Peptide Mapping 101: Essential Tools for Effective Development and Characterization. Part 1:Introduction to Peptide Mapping



Similar documents
of Biologics Using LC/MS in Compliant Laboratories

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

Robert Birdsall, Eoin Cosgrave, Henry Shion, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Protein Separation Technology

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

A Guide to the Analysis and Purification of Proteins and Peptides by Reversed-Phase HPLC

ProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

PROTEIN SEQUENCING. First Sequence

Retrospective Analysis of a Host Cell Protein Perfect Storm: Identifying Immunogenic Proteins and Fixing the Problem

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

Monoclonal Antibody Fragment Separation and Characterization Using Size Exclusion Chromatography Coupled with Mass Spectrometry

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

ProteinPilot Report for ProteinPilot Software

Biopharmaceutical Glycosylation Analysis

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Application Note. Separation of three monoclonal antibody variants using MCSGP. Summary

Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary

Industry Perspective: Advantages of Open Access and Walkup LC/ MS Supporting Protein Drug Discovery and Development

Error Tolerant Searching of Uninterpreted MS/MS Data

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

The Theory of HPLC. Gradient HPLC

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

NUVISAN Pharma Services

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

How To Use An Acquity Qda Detector

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software

Guideline on similar biological medicinal products containing biotechnology-derived proteins as active substance: quality issues (revision 1)

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC

Separation of Peptides from Enzymatic Digestion on Different Acclaim Columns: A Comparative Study

Tackling the data analysis challenge for characterisation of biotherapeutics

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Agilent s Solutions for: Biosimilars & Antibody Drug Conjugates

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Laboration 1. Identifiering av proteiner med Mass Spektrometri. Klinisk Kemisk Diagnostik

How To Make A Drug From A Peptide

Electrospray Ion Trap Mass Spectrometry. Introduction

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics

This handbook presents the basic principles of reversed-phase HPLC for the

Thermo Scientific BioLC Columns. Innovative solutions for mab analysis and characterization

ASMS Regulated Bioanalysis Interest Group (RBIG) Workshop. Antibody-Drug Conjugates (ADC) A Complex Problem in Regulated Bioanalysis.

Introduction to the ACQUITY UPLC H-Class Amino Acid Analysis System

TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature

Quantification of Multiple Therapeutic mabs in Serum Using microlc-esi-q-tof Mass Spectrometry

2011 Waters Corporation

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes

Optimal Conditions for F(ab ) 2 Antibody Fragment Production from Mouse IgG2a

Application Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008

Contents. List of Contributors XIII The structure of The HPLC-Expert" Preface XVII

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Choices, choices, choices... Which sequence database? Which modifications? What mass tolerance?

LC-MS/MS STRATEGIES FOR QUANTIFICATION OF THERAPEUTIC ANTIBODIES TO SUPPORT CLINICAL

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Introduction to Bioprocessing

Implementing New USP Chapters for Analytical Method Validation

Application Note. Increasing the activity of monoclonal antibody isoforms by MCSGP. Summary

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM

Comparison of the Reversed-Phase Selectivity of Solid Core HPLC Columns

La Protéomique : Etat de l art et perspectives

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA

(c) How would your answers to problem (a) change if the molecular weight of the protein was 100,000 Dalton?

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

Aiping Lu. Key Laboratory of System Biology Chinese Academic Society

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Appendix 5 Overview of requirements in English

Methods for Protein Analysis

Monoclonal Antibody Characterization Achieving Higher Throughput and Productivity

Introduction to Proteomics 1.0

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

Introduction to Proteomics

Quantitative proteomics background

amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites

Waters Integrated Solutions El presente y el Futuro de las estrategias Informáticas de Laboratorio

Transcription:

Thank you for joining us! The Webinar will begin shortly Peptide Mapping 101: Essential Tools for Effective Development and Characterization Part 1:Introduction to Peptide Mapping Stephan M. Koza, Ph. D. Principal Applications Chemist Waters Technologies Corporation 2013 Waters Corporation 1

Friendly Reminders We will have LIVE Technical Support available to address your questions. Please use text chat functionality to submit questions during the Webinar. Upon conclusion, follow up information will be available: http://www.waters.com/aug20 Recorded version of today s presentation Copies of today s slides Product specific discount offers Product specific information Categorized reference materials 2013 Waters Corporation 2

Agenda What is Peptide Mapping and Why Do It? Protein Digestion Peptide Separations 2013 Waters Corporation 3

What is Peptide Mapping? For biotherapeutic proteins and peptides peptide mapping is: The chemical or enzymatic treatment of a protein to produce peptide fragments Separation and identification of these fragments in a reproducible manner In-depth analysis that can identify minor and even isobaric differences in protein primary structure such as errors in the transcription of complementary DNA, point mutations., and PTMs (CQAs) Due to the complexity and inherent variability of the method peptide mapping is generally a comparative procedure where the peptide map of the test sample is compared to that of a reference substance prepared in a side-by-side experiment. 2013 Waters Corporation 4

Uses of Peptide Mapping Proteomics Studies Protein Biopharmaceutical Analysis Structural characterization o Pattern conforms to primary structure o Used with MS for primary structure determination o Non-Reduced Mapping for Disulfide Bond Assignment Protein modification o Identify post-translational modifications Glycosylation, substitution, truncation o Determine product related impurities: deamidation, oxidation, etc. o Characterization of variants observed in other methods (IEX, SEC) Protein identity o Confirm presence of signature peptides o Product integrity lot-to-lot analysis 2013 Waters Corporation 5

Biopharmaceutical Classes That Use Peptide Mapping Methods Peptides/Proteins derived through recombinant DNA-based processes Insulin Diabetes Erythropoietin Cancer Monoclonal antibodies derived by recombinant DNA processes, and their derivatives o Herceptin Cancer Protein Conjugates ADC PEGylated proteins Synthetic peptides Oligonucleotides/siRNA Vaccines Gene therapy Cells/Stem cells 2013 Waters Corporation 6

Why Do We Develop Peptide Maps for Biotherapeutic Proteins? Guidance for Industry Q6B Specifications: Test Procedures and Acceptance Criteria for Biotechnological/Biological Products 1. Structural Characterization and Confirmation (6.1.1) d. Peptide map Selective fragmentation of the product into discrete peptides is performed using suitable enzymes or chemicals.peptide mapping of the drug substance or drug product using an appropriately validated procedure is a method that is frequently used to confirm desired product structure for lot release purposes. 2013 Waters Corporation 7

Agenda What is Peptide Mapping and Why Do It? Protein Digestion Peptide Separations On to Part II 2013 Waters Corporation 8

Enzymes and Chemistries for Protein Digestion EUROPEAN PHARMACOPOEIA 5.0, 2.2.55. PEPTIDE MAPPING 2013 Waters Corporation 9

Enzymes and Chemistries for Protein Digestion EUROPEAN PHARMACOPOEIA 5.0, 2.2.55. PEPTIDE MAPPING Trypsin, Lys-C,and Asp-N are most commonly used and can provide high fidelity digestions for reproducible peptide maps. 2013 Waters Corporation 10

In silico Digestion Tools for Selecting an Enzyme (or Chemical):MassLynx Protein/Peptide Editor 2013 Waters Corporation 11

In silico Digestion Tools for Selecting an Enzyme (or Chemical) Trypsin results in 2 amino acids and 1 di-peptide, Lys-C might be a better choice as it generates 3 manageable peptides Further digestion would be needed to assign disulfide bonds in this example 2013 Waters Corporation 12

Flow Chart of Peptide Mapping Protein (e.g. antibody) Denaturation, Disulfide Reduction/Alkylation, Buffer Exchange Enzymatic Digestion (e.g. Trypsin) Peptide Map Analysis UPLC/ UV UPLC/ MS UPLC/ MS/MS UPLC/ MS E 2013 Waters Corporation 13

What s RapiGest SF Anionic detergent that improves solubility and digestion of many proteins for improved enzymatic digests. Unlike conventional denaturants, RapiGest SF does not inhibit enzyme activities so it can reduce digestion times and reduces the amount of enzyme used. It does not cause protein modifications (e.g., urea causing carbamylation) unlike some other protein denaturants. It s an acid labile surfactant whose degradation products do not interfere with LC/MS or MALDI MS analysis. 2013 Waters Corporation 14

Reproducible Peptide Mapping Pitfalls of Peptide Mapping that can affect robustness, reproducibility and accuracy: Sample preparation o Incomplete digestion o Non-reproducible digestion conditions o Non-specific cleavages (over-digestion) o Enzyme lot-to-lot variability (activity units or mass?) Non-reproducible chromatography It is critical that SOPs be written clearly and transferred precisely in order for peptide maps to be reproducible between different labs or even analysts Preparing a blank digest is always recommended for troubleshooting purposes 2013 Waters Corporation 15

Agenda What is Peptide Mapping and Why Do It? Protein Digestion Peptide Separations On to Part II 2013 Waters Corporation 16

Peptide Separations Column Selection Ethylene Bridged Hybrid (BEH) Particle Technology UPLC vs HPLC Charged Surface Hybrid Technology Fine Tuning Your Separation 2013 Waters Corporation 17

Ethylene Bridged Hybrid - BEH Technology U.S. Patent No. 6,686,035 B2 and others patent pending Bridged Ethanes In Silica Matrix Organo Silica Hybrid Particles ph stability Reduced ionic interactions Basis of Peptide Separation Technology EtO CH 2 CH 2 OEt OEt O Si Si Si O O Si O O Si O O Si O EtO OEt OEt Polyethoxysilane Et Et n EtO EtO OEt 4 EtO CH 2 Si + Si EtO OEt CH Si OEt 2 EtO EtO OEt Tetraethoxysilane Bis(triethoxysilyl)ethane 2013 Waters Corporation Anal. Chem. 2003, 75, 6781-6788 18

Small Particle Size Mobile Phase Peptides 1500 Da Peptide 2 3.5 µm Porous Particle Diffusion-related band broadening Adsorption Equilibria H (mm) 1 Diffusion distances decrease Reduced Eddy diffusion, A-Term Improved mass transfer kinetics, C-Term 1.7 µm Column efficiency Narrower peaks 40 µl/min 2.1 mm ID 400 µl/min 0 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 Velocity (mm/sec) 2013 Waters Corporation 19

Why UPLC for peptide mapping More resolution even using a shorter gradient 7.0e-2 6.0e-2 HPLC 2.1 x 300 mm, 3.5 µ 90 min 5.0e-2 AU 4.0e-2 3.0e-2 2.0e-2 1.0e-2 Time 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00 70.00 75.00 80.00 85.00 90.00 1.0e-1 9.0e-2 UPLC 2.1 x 150 mm, 1.7 µ 55 min 8.0e-2 7.0e-2 AU 6.0e-2 5.0e-2 4.0e-2 3.0e-2 2.0e-2 Time 20.00 22.00 24.00 26.00 28.00 30.00 32.00 34.00 36.00 38.00 40.00 42.00 44.00 46.00 48.00 50.00 52.00 54.00 56.00 2013 Waters Corporation 20

Charged Surface Hybrid (CSH) Technology patent pending Charged Surface Hybrid (CSH) Technology and Its Use in Liquid Chromatography. P.C. Iraneta, K.D. Wyndham, D.R. McCabe, and T.H. Walter Waters White Paper 720003929EN 2011 Expands upon the robust BEH particle technology CSH130 C18 = BEH130 base particle + low level of basic moieties + trifunctional C18/end cap Acidic ph Positive Surface Charge Peptide 2013 Waters Corporation 21

Peak Capacity Peak Capacity = The number of peaks that can be separated within a retention window Neue, U. D., J Chromatogr A 2005, 1079 (1-2), 153-61. The best metric for determining the quality of gradient separations 100% 9 peaks could resolve ~300-400 Peak Height 50% 2.35σ w h 13.4% 4σ w 4σ 0% t gradient 2013 Waters Corporation 22

A Novel Column Chemistry: CSH130 C18 (0.1% TFA) BEH130 C18 Competitor s Industry Standard C18 Porous (130Å) 1.7 µm 2.1 x 150 mm UV absorbance (214 nm) Porous (300Å) 5 µm 2.1 x 250 mm 10 10 50 10 Time (min) 20 30 Time (min) 40 50 50 CSH130 C18 Competitor s Superficially Porous Peptide C18 Porous (130Å) 1.7 µm 2.1 x 150 mm SPP (100Å) 1.7 µm 2.1 x 150 mm 1 10 50 Time (min) 2013 Waters Corporation 2 50 10 Time (min) 23

Peak Capacity - FA vs TFA 370 320 270 P c,4σ 220 170 120 Competitor s Industry Standard Silica C18 5 µm 2.1 x 250 mm 70 20 0.00 0.05 0.10 FA TFA Percent TFA % TFA 0.00 0.05 0.10 % FA 0.10 0.05 0.00 2013 Waters Corporation 24

Peak Capacity - FA vs TFA 370 320 270 BEH130 C18 1.7 µm 2.1 x 150 mm P c,4σ 220 170 120 Competitor s Industry Standard Silica C18 5 µm 2.1 x 250 mm 70 20 0.00 0.05 0.10 FA TFA Percent TFA % TFA 0.00 0.05 0.10 % FA 0.10 0.05 0.00 2013 Waters Corporation 25

Peak Capacity - FA vs TFA 370 320 270 Competitor s SPP Peptide C18 1.7 µm 2.1 x 150 mm BEH130 C18 1.7 µm 2.1 x 150 mm P c,4σ 220 170 120 Competitor s Industry Standard Silica C18 5 µm 2.1 x 250 mm 70 20 0.00 0.05 0.10 FA TFA Percent TFA % TFA 0.00 0.05 0.10 % FA 0.10 0.05 0.00 2013 Waters Corporation 26

Peak Capacity - FA vs TFA 370 CSH130 C18 1.7 µm 2.1 x 150 mm 320 270 90% Competitor s SPP Peptide C18 1.7 µm 2.1 x 150 mm BEH130 C18 1.7 µm 2.1 x 150 mm 20% P c,4σ 220 170 120 Competitor s Industry Standard Silica C18 5 µm 2.1 x 250 mm 70 20 0.00 0.05 0.10 FA TFA Percent TFA % TFA 0.00 0.05 0.10 % FA 0.10 0.05 0.00 2013 Waters Corporation 27

Loadability Attribute how much analyte can be loaded before peak shape deteriorates CSH C18 BEH C18 450 400 B Low Mass Load 0.6 µg 0.6 of Low µg mixture Mass of Load mixture (Equivalent to ~ 4.5 µg of a mab) CSH C18 CSH130 BEH C18 1.7 µm 450 400 A Typical Mass Load 6 µg 6 µg of mixture High of Mass mixture Load (Equivalent to ~ 45 µg of a mab) CSH C18 BEH C18 CSH130 C18 1.7 µm 350 BEH130 C18 1.7 µm 350 P c,4σ 300 P c,4σ 300 BEH130 C18 1.7 µm P c,4σ 250 250 200 200 *Previously shown 0.10 0.00 % TFA % FA 150 0.00 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 TFA 0.10 0.10 FA Percent 0.05 TFA 0.00 0.00 0.05 0.10 0.10 0.05 0.00 150 % TFA 0.00 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 TFA0.10 % FA 0.10 FA Percent 0.05TFA 0.00 % TFA 0.00 0.05 0.10 % FA 0.10 0.05 0.00 2013 Waters Corporation 28

Which Column do I choose CSH130 C18 or BEH130 C18? Intensity 2E+6 1E+6 0.1% FA BEH130 C18 C18 1.7 1.7 µm µm P c,4σ = P c,4σ 399= 399 0E+0 0 10 20 30 40 50 60 Intensity 2E+6 1E+6 0 10 20 30 40 50 60 Time(min) CSH130 C18 C18 1.7 1.7 µm µm P c,4σ P= c,4σ 532 = 532 0E+0 0 10 20 30 40 50 60 Time(min) 2013 Waters Corporation 29

LC-MS Retention and Selectivity BEH130 C18 1E+6 T3 SVYDSR T5 GVFR T12 ANIDVK T19 HLADSK More positive T5/ T10 GVLHAVK T12 T19 charge T3 T10 T40 IATAIEK 0E+0 5.5 6.5 7.5 8.5 9.5 10.5 11.5 12.5 Time(min) T40 1E+6 CSH130 C18 T5/ T19 T12 T10 T3 T40 0E+0 5.5 6.5 7.5 8.5 9.5 10.5 11.5 12.5 13.5 Time(min) 2013 Waters Corporation 30

UPLC and HPLC CSH130 C18 2.1 x150 mm 1.7 µm A 214 1.0 0.8 0.6 0.4 High peak capacity separations not limited to UPLC 0.1 % FA ~8000 psi A 214 1.0 0.8 0.6 0.4 0.1 % TFA 0.2 0.2 Method Transfer 0.0 10 20 30 40 50 1.0 Time (min) 0.0 10 20 30 40 50 1.0 Time (min) 0.8 0.8 2.5 µm XP A 214 0.6 0.4 ~3000 psi A 214 0.6 0.4 Longer Run Time Lower Pressure 0.2 0.0 14.5 24.5 34.5 44.5 54.5 64.5 74.5 Time (min) 0.2 0.0 13.5 23.5 33.5 43.5 53.5 63.5 73.5 Time (min) 2013 Waters Corporation 31

UPLC and HPLC CSH130 C18 2.1 x150 mm 1.7 µm A 214 1.0 0.8 0.6 High peak capacity separations not limited to UPLC 0.1 % FA ~8000 psi 0.4 0.4 CSH130 C18 Peptide Separation Technology Columns A 214 1.0 0.8 0.6 0.1 % TFA Method Transfer 2.5 µm XP Longer Run Time Lower Pressure 0.2 Available 0.0 Now: Analytical Columns 1.0 Time (min) 1.7 µm 0.8 2.5 µm XP 0.6 3.5 µm Prep 0.4 Columns (5 µm) A 214 0.2 10 20 30 40 50 0.0 14.5 24.5 34.5 44.5 54.5 64.5 74.5 Time (min) ~3000 psi 0.2 Upcoming: 0.0 Nano (75, 150, 300 µm ID) A 214 1.0 10 20 30 40 50 Time (min) 0.8 0.6 0.4 0.2 0.0 13.5 23.5 33.5 43.5 53.5 63.5 73.5 Time (min) 2013 Waters Corporation 32

Three Outstanding Peptide Separation Technology Columns Peptide/Protein kda 1 Bradykinin 1.1 130 Å A 214 1.2 1 0.8 0.6 0.4 0.2 CSH130 C18 1.7 µm 1 2 0 1.2 10 15 20 25 30 35 40 45 50 BEH130 C18 1.7 µm Time (min) 1 3 4 5 6 2 Renin Substrate 1.8 3 Ubiquitin 8.6 4 5 Cytochrome C (Equine) Insulin (Bovine) 12.4 5.7 6 Melittin 2.8 0.8 A 214 0.6 0.4 0.2 1 2 4 3 5 6 300 Å A 214 0 1.2 10 15 20 25 30 35 40 45 50 BEH300 C18 1.7 µm Time (min) 1 2.1 x 150 mm columns 2% ACN for 1 min, 0.8 then to 50% ACN over 60 min 5 0.3 ml/min 0.6 40 C 0.4 0.2 1 2 4 3 6 ACQUITY UPLC H-Class Bio UV @ 214 nm / Xevo G2 QTOF 1 µg each component 0 10 15 20 25 30 35 40 45 50 Time (min) 2013 Waters Corporation 33

New Addition to the Suite of Waters Peptide Separation Technology Peptide Separation Technology Peptide C18 Columns QC Tested with Digests BEH Technology BEH130 C18 and BEH300 C18 Outstanding Performance for Most Applications Two Pore Sizes Particle Sizes: 1.7 µm, 3.5 µm, 5 µm Analytical, Nano and Prep Columns Now even more tools in the toolbox CSH Technology CSH130 C18 Highest peak capacities in TFA and FA mobile phases. Unique selectivity Particle Sizes: 1.7 µm, 2.5 XP, 3.5 µm, 5 µm Analytical and Prep Columns (Nano in development) 2013 Waters Corporation 34

All Waters Peptide SeparationColumns are Quality Control Tested with Tryptic Digest of Cytochrome c CSH130 C18 0.1 % Formic Acid 2013 Waters Corporation 35

Fine Tuning Your Separation Parameters that Influence Selectivity Ion Pairing Reagent (TFA, HFBA, etc.) and Concentration Organic Eluent (MeCN, MeOH, IPA) Column Temperature Gradient Slope/Column Length Peak Tracking Ideally using LC-MS can expedite separation optimization Make several incremental changes Peak areas and A280/A214 UV absorbance ratios 2013 Waters Corporation 36

Method Optimization: Gradient Slope Rate of Change 0.75%/ col. vol. % * * * 1 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00 32.00 % * * Rate of Change 1.5%/ col. vol. * 1 Time 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00 32.00 2013 Waters Corporation 37

Why Does This Switch in Elution Order Occur? Log k Elution at Lower % MeCN w/ Shallower Gradient Elution at Higher % MeCN w/ Steeper Gradient % MeCN Adapted from: Spicer, V., Grigoryan, M., Gotfrid, A., Standing, K. G., & Krokhin, O. V. (2010). Predicting retention time shifts associated with variation of the gradient slope in peptide RP-HPLC. Analytical chemistry, 82(23), 9678-9685. 2013 Waters Corporation 38

Gradient Slope and Segmented Gradients Changes in gradient slope should occur in regions of separation where there are no peaks of interest Potential selectivity differences should be tracked Approach could also be used to generate a focused gradient if only specific peptides are of interest 2013 Waters Corporation 39

CSH130 C18 Useful Current Literature/Resources http://pubs.acs.org/doi/abs/10.1021/ac401481z Previously recorded webinar available: http://www.waters.com/waters/promotiondetail.htm?id=134727909 Recent Application Notes Increasing Peak Capacity in Reversed Phase Peptide Separations with Charged Surface Hybrid (CSH) C18 Columns M.A. Lauber, S.M. Koza, K.J. Fountain Waters Application Note 720004568EN 2013 Peptide Mapping and Small Protein Separations with Charged Surface Hybrid (CSH) C18 and TFA-Free Mobile Phases M.A. Lauber, S.M. Koza, K.J. Fountain Waters Application Note 720004571EN 2013 High Mass Loading of Peptides with Hybrid Particle C18 Columns and Acetic Acid Mobile Phases M.A. Lauber, S.M. Koza, K.J. Fountain Waters Application Note 720004674EN 2013 2013 Waters Corporation 40

End of Part 1 2013 Waters Corporation 41

Part 2 - Gaining Efficiency: Instrumentation and Informatics Platforms for Peptide Mapping Asish Chakraborty, Ph.D Asish_Chakraborty@waters.com 2013 Waters Corporation 42

A History of Relieving the Pressure on Analysts Sample Generation Sample Preparation Acquisition Data Analysis Report Generation Chemistries Sample Generation Instrumentation and Automation Data Analysis Informatics Report Generation Sample Generation Report Generation Sample Generation Now It Becomes Routine 2013 Waters Corporation 43

Performance and Usability through Engineered Simplicity Automatically ensuring the system is ready to run 2013 Waters Corporation 44

Automating batch processing, annotatation, and comparison tools in BiopharmaLynx TM increases productivity First shown at WCBP 2007 Meeting 2013 Waters Corporation 45

Biopharmaceutical Platform Solution with UNIFI 1.7 An analytical system for biotherapeutic analysis integrating UPLC/UV and UPLC/MS Intact Protein Mass Peptide Mapping Biopharmaceutical Platform Solution DDA (Peptide & Glycan) Released Glycan GU + Mass Xevo G2-S QTof ACQUITY UPLC H-Class & H-class BIO Bioseparations Size Exclusion (UV) Intact Protein: Peptide Mapping: Released Glycan: Bioseparations: TUV, MS TUV, MSE, MS/MS FLR (+MS, NIBRT Library), MS/MS TUV, FLR Workstation or Workgroup (Compliance) 2013 Waters Corporation 46

Deploy high resolution analytics across a biotherapeutic organization Few compliance issues GxP Labs Regulatory Compliance Discovery Development Production QC/QA Post-Approval Characterization Monitoring Release 2011 Waters Corporation 47

The Biopharmaceutical Workgroup Office PC Office PC Intact Mass, Peptide Mapping Lab PC UPLC-TUV- Xevo G2-S Released Glycan Analysis Lab PC UPLC-FLR- Xevo G2-S Office PC Lab Network Device (LND) LND LABORATORY NETWORK Lab PC LND Lab PC LND UPLC-TUV UPLC-TUV UPLC-FLR UPLC-FLR Data Processing & Database Storage Server Bioseparations 2013 Waters Corporation 48

UNIFI Meets the Biopharmaceutical industry s Global reach A Scientific management system for the global nature of the Biopharmaceutical business 2013 Waters Corporation 49

Peptide Mapping in UNIFI TM Advanced Reporting Capabilities in a GxP-ready Environment 2013 Waters Corporation 50

Experimental Setup for Peptide Mapping LC/MS E Reduced Peptide Map Reduction & Alkylation Therapeutic Proteins Denature & Alkylate LC/MS E Non-Reduced Peptide Map Trypsin Digest UNIFI Scientific Information System UPLC BEH300 C18, 1.7 µm, 2.1 x 100 mm 2011 Waters Corporation 51

UPLC/MS E Comprehensively Analyzes Complex Samples UPLC/MS E is a simple method of unbiased data acquisition that comprehensively analyzes all components in a single analysis. 2011 Waters Corporation 52

Surveying Chromatography and Complexity in Peptide Mapping Data Chromatogram with Peak Assignments 2013 Waters Corporation 3D Chromatogram 53

Peptide Mapping Data Assignments Annotated Chromatograms Fragment ions Spectrum 2013 Waters Corporation 54

Peptide Mapping Data Data Table (linked to Coverage Map) Assignments Coverage Map Fragment ions Spectra 2013 Waters Corporation 55

Access to both raw and processed data 2013 Waters Corporation 56

Case Study 1 2011 Waters Corporation 57

Experimental Setup for Peptide Mapping LC/MS E Reduced Peptide Map Reduction & Alkylation Therapeutic Proteins Denature & Alkylate LC/MS E Non-Reduced Peptide Map Trypsin Digest UNIFI Scientific Information System UPLC BEH300 C18, 1.7 µm, 2.1 x 100 mm 2011 Waters Corporation 58

Equivalent protein coverage was obtained for innovator and biosimilar Innovator Innovator HC Biosimilar Biosimilar HC Innovator LC Biosimilar LC BEH, C18, 1.7 µm, 130, 2.1x 100 mm, Gradient 1 to 35% ACN, 0.05%TFA, 60 min 2011 Waters Corporation 59

Asp Isomerization of Peptide T24 (FNWYVDGVEVHNAK) XIC Iso ASP Innovator Biosimilar Isomerization: Asp to iso-asp (no mass difference). isoasp is not a natural amino acid and can potentially be immunogenic. 2013 Waters Corporation 60

Oxidation of HC Peptide T42 Biosimilar Batch I % Oxidation Innovator Batch I Sample Injections 2013 Waters Corporation 61

Peptide Maps Report: Unifi enables researchers to focus on critical attributes of a molecule Analysis Information LC Coverage HC Coverage % Oxidation Biosimilar Batch I Innovator Batch I 2011 Waters Corporation 62

Experimental setup for disulfide bond mapping LC/MS E Reduced Peptide Map Reduction & Alkylation Therapeutic Proteins Denature & Alkylate LC/MS E Non-Reduced Peptide Map Trypsin Digest UNIFI Scientific Information System UPLC BEH300 C18, 1.7 µm, 2.1 x 150 mm 2011 Waters Corporation 63

Expected disulfide bonds in IgG1 Antibody Trypsin Digest Heavy chain IgG1 mab contains 16 S-S bonds (12 intra, and 4 inter) Light chain S - S Light Chain (1) S - S VH C H 1 S - S CHO Heavy Chain (2) S - S S-S S-S S-S -S-S- S-S CH2 CH3 S S S S S S S S S-S S- S S- S Humanized IgG K K C L S- S CHO S- S V L Light Chain (4) Heavy Chain (3) Digestion Enzyme: Trypsin Symmetry of IgG1 molecule provides redundancy in mass-based search 8 unique S-S bonded peptides LC: 2 Intra, HC: 4 Intra, HC-HC(Hinge): 1 inter HC-LC:1 inter 2013 Waters Corporation 64

Nonreduced peptide mapping enabled ID of all canonical S-S S peptides A simple filter to only display disulfide containing peptides Disulfide Containing Peptides 2:T21-3:T21 Additional studies show there are no scrambled disulfide presence 2:T21-3:T21 MS E Fragment Ions 2:T21-3:T21 2:T21-3:T21 2:T21-3:T21 UNIFI enables researchers to focus on critical attributes of a molecule 2013 Waters Corporation 65

Disulfide Bonds Report: Unifi enables researchers to focus on critical attributes of a molecule Component Plot for S-S peptides Analysis Information KK Disulfide containing peptides identified in both innovator and biosimilar mab samples Component Summary 2011 Waters Corporation 66

Case Study 2 2013 Waters Corporation 67

Automated Processing and Reporting with UNIFI : Intact Protein Analysis INTACT PROTEIN ANALYSIS MaxEnt1 deconvoluted mass spectra in compare mode G0F/G0F G0F/G1F G1F/G1F G0F/G2F Innovator G0F/G0F G0F/G1F G1F/G1F G0F/G2F Innovator G1F/G2F G1F/G2F G0/G0F G2F/G2F G0/G0F G2F/G2F Biosimilar 1 Biosimilar 2 m = 56 Da Biosimilar 1 glycoforms broadly match the innovator Biosimilar 2 glycoforms have a systematic mass shift of 56 Da compared to innovator mab Discrepancy needs to be explained UNIFI workflow automatically acquires, processes and reports the intact mass Deconvolution with MaxEnt Reporting with Flexible templates Ivleva et al Poster - ASMS 2012; 2011 Waters Corporation http://www.waters.com/webassets/cms/library/docs/2012asms_ivleva_rituximab.pdf 68

Reduced Protein Analysis LCs identical Biosimilar 2 LC w/ PyQ HC Reduction + Biosimilar 1 Mass Analysis of the Light Chain Light Chain Innovator MaxEnt1 deconvoluted mass spectra in compare mode LC Light Chain Innovator Innovator Biosimilar 1 Biosimilar 2 Automated Processing and Reporting Light chain masses are identical for Innovator, Biosimilar 1 and Biosimilar 2 2011 Waters Corporation 69

Reduced Protein Analysis of Heavy Chain G0F G1F MS Response Biosimilar 1 G0 G0 G0F+K Innovator G2F G1F+K 7.00E+06 6.00E+06 5.00E+06 4.00E+06 3.00E+06 Innovator Quantification of C-terminal Lys Variation G0 G0F G0F+K G1F Biosimilar 1 Innovator G2F G1F+K Summary plots Based on UNIFI results 2.00E+06 1.00E+06 0.00E+00 MS Response 1.20E+07 1.00E+07 8.00E+06 6.00E+06 Biosimilar 2 1 2 3 4 5 6 7 8 9 Distribution of G0 Glycoform Biosimilar 2 G0 G1F Biosimilar 2 4.00E+06 2.00E+06 Innovator Biosimilar 1 G0F G0F m = 28 Da 0.00E+00 Automated Processing and Reporting 1 2 3 4 5 6 7 8 9 Detailed Information automatically reported in UNIFI Multiple aspects available from the dataset Response for each batch of each protein measured and compared 2011 Waters Corporation 70

LC/MS E Tryptic peptide mapping to locate the sequence variance Tryptic Digest comparison between Innovator and Biosimilar 2 does not show sequence differences in Light Chain Light Chain - Innovator Compare mode Biosimilar 2 Coverage Map Automated Reporting 2013 Waters Corporation 71

LC/MS E Tryptic peptide mapping to locate the sequence variance Compare mode Heavy Chain - Innovator HC - Biosimilar 2 Biosimilar Coverage Map: shows section where no sequence match is made 2013 Waters Corporation 72

LC/MS E Tryptic peptide mapping to locate the sequence variance Compare mode Heavy Chain - Innovator Biosimilar 2 Coverage Map Alternative Enzyme to Trypsin needed to ascertain if there is a different sequence here in Biosimilar 2 2013 Waters Corporation 73

LC/MS E Chymotryptic peptide mapping analysis of Biosimilar 2 Innovator Innovator Biosimilar 2 BPI Peptide Map Biosimilar 2 m = 28 Da HC Coverage Map Chymotryptic Digest used to reveal differences Peptides are highlighted in the coverage map as each is selected by the user Amino Acid Substitution can be identified 2011 Waters Corporation 74

UNIFI Peptide Map Workflow: MS E data confirming the sequence Variant K 218 R 218 Automated fragment information from MS E data MS E Spectrum of chymotryptic digest confirms amino acid substitution K for R at position 218 2011 Waters Corporation 75

Summary of the Structural Analysis of Rituximab by LC/MS Approach The differences between Innovator and Biosimilars Rituximab candidates are: Biosimilar 1 vs. Innovator o Same AA Sequences o Higher percentage of C- terminal variation o Increased G0 glycoform o Different percentage of pyroglutamation at the N- termini of both LC and HC Biosimilar 2 vs. Innovator o Sequence Variant in HC, K 218 > R 218 o Lower percentage of C-terminal Lys variation o Much higher percentage of G0 o Different percentage of pyroglutamation at the N-termini of both LC and HC 2011 Waters Corporation 76

Summary Intact Mass Analysis Reduced Peptide Mapping Glycan Analysis Reduced LC Mass Analysis Non-Reduced Peptide Mapping Aggregate Analysis Reduced HC Mass Analysis Charge variant Analysis 2013 Waters Corporation 77

2013 Waters Biopharmaceutical Application Notebook www.waters.com/ biopharmbook 2013 Waters Corporation 78

Thank You! Questions? Landing Page http://www.waters.com/aug20 Promotional Discount Offers on Peptide Separation Columns PDF Slide Deck Full Webinar Recording of Today s Session Compilation of Literature, White Papers, Brochures General Questions email: mychemrep@waters.com 2013 Waters Corporation 79