ANALYSIS OF WATER FOR PPB RANGE CHLORINATED ORGANICS USING A TOTAL ORGANIC CHLORINE ANALYZER



Similar documents
METHOD 3510C SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION

Introduction. The following definitions may help you better understand the components of the data report.

PCB SPIU CLEANUP CAUSE AND EFFECT DIAGRAM. Bruce A. Bohnen Technical Director Integrated Chemistries, Incorporated 1970 Oakcrest Avenue, Suite 215

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

XI (a). WASTE DISPOSAL -- CHEMICAL WASTE

This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.

Appendix C Data Validation Report

METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY

NWTPH-HCID. Hydrocarbon Identification Method for Soil and Water

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

6. ANALYTICAL METHODS

The purpose of this bid is to procure a contract with a laboratory to provide analytical testing services on an as-needed basis.

Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS No. SSI-GCMS Shilpi Chopra, Ph.D.

METHOD 8082A. POLYCHLORINATED BIPHENYLS (PCBs) BY GAS CHROMATOGRAPHY

Hazardous Waste Containers. Hazardous Waste Disposal. Examples of HW Containers. Typical Wastes in Chemistry

Water Analysis Report Report Date: March 31, 2012 Testing Period: 1st - 4th Quarter 2011

Separation by Solvent Extraction

Determination of calcium by Standardized EDTA Solution

Quantitation of Drugs in Dried Bloodstains. Thomas Meyer Anna Gomenyuk Nadiah Lester

TERRA TEK LIMITED GUIDE TO SAMPLE HANDLING. Version 1.06

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

APPENDIX II TIER I CHECKLIST

Appendix D lists the Field Services Standard Operating Procedures. Appendix E lists the Biological Monitoring Standard Operating Procedures.

Measurement by Ion Selective Electrodes

Extraction Theory. Org I Lab W. J. Kelly. Liquid-liquid extraction is a useful method to separate components (compounds) of a mixture

Experiment 8 Preparation of Cyclohexanone by Hypochlorite Oxidation

Pace Analytical e-report

How To Test For Contamination In Large Volume Water

DOWEX Resins as Organic Solvent Desiccants

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Name Lab #3: Solubility of Organic Compounds Objectives: Introduction: soluble insoluble partially soluble miscible immiscible

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston

A Potentiometric Analysis of Fluoride Ion in Toothpaste

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Formulas, Equations and Moles

Draeger-Tubes and accuro Pump

The Empirical Formula of a Compound

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

ANALYTICAL DETECTION LIMIT GUIDANCE. & Laboratory Guide for Determining Method Detection Limits

A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas

Chapter 5 Classification of Organic Compounds by Solubility

Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup

Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance

Aliphatic/Aromatic Separation of Extractable Petroleum Hydrocarbons in Solids or Water by Silica gel Fractionation

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil.

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

Removing Heavy Metals from Wastewater

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE

Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis

Calculation of Molar Masses. Molar Mass. Solutions. Solutions

SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0)

Forensic Fingerprinting Petroleum Contaminants using UVF Field Screening Technology

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

FACTS in Drinking Water

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

AMD Analysis & Technology AG

Soil, Ground Water and Sediment Standards for Use Under Part XV.1 of the Environmental Protection Act Ministry of the Environment April 15, 2011

Fractional Distillation and Gas Chromatography

Coating and Extraction of Honeycomb Denuders

Flow Injection Analysis

Holcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants

PROCEDURES FOR LABORATORY CHEMICAL WASTE DISPOSAL

A Comparison of EPA Methods 8260B and TO-14A/15 for the analysis of VOCs in Soil Gas: Application to Upward Vapor Intrusion Risk Studies

HAZARDOUS CHEMICAL WASTE DISPOSAL SECTION 7

Inferred ph in Steam Plant Water Chemistry Monitoring

FerroVer Method 1 Method to 3.0, 1.0 to 30.0 and 10.0 to mg/l Fe Powder Pillows

Chapter 1: Moles and equations. Learning outcomes. you should be able to:

DATA VALIDATION STANDARD OPERATING PROCEDURES FOR CONTRACT LABORATORY PROGRAM ROUTINE ANALYTICAL SERVICES

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

CCME. Reference Method for the Canada- Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method

Acid-Base Extraction.

F The following spent

General Chemistry I (FC, 09-10) Lab #3: The Empirical Formula of a Compound. Introduction

6 H2O + 6 CO 2 (g) + energy

Neutralizing an Acid and a Base

Experiment 3 Limiting Reactants

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

S olid Ph ase Extraction (SP E) Method:

Version 1.0 September, Jean W. Munch and Paul E. Grimmett

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

EXPERIMENT 12: Empirical Formula of a Compound

Pennsylvania Marcellus Shale Short- Term Ambient Air Sampling Project

Transcription:

DTR-21-01 1 ANALYSIS OF WATER FOR PPB RANGE CHLORINATED ORGANICS USING A TOTAL ORGANIC CHLORINE ANALYZER Theodore B. Lynn, Ph.D., Mary Kate Boggiano, Larry M. Sacramone, David L. Balog, Andrew C. Lynn, Dexsil, Corporation, One Hamden Park Drive, Hamden, Connecticut 06517 Presented at the Waste Testing and Quality Assurance Symposium 2001, August 13-16, 2001, Arlington, A V ABSTRACT Dexsil Corporation has recently developed a total halogen method for the analysis of water samples for chlorinated organics. The new procedure is an addition to the existing methods for the L2000DX Chloride Analyzer. The L2000DX has been in use in the field for 12 years and can be used for the analysis of transformer oil, surface wipes and soils. The soil method is the basis for SW-846 Method 9078 "Screening Test Method for Polychlorinated Biphenyls in Soil." Using the same reaction and quantification steps as the established L2000DX methods, the new method uses a liquid-liquid extraction step to achieve method detection limits (MDLs) in the 15-30 ppb range for most chlorinated solvents. This new method will allow nearly real time data to be collected from ground water monitoring wells at a fraction of the cost of laboratory analysis. This paper describes the new method, the development work in establishing the feasibility of the method and the comparison with SW-846 laboratory methods for spiked samples. Recoveries and MDLs have been determined for the following analytes: Carbon Tetrachloride, 1,2-Dichloroethane (DCEA), cis-1,2-dichloroethylene (DCEE), trans-1,2- Dichloroethylene (DCEE), 1,2-Dichloroethylene (cis/trans mix), Methylene Chloride, Pentachlorophenol (PCP), Tetrachloroethylene (PCE), 1,1,1-Trichloroethane (TCEA), Trichloroethylene (TCEE), Vinyl Chloride. As expected, the extraction recoveries vary inversely with the solubility of the analyte in water which, in turn, affects the MDL for the analyte. The MDLs for these analytes range from 16 ppb for carbon tetrachloride to 157 ppb for methylene chloride with the majority around 25 ppb. The method development data indicate that the L2000DX water method should be a useful tool for monitoring ground water in areas contaminated with chlorinated solvents. Now, one instrument can be used to track contamination in both soil and water. This type of screening can be used for checking the progress of a treatment process, defining the limits of a contamination plume, as well as, monitoring ground water over long periods of time. All of these applications do not require expensive laboratory analysis and the L2000DX can provide the information in the field in 10 minutes INTRODUCTION The L2000DX is a screening tool for chlorinated o rganic compounds and has been used in the field for the analysis of transformer oil for PCB contamination since 1989. Shortly after its introduction, a method was developed to use the L2000DX for the s c reening of soils for PCBs. This method eventually became SW-846 Method 90781and has been evaluated in two diff e rent forms, first, through the SITE p rogram and, most re c e n t l y, under the ETV program2 both run by the USEPA. The pro c e d u re for PCBs in transformer oil has just recently been validated through the ETV program and a method has also been developed for use on wipe samples.3 *Author to whom correspondence should be addressed.

DTR-21-01 2 In the late 90s, with the rise of chlorinated solvents and pesticides as contaminants in the enviro n m e n t, the demand grew for the development of L2000DX based methods for other chlorinated compounds in soil. Laboratory experiments and field trials have since demonstrated that the L2000DX, following suitable matrix preparation, is suited to the analysis of basically any extractable chlorinated organic in s o i l. 4 As chlorinated solvents leach into ground water and i n t e rest in cleaning up old dry cleaning and manufacturing sites grows, so has the need for an L2000DX method for water analysis. The objective of this work was to develop and document a water extraction technique for concentrating org a n o - c h l o r i n e contaminants into an organic solvent suitable for i n t roduction into the L2000DX system. The L2000DX is a total chlorine analyzer. A re q u i rement for the system is that all chlorine must be chemically converted to inorganic chloride for quantification. By changing the sample preparation steps either total chlorine or total organic chlorine will be m e a s u red. There are three elements to the analysis: sample preparation/extraction, converting the org a n i c chlorine present into inorganic chloride using metallic sodium and the quantification of the re s u l t i n g chloride using a chloride ion selective electro d e (ISE). Over the 12 years of its use in the field, the L2000DX has proven to be a very reliable instru m e n t and the ISE based system has been shown to be accurate and relatively free of interf e rences. (NOTE: All o rganic chlorine is quantified as the target analyte, h o w e v e r, this is the nature of a TOC measure m e n t and is not an interf e rence, per se. Inorganic chloride can be an interf e rence, but can be removed in the matrix cleanup step and, there f o re, would not interf e re. ) matrix preparation steps and the conversion to the a p p ropriate units. The fourth correction necessary to obtain a useful result is derived from empirical data on the overall efficiency of the analytical steps, including the extraction efficiency from the original matrix, the reaction efficiency for the conversion to chloride, the re c o v e ry of the chloride ions in the final b u ffer solution, etc. The theoretical conversion factors can be calculated for all analytes and tabulated for each matrix, but the correction for re c o v e r i e s must be experimentally determined for each matrix and each analyte. After some experimentation, it was determined that a simple liquid/liquid extraction would be the simplest and also effective. Of the commonly used solvents, e.g., hexane, heptane, cyclohexane, it was determ i n e d that 2,2,4-trimethylpentane (isooctane) worked well and had the advantage of being less volatile. A liquid/liquid extraction will accomplish two things: by choosing the appropriate solvent, it eliminates interf e rences due to inorganic chloride and by choosing the correct solvent to sample ratio, a concentration of 1:100 can easily be accomplished. The effectiveness of a liquid/liquid extraction is determined primarily by the extracted compound and is, t h e re f o re, not constant over all possible contaminants. In general, compounds with a high water solubility will tend to have low recoveries in the solvent l a y e r. Extraction efficiency can also be affected by water quality parameters, e.g., ph, salinity, ionic s t rength, etc. In this work we set out to determ i n e the extraction efficiency for a liquid/liquid extraction combined with the L2000DX sample preparation as well as the MDL for the quantification of enviro n- mentally significant organo-chlorine contaminants. T h e re are four basic corrections necessary to convert the final chloride reading into the equivalent analyte concentration in the desired units. Three of the corrections are derived corrections calculated from the known chlorine composition of the analyte (perc e n t chlorine), the dilution/extraction volumes of the

DTR-21-01 3 EXPERIMENTAL All samples were prepared in 965 ml of deionized water, cooled overnight, in 1-quart glass jars from Quality Environmental Containers. The majority of the water samples were between 11 and 14 EC when extracted, determined after the extraction was complete. The temperature was varied between 2EC and 45EC for water samples used to determine the effects of temperature on the extraction of methylene chloride. To study the effects of ionic strength, ph and salinity on the recovery of methylene chloride, deionized water was first spiked with either 1% sodium sulfate, hydrochloric acid (ph < 2), sodium hydroxide (ph>12) or 1% sodium chloride, shaken and then refrigerated. Recovery experiments using PCP were also conducted on ph adjusted water. Each day, a fresh stock solution of approximately 2500 ppm of a single analyte was prepared in methanol. Appropriate volumes of stock solution were injected by syringe into the cooled water to yield the desired analyte concentrations. Each jar was gently turned to ensure mixing of the analyte, while minimizing its partitioning into the headspace. To determine the method recoveries and linearity, duplicate samples were prepared at the following concentrations: 0.15, 0.25, 0.50, 1.0, 2.0, 5.0, and 10.0 ppm. Additionally, the MDL was determined for each analyte according to 40 CFR Part 136.5 The seven replicate water samples for each analyte were prepared in the same way at concentrations within a factor of 3-5 of the estimated MDL (See Table 1). For chloride analysis by L2000DX, the analyte of each sample was extracted into 10 ml isooctane, followed by vigorous shaking for two minutes. The sample jars were then filled to zero headspace with deionized water, and the isooctane was allowed to settle in the neck of the bottle for at least three minutes. After the isooctane extract had settled, 5 ml were transferred into an L2000DX reaction tube. The samples were further prepared according to the instructions provided with the L2000DX instruction manual.6 This preparation involved the complete isolation of chloride ions from the analyte. The L2000DX instrument was calibrated with a standard 50 ppm chloride solution. Before the samples were tested, standard 10 ppm and 1000 ppm chloride solutions were checked. The chloride content of all samples were measured using the uncorrected L2000DX Chloride Method. Initial preparation of samples for GC analysis was identical to that of the L2000DX samples. For the GC readings, however, the samples were injected into a purge-and-trap device after the chlorinated solvent had been gently mixed into the 965 ml of water. For analyte concentrations up to 2 ppm, 5 ml of sample were drawn into a syringe with a Luer-Lock connection without a needle. Into this solution, 5 ml of an internal standard was injected. The internal standard consisted of a 2000 mg/ml solution of 2-bromo-1- chloropropane and fluorobenzene in methanol. This combined solution with internal standard was injected into the purge-and-trap device. For 5 ppm and 10 ppm solutions, 1mL of sample was diluted to 5 ml with deionized water, then injected into the purgeand-trap along with 5 ml internal standard. The purge-and-trap served to extract all the low-boiling solvents out of the sample into the gas phase by bubbling an inert purge gas through the water. The analyte and internal standards were collected in the trap. After the purging was complete, all analytes that had collected into the trap were desorbed onto the GC column. A calibration curve of the analytes studied was made from 0.15 ppm to 2.0 ppm. The calibration curve and all samples were run using the EPA Volatiles 502.2 method. PERC was additionally analyzed by GCMS, using EPA VOC Method 8260. The GCMS samples were also injected via purge-and-trap and most were diluted by a factor of 10 or 100, depending on the initial concentration of the samples. This method used an internal standard mixture of pentafluorobenzene, 1,4-

DTR-21-01 4 difluorobenzene, chlorobenzene-d5, and 1,4- dichlorobenzene at a fixed concentration reading of 50.00 mg/l. In the internal standard mixture, the surrogate analytes present were dibromofluoromethane, toluene-d8, and 4-bromofluorobenzene at concentrations of approximately 100.0 mg/l. A calibration curve for PERC was made of concentrations ranging from 5 ppb to 200 ppb. RESULTS AND DISCUSSION The methylene chloride experiments indicate that v a rying water quality parameters have little effect on the re c o v e ry of chlorinated solvents. While this means that the re c o v e ry of methylene chloride cannot be easily improved, it also indicates that variations in g round water should not significantly affect re c o v e r- ies in the field. As expected, ph did affect the re c o v- e ry of PCP. The re c o v e ry improved from 54% to 60% when the ph was reduced from neutral to less than 2. The one water parameter that did have an affect on re c o v e ry of chlorinated solvents was temperature. Spiking water of various temperatures at 1 ppm with PCE, resulted in recoveries ranging from 35% at 2EC to approximately 60-65% at 12EC. Above 12EC the recoveries remained more or less constant in the 60-70% range, up to the highest temperature tested at 45 EC. In order to determine the recoveries, theoretical corrections were made to the instrumental readings to account for percent chloride (on L2000DX re a d i n g s only) and dilution/concentration factors. While the GC experiments analyzed the VOC analyte concentrations dire c t l y, the L2000DX uses a chloride-specific electrode, which measures free chloride concentration. The L2000DX readings were converted fro m chloride concentrations to analyte concentrations by dividing the measurement by the percent chlorine in the analyte. The extraction of the analyte from water to isooctane produced a size multiplier for the L2000DX samples of 965/10 = 96.5. Although there was no extraction involved in the GC samples, a size multiplier of 5 was used for the most concentrated samples since these samples were diluted by a factor of five. Similarly for GCMS samples, the dilution factor was the size multiplier. Once these theore t i c a l c o rrections had been made, a re g ression analysis of the corrected instrument response versus the theore t- ical concentration for each analyte indicates the linearity of the response. Table 1 shows the comparison of the coefficient of re g ression for both the L2000DX and the GC/GCMS response curves. Figure 1 showing the measured concentration versus the theore t i c a l concentration for PCE, illustrates a typical re s p o n s e c u rve for both methods. The slopes of the re g re s s i o n line can be taken as the average percent recoveries for each analyte, as shown in Table 2. A plot of the average re c o v e ry versus the analyte solubility in water for all analytes tested, indicates that the re c o v e ry is inversely pro p o rtional to an analyte s solubility (See Figure 2). Using published data, it can be seen in Figure 3 that a reasonable appro x i m a- tion of the re c o v e ry can also be made based on an analytes octanol/water partition coeff i c i e n t. 7 The ultimate sensitivity of the L2000DX method is d e t e rmined by the sensitivity of the ISE. The practical limit for chloride detection, with some low-end c o rrections, is 1 ppm in the final extract solution used for quantification. Since all of the pre p a r a t i o n steps are the same for each analyte, and the final extraction multiplier is determined by the extraction e ff i c i e n c y, the achievable MDL for each analyte will, t h e re f o re, be a function of the re c o v e ry. The re s u l t s f rom the MDL determinations, listed in Table 2, conf i rm this general trend with the lowest MDL of 16 ppb achieved for carbon tetrachloride ( 58% re c o v- e ry) and the highest of 157 ppb for methylene chloride (9% re c o v e ry ).

DTR-21-01 5 SUMMARY The results from the response range experiments indicate that the L2000DX liquid/liquid extraction method for water is has a linear range up to 10 ppm for all of the analytes tested. The coefficient of regression was greater than 0.96 for all analytes including methylene chloride, the hardest to extract. The recovery for each of the analytes has been demonstrated to be consistent and reproducible enough to make this method a suitable method for field use. Typical MDLs are in the 25 ppb range for most of the analytes of interest which should make the L2000DX water method a useful tool for monitoring ground water in areas contaminated with chlorinated solvents. Now, one instrument can be used to track contamination in both soil and water. This type of screening can be used for checking the progress of a treatment process, defining the limits of a contamination plume, as well as, monitoring ground water over long periods of time. All of these applications do not require expensive laboratory analysis and the L2000DX can provide the information in the field in 10 minutes. REFERENCES 1. US-EPA (1998) Screening Test Method for Polychlorinated Biphenyls in Soil, Test Method for Evaluating Solid Waste, Physical/Chemical Methods (SW-846), Update IVA, U.S. Government Printing Office, Washington, D.C. 2. Dindal, A.B., Billetts, S, (1998) Electrochemical Technique/Ion Specific Electrode Dexsil Corporation L2000 PCB/Chloride Analyzer, Environmental Technology Verification Report, US-EPA, Office of Research and Development, Washington, D.C., EPA/600/R-98/109, August 3. Dindal, A.B., Bayne, C.K., Koglin, E.N., (2000) PCB Detection Technology Dexsil Corporation L2000DX Analyzer, Draft Environmental Technology Verification Report, US-EPA Office of Research and Development, Washington, D.C. Report (in Press) 4. Lynn, T.B., Kneece, J.C., Meyer, B.J., Lynn, A.C. (1997) Determination of Chlorinated Hydrocarbon Concentrations in Soil Using a Total Organic Halogen Method, Presented at the 13th Annual Waste Testing and Quality Assurance Symposium, July 6-9, Arlington, VA 5. Definition and Procedure for the Determination of the Method Detection Limit, US-EPA Publication, Code of Federal Regulations, 40 CFR Ch. 1 (7-1-89 Edition), Appendix B, part 136, pp. 525-527 6. Dexsil Corporation (2000) L2000DX Users Manual, Hamden, CT 7. Mackay, D., Shiu, W. Y., Ma, C. K., (1993) Illustrated Handbook of Physical-Chemical Properties and Environmental Fate for Organic Chemicals, Volume 3, Lewis Publishers, Chelsea, Michigan

DTR-21-01 6 Table 1: Comparison of Recoveries and Regression Coefficients for GC and L2000DX Response Curves Analyte Carbon Tetrachloride 1,2-Dichloroethane (DCEA) cis-1,2-dichloroethylene (DCEE) trans-1,2-dichloroethylene (DCEE) 1,2-Dichloroethylene (cis/transmix) Methylene Chloride Pentachlorophenol (PCP) Tetrachloroethylene (PCE) 1,1,1-Trichloroethane (TCEA) Trichloroethylene (TCEE) Vinyl Chloride L2000DX R2 0.974 0.992 0.987 0.987 0.990 0.992 0.996 0.969 0.989 0.965 0.990 L2000DX Recovery 0.58 0.18 0.26 0.41 0.26 0.09 0.54 (0.6@pH<2) 0.59 0.37 0.41 0.11 GC/GCMS R2 0.991 0.990 0.973/0.990 0.991 0.996 GC Recovery 0.95 1.15 1.20/1.19 1.16 1.14 Table 2: Physical Data and L2000DX MDLs and Recoveries Analyte Carbon Tetrachloride 1,2-Dichloroethane (DCEA) cis-1,2-dichloroethylene (DCEE) trans-1,2-dichloroethylene (DCEE) 1,2-Dichloroethylene (cis/transmix) Methylene Chloride Pentachlorophenol (PCP) Tetrachloroethylene (PCE) 1,1,1-Trichloroethane (TCEA) Trichloroethylene (TCEE) Vinyl Chloride H2O Sol. (wt %) 0.08 0.81 0.35 0.63 1.32 0.008 0.015 0.13 0.11 0.88 Typical Log Kow 2.6 (2.2-3) 1.5 (1.4-1.8) 1.7 (1.5-1.9) 1.9 (1.5-2.1) 1.2 (1-1.5) 3 (1.3-5.9) 2.7 (2.4-3.4) 2.5 (2-2.6) 2.3 (2-3.3) 1 (0.6-1.4) Recovery 0.58 0.18 0.26 0.41 0.26 0.09 0.54(0.6@pH<2) 0.59 0.37 0.41 0.11 Spike Level (ppb) 51 250 150 377 72 53 58 63 207 MDL (ppb) 16 71 86 157 37 22 25 23 94

DTR-21-01 7 Figure 1: Recovery of PCE from Water (omitting 2 outliers) 14 DX 12 GCMS Measured Concentration (ppm) 10 8 6 4 2 y = 1.186x - 0.0149 R 2 = 0.9902 y = x - 0.0223 R 2 = 0.9692 0 0 2 4 6 8 10 12 Theoretical Concentration (ppm) Figure 2: Recoveries Vs Solubilities (all analytes) 0.7 0.6 0.5 y = -0.366x + 0.5067 R 2 = 0.7628 Recoveries 0.4 0.3 0.2 0.1 0.0 0 0.2 0.4 0.6 0.8 1 1.2 1.4 Solubility in Water (wt%)

DTR-21-01 8 Figure 3: Recoveries Vs Log Kow (all analytes) 0.7 0.6 0.5 Recoveries 0.4 0.3 0.2 0.1 y = 0.2614x - 0.1793 R 2 = 0.8781 0.0 1 1.5 2 2.5 3 Log Kow