1 Test Procedure for VERIFYING CALIBRATION OF WEIGHING DEVICES USED FOR Texas Department of Transportation TxDOT Designation: Tex-901-K Effective Date: August SCOPE 1.1 All weighing devices shall meet the requirements of the test procedure using it. Some general-purpose scales and balances may not have sufficient accuracy for many tests. This will be especially true when large quantities of material are to be weighed to high degrees of accuracy. The test procedure may specify requirements for the weighing device, such as the capacity, accuracy, sensitivity, and readability. 1.2 Many laboratory balances have more than one range. Each range will have a different accuracy, readability, and sensitivity. The user must understand the test requirements and the capabilities of each weighing device. 1.3 The values given in parentheses (if provided) are not standard and may not be exact mathematical conversions. Use each system of units separately. Combining values from the two systems may result in nonconformance with the standard. 2. REFERENCE DOCUMENTS 2.1 AASHTO M ASTM E CLASSES OF WEIGHING DEVICES 3.1 Weighing devices are divided into two groups based on accuracy, readability, and sensitivity. Analytical balances are further divided into two classes (See Table 1) and general-purpose balances into five classes based on accuracy (See Table 2). 3.2 Analytical Balances: Analytical balances should conform to the accuracy, sensitivity, and readability requirements of a class specified in Table Place a label on each balance stating the class of the balance in a location easily seen by the user. CONSTRUCTION DIVISION 1 5 EFFECTIVE DATE: AUGUST 1999

2 3.2.3 Table 1 lists weight classes used for calibration, either as external or internal balance weights mechanically selected by the operator Verify counter weights in accordance with Tex-900-K. 3.3 General-Purpose Balances and Scales: General-purpose balances and scales should conform to the accuracy, sensitivity, and readability requirements of Table Place a label on each balance stating the class of the balance in a location easily seen by the user Select general-purpose weighing devices according to the sample weight. Do not use the weighing device for loads less than ten percent of the capacity of the range in use. Select a weighing device of a lesser capacity for small loads. 4. VERIFYING ACCURACY 4.1 Verify accuracy at a minimum of three points for each range or beam-arm of the weighing device, with a minimum of five points for the device. Checking more points may be necessary to determine if the instrument is accurate over its full range. This is especially true of two-pan weighing devices used with a variety of counter weights and with multiple beam-arm mechanical devices. 4.2 Apply verification test loads in order of increasing value. 4.3 For weighing devices with multiple ranges or beam-arms, verify each range and beamarm, at a minimum, with test loads of approximately 10, 50, and 100% of the range or capacity. The test load may be values presented by a minimum number of weights, not the exact value calculated from the capacity. EXAMPLE: For a range of 600 g capacity, the approximate 10 and 50% loads could be 50 and 300 g. Use a single 50 g weight and two or three weights to make the 300 g load. 4.4 For single range weighing devices, the minimum number of verification points should be at loads of approximately 10, 20, 50, 75, and 100% of the device capacity. The actual test load should be values presented by a minimum number of weights, not the exact value calculated from the capacity. EXAMPLE: For a scale of 4600 g capacity, the approximate 10% and 50% loads would NOT be 460 g and 2300 g. Use a single 500 g weight and several weights to make a 2500 g load. Verification at more than these five points is desirable. 4.5 Verify beam-arm hanger weights individually. Use a beam-arm slider to determine the hanger weight's deviation. CONSTRUCTION DIVISION 2 5 EFFECTIVE DATE: AUGUST 1999

3 4.6 Procedure: Clean the platform or pan, beams, display, etc. as needed. On laboratory balances, be sure that no debris or portion of a plastic cover restricts pan movement Level the weighing device If the weighing device has a platform or beam lock, use the lock each time placing a weight on the pan. When using an analytical balance with a lock, lock the balance before applying or removing any weight With the lock off, zero the load indicator or balance the beam at no load Make a test weighing with a load of from 50 to 100% of the capacity. If the indicated load is out of tolerance, record the values in an "as found" column. Adjust the device to meet the required accuracy When the test load is within specifications, verify the device on all ranges. Record the values in the "as found" column if the device was not adjusted and in the "adjusted" column if adjustments were made Report the weighing device as failing to meet the accuracy specifications if any range cannot meet the accuracy requirements for the class best describing that range Report the weighing device as failing to meet the specifications if it fails to operate properly (according to the manufacturer's instruction manual) on all ranges. 5. VERIFYING ECCENTRIC LOAD ACCURACY 5.1 Verify the eccentric or off-center load accuracy at a load of approximately 50% of the capacity of the weighing device. 5.2 Check additional load points if the device is suspected of being inaccurate over some portion of its load range. 5.3 Use the least number of weights practical for this test load. EXAMPLE: For a Torsion two-pan balance of 4500 g capacity, use a 2 kg (4.4 lb.) weight or stack two 1 kg (2.2 lb.) weights. 5.4 Compare the indicated value of the test load at the center of the loading pan or platform with the indicated value with the loads placed one-half of the distance from the center of the pan to the edge. Perform this test at four off-center points: one toward the operator one away from the operator one to the left, and one to the right. CONSTRUCTION DIVISION 3 5 EFFECTIVE DATE: AUGUST 1999

4 5.5 Procedure: Level the weighing device If the weighing device used has a platform or beam lock, lock the device before applying any weights. When using an analytical balance with a lock, lock the balance before applying or removing any weights With the lock off, zero the load indicator or balance the beam at no load Place the test load on the center of the pan or platform of the device and note the indicated weight Remove the load, zero the instrument (reset the balance), and then apply the load at one of the off-center loading points. Note the indicated weight. Repeat for the remaining three off-center loading points Report the weighing device as out of specification if the difference between the indicated weight for the test load centered and that for any off-center load position exceeds the tolerance for its class. 6. VERIFYING SENSITIVITY 6.1 Sensitivity tests determine the minimum weight change that the weighing device can detect. 6.2 Procedure: Level the weighing device With the lock off, zero the load indicator or balance the beam at no load Place a load of about 3/4 of the device capacity on the pan or platform and add small weights to the load until the weighing device detects a change in the reading. Do not lock the beam or platform while performing this part of the test. Note the minimum value of weights to affect a change in reading If the minimum change in the load to effect a change in the indicated load is greater than the required accuracy, report the device as failing the sensitivity test. 7. CERTIFICATION 7.1 Place a certification label on those balances or scales that meet the class requirements in this procedure Place the label so that the user can see it easily The serial number of the balance or scale should be on the label. CONSTRUCTION DIVISION 4 5 EFFECTIVE DATE: AUGUST 1999

5 7.1.3 The technician performing the verification should sign and date the label The technician should verify that a label states the correct class of the weighing device. 8. CALIBRATION REPORT 8.1 Issue a calibration report containing the following: Weighing device, brand, serial number, capacity, and class Whether the weighing device passes or fails the requirements for the stated class the cause of failure, if the weighing device fails. 9. TABLE REFERENCES Table 1 Classes of Analytical Balances Class Accuracy in grams Sensitivity and Readability in grams Class of Weight for Calibration Max. Load in grams (a) (b) (c) A ,2,3 200 B (a) Readability of digital readout or graduations on mechanical balances. (b) Suggested class of calibration weight (Tex-900-K, Table 1) (c) Approximate maximum capacity generally available with balances of this class. Table 2 Classes of General Purpose Balances and Scales Balance Class Readability and Sensitivity in grams Accuracy in grams (a) G ± 0.1 % G ± 0.1 % G ± 0.1 % G ± 0.1 % G ± 0.1 % (a) Accuracy throughout the range of use shall be equal to or less than the mass stated or 0.1 % of the test load, whichever is greater. CONSTRUCTION DIVISION 5 5 EFFECTIVE DATE: AUGUST 1999

Testing Your Laboratory Balance

Testing Your Laboratory Balance Testing Your Laboratory Balance If you use a laboratory balance in your day-to-day work, you may have wondered whether it is accurate. Here are four tests that you can use to confirm your instrument's

More information

Estimates of Uncertainty of the Calibration of Balances

Estimates of Uncertainty of the Calibration of Balances Estimates of Uncertainty of the Calibration of Balances Chart shows the uncertainty (estimates) of the calibration of representative electronic balances. Chart. Uncertainty of calibration (estimates) Minimum

More information

PART I SIEVE ANALYSIS OF MATERIAL RETAINED ON THE 425 M (NO. 40) SIEVE

PART I SIEVE ANALYSIS OF MATERIAL RETAINED ON THE 425 M (NO. 40) SIEVE Test Procedure for PARTICLE SIZE ANALYSIS OF SOILS TxDOT Designation: Tex-110-E Effective Date: August 1999 1. SCOPE 1.1 This method covers the quantitative determination of the distribution of particle

More information

Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance

Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance INL/CON-10-18643 PREPRINT Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance 2010 NCSL International Workshop and Symposium Michael Stears July 2010 This is

More information

DETERMINING ASPHALT CONTENT FROM ASPHALT PAVING MIXTURES BY THE IGNITION METHOD

DETERMINING ASPHALT CONTENT FROM ASPHALT PAVING MIXTURES BY THE IGNITION METHOD Test Procedure for DETERMINING ASPHALT CONTENT FROM ASPHALT PAVING MIXTURES BY THE IGNITION METHOD TxDOT Designation: Tex-236-F Effective Date: March 2016 1. SCOPE 1.1 Use this test method to determine

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

NITP 6.1 to 6.4 National Instrument Test Procedures for Non-automatic Weighing Instruments

NITP 6.1 to 6.4 National Instrument Test Procedures for Non-automatic Weighing Instruments NITP 6.1 to 6.4 National Instrument Test Procedures for Non-automatic Weighing Instruments Commonwealth of Australia 2012 NSC V 1 First edition 2000 Second edition October 2002 Third edition February 2004

More information

Measuring equipment in a laboratory.

Measuring equipment in a laboratory. Measuring equipment in a laboratory. Laboratory and metrology are two words that practically do not function separately. They are integrally connected to measuring equipment for tests and measurements.

More information

METHOD OF TEST FOR UNIT WEIGHT OF FRESH CONCRETE

METHOD OF TEST FOR UNIT WEIGHT OF FRESH CONCRETE STATE OF CALIFORNIA BUSINESS, TRANSPORTATION AND HOUSING AGENCY DEPARTMENT OF TRANSPORTATION DIVISION OF ENGINEERING SERVICES Transportation Laboratory 5900 Folsom Blvd. Sacramento, California 95819-4612

More information

SIEVE ANALYSIS OF FINE AND COARSE AGGREGATES

SIEVE ANALYSIS OF FINE AND COARSE AGGREGATES Test Procedure for SIEVE ANALYSIS OF FINE AND COARSE AGGREGATES TxDOT Designation: Tex-200-F Effective Date: January 2016 1. SCOPE 1.1 Use this test method to determine the particle size distribution of

More information

METHOD OF TEST FOR SAMPLING AND TESTING CRUMB RUBBER MODIFIER

METHOD OF TEST FOR SAMPLING AND TESTING CRUMB RUBBER MODIFIER STATE OF CALIFORNIA BUSINESS, TRANSPORTATION AND HOUSING AGENCY DEPARTMENT OF TRANSPORTATION DIVISION OF ENGINEERING SERVICES Transportation Laboratory 5900 Folsom Boulevard Sacramento, California 95819-4612

More information

User checks and maintenance of laboratory balances

User checks and maintenance of laboratory balances Technical Note 13 March 2014 Issued: July 1995 Amended and reissued: July 2005, July 2007, October 2007, May 2010, March 2013, March 2014 User checks and maintenance of laboratory balances Copyright National

More information

COMPACTING BITUMINOUS SPECIMENS USING THE SUPERPAVE GYRATORY COMPACTOR (SGC)

COMPACTING BITUMINOUS SPECIMENS USING THE SUPERPAVE GYRATORY COMPACTOR (SGC) Test Procedure for COMPACTING BITUMINOUS SPECIMENS USING THE SUPERPAVE TxDOT Designation: Tex-241-F Effective Date: December 2015 1. SCOPE 1.1 Use this test method to: compact cylindrical specimens of

More information

PREPARATION. INTERNAL CALIBRATION (No external weight needed.) SPAN CALIBRATION

PREPARATION. INTERNAL CALIBRATION (No external weight needed.) SPAN CALIBRATION Calibration Guide PREPARATION Adventurer Pro Adventurer Pro balances have two calibration methods: Span and Linearity. Some models have Internal Calibration (InCal ) for Span calibration. Ensure that the

More information

Figure 1. A typical Laboratory Thermometer graduated in C.

Figure 1. A typical Laboratory Thermometer graduated in C. SIGNIFICANT FIGURES, EXPONENTS, AND SCIENTIFIC NOTATION 2004, 1990 by David A. Katz. All rights reserved. Permission for classroom use as long as the original copyright is included. 1. SIGNIFICANT FIGURES

More information

Calibration Service Selection Guide

Calibration Service Selection Guide Calibration Service Selection Guide Certification of Accreditation available for download at www.ricelake.com/accreditation. Answer Questions to determine the calibration service you need Need accredited*

More information

Procedure for Equipment Calibration and Maintenance

Procedure for Equipment Calibration and Maintenance Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory

More information

Lab 1: The metric system measurement of length and weight

Lab 1: The metric system measurement of length and weight Lab 1: The metric system measurement of length and weight Introduction The scientific community and the majority of nations throughout the world use the metric system to record quantities such as length,

More information

QUALIFIED TECHNICIAN PROGRAM

QUALIFIED TECHNICIAN PROGRAM VERMONT AGENCY OF TRANSPORTATION QUALIFIED TECHNICIAN PROGRAM William E. Ahearn, P.E. Materials and Research Engineer March 1, 2010 1.0 INTRODUCTION This document has been developed to outline the minimum

More information

SSI. Modbus Analogue Output DEVICENET ETHERNETIP. Zero Slip, Zero Wear. Exact Measurement. No Moving Parts. Permanently Calibrated.

SSI. Modbus Analogue Output DEVICENET ETHERNETIP. Zero Slip, Zero Wear. Exact Measurement. No Moving Parts. Permanently Calibrated. Contact Wheel / Encoder Counter Length & Speed Errors through slippage and wear, result in Short Lengths and Give Away. Maintenance Costs, through calibration downtime and replacement parts. Marking and

More information

Weight Calibration ComparatorPac TM Simple Calibration Efficient Workflow Reliable Results

Weight Calibration ComparatorPac TM Simple Calibration Efficient Workflow Reliable Results Weight Calibration ComparatorPac TM Simple Calibration Efficient Workflow Reliable Results Safe and Simple Weight Calibration with ComparatorPac TM ComparatorPac Weight Calibration for Traceable Results

More information

Inspections: Central Laboratory of the Regions

Inspections: Central Laboratory of the Regions Page 1 Inspections: Central Laboratory of the Regions Protocol for Open Laboratory Review of a CDOT Materials Testing Laboratory Protocol for the Inspection of Region Materials Laboratories by the Central

More information

PT-MT3 Magnetic Tablet for dynamic calibration and perodical qualification of Tablet Harndness Testing Instruments

PT-MT3 Magnetic Tablet for dynamic calibration and perodical qualification of Tablet Harndness Testing Instruments PT-MT3 Magnetic Tablet for dynamic calibration and perodical qualification of Tablet Harndness Testing Instruments What is a PT-MT3 and why to use it? This instrument represents a system for calibrating

More information

Assessment of Accuracy and Precision

Assessment of Accuracy and Precision 2 chapter Assessment of Accuracy and Precision S.S. Nielsen, Food Analysis Laboratory Manual, Food Science Texts Series, DOI 10.1007/978-1-4419-1463-7_2, Springer Science+Business Media, LLC 2010 9 Chapter

More information

Grammage of paper and paperboard (weight per unit area) (Revision of T 410 om-08) (underscores and strikeouts indicate changes from Draft 1)

Grammage of paper and paperboard (weight per unit area) (Revision of T 410 om-08) (underscores and strikeouts indicate changes from Draft 1) NOTICE: This is a DRAFT of a TAPPI Standard in ballot. Although available for public viewing, it is still under TAPPI s copyright and may not be reproduced or distributed without permission of TAPPI. This

More information

Apr 17, 2000 LAB MANUAL 1302.0. 1302 PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified)

Apr 17, 2000 LAB MANUAL 1302.0. 1302 PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified) Apr 17, 2000 LAB MANUAL 1302.0 1302 PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified) 1302.1 SCOPE This method describes a procedure for the quantitative determination of the distribution

More information

Quality Management System. Manual

Quality Management System. Manual Quality Manual ISO 9001 Quality Approved by: JTH Page 1 of 7 Introduction The adoption of a Quality is a strategic decision of an organization. Industrial Scale Ltd. has instituted a formal Quality as

More information

SOIL-LIME TESTING. Test Procedure for. TxDOT Designation: Tex-121-E 1. SCOPE 2. APPARATUS 3. MATERIALS TXDOT DESIGNATION: TEX-121-E

SOIL-LIME TESTING. Test Procedure for. TxDOT Designation: Tex-121-E 1. SCOPE 2. APPARATUS 3. MATERIALS TXDOT DESIGNATION: TEX-121-E Test Procedure for SOIL-LIME TESTING TxDOT Designation: Tex-121-E Effective Date: August 2002 1. SCOPE 1.1 This method consists of three parts. 1.1.1 Part I determines the unconfined compressive strength

More information

Calibration Certificate

Calibration Certificate P.O. Box 87-0087 Phone:856-686-1600 Fax:856-686-1601 www.troemner.com e-mail: troemner@troemner.com SECTION 1: Page 1 of 7 Pages SECTION 2: APPROVED SIGNATORY Joseph Moran SECTION 3: PERSON PERFORMING

More information

Spreadsheets and Laboratory Data Analysis: Excel 2003 Version (Excel 2007 is only slightly different)

Spreadsheets and Laboratory Data Analysis: Excel 2003 Version (Excel 2007 is only slightly different) Spreadsheets and Laboratory Data Analysis: Excel 2003 Version (Excel 2007 is only slightly different) Spreadsheets are computer programs that allow the user to enter and manipulate numbers. They are capable

More information

Supplementary Certificate of Approval NMI S669

Supplementary Certificate of Approval NMI S669 Bradfield Road, West Lindfield NSW 2070 Supplementary Certificate of Approval Issued by the Chief Metrologist under Regulation 60 of the National Measurement Regulations 1999 This is to certify that an

More information

Good Weighing Practice In Pharmaceutical Industry. Pharmacy Risk Assessment For Mass Measurement Processes - 1 -

Good Weighing Practice In Pharmaceutical Industry. Pharmacy Risk Assessment For Mass Measurement Processes - 1 - Good Weighing Practice In Pharmaceutical Industry Pharmacy Risk Assessment For Mass Measurement Processes - 1 - Content Introduction...3 2. Measurement errors...4 2.1. Gross error...4 2.2. Systematic error...4

More information

WIKA INSTRUMENT CORPORATION

WIKA INSTRUMENT CORPORATION WIKA INSTRUMENT CORPORATION Instruction Manual GAUGE PRESSURE INDICATOR Series 1500 Series 1000 Series 300 WIKA Instrument Corporation 1000 Wiegand Boulevard Lawrenceville, GA 30043 1-888-945-2872 http://www.wika.com

More information

Calibration of Glassware

Calibration of Glassware Calibration of Glassware Introduction Glassware is commonly calibrated using a liquid of known, specific density, and an analytical balance. The procedure is to determine the mass of liquid the glassware

More information

Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less

Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less 3A2 008 ISTA 3 Series General Simulation Performance Procedure VERSION DATE Last TECHNICAL Change: FEBRUARY 2008 Last EDITORIAL

More information

The Analytical Balance

The Analytical Balance Chemistry 119: Experiment 1 The Analytical Balance Operation of the Single-Pan Analytical Balance Receive instruction from your teaching assistant concerning the proper operation of the Sartorius BP 210S

More information

MINIMUM AUTOMOTIVE QUALITY MANAGEMENT SYSTEM REQUIREMENTS FOR SUB-TIER SUPPLIERS

MINIMUM AUTOMOTIVE QUALITY MANAGEMENT SYSTEM REQUIREMENTS FOR SUB-TIER SUPPLIERS MINIMUM AUTOMOTIVE QUALITY MANAGEMENT SYSTEM REQUIREMENTS FOR SUB-TIER SUPPLIERS CONTENTS 1. CONTROL PLANS 2. PROCESS APPROACH 3. PERFORMANCE 4. INTERNAL AUDITING 5. CONTROL OF NON-CONFORMING PRODUCT 6.

More information

Measurement and Calibration

Measurement and Calibration Adapted from: H. A. Neidig and J. N. Spencer Modular Laboratory Program in Chemistry Thompson Learning;, University of Pittsburgh Chemistry 0110 Laboratory Manual, 1998. Purpose To gain an understanding

More information

A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion

A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion Objective In the experiment you will determine the cart acceleration, a, and the friction force, f, experimentally for

More information

Measuring instruments in industry.

Measuring instruments in industry. Measuring instruments in industry. Measuring instruments that are present in any branch of industry, are subject to different levels of supervision, like in case of laboratories. Supervision relates to

More information

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement Chapter 2 Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement 2.1 Introduction and Objectives This laboratory exercise involves the static calibration

More information

My Weigh Triple Beam Balance User Manual

My Weigh Triple Beam Balance User Manual My Weigh Triple Beam Balance User Manual Troubleshooting: Before troubleshooting please take these words of "Balance Beam Wisdom": Balance Beams are a very old and basic design of scale. They date back

More information

DryWeight BulkVolume

DryWeight BulkVolume Test Procedure for BULK SPECIFIC GRAVITY AND WATER ABSORPTION OF AGGREGATE TxDOT Designation: Tex-201-F Effective Date: January 2016 1. SCOPE 1.1 Use this method to determine the bulk specific gravity

More information

RTP s NUCLEAR QUALITY ASSURANCE PROGRAM

RTP s NUCLEAR QUALITY ASSURANCE PROGRAM RTP s NUCLEAR QUALITY ASSURANCE PROGRAM RTP operates under one quality program, whether you purchase products that are commercial grade, nuclear safety-related or industrial safety compliant (IEC 61508).

More information

EUROPEAN COMMISSION HEALTH AND CONSUMERS DIRECTORATE-GENERAL. EudraLex The Rules Governing Medicinal Products in the European Union

EUROPEAN COMMISSION HEALTH AND CONSUMERS DIRECTORATE-GENERAL. EudraLex The Rules Governing Medicinal Products in the European Union EUROPEAN COMMISSION HEALTH AND CONSUMERS DIRECTORATE-GENERAL Public Health and Risk Assessment Pharmaceuticals Brussels, SANCO/C8/AM/sl/ares(2010)1064587 EudraLex The Rules Governing Medicinal Products

More information

5.2.7.4 CONCRETE: CONTRACTOR S QUALITY CONTROL PLAN

5.2.7.4 CONCRETE: CONTRACTOR S QUALITY CONTROL PLAN 5.2.7.4 CONCRETE: CONTRACTOR S QUALITY CONTROL PLAN NOTE: This document is generally written in the imperative mood. The subject, "the Contractor" is implied. Also implied in this language are "shall",

More information

Estimating Weighing Uncertainty From Balance Data Sheet Specifications

Estimating Weighing Uncertainty From Balance Data Sheet Specifications Estimating Weighing Uncertainty From Balance Data Sheet Specifications Sources Of Measurement Deviations And Uncertainties Determination Of The Combined Measurement Bias Estimation Of The Combined Measurement

More information

ALACC Frequently Asked Questions (FAQs)

ALACC Frequently Asked Questions (FAQs) Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC

More information

Standard Test Method for Consistency of Paints Measuring Krebs Unit (KU) Viscosity Using a Stormer-Type Viscometer 1

Standard Test Method for Consistency of Paints Measuring Krebs Unit (KU) Viscosity Using a Stormer-Type Viscometer 1 Designation: D562 01 (Reapproved 2005) Standard Test Method for Consistency of Paints Measuring Krebs Unit () Viscosity Using a Stormer-Type Viscometer 1 This standard is issued under the fixed designation

More information

MATERIALS TESTING INSTRUMENTATION

MATERIALS TESTING INSTRUMENTATION MATERIALS TESTING INSTRUMENTATION A critical element in any materials testing environment is the accurate capture, display, analysis and presentation of test data. United offers a complete line of materials

More information

Copyright 2007 Casa Software Ltd. www.casaxps.com. ToF Mass Calibration

Copyright 2007 Casa Software Ltd. www.casaxps.com. ToF Mass Calibration ToF Mass Calibration Essentially, the relationship between the mass m of an ion and the time taken for the ion of a given charge to travel a fixed distance is quadratic in the flight time t. For an ideal

More information

1 Global Warranty Service Agent Guide. Service Agent Help Guide. Service Agent Starts the Claim

1 Global Warranty Service Agent Guide. Service Agent Help Guide. Service Agent Starts the Claim 1 Global Warranty Service Agent Guide Service Agent Help Guide Service Agent Starts the Claim 1 2 Global Warranty Service Agent Guide Table of Contents Warranty Guidelines... 3 Entering a Claim... 4 Log

More information

Handbook of Weighing Applications Balances and Scales Used as Measuring and Test Equipment in a Quality System

Handbook of Weighing Applications Balances and Scales Used as Measuring and Test Equipment in a Quality System Handbook of Weighing Applications Balances and Scales Used as Measuring and Test Equipment in a Quality System turning science into solutions Preliminary Considerations In many areas of applications, the

More information

The entire document shall be read and understood before proceeding with a test. ISTA 3B 2013 - Page 1 of 35

The entire document shall be read and understood before proceeding with a test. ISTA 3B 2013 - Page 1 of 35 Packaged-Products for Less-Than-Truckload (LTL) Shipment ISTA 3 Series General Simulation Performance Test PROCEDURE VERSION DATE Last TECHNICAL Change: NOVEMBER 2012 Last EDITORIAL Change: JANUARY 2013

More information

Peel Back Force Tester GPD s Peel Back Force Tester with FORCEWare software is being used to set peel back standards for the entire industry

Peel Back Force Tester GPD s Peel Back Force Tester with FORCEWare software is being used to set peel back standards for the entire industry PBF T Peel Back Force Tester GPD s Peel Back Force Tester with FORCEWare software is being used to set peel back standards for the entire industry 611 Hollingsworth Street, Grand Junction, CO 81505 tel:

More information

STATE OF OHIO DEPARTMENT OF TRANSPORTATION SUPPLEMENTAL SPECIFICATION 888 PORTLAND CEMENT CONCRETE PAVEMENT USING QC/QA.

STATE OF OHIO DEPARTMENT OF TRANSPORTATION SUPPLEMENTAL SPECIFICATION 888 PORTLAND CEMENT CONCRETE PAVEMENT USING QC/QA. STATE OF OHIO DEPARTMENT OF TRANSPORTATION SUPPLEMENTAL SPECIFICATION 888 PORTLAND CEMENT CONCRETE PAVEMENT USING QC/QA October 21, 2011 888.01 General 888.02 Materials 888.03 Concrete Proportioning 888.04

More information

Series 6000 Torque measured metal bellow coupling

Series 6000 Torque measured metal bellow coupling Properties Free of float metal bellow coupling with integrated torque measurement Non-contact measurement system, high robustness High torsional stiffness Limited torque of inertia Performance Measurement

More information

Process Validation for Medical Devices

Process Validation for Medical Devices Process Validation for Medical Devices Dan O Leary CBA, CQE, CRE, CSSBB, CIRM, LLC 603-209-0600 OmbuEnterprises@msn.com Copyright 2010 by, LLC Process Validation for Medical Devices 1 Instructor Introduction

More information

SIEVE ANALYSIS OF FINE AND COARSE AGGREGATES FOP FOR AASHTO T 27 (11)

SIEVE ANALYSIS OF FINE AND COARSE AGGREGATES FOP FOR AASHTO T 27 (11) SIEVE ANALYSIS OF FINE AND COARSE AGGREGATES FOP FOR AASHTO T 27 (11) MATERIALS FINER THAN 75 µm (No. 200) SIEVE IN MINERAL AGGREGATE BY WASHING FOP FOR AASHTO T 11 (11) Scope Sieve analysis determines

More information

BW-500 BED WEIGHER USER MANUAL

BW-500 BED WEIGHER USER MANUAL BW-500 BED WEIGHER USER MANUAL 0 1 TABLE OF CONTENTS SPECIFICATION... 3 Equipment... 4 INSTALLING WEIGH BEAM... 5 PREPARATION FOR MEASUREMENT... 6 DP3800 Indicator... 8 POWER SUPPLY... 8 PANEL... 9 KEY

More information

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Data Quality Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Why do we care? Program goals are key what do you want to do with the data? Data utility increases directly

More information

LAB14 EDITION 4 November 2006

LAB14 EDITION 4 November 2006 LAB14 EDITION 4 November 2006 Calibration of Weighing Machines CONTENTS SECTION PAGE 1 Introduction 2 2 Calibration of Weights 2 3 Considerations in the use of weighing machines 4 4 In-house calibration

More information

INDIANA DEPARTMENT OF TRANSPORTATION QUALIFIED LABORATORY AND TECHNICIAN PROGRAM

INDIANA DEPARTMENT OF TRANSPORTATION QUALIFIED LABORATORY AND TECHNICIAN PROGRAM 1/1/13 INDIANA DEPARTMENT OF TRANSPORTATION QUALIFIED LABORATORY AND TECHNICIAN PROGRAM The purpose of the Qualified Laboratory and Technician Program is to establish policies, procedures, and guidelines

More information

The Facts on Equine Drug Testing

The Facts on Equine Drug Testing The Facts on Equine Drug Testing CONTACT Travis Mays, MS Interim Section Head Analytical Chemistry, Drug Testing Laboratory Toxicology Laboratory 979.845.3414 tmays@ LIMITATIONS IN DRUG TESTING While advancements

More information

Field Density Determination of Asphalt Concrete by the Coring Method

Field Density Determination of Asphalt Concrete by the Coring Method Field Density Determination of Asphalt Concrete by the Coring Method 1. Scope: This procedure is for determining the density of in place asphalt concrete pavement. 2. Apparatus: 2.1 Scale or balance having

More information

GE Power Electronics Business Total Quality Management for Suppliers

GE Power Electronics Business Total Quality Management for Suppliers GE Power Electronics Business Total Quality Author: Reviewers: Contributors: Alessandro A. Cervone Technical Manager Component & Reliability Engineering Alicia Scott Title Erika Nannis Title Robert Roessler

More information

The University of Toledo Soil Mechanics Laboratory

The University of Toledo Soil Mechanics Laboratory 1 Grain Size Distribution Sieve Analysis The University of Toledo Soil Mechanics Laboratory Introduction The grain size distribution is a representation of the approximate distribution of soil grain sizes

More information

SERVICE MANUAL Valor 3000 SCALES

SERVICE MANUAL Valor 3000 SCALES SERVICE MANUAL Valor 3000 SCALES Ohaus Corporation, 7 Campus Drive, Suite 310, Parsippany, NJ 07054 (973) 377-9000 SERVICE MANUAL Valor 3000 SCALES The information contained in this manual is believed

More information

Apr 17, 2000 LAB MANUAL 1811.0

Apr 17, 2000 LAB MANUAL 1811.0 Apr 17, 2000 LAB MANUAL 1811.0 1811 BULK SPECIFIC GRAVITY (GMB) AND DENSITY OF COMPACTED BITUMINOUS SPECIMENS USING PARAFFIN OR PARAFILM ASTM Designation D 1188 (MN/DOT Modified) 1811.1 SCOPE This test

More information

OPERATOR S MANUAL SUPPORTED TORQUE CALIBRATION BEAM. Part Number 34320 Issue 4 Original Instructions (English)

OPERATOR S MANUAL SUPPORTED TORQUE CALIBRATION BEAM. Part Number 34320 Issue 4 Original Instructions (English) OPERATOR S MANUAL SUPPORTED TORQUE CALIBRATION BEAM Part Number 34320 Issue 4 Original Instructions (English) CONTENTS Part Numbers Covered by This Manual 2 Safety 2 Introduction 2 Parts Included 2 Accessories

More information

NRMCA Quality Certification. Ready Mixed Concrete Quality Management System. Certification Criteria Document

NRMCA Quality Certification. Ready Mixed Concrete Quality Management System. Certification Criteria Document NRMCA Quality Certification Ready Mixed Concrete Quality Management System Certification Criteria Document Version 1 February 2014 NRMCA Quality Certification Ready Mixed Concrete Quality Management System

More information

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history. Revision History COPY The top row of this table shows the most recent changes to this controlled document. For previous revision history information, archived versions of this document are maintained by

More information

E X P E R I M E N T 8

E X P E R I M E N T 8 E X P E R I M E N T 8 Torque, Equilibrium & Center of Gravity Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 8:

More information

Your essential guide to CE Marking

Your essential guide to CE Marking Your essential guide to CE Marking CE marking A guide for window and door fabricators This guide provides essential information that will help you with your preparations for CE marking. A flowchart is

More information

NMI R 51 Automatic Catchweighing Instruments

NMI R 51 Automatic Catchweighing Instruments NMI R 51 Automatic Catchweighing Instruments (OIML R 51-1:1996(E), IDT and OIML R 51-2:1996(E), IDT) The English version of international standard OIML R 51:1996 Automatic Catchweighing Instruments is

More information

Calibration of Volumetric Glassware

Calibration of Volumetric Glassware CHEM 311L Quantitative Analysis Laboratory Revision 2.3 Calibration of Volumetric Glassware In this laboratory exercise, we will calibrate the three types of glassware typically used by an analytical chemist;

More information

ISO/IEC 17025 QUALITY MANUAL

ISO/IEC 17025 QUALITY MANUAL 1800 NW 169 th Pl, Beaverton, OR 97006 Revision F Date: 9/18/06 PAGE 1 OF 18 TABLE OF CONTENTS Quality Manual Section Applicable ISO/IEC 17025:2005 clause(s) Page Quality Policy 4.2.2 3 Introduction 4

More information

Inventory Control. Inventory Control Tank & Silo Weighing

Inventory Control. Inventory Control Tank & Silo Weighing Inventory Control Tank & Silo Weighing Gravimetric level control is one of the most accurate inventory control methods available. Its high degree of accuracy makes it particularly helpful when measuring

More information

Quality Assurance Program. June 2005 2005 by Texas Department of Transportation (512) 416-2055 all rights reserved

Quality Assurance Program. June 2005 2005 by Texas Department of Transportation (512) 416-2055 all rights reserved Quality Assurance Program June 2005 2005 by Texas Department of Transportation (512) 416-2055 all rights reserved Quality Assurance Program June 2005 Manual Notices Manual Notice 2005-1To: Users of the

More information

Vehicle Lift Safety The Fleet Manager s Responsibility Presented by: Steve Perlstein

Vehicle Lift Safety The Fleet Manager s Responsibility Presented by: Steve Perlstein Vehicle Lift Safety The Fleet Manager s Responsibility Presented by: Steve Perlstein An Overview of Your Responsibilities Annual lift inspection by an experienced lift inspector Annual lift training to

More information

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry

More information

Anti-Counterfeit Policy Mouser has adopted this Anti-Counterfeit Policy to eliminate the impact of counterfeit products on Mouser and its customers.

Anti-Counterfeit Policy Mouser has adopted this Anti-Counterfeit Policy to eliminate the impact of counterfeit products on Mouser and its customers. QS-PR-024 Anti-Counterfeit Control Plan Approved by: Chuck Amsden Approved 1/19/15 This procedure is current and approved when viewed on-line. The procedure becomes uncontrolled when printed. Uncontrolled

More information

UNIFORM DATA COLLECTION PROCEDURES

UNIFORM DATA COLLECTION PROCEDURES UNIFORM DATA COLLECTION PROCEDURES PURPOSE: The purpose of these procedures is to provide the framework for a uniform, accurate record system that will increase dairy farmers' net profit. The uniform records

More information

NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING

NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING Instructions to the Assessor: This checklist addresses specific accreditation criteria prescribed in NIST Handbook 150-5, Construction Materials

More information

ANALYTICAL BALANCES OHAUS EXPLORER. Intelligent Performance

ANALYTICAL BALANCES OHAUS EXPLORER. Intelligent Performance ANALYTICAL BALANCES OHAUS EXPLORER Intelligent Performance Explorer delivers accurate results within seconds, improving operator efficiency, productivity, and throughput, with a stabilization time up to

More information

PRESSURE REGULATOR SERIES 2100-4100 - 5100 R 4100 R 2100 R 5100 VIRTUAL CATALOGUE FAIL TO OPEN REGULATOR BALANCED VALVE HIGH FLOW COEFFICIENT

PRESSURE REGULATOR SERIES 2100-4100 - 5100 R 4100 R 2100 R 5100 VIRTUAL CATALOGUE FAIL TO OPEN REGULATOR BALANCED VALVE HIGH FLOW COEFFICIENT VIRTUAL CATALOGUE FAIL TO OPEN REGULATOR BALANCED VALVE HIGH FLOW COEFFICIENT PRESSURE REGULATOR WIDE PRESSURE-REGULATION RANGE CAN BE SUPPLIED WITH MINIMUM / MAXIMUM PRESSURE SLAM-SHUT VALVE R 4100 POSSIBILITY

More information

HOOKE S LAW AND OSCILLATIONS

HOOKE S LAW AND OSCILLATIONS 9 HOOKE S LAW AND OSCILLATIONS OBJECTIVE To measure the effect of amplitude, mass, and spring constant on the period of a spring-mass oscillator. INTRODUCTION The force which restores a spring to its equilibrium

More information

TEMPORARY AND PERMANENT SUPPORT OF HYDRO OTTAWA DUCT BANKS WHEN UNDERCUT BY AN EXCAVATION

TEMPORARY AND PERMANENT SUPPORT OF HYDRO OTTAWA DUCT BANKS WHEN UNDERCUT BY AN EXCAVATION TITLE: Engineering Specification RECOMMENDED: C. Proulx APPROVED: C. Malone REV. DATE: 2002-09-12 NO: UDS0013 1 OF 5 REV: 0 TEMPORARY AND PERMANENT SUPPORT OF HYDRO OTTAWA DUCT BANKS WHEN UNDERCUT BY AN

More information

Training Material for Balances (1) Prepared by R&D Division 1. Marketing Department A&D Company, Limited

Training Material for Balances (1) Prepared by R&D Division 1. Marketing Department A&D Company, Limited Training Material for Balances (1) Prepared by R&D Division 1 Marketing Department A&D Company, Limited 1 Contents Things you should know before using a balance --------------------------- P.3 1. Convection

More information

Standard Practice for Quality Control Systems for Nondestructive Testing Agencies 1

Standard Practice for Quality Control Systems for Nondestructive Testing Agencies 1 Designation: E 1212 99 AMERICAN SOCIETY FOR TESTING AND MATERIALS 100 Barr Harbor Dr., West Conshohocken, PA 19428 Reprinted from the Annual Book of ASTM Standards. Copyright ASTM Standard Practice for

More information

AP Physics 1 and 2 Lab Investigations

AP Physics 1 and 2 Lab Investigations AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks

More information

CountEasy Retail Advanced Configuration Manual

CountEasy Retail Advanced Configuration Manual CountEasy Retail Advanced Configuration Manual The CountEasy TM advanced configuration options affect the function and operation of the machine. They are options that are not normally required to get CountEasy

More information

Rounding and Significant Figures

Rounding and Significant Figures National Renewable Energy Laboratory Innovation for Our Energy Future A national laboratory of the U.S. Department of Energy Office of Energy Efficiency & Renewable Energy Rounding and Significant Figures

More information

DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements

DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements DMS 680 Universal length measuring system gauge inspection for»iso 9000«requirements Universal length measuring system DMS 680 Wide range of applications For periodic inspection of gauges, reference gauges,

More information

ERRATUM: OIML R 60 RECOMMENDATION. Edition 2000 (E) ORGANISATION INTERNATIONALE ORGANIZATION INTERNATIONAL. Metrological regulation for load cells

ERRATUM: OIML R 60 RECOMMENDATION. Edition 2000 (E) ORGANISATION INTERNATIONALE ORGANIZATION INTERNATIONAL. Metrological regulation for load cells ERRATUM: INTERNATIONAL RECOMMENDATION OIML R 60 Edition 2000 (E) Metrological regulation for load cells THIS ERRATUM ONLY CONCERNS THE ENGLISH VERSION OF R 60 (2000) ORGANISATION INTERNATIONALE DE MÉTROLOGIE

More information

GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards

GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards March 2003 GMP 11 Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards 1. Introduction 1.1. Purpose Measurement processes are dynamic systems and often

More information

Content Sheet 3-1: Equipment Management Overview

Content Sheet 3-1: Equipment Management Overview Content Sheet 3-1: Equipment Management Overview Role in quality management system Equipment management is one of the essential elements of a quality management system. Proper management of the equipment

More information

Electronic Cash Registers (ECR) and Point-of-Sale Systems (POS) Interfaced with Scales Part 2 - Examination By Steve Cook

Electronic Cash Registers (ECR) and Point-of-Sale Systems (POS) Interfaced with Scales Part 2 - Examination By Steve Cook September 2006 Electronic Cash Registers (ECR) and Point-of-Sale Systems (POS) Interfaced with Scales Part 2 - Examination By Steve Cook This is the second part of a two-part article on Electronic Cash

More information

GENERAL SCIENCE LABORATORY 1110L Lab Experiment 5 THE SPRING CONSTANT

GENERAL SCIENCE LABORATORY 1110L Lab Experiment 5 THE SPRING CONSTANT GENERAL SCIENCE LABORATORY 1110L Lab Experiment 5 THE SPRING CONSTANT Objective: To determine the spring constant of a spiral spring Apparatus: Pendulum clamp, aluminum pole, large clamp, assorted masses,

More information

Operating instructions Platform/floor scales

Operating instructions Platform/floor scales KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instructions Platform/floor scales KERN

More information