Automated Calibration Software Guardbanding. Author: Matt Nicholas Fluke Corporation. Speaker: Lars Andersson Fluke Europe B.V



Similar documents
Guidelines on the Reporting of Compliance with Specification

CALIBRATION DATA MANAGEMENT: MEETING THE REPORTING REQUIREMENTS OF ISO/IEC FDIS

MET/TEAM Test Equipment Asset Management Software

calibrate confidence with Risk Management: The Critical Role of Measurement and Calibration WHITE PAPER SERVICES

How to calibrate an RTD or Platinum Resistance Thermometer (PRT)

Metrology, Quality and ISO 9000

Field Calibration Software

calibrate confidence with Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Abstract Introduction White Paper

R214 SPECIFIC REQUIREMENTS: INFORMATION TECHNOLOGY TESTING LABORATORY ACCREDITATION PROGRAM

Software Verification and Validation

CONTROL CHARTING LABORATORY REFERENCES FOR QUALITY CONTROL

Measurement Information Model

NATO STANDARD AQAP-2310 NATO QUALITY MANAGEMENT SYSTEM REQUIREMENTS FOR AVIATION, SPACE AND DEFENCE SUPPLIERS

CALIBRATION PRINCIPLES

Concept for an Algorithm Testing and Evaluation Program at NIST

Regulatory Requirements for Medical Device Calibration Programs

Number of hours in the semester L Ex. Lab. Projects SEMESTER I 1. Economy Philosophy Mathematical Analysis Exam

Machine Tool Inspection & Analyzer solutions

Attributes Acceptance Sampling Understanding How it Works

Migrating from dc voltage dividers to modern reference multimeters

Measuring equipment in a laboratory.

ISO Information Technology Service Management Systems Professional

IMQ RULES. Contents. IMQ Rules - Product certification

ISO/IEC QUALITY MANUAL

Administrator s Guide ALMComplete Support Ticket Manager

TOTAL QUALITY MANAGEMENT II QUALITY AUDIT

FSI Machine Vision Training Programs

ISMS Implementation Guide

Comparison between FDA QSR and ISO 13485

OPERATIONAL STANDARD

ISO 9001:2008 Audit Checklist

Quality Management. Lecture 12 Software quality management

COMBINE. Part B. Manual for Marine Monitoring in the. Programme of HELCOM. General guidelines on quality assurance for monitoring in the Baltic Sea

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Agilent Measuring Noninsertable Devices

5320A Multifunction Electrical Tester Calibrator. Verify and calibrate electrical test tools with a single instrument

Quality management systems

Precision Manufacturing Program

Peer Review Process Description

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x

ALACC Frequently Asked Questions (FAQs)

Calibration Control. Calibration Management Software. Tools for Management Systems

JOURNAL OF MEDICAL INFORMATICS & TECHNOLOGIES Vol. 21/2012, ISSN

This procedure applies to the following categories of documents:

1 For more information T: / E: DAP@ul.com / W: ul.com/dap

International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research)

Peer Review Process Description

NATO QUALITY ASSURANCE REQUIREMENTS FOR DESIGN, DEVELOPMENT AND PRODUCTION

Make Better Decisions with Optimization

Multiple Optimization Using the JMP Statistical Software Kodak Research Conference May 9, 2005

EPEAT Program Guidelines

ORACLE CONSULTING GROUP

Welcome! Frank Teleso. Quality Manager

IFS Food Safety and Quality Management System

The PaperCept Journal Submission and Review Management System

How to Verify Performance Specifications

Calibration of Electronic Test Equipment: An Introduction to ISO & Accreditation

ISO 9000 Quality Standard Background Information Quality Systems Implementation

GS1 bar code validation and verification

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

EXAM PREPARATION GUIDE

Taking the pain out of Risk and Compliance Management Systems. Presented by Andrew Batten 23 April 2015

Unit 6: INTRODUCTION TO QUALITY ASSURANCE and TOTAL QUALITY MANAGEMENT (key-words: pre-fabrication, site assembly, integrated systems)

The Challenge of Productivity Measurement

ENVIRONMENTAL MANAGEMENT SYSTEM PLANNING GUIDANCE UNDER THE TOXICS USE REDUCTION ACT (TURA)

4Sight Calibration Management Software

Anwendung von Polyspace im Software Entwicklungsprozess nach IEC München, , Dr.-Ing. Jörg Barrho

B i o s o l u t i o n s

Supplier Quality Agreements

Table of Contents. Preface 1.0 Introduction 2.0 Scope 3.0 Purpose 4.0 Rationale 5.0 References 6.0 Definitions

Software Metrics & Software Metrology. Alain Abran. Chapter 4 Quantification and Measurement are Not the Same!

How To Check For Differences In The One Way Anova

control manual Optimal maintenance decisions for asset managers by andrew k.s. jardine and neil montgomery 44 Industrial Engineer

CHAPTER 461b. SLOT MACHINE TOWER LIGHTS AND ERROR CONDITIONS TECHNICAL STANDARD. 461b.2. Slot machine tower lights and error conditions.

Quality Management System Policy Manual

Accreditation and Competence?

FMC Technologies Measurement Solutions Inc.

PROJECT QUALITY MANAGEMENT

Measurement Systems Correlation MSC for Suppliers

Chapter 14 Managing Operational Risks with Bayesian Networks

Self Testing and Product Qualification Processes

Implementing ISO 9000 Quality Management System

Detecting Fraud in Health Insurance Data: Learning to Model Incomplete Benford s Law Distributions

Pick a Palette and Beyond: REPORT GENERATION with LABVIEW

EXAM PREPARATION GUIDE

Quality Assurance Manual January 2008 Revision I Revision date of 01/17/13. Pi Tape Corporation. 344 North Vinewood St. Escondido CA USA

How DCMA Helps To Ensure Good Measurements

STEEL fabrication quality SySTEmS guideline

DEC STD ISO Quality Systems - Model for Quality Assurance in Production and Installation

The Standard Normal distribution

Raad voor Accreditatie (Dutch Accreditation Council RvA) Assessment of Conformity Assessment Schemes

EffiValidation 3.0 software basic training module

Check-list for auditors. Audit visit at candidate registered laboratories

How To Test Your Web Site On Wapt On A Pc Or Mac Or Mac (Or Mac) On A Mac Or Ipad Or Ipa (Or Ipa) On Pc Or Ipam (Or Pc Or Pc) On An Ip

NETWORK DESIGN & ANALYSIS PTIIK

ISO 9001:2008 QUALITY MANUAL. Revision B

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

CPA SECURITY CHARACTERISTIC DATA SANITISATION - FLASH BASED STORAGE

Transcription:

Automated Calibration Software Guardbanding Author: Matt Nicholas Fluke Corporation Speaker: Lars Andersson Fluke Europe B.V 1 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

ABSTRACT The ISO 17025 standard ( General Requirements for the Competence of Testing and Calibration Laboratories ) requires that measurement uncertainty be taken into account when statements of compliance are made. In other words, when a calibration laboratory calibrates an instrument and produces a calibration certificate indicating that the calibration verification procedure passed or failed, it is important that, for each test point, the measurement uncertainty be first calculated and then used in the determination of the test result. 2 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

ABSTRACT (continued) Guardbanding is a primary technique for assuring compliance with this 17025 requirement. Considerations of efficiency and productivity in calibration laboratories require full automation whenever possible. It is therefore desirable for automated calibration software to include guardbanding capability. This paper describes a flexible, configurable implementation of guardbanding in an automated calibration software system. A number of techniques are discussed, including both table-based and formula-based methods. Facilities for customization of the guardbanding algorithm are presented. A description of generated result data is included. 3 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Introduction Overview Motivation for Guarbanding Guardbanding Strategies Automation of Guardbanding Comments & Questions 4 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

What is Guardbanding? The fundamental concept is to restrict the pass/fail limits applied to a calibration test based on some criterion. The purpose of the restriction is to control the risk of accepting an out-of-tolerance unit, or rejecting an in-tolerance unit. Example: Test Point = 10 V Test Specification = 1% Test Limits are 9.9 V and 10.1 V Measurement Uncertainty = 0.03 V UUT Reading = 9.92 V The reading is within the test limits, but if the uncertainty is considered, the true value is 9.92 +/- 0.03 V. The true value may be as low as 9.89 V, which is below the lower test limit. 5 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Guardbanding Example Measurement Uncertainty = 0.03 Indeterminate regions based on 100% of measurement uncertainty. Indeterminate Indeterminate +9.90 +10.00 +10.10 9.89 9.95 9.92 (Measurement) 6 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Consumer Risk and Producer Risk as a function of TUR 7 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Guardbanding Motivation ISO 17025 Section 5.10.4.2 When statements of compliance are made, the uncertainty of the measurement shall be taken into account. The current statement is open to some debate. If one simply reports the measurement uncertainty, leaving it to the customer s discretion to interpret the value, has the uncertainty been taken into account? Other Standards An organization may be subject to or claim compliance with standards other than ISO/IEC 17025: * ISO 14253-1 Geometrical Product Specifications (GPS) -- Inspection by measurement of workpieces and measuring equipment -- Part 1: Decision rules for proving conformance or non-conformance with specifications. * ILAC-G8:1996 Guidelines on Assessment and Reporting of Compliance with Specification An organization may apply internal quality control standards which required guardbanding, or where guardbanding assists in meeting quality criteria. Guardbanding affects the result of a calibration test! Setting different test limits influences the risk of accepting defective units or rejecting conforming units 8 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Automated Software - Guardbanding Strategies Strategies range from conservative to not-so-conservative. Most conservative is to tighten the specification limits by the full amount of the measurement uncertainty. Must be able to calculate the uncertainty. Calibration software which calculates uncertainty is a pre-requisite for full automation. Least conservative is to do nothing. Specification limits are used as is. This is traditional, but takes no account of the measurement uncertainty. MIL STD 45662A, which requires a 4:1 TSR, follows this approach. * the point was found not to be out of tolerance rather than within specified tolerances. This takes into account that any parameters which are accepted under this criteria which are actually out of tolerance, are not likely to be very far out of tolerance. * If reference specification are 3σ there s a 0.06% consumer risk * TSRs between 1.5:1 and 4:1, parameters with measured values which are less than 80% of the specification limits may be declared in tolerance 9 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Guardbanding Strategies (continued) What are the input parameters? Measurement uncertainty Test uncertainty ratio (TUR) Test specification ratio (TSR) Acceptable risk. Historical data for references and UUT * Requires access to database containing extended calibration data. * Access to historical data may be impractical for 3rd-party cal labs. Population data for references and UUT * It may be difficult to obtain data, except perhaps for manufacturer. * Population data for the UUT may be used to evaluate the likely success of a particular guardbanding strategy. 10 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Manual Guardbanding Automated calibration software is not used. Guardbanded test limits must be calculated prior to the calibration. Software packages like Excel may be used to assist in the calculation of the guardbanded limits. The calibration procedure is performed manually, and results are recorded on paper. Operator to determine each test to be Pass, Fail or Indeterminate 11 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Semi-Automated Guardbanding Adjust Test Tolerances Example: * Change DMM 10mV 1% to DMM 10mV 0.8% Once written, procedure may be re-used. May be possible to use software facilities not explicitly designed for guardbanding: Marginal Pass indication. Significantly Out of Tolerance indication. Requires significant effort to update procedure Static, e.g. guardbanding factor difficult to vary based upon expanded measurement uncertainty Reporting system may not be able to easily distinguish between a cal performed with or without guardbanding 12 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Software Requirements An ideal automated guardbanding solution would: Automate the calculation of the guardbanded test limits. Support a number of canned methods. Allow the procedure writer to enable or disable guardbanding. Save result data to satisfy an auditor s requirements. Require little extra work on the part of the procedure writer. Flexibility Customizable -- configurable on a per-test, per-procedure, perworkstation, or per-site basis. It should be possible to override built-in defaults for all parameterized values. Compatibility It should be possible to use existing automated calibration procedures with little or no modification. 13 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

TSR-Based Methods Method 1: Recall that TSR = (Test Specification) / (Reference Specification) * Nomenclature has been inconsistent. Often called TAR, sometimes TUR. If TSR between 1.5 and 4, use 80% of specification limits If TSR > 4, use 100% of specification limits (i.e., no guardbanding) If the TSR is less than 1.5, the validity of the test may be called into question. One may, however, wish to define a guardbanding factor (less than 80%) for such tests. This method is easy to describe, and relatively easy to implement. * Does not require that the software have the ability to compute the measurement uncertainty 14 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

TSR-Based Methods (continued) The first generalization is to parameterize this method and make all parameters software-settable: Make the TSR thresholds (4 and 1.5) settable by the procedure writer. Make the guardbanding factor (80%) settable by the procedure writer. 15 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

TSR-Based Methods (continued) Further generalize to allow a user-specified table of TSR ranges with a corresponding guardbanding factor for each range. Supports an arbitrary number of points in the table. Provide a software-settable option to either: Linearly interpolate between the points in the table, or Interpret the table as a step function, choosing the more conservative guardband factor for values which fall between points in the table. 16 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

TUR-Based Strategies Method 2: Same flexibility as TSR-based, but uses the TUR rather than the TSR. TUR = (Test Specification) / (Measurement Uncertainty) Software must be able to calculate the measurement uncertainty in order to use the TUR as an input parameter to the guardbanding algorithm. 17 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Normalized TUR-Based Strategies Method 3: Same as TUR-Based, but normalize the numerator and the denominator to 1 sigma: Normalized TUR = (1-sigma test specification) / (standard uncertainty) The software package must allow the procedure writer to specify the confidence associated with the test specification. Default confidence value is 2 sigma. 18 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Uncertainty-Based Strategies Method 4: Use the measurement uncertainty directly: Guardbanded Upper Limit = Specification Upper Limit - (GBF * Measurement Uncertainty) Guardbanded Lower Limit = Specification Lower Limit + (GBF * Measurement Uncertainty) where GBF is the guardbanding factor, specifiable by the procedure writer. If GBF is set to 1, this implements the Guide 25 Draft 5 recommendation. Setting GBF to a value less than 1 requires justification by a metrologist. 19 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Result Data Specification Test Limits Guardbanding Flag Guardbanding Method Guardbanded Test Limits Guardband Factor Overall Result for Each Test Pass Fail Indeterminate Overall Result of Calibration Procedure 20 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Calibration Procedure Result What is the overall result of the procedure (pass, fail, or indeterminate)? Configurable: Method A: Indeterminate equivalent to Pass Method B: Indeterminate equivalent to Fail Method C: Indeterminate remains Indeterminate Example Suppose 10 tests, with 6 PASS and 4 INDETERMINATE Method A: Overall result is PASS Method B: Overall result is FAIL Method C: Overall result is INDETERMINATE 21 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Software Review Guardbanding Control Enable or disable guardbanding. Specify the guardbanding method. Specify a table of parameter thresholds with associated guardband factors. Choose whether or not to interpolate table values. Supports asymmetrical tolerance specifications. Specify how indeterminate test results are handled. Call external programs to provide data on GBFs & other parameters. Change guardbanding parameters on a per-site, per-workstation, per-procedure, or per-test basis. Report results in a SQL database, for easy access by report generators or other software tools. 22 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Conclusion Guardbanding now implemented in MET/CAL Built-In Calculations & Default Values (adequate in Most Cases) * Use Existing Procedures without Modification Flexible Implementation * Support for Different Guardbanding Strategies * Customizable by the User Opportunities for improvements * The table-based method described here, where the independent parameter is determined by the user, and an associated guardband factor is specified for each row in the table, could be made more general by allowing a mathematical expression to determine the guardband factor on each row of the table. * More work should be done on guardbanding methods that directly maintain a constant risk of accepting an out-oftolerance UUT. RESULT Pass Indeterminate Pass Indeterminate Fail Fail RSLT & Brace Text COLOUR Green Light Green Light Red Red Cyan 23 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004

Any Questions? 24 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004