Creating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II



Similar documents
Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation

Budgeting in QuickBooks

Lab: Data Backup and Recovery in Windows XP

Lab - Data Backup and Recovery in Windows XP

Steps to Create a Database

Overview... 2 How to Add New Documents... 3 Adding a Note / SMS or Phone Message... 3 Adding a New Letter How to Create Letter Templates...

ICP Data Entry Module Training document. HHC Data Entry Module Training Document

Appendix A How to create a data-sharing lab

Creating a Newsletter with Microsoft Word

Search help. More on Office.com: images templates. Here are some basic tasks that you can do in Microsoft Excel 2010.

Fluorescent dyes for use with the

[COGNOS DATA TRAINING FAQS] This is a list of frequently asked questions for a Cognos user

Using an Automatic Back Up for Outlook 2003 and Outlook 2007 Personal Folders

To reuse a template that you ve recently used, click Recent Templates, click the template that you want, and then click Create.

Setting Up Power POS in Windows 95/98

DIRECTIONS FOR SETTING UP LABELS FOR MARCO S INSERT STOCK IN WORD PERFECT, MS WORD AND ACCESS

To export data formatted for Avery labels -

Shimadzu UV-VIS User s Guide

Excel 2010: Create your first spreadsheet

Mail Merge Creating Mailing Labels 3/23/2011

FRONTPAGE FORMS

emarketing Manual- Creating a New

Personal Portfolios on Blackboard

ING USING HOTMAIL

BD FACSComp Software Tutorial

To change title of module, click on settings

FrontPage 2003: Forms

RealStar HBV PCR Kit /2012

Implementing Mission Control in Microsoft Outlook 2010

How do I create a Peachtree (Sage 50) Payroll export file?

2. PMP New Computer Installation. & Networking Instructions

Remote Viewer Recording Backup

Quick Start Guide. Highly customizable automated trading Automate your trades according to rules and models you create.

MY WORLD GIS. Installation Instructions

Installing a printer in Windows 95.

How To Use Ticket Validation Software On A Pc Or Mac Or Macbook Or Ipad (For Acedo) On A Computer Or Ipa (For An Ipa) On An Ipad Or Macintosh (For Macintosh) On Pc

Snap 9 Professional s Scanning Module

INTERCALL ONLINE Administrator Invoices User Guide

The tool also provides an overview of date criteria set against all items within your module.

Don't have Outlook? Download and configure the Microsoft Office Suite (which includes Outlook)!

Microsoft Outlook 2003 Quick How-to Guide

Quick Start Using DASYLab with your Measurement Computing USB device

OUTLOOK GETTING STARTED

Managing Contacts in Outlook

Enhanced Formatting and Document Management. Word Unit 3 Module 3. Diocese of St. Petersburg Office of Training Training@dosp.

For example, within General Settings, you can change the default language from English to Spanish and change the size of the text that is displayed.

QIAGEN Rotor Gene Q: Software V2.0

Company Setup 401k Tab

The How Do Guide to AxioVision. Author: Brian Svedberg Image Analysis Specialist Carl Zeiss MicroImaging, Inc.

Creating a Gradebook in Excel

Outlook 2013 Delegate Access Managing Someone Else's Calendar and Inbox

Filtering with Microsoft Outlook

Backing Up and Restoring FrontDesk Data Files

Microsoft Access Rollup Procedure for Microsoft Office Click on Blank Database and name it something appropriate.

Write & Print Checks

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab

How to make a backup copy of a.pst file

Creating a New Search

Call Recorder Quick CD Access System

Microsoft Access 2010 handout

Using Outlook Web Access

Discoverer Training Guide

This guide provides step by step instructions for using the IMF elibrary Data - My Data area. In this guide, you ll learn how to:

Plexor Systems Instrument Setup and Data Analysis for the Roche LightCycler 2.0 and LightCycler 1.5 Systems using LightCycler Software Version 4.

Creating Your Personal Website

WHAT S NEW IN MS EXCEL 2013

HHC Compensation Module Training Document

Setting Preferences in QuickBooks

Basics. a. Click the arrow to the right of the Options button, and then click Bcc.

ICE.TCP Pro Update Installation Notes

Junk Settings. Options

Creating Pivot Tables

Recruiter s Resource Guide

Computer Training Centre University College Cork. Excel 2013 Pivot Tables

ORACLE BUSINESS INTELLIGENCE WORKSHOP

Instructions for Using Excel as a Grade Book

DEPLOYING A VISUAL BASIC.NET APPLICATION

A. I do not have my own personal certificate I am a new client or want to download a new certificate

Organizing and Managing

REMOTE DESKTOP IN WINDOWS XP

Excel 2013 What s New. Introduction. Modified Backstage View. Viewing the Backstage. Process Summary Introduction. Modified Backstage View

How to Install Microsoft Windows Server 2008 R2 in VMware ESXi

MICROSOFT OUTLOOK 2011 READ, SEARCH AND PRINT S

Microinvest Warehouse Pro Light Restaurant is designed to work in tandem with Microinvest Warehouse Pro which provides all back office functions.

Introduction to Microsoft Access 2007

CREATING EXCEL PIVOT TABLES AND PIVOT CHARTS FOR LIBRARY QUESTIONNAIRE RESULTS

USB Flash Drive. The New Age of Data Storage Future. 1990s

Using JCPS Online for Websites

ANYWHERE POLLING - POLLING WITH A QUESTION LIST

Check out our website!

All V7 registers support barcode printing, except the Sharp 410/420 1A ROM and that limitation is based upon the register.

Navios Quick Reference

Scientific Graphing in Excel 2010

Microsoft Access 2000

WINDOWS POINT OF SALE (WinPOS) Property Tax Entry Module

SENDING S & MESSAGES TO GROUPS

One Drive for Business User Guide

You must have at least Editor access to your own mail database to run archiving.

Outlook Mail, Calendar, Contacts, Notes & Tasks. User Guide

Word 2010: Mail Merge to with Attachments

Transcription:

Creating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II Due to ROX and Cy5 dye emmision spectra overlap Roche LightCycler 480 can pick up signals from a dye measured by another channel. This phennomen is called crosstalk and can lead to misleading results. To correct for crosstalk Roche recommends to create a so-called Color Compensation file. This permits the LightCycler 480 software to compensate for fluorescence crosstalk. For further details, please refer to the Roche LightCycler 480 Instrument Operator Manual. Note: The detection format for the Color Compensation must be identical to the format used when performing a HemaVision 7Q/28Q qpcr. Also a Color Compensation file can only be applied to experiments that were run on the same LightCycler 480 instrument. Note: The settings given below are recommendations only. Content: The HemaVision 7Q/28Q Color Compensation kit for Roche LightCycler 480 includes four 8-tube strips with dye. Each strip contains 5 replicates of the same dye, either FAM, ROX or Cy5 dye or no dye in a volume of 20 L HemaVision 7Q/28Q qpcr mix. The dye content of the strip is marked with a white label on the caps of the 3 empty tubes in each 8-tube strip. Creating a Color Compensation file: 1. Turn on the LightCycler480 and let it warm-up for at least 10 min. 2. Prior to use, spin the 8-tube strips quickly to collect the dye solution at the bottom of the tubes. 3. In the software window, select the Tools Icon (the wrench symbol). #8 #4 #6 #5 #4 #7 4. Select Detection Format, and then choose New. 1

5. Select the excitation/emission filter combinations 483/533 nm for FAM, 558/610 nm for ROX and 618/660 nm for Cy5 as shown above. 6. Type in a name that describes the calibration series, for instance HemaVision 7Q. 7. Click Close. 8. From the Software Overview window, choose New experiement. 9. Under Setup, Detection Format choose the name of the filter combination you created in step 5-6. 10. Click the Customize tab to ensure the right filter combinations have been choosen. 11. Set up a Melt Curve program as shown below: #9 #10 #11 #11 #11 #11 #11 12. Introduce a Cool Down step by pressing +, write Cool Down and set Target to 40 for 10 seconds. 2

#12 #12 #12 13. In the Subset Editor on the left hand column, choose a new subset by pressing the + tab in the lower left corner. Name the subset Color Compensation and mark 4 rows with each 5 wells. Hold down the control button on your keyboard to select multiple rows. Once done click Apply. 14. Click the Sample Editor in the left hand column. 15. Under Step 1: Select Workflow click Color Comp 16. Under Select Filter Combinations check that the correct filter combinations have been chosen. 17. Under Step 2: Select Samples select the sample subset created in step 12. 3

18. Under Step 3: Edit Color Comp Properties, use the drop down menu to select the dominant channel for each of the 5 dye strips. In the strip without dye, choose water as dominant channel (see figure below). #15 #16 #14 #17 #18 #19 #18 19. In the column to the right, click the floppy disk icon to save the experiment. In the folder Experiments save the experiment under an appropriate name. 20. Place the calibration dye strips in the adapter plate. Make sure that the orientation is the same as you previously have indicated in the Sample Editor. The 5 dye containing tubes of each strip are not covered by the white label. Well A1 must be located in the top left corner and the cut corner in the lower right of the plate. 21. Insert the adapter plate and calibration dyes in the machine. 22. In the Experiment window press Start. 23. When the experiment has finished click Analysis in the left hand column. 4

#24 #24 #23 #24 24. In Create New Analysis click Color Compensation and click the OK symbol. 25. Click Calculate to perform the analysis. #25 #26 26. Click Save CC Object to create a Color Compensation file. By default the file is saved under Special Data and CCC. 27. You can now apply the Color Compensation data to new experiments or re-analyse previous experiments. If the lamp of the Roche LightCycler 480 apparatus has been changed a new Color Compensation file is required. The Color Compensation kit can be re-used a limited number of times. Store the kit dark at -20 0 C. 5

Applying Color Compensation 1. Select the filter combination you want to display and compensate by using the Filter Comb button. #3 #2 #1 2. Click the arrow-down button Color Comp and select In Use or In Database to find the Color Compensation file you have created and click OK. 3. The amplification curves are now changed using the compensated data. Note that the Color Comp button now is labeled (On). In case of insufficient Color Compensation leading to either overcompensation or remaining crosstalk, please refer refer to the LightCycler 480 Instrument Operator Manual. 6