Cryogenic Performance of Millitech Model FBI-22 Waveguide Isolator.



Similar documents
MEASUREMENT OF CRYOGENIC PERFORMANCE OF 4-8 GHz PAMTECH ISOLATORS S/N Juan Daniel Gallego Isaac López Fernández Carmen Diez González

Informe Técnico IT - CDT

General Information. Lower Frequency Models Custom Designs Special Flanges Narrow Band Specials VSWR. Frequency ( GHz) COMMON SPECIFICATIONS

ELECTRICAL CHARACTERISATION OF SMA CONNECTORS FOR CRYOGENIC AMPLIFIERS. Carmen Diez Juan Daniel Gallego Isaac López Rafael García.

PHASE SHIFT DUPLEXERS

Orthomode Transducer at 43GHz

Measurement & Modeling Device Characterization Solutions

Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows

W1 Connector W1 Adapter W1 Termination

ELECTRICAL CHARACTERISATION OF SEMI-RIGID COAXIAL CABLES WITH SMA AND K CONNECTORS. Carmen Diez Rafael García Juan Daniel Gallego.

S-PARAMETER MEASUREMENTS OF MEMS SWITCHES

Thermal Storage Unit Using the Triple Point of Hydrogen

February 2010 Number Vector Network Analyzer upgrade to SureCAL Power Sensor and RF Components Packages

Improving Temperature Measurement to Optimize Inventory Control and Custody Transfer Systems

Agilent Measuring Noninsertable Devices

The infrared photon trap F. Caspers

FIVE PROJECTS OF HIGH RF POWER INPUT COUPLERS & WINDOWS FOR SRF ACCELERATORS*

LOW GAIN BASE STATION ANTENNA AV1312-1A

"FP", "FR", "FQ" Series Bandpass Filters

5 Waveguide to Coaxial Adapters

R&S ZCxxx Millimeter-Wave Converters Specifications

Diaphragm Seal with flush diaphragm and flange acc.to SMS. Completed with pressure gauge in stainless steel, liquid filled or pressure transmitter

THE MMT MEGACAM Focal Plane Design and Performance

THE ALIGNMENT STRATEGY FOR THE SPIRAL2 SUPERCONDUCTING LINAC CRYOMODULES

Agilent X/P/K281C Adapters

1. Straight Sections

The Fundamentals of Thermoelectrics

MAURY. VNA Calibration Kits, Microwave Components & Adapters. Calibrate With Confidence IN THIS CATALOG:

Technical Specification for the Cryogenic Testing of HTS Current Leads

BB-18 Black Body High Vacuum System Technical Description

MMIC packaging. 1. Introduction 2. Data interface. Data submittal methods. Data formats. Single chip & MCM solutions. Contents

R&S ZVA-Z75, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500 Converters Quick Start Guide

RESOLUTION CHARTS AND GRATINGS RESOLUTION CHARTS AND GRATINGS RESOLUTION CHARTS AND GRATINGS RESOLUTION CHARTS AND GRATINGS RESOLUTION CHARTS AND

Using Thermocouple Sensors Connecting Grounded and Floating Thermocouples

MGO Insulated Thermocouples Index

Temperature. LumaSense Technologies, Inc. supplies a unique selection of very precise calibration

75 Ω Transmission System

2. The Vector Network Analyzer

T7079A,B Solid State Remote Temperature Controllers

Low noise millimeter wave receivers for Cosmic Microwave Background radiometers

Herschel PACS Test Procedure STM PACS-KT-PR-011 Issue: 1 Revision: -

Introduction to vacuum gauges. Vacuum Gauges where the Pressure Readings are Independent of the Type of Gas (Mechanical Vacuum Gauges)

Complimentary Reference Material

Shielding Effectiveness Test Method. Harbour s LL, SB, and SS Coaxial Cables. Designs for Improved Shielding Effectiveness

Technical Data. Dimensions

Digital Magnetometer Systems - DM Series

AUTOMATIC NETWORK ANALYZER PROCEDURES FOR 5045 KLYSTRON CAVITIES * J.G. Judkins

Solar Hot Water Heaters

JITAI Technology Co., LTD. Your 1 st Partner. Product Catalog. Microwave Passive Device Design and Manufacture

V281C/D and W281C/D 1.0 mm Coaxial to Waveguide Adapters. Operating and Service Manual

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

MECHANICAL AND THERMAL ANALYSES OF THE CABLE/ STRAND STRAIN TEST FIXTURE

Turbine Flow Sensor Series VTR

ELECTRONIC THERMOSTAT AND THERMOMETER With SPEED CONTROL

Microwave Components. J & N Microwave and Engineering. Product Catalogue 2010 Version 2.0

WAVEGUIDE-COAXIAL LINE TRANSITIONS

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control Solutions. Benefits.

The Design & Test of Broadband Launches up to 50 GHz on Thin & Thick Substrates

Between Series Adapters

Penn Engineering PENN. Quality Microwave Components ENGINEERING. ~since 1971~ (661) Avenue Penn Valencia, California

The multibeam receiver for SRT, the Sardinia Radio Telescope

Agilent 8753ET/8753ES Network Analyzers

3.3 Calibration standards

Brookhaven Magnet Division - Nuclear Physics Program Support Activities

MEASUREMENT OF COMPLEX PERMITTIVITY OF LIQUIDS USING WAVEGUIDE TECHNIQUES

Applying Error Correction to Network Analyzer Measurements. Application Note Table of Contents. Page

ENAiKOON ultrasonic fuel-level sensor

Reprinted from Winter2010 / lngindustry /

Infrared Thermometry. Introduction, History, and Applications. Optical Pyrometry. Jason Mershon, Advanced Energy Industries, Inc

DEPARTMENT OF DEFENSE TEST METHOD STANDARD METHOD OF INSERTION LOSS MEASUREMENT

Index mm Connector Series mm Connectors (50 GHz)... Page. Interface Standards Specifications Connector Numbers...

EVAC ONLINEVAC TECHNICAL DATA EVAC ONLINEVAC 16, 24 V DC. Required maintenance space 700 mm. Required maintenance space 200 mm 360 Ø13

AMSC s Superconductor Cable Technologies for Electric Utilities

MILLIMETER-WAVE DEVICES

Semi-rigid Coaxial Cables

WATERBATHS AND OILBATHS

EMC Test Chamber Solutions RF Cable Assemblies and Connectors

Times Microwave Systems Hermetically Sealed Assemblies

Chalmers Publication Library

RF SYSTEM FOR VEPP-5 DAMPING RING

Comparison of Vector Network Analyzer and TDA Systems IConnect Generated S-Parameters

James M. Pleasants Company

SMT POWER THERMISTORS

CONTENTS. Elmika warrants each product of its manufacture to be free from defects in material and workmanship.

UV Disinfection Systems A Series. Main Applications: Drinking Water, Process Water, Warm Water / Legionellae. IT T Industries Engineered for life

Thermometer in stainless steel with capillary and switch contacts

Experiment 7: Familiarization with the Network Analyzer

Michael Hiebel. Fundamentals of Vector Network Analysis

R&S SMZ Frequency Multiplier Specifications

A New Concept of PTP Vector Network Analyzer

COIL INPUT SCREENS. ITEM NUMBER: Always use a 1 through? In this field you might also use a dash and place quantity of coils here.

Solid Core and Split Core Adjustable Current Status Switches CSS-O, CSS-C; CSP-O, CSP-C

User s and Service Guide

High Power Tests of Beryllium Oxide Windows to the Lower Hybrid Current Drive Launcher in JET

Software Utility VNA Frequency Converter Leveling Tool Getting Started

REGULATORS 175/153 Series

Transcription:

THE AUSTRALIA TELESCOPE NATIONAL FACILITY AT Technical Document Series AT/39.3/127 May 2, 2006 Cryogenic Performance of Millitech Model FBI-22 Waveguide Isolator. Russell Bolton Russell Gough Two model FBI-22 waveguide isolators, serial numbers 574 and 575, were supplied by the manufacturer for testing at cryogenic temperatures. This report describes tests performed on the isolators. Each isolator was installed in a cryostat with lengths of WR-22 waveguide providing input and output connection to room temperature Agilent, model Q281A, waveguide to coaxial adapters. Fig. 1 is a photograph of the test set-up. The waveguides in the dewar use custom 19.1 mm diameter flanges. Adapters are used to connect the 19.1 mm diameter flanges to standard WR-22 flanges. 50 mm long, thin-walled stainless steel, waveguides are connected to both flanges of the isolator to provide thermal isolation and copper straps are used to connect the cold ends of the stainless steel waveguides to the cold plate. Waveguide vacuum windows were installed in the ends of the sliding waveguide sections so the dewar could be evacuated prior to cooling. The operating temperature of the isolator is measured using a temperature sensor on the isolator input waveguide. When the cold mounting plate within the dewar is cooled to 12 Kelvin, the isolator temperature reaches 19 Kelvin. S-parameters of the isolator were measured using an Anritsu model 37377D vector network analyzer. The network analyzer was calibrated, at the flanges where the isolator is connected, using the LRL/LRM method. The waveguide calibration standards used were a through connection, a 3.039 mm length of WR-22 waveguide and a short circuit. Only one calibration was performed, but the calibration was checked after the series of tests. The calibration of S 21 was found to have drifted about 0.08 db higher mid-band and 0.2 db higher at the band edges over the 7 day duration of the tests. Note that when the isolator is cooled, the waveguide sections operate at a lower temperature than when the calibration was performed. This effect is mainly associated with the stainless steel waveguide sections. When the waveguides are cooled, their loss decreases by about 0.1 db at 30 GHz, but is unchanged at 50 GHz. This results in about a 0.15 db underestimation of the loss of the cooled isolator at 30 GHz, and a 0.1 db underestimation of loss mid-band. Isolator serial number 574 was cooled to 19 Kelvin once and isolator serial number 575 was cooled to 19 Kelvin three times. As we were not certain the isolators would operate successfully at cryogenic temperatures, the performance of the first isolator (serial number 574) was measured as it was cooled, at 250 Kelvin, 200 Kelvin, 150 Kelvin, 100 Kelvin, 77 Kelvin, 50 Kelvin and finally at 19 Kelvin. Fig. 2 shows the performance of this isolator at room temperature, as supplied by the manufacturer, and at 19 Kelvin. - 1 -

Fig. 3 shows the performance of isolator serial number 575 at room temperature, as supplied by the manufacturer, and at 19 Kelvin after the third cooldown. The isolators were measured at room temperature after each cooldown and no significant change in performance was discernable. Fig. 4 shows the performance of isolator serial number 575 at room temperature after the third cooldown. The difference in S 21 between Fig. 3(a) and Fig. 4 is less than the drift in the calibration, so these tests show there was no significant change in the performance of the isolator as a result of it being cooled down to 19 Kelvin and warmed up three times. (a) Fig 1 Measurement setup for cryogenic testing of isolator model FBI-22 (a) view showing test dewar and vector network analyser, and view showing detail of isolator and temperature sensor in test dewar. - 2 -

(a) Fig 2. Isolator model FBI-22, serial number 574 (a) at room temperature, as supplied, and at 19 Kelvin. - 3 -

(a) Fig 3. Isolator model FBI-22, serial number 575 (a) at room temperature, as supplied, and at 19 Kelvin, after the third cooldown. - 4 -

Fig 4. Isolator model FBI-22, serial number 575, at room temperature, after the third cooldown. - 5 -