Determination of the Effective Energy in X-rays Standard Beams, Mammography Level



Similar documents
DEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR

PERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS

IMPROVEMENT OF GAMMA CALIBRATION PROCEDURES WITH COMMERCIAL MANAGEMENT SOFTWARE

DOSIMETRIC CHARACTERIZATION OF DYED PMMA SOLID DOSIMETERS FOR GAMMA RADIATION

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY

DEVELOPMENT OF A QUALITY MANAGEMENT SYSTEM FOR BRAZILIAN NUCLEAR INSTALLATIONS

IMPLEMENTATION OF A MANAGEMENT SYSTEM FOR OPERATING ORGANIZATIONS OF RESEARCH REACTORS

PROCESSING OF AISI M2 HSS WITH ADDITION OF NbC BY MECHANICAL ALLOYING USING TWO DIFFERENT TYPES OF ATTRITOR MILLS

APMP/TCRI Key Comparison Report of Measurement of Air Kerma for Medium-energy X-Rays (APMP.RI(I)-K3)

STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE

Ra, 232 Th and 40 K analysis in sand samples from some beaches of Great Vitória, Espírito Santo, Brazil: Preliminary results.

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

TLD ALBEDO DOSIMETER PERFORMANCE ON A Hp(10) NEUTRON DOSE IAEA INTERCOMPARISON

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

José Egidio M.Tondin 1, Mario O. de Menezes 1, Marina B.A.Vasconcellos 1 and João A.Osso Jr. 1 ABSTRACT 1. INTRODUCTION

Public Funding and the Beginning of a New Era in Higher Education in Brazil

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Research Topics in Renewable Energies in Brazil

A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS

EFFECT OF MICROSTRUCTURE ON PROPERTIES AND PERFORMANCE OF SPRAY FORMED AISI M2 HIGH-SPEED STEEL

STUDY OF MAGNETISM IN GDTIN (T = NI, PD, CU) COMPOUNDS BY PERTURBED GAMMA-GAMMA ANGULAR CORRELATION SPECTROSCOPY

Test Section for Experimental Simulation of Loss of Coolant Accident in an Instrumented Fuel Assembly Irradiated in the IEA-R1 Reactor

Improvement and storage of conventional mammography images using techniques of digital image processing

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

arxiv: v1 [stat.ml] 21 Jan 2013

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Cu Nanoparticles Enables Plasmonic-Improved Silicon Photovoltaic Devices

Production of X-rays and Interactions of X-rays with Matter

Crystalline perfection of an aluminium single crystal determined by neutron diffraction

THE EFFECT OF HIGH ENERGY MILLING ON THE MICROSTRUCTURE AND MECHANICAL PROPERTIES OF A Ti-13Nb-13Zr ALLOY PRODUCED BY POWDER METALLURGY

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

National Laboratory for Metrology of Ionizing Radiation (LNMRI) 1

Atomic and Nuclear Physics Laboratory (Physics 4780)

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Mconf: Global Webconferencing based on Open Source

THE EFFECT OF THE HDDR PROCESS ON THE PRODUCTION OF Pr-Fe-Co-B-Nb SINTERED MAGNETS

Lasers and Specturbility of Light Scattering Liquids - A Review

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Development of a parallel plate ion chamber for radiation protection level

7üVNORD. TÜV NORD Röntgentechnik

Study of the Natural Circulation Phenomenon for Nuclear Reactors

Simone Alberigi and Brigitte Roxana Soreanu Pecequilo

NUCLEAR ENERGY EDUCATION SCENARIO AROUND THE WORLD

Manual for simulation of EB processing. Software ModeRTL

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

AML/ /06/18414/II-0545-FA-FCD-FI SUPPORT (Sustainable Use of Photosynthesis Products & Optimum Resource Transformation)

"I" The Technology Park of Sio Paulo a project to stimulate the local knowledge-based innovation system Track

Introduction to Error Analysis

The current situation of sanitary landfills in Brazil and the application of economic models

International Scientific Cooperation in Neglected Tropical Diseases: Portuguese Participation in EDCTP-2

One approach to the use of the practices of CMMI- DEV V1.3 level 2 in a process of development of Embedded Systems

How To Test Water Penetration On Concrete Block And Mortar

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Voltage nominal maximum for test. Inverse voltage. nominal maximum for test. Focal spots (IEC 60336:2005) small large

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

PLANMED SOPHIE MAMMOGRAPHY X-RAY UNIT TECHNICAL MANUAL /11 LAT LAT SIN DEX. kv kv. mas. mas CTL

COMPARATION BETWEEN PIII SUPERFICIAL TREATMENT AND CERAMIC COATING IN CREEP TEST OF Ti-6Al-4V ALLOY

CFD SIMULATION OF IPR-R1 TRIGA SUBCHANNELS FLUID FLOW

Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY

VoxelDose: A SOFTWARE FOR DOSIMETRIC EVALUATIONS IN X-RAY DIAGNOSES

A BLOCK PROGRAMMING INTERFACE FOR EDUCATIONAL MOBILE ROBOTS

Coating Thickness and Composition Analysis by Micro-EDXRF

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

ASPECT OF SUCROSE AND ITS MONOMERS FROM SUGARCANE JUICE SUBMITTED TO DIFFERENT DOSES OF COBALT-60 IRADDIATION

Diagnostic Exposure Tracking in the Medical Record

EQUIPMENT REQUIREMENTS AND QUALITY CONTROL FOR MAMMOGRAPHY

On the re-calibration of Ni-foil reference standards

Barracuda & QABrowser

WebBee a Web-based Information Network on Bees

HYDROGEN STORAGE AND MICROSTRUCTURE INVESTIGATIONS OF La 0.7- Av. Prof. Lineu Prestes, 2242, ZIP , São Paulo, Brazil.

Physics testing of image detectors

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards

ACCELERATORS AND MEDICAL PHYSICS 2

Effect of Gamma Ray Energies and Addition of Iron Slag by weight to Portland Cements on Mass Attenuation Coefficient

2 Absorbing Solar Energy

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

EVALUATION OF ANALYSIS METHOD STANDARDLESS BY WDXRF AND EDXRF OF ALUMINUM POWDER USED IN MTR TYPE FUEL

portable x-ray survey report

Transcription:

Determination of the Effective Energy in X-rays Standard Beams, Mammography Level Eduardo de Lima Corrêa 1, Vitor Vivolo 1, Maria da Penha A. Potiens 1 1 Instituto de Pesquisas Energéticas e Nucleares Av. Prof. Lineu Prestes, 2242, Cidade Universitária, 05508-000, Brazil educorrea1905@gmail.com, vivolo@ipen.br, mppalbu@ipen.br Abstract. The X-rays beams used in diagnostic radiology are heterogeneous. This means that, in a radiological beam, it can be found photons with different energies. Because of that is common to work with the concept of effective energy. In this study the effective energy of an X-rays system used in instruments calibration was determined, as part of the mammography radiation qualities establishment. The procedure presented here was developed based on information found in the literature. The X-ray mass attenuation coefficients for aluminum, given by NIST website, were used and the mathematical adjusts were done using the Origin 8.0 program. The results are part of the mammographic X-rays beams characteristics determination and it is important to keep the quality of this reference system. 1 Introduction The X-rays beams used in diagnostic radiology are heterogeneous. This means that, in an X-ray beam, can be found photons with different energies. This phenomenon happens because, in the interaction of the electrons with the anode, both characteristic and Bremmstrahlung radiation may occur. Because of this it is common to work with the concept of effective energy, defined as the energy of a heterogeneous beam with the same HVL (half-value layer) as a homogeneous beam[1]. To determine the effective energy is necessary to use equations and values available in literature[2,3]. The effective energy determination is an important issue for the characterization of radiation qualities established in X-ray systems used to instruments calibration.

2 Materials and Method 2.1 Materials The effective energy was determined for the mammography radiation qualities established in a Pantak/Seifert X-rays system, with tungsten (W) target (anode), inherent filtration of 0.138 mm of aluminum (Al)[4] and a window of 0.8 mm beryllium (Be)[5]. The mammography qualities were already established in this system using Al and molybdenum (Mo) filters as additional filtration. The tube voltages used were 25 kv, 28 kv, 30 kv and 35 kv. The mathematical analysis was done using the Origin 8.0 program. 2.2 Methods The first step in this procedure was to determine the attenuation coefficient ( ) of the material used to determine the HVL (in this case, aluminum), for each established radiation quality. The equation 1 was used in this case. I o is the beam intensity without additional filtration, and I is the intensity after the insertion of attenuate material with thickness x. From the equation 1, when is using the HVL values, and applying the natural logarithm, the value can be determined using the equation 2. The next step was to obtain the ratio /, in which is the material specific mass. It can be found in some books or specifics tables[2]. At National Institute of Standards and Technology (NIST)[3], it is presented the graphic that relates the ratio µ/ with the effective energy. This graphic is shown at figure 1... (1) (2) FIG. 1. Graphic that relates, cm²/g, with photon energy in MeV (continuous line).

(cm²/g) The NIST also provides the values used to build this graphic, which makes the link between the ratio / and the respective effective energy. Comparing the calculated ratio with the values given by NIST it is possible to determine the effective energy for each one of the established radiation qualities. 3 Results The mammography radiation qualities were established using the HVL values presented by the German Primary Standard Dosimetry Laboratory Physikalisch- Technische Bundesanstalt (PTB). The calculated values (ln2/hvl) were divided by the Al specific mass ( = 2.699 g/cm³). The results ( were compared with those presented by NIST[3]. To find the effective energy values related to the values calculated in this work it was developed a special procedure. Usually in the range of mammography beams, the photon energy is between 14 kev and 25 kev[1,6]. In the figure 1, the graphic section referent to this energy shows a linear tendency. Using the values which are in this part of the graphic(figure 1), it was built another graphic, figure 2. Adjusting the values it was drawing the curve in order to obtain the energy values. 10 1 0,1 10 20 30 40 50 60 Energia (kev) Fig. 2. Fit linking the points obtained from the table given by NIST As the both axes are in logarithmic scale, it is not possible to find a straight line doing a linear fit, because of that the option ExpDec2 of the Origin 8.0 it was used to fit the curve. The equation found after the adjustment was:. (3)

The adjusted parameters values are shown in table 1: Table 1. Parameters values of the equation given by Origin in the exponential fit Parameter Value Uncertain y o 0.2976 0.0277 A1 633.1038 44.9821 t1 2.77845 0.08632 A2 27.84462 3.97565 t2 8.5271 0.4431 Knowing the parameters of the equation 3 it is possible to determine the effective energy. However, this equation is not simple to solve. The best way to obtain the energy value is fitting the graphic, in the program Origin, using a great number of points (about 50000). Therefore, using the Data Reader tool, is possible to obtain the x and y values at a specific point just putting the mouse pointer over the curve. Using this procedure it was possible to determine the effective energy values. The obtained results, for entrance (WAV with Al and WMV with Mo) and attenuated (WAH with Al and WMH with Mo) radiation qualities, are shown in tables 2, 3, 4, and 5 Table 2. Ratio / and effective energy values for WAV radiation qualities (kv) (mmal) (mmal) (cm²/g) (kev) WAV 25 25 0.57 0.35 7.338 15.4 ± 6.0 WAV 28 28 0.57 0.40 6.421 16.1 ± 6.3 WAV 30 30 0.58 0.43 5.973 16.5 ± 6.4 WAV 35 35 0.62 0.51 5.036 17.5 ± 6.8 Table 3. Ratio / and effective energy values for WMV radiation qualities (kv) (mmmo) (mmal) (cm²/g) (kev) WMV 25 25 0.07 0.36 7.134 15.6 ± 6.0 WMV 28 28 0.07 0.37 6.941 15.7 ± 6.1 WMV 30 30 0.07 0.38 6.758 15.8 ± 6.2 WMV 35 35 0.07 0.41 6.264 16.3 ± 6.3 Table 4. Ratio / and effective energy values for WAH radiation qualities (kv) (mmal) (mmal) (cm²/g) (kev) WAH 28 28 0.57 + 2 0.88 2.918 21.2 ± 8.2 WAH 30 30 0.58 + 2 0.97 2.648 22.0 ± 8.5 WAH 35 35 0.62 + 2 1.21 2.122 23.8 ± 9.3

Table 5. Ratio / and effective energy values for WMH radiation qualities (kv) (mmmo) (mmal) (mmal) (cm²/g) (kev) WMH 25 25 0.07 2 0.56 4.586 18.1 ± 7.0 WMH 28 28 0.07 2 0.61 4.210 18.6 ± 7.2 WMH 30 30 0.07 2 0.68 3.777 19.4 ± 7.5 WMH 35 35 0.07 2 0.93 2.761 21.6 ± 8.4 4 Conclusion The procedure used allowed to obtain the effective energy values however the uncertainties are high (about 40%). This happened because uncertainty values used for the combined uncertainties calculation (those given by the program Origin ) are also very high. In this case it was not performed a set of measurements, so it was not possible to determine the coverage factor. Thus, there has been used k = 2, because this value is more commonly used, representing 95 % of level of confidence. The determination of these values was so important to cover all factors needed to have this beam well known, increasing the quality and reliability of the calibrations made in this system Acknowledgments The authors acknowledge the partial financial support of the Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq), Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) and Ministério da Ciência e Tecnologia (MCT, Project: Instituto Nacional de Ciência e Tecnologia (INCT) em Metrologia das Radiações na Medicina), Brazil. References 1. GUERRA, A. B. Estabelecimento e controle de qualidade de feixes padrões de radiação X para calibração de instrumentos, nível mamografia. 2001. Ph.D. Thesis Instituto de Pesquisas Energéticas e Nucleares, São Paulo, SP. 2. JOHNS, H. E.; CUNNINGHAM, J. R. The Physics of Radiology. 4ª Edição. Ed. Charles C. Thomas. U.S.A. 1983 3. NIST X-ray Mass Attenuation Coefficients Available in: <http://physics.nist.gov/physrefdata/xraymasscoef/elemtab/z13.html> Accessed on: November 09 th 2010

4.FRANCISCATTO, P. C. Caracterização das qualidades de radiação X seguindo as recomendações da norma IEC 61267 no laboratório de calibração do IPEN. 2009. Master Thesis Instituto de Pesquisas Energéticas e Nucleares, Universidade de São Paulo, São Paulo, SP. 5.MAIA, A. F. Padronização de feixes e metodologia dosimétrica em tomografia computadorizada. 2005. Ph.D. Thesis Instituto de Pesquisas Energéticas e Nucleares, Universidade de São Paulo, São Paulo, SP. 6. Established qualities at the German standard primary laboratory Physikalisch-Technische Bundesanstalt (PTB). Available in: <http://www.ptb.de/de/org/6/62/625/pdf/strhlq.pdf> Accessed on: April 1 st 2011