: When you return to perform a similar experiment and you want the setting to be relevant to previous runs. 1. Import the experiment template or create a blank experiment 2. Right-click Cytometer Settings > Application Settings > Apply 3. Select your application setting and Ok -if necessary click on Overwrite and Yes-If the 4. Confirm dialog appears and you have a new set of compensation tubes, click Set all compensation to zero and proceed with step 9 above. Alternatively, you can keep compensation values from a previous run or Duplicate Experiment Without Data, but in the latter two examples the settings won t be relative to the current CS&T.

CALCULATING COMPENSATION 26. Create compensation tubes including a tube for unstained cells/beads under Experiment > Compensation Setup > Create. Add or change labels for tandems. 27. Acquire and record unstained and single stained compensation samples. 28. Check the gates for all tubes, excluding outlier events as needed. 29. Calculate compensation under Experiment > Compensation Setup > Calculate Compensation. Name Compensation Set-Up. Choose Link & Save only if you are running a similar experiment during the same CST period (24hours). Otherwise select Apply Only. 30. Develop the Global Worksheet with plots and stats. 31. Add additional tubes to the experiment and fill in Experiment > Experiment Layout. Change the axis labels in your global experiment. 32. Turn FSC voltage down to the original noted value when switching from compensation beads to cells. ACQUISITION: 1. Experiment > Experiment Layout > Acquisition Tab. Click Global Worksheet column header then select Global Sheet1 on the Global Worksheet pulldown tab. Select the number of events to acquire as well as the stopping gate by selecting the appropriate column header first. 2. Experiment > Experiment Layout > Keyword. You can enter patient ID, Sample ID, Project number, etc... 3. Set up the global worksheet and turn on the biexponential display for the fluorescent log plots 4. Run a fully stained sample and set gates. IF USING SINGLE TUBES: a. Remove the Tube Guide if running single tubes b. Run tubes. c. Record data. IF USING CAROUSEL: a. Carousel > carousel setup. Under the carousel ID tab select the carousel # and under the Export folder click to select the location for the Data file to be exported (FCS files are exported during experiment acquisition. Once recorded, each tube is exported). b. Select appropriate Run setting and Tube pressurization error handling parameters and click ok when finished. c. Make sure the Tube Guide is on when using the carousel d. Click on Run Carousel. Note 1: Export the Experiment Template to the hard drive for future use: File > Export > Experiment Template. Name, etc Note 2: Even though the saved experiment template contains Application Settings, these Application Settings are not automatically synchronized with current CS&T every time a new document is created from the template. When you return to perform a similar experiment and you want the settings to be relevant to previous runs, import the experiment template, right-click Cytometer Settings > Application Settings > Apply. If the Confirm dialog appears and you have a new set of compensation tubes, click Set all compensation to zero and proceed with step 9 above. Alternatively, you can keep compensation values from a previous run or Duplicate Experiment Without Data, but in the latter two examples the settings won t be relative to the current CST

EXPORTING DATA: 1. For single tubes: Export > experiment to your computer folder. For Carousel: Make sure that FCS files have been exported to your designated FCS folder on the computer folder 2. Connect to your online tools (= Cytobank.org with your personnel username and password) to upload the folder on this online server. CLEANING BETWEEN USERS: 1. For single tubes: Clean the fluidics: For Carousel: Clean the fluidics: 2. Wipe down workspace. 3. Check sheath and waste levels. 4. Logout of PPMS DAILY SHUTDOWN: 5. For single tubes: Clean the fluidics: For Carousel: Clean the fluidics: 6. Performs Fluidics Shutdown (Cytometer>Fluidics Shutdown) 7. Wipe down workspace. 8. Check sheath and waste levels. 9. Turn off cytometer. 10. Logout of PPMS