User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending



Similar documents
ExStik TM ph Waterproof Meters

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

User's Guide. Digital Viscotesters. Model High Range (0.3 to 4000 dpas) Model Low Range (1.5 to 330 mpas) Introduction

Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented

User's Manual. Model Laser Photo / Contact Tachometer. Introduction

User Manual. Humidity-Temperature Chart Recorder. Model RH520

Dual Laser InfraRed (IR) Thermometer with Color Alert

Dual Laser InfraRed (IR) Thermometer

Dual Laser InfraRed (IR) Thermometer

EzFlex Combustible Gas Leak Detector Model EZ40

Ground Resistance Clamp-On Tester

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

User's Guide. Integrating Sound Level Datalogger. Model Introduction

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

High Temperature InfraRed Thermometer with Laser Pointer

User's Guide Integrating Sound Level Datalogger Model

Care and use guide. Contents. a handy solution made easy. page

Milwaukee USER MANUAL

Instruction Manual HI Portable Waterproof Microprocessor Dissolved Oxygen Meter

Global Water Instrumentation, Inc.

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

Instruction Manual Temp 4/5/6 Temperature Meter Series

UltraBasic Benchtop ph Meter Operation Manual

INSTRUCTION MANUAL DTH Fax: (503)

Care and use guide. Contents. page

Care and use guide. Contents. a handy solution made easy. page

Ground Resistance Clamp On Tester

ph 200 ph-/redox-/temperature Meter Operating instructions

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL

Hygro-Thermometer + InfraRed Thermometer Model RH101

RI-215A Operator s Manual. Part Number: RK Revision 0 Released: 10/3/05

User's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN

User Guide. Temperature and Humidity Datalogger. Model 42280

DO-6. Dissolved Oxygen Meter. Newport Electronics, Inc. Operator s Manual

el-720 Carbon Dioxide Gas Leak Detector User s Guide

4-Channel Thermometer / Datalogger

Microwave Meter. Instruction Manual

User s Manual. Bluetooth Calculator Keypad. Page

ph Measurement and Control

Electrical Stability Tester, Model 23E Instruction Manual Manual No , Revision B Instrument No

User's Guide. Pinless Moisture Meter. Model MO257

Temperature Humidity Meter

SenTix 950 SenTix 980

AutoRanging Digital MultiMeter

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:

SINGLE RANGE DS METERS Operation Instructions

Soil Sensor Reader PRODUCT MANUAL

Light Meter (Foot Candles)

DICKSON PW4 DICKSON. Pressure Chart Recorders. Applications & Specifications. Product. Getting Started, Charts & Accessories.

Temperature, Humidity & Dew Point Chart Recorder

Pro1030 USER MANUAL. English

AIRMoN, ph and Conductivity Field Measurement

User s Guide HHAQ IN-1 Psychrometer With Enthalpy. Shop online at omega.com SM

Periodic Verification System Model 775PVS

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

USEPA Electrode Method Method 8156

DISSOLVED OXYGEN METER Model : DO-5509

Alarm Clock USER GUIDE

Professional Thermohygrometer

DICKSON PR100/PR300/PR500 DICKSON. Pressure Logger Operation. Contents:

EE7016 EchoStream Survey Kit

Portable Waterproof ph/dissolved Oxygen Meter

DSM155 Digital LED Light Meter User s Manual DSM155

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen

User's Guide. Digital Sound Level Meter. Model

Tympanic Thermometer

Bluelab Soil ph Meter Instruction Manual

Intelli-Time Alarm Clock model 13027

Drop Counter Sensor Product Number: ENDRP-AD100

MP-4V Heavy Duty Riveter / 39048

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

CABLE MODEM QUICK START

MODEL G300 BRAKE BLEEDER

Wolverine Data PASS Portable Scanner User Manual V1.0

INSTALLER S & OWNER S MANUAL

User's Guide. True RMS Industrial Multimeter

Part I - Installation

CM100. Carbon Monoxide Meter. Users Manual Mode d emploi Bedienungshandbuch Manual d Uso Manual de uso Användarhandbok

Mini Digital Altimeter & Climb rate & Barometer & Thermometer & Compass & Weather forecast & Time

Automatic Phone-Out Home Monitoring Systems

Thermistor Thermometer MODEL NO

HVAC Clamp Meter. Users Manual

Combustible Gas Leak Detector 720b

Refer to page 6 to page 19 for detailed instructions.

Water Leak Detection System

Operating manual. ph 7310 ph 7310P. Printermodul. ph 7310P. ph ph meter

USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model #

Appendix C. Vernier Tutorial

Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan

INSTALLATION AND OPERATING INSTRUCTIONS

Atomic Alarm Clock model 13131W2

Smart-UPS RT External Battery Pack Stack/Rack-Mount 6U

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

In-Situ Inc. Warranty / Terms & Conditions / Software License. This warranty policy applies to items that shipped prior to January 1, 2012.

About Your NetBotz 4-20mA Sensor Pod

ADVANCED INTERNET DOWNLOADABLE POSTAL & SHIPPING SCALE

2087DF-BD / 2092DF-BD

Transcription:

User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending

Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference junction Before first use or after storage, soak the electrode (with its cap removed) in a ph 4 solution for about 10 minutes (PH100 only) White KCL crystals may be present in the cap. These crystals will dissolve in the soak or they can be simply rinsed with tap water Always calibrate close to the expected measurement value Twist off the battery compartment cap to access the battery tab; remove this plastic tab before first use To preserve Electrode life keep the sponge in the protective cap soaked with a ph 4 solution (PH100), deionized, distilled water (RE300) when not in use (store vertically) Specifications Display Multifunction LCD with Bargraph Operating conditions 32 to 122 o F (0 to 50 o C) / < 80% RH ph Range and Accuracy 0.00 to 14.99 / ± 0.01pH typical Temp. Compensation Automatic from 32 to 194 o F (0 to 90 o C) Compensation Accuracy ± 0.9 o F (0.5 o C) ORP Range / Accuracy ± 999mV / ± 4mV Temperature Range 23 to 194 o F (-5 to 90 o C) Temperature Resolution 0.1 o up to 99.9 then 1 o thereafter Temperature Accuracy ± 1.8 o F / 1 o C [from 23 to 122 o F (-5 to 50 o C)] ± 5.4 o F / 3 o C [from 122 to 194 o F (50 to 90 o C)] Measurement storage 15 tagged (numbered) readings Power Four (4) SR-44 button batteries (See note on page 6) Low battery indication BAT appears on the LCD Auto power off After 10 minutes of inactivity Battery Life 200 to 400 tests, assuming test time of < 10 minutes and use of SR44 silver oxide batteries. Use of LR44 alkaline will reduce battery life. Page 2

ExStik TM Description 1. Battery compartment cap 2. LCD Display 3. MODE button 4. CAL button 5. ON/OFF button 6. Electrode collar 7. Electrode (refillable on PH110) (Electrode cap is not shown) 1 2 3 4 5 6 7 ExStik TM Display 1. Bargraph reading 2. Measurement reading 3. BAT (low battery) and HOLD (data hold) indicators 4. Temperature display 5. Bargraph scale designations 6. Units of measure 7. Calibration indicators 8. RENEW and CAL indicators 5 6 7 8 1 2 3 4 ph Overview ph is a unit of measure (ranging from 0 to 14pH) indicating the degree of acidity or alkalinity of a solution. ph tests are the most commonly performed measurements in water analysis, using the hydrogen ion activity of a solution to determine acidity or alkalinity. Solutions with a ph less than 7 are considered acidic, solutions with a ph higher than 7 are known as bases, and solutions with a ph of exactly 7 are neutral. The ph scale is logarithmic so, for example, if sample A is 1 ph less than Sample B, this means that Sample A is 10 times more acidic then Sample B. A difference of 1 ph represents a ten-fold difference in acidity. Page 3

ORP/REDOX Overview ORP stands for Oxidation / Reduction Potential and represents the oxidizing or reducing nature of a solution. The overall reactive tendency of a solution can be determined from ORP measurements. ORP testing is becoming increasingly more common in waste-water measurement regimens. The unit of measurement for ORP is mv. Replacing Electrodes The ExStik TM is shipped with an electrode attached. If the electrode is to be replaced, follow these steps for removing and connecting electrodes. Note that the PH110 has a refillable electrode and the PH100 does not. 1. To remove an electrode, unscrew and completely remove the electrode collar (turn the collar counter-clockwise to remove). 2. Gently rock the electrode from side to side, pulling it downwards, until it disconnects from the meter. 3. To attach an electrode, carefully plug the electrode into the meter socket (note that the electrode connector is keyed, ensuring proper connection). 4. Tighten the electrode collar firmly enough to make a good seal (a rubber gasket seals the electrode with the meter). Automatic Electrode Recognition When the ExStik TM is turned on, it recognizes the type of electrode that is connected and displays the appropriate unit of measure. Attach electrode before turning the ExStik TM on. Powering the ExStik TM The ExStik TM uses four (4) SR-44 batteries (included). The insulating tab located in the battery compartment must be removed before first use. If the batteries are weak, the BAT indicator appears on the LCD. Press the ON/OFF key to turn the ExStik TM on or off. The auto power off feature shuts the ExStik TM off automatically after 10 minutes of inactivity to preserve battery life. ph Operation When the electrode is placed in a solution, the main display and bargraph indicate the ph reading while the lower display reads temperature (readings flash until they have stabilized). The bargraph is center zero, i.e. at ph 7 there is no display. As the ph rises, the bar moves from the center to the right. If the ph drops, the bar moves from the center to the left. ph Calibration (1, 2, or 3 points) 1. Place the electrode into a buffer solution (4, 7, or 10) and press the CAL key. Typically, ph 7 is calibrated first then 4 or 10 ph. 2. The ExStik TM automatically recognizes the solution and calibrates itself to that value (the circled number on the LCD will match the solution). Note that if the solution is more than 1pH off from the 4, 7, or 10pH standard, the ExStik TM will assume an error and abort the calibration (CAL and END will be displayed). 3. During calibration, the ph reading flashes on the main display. 4. When calibration is complete, the ExStik TM automatically displays END and returns to normal operation mode. 5. The appropriate circled indicator (4, 7, or 10) appears on the LCD when a particular calibration has been completed. When the ExStik TM is turned off, the circled indicator configuration and the calibration data will be memorized. 6. For a two or three point calibration, repeat steps 1-5. Page 4

Changing the Displayed Temperature Units Press and hold the CAL button for approx. 3 seconds. The o C or o F icon will change first and then the numerical temperature value will change after the button is released. If the Calibration mode is accidentally accessed ( CAL appears on the LCD) simply turn the ExStik TM off and start again. ORP Operation When an ORP electrode is connected and the ExStik TM is turned on, it enters the Automatic Calibration mode. SELF and CAL will appear on the LCD while calibration is in progress. After the calibration is complete, SELF and CAL will extinguish and both the main display and bargraph will read ORP in mv units. The bargraph reads 0mV (far left), 500mV (center), and 1000mV (far right). Data Hold Press the MODE button to freeze the current reading. The HOLD display icon will appear along with the held reading. The held reading will be stored in the next 15-reading storage location (see below). Press the MODE key to return to normal operation. 15-Reading Measurement Storage 1. Press the MODE button to store the current reading (Data Hold will activate). Press MODE again to return to normal operation. Repeat step 1 above to store the next reading and so on. 2. If an attempt is made to store more than 15 readings, the stored readings (starting with the first reading) will be overwritten. Recalling Stored Readings Note: First ensure that the HOLD symbol is not displayed. If it is, exit the HOLD function by pressing the MODE button. 1. Press the CAL button once and then press the MODE button immediately after CAL is displayed; the location number (1 through 15) will flash. If the CAL mode is accidentally accessed (display flashing), press the CAL button again to exit. 2. The last stored reading taken will be displayed first. To advance through the stored readings, press the MODE button. The location number is displayed first, followed by the reading stored in that location. 3. To exit the storage mode, press the CAL button and the ExStik TM will return to normal operation. CAL Reminder Display When the ExStik TM is turned on (in ph models) for the 15th time, the CAL icon appears on the LCD indicating that the ExStik TM may require calibration. Some applications may require recalibration of the electrode sooner than others. The CAL display is simply a reminder and will turn off when the ph electrode is recalibrated. RENEW Display A flashing RENEW warning indicates that the probe may be nearing the end of its useful life. If cleaning or recalibration does not cause the RENEW icon to disappear, replace the probe (see optional accessories below). The RENEW display appears as a result of the ph electrode slope falling below 75% of a nominal slope. Page 5

Battery Replacement 1. Twist off the battery compartment cap 2. Replace the four (4) SR-44 button batteries observing polarity Rear Top View of ExStik TM Front (LCD Side) 3. Replace the battery compartment cap Battery Note: For maximum battery life, use battery style SR-44 (Silver Oxide) such Energizer TM No. 356 or Duracell TM No. 303/357. Electrode Refilling The electrode does not need to be detached from the body of the ExStik TM in order to refill it as long as the electrode is of the refillable type. Refillable electrodes have a removable reference junction (slotted) and the word REFILLABLE on the side of the electrode casing. Instructions for removing the reference junction and filling the electrode are provided with the refill kit. See Optional Accessories below for refill kits and refillable electrodes. Optional Accessories Tripak buffers with 4, 7 & 10 ph capsules (6 each) plus two rinsing solutions (Part Number: PH103) Spare ph Electrode non-refillable (Part Number: PH105) Spare ph Electrode - refillable (Part number PH115) Electrode Refill solution (Part number PH113) Spare ORP electrode (Part Number: RE305) Spare Chlorine electrode (Part Number: CL205) Weighted base with 5 solution cups (Part Number: EX006) Note: If the unit is to be converted for Chlorine use by attaching the chlorine electrode, please request a CL200 User Guide Page 6

Measurement and Display Considerations If the unit appears to be locked (display frozen). It is possible that the Data Hold mode has been inadvertently accessed by pressing the MODE button. Simply press the MODE button again or turn the meter off and restart if the display appears frozen. If the meter does latch up and no button presses revive it, remove the batteries and restart. Note that if the batteries are removed, any stored readings will be discarded. Also, the calibration for ph will be cleared requiring recalibration. Calibration data for ORP or Chlorine will be retained however. Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies on sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our website at www.extech.com (click on Contact Extech and go to Service Department to request an RA number). A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied. Calibration and Repair Services Extech offers complete repair and calibration services for all of the products we sell. For periodic calibration, NIST certification or repair of any Extech product, call customer service for details on services available. Extech recommends that calibration be performed on an annual basis to ensure calibration integrity. Support Hotline (781) 890-7440 Tech support: Ext. 200; Email: support@extech.com Repair/Returns: Ext. 210; Email: repair@extech.com Website: www.extech.com Copyright 2003 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form. Page 7

Notes Page 8