LINAC-based stereotactic radiosurgery for treatment of trigeminal neuralgia



Similar documents
Gamma Knife and Axesse Radiosurgery

Quality Assurance in Stereotactic. Radiotherapy

THE DOSIMETRIC EFFECTS OF

Radiosurgery 8/16/2011. Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.

Our Department: structure and organization

1. Provide clinical training in radiation oncology physics within a structured clinical environment.

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.

STEREOTACTIC RADIOSURGERY COMMISSIONING POLICY

Quality Assurance of Radiotherapy Equipment

External dosimetry Dosimetry in new radiotherapeutic techniques

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Path Finder. CMA Engineering Designing & Consulting Inc. Miami, Florida

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong

02 CyberKnife: Treatment Delivery

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

Baylor Radiosurgery Center

Dosimetric impact of the 160 MLC on head and neck IMRT treatments

Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA USA phone NOMOS

Basic Radiation Therapy Terms

QA of intensity-modulated beams using dynamic MLC log files

IBA Proton Therapy. Biomed days Vincent Bossier. System Architect Protect, Enhance and Save Lives

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y

Gated Radiotherapy for Lung Cancer

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers

First Three Years After Project Proton Therapy Facility:

Prepublication Requirements

TITLE: Comparison of the dosimetric planning of partial breast irradiation with and without the aid of 3D virtual reality simulation (VRS) software.

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

Radiation Oncology Centers Participating in MassHealth. Daniel Tsai, Assistant Secretary and Director of MassHealth

Accreditation Is Coming

Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021

Key words: treatment planning, quality assurance, 3D treatment planning

Trigeminal neuralgia (TN) is a common CYBERKNIFE RADIOSURGERY FOR TRIGEMINAL NEURALGIA TREATMENT: A PRELIMINARY MULTICENTER EXPERIENCE

Volumetric modulated arc therapy (VMAT) for prostate cancer

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

A Thesis. entitled. Retrofitted Micro-MLC SRS System. George H. Hancock

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

FEATURE ARTICLE. One precise dose

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015

At the intersection of innovation and medical insight, patients find the new standard for excellence in neurosurgery.

IMRT for Prostate Cancer. Robert A. Price Jr., Ph.D. Philadelphia, PA

Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

Proton Therapy for Prostate Cancer

SAFE HARBOR STATEMENT

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Clinical Photon Beams

Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT

WELCOME. Massachusetts General Hospital Cancer Center

Department of Radiation Oncology

Gamma Knife Patient Treatment Guide

Evolution of Head and Neck Treatment Using Protons. Mayank Amin, M.Sc,CMD

Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care

A SAFE, NON-INVASIVE TREATMENT OPTION: GAMMA KNIFE PERFEXION

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Stereotactic Radiosurgery & Stereotactic Body Radiation Therapy - Billing Basics. Presented: June 19, 2013 AAMD Annual Meeting San Antonio, TX

What is the CyberKnife System?

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Elekta. Product Overview. Human care makes the future possible

Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer

Design and Scheduling of Proton Therapy Treatment Centers

Protons vs. CyberKnife. Protons vs. CyberKnife. Page 1 UC SF. What are. Alexander R. Gottschalk, M.D., Ph.D.

Proton Therapy. What is proton therapy and how is it used?

Brain Tumor Treatment

Quantitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

Gamma Knife Surgery for Trigeminal Neuralgia: Outcomes and Prognostic Factors. Jason Sheehan, MD, PhD, Hung-Chuan Pan, MD,

Radiation Therapy Coverage, Coding, and Reimbursement for New Technologies

Radiation Protection in Radiotherapy

Grazing incidence wavefront sensing and verification of X-ray optics performance

Challenges in small field MV photon dosimetry

Secondary Neutrons in Proton and Ion Therapy

ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS. Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic

Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD

MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015)

IAEA-TECDOC Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems

MEDICAL DOSIMETRY. COLLEGE OF APPLIED SCIENCES AND ARTS Graduate Faculty: therapy.

Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY

Interactive X-Ray and Proton Therapy Training and Simulation

Clinical Rotation 4: PHYS 707 Spring 2016 (Feb. 25, 2016 to Aug. 25, 2016) COURSE INFORMATION

Performance evaluation of an automated image registration algorithm using an integrated kilovoltage imaging and guidance system

Horizon Blue Cross Blue Shield of New Jersey 2012 Radiation Therapy Payment Rules

SITE IMAGING MANUAL ACRIN 6698

Introduction to Radiation Oncology

Potential Benefits of Proton Therapy in Clinic

Analysis of Trajectory Log Files of TrueBeam Medical Electron Linear Accelerator for Patient Specific IMRT QA

REPORT ORGANIZATION ABSTRACT I. INTRODUCTION CALCULATION OF EFFECTIVE DOSE PER IMAGE ACQUISITION CALCULATION OF EFFECTIVE DOSE PER TREATMENT COURSE

World-first Proton Pencil Beam Scanning System with FDA Clearance

HITACHI Proton Beam Therapy System

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

Oftentimes, as implant surgeons, we are

Jay Hoey. Executive Vice President. Region North America

Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT)

Robot Perception Continued

RADIATION-INDUCED CANCERS FROM MODERN RADIOTHERAPY TECHNIQUES: INTENSITY-MODULATED RADIOTHERAPY VERSUS PROTON THERAPY

The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning.

Transcription:

LINAC-based stereotactic radiosurgery for treatment of trigeminal neuralgia Bruce J. Gerbi, Patrick D. Higgins, 1 Kwan H. Cho, 2 and Walter A. Hall 3 Department of Therapeutic Radiology-Radiation Oncology, 1 University of Minnesota, Mayo Mail Code 494, 420 Delaware St. S.E., Minneapolis, Minnesota 55455 U.S.A. National Cancer Center, 2 Center for Proton Therapy, Madu 1-dong, Ilsan Goyang, Gye onggi, 411-764, Korea Department of Neurosurgery, 3 University of Minnesota, Mayo Mail Code 96, 420 Delaware St. S.E., Minneapolis, Minnesota 55455 U.S.A. gerbi001@umn.edu; higgi010@umn.edu; kwancho@ncc.re.kr; hallx003@umn.edu (Received 18 May 2004; accepted 24 May 2004) Trigeminal neuralgia (TN) is a disabling pain condition that has classically been treated using either surgical or medical techniques. Several researchers have shown that stereotactically delivered radiation can be an effective tool in the amelioration of this condition. For these studies, the Gamma Knife was used to deliver the radiation treatment. The target location was designated as the proximal nerve at the root entry zone, and doses greater than 70 Gy to the maximum point in a single fraction were found to be effective in controlling pain in 80% of the patients treated. LINAC-based stereotactic radiosurgery has been notably absent from the treatment of TN, even though it has many similarities to Gamma Knife-based stereotactic radiosurgery. The aim of this paper is to describe our LINAC-based stereotactic technique for treatment of TN. We also compare treatment of TN using our technique to that using the Gamma Knife. We found that a LINAC-based treatment of TN can be accomplished with accuracy comparable to treatments delivered using the Gamma Knife. The dose distributions are essentially equivalent for the two treatment approaches. The LINAC-based system is easy to plan and offers the ability to reduce the involvement of sensitive structures from the treatment fields as well as the Gamma Knife system does. A disadvantage of the LINAC-based system is the time involved for treatment. PACS number: 87.53.Tf 87.56.Da Key words: trigeminal neuralgia, stereotactic radiosurgery, Gamma Knife, linear accelerator, floor stand, facial pain I. INTRODUCTION Trigeminal neuralgia (TN) is a disabling pain condition that has classically been treated using either surgical or medical techniques. Recently, research in several publications has shown that stereotactic radiosurgery can be an effective tool in the amelioration of this condition. (1 8) In the majority of the studies, the Gamma Knife was used to deliver the radiation treatment. The 4-mm aperture was used for these treatments, and the target 80 2004 Am. Coll. Med. Phys. 80

81 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 81 location was designated as the proximal nerve at the root entry zone. Doses greater than 70 Gy to the maximum point (100%) in a single fraction were found to be effective in controlling pain in 80% of the patients treated. Patients with no prior surgery had complete or near complete pain relief, while complications due to treatment were nearly nonexistent in the published studies. What makes this particular treatment unique from other stereotactic treatments is the extreme accuracy that is required for all steps for successful treatment. This is primarily due to the small size of the target, the difficulty in identifying the target region, the small cone sizes used for treatment, and the extremely high singleshot doses >70 Gy that could result in neural injury if they were misguided (or misdirected). The use of LINAC-based stereotactic radiosurgery for the treatment of TN has been limited in the historical literature, (6 8) even though many of the steps in the process of this stereotactic treatment are similar, if not identical, to those undertaken for a Gamma Knife treatment. Due to the effectiveness of this treatment for pain relief, we initiated a program of LINAC-based stereotactic radiosurgery at our institution. We evaluated our technique to determine if our LINAC-based system had the required accuracy to properly treat such small target volumes. Positioning accuracy and dose distributions are compared against what is expected for Gamma Knife treatments, and application of LINAC-based stereotactic radiosurgery for this clinical problem is discussed. II. METHODS We compared the mechanical accuracy of the two systems, the imaging accuracy of the techniques, and the dosimetric aspects of the treatment including coverage of the target, and the dose to uninvolved normal structures. A. Description of the two techniques Our LINAC-based system uses the Radionics, Inc. MRI-compatible head ring for patient fixation; the larger diameter Brown-Roberts-Wells (BRW) head ring is attached to the MRI head ring for CT scanning and to secure the patient to the treatment stand. Depth helmet measurements are taken after the head ring is fixed initially to the patient and immediately before treatment to ensure that ring placement has remained constant throughout the process. The high-precision beam delivery is provided by a Philips (Elekta) SRS 200 independent support stand and subgantry assembly installed on a 6-MV X-ray Varian 6/100 LINAC. A 5-mm circular cone is used for the treatments, with the LINAC collimators set for a 5 5 cm 2 field. The Philips SRS 200 mechanical system, which is the key source of the mechanical accuracy required by this technique, consists of an independent subgantry assembly, which has a head support system that is independent of the treatment couch, and a secondary gantry, which holds the cones for patient treatment. This independent subgantry is mated to the LINAC through a gimbal assembly. The force to rotate the subgantry assembly is supplied by the LINAC and transmitted to the subgantry via the gimbal assembly. Any small misalignment between the LINAC rotation and the subgantry rotation is handled by the play available in the gimbal assembly. This integrated subgantry is the source of the extreme accuracy required for this particular radiosurgery technique. The independent head support is much more accurate for head positioning than the LINAC couch, while the subgantry is much more accurate than the gantry rotation of a standard LINAC. The capabilities of the system are well documented (9) and easily meet the manufacturer s mechanical accuracy design specifications of 0.3 mm average with no more than 0.5 mm

82 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 82 deviation at any one point using the tests described in the literature. (10,11) It is the extreme accuracy of this system that allows for the treatment of TN using a LINAC; couch-based stereotactic systems are completely dependent on the rotational accuracy of the couch and gantry of the LINAC itself. This Philips SRS 200 system, no longer marketed, is the precursor to the LINAC Scalpel System currently available from Zmed, Inc. (recently acquired by Varian Medical Systems, Inc.). In our procedure, we first obtain MRI scans of the head with the MR head ring and universal localizer (UCLF) on the patient using a 3D data acquisition sequence (MP- RAGE) on a Siemens Magneton scanner. Author Sagittal, axial, and coronal scans of 1 mm thickness at 1 mm spacing are reconstructed from this data acquisition sequence. CT scans are then acquired with the BRW head ring attached to the MR head ring with the CT fiducial localizer in place. CT scans are obtained at 1 mm spacing through the target region with a Siemens Somatom Plus scanner with 2-mm scans taken outside the treatment region. Additional checks are performed to ensure the accurate transfer of the MR and CT scan data from the scanners to the treatment-planning system. The target is determined using the sagittal and axial MR scans. The target volumes are drawn using the MR image sets obtained with the Radionics universal localizer (UCLF) in place during the MR scans and using the Siemens MR head coil. The MR image sets are transformed into the BRW coordinate system using the fiducials on the MR images. The CT data set is required for planning by the Radionics software, and we use the CT image set to delineate the external contour, the eyes, optic nerves, and brain stem. The treatment field arrangement for our LINAC-based treatment contains 13 arcs spaced at 15 intervals for approximately 1300 total degrees of arc rotation (Fig. 1). Treatment planning is done to deliver 7000 cgy to 8750 cgy to the maximum using Xknife-4 software supplied by Radionics. When the patient is ready for treatment and attached to the independent support stand, we ensure that the plan specified treatment coordinates are properly set.

83 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 83 FIG 1. LINAC-based stereotactic radiosurgery arcs for treating trigeminal neuralgia. Thirteen 6-MV X-ray arcs with a total of approximately 1300 arcs degrees of rotation are represented in this plan. The 5-mm cone is used for treatment, and the arc separation is 15º. To evaluate the dosimetric component of our treatment technique, we compared our planning results to a Gamma Knife B unit that we simulated in our Xknife planning computer. The Gamma Knife unit consisted of 201 individual cobalt-60 sources arranged radially in five rings of 35, 39, 39, 44, and 44 sources (Fig. 2). We obtained the collimator location and patterns for an installed Gamma Knife B unit along with tissue-maximum ratio (TMR) and dose profile data for their 4-mm collimators. This information was entered into the Xknife 4 planning system in the same manner as our LINAC data. Once this Gamma Knife beam data was in the program, it could be used directly in the Xknife planning software in the same manner as the LINAC information. Thus a direct comparison of the Gamma Knife system and our LINAC-based system could be conducted on the same patient-related image data sets.

84 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 84 FIG 2. A Gamma Knife B unit simulated using the Radionics Xknife software. The 201 4-mm cobalt-60 beams are simulated using actual off-axis ratio, TMR, and beam orientation information for this unit. The accuracy of the beam modeling for the Gamma Knife data in Xknife was verified before doing the test comparisons of dose distribution for the LINAC and Gamma Knife trigeminal treatments. We first tested the accuracy of the representation of a single 4-mm Gamma Knife Co-60 beam in the Xknife planning system. This Gamma Knife beam was simulated in the Xknife software as a single beam having a one-degree arc. At the isocenter, the profile obtained from this test matched the data of the input dose profile. The treatment plan of the trigeminal nerve using the Gamma Knife data was simulated in Xknife using 201 single arc beams with the same isocenter location (Fig. 2) as our LINACbased plan using 13 arcs. We did not compare the plan using the Xknife representation of a Gamma Knife unit directly to a plan calculated from a Gamma Knife treatment planning computer. Our treatment technique using 13 arcs totaling approximately 1300 arc degrees of rotation was designed to deliver approximately the same amount of dose along the arc path as that given along a Gamma Knife beam. Our rationale for choosing this beam arrangement was based on the assessment that each of the individual 201 fields of the Gamma knife could be closely approximated by a 5 arc, and that our 1300 degrees of rotation would equal about 260 five-degree arc segments. Each Gamma Knife cobalt field would essentially deliver 7000 cgy/201 beams, or approximately 35 cgy to the isocenter per beam. 7000 cgy delivered using 260 five-degree segments would deliver about 27 cgy per 5º arc, close to the 35 cgy per beam for the Gamma Knife unit. Thus the dose through

85 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 85 normal brain tissue for each of these 5 segments would be close to the dose delivered along the track of each Gamma Knife beam. B. Mechanical accuracy of treatment delivery B.1 Isocenter alignment The mechanical isocenter accuracy of the SRS 200 system comes from the design, which minimizes the couch inaccuracy effect by supporting the patient s head independent from the LINAC treatment table. In addition, the subgantry system minimizes the inaccuracies of gantry rotation. (9) Using the standard Lutz alignment test, (10) this system exhibits a mechanical accuracy of <0.3 mm deviation average with a maximum deviation of 0.5 mm. The mechanical isocenter accuracy of the Gamma Knife system is approximately 0.25 mm, as reported in the literature. (12) B.2 Frame system accuracy The application accuracy (13) of the various head frames has been reported in the literature. (13,14) This represents the total clinically relevant uncertainty and includes image reconstruction accuracy, coordinate system transformation, and mechanical errors. The application accuracy of the Leksell frame used with the Gamma Knife was reported to be 1.7 mm ± 1.0 mm (SD), while the BRW frame was 1.9 mm ± 1.0 mm (SD). (13) The general conclusion of these investigations is that there is no major difference in either accuracy or precision between the Leksell or the BRW frame systems when the CT slice thickness is 1 mm. B.3 Imaging accuracy Imaging accuracy depends on the scanning system employed, the coordinate transformation accuracy of the planning system, and the accuracy of the transfer of imaging data from the scanner to the planning computer. Imaging accuracy for our system was determined using the Radionics head test phantom as the standard of comparison. The Radionics test phantom consists of four internal structures a cube, cone, cylinder, and sphere whose location in BRW space is very accurately known (Fig. 3). For tests of both the CT and MR scan set accuracy, the phantom was scanned on the appropriate unit, and these scans were transferred to the planning computer. The internal structures of the phantom were contoured, and the location of the tops of these structures in stereotactic space was determined. These identified locations were then compared against the known coordinates of these structures supplied by the manufacturer. The assessment of accuracy is determined as the difference between the expected location of the tops of these structures and the reconstructed x, y, and z (anterior, lateral, vertical) coordinates of these structures in the transformed computer stereotactic coordinate system.

86 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 86 FIG 3. The Radionics test phantom showing the sphere, cube, cylinder (in center), and cone (to the back of the image). CT resolution for the Gamma Knife system has been reported to be approximately 1.7 mm. (12) Our tests of CT accuracy done for the BRW system using the Radionics test phantom showed a mean difference of 1.1 mm between the expected target coordinates of the test objects and the measured coordinates on images transferred to the Xknife treatment planning computer. Table 1 shows test results of MR accuracy done using 1-mm reconstructions from a 3D volume data acquisition sequence (MP-RAGE) for two Siemens Magneton MR scanners. For this test, the standard Siemens head coil was used with the Radionics test phantom secured in the MR head ring and with the UCLF fiducial localizer in place. We found average uncertainties in position of the four phantom test objects ranging from 0.5 mm to 1.2 mm, which is similar to the results for CT reconstruction. The root-mean-square error (RMS) for the individual scan orientations (either a coronal, sagittal, or axial scan set) ranged from 1.0 mm to 1.4 mm. Although inaccuracies experienced with stereotactic MR scans are well documented in the literature, (15,16) our tests of the MR accuracy of the UCLF localizer have consistently yielded the indicated results. For each patient treatment, the reconstructed locations of the localizer frame are checked, and they routinely match the theoretical location of the localizer rods.

87 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 87 Scan type AP difference Lateral difference Vertical difference Root-meansquare (avg. ± 1 SD) (avg. ± 1 SD) (avg. ± 1 SD) error MR axial 0.29 ± 0.29 mm 0.31 ± 0.27 mm 0.93 ± 0.44 mm 1.02 ± 0.59 mm MR coronal 0.32 ± 0.18 mm 0.20 ± 0.19 mm 1.23 ± 0.50 mm 1.29 ± 0.56 mm MR sagittal 0.33 ± 0.20 mm 0.48 ± 0.36 mm 1.22 ± 0.61 mm 1.35 ± 0.74 mm Table 1. The average difference and the standard deviation (SD) between the expected AP, lateral, and vertical BRW coordinate of the four test phantom objects for MR scans done using the Siemens Magneton MR MP-RAGE acquisition sequence. These data show the accuracy of the image data transferred and analyzed on the Xknife treatment-planning computer. C. Dosimetric comparisons Calculated dose comparisons were done between the 5-mm, 6-MV X-ray LINAC-based beam and the 4-mm Co-60 fields from the Gamma Knife unit. Figure 4(a) shows the offaxis ratio (OAR) data for the two individual beams as a function of the distance from the center of the field. Figure 4(b) is the same data normalized to the 50% beam intensity location. The penumbral widths (80% 20%) were 1 mm for the Co-60 beams and 2.1 mm for the 6-MV field, indicating that the Gamma Knife beam had a sharper falloff.

88 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 88 (a) (b) FIG 4. A comparison of the off-axis ratios for a single 4-mm Gamma Knife beam versus the 6-MV X-ray 5-mm collimator LINAC beam. (a) The off-axis ratios plotted as a function of distance from the center of the collimator; (b) the off-axis ratios normalized to the fraction of the cone radius. A test case on which to make the comparison was a patient treated for TN using the LINAC-based system. Dose distributions were done for the 13-arc LINAC-based treatment approach and compared to the Gamma Knife plan. The locations of the 201 Co-60 sources were from a B model unit, and the beams were represented as 1 arcs in the treatment-

89 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 89 planning system. As stated above, actual OAR and TMR data from an existing clinical Gamma Knife B unit were input into the system to provide a realistic comparison of delivered dose. III. RESULTS AND DISCUSSION Table 2 shows the calculated dose to the various structures for the LINAC-based system and the Gamma Knife system when 70 Gy is delivered to the maximum point for each treatment system. The doses to uninvolved structures are generally higher for the LINACbased system but not significantly so. From an imaging, mechanical accuracy, dosimetric, and clinical standpoint, we are encouraged by the results of the comparison between our LINAC-based stereotactic dose delivery system for treatment of TN. Radiobiological concerns are a factor with our LINAC-based system, since treatment time is typically 2.5 h on a Varian 6/100 running at 200 monitor units (MU) per minute and limited to 2.8 MU/degree. On a more modern LINAC running at 400 MU/min and capable of delivering 10 MU/degree, the treatment time could be reduced by a factor of 3, which would give treatment times more in line with those of a Gamma Knife unit. anatomical structure brain stem, axial MR defined optic chiasm, coronal MR defined left optic nerve, CT defined right optic nerve, CT defined left eye, CT defined right eye, CT defined Gamma Knife Dose summary LINAC Dose summary Gamma Knife B unit, Co-60 gamma rays, 4- LINAC-based treatment regimen: 6-MV mm collimators, 201 individual sources X-rays, 5-mm cone, 13 arcs, ~1300 total arc degrees, 15 spacing between arcs Avg. Min/Max Avg. Min/Max (cgy) (cgy) (cgy) (cgy) 92 13/2746 119 10/3144 58 24/84 76 38/114 16 12/37 10 10/10 14 12/28 10 9/13 13 10/27 9 10/10 13 9/25 9 8/9 Table 2. The dose delivered to various structures for the Gamma Knife and LINAC-based system when 7000 cgy is delivered to the maximum point IV. CONCLUSION By comparison of imaging accuracy, mechanical accuracy, dosimetric comparisons, and clinical results, we feel that LINAC-based stereotactic radiosurgery using this subgantry assembly is an effective means of treating TN and can be comparable to Gamma Knife treatments.

90 Gerbi et al.: LINAC-based SRS for trigeminal neuralgia... 90 REFERENCES 1. Kondziolka D, Perez B, Flickinger JC, Habeck M, Lunsford LD. Gamma knife radiosurgery for trigeminal neuralgia. Results and expectations. Arch Neurol 1998;55:1524 1529. 2. Kondziolka D, Lunsford LD, Habeck M, Flickinger JC. Gamma knife radiosurgery for trigeminal neuralgia. Neurosurg Clin N Am 1997;8:79 85. 3. Young RF, Vermeulen SS, Grimm P, Blasko J, Posewitz AA. Gamma Knife radiosurgery for treatment of trigeminal neuralgia: Idiopathic and tumor related. Neurology 1997;48:608 614. 4. Kondziolka D, Lunsford LD, Flickinger JC, Young RF, Vermeulen S, Duma CM. Stereotactic radiosurgery for trigeminal neuralgia: A multi-institutional study using the gamma unit. J Neurosurg 1996;84:940 945. 5. Kondziolka D, Flickinger JC, Lunsford L, Habeck M. Trigeminal neuralgia radiosurgery: The University of Pittsburgh experience. Stereo Func Neurosurg 1996;66(1):343 348. 6. De Salles AAF, Buxton W, Solberg T, et al. Linear accelerator radiosurgery for trigeminal neuralgia. In: Kondziolka D, ed. Radiosurgery, Basel Karger;1998:173 182. 7. Goss BW, Frighetto L, DeSalles AAF, Smith Z, Solberg T, Selch M. Linear accelerator radiosurgery using 90 Gray for essential trigeminal neuralgia: Results and dose volume histogram analysis. Neurosurgery 2003;53(4):823 828. 8. Smith Z, DeSalles AAF, Frighetto L, et al. Dedicated linear accelerator radiosurgery for the treatment of trigeminal neuralgia. J Neurosurg 2003;99:511 516. 9. Friedman WF, Bova FJ. The University of Florida radiosurgery system. Surg Neurol 1989;32:334 342. 10. Lutz W, Winston KR, Maleki N. A system for stereotactic radiosurgery with a linear accelerator. Int J Radiat Oncol Biol Phys 1988;14:373. 11. Winston KR, Lutz W. Linear accelerator as a neurosurgical tool for stereotactic radiosurgery. Neurosurgery 1988;22:454 464. 12. Wu A. Physics and dosimetry of the gamma knife. Neurosurg Clin N Am 1992;3:35 50. 13. Maciunas RJ, Galloway RL, Latimer JW. The application accuracy of stereotactic frames. Neurosurgery 1994;35:682 694. 14. Galloway RL, Maciunas RJ, Latimer JW. The accuracies of four stereotactic frame systems: An independent assessment. Biomed Instrument Tech 1991;25:457 460. 15. Sumanaweera TS, Adler JR, Napel S, Glover GH. Characterization of spatial distortion in magnetic resonance imaging and its implications for stereotactic surgery. Neurosurgery 1994;35:696 703. 16. Sumanaweera T, Glover G, Song S, Adler J, Napel S. Quantifying MRI geometric distortion in tissue. Magn Reson Med 1994;31:40 47.