Direct Analysis of Opiates in Urine by Liquid Chromatography-Tandem Mass Spectrometry
|
|
|
- Lucinda Rodgers
- 10 years ago
- Views:
Transcription
1 [ TechnicalNote Direct Analysis of Opiates in Urine by Liquid Chromatography-Tandem Mass Spectrometry Leslie E. Edinboro 1,2,*, Ronald C. Backer 3, and Alphonse Poklis 2 7Department of Pharmaceutics, School of Pharmacy, 2Department of Pathology, School of Medicine, Virginia Commonwealth University, Richmond, Virginia and 3Ameritox, Ltd., Midland, Texas Abstract I A method for the direct analysis of 10 opiate compounds in urine was developed using liquid chromatography-mass spectrometry-mass spectrometry (LC-MS--MS) with electrospray ionization interface (ESI). Opiates included were morphine-3-~glucuronide, morphine-6-~glucuronide, morphine, oxymorphone, hydromorphone, norcodeine, codeine, oxycodone, 6-monoacetylmorphine (6MAM), and hydrocodone. Urine samples were prepared by centrifugation to remove large particles and direct injection into the LC-MS-MS. Separation and detection of all compounds was accomplished within 6 rain. Linearity was established for all opiates except 6MAM from 50 ng/ml to 10,000 ng/mt; 6MAM from 0.25 ng/mt to 50 ng/ml with all correlation coefficients (r) > Interrun precision (%CV) ranged from 1.1 % to 16.7%, and intrarun precision ranged from 1.3% to 16.3%. Accuracy (% bias) ranged from -7.3% to 13.6% and -8.5 % to l!.8 for inter- and intrarun, respectively. Eighty-nine urine samples previously analyzed by gas chromatography-ms were re-analyzed by the LC-MS-MS method. The qualitative results found an 88% agreement for negative samples between the two methods and 94% for positive samples. The LC-MS-MS method identified 19 samples with additional opiates in the positive samples. Overall, the direct injection LC-MS-MS method performed well and permitted the rapid analysis of urine samples for several opiates simultaneously without extensive sample preparation. Introduction The use of liquid chromatography-mass spectrometry (LC-MS) for the analysis of opiates has increased since first being reported in 1995 (1). One application has been in the area of pharmacokinetic studies (1-5). These reports have primarily focused on morphine and its glucuronide metabolites. Oxycodone (6), hydrocodone, and hydromorphone (7) methods have been reported as well. As opiates play a significant role in 9 Author to whom correspondence should be addressed. Current address: Commonwealth of Virginia, Department of Forensic Sciences, Central Laboratory, 700 North 5lh Street, Richmond, Virginia [email protected]. the field of forensic toxicology, methods for their determination in various biological matrices of forensic importance have been published (8-12). Most of these methods employ some kind of sample preparation technique such as liquid-liquid or solid-phase extraction. Fitzgerald et al. (10) utilized a sophisticated column switching technique to purify directly injected urine prior to introduction into the LC-MS. This application was for broad urine drug screening and quantitation of the screened opiates was not presented. More recently, Dams et al. (11) reported a simple protein precipitation using acetonitrile for the analysis of 27 compounds in oral fluids. Heroin, morphine, codeine, and their metabolites were included in this application. Protein precipitation techniques or "dilute-andshoot" provide a rapid and simple sample pretreatment, but suffer from the potentially adverse effects of ion suppression discussed by Darns et al. (11). Ion suppression effects or matrix effects result in loss of signal due to competition for available charge within the ionization source. Electrospray ionization (ESI) is more susceptible to matrix effects; therefore, Dams et al. (12) applied atmospheric pressure chemical ionization (APCI) LC-MS-MS in developing a urine drug screening method utilizing the direct injection of urine. The opiates included in this method were morphine, normorphine, codeine, norcodeine, acetylcodeine, heroin, and 6-monoacetylmorphine (6MAM). This report presents a method for the analysis of 10 opiates, including the glucuronide metabolites of morphine, by the direct injection of urine and ESI-LC-MS-MS. Experimental Materials LC-MS-MS. Hydrocodone bitartrate salt, oxycodone HC1, morphine-3-glucuronide (M3G), and morphine-6-glucuronide (M6G) were purchased from Sigma (St. Louis, MO). Morphine, oxymorphone, codeine, norcodeine, and 6MAM were purchased from Cerilliant Corporation (Round Rock, TX) as 1.0 mg/ml standards in methanol except 6MAM, which is in acetonitrile. 704 Reproduction (photocopying) of editorial content of this journal is prohibited without publisher's permission.
2 Table I. LC-MS-MS Acquisition Parameters Dwell Collision Cone Segment Precursor Product Time Energy Voltage Time Channel (m/z) (m/z) (s) (ev) (V) (min) M3G M6G Morphine-d Morphine Oxymorphone Hydromorphone Norcodeine Codeine Oxycodone MAM-d MAM Hydrocodone R S MONOACETYLMORPHINE- d~ i... ~... ~... t... ~..., i...,, ~ 6-MONOACETYLMORPHINE 100~ OXYCODONE lt~'/on0 ~ HYDROCODONE , 6.72e3 4: MRM 0{' 4 Channels ES > e \ 2.81e4 ln., > CODEINE // 1.42e4... i... i... p... i... i... i... q... r... r... i... ~... i... J... i... J lnn_ o;o. ~o~),~ HYDROMORPHONE A 2.86e4... t... i... i... i... ~...,... i... i... f... i... ~... i... ~... J..., 101._ ;~ NORCODEINE 4.25A 1.81e3 ol... ~.J,t MIP ,,,n~ OXYMCRPHONE.~ 3.t4e3 10 n :/:~ MORPHINE-d3 A 0... i... i... i... i... i... ~... i..., /o0 ~ MORPHINE... ~,, i... i... i... i... i... r... i e3 0 `1... i... i... i... ~... J... J..,... ~... i... ~... J... i... J... J 100-~ > 286 'Y~ MORPHIN E-6-GLUCURON IDE A ,,-,~, ~, i,.i... i... i,, J, '..i... ~... i... i... i...,... i... i 1: MRM of 1 Channel ES ~AO~ A MORPHINE',.~GLUCURONIDE 462 > ~! \ 3.oi Time (min) Figure 1, Non-extracted standard containing 0.25 ng/ml 6MAM and 50 ng/ml all other opiates. Deuterated internal standards (IS), morphine-d3 (1.0 mg/ml in methanol), and 6MAM-d6 (100 IJg/mL in acetonitrile) were purchased from Cerilliant. Blank pooled urine was provided by the Bioanalytical Laboratory, Virginia Commonwealth University and Quality Assurance Services (QAS, Augusta, GA). Formic acid (99%) and ammonium formate were purchased from Aldrich Chemicals (Milwaukee, WI). Acetonitrile and methanol were B&J brand high-purity solvents from Burdick and Jackson (Muskegon, MI). Deionized water (DI H20) was produced by a NANOpure Diamond TM water purification system from Barnstead International (Dubuque, IA). Gas chromatography (GC)-MS. All primary reference materials of oxycodone, oxymorphone, and other drugs were obtained as 1.0 mg/ml methanolic solutions from Cerilliant Corporation. Drug-free urine and GC-MS urine controls were obtained from QAS. Apparatus LC-MS-MS. Two Shimadzu (Columbia, MD) LC-10AVP pumps were connected through a high-pressure mixing tee with a 0.2-1Jm inline filter prior to the autosampler. Mobile phases were passed through a Shimadzu DGU-14 solvent degasser prior to entering the pumps. The autosampler was from LEAP Technologies (Carrboro, NC) with CTC Analytics HTS PAL with Peltier cooling stacks and a TRIOS switching valve. A 10-1JL syringe was fitted in the autosampler. A Shimadzu SCL-10 controller was integrated to control communication with the pumps, autosampler, and the MS. The MS was a Waters (Milford, MA) Micromass Quattro LC triple-quadrupole with an electrospray ionization (ESI) interface. Liquid N2 was used to produce the desolvation/nebulizer gases needed. Argon was used as the collision gas. MassLynx version 3.5 was installed as the operating software performing data acquisition and processing functions. The analytical column was an Agilent Zorbax SB Phenyl (2.1 mm 150 mm, 5-1Jm particle size). A 0.2-1Jm inline filter was connected directly to the frontend of the analytical column using a zero dead volume connector. The HPLC column was placed in a Jones Chromatograhy (Cardiff, Wales) column heater and maintained at 30~ The mobile 705
3 phase was established as A: 95% (0.05% formic acid with 10mM ammonium formate in DI-H20) and 5% (0.05% formic acid in acetonitrile) and B: 95% (0.05% formic acid in acetonitrile) and 5% (0.05% formic acid with 10mM ammonium formate in DI- H20). Flow was 0.25 ml/min. The following gradient was used: min, 0% B, linear gradient to 18% B at 3.00 min, hold 18% B for 2.5 rain, then return to 0% B. Injection volume was 5 tjl, and autosampler temperature was set at 6~ Autosampler wash solutions were Wash 1: methanol and Wash 2: Mobile Phase A. An Eppendorf refrigerated centrifuge was used to centrifuge all samples prior to injection. GC-MS. The samples were analyzed on an Agilent (Pale Alto, CA) 6890 GC with a split/splitless injection port, a 7673 auto-sampler and a 5973A mass selective detector (MSD). The column was an HP-5 capillary column (5.0 m x 0.1-mm i.d., Jm film thickness). Column flow rate was 1.0 ml/min; inlet pressure at psi; and injection was in the 4:1 split mode with a split flow of 4.0 ml/min. The oven temperature program was as follows: initial 170~ for 0 min, then ramped to 280~ at 30~ and held for 0.33 min. Under these conditions, the retention times of opiate TMS derivatives were 2.85 min for codeine; 3.02 min for morphine; 3.10 min for hydrocodone-oxime; 3.17 min for hydromorphone-oxime; and 3.32 min for oxycodone-oxime. The MSD was operated in the SIM mode using the following ions (quantitative ions are in parentheses): 234, 343, (371) for codeine; 346 (374) for codeine-d3; 234, 401, (429) for morphine; 417, (432) for morphine-d3; 297, 371, (386) for hydrocodone-oxime-tms; 368, 444 (459) for oxycodone-oxime-tms; and 429, 444, (355) for hydromorphoneoxime-tms. Methods Stock solutions. Standard 1.0 mg/ml solutions in methanol were prepared for all solid compounds with the exception of M3G and M6G, which were prepared in DI H20. A 0.01 mg/ml 6MAM standard was prepared by serial dilution with methanol of the 1.0 mg/ml stock. A 0.01 mg/ml 6MAM-d~ solution was prepared by dilution of the stock standard. Separate stock solutions were prepared for preparation of calibrators and controls. Final mixed internal standard solution was prepared by transferring 100 IJL of each of the 1.0 mg/ml morphine-d3 and the 0.01 mg/ml 6MAM-d~ into a ]2-ram x 75-ram glass tube. The methanol was evaporated at 40~ under N2. A 1.0 ml solution of 95:5 DI H20/ACN was added to the tube, which was then vortex mixed and capped. Final IS concentrations are 100 tjg/ml 706 morphine-d3 and 1 IJg/mL 6MAM-d6. Urine calibrators~controls. A 10,000/50 (opiates/6mam) ng/ml mixed calibration urine standard was prepared in a single 20.0-mL volumetric flask using the appropriate dilutions of the stock standards with drug-free pooled urine as the diluent. Six additional calibration curve standards were prepared by serial dilution of 10,000/50 ng/ml mixed calibration standard. An additional 10,000/50 ng/ml mixed urine control solution was prepared separately from the primary reference stock solutions as described for the calibrators. Additional urine controls at 1000/5.0 ng/ml and 100/0.5 ng/ml were prepared. A 10,000 ng/ml mixed standard was prepared as described except that 95:5 DI H~O/ACN was used as the diluent. Final urine calibrators were prepared by transferring 1.0 ml of the appropriate calibration standard to a 1.5-mL autosampler vial. Ten microliters of IS solution was then added, and the vial was capped and vortex mixed. Samples were then placed in an Eppendorf refrigerated centrifuge (15~ and spun for 10 rain U1 BLANK 1 4: M RM of 4 Channels ES ~ 1( '84~ 7 ~ ^, "u' e3 0 ~.... i.... i.... i.... i.... L.... i.... L.... i.... i i.... [.... i % "85" > i.... i....,.... i.... i.... J.... ~.... F.... i.... i.... i.... i.... i.... J ~ 5.67~ 5: % 5.01~,~'-~f?" 4.98e4.... i.... i.... i.. ~ w ~ - ~.. i.... i.... b.... i.... =.... i.... i.... i.... i.... i.... i 4; MRM of 4 Channels ES nl 54&5.85-, o,fl v~vr 2.71e5 o, ~ 4 174A~ 4.654,98 300> ~ off o r.... i.... i....,.... j....,.... i.... i.... J.... i.... i.... i.... i 2: MRM of 3 Channels ES+ 1007, 3.77,,&86 _ %i ~1~ 2.99e i.... i.... i, i.... i.... i.... i.... i.... ~.... i.... ~,,, J.... i.... i.... i 2: MRM of 3 Channels ES+ 100~ :1 oo9 ~,. i.... J.... i.... i.... i.... i.... J.... J.... J.... i.... i.... i.... i.... i.... i 2: MFIM of 3 Channels ES+ 1001~ 366~, ~P/~1"3"86~3' ~ i.... i....,.... t.... i.... J....,.... ~.... ~.... i.... ~.... r.... =.... i.... 1: MRM of 1 Channel ES %0~,04 0,s ~.~ 6.73e4 0~... i 2 ~ , Time (min) Figure 2. Evaluation of ion suppression, Ion channels as described in Figure 1.
4 "L Journal of Analytical Toxicology, Vol. 29, October 2005 U BLANK MIP110903D91 4: MRM of 4 Channels E$ >1649 1~176 "-. "L5.6% ~ip5 1~8 MtPl 10g03D0"l , > i.... i.... i.... i.... i... ~.... i.... i.... i.... i.... i.... ~.... J.... i.... i 100--~ ~ i i i.. b, i i... i... ~ i i. i.... i.... i.... i ~.... 4: MRM of 4 Channels lee+ 10 ~ ,03e ,....,....,....,....,....,....,....,.... ~... : 7 ~ ' ; ~ '. ~,.. " ~ : ' ~....,....,...., 3: MRM of 3 Channels ES 10~ ~ , e3... i... ~... i... f... i... i... i... J... i... i... i... i...,... i... i MIP D01 ~0~ > 184,5 1 4 ;6 ~ ~ ~ ' "~.~l ~.~ ~ ~,",, ~... 1~ > 225 MIP ~ : MRM of 4 Chennels ES > %0~ '77 ' LLJ MIP D01 2: M RM Of 4 Channels ES+ 10 P~ > 168 ~:,~ "t 14.~4,340 MIP11GgO3D , ~... ~... ~...,... ~... ~,,"]K"ET~ v. ",~ 7. 9.,... ~...,...,...,...,..., 1003,~ o " " " '.15 1 : MRM of 1 Channel ES O Time (min) Figure 3. Direct injection of drug-free urine. Ion channels as described in Figure 1. Table II. Method Validation Parameters* Precision (%CV) Accuracy (%Bias) Opiate r Inter lntra Inter Intra M3G M6G Morphine Oxyrnorphone Hydromorphone ,4 Norcodeine t Codeine , Oxycodone MAM Hydrocodone * n = 6 for all determinations. at 4500 x g. Urine controls were processed in the same manner. Tuning solutions. Tuning solutions for the LC-MS-MS were made for each compound by preparing serial dilutions of the 1.0 mg/ml stocks to 1 pg/ml in 50:50 acetonitrile and water. A 500-1JL Hamilton gas tight syringe filled with tuning solution was installed in a Harvard Infusion Pump connected into the mobile phase stream via a tee connector. The infusion rate was set at 10 IJL/min into the mobile phase flow at 0.25 ml/min. Final tuning parameters for MRM acquisition are listed in Table I. Ion suppression~matrix effect. An ion suppression experiment was conducted to determine the presence of a matrix effect that would adversely affect the results. In this experiment the infusion pump is connected post-column via a tee connector. The 10,000 ng/ml mixed standard in solvent was diluted 1:100 with 95:5 DI H20/ACN to yield a 100 ng/ml solution. This solution was loaded in the gas-tight syringe, and the infusion rate was set at 10 IJL/min. The baseline was monitored until it became stable. Injections of drug-free urine were then made as normal acquisitions. LC--MS-MS sample analysis. An aliquot of urine was centrifuged in the Eppendorf refrigerated centrifuge at 4500 x g for 10 min at 15~ to pellet large particles; 1.0 ml of sample was then transferred with a micropipette to a 1.5-mL glass autosampler vial containing 10 ljl of IS solution. The vial was capped and vortex mixed. All specimens that exceeded the upper limit of linearity were diluted with drug-free human urine and reanalyzed. GC-MS sample analysis. To 2.0 ml urine calibrators, controls, and diluted patient sample (500 I~L of urine was added to 1.5 ml of distilled water), 20 l~l (33 ng/ljl) of tri-deuterated internal standards of codeine and morphine was added, followed by 75 IJL of [3-glucuronidase and 300 IJL of 2M acetate buffer/10% hydroxylamine mixture. The mixture was vortex mixed and placed in a water bath at 55~ for 2 h to hydrolyze opiate glucuronide conjugates and convert the ketone group on the phenanthrene ring to an oxime to prevent tautomeric shifts. The mixture was removed from the water bath and diluted with 5.0 ml of ph 9.1 buffer (Biochemical Diagnostics Corp., Edgewood, 707
5 NY). Opiates were isolated from urine by solid-phase extraction in a Detectabuse TM Gravity GV-65 column as described by the manufacturer (Biochemical Diagnostics Corp.). The final column wash (acetonitrile/n-butylchloride, 30:70) containing extracted opiates was evaporated to dryness under nitrogen with low heat. The residue was dissolved in 50 pl of ethyl acetate, and the opiates were derivatized by the addition of 50 pl ofn, O-bis(trimethylsilyl)trifluoroacetamide (BSTFA). The mixture was transferred to autosampler vials that were heated in a bead bath at 75~ (+ 20%) for 15 min. One to 3 pl of the mixture was injected into the GC-MS. The GC-MS method was previously validated at Ameritox. The linear range was 80 ng/ml to 20,000 ng/ml with the lower limit of quantitation for each opiate 80 ng/ml. Oxymorphone and 6MAM were not analyzed by this method. Results LC-MS-MS method development and validation Initial development of the method used isocratic conditions in order to decrease retention time and take advantage of the capabilities of the LC-MS-MS to spectrally separate the opiates. Product ions were chosen that provided characteristic fragments of the precursor ion and provided maximal response. Morphine and hydromorphone, however, share common precursors and products ions, and isocratic conditions did not provide sufficient chromatographic separation of morphine and hydromorphone. Therefore, a gradient was instituted that started with 5% organic solvent, held briefly, and then increased to approximately 20% over a short period of time. The final separation is shown in Figure 1. In this system, morphine and oxymorphone are not totally separated in time, but spectrally they were distinguishable. Oxymorphone's parent mass-tocharge ratio is versus for morphine, and they do not share common product ions. Hydromorphone was well separated from morphine and other components. Single product ion transitions were considered acceptable for forensic identification purposes as the urine specimens used in the cross-validation were obtained from a pain management clinic and had been submitted specifically for opiate analysis. In the event of the analysis of an unknown specimen, an additional transition could be added to provide additional confirmation. Selectivity of the method was determined in two ways: injection of multiple drug-free urines specimens obtained from different donors and determination of matrix effects on ion suppression. Ten different urine specimen pools were tested. In the infusion experiment, some ion suppression was noted although not all the compounds may potentially be affected. Figure 2 demonstrates a typical ion suppression pattern for a blank urine sample. A drop in the baseline indicates potential ion suppression. Based upon the retention times of all compounds, hydromorphone, norcodeine, codeine, and oxycodone appear to have the greatest potential for ion suppression issues. Figure 3 depicts the same drug-free urine presented in Figure 2 injected using the MRM conditions. No peaks were observed that would interfere with the analysis of any of the compounds. Name: MIP110903D02 Textr U $1 50/0.25 i: Mor 3-B-glucur..2: Morphine - d3 IS3: Mor 6-B-glucur.4: Morphine 5: 0xymorphone 6: Hydromorphone MIPl10903D02 F11 MIP110903D02 F2 MIP110903D02 F2 MIP110903D02 F; MIP110903D02 F2 I MIP110903D02 F3 100~ 462 > > [ ) > 200! C > @ > q 1.53e:t 1i ',891.48e[ > e,~ ~[ 6.68e3 2.79e4 Are~ / Are~ Are~ %H 0,~,,,~;,;,~ Time ],,,,,,,,.,,,,,,~,, Time %-t 0...,'.-.,... "rime , ,00 %-1 T,me ~ ~ Time I %. 0,.,.,.~..,,,,,,, Time : Norcodeine 8: Codeine 9: Hydrocodone i0: 6-MAM ii: 6-MAM-d6 12: Oxycodone MIP110903D02 F3 MIP110903D02 F3 MIPl10903D02 F4I MIP110903D02 F4 MIP110903D02 F4 I MIP110903D02 F > > i > ~ 334> I > 240.s 100] 1.27e: 1o, l 3c, e 3.88e ~ 5.81e3 I 7.21 e3 Are; Are~ l %4 %-1 %- 0 'nme ~d ~- 9,- Time ,,~,,,,,,Lr~-., Time 0% Time 0 ',... ~... Time 0- ~ Time Figure 4. Direct injection of LOQ urine calibrator containing 0.25 ng/ml 6MAM and 50 ng/ml all other opiates. 708
6 None of the 10 drug-free urine pools tested demonstrated peaks that interfered with analysis of the opiate analytes. All the calibrations of the opiate analytes demonstrated a correlation coefficient (r) greater than 0.99 (Table II). A typical chromatogram of the lowest opiate calibrator is shown in Figure 4. Interassay accuracy and precision were determined from the back-calculated calibrators (Table II). The method demonstrates acceptable accuracy and precision. The greatest bias, 13.60%, was found with the 0.25 ng/ml 6MAM calibrator, and the greatest CV was the norcodeine 50 ng/ml calibrator. All others demonstrated good accuracy and precision. Intrarun accuracy and precision was demonstrated by six replicates at three concentrations, 100 ng/ml, 1000 ng/ml, and 10,000 ng/ml (Table II). The greatest bias was -9.59% with the hydrocodone 1000 ng/ml control. 6MAM at 0.50 ng/ml demonstrated the greatest imprecision at 16.93% CV. The lower limits of quantitation (LOQ) were arbitrarily established at the lowest calibrator for each opiate analyte. At these concentrations, the accepted requirements for LOQ were met: signal-to-noise ratio greater than 10 and precision and accuracy less than 20%. Cross-validation with GC-MS Eighty-nine urine specimens previously analyzed by GC-MS were re-analyzed by LC-MS-MS. The LC-MS-MS results yielded an 88% agreement with GC-MS results for opiate negative samples and 94% agreement for opiate positive samples. The LC-MS-MS method identified additional opiates in 19 specimens opiate positive by GC-MS. Oxymorphone was identified by LC-MS-MS in 10 specimens that the GC-MS method was not set up to detect. However, 9 of these 10 specimens were positive for oxycodone by GC-MS. Because oxymorphone is a metabolite of oxycodone, the presence of the parent opiate was not missed. Similarly hydromorphone was detected by LC-MS-MS in three specimens that were positive for hydrocodone by GC-MS, which can also be explained by hydrocodone metabolism. The remaining LC-MS-MS opiate-positive specimens not detected by GC-MS contained oxymorphone, oxycodone, hydrocodone, and hydromorphone. Quantitatively, it was difficult to compare the two methods because the GC-MS method utilized a glucuronidase step that was not included in the LC-MS-MS method. Therefore, only the morphine results could be compared if the LC-MS-MS results for the glucuronide metabolites were summed on a molar basis with the morphine. A scatter-plot of GC-MS results versus LC-MS-MS results yielded a coefficient of determination (r ~) equal to Assuming 100% hydrolysis, the M3G and M6G (on a molar basis) were added to the morphine value and the scatter-plot was redrawn. The r 2 increased to Discussion With the exception of 6MAM, the method was linear for all opiate analytes from 50 ng/ml to 10,000 ng/ml. 6MAM was linear from 0.25 ng/ml to 50 ng/ml. The r was > 0.99 for all analytes. Accuracy and precision at all concentrations were within 20% for bias and a CV of 20% is a normal accep- tance criterion for bioanalytical methods. LOQ for all opiates but 6MAM was 50 ng/ml; LOQ for 6MAM was 0.25 ng/ml. Dams et al. (12) reported a direct injection LC-MS--MS method using urine as the matrix and several drugs, including opiates were analyzed. Sample preparation consisted of adding internal standard, mixing, centrifuging, and transferring to the LC-MS for analysis. APCI was the interface chosen because of the decreased ion suppression caused by the urine matrix. ESI typically has greater ion suppression in a urine matrix because several endogenous compounds found in urine are polar by nature of the excretion process. The competition of these polar compounds for charge created in the ESI probe is the suspected mechanism of the ion suppression. ESI, however, provides a greater response to ionized compounds then does APCI. Therefore, ESI was chosen for this method to try to provide sufficient sensitivity with direct injection for 6MAM below the 10 ng/ml reported by Dams and colleagues (12). To try to circumvent the ion suppression problem, a high aqueous initial mobile phase coupled with a gradient was used. The high aqueous environment would permit the opiates to be retained on the HPLC column, whereas some of the more polar compounds would be passed directly through the column. After an initial 's~vashing" period, the gradient is started to elute the opiates from the column. Of course this methodology is not selective, but it was hoped that it would simply decrease the ion suppression sufficiently to allow the analysis to occur. Considering that LC-MS-MS methods are routinely reported in the picograms-per-milliliter range, 10 ng/ml is a significant concentration to use. The gradient would further permit the separation of the critical analyte pair of morphine and hydromorphone. The results of the validation demonstrate that the method is usable for the determination of opiates by direct injection of urine into the LC-MS-MS. Stability of the method was not evaluated, however, for the following reasons. It is well documented in the literature that 6MAM is not stable for long periods of time when stored in urine. Rop et al. (13) found that 6MAM stored at 20~ lost 10% within 4 h, and within 24 h, 65% of the original amount was lost. Increasing the pi-i of the sample accelerated the loss of 6MAM. If the samples were stored at -20~ loss was limited to 9%. Additionally, when the samples were treated with ~-glucuronidase to hydrolyze the glucuronide conjugates of morphine, 100% of the 6MAM was lost within 8 h. The loss of morphine under the same conditions (-20~ and hydrolysis) was 0%. Stored methanolic solutions of both morphine and 6MAM were stable for at least 14 days in that study. Stability of methanolic solutions must be longer because Cerilliant, the manufacturer of the standards used in this study, includes one-year expiration dates on their products. Given the ease with which the calibration curves can be prepared, it seems more practical to make the calibration curves for each analysis on an as needed basis. Although this may potentially cause a decrease in the interrun performance of the method, that was not a factor in this validation. To further validate this method, a cross-validation was performed, by analyzing samples previously reported by a GC-MS method. Initially, a poor correlation was found between the two methods. Removing the compounds for which no glucuronide 709
7 standards were available improved the r 2. Despite the correction used by summing morphine and its glucuronide metabolites, some samples still were vastly different between GC-MS and LC-MS-MS. For example, one sample by GC-MS was morphine positive with 18,350 ng/ml and LC-MS-MS positive with 37,685 ng/ml. This large difference cannot be explained, but it appears to be reproducible. Only one other direct injection method for urine using LC-MS-MS has been reported (12). APCI interface was used, however, although similar LOQs and upper limits of linearity were reported. The LOQ for 6MAM at 0.25 ng/ml for our method is far superior to the 10 ng/ml reported by Dams et al. (12). In that report, the glucuronide metabolites of morphine, hydromorphone, hydrocodone, oxycodone, norcodeine, and oxymorphone were not included in the method. Normorphine, heroin, and acetycodeine were included, and our method does not analyze for those opioids. Another advantage of this new method is shorter run times. In this method, all analyses are complete within 6.00 rain. In the Dams et al. (12) report, the last opioid is eluted at 19.3 rain. Conclusions A method for the analysis to 10 opiates by direct injection of urine and detection by LC-MS-MS has been developed. The method is simple and provides a rapid analysis of the 10 opiate analytes included. The LOQ, linear range, precision, and accuracy validated for this method would permit its use for a variety of bioanalytical applications. The LOQ for 6MAM, 0.25 ng/ml, presents an improvement in previously reported limits for this analyte. References 1. R. Pacifici, S. Pichini, I. Altieri, A. Caronna, A.R. Passa, and P. Zuccaro. High-performance liquid chromatographic-electrospray mass spectrometric determination of morphine and its 3- and 6-glucuronides:application to pharmacokinetic studies. J. Chromatogr. B 664: (1995). 2. N. Tyrefors, B. Hyllbrant, L. Ekman, M. Johansson, and B. Langstrom. Determination of morphine, morphine-3-glucuronide and morphine-6-glucuronide in human serum by solidphase extraction and liquid chromatography-mass spectrometry wih electrospray ionisation. J. Chromatogr. 729: (1996). 3. W. Naidong, J.W. Lee, X. Jiang, M. Wehling, J.D. Hulse, and P.P. Lin. Simultaneous assay of morphine, morphine-3-glucuronide and morphine-6-glucuronide in human plasma using normal-phase liquid chromatography-tandem mass spectrometry with a silica column and an aqueous organic mobile phase. J. Chromatogr. B 735: (1999). 4. G. Schanzle, S. Li, G. Mikus, and U. Hofmann. Rapid, highly sensitive method for the determination of morphine and its metabolites in body fluids by liquid chromatography-mass spectrometry. J. Chromatogr. B 721 : (1999). 5. W.Z. Shou, M. Pelzer, T. Addison, X. Jiang, and W. Naidong. An automated 96-well solid phase extraction and liquid chromatography-tandem mass spectrometry method for the analysis of morphine, morphine-3-glucuronide and morphine-6-glucuronide in human plasma. J. Pharm. Biomed. Anal. 27: (2002). 6. M. Dawson, F. Bronwyn, T. Kelly, J. Keegan, and L.E. Mather. A rapid and sensitive high-performance liquid chromatographyelectrospray ionization-triple quadrupole mass spectrometry method for the quantitation of oxycodone in human plasma. ]. Chromatogr. 5ci. 40:40-44 (2002). 7. Y.L. Chen, G.D. Hanson, X. Jiang, and W. Naidong. Simultaneous determination of hydrocodone and hydromorphone in human plasma by liquid chromatography with tandem mass spectrometric detection. J. Chromatogr. B 769(1): (2002). 8. M.J. Bogusz, R.D. Maier, and S. Driessen. Morphine, morphine- 3-glucuronide, morphine-6-glucuronide, and 6-monoacetylmorphine determined by means of atmospheric pressure chemical ionization-mass spectrometry-liquid chromatography in body fluids of heroin victims. J. Anal. Toxicol. 21: (1997). 9. M.J. Bogusz, R.D. Maier, M. Erkens, and S. Driessen. Determination of morphine and its 3- and 6-glucuronides, codeine, codeine-glucuronide and 6-monoacetylmorphine in body fluids by liquid chromatography atmospheric pressure chemical ionization mass spectrometry. J. Chromatogr. B 703: (1997). 10. R.L. Fitzgerald, J.D. Rivera, and D.A. Herold. Broad spectrum drug identification directly from urine, using liquid chromatography-tandem mass spectrometry. Clin. Chem. 45" (1999). 11. R. Dams, C.M. Murphy, R.E. Choo, W.E. Lambert, A.P. De Leenheer, and M.A. Huestis. LC-atmospheric pressure chemical ionization-ms/ms analysis of multiple illicit drugs, methadone, and their metabolites in oral fluids following protein precipitation. Anal. Chem. 75: (2003). 12. R. Dams, C.M. Murphy, W.E. Lambert, and M.A. Huestis. Urine drug testing for opioids, cocaine, and metabolites by direct injection liquid chromatography/tandem mass spectrometry. Rapid Commun. Mass Spectrom. 17: (2003). 13. P.P. Rop, F. Grimaldi, J. Burle, M.N. De Saint Leger, and A. Viala. Determination of 6-monoacetylmorphine and morphine in plasma, whole blood and urine using high-performance liquid chromatography with electrochemical detection. J. Chromatogr. B 661 : (1994). 710
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with the sensitivity and selectivity
One Source Toxicology Laboratory, 1213 Genoa Red Bluff, Pasadena, Texas 77504
Validation of Analysis of Amphetamines, Opiates, Phencyclidine, Cocaine, and Benzoylecgonine in Oral Fluids by Liquid Chromatography Tandem Mass Spectrometry Subbarao V. Kala*, Steve E. Harris, Tom D.
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
Matrix Effect in Bio-Analysis of Illicit Drugs with LC-MS/MS: Influence of Ionization Type, Sample Preparation, and Biofluid
Matrix Effect in Bio-Analysis of Illicit Drugs with LC-MS/MS: Influence of Ionization Type, Sample Preparation, and Biofluid Riet Dams and Marilyn A. Huestis Chemistry and Drug Metabolism, National Institute
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT
AppNote 1/2014 Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine Fred D. Foster, John R. Stuff, Edward A. Pfannkoch GERSTEL, Inc., 701 Digital Dr. Suite J,
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis
Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis 1.0 Purpose - This procedure specifies the required elements for the solid phase extraction of acidic,
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
TECHNICAL NOTE Toxicology Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with
Overview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS
AppNote 1/2012 Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS Oscar G. Cabrices, Fred D. Foster, John R. Stuff, Edward
Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
Opiates in Urine by SAMHSA GC/MS
application Note Gas Chromatography/ Mass Spectrometry Author Timothy D. Ruppel PerkinElmer, Inc. Shelton, CT 06484 USA Opiates in Urine by SAMHSA GC/MS Introduction The United States Department of Health
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System
Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection
Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection Derek Hillbeck, Thermo Fisher Scientific, Runcorn, UK Application Note 2999 Key
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine
Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine Application Note Authors Peter JW Stone and Kevin McCann Agilent Technologies, Inc. Santa Clara,
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
Oasis HLB Cartridges and 96-Well Plates
CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
Riet Dams,, Constance M. Murphy,*, Robin E. Choo, Willy E. Lambert, Andre P. De Leenheer, and Marilyn A. Huestis
Anal. Chem. 2003, 75, 798-804 LC-Atmospheric Pressure Chemical Ionization-MS/ MS Analysis of Multiple Illicit Drugs, Methadone, and Their Metabolites in Oral Fluid Following Protein Precipitation Riet
Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS
Therapeutic Drug Monitoring of Antiretroviral Drugs with PLC-M Ursula Gutteck-Amsler, Katharina M. Rentsch Abstract Prospective and retrospective studies have provided some evidence of the clinical and
Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.
Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection Application Note Food Testing & Agriculture Authors Edgar Naegele and Thomas
CORESTA RECOMMENDED METHOD N 72
CORESTA RECOMMENDED METHOD N 72 DETERMINATION OF TOBACCO-SPECIFIC NITROSAMINES IN SMOKELESS TOBACCO PRODUCTS BY LC-MS/MS (February 216). INTRODUCTION The CORESTA Smokeless Tobacco Sub-Group studied various
Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum
Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in uman Serum Technical verview Clinical Research Authors Anthony Macherone, Ph.D. Agilent Technologies,
Interlaboratory Studies on the Analysis of Hair for Drugs of Abuse: Results from the Fourth Exercise
Interlaboratory Studies on the Analysis of Hair for Drugs of Abuse: Results from the Fourth Exercise Lorna T. Sniegoski and Michael J. Welch Analytical Chemistry Division, Chemical Science and Technology
SCREENING FOR DRUGS IN SERUM AND URINE BY LC/ESI/CID-MS AND MS/MS WITH LIBRARY SEARCHING *
SCREENING FOR DRUGS IN SERUM AND URINE BY LC/ESI/CID-MS AND MS/MS WITH LIBRARY SEARCHING * Wolfgang WEINMANN 1, Natalie LEHMANN 1, Michaela RENZ 1, Alexander WIEDEMANN 1, Michal SVOBODA 2 1 Institute of
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
Determination of Tapentadol and its Metabolite N
Determination of Tapentadol and its Metabolite N-Desmethyltapentadol in Urine and Oral Fluid using Liquid Chromatography with Tandem Mass Spectral Detection Cynthia Coulter, Margaux Taruc, James Tuyay,
Supplementary Materials and Methods (Metabolomics analysis)
Supplementary Materials and Methods (Metabolomics analysis) Metabolite extraction and mass spectrometry analysis. 12 Vldlr-/- knock out and 12 wild type (WT) retinas were separately frozen at -80 ºC in
Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report
of melamine-ware by GC-MS and LC-MS/MS Page 1 of 15 Method development for analysis of formaldehyde in foodsimulant extracts of melamine-ware by GC-MS and LC-MS/MS December 2012 Contact Point: Chris Hopley
AppNote 6/2011. Determination of Pain Management Drugs using Automated Disposable Pipette Extraction and LC-MS/MS KEYWORDS ABSTRACT
AppNote 6/2011 Determination of Pain Management Drugs using Automated Disposable Pipette Extraction and LC-MS/MS Fred D. Foster, John R. Stuff, Edward A. Pfannkoch Gerstel, Inc., 701 Digital Dr. Suite
Improved Method for the Analysis of 31 Drugs of Abuse in Oral Fluid samples using the Thomson extreme FV by LC-MS/MS
htslabs.com [email protected] 800 541.4792 760 757.8080 760 757.9367 Improved Method for the Analysis of 31 Drugs of Abuse in Oral Fluid samples using the Thomson extreme FV by LC-MS/MS Nadine Koenig 2,
Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs
Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan
AppNote 7/2009. Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS KEYWORDS ABSTRACT
AppNote 7/2009 Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS Fred D. Foster, John R. Stuff, Edward A. Pfannkoch Gerstel, Inc., 701 Digital Dr. Suite J, Linthicum,
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Micromass LCT User s Guide
Micromass LCT User s Guide 1) Log on to MassLynx with your username & password. 2) After you have logged in, the MassLynx software will automatically run. 3) After MassLynx has come up, open your project
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,
APPLICATIONS MANUAL. Ifosfamide in blood serum... 2. Ingredients in blood serum... 3. Organophosphorus pesticides in tea leaf... 5
1 APPLICATIONS MANUAL CONTENTS Ifosfamide in blood serum... 2 Ingredients in blood serum... 3 Organophosphorus pesticides in tea leaf... 5 Sudan i ii iii iv in chilli sauce... 6 Pah in water... 7 Phenols
Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS
Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS Solfrid Hegstad, PhD Department of Clinical Pharmacology St.Olavs Hospital St.Olavs Hospital Department of Clinical Pharmacology
AppNote 4/2014. Prep-and-Shoot : The Completely Automated Workfl ow for the Hydrolysis and Analysis of Urine Samples by LC/MS/MS.
AppNote 4/2014 Prep-and-Shoot : The Completely Automated Workfl ow for the Hydrolysis and Analysis of Urine Samples by LC/MS/MS. Oscar G. Cabrices, Fred D. Foster Gerstel, Inc., 701 Digital Dr. Suite J,
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
Instrumentation. Sample pretreatment
Analysis of Propofol (2,6-Diisopropylphenol) and its Metabolites in One Injection using the DuoSpray Ion Source AB SCIEX QTRAP 500 LC-MS/MS System and Gerstel, Inc. MultiPurpose Sampler Adrian M. Taylor,
SIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS
SIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS Russell Watts 1, Allan Traynor 2, Michelle Wood 1. 1 Waters Corporation, MS Technologies Centre,
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS Mikaël LEVI Shimadzu France Welcome Seminar Barcelona 2013 Interest in Steroids assay by LC-MS/MS Steroids measurement is essential
Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow
Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow Abstract There is high demand in clinical research and forensic toxicology for comprehensive,
Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
SCREENING FOR THE PRESENCE OF PARA-METHYLTHIOAMPHETAMINE IN URINE BY SOME COMMERCIAL IMMUNOASSAYS AND CONFIRMATION BY GC/MS
SCREENING FOR THE PRESENCE OF PARA-METHYLTHIOAMPHETAMINE IN URINE BY SOME COMMERCIAL IMMUNOASSAYS AND CONFIRMATION BY GC/MS Ingrid J. BOSMAN 1, Douwe DE BOER 2, Robert A. A. MAES 1 1 Department of Human
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2
Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2 1 OpAns LLC, Durham, NC, USA; 2 Ionics Mass Spectrometry
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection LGC/R/2011/181 LGC Limited 2011 GC11/0005A A feasibility study of the determination of B-group
Sex Hormone Testing by Mass Spectrometry
Sex Hormone Testing by Mass Spectrometry Robert L. Fitzgerald, PhD, DABCC Professor of Pathology University of California-San Diego San Diego, CA, 92161 [email protected] Learning Objectives After this
Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance
Thermo Scientific SOLA SPE cartridges and plates Technical Guide Join the revolution unparalleled performance Join the revolution next-generation SPE Thermo Scientific SOLA products revolutionize Solid
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
Introduction. Methods. Sample Processing
PO-CON1656 High-sensitivity, high-throughput quantitation of catecholamines and metanephrine in plasma by automated WCX-SP coupled to LC/MS/MS for clinical research ASMS 2016 MP-080 Ichiro HIRANO 1, Atsuhiko
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
Rapid LC/MS/MS Method Development for Drug Discovery
Rapid LC/MS/MS Method Development for Drug Discovery A three-checkpoint paradigm offers a way to rapidly develop a very reliable bioanalytical method when LC/MS/MS is used in the drug discovery process.
United States Department of Agriculture Food Safety and Inspection Service, Office of Public Health Science
CLG-MG/CV2.00 Page 1 of 15 Contents A. INTRODUCTION... 2 B. EQUIPMENT... 3 C. REAGENTS AND SOLUTIONS... 4 D. STANDARDS... 5 E. SAMPLE PREPARATION... 6 F. ANALYTICAL PROCEDURE... 6 G. CONFIRMATION... 8
MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process
- Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
Expectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation
Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Craig Aurand 1, David Bell 1, Anders Fridstrom 2, and Klaus Buckendahl 2 1 Supelco, Div.
Electrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
What Do We Learn about Hepatotoxicity Using Coumarin-Treated Rat Model?
What Do We Learn about Hepatotoxicity Using Coumarin-Treated Rat Model? authors M. David Ho 1, Bob Xiong 1, S. Stellar 2, J. Proctor 2, J. Silva 2, H.K. Lim 2, Patrick Bennett 1, and Lily Li 1 Tandem Labs,
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
The basic division explosives for forensic purposes is shown in Fig. 1
Isolation, concentration and determination of post blast residua by liquid chromatography with DAD detector capillary electrophoresis with DAD detector Firearm fumes (gunshot residua, GSR) - various metallic
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis