Evaluation of headspace Solid Phase Micro-extraction method for analysis of phosphine residues in wheat



Similar documents
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS No. SSI-GCMS Shilpi Chopra, Ph.D.

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

SUCRALOSE. White to off-white, practically odourless crystalline powder

POLYVINYL ALCOHOL. SYNONYMS Vinyl alcohol polymer, PVOH, INS No DEFINITION DESCRIPTION FUNCTIONAL USES CHARACTERISTICS

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

ß-CYCLODEXTRIN SYNONYMS

With proper calibration, the upper limit may be extended as needed.

Gas Chromatography Liner Selection Guide

Expectations for GC-MS Lab

Split / Splitless Injection for Capillary GC

Aromatic plants: from extraction to analysis

Ethyl formate fumigation as a replacement for methyl bromide fumigation in dried vine fruit.

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

Purification of reaction mixtures using flash chromatography.

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

AMD Analysis & Technology AG

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup

GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits

Dissolved and precipitated oxalate

Chapter 28: High-Performance Liquid Chromatography (HPLC)

ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC)

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles

105 Adopted:

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Analyzing Small Molecules by EI and GC-MS. July 2014

A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Background Information

Version 1.0 September, Jean W. Munch and Paul E. Grimmett

Pesticide Analysis by Mass Spectrometry

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2

Analysis of Liquid Samples on the Agilent GC-MS

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

Air Sampling for Gases and Vapors. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University

CARBON DIOXIDE CONCENTRATION IN HERMETIC STORAGE OF SOYBEAN (GLYCINE MAX) IN SMALL GLASS JARS

HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.

Enhancing Utility and Throughput of SPME using Advanced Automation Strategies

ORGANIC SAMPLE PREPARATION

CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA ABSTRACT

Vitamin C quantification using reversed-phase ion-pairing HPLC

SCION SQ Series GC-MS

EnAlgae SOP: Biogas analysis

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase

How To Test For Contamination In Large Volume Water

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Bologna Maggio 2011; 1 MS Environmental Day

TANNIC ACID. SYNONYMS Tannins (food grade), gallotannic acid, INS No. 181 DEFINITION DESCRIPTION

A new concept for isotope ratio monitoring LC/MS. A Wide Range of Applications

Novel Method for Solid Phase Peptide Synthesis Using Microwave Energy

MILESTONE H E L P I N G C H E M I S T S

PLOT Columns. Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Multi-Shot Pyrolyzer EGA/PY-3030D. Flexible Versatile Reproducible

(S&G 5th ed. Expt 27, 6th, 7th & 8th eds. Expt 26)

Gas Chromatograph GC Consumables. p/n:

Determination of Calcium and Magnesium in Brine

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY. Andrew Tipler

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

Identification of Arson Accelerants by Gas Chromatography

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Micro Cryo-Trap Model 971

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Development and Validation of Gas Chromatographic Analytical Method for Estimation of Residual Solvents in Excipients (Klucell, Triacetan, HPMC)

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

On-line Dissolved Gas Analysis

SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0)

23 The Thermal Conductivity Detector

6. ANALYTICAL METHODS

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder

This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA

Transcription:

Evaluation of headspace Solid Phase Micro-extraction method for analysis of phosphine residues in wheat Ren, Y.L. 1,2,3 *#, Padovan, B. 4 1 Cooperative Research Centre for National Plant Biosecurity, Australia, Email: y.ren@murdoch.edu.au 2 School of Biological Science and Biotechnology, Murdoch University, South Street, Murdoch, WA, 6150 Australia 3 Department of Agriculture and Food, Western Australia, 3 Baron Hay Court South Perth WA 6151, Australia 4 CSIRO Entomology, GPO Box 1700, Canberra ACT 2601, Australia * Corresponding author # Presenting author DOI: 10.5073/jka.2010.425.095 Abstract This new method utilizes headspace-solid-phase micro extraction (HS-SPME) for pre-concentration of PH 3. Phosphine was determined with gas chromatography/pulsed flame photometric detector (PFPD). Spiked samples were used for calculation of phosphine residue in grain. Four types of fibres (100μm- PDMS, 85µm-CAR/PDMS, 75µm-CAR/PDMS and 65μm-PDMS/DVB) were tested. The bipolar fibres (CAR/PDMS and PDMS/DVB) can extract PH 3, but the non-polar fibre (PDMS) did not. Larger size fibres extracted PH 3 more efficiently than the smaller size fibres (e.g., 85 µm > 75 µm > 65 µm). The 85µm CAR/PDMS fibre was used to optimize the different parameters that affect the SPME extraction efficiency of PH 3. In the validation study, 50 grams of wheat in a 250 ml glass flask and capped with an open-top screw cap and PTFE/Silicon septa were spiked at 0.02 ng PH 3 /g of wheat. The flask was then heated to 45 C in an oil bath for 45 min, after which time the 85 µm CAR/PDMS fibre was exposed for 20 min and then exposed in the heated injection port of a GC/PFPD and desorbed for 2 min. Under conditions of the validation study, the limit of detection (LOD) or level of quantification (LOQ) was in the range of 0.005 0.01 ng PH 3 /g of wheat. Keywords: Fumigant, Phosphine, Residue, SPME, HS-SPME 1. Introduction The world grain industry relies heavily on chemicals for grain treatment. Currently, phosphine (PH 3 ) is the only fumigant available to treat bulk grains and oil seeds (more than 85% grains are treated/re-treated with PH 3 ) in each of the linkages from on-farm storage to central storage (Collins et al., 2000; Collins et al., 2002; Ren and Mahon, 2007). However, a restrictive Codex Maximum Residue Limit (MRL) of 0.1 mg/kg, and lower limits set by some purchasers, are challenging the use of PH 3 for treatment of grain. Phosphine residues can remain in the grain for several months. Multi-fumigation with PH 3 can also cause accumulation of residues in the grain exceeding the MRLs. It is also important to note that levels of PH 3 residues in grain, producing low aerial concentrations of PH 3, can also facilitate resistance. Therefore, it is necessary to understand PH 3 residues in bulk grain to establish better procedures for multi-fumigation. This will guide industry in the conduct of good PH 3 fumigation practice to minimise PH 3 residue in fumigated grain. These include application methods and choice of the right periods of exposure and airing. Typical procedures for analysis of PH 3 residues include: 1) removal from the commodity matrix by either purge and trap techniques or by solvent extraction (Miyahara and Saito, 1994), and 2) and then analysis by gas chromatography (GC) or gas chromatography-mass spectra (GC-MS). However, purge and trap methods are not suitable for highly volatile fumigants such as PH 3 as they are unable to trap all fumigant (Ren and Desmarchelier, 1998). Solvent extraction has the problem of solvent interference and is timeconsuming. None of the above methods are suitable for analysis of PH 3 residues in grains, particularly at very low levels. Microwave irradiation is being increasingly used in the digestion of samples (Ren, 2001; Desmarchelier et al., 1998), and, in recent work (Chemfate 2004), excellent recoveries and precision have been obtained from microwave extractions of PH 3 from wheat and hay (Ren and Mahon, 2007). However, care is required in selecting the appropriate power setting and the safety implications of heating sealed flasks in microwave ovens is of concern, and water vapour generated from microwave irradiation can significantly interfere with GC results (Ren and Mahon, 2007). Julius-Kühn-Archiv, 425, 2010 407

Solid-phase microextraction (SPME) in combination with head space (SH-SPME) analysis by gas chromatography is a convenient alternative method for volatiles. Solid-phase micro extraction is a simple, sensitive, and solvent-free technique that has become popular in a wide range of applications (Penalver et al., 2001). It has been used to study pollutant gases and volatile degradation products (Lattuai-Derieux, et al., 2004). For example residues of methyl isothiocyante (MITC) were analysed by HS-SPME with the 85 µm Polyacrylate (PA) fibre (Ren et al., 2008a). Using the HS-SPME method, MITC residues from wheat can be successfully analysed at levels below 0.1 ng/g, compared with purge and trap methods where the limit of detection is 10 ng/g (Ren et al., 2008a,b). This paper describes a new method for analysis of PH 3 residues in wheat which utilises headspace-solid-phase micro extraction (HS- SPME). It includes selection of the SPME fibre and factors that affect the extraction efficacy of PH 3. 2. Materials and methods 2.1. Reagents and apparatus One litre Erlenmeyer flasks (Bibby Sterilin, Staffordshire, Cat. No. FE 1 L/3) were used for fumigation of wheat and preparation of standards. The measured volume of each Erlenmeyer flask and inlet system was calculated from the weight of water required to fill the container and was used for calculations. Bottles of 250 ml (Alltech Cat. No. 9535) were used for the microwave extraction. Each bottle was fitted with a Mininert valve equipped with septa (Alltech Cat. No. 95326). A 100 μl air tight syringe with valve (SGE, Melbourne, Australia; Cat. No 005279) was used for GOW-MAC gas density balance injection and transfer of PH 3 from source to fumigation chambers and flasks. The SPME fibres used were coated with 85 µm Carboxen/Polydimethylsiloxane (CAR/PDMS) fibre (Sigma-Aldrich Australia, Cat. 57334-U), 75 µm Carboxen/Polydimethylsiloxane (CAR/PDMS) fibre (Sigma-Aldrich Australia, Cat. 57344-U), 65μm Polydimethylsiloxane/Divinylbenzene (PDMS/DVB) fibre (Sigma-Aldrich Australia, Cat. 57310-U) and 100 μm Polydimethylsiloxane (PDMS) fibre (Sigma- Aldrich Australia, Cat. 57301), respectively. All fibres were conditioned at 270 C for 1 h prior to use in accordance with manufacturers recommendations. Phosphine (85.0% PH 3 and 15.0% air and CO 2 ) was laboratory prepared by the FAO method (FAO, 1975). The purity of PH 3 was determined using a GOW-MAC gas density balance (GOW-MAC Instrument Co., Madison, N.J.) after separation of the gases on a 1 m 5 mm i.d. Porapak Q 100/120 mesh (Alltech Associates, Cat. No. 2702) column at 105 C with a carrier flow (N 2 ) of 150 ml/min. The reference gas used was tetrafluoroethane (> 99.9%), which was supplied by ACTROL Ltd, Australia. Phosphine (PH 3 ) was determined on a Varian CP-3800 (Varian Instruments, Sunnyvale, CA), equipped with a pulsed flame photometric detector (PFPD) with phosphorus filter. Separation was achieved on a 30 m 0.53 mm ID, AT-Q column (Alltech Associates, Cat. No. 0810025, Part No. 13939) at 125 C and carrier flow (N 2 ) of 5.0 ml/min at 5.0 psi. Injector and detector temperatures were 200 C. Injection volumes of gases were 40 µl. A minimum interval of 5 min was kept between injections, in order to elute interfering chemicals. 2.2. Wheat sample and fumigation of wheat Wheat used was Australian standard white wheat, 10.9% moisture content, w/w wet basis. The moisture content of the wheat was measured by oven drying at 105 C for 2 hours (ISO 712 International Standard 1998). In order to ensure the moisture result is precise, 500 kg wheat sample was quartered equally using a Boerner grain divider. One quarter sample (125 g) was quartered equally again and then one quarter sample of about 30 g was collected and ground with a laboratory sealed metal grinder for measurement of moisture content. Four replicates samples were dried at 105 C in mechanical convection oven ( DK 62 American Scientific Products Columbus, OH, USA) and the loss of weight was used to calculate the moisture content of the sample. Wheat samples (860 g) were fumigated in an Erlenmeyer flask (1 L) equipped with a lid fitted a septum injection system for 7 days at concentrations of 0.5 mg (PH 3 )/L. After 7 days exposure, the flask was opened and aired for 7 days in a fume hood to obtain samples containing residual fumigant. 408 Julius-Kühn-Archiv, 425, 2010

2.3. Preparation of diluted PH 3 gas and spiking standards Diluted PH 3 gases were prepared by first removing the same volume of air as the known volume (0.5 ml) of concentrated PH 3 to be injected into an Erlenmeyer flask (1 L) containing six glass beads (2-3 mm OD). For validation study, the limit of detection (LOD) or level of quantification (LOQ) was investigated using spiked standards. Spiked wheat samples at and 0.02 and 0.5 ng PH 3 /g of wheat were prepared by adding appropriate volumes of PH 3 into sealed flasks (250 ml) containing wheat (50 g) 3 min before oil bath heating. Triplicate samples were used and each sample was injected duplicates (n=6). The dosages and required volumes for PH 3 concentrations, calibrated to the current laboratory temperature and pressure, were calculated from Eq. 1 Where: 4 T 1.7 10 C V V f = (1 ) ( ) Eq. 1 273 P M N V is volume of fumigation container (L); P is pressure (mm Hg); T is temperature ( C); C is the intended concentration of methyl bromide (mg/l); V f is dosage volume of fumigant (ml); M is molecule weight of fumigant; N is purity of gas (%). 2.4. Selection of fibre and HS-SPME extraction time Four different types of fibres (100 μm-pdms, 85 µm-car/pdms, 75 µm-car/pdms and 65 μm- PDMS/DVB) were tested at 20 min extraction time. Diluted PH 3 gas was injected into a 250 ml flask fitted with a sample port to obtain 0.3, 0.1, 0.05 and 0.01 ppm of PH 3, respectively. Headspace samplings were carried out on all four types of fibres. The needle was carefully inserted into the headspace of the flask and the fibre exposed into the headspace for 20 min. At the end of the defined extraction time, the fibre was withdrawn from the headspace into the needle. The fibre holder was removed from the extraction flask and inserted into the injection port. The fibre was extended into a GC-PFPD inlet where sample components were desorbed at 200 C for 5 min to clean it between extractions. For evaluation of HS-SPME extraction time on the efficacy of PH 3 extraction, the SPME equilibration time was determined by exposing the fibres (85 µm-car/pdms, 75 µm-car/pdms and 65 μm- PDMS/DVB) to 0.3 ppm of PH 3 for 1, 5, 10, 15, 20 and 30 min, respectively. Finally the fibre was retracted and then exposed in the heated injection port (200 C) of a GC-PFPD and desorbed for 2 min. Results are the mean of duplicate samples and injections for each sample (n=4). 2.5. Evaluation of GC Injector temperature for desorption of PH 3 from fibre (85µm CAR/PDMS) Efficient thermal desorption of the analysis in a GC injection port is dependent on the type of fibres and on the GC injector temperature. The 85 µm CAR/PDMS fibre was used to optimize the injector temperature. It was inserted into the headspace of the flask containing 0.1 ppm of PH 3 and exposed for 20 min. The fibre was carefully injected into the GC at different injector temperatures (100,150, 200 and 250 C) and desorbed for 5 min. Results are the mean of duplicate samples and injections for each sample (n=4). 2.6. Evaluation of wheat treatment and HS-SPME extraction temperature for extraction efficacy of PH 3 A fumigated wheat sample of 50 g in a 250 ml flask fitted with a sample port was immersed in a heated oil bath at 30 C for 30 min, after which time the fibre was inserted into the headspace of the flask and exposed for 20 min. Equivalent flasks were heated a 35, 40, 45 and 50 C. The fibre was injected (200 C) into a GC-PFPD and desorbed for 5 min. Spiked wheat samples were treated in exactly the same manner. Levels of PH 3 residue were determined against spiked standards. The peak areas were calibrated periodically using a fortified standard, and the data presented are the mean of duplicate samples and injections for each sample (n=4). Julius-Kühn-Archiv, 425, 2010 409

2.7. Statistical analysis The GC readings of fumigant concentration were averaged respectively. The average variation of GC readings and PH 3 concentration and standard deviations (SD) between the duplicate treatments and injections were analysed with Microsoft Excel. 3. Result and discussion 3.1. Effect of fibre types and HS-SPME extraction time on extraction efficacy of PH 3 The amounts of PH 3 absorbed by each of four different SPME fibres are shown in Figure 1 for each of four PH 3 concentrations. These four fibres showed different efficiency and selectivity in extracting PH 3 in the headspace. The bipolar fibres (CAR/PDMS and PDMS/DVB) which dispersed solid adsorbents in polymer extracted PH 3, but the nonpolar fibres (PDMS) extracted no PH 3. The larger size of polymer fibre (e.g., 85 µm) extracted PH 3 more efficiently than the smaller size of polymer fibre (e.g., 75 µm and 65 µm). Figure1 Fibre types (100 µm PDMS, 85 µm CAR/PDMS, 75 µm CAR/PDMS and 65 µm PDMS/DVB) affect the extraction efficacy of PH 3 after 20 minutes extraction. The effect of time of extraction of PH 3 is shown in Figure 2. The responses of GC peak areas progressively increased with increasing the extraction time, but no increase in the response occurred after 20 min. That is, the amount of PH 3 absorbed by the tested fibres increased over a period of 20 min and then attained equilibrium. After 30 min extraction, the amount of absorption was 85 µm CAR/PDMS > 75 µm CAR/PDMS > 65 µm PDMS/DVB. The 85 µm CAR/PDMS fibre was used in this study to optimize the different parameters that affect the SPME extraction efficiency of PH 3. 410 Julius-Kühn-Archiv, 425, 2010

Figure 2 Time of extraction with different type of fibres (85 µm CAR/PDMS, 75 µm CAR/PDMS and 65 µm PDMS/DVB) affects the extraction efficacy of PH 3 (0.3 ppm) at 1, 5, 10, 15, 20 and 30 min. 3.2. Effect of GC Injector temperature for desorption of PH 3 from fibre (85 µm CAR/PDMS) The effects of GC injector or desorption temperature were evaluated by varying the temperature from 100 to 250ºC. The injector temperature profile obtained is shown in Figure 3. The amount of desorbed PH 3 increased with increasing the injector temperature and then reached a maximum at 200ºC. The GC used had split/splitless capillary injectors. It was suitable for direct introduction of the fibre into the injection port. In this study, the CG injector operated at 200ºC of optimal desorption temperature and did not cause peak broadening and tailing (Fig. 4). 50000 GC readings (m Vols) 40000 30000 20000 10000 0 50 100 150 200 250 300 GC injector temperature (ºC) Figure 3 The GC injector temperature affects desorption of PH 3 from fibre (85 µm CAR/PDMS) at 0.1 ppm of PH 3 after 20 minutes extraction. Julius-Kühn-Archiv, 425, 2010 411

Figure 4 GC spectra with fibre (85 um CAR/PDMS) after 20 min extraction at levels of PH 3 at 0.5 ng/g (A) and 0.02 ng/g (B). 3.3. Effect of wheat treatment and HS-SPME extraction temperature on extraction efficacy of PH 3 The effects of temperature for treatment of wheat sample and extraction efficacy were evaluated by varying the temperature from 30 to 50ºC. The temperature profile for PH 3 released from fumigated wheat into the headspace and absorption of PH 3 on the fibre (85 µm CAR/PDMS) obtained is shown in Figure 5. 400000 350000 GC readings (m Vols) 300000 250000 200000 150000 100000 50000 0 25 30 35 40 45 50 55 Extraction temperature (ºC) Figure 5 Temperature of extraction with fibre (85 µm CAR/PDMS) for 20 minutes extraction affects the extraction efficacy of PH 3. The amount of absorbed PH 3 increased with increasing the injector temperature and then reached a maximum at 45ºC. As the temperature increases, the liquid and gas phases equilibrate faster and drive more PH 3 partitioning into the headspace, but increasing fibre temperature causes a competitive effect of 412 Julius-Kühn-Archiv, 425, 2010

desorption from the fibre, ultimately limiting its analytical sensitivity of PH 3. This is consistent with the results of Vás and Vékey (2004). Therefore, we used the low temperature (45ºC) and long term (30 min) extraction for treatment of wheat samples. The limit of detection (LOD) or level of quantification (LOQ) was estimated based on a signal to noise ratio shown in Figure 5 (B) in the range of 0.005 0.01 ng/g. Although the determination of PH 3 in grain is well established in the scientific literature, our results have demonstrated that the combination of SPME with headspace sampling is very effective for determining trace levels of PH 3. Acknowledgements The authors would like to acknowledge the support the Australian Government s Cooperative Research Centres Program and Jim Desmarchelier for helpful contributions. References Chemfate, 2004. Environmental Fate Database. Syracuse Research Corporation, Syracuse, NY. USA. accessed June 2004www.syrres.com/esc/chemfate.htm. Collins, P.J., Daglish, G.J., Pavic, H., Lambkin, T.M., Kopittke, R. 2000. Combating strong resistance to phosphine in stored grain pests in Australia. In: Wright, E.J., Banks, H.J., Highley, E. (Eds), Stored Grain in Australia. Proceedings of the Australian Postharvest Technical Conference, 1-4 August 2000, Adelaide, Australia, pp. 109-112. Collins, P.J., Emery, R.N., Wallbank, B.E. 2002. Two decades of monitoring and managing phosphine resistance in Australia. In: Credland, P.F.A., Armitage, D.M., Bell, C.H., Cogan, P.M., Highley, E. (Eds), Proceedings of the Eighth, International Working Conference on Stored-product Protection, 22-26 July 2002, York, UK, CAB International, Wallingford, UK, pp. 570-575. Desmarchelier, J. M., Allen, S. E., Ren, Y. L., 1998. Modifications of a method for determining multifumigant residues. Journal of the Association of Official Analytical Chemists, International 81, 638-644. FAO, 1975. Recommended methods for the detection and measurement of resistance of agricultural pests to pesticides. Tentative method for adults of some major pest species of stored cereals, with methyl bromide and phosphine. FAO Method NO. 16, 1975, Plant Protection Bulletin No. 23, 12-25. Vás, G., Vékey, K., 2004. Solid-phase microextraction: a powerful sample preparation tool prior to mass spectrometric analysis. Journal of Mass Spectrometry, 2004, 39, 233 254. Miyahara, M., Saito, Y., 1994. Determination of bromide ions in food by unsuppressed ion chromatography with ultraviolet detection after microwave digestion in a sealed PTHE vessel. Journal of Agricultural and Food Chemistry 42, 1126-1131. Penalver, A., Pocurull, E., Aguilar, C., Borrull, F., Marce, R. M., 2000. Comparison of different fibres in the solid phase microextraction of phthalate esters from water samples. Journal of Chromatography A 922, 377-384. Ren, Y. L., 2001. Comparison of solvent extraction and microwave extraction for release of dimethyl sulfide from cereals and canola. Journal of Agricultural and food Chemistry 49, 1737-1739. Ren, Y.L., Desmarchelier, J.M., 1998. Release of Fumigant Residue from Grain by Microwave Irradiation. Journal of the Association of Official Analytical Chemists, International 81, 673-678. Ren, Y.L., Mahon, D., 2007. Evaluation of microwave irradiation for analysis of carbonyl sulfide, carbon disulfide, cyanogen, ethyl formate, methyl bromide, sulfuryl fluoride, propylene oxide, and phosphine in hay. Journal of Agricultural and Food Chemistry 55, 32-37. Ren, Y.L., van Emmerik, T., Mahon, D., Lee, B.H., Padovan, B., 2008a. Ethyl formate plus methyl isothiocyanate is a potential liquid fumigant for stored grains. In: Guo, D.L., Navarro, S., Jian, Y., Cheng, T., Zuxun, J., Yue, L., Yang, L., Haipeng, W. (Eds), Proceedings of the 8th International Conference on Controlled Atmospheres and Fumigation in Stored Products, 21-26 September 2008, Chengdu, China, Sichuan Publishing House of Science and Technology, Chengdu, China, pp. 82-87. Ren, Y.L,, Lee, B.H., Mahon, D.A., Xin, N., Head, M., Reid, R., 2008b. Fumigation of wheat using liquid ethyl formate plus methyl isothiocyanate in 50 tonne farm bins. Journal of Economic Entomology 101, 623-630. Julius-Kühn-Archiv, 425, 2010 413