DETECTION OF SUBSURFACE DAMAGE IN OPTICAL TRANSPARENT MATERIALSS USING SHORT COHERENCE TOMOGRAPHY. Rainer Boerret, Dominik Wiedemann, Andreas Kelm

Similar documents
Application Report: Running µshape TM on a VF-20 Interferometer

Optical Coherence Tomography OCT. 3D Imaging in Medical Technology and Quality Control

ENGINEERING METROLOGY

FTIR Instrumentation

Prototyping to Production

Measuring large areas by white light interferometry at the nanopositioning and nanomeasuring machine (NPMM)

The Nurse/Technician Role Within the Emerging Ophthalmic Technology - OCTs/B-Scan

It has long been a goal to achieve higher spatial resolution in optical imaging and

Optical thickness measurement of substrates using a transmitted wavefront test at two wavelengths to average out multiple reflection errors

Metrology of silicon photovoltaic cells using coherence correlation interferometry

One example: Michelson interferometer

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

Interferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1)

Introduction to Fourier Transform Infrared Spectrometry

Modern Classical Optics

Plastic Film Texture Measurement With 3D Profilometry

4.3.5: High Temperature Test 3

Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated testing of square and round bars

Physics 441/2: Transmission Electron Microscope

Scanning Near Field Optical Microscopy: Principle, Instrumentation and Applications

Optical coherence tomography as a tool for non destructive quality control of multi-layered foils

Experiment 5. Lasers and laser mode structure

Einsatz und Nutzen der optischen Kohärenztomografie (OCT) bei technischen Bauteilen

Measuring Miniature Lens Radius of Curvature and Refractive Index with White Light Optical Profiler

Optical Metrology. Third Edition. Kjell J. Gasvik Spectra Vision AS, Trondheim, Norway JOHN WILEY & SONS, LTD

Resolution for Color photography

Components for Infrared Spectroscopy. Dispersive IR Spectroscopy

Femtosecond laser-induced silicon surface morphology in water confinement

Acoustic GHz-Microscopy: Potential, Challenges and Applications

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

A Guide to Acousto-Optic Modulators

RECOMMENDATION ITU-R SM Measuring sideband emissions of T-DAB and DVB-T transmitters for monitoring purposes

Introduction to Optics

Fiber Optics: Engineering from Global to Nanometer Dimensions

Today. next two weeks

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications

Fig.1. The DAWN spacecraft

3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY

Nanoscale Resolution Options for Optical Localization Techniques. C. Boit TU Berlin Chair of Semiconductor Devices

Non-Contact Vibration Measurement of Micro-Structures

CT Traceability - Prof. Wim Dewulf, Group T - KU Leuven

DIODE LASER BASED PHOTOACOUSTIC SYSTEM FOR ATMOSPHERIC WATER VAPOR MEASUREMENTS

ING LA PALMA TECHNICAL NOTE No Investigation of Low Fringing Detectors on the ISIS Spectrograph.

Near-field scanning optical microscopy (SNOM)

PROVE, the next generation registration metrology tool, status report

Sensor/Raft Metrology Status

APPLICATION OF X-RAY COMPUTED TOMOGRAPHY IN SILICON SOLAR CELLS

Self-Mixing Laser Diode Vibrometer with Wide Dynamic Range

PCB Component Placement Inspection

Agilent Cary 60 UV-Vis

ZEISS T-SCAN Automated / COMET Automated 3D Digitizing - Laserscanning / Fringe Projection Automated solutions for efficient 3D data capture

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F)

Accuracy of SpotOptics wavefront sensors. June 2010 Version 4.0

A Measurement of 3-D Water Velocity Components During ROV Tether Simulations in a Test Tank Using Hydroacoustic Doppler Velocimeter

Raman Spectroscopy Basics

Optical Communications

Proposed experiment to test the non-locality hypothesis in transient light-interference phenomena

Lab 9: The Acousto-Optic Effect

Sensori ottici e laser nelle applicazioni industriali

Nederland België / Belgique

MACHINE VISION FOR SMARTPHONES. Essential machine vision camera requirements to fulfill the needs of our society

Prospects of Laser Polishing for Small and Complexly Shaped Parts

Alignement of a ring cavity laser

Realization of a UV fisheye hyperspectral camera

RAY TRACING UNIFIED FIELD TRACING

Defect studies of optical materials using near-field scanning optical microscopy and spectroscopy

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale

Self-Mixing Differential Laser Vibrometer

Optical Methods of Surface Measurement

An Introduction to the MTG-IRS Mission

The coherence length of black-body radiation

Laser-induced surface phonons and their excitation of nanostructures

Interference. Physics 102 Workshop #3. General Instructions

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

INFITEC - A NEW STEREOSCOPIC VISUALISATION TOOL BY WAVELENGTH MULTIPLEX IMAGING

Application Note #503 Comparing 3D Optical Microscopy Techniques for Metrology Applications

Scanning Acoustic Microscopy Training

GRID AND PRISM SPECTROMETERS

Master Degree Program

ME 472 Engineering Metrology

5-Axis Test-Piece Influence of Machining Position

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

Apertureless Near-Field Optical Microscopy

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Flexible Measurements of Shapes and Roughness of Aspheres and Optical Freeforms

Module 13 : Measurements on Fiber Optic Systems

DNA NANOWIRES USING NANOPARTICLES ECG653 Project Report submitted by GOPI

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Metrology for Characterization of Wafer Thickness Uniformity During 3D-IC Processing

Scanning Near-Field Optical Microscopy for Measuring Materials Properties at the Nanoscale

Efficiency, Dispersion and Straylight Performance Tests of Immersed Gratings for High Resolution Spectroscopy in the Near Infra-red

NEAR FIELD OPTICAL MICROSCOPY AND SPECTROSCOPY WITH STM AND AFM PROBES

The Basics of Scanning Electron Microscopy

Holography 1 HOLOGRAPHY

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm

Friday 18 January 2013 Morning

Scanning Surface Inspection System with Defect-review SEM and Analysis System Solutions

Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center

Transcription:

URN (Paper): urn:nbn:de:gbv:ilm1-2014iwk-199:0 58 th ILMENAU SCIENTIFIC COLLOQUIUM Technische Universität Ilmenau, 08 12 September 2014 URN: urn:nbn:de:gbv:ilm1-2014iwk:3 DETECTION OF SUBSURFACE DAMAGE IN OPTICAL TRANSPARENT MATERIALSS USING SHORT COHERENCE TOMOGRAPHY Rainer Boerret, Dominik Wiedemann, Andreas Kelm HTW Aalen, Anton-Huber Str. 21, 73430 Aalen, Germany ABSTRACT This paper describes the use of short coherence interferometry for the detection of subsurface damage (SSD) in optical transparent materials. The method is based on an interferometer using a light source with a short coherence length. A time domain (TD) and a frequency domain (FD) approach are explained. The OCT method allows a nondestructive measurement of the SSD compared to the state of the art SSD metrology, where the sample is destroyed. Compared to previous results [Ser10] a frequency domain (FD) setup is build up which allows SSD measurements under production environment conditions. The results of the FD set are presented and discussed. 1. INTRODUCTION The generation of optical surfaces is often done by a grinding process. In brittle materials like glass, the grinding process can introduce cracks, spreading in the brittle material below the surface. These are called subsurface damage (SSD). In Fig. 1 the process is sketched. Image 1 shows the unprocessed blank and image 2 the surface after grinding that contains small cracks spreading in the material. The depth of these cracks is related to the processing tools and parameters and determines the amount of material, which have to be removed during the subsequent polishing process. Fig. 1 Generation of an optical surface 1. Unprocessed blank, 2. Ground blank, 3. Cracks in surface (schematic) [Wi11] To remove the material layer on the optical surface containing SSD the depth of these cracks needs to be detected. There are several destructive methods available that do a partial polishing of sample parts followed by an inspection of the partly polished surfaces. Another method uses 2014 - TU Ilmenau

corrosive liquids like the haloalkanes to etch down the surface stepwise, inspecting it afterwards. Due to the use of these chemical liquids, the operator itself has to be very careful in handling them. A non-destructive approach is the use of short coherence interferometry. This technique is known as Optical Coherence Tomography (OCT) in clinical imaging [Bre05, chapter 12] [Dre08]. In this paper two different technologies for OCT are described: the time domain [Ser10] and the frequency domain approach. 2. TIME DOMAIN (TD) TECHNOLOGY The TD technology is using a Michelson interferometer as sketched in Fig. 2. For the TD measurement the light source (1), shown in Fig. 2, consists of a source with short spatial coherence, e.g. a superluminescence diode (SLD). The beam splitter (2) divides the light between the reference mirror (3) and the device under test (4). The reflected light from (3) and (4) interferes on the detector (5) when the optical path length of both arms LL RR and LL TT in Fig. 2 are equal. As long as the material is optical transparent the point of reflection (4) can lay inside the material. Fig. 2 Michelson interferometer, 1. Light source, 2. Beam splitter, 3. Reference mirror, 4. Device under test, 5. Detector [Wi11] Changing LL RR by moving (3) in a controlled way the detector signal from (5) can be assigned to a known position of (3) and therefore providing a depth scan. This provides information of different layers of reflective structures on a single point of the surface. 2014 - TU Ilmenau 2

3. FREQUENCY DOMAIN (FD) TECHNOLOGY As for the TD technology the FD technology is based on an interferometer. For the measurement the optical path lengths LL RR and LL TT are a little differing. The setup is sketched in Fig. 3. The point of equal optical path length is equal the focus of the lens (1). The reflected light from (5) and (6) interferes on the detector. The detector consists of a spectrometer that is able to detect the amount of interference in relation to the wavelength. Fig. 3 FD setup; 1. Focus of lens, 2. Distance to surface, 3. Distance to SSD, 4. Interferometer, 5. Surface of lens, 6. Crack inside material. [Wi11] Depending on the position of the reflective structures (2) and (3) the spectrum of the light source is modulated with a distance-dependent frequency onto the light source spectrum. The depth s information can be immediately calculated by a Fourier-transformation from the acquired spectrum, without movement of the reference arm [Sch99]. 4. MEASUREMENT To practically evaluate the accuracy of both technologies a reference measurement of a known structure was done. The type of structure is shown by Fig. 4 and consists of a rectangular shape with a spacing of 40µm and a depth of 5 µm. The measurement was obtained using a Zygo NewView microinterferometer. Fig. 4 structures on reference waver with 40µm width [Wi11] 2014 - TU Ilmenau 3

The measurement results of the TD technology is shown in Fig. 5, top, and of the FD technology in Fig. 5, bottom. Fig. 5 measurement of reference structure with 40µm width and 5µm height, top: TD measurement, bottom: FD measurement [Wi11] As shown in Fig. 5, the Reference Structure can be reproduced. Due to the evaluation process of both methods, the results of the Time Domain technology contain analog noise, which can be clearly seen in the results. The Frequency Domain technology on the other hand only has digital noise, which originates from the resolution of the spectrometer itself. 4.1 Axial resolution To determine the axial resolution a flat reference made of silicon with a surface irregularity of 50,6nm measured from lowest to highest point (peak to valley, pv). A measurement of that surface is shown in Fig. 6, left side, using the Zygo NewView microinterferometer. The slice in Fig. 6 right side, demonstrates the high quality of that surface with about 2nm pv of irregularities. Fig. 6 flat reference surface to measure the axial resolution 2014 - TU Ilmenau 4

Using this reference flat a measurement was done with the TD as well as the FD technology. A slice with a length of about 100µm was measured. The result of the TD technology can be seen in the upper part of Fig. 7. The graph shows a deviation of about 1.51µm pv. Because the quality of the reference surface is much higher it can be assumed that this shows the analog noise present in the TD technology. The same amount of noise can be seen in the measurement of the reference structure in the upper part of Fig. 5. Fig. 7 Fig. 5 measurement of flat surface of 50,6nm pv, top: TD measurement, bottom: FD measurement [Wi11] The measurement with the FD technology is shown in the lower part of Fig. 7. It shows a different structure with a deviation of about 0.74µm pv originating from the resolution of spectrometer. The same structure is present in the lower part of Fig. 5. Both measurements show the limitation of the axial resolution of the technology which lies in the range of 1µm for the FD and about 2µm for the TD technology. 4.2 Measurement of SSD To verify the measurement of SSD using the OCT a reproducible reference is needed. Therefore a ground surface was used that was partly polished with a wedge. Fig. 8, right side, shows a tactile measurement with the UPMC 550 CARAT from Zeiss. The right part of the surface shows a polished wedge with a maximum depth of about 60µm on the right side. The left side of the surface up to the center is not polished. The tactile measurement shows an almost flat surface in that region. The left side of Fig. 8 shows an OCT measurement using the TD technology. The wedge starting from the center of the surface to the right is clearly visible and similar to the tactile measurement. In contrast to that the right side of the OCT measurement shows a lot more variation with an amplitude of about 30µm pv. In addition to the roughness of the ground surface spikes can be seen ranging from 0µm down to -18µm on the y-axis marked by the yellow area. This spikes are SSD that is detected by the OCT. 2014 - TU Ilmenau 5

Fig. 8 ground surface, partly polished for detection of SSD, left: measurement with TD technology, right: tactile measurement In addition to the OCT measurement the SSD were analyzed optically on the wedge of the polished part. The measurement results of both technologies correlate with each other. Fig. 9 shows the FD OCT measurement of a flat ground optic consisting of zerodur. The measurement shows, that SSD can be seen within the results, having a depth of approximately 16µm. Fig. 9 measurement of ground surface To show the capabilities of the FD OCT on rough surfaces a measurement of sawed raw flat glass optic prior to the grinding process was done. Fig. 10 shows a measurement of such a BK7 lens, which has deviation of 38.3µm pv. Fig. 10 measurement of a sawn surface 2014 - TU Ilmenau 6

The sharp spikes down to below -20µm on the y-axis again are SSD e.g. at x=-0.1mm. This measurement shows that SSD can be detected even on raw surfaces. 5. DISCUSSION The results of the OCT measurement of a flat reference surface show a limited axial resolution of about 2µm for the TD technology and 1µm for the FD technology. Both technologies have shown the ability to detect SSD in optical transparent materials. The results of both technologies correlate with each other, the correlation of the TD technology with results obtained by different methods is shown in [Ser10]. Therefore both technologies, TD and FD, can be used to detect SSD. 6. CONCLUSION For the measurement of subsurface damage in brittle materials the nondestructive technique of Optical coherence Tomography was investigated. Two different techniques where examined and found to deliver comparable results that correlate with existing measurements. The measurement of polished as well as raw surfaces is possible enabling the control of SSD even on rough surfaces. REFERENCES [Bre05] Brezinski, M., Optical Coherence Tomography - Principles and Applications, Academic Press, 2005. [Dre08] Drexler, W., Fujimoto, J. G., Optical Coherence Tomopgraphy, - Technology and Applications, Springer Verlag, 2008. [Sch99]Schmitt, J, "Optical coherence tomography (OCT): a review". IEEE Journal of Selected Topics in Quantum Electronics 5 (4): 1205.,", 1999. [Ser10] M. Sergeeva, K. Khrenikov, T. Hellmuth, R. Boerret, "Sub surface damage measurements based on short coherent interferometry", J. Europ. Opt. Soc. Rap. Public. 10003 Vol 5, 2010. [Wi11] Wiedemann, Dominik. Development and test of a frequency domain OCT for subsurface damage measurement, Master thesis, HTW Aalen, 2011. CONTACTS Prof. Dr. Rainer Börret Rainer.Boerret@htw-aalen.de 2014 - TU Ilmenau 7