CA12 HANDY CAL Temperature Calibrator Model
|
|
|
- Jean Douglas
- 10 years ago
- Views:
Transcription
1 Instruction Manual CA12 HANDY CAL Temperature Calibrator Model Thank you for purchasing the CA12 HANDY CAL. To fully utilize all of the features of this instrument, read this Instruction Manual carefully and use the instrument accordingly. 2nd Edition: Aug All rights reserved, Copyright 1997, Yokogawa M&C Corporation IM CA12-E 2nd Edition
2 Contents 1. Safety Use Names and Functions of Parts Replacing the Batteries Turning the Power On/Off Generation Connecting the Output Terminals Generating DC Voltage Generating Resistance or RTD Signal Generating a Thermocouple (TC) Signal Measurement Connecting Procedure Measuring DC Voltage, Resistance, or an RTD Signal Measuring a Thermocouple (TC) Signal
3 7. Other Features Internal Reference Junction Compensation Disabling Automatic Power off Specifications How to Use the Carry Case Calibration Procedure Selecting the Standards Environmental Conditions for Calibration Calibration Points Precautions for Calibration Assignment of Keys for Calibraion Calibrating the Generation feature Calibrating the Measurement Feature Notice of the Instruction Manual
4 1. Safety Use The following symbols are used on the instrument and in the Instruction Manual to ensure safe use. WARNING Indicates that there is a possibility of serious personal injury or death if the operating procedure is not followed correctly and describes the precautions for avoiding such injury or death. CAUTION Indicates that there is a possibility of personnel injury or damage to the instrument if the operating procedure is not followed correctly and describes the precautions for avoiding such injury or damage. NOTE Draws attention to information essential for understanding the operation and features. 4
5 Damage to the instrument or personal injury or even death may result from electrical shock or other factors. To avoid this, follow the precautions below. WARNING Use in gases Do not operate this instrument in areas where inflammable or explosive gases or vapor exists. It is extremely hazardous to use the instrument under such environments. External connection If you need to touch a circuit for external connection, cut off the power from that circuit and make sure that no voltage is being supplied. Then carry out the connection. When replacing the batteries, always disconnect lead cables. CAUTION Disassembly No person other than our service personnel should open the cover. 5
6 ON Names and Functions of Parts Front View Side View Rear View (with the rear cover removed) <3> POWER key <4> C/ F selection key <10> AC adapter connection <1> Display unit jack <11> RJC sensor input connector <2> Output setting keys <12> DIP switches <13> Battery holder <5> Range selection rotary switch <6> MEASURE (measurement)/ SOURCE (generation) selection switch <8> Three-wire input terminal <7> Voltage and resistance input/ output terminals <9> Terminal adapter red black <14>Lead cables black 6
7 j i h a b c d g e f <1> Display unit a MEASURE Lights when MEASURE (measurement) is selected using the selection switch <6>. 7
8 b SOURCE Lights when SOURCE (generation) is selected using the selection switch <6>. c CAL Lights in the calibration mode. d 0 FS Lights or blinks when the offset or full scale is calibrated in the calibration mode. e + - This mark informs the battery s status. When lit, it indicates that the batteries will soon need replacing and when blinking, it indicates that they must be replaced. (see Section 3, Replacing the Batteries. ) f Displays a measured value or generated value. 8
9 g F appears when C/ F selection key is selected to F at temperature range. It also displays the lower two digits of a measured or generated value in the adjustment mode. h Shows the unit of the range selected. i RJ-ON Lights when the reference junction compensation is being calculated in signal generation. j ON Lights during signal generation. 9
10 <2> Output value setting keys Sets an output value for signal generation. The [ ]/[ ] keys are provided for each digit, which increase or decrease the value one by one. Carry and borrow of the digits is applied respectively according to the incrementing of 9 and decrementing of 0. <3> POWER key Turns on/off the power supply. For more information, see Section 4, Turning the Power On/Off. <4> C/ F selection key Selects C or F unit. <5> Range selection rotary switch Selects a range for generation or measurement. <6> MEASURE/SOURCE selection switch Selects SOURCE (generation) or MEASURE (measurement). 10
11 <7> I/O terminals Used for generation and measurement in each range. <8> Three-wire input terminal Used for measurement in three-wire connection in the resistance or RTD range. <9> Terminal adapter Attached to the instrument s terminals for use when a thermocouple signal is measured or when lead wires are to be connected directly to the terminals. <10> AC adapter connection jack Used to connect an AC adapter (optional). <11> RJC sensor input connector When using an external RJC sensor (optional), connect it here. 11
12 <12> DIP switches See Section 7, Other Features. <13> Battery holder Contains four AA-size batteries. See Section 3, Replacing the Batteries. <14> Lead cables for measurement or generation Used to connect the instrument to the device under measurement/ generation. 12
13 3. Replacing the Batteries If the + - mark on the display unit starts blinking, the batteries have been used up. Follow the procedure below to replace the batteries. <1> Check that the power is off. <2> Slide the rear cover at the back of the instrument to remove it. <3> Replace all four batteries with new ones. Place them in their holder according to the polarity directions shown inside the holder. <4> After replacing the batteries, put the rear cover back on the instrument. Connecting the AC Power (optional) Before connecting the AC power Perform the following precautions to avoid electrical shock or damage to the instrument. 13
14 WARNING Do not use any AC adapter other than the dedicated AC adapter (See Page 34) Before connecting the power cord, check that the voltage of the supply side matches the rated voltage of the instrument. Before connecting the power cord, check that the power of the instrument is off. Connecting procedure: <1> Check that the POWER key of the instrument is off. <2> Connect the AC adapter (optional) to the AC adapter connection jack in the instrument. (Note that unless the AC adapter is connected to the power outlet, the power cannot be turned on.) 14
15 4. Turning the Power On/Off Operating the POWER Key When the instrument s power is off, pressing the POWER key on the front panel for more than 1 second causes the power to be turned on. Pressing this key again causes it to be turned off. When the power is turned on, the instrument starts a self-test and displays CA12. The features selected by the range selection rotary switch and MEASURE/SOURCE selection switch starts functioning. For battery-driven operations, disconnect the AC adapter from the instrument. 15
16 Automatic Power off In the factory-shipped setting, all indications start blinking if the instrument has not been operated for about 9.5 minutes. Then, if the instrument is not operated for another 30 seconds, it automatically turns off. To disable this automatic Power off feature, refer to Section 7, Other Features. If you wish to keep the instrument turned on when the indications start blinking, press the [POWER] key. This causes blinking to normal lighting, without changing the previous status. 16
17 5. Generation 5.1 Connecting the Output Terminals <1> Insert the plugs of the lead cables supplied into the output terminals of the instrument. <2> Connect the clips on the other ends of the cables to the input terminals of the device under measurement/generation. (a) Voltage (Thermocouple), 2W Resistance (RTD) OUTPUT 3W Lo Lo Hi Lo Hi Terminal adaptor (b-1) 3W Resistance (RTD) OUTPUT 3W Lo Lo Hi 3W Lo Lo 17 Hi (b-2) 3W Resistance (RTD) OUTPUT 3W ; Lo Lo Hi Short bar (supplied) Terminal adaptor 3W Lo Hi Lo When carrying out calibration on a resistance temperature detector or a resistance measurement unit in three-wire connection, the supplied terminal adapter can be used to achieve a three-wire configuration (Fig.b-1). 3W Lo and Lo terminals can be shorted (Fig.b-2). Otherwise connect as Fig.a.
18 CAUTION Do not apply any voltage to the output terminals during the generation operation. If voltage is applied inadvertently, the internal circuit may be damaged. NOTE As this instrument is calibrated without the voltage drop of the lead cables, error due to the resistance of the lead cables (about 0.1 Ω for go and return) must be considered. 18
19 5.2 Generating DC Voltage The instrument generates a voltage or resistance of a specified value from the output terminals. Normal procedure: <1> Switch the MEASURE/SOURCE selection switch to SOURCE (generation). This causes the display unit to light SOURCE and ON. <2> Select the range to be generated using the range selection rotary switch. The display unit shows an initial value and the unit for each range. <3> Press the [ ]/[ ] keys for each digit to set an output value. 19
20 5.3 Generating Resistance or RTD Signal The instrument generates a required value of simulated resistance output at the 400 Ω resistance range and at PT100 RTD (resistance temperature detector) range which corresponds to the required temperature value of the PT100. NOTE The method of simulating resistance output is by generating voltage according to the excitation current I received from the device under calibration, due to the equation: R (required resistance) = V (generated voltage) / I (current received from device) Therefore the device to be calibrated must have the excitation current for resistance measurement. For correct simulation output, the excitation current should be within a range of 0.5 ma to 2 ma (please refer Section 8 for more information). 20
21 NOTE On initial calibration conditions from the factory, the generated resistance value is calibrated without considering the voltage drop due to the lead cables (approx 0.1 Ω for the supplied), so the increase of error rate should be considered at the opposite end of the lead cables. If capacitance between the terminals of the resistance-measuring device under calibration becomes 0.1 µf or more, the CA12 may not be able to generate the correct resistance value. 21
22 5.4 Generating a Thermocouple (TC) Signal The instrument generates thermo-electromotive force from the output terminals, corresponding to the required temperature of a specified thermocouple (TC). Reference junction compensation When you directly calibrate a thermometer with a function of a built-in reference junction compensation, use an RJC sensor (optional) as follows. (You can also use the built-in RJC sensor to carry out calibration (see Section 7, Other Features ).) <1> Connect the RJC sensor to the RJC sensor input connector of the instrument. Insert the sensor such that the claw at the bottom of the connector is locked in. To release the connector, gently press the locking claw to unlock the connector and then remove it. 22
23 <2> When the sensor is connected, the instrument automatically enters RJ-ON status and outputs a thermo-electromotive force based on the output value settings and compensation of the temperature detected by the RJC sensor. The instrument displays RJ-ON. (For the accuracy of the reference temperature measurement using the RJC sensor, see Section 8, Specifications. ) The thermo-electromotive force is obtained by subtracting calculated value of the RJC sensor from the calculated thermo-electromotive force without RJC sensor. Compensation of output voltage by the RJC sensor is achieved by a sampling approximately every 4 seconds. Thus, there is a maximum delay of 4 seconds between the connection of the connector and the start of compensation. For accurate compensation, a certain amount of time (about 5 minutes) must be reserved. 23
24 NOTE If reference junction compensation is not required, the RJC sensor should always be disconnected from the connector of the instrument. 24
25 6. Measurement WARNING When connecting the device under measurement, turn off the power for the device. Connecting/disconnecting the lead cables for measurement without turning off the power of the device under measurement may be extremely dangerous. Special care should be taken to avoid connecting a current circuit to the input terminals. Inadvertent connection may not only cause damage to the circuit or device under measurement and this instrument, but may also be hazardous to personnel. The maximum allowable voltage between all input/output terminals and ground is 42 V. Any voltage exceeding this level may not only damage the instrument but also cause injury to personnel. Never attempt to apply such voltage. 25
26 CAUTION Do not apply voltage exceeding the measuring range to the input terminals. This may damage the instrument. 6.1 Connecting Procedure <1> When connecting a thermocouple or the lead wires, attach them to the terminal adapter supplied with the instrument. <2> For measurements of the voltage or the resistance or RTD in a twowire connection, connect the supplied lead cables to the terminals of the instrument or the lead wires to the terminal adapter (see following figure a ). For measurements of the resistance or RTD in a three-wire connection, connect the lead cables to the terminals of the instrument or connect the lead wires to the terminal adapter (see following figure b ). 26
27 Remove the shortbar (supplied) when measuring with a three-wire commection. For the measurements of the resistance or the RTD in a two-wire connection, do not connect any point to the 3W Lo terminal except the Lo terminal (middle terminal). a Connection of the input terminals for measuring voltage, a thermocouple signal, the resistance in a two-wire connection, or an RTD signal in a twowire connection 3W Lo Lo Hi b Connection of the input terminals for measuring the resistance in a three-wire connection and an RTD signal in a threewire connection 3W Lo Lo Hi Terminal adaptor Terminal adaptor Lo Hi 27
28 6.2 Measuring DC Voltage, Resistance, or an RTD Signal <1> Switch the MEASURE/SOURCE selection switch to MEASURE (measurement). MEASURE lights up on the display unit. <2> Select the range to be measured using the range selection rotary switch. The display unit shows the measurement results (such as the corresponding temperature for the RTD). The measurement results display is updated approximately every second. If the measurement data is over range, the display unit shows
29 6.3 Measuring a Thermocouple (TC) Signal <1> Switch the MEASURE/SOURCE selection switch to MEASURE (measurement). MEASURE lights up on the display unit. <2> Select the thermocouple to be measured using the range selection rotary switch. The display unit shows the results of the corresponding temperature, on which the internal RJC sensor-measured temperature compensation was based. If the input terminals are open, the display unit shows -bo- (burnout). If the measurement data is over range, the display unit shows If the temperature in the operating environment of the instrument changes rapidly, wait until RJ compensation stabilizes and then use the instrument. 29
30 7. Other Features The following features are available depending on the setting of the DIP switches above the battery holder. These switches are accessed by removing the rear cover from the back of the instrument. DIP switches ON : When generating: Internal RJC 2: Unused 3: Unused 4: Disabling automatic power off 7.1 Internal Reference Junction Compensation When generating a thermocouple signal, setting the DIP switch 1 on enables the instrument s internal RJC sensor to generate a temperature compensated output. 30
31 7.2 Disabling Automatic Power off At the factory shipment, the instrument is initialized to turn off automatically if not operated for a period of 10 minutes or more. If you wish to use it continuously without automatic power off, set the DIP switch 4 to on. This disables the automatic power off feature. However, when the instrument is battery-driven, it is recommended that this switch be generally set to off in order to prevent the batteries from being used up. 31
32 8. Specifications Source/Measurement Functions Accuracy: ±(% of set value or reading + C, µv, or Ω), at 23 ± 5 C for one year Range Selection TC Range of Source/measure Source Accuracy Measurement Resolution K E to º C to º C -328 to 2498º F -328 to 1832º F 0.05%+1ºC *1 0.07%+1.5ºC J to º C -328 to 2192º F (at -100ºC or greater) (at -100ºC or greater) 0.05%+2ºC 0.07%+2ºC 0.1ºC/1ºF T to 400.0º C -328 to 752º F (at less than -100ºC) (at less than -100ºC) N to º C -328 to 2372º F R 0 to 100ºC 32 to 212º F 0.05%+3ºC * %+3º C 100 to 1768ºC 212 to 3214º F 0.05%+2ºC * %+2º C 1ºC/1ºF 100mV to m V 0.05%+30µV 0.05%+30µV 10µV RTD Pt to 850.0ºC -328 to 1562º F 0.05%+0.6ºC * %+0.6º C 0.1ºC/1º F 400Ω 0 to Ω 0.05%+0. 2Ω * %+0. 2Ω 0. 1Ω Temperature effect: 1/10 of accuracy / ºC. Remarks *1 Accuracy of RJC is excluded. *2: External exitation current 0.5 to 2 ma; A dd 0.05% +1ºC (0.4Ω ) when it is 0.1 ma. 32
33 Power supply : Four 1.5-V alkaline batteries (ANSI AA-size) or dedicated AC adapter (sold separately) Battery life : Approximately 80 hours (when running on alkaline dry batteries) Automatic Power off : After a period of approx. 10 minutes with no operations Generation Signal Level Setting : By four sets of up and down keys Response of Generator : Approximately 20 milliseconds Loading conditions : Less than 0.1 µf Measured-value indication updating intervals : Every second (approx.) Display : 7-segment-by-5-digit LCD Maximum allowable applied voltage : 42 V between each terminal and ground Operating temperature and humidity range : 0 to 50 C, 20 to 80% R.H (no condensation allowed) Storage temperature and humidity range : -20 to 50 C, 90% R.H or less (no condensation allowed) Dimension : Approximately 192 (H) 92 (W) 42 (D) mm (excluding protrusions) Weight : About 440 g Safety : EN :1993 When using with AC adaptor (optional), only B9108WB conforms to the safety regulation (A1020UP/A1022UP are excluded). EMC : EN55011:1991 Group 1 Class B EN :1992 Influence under RF field : ±1% of range 33
34 Accessories Optional accessories : Measurement/generation lead cables (B9108MT), a set of red and 2black wires Carry case (B9108NK) Terminal adapter (B9108KF) : Dedicated AC adapter (A1020UP AC 100 V A1022UP AC 120 V B9108WB AC V) RJC (reference junction compensation) sensor (B9638CR) Specifications of the RJC Sensor (for Reference Junction Compensation of TC Signal Generation) M easuring range Accuracy (when combined with the instrument) - 10 to 50 C 23± 5 C - 10 to +50 C :± 0. 5 C :± 1 C Length of cord Approximately 1.5 m Compensation by the built-in sensor is also possible by adjusting the setting of the internal DIP switch on the instrument. 34
35 9. How to Use the Carry Case The carry case may be used as described below: (1) Release the strap-side snap fasteners and the other side snap fasteners of the case cover to open it. (2) Turn the case cover over by using the logo-side snap fastener as the center. (3) Re-fasten the strap-side snap fasteners and side snap fasteners of the cover. Other features and notes of the carry case * The instrument can be placed into the housing of the back side of the case cover, with the lead cables connected to the terminals of the instrument. * The strap allows the instrument to be used or stored by suspending it on a hook, rod, and others. * Note that the logo-side fastener of the case cover cannot be released. 35
36 10. Calibration Procedure To maintain a high level of accuracy, it is recommended that the CA12 HANDY CAL be calibrated annually. Requests for calibration work can also be made to our service representatives. The following is examples of calibrations that use the standards recommended in Selecting the Standards Selecting the Standards Generation feature Items to be calibrated DCV Ω Names of standards Digital multimeter Digital multimeter Range Measuring range Accuracy 100mV 110 mv max. ± ( 0.005%+ 5µ V) 400Ω 440 Ω max., Measuring current: 1 ma ± ( 0.005%+0.02Ω) Remarks Model 1271/7561 (Yokogawa) or equivalent Model 1271/7561 (Yokogawa) or equivalent 36
37 Measurement feature Items to be calibrated DCV Ω Names of standards Standard DC voltage generator Decade resistance box Range Generated value 100mV 100mV ± 0.01% 400Ω 400Ω ± 0.01% Accuracy Remarks Model 9100/2552 (Yokogawa) or equivalent Model (Yokogawa) or equivalent 10.2 Environmental Conditions for Calibration Ambient temperature : 23 ± 1 C Relative humidity : 45 to 75% R.H Warm-up : Warm-up time specified for the standard 37
38 10.3 Calibration Points The calibration points are as specified in the tables below. It is possible to select the necessary range to be calibrated independently. Always calibrate the zero (0) point and full scale (FS) point in pair for generation. Generation: Measurement: Calibration points Standard value* 1 100mV 400Ω 0 0mV FS 100mV 0 0Ω FS 400Ω *1 Make adjustments to CA12 according to the reading of the standard (CA12 output value), as specified in the table. Calibration points Standard value* 2 100mV FS 100mV 400Ω FS 400Ω *2 Set the value to the standard as specified in the table. 38
39 10.4 Precautions for Calibration CAUTION Do not apply a voltage exceeding the maximum input voltage; otherwise, the input part may be damaged. Do not short-circuit or apply an external voltage to output terminals of the instrument or standard equipment, or else their internal circuitry may be damaged. (1) Calibrating 400 Ω generation Set resistance-measuring current to 1 ma (*). (*) Check the specifications of the digital multimeter (DMM) to be used as the standard. Example: Resistance-measuring current is 1 ma for the 1 kω range of 1271 (Yokogawa) or the 2000 Ω range of 7561 (Yokogawa). 39
40 Connection method Connect the CA12 calibrator to the standard in four-wire connection for calibration as shown below: CA12 Input terminal Sensing terminal 3W L L H H L DMM 2000 Ω range Standard 40
41 (2) Calibrating 400 Ω measurement Connection method Connect the CA12 calibrator to the standard in three-wire connection for calibration as shown below: CA12 3W L L H Decade resistance box 400 Ω Standard 41
42 (3) Calibrating the internal reference junction compensation feature Because this calibration requires special equipment (a K-type thermocouple and 0 C reference temperature chamber), contact our service representatives for calibration. 42
43 10.5 Assignment of Keys for Calibration Input cancellation key Input determination (ENTER) key Switching the mode (calibration or normal) (Press two keys simultaneously for 2 seconds.) Adjustment keys (6 keys) Digits to be adjusted by adjustment keys [ 4/ 4] 43
44 10.6 Calibrating the Generation feature Operation procedure: <1> Warm up the standard. <2> Before turning on the power of the CA12 calibrator, connect it to the standard. Make sure that the external RJC sensor is disconnected. <3> Turn on the [POWER] key of CA12, and warm up the instrument. <4> Press the [ 1] and [ 4] keys shown in the previous figure simultaneously for about 2 seconds to enter the calibration mode. <5> Select the generation range to calibrate, using the MEASURE/ SOURCE selection switch and range selection rotary switch. The display unit shows CAL, SOURCE, ON, 0, and a lower limit. 44
45 <6> Read the output value of the CA12 using the standard (digital multimeter) and adjust the CA12 so that the output value is set to the offset value using the [ ] and [ ] adjustment keys. Then, press the [ 1] input determination (ENTER) key to fix the setting. After fixing the setting, the display unit changes its contents to CAL, SOURCE, ON, FS, and a full scale value of the range. <7> Read the output value of the CA12 using the standard (digital multimeter) and adjust the CA12 so that the output value is set to the full scale value using the [ ] and [ ] adjustment keys. Then, press the [ 1] input determination (ENTER) key for about 1 second to fix the setting. After fixing the setting, the display unit shows 0 FS blinking. Repress the [ 1] input determination (ENTER) key for about 1 second to write the calibration coefficients into the EEPROM of the instrument. (This overwrites the previous calibration coefficients.) When this is complete, the instrument returns to the status in step 5. <8> Repeat steps 5 to 7 for each range to calibrate. 45
46 To return to the previous step: <9> To return to the previous step without fixing the setting, press the [ 1] input cancellation key. Note that this is possible only before writing into EEPROM. To return to the normal operation mode: <10> Press the [ 1] and [ 4] keys shown in the previous figure simultaneously for about 2 seconds, or press the [POWER] key to turn off the power once and then press again to turn back on. 46
47 10.7 Calibrating the Measurement Feature Operation procedure: <1> Warm up the standard. <2> Before turning on the power of the CA12 calibrator, connect it to the standard. Make sure that the external RJC sensor is disconnected. <3> Turn on the POWER key of the instrument, and warm up the instrument. <4> Press the [ 1] and [ 4] keys shown in the previous figure simultaneously for about 2 seconds to enter the calibration mode. <5> Select the measurement range to calibrate, using the MEASURE/ SOURCE selection switch and range selection rotary switch. CAL and MEASURE appear and FS blinks on the display unit. (If a value nearly equivalent to full scale has been input, a measured value and FS appear.) 47
48 <6> Set up the standard in order to input the full scale to the CA12. Wait until the reading stabilizes, then press the [ 1] input determination (ENTER) key for about 1 second to fix the setting. <7> After fixing the setting, 0 and FS indications on the display unit start blinking. Re-press the [ 1] input determination (ENTER) key for about 1 second to write the calibration coefficients into the EEPROM of the instrument. (This overwrites the previous calibration coefficients.) <8> Repeat steps 5 to 7 for each range to calibrate. To return to the previous step: <9> To return to the previous step without fixing the setting, press the [ 1] input cancellation key. Note that this is possible only before writing into the EEPROM. To return to the normal operation mode: <10>Press the [ 1] and [ 4] keys shown in the previous figure simultaneously for about 2 seconds, or press the [POWER] key to turn off the power once and then press again to turn back on. 48
49 11. Notice of the Instruction Manual <1> The information contained in this Instruction Manual is subject to change without notice. <2> The information contained herein is assumed to be accurate. However, should any doubt, errors, omission, or come to your attention, please let inform us. <3> Yokogawa M&C assumes no responsibility or liability for damages resulting from the customer s misuse or inadvertent operations. <4> This Instruction Manual describes the details of the functions of the product and does not warrant the product for any particular purposes. 49
50 YOKOGAWA M&C CORPORATION International Operations Kojimachi-Tokyu Bldg. 3F 6-6 Koji-machi, Chiyoda-ku, Tokyo, Japan Phone : Facsimile : YOKOGAWA CORPORATION OF AMERICA 2 Dart Road, Newnan, Ga , U.S.A. YOKOGAWA TAIWAN CORPORATION Fu Chi Bldg. 6th Floor, 99Jon-ai Road, Phone : Facsimile : nd Section, Taipei, TAIWAN YOKOGAWA EUROPE B. V. Phone : Facsimile : Radiumweg 30, 3812 RA, Amersfoort, THE NETHERLANDS YOKOGAWA AUSTRALIA PTY. LTD. Phone : Facsimile : Private mail bag 24, Centre Court D3, Telex : YEF NL Paul Street North, North Ryde, N. S. W. YOKOGAWA AMERICA DO SUL S. A. Avenida Jura, 149-Alphaville, 2113, AUSTRALIA Phone : Facsimile : Barueri, Sao Paulo, BRAZIL YOKOGAWA BLUE STAR LTD. Phone : Facsimile : /4 Lavelle Road, Bangalore , INDIA YOKOGAWA ENGINEERING ASIA PTE. LTD. Phone : Facsimile : Tampines Street 92, Singapore , SINGAPORE LTD. YOKOGAWA ELECTRIC Phone : Facsimile : Hotel International No. 2, Room 1047, Krasnopresnenskaja Nab. Telex : RS YASSIN 12, Moscow , RUSSIAN FEDERATION YOKOGAWA MEASURING INSTRUMENTS KOREA CORPORATION City Air Tevminel Bldg., 405-9, #159-6, Samsung-dong, Kangnam-ku, Seoul, KOREA Phone : to 0664 Facsimile : Phone : Facsimile :
VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421
VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 TABLE OF CONTENTS 1. FEATURES... 1 2. SPECIFICATIONS... 1 2-1 General Specifications...1 2-2 Electrical Specifications... 2 3. FRONT PANEL DESCRIPTION... 4 3-1
Model UT713 OPERATING MANUAL
Model UT713 OPERATING MANUAL TABLE OF CONTENTS TITLE PAGE Introduction Unpacking Inspection Safety Information Turning the Calibrator On Simulating a Thermocouple Measuring a Thermocouple Simulating TC
Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011
Instruction Manual 2in1 LAN Tester & Multimeter Model: LA-1011 1 Contents Introduction... Features... Safety Precautions.. Meter Description... Electrical Specification... Operation.. AutoRanging Multimeter.
Model UT33A: OPERATING MANUAL Table of Contents (1)
Table of Contents (1) Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter structure Functional Buttons LCD Display Measurement Operation
The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.
i800 AC Current Clamp Instruction Sheet Introduction The i800 AC Current Clamp, the Clamp, has been designed for use with multimeters, recorders, power analyzers, safety testers, etc., for accurate non-intrusive
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-763 61-765
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-763 61-765 The Service Information provides the following information: Precautions and safety information Specifications Performance test procedure Calibration
UT202A Operating Manual. Contents
Title Contents Page Overview Unpacking Inspection Safety Information Rules for Safe Operation International Electrical Symbols The Meter Structure Functional Buttons and auto power off Display Symbols
User's Guide. True RMS Industrial Multimeter
User's Guide 97650 True RMS Industrial Multimeter Ω C ã F ã 10A V µ 10A V ã ã ma A Introduction This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency (electrical & electronic),
Manual Ranging MultiMeter
Owner s Manual Manual Ranging MultiMeter Model 82345 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product.! Safety! Operation! Maintenance!
AutoRanging Digital MultiMeter
Owner's Manual AutoRanging Digital MultiMeter Model No. 82139 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance
Multi Media Cable Tester TM-901N/TM-902/TM-903
Multi Media Cable Tester TM-901N/TM-902/TM-903 User Manual HB2TM9020010 TENMARS ELECTRONICS CO., LTD Table of Contents 1 Introduction... 1 2 Safety Precaution... 1 3 The verifier does the following:...
DC POWER SUPPLY ALIMENTATION C.C.
DC POWER SUPPLY ALIMENTATION C.C. ISO-TECH IPS 303A 201-3424 ISO-TECH IPS 601A 201-3446 82IP-303A0ME E1 SAFETY TERMS AND SYMBOLS These terms may appear in this manual or on the product: WARNING. Warning
Model UT33B/C/D: OPERATING MANUAL Table of Contents (1)
Table of Contents (1) Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter structure Functional Buttons Measurement Operation A. DC
User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
Thermistor Thermometer MODEL NO. 93210-00
Thermistor Thermometer MODEL NO. 93210-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail:
RIGOL. Quick Guide. DS1000CA Series Oscilloscope. Aug. 2011. RIGOL Technologies, Inc.
Quick Guide DS1000CA Series Oscilloscope Aug. 2011 Technologies, Inc. Guaranty and Declaration Copyright 2011 Technologies, Inc. All Rights Reserved. Trademark Information is a registered trademark of
User Guide. Model 380260 Insulation Tester / Megohmmeter
User Guide Model 380260 Insulation Tester / Megohmmeter Introduction Congratulations on your purchase of Extech s Insulation Tester/Megohmmeter. The Model 380260 provides three test ranges plus continuity
4-Channel Thermometer / Datalogger
USER GUIDE 4-Channel Thermometer / Datalogger RTD and Thermocouple Inputs Model SDL200 Introduction Congratulations on your purchase of the Extech SDL200 Thermometer, an SD Logger Series meter. This meter
Automatic Phone-Out Home Monitoring Systems
Automatic Phone-Out Home Monitoring Systems Power Outage and Freeze Alarm Model Number: THP202 Power Outage, Freeze and Flood Alarm Product Description Model Number: THP201 These monitoring systems are
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
How To Control A Power Supply With A Mini Transmitter And Switch (Power Supply)
transmitter (2 wire) / switch for continuous or On/Off Control Indication, monitoring, transmitting and continuous or On/Off control in one device Output signal 4...20 ma, 2-wire for continuous control
User Guide. www.burntec.com. 4 wire Earth Resistance Tester Model GRT300
User Guide 4 wire Earth Resistance Tester Model GRT300 Introduction Congratulations on your purchase of Extech s 4 Wire Earth Resistance Tester. The Model GRT300 has been designed and tested according
M 140i. Multifunction Calibrator. Operation Manual
M 140i Multifunction Calibrator Operation Manual M-140i Multifunction Calibrator MEATEST, s.r.o. 2 Operation Manual v33 MEATEST, s.r.o. M-140i Multifunction Calibrator Content Operation Manual... 1 Basic
Multifunction Process Calibrator
725 Multifunction Process Calibrator Product Overview PN 1549644 English January 2000 Rev.2, 8/05 2000-2005 Fluke Corporation. All rights reserved. Printed in USA All product names are trademarks of their
User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction
User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 407780 Product Page Integrating Sound Level Datalogger
User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
Differential Pressure Sensors for air and non-corrosive gases, with calibration certificate
1 919 1919P01 1919P03 QBM65/C, QBM75-1U/C Differential Pressure Sensors for air and non-corrosive gases, with calibration certificate QBM751-1/C QBM65/C QBM75/C Highly accurate measurements with calibration
Agilent 4338B Milliohm Meter
Agilent 4338B Milliohm Meter 10 µω to 100 kω Technical Overview Introduction Ideal for precise measurements of extremely low resistances using an ac test signal, the Agilent Technologies 4338B suits bench-top
TMS TANK MANAGEMENT SYSTEM
TMS TANK MANAGEMENT SYSTEM Page 1 of 9 Operating Instructions GENERAL The Tank Management System is a bespoke design to control, monitor and accommodate efficient storage and dispensing of TMS. FUNCTIONS
Ground Resistance Clamp On Tester
USER MANUAL Ground Resistance Clamp On Tester MODEL 382357 Introduction Congratulations on your purchase of Extech s 382357 Ground Resistance Tester. This Clamp on device allows the user to measure ground
User Manual. Humidity-Temperature Chart Recorder. Model RH520
User Manual Humidity-Temperature Chart Recorder Model RH520 Introduction Congratulations on your purchase of the Extech RH520 Temperature + Humidity Chart Recorder. The RH520 measures and displays Temperature,
OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S
Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and
EzFlex Combustible Gas Leak Detector Model EZ40
User's Guide EzFlex Combustible Gas Leak Detector Model EZ40 Introduction Congratulations on your purchase of the Extech EZ40 Combustible Gas Leak Detector. The EZ40 detects the presence of combustible
ITEM MODEL QTY. Filter 3910-04 1. ITEM MODEL QTY USB-RS485 Converter (with 3716A-21) 3716A-20 1
Component List Standard ITEM MODEL QTY Main Unit (3716A/3717A) 3716A-00 3717A-00 AC Adapter - 1 Filter 3910-04 1 Joint - 1 Tygon Tube - 1 User s Manual - 1 Settings Software (DVD) 3716A-40 1 1 Optional
Multi-Range Programmable DC Power Supplies 9115 Series
Data Sheet 1200 W Multi-Range DC Power Supplies Features & Benefits Any 9115 series model can replace several supplies on your bench or in your rack. Unlike conventional supplies with fixed output ratings,
Hygro-Thermometer + InfraRed Thermometer Model RH101
User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative
323/324/325. Calibration Manual. Clamp Meter
323/324/325 Clamp Meter Calibration Manual February 2013 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective
SDX Submersible Depth Transmitter User Manual
SDX Submersible Depth Transmitter User Manual January 2011 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.
Automation System TROVIS 6400 TROVIS 6493 Compact Controller
Automation System TROVIS 6400 TROVIS 6493 Compact Controller For panel mounting (front frame 48 x 96 mm/1.89 x 3.78 inch) Application Digital controller to automate industrial and process plants for general
i410/i1010 AC/DC Current Clamp
/i1010 AC/DC Current Clamp Instruction Sheet Safety Information XW Read First: Safety Information To ensure safe operation and service of the current clamp, follow these instructions: Read all operating
HVAC Clamp Meter. Users Manual
902 HVAC Clamp Meter Users Manual PN 2547887 May 2006 Rev. 1, 3/07 2006-2007 Fluke Corporation. All rights reserved. Printed in China. All product names are trademarks of their respective companies. LIMITED
7 FUNCTION DIGITAL MULTIMETER
7 FUNCTION DIGITAL MULTIMETER 90899 OPERATING INSTRUCTIONS 3491 Mission Oaks Blvd., Camarillo, CA 93011 Visit our Web site at http://www.harborfreight.com Copyright 2004 by Harbor Freight Tools. All rights
CONTENTS. PAGE Chapter 1 General Information 1. Chapter 2 Setup 4 DP450-VDC 7 DP450-VAC 7 DP450-DCC 7 DP450-F 8 DP450-HACC 9 DP450-HDCC 11 DP450-E 13
CONTENTS PAGE Chapter 1 General Information 1 Chapter 2 Setup 4 DP450-VDC 7 DP450-VAC 7 DP450-DCC 7 DP450-F 8 DP450-HACC 9 DP450-HDCC 11 DP450-E 13 Chapter 3 Alarm/Excitation Options 15 Wiring 17 Chapter
Ground Resistance Clamp-On Tester
User s Manual Ground Resistance Clamp-On Tester MODEL 382357 Warranty EXTECH INSTRUMENTS CORPORATION (a FLIR company) warrants the basic instrument to be free of defects in parts and workmanship for one
Hand-held thermometer Model CTH7000
Calibration technology Hand-held thermometer Model CTH7000 WIKA data sheet CT 55.50 Applications Precision thermometer for very accurate temperature measurements in a range of -200... +962 C Reference
Square D Clipsal DIN-Rail Four-Channel Auxiliary Input Unit
Square D Clipsal DIN-Rail Four-Channel Auxiliary Input Unit SLCLE5504AUX for Use with Wired C-Bus Networks Instruction Bulletin Retain for future use. Square D Clipsal DIN-Rail Four-Channel Auxiliary Input
User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
CL90i. 77-021 Please read these instructions before operating the product. 3 - Beam Self-Leveling Cross Line Laser
3 - Beam Self-Leveling Cross Line Laser CL90i 77-01 Please read these instructions before operating the product Self-Leveling GB D F I E PT NL DK SE FIN NO PL GR CZ RU HU SK SI BG RO EE LV LT Contents
Instruction Manual Temp 4/5/6 Temperature Meter Series
Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.
Thermocouple Thermometer Models
Technology Made Easy... Thermocouple Thermometer Models 60010-00 (Type J) 60010-10 (Type K) 60010-20 (Type T) 68X309905 Rev 1 04/04 TABLE OF CONTENTS 1. INTRODUCTION 4 2. SAFETY PRECAUTIONS 6 3. SPECIFICATIONS
Part No. IA005001. Feb. 2013 USER S MANUAL INSULATION RESISTANCE TESTER TOS7200
Part No. IA005001 Feb. 2013 USER S MANUAL INSULATION RESISTANCE TESTER TOS7200 DANGER This Tester generates high voltage! Any incorrect handling may cause death. Read Chapter 2 PRECAUTIONS ON HANDLING
Modular I/O System Analog and Digital Interface Modules
OPERATING INSTRUCTIONS Modular I/O System Analog and Digital Interface Modules Installation Operation Maintenance Document Information Document ID Title: Operating Instructions Modular I/O System Part
Adjustment functions for both span and shift have been incorporated
SENSORS FOR SERIES LED Type Wafer Alignment Sensor FX-0-F FT-F9 FD-F7 EX-F70/F60 M SH-7 FD-L4 M-DW The use of a safe LED light beam now allows for high precision detection with a resolution of 0!m (.8
EM90 INFRA-RED THERMOMETER & THERMOCOUPLE METER
EM90 INFRA-RED THERMOMETER & THERMOCOUPLE METER Users Manual - 1 - .Introduction Thank you for purchasing the SUPCO infrared (IR) thermometer and thermocouple meter. Please take a few minutes to browse
Model UT201/202: OPERATING MANUAL. Table of Contents
Table of Contents Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons The Effectiveness
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA &
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
TS1200 Intelligent Lead Free Soldering Station. Operation Manual
TS1200 Intelligent Lead Free Soldering Station Operation Manual Thank you for purchasing an Intelligent Lead Free Soldering Station. It is designed for lead free soldering. Please read this manual before
Congratulations! Thank you!
TM-47 ORDERCODE D1370 Congratulations! You have bought a great, innovative product from DAP Audio. The DAP Audio Microphone range brings excitement to any venue. Whether you want simple plug-&-play action
Digital display DA 48
Loop powered plug-in LCD display No auxiliary energy source needed Scale freely adjustable on site Can be turned to any position Suitable for any 0 ma transmitter with EN 175301-803A connector Digital
Optical Power Meter. Specification & User Manual
Optical Power Meter Specification & User Manual Page 1 of 9 Copyright 2011 reserves the right to modify specifications without prior notice Table of Contents 1. Description and Features.......3 2. Specification......4
HP UPS R1500 Generation 3
HP UPS R1500 Generation 3 Installation Instructions Part Number 650952-001 NOTE: The rating label on the device provides the class (A or B) of the equipment. Class B devices have a Federal Communications
GSM HOME SECURITY SYSTEM
Cell /Mobile phone home security system GSM HOME SECURITY SYSTEM Model : GSM-120 TABLE OF CONTENTS 1. FEATURES... 1 2. APPLICATION... 2 3. SPECIFICATIONS... 3 4. FRONT PANEL & LAYOUT DESCRIPTION...6 5.
±15 V Current-Limited Power Supply
USER S GUIDE ±15 V Current-Limited Power Supply Model 0901 Newport Corporation 3635 Peterson Way Santa Clara, CA 95054 www.newport.com/newfocus Warranty Newport Corporation guarantees its products to be
Advantium 2 Plus Alarm
ADI 9510-B Advantium 2 Plus Alarm INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501 Harpers Road Virginia
ExStik TM ph Waterproof Meters
User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL
Electronic Pressure Switch EDS 300
Electronic Pressure Switch EDS 300 User manual Page 2 of 16 Content 1. Functions of the EDS 300...3 2. Mounting...3 3. Operating keys on the membrane keypad...4 4. Digital display...4 5. Output function...5
T0118 T2118 T3118. Instruction Manual
Programmable indoor transmitter of temperature T0118 Programmable indoor transmitter of atmospheric pressure T2118 Programmable indoor transmitter of temperature, relative humidity and other derived humidity
+ A + + - B V B - - + + + NO NC V COM - - - PV PV
+ - A + B B - V + - NO NC COM + - + - V + - PV PV Ramp Soak 100 70 50 30 3 5 3 8 5 3 Minute Steam Heating Sensor Cooling Water PID Hotplate Specifications Input power supply AC 100 to 240V, 50/60Hz, DC
Agilent U1230 Series Handheld Digital Multimeters (DMMs)
Agilent U1230 Series Handheld Digital Multimeters (DMMs) Data Sheet Retool your expectations with the new Agilent U1230 Series Handheld DMMsthe first to combine a built-in LED flashlight, both audible
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
LOXONE 12 Channel Amplifier
LOXONE 12 Channel Amplifier Item no.: 200110 Thank you for purchasing the Loxone Twelve Channel Amplifier. The versatility of the Amplifier makes it the perfect choice for almost every type of custom multi-room
Duct Humidity Transmitter
SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers
Manual. Solar Fountain Mobile Phone Charger
EN Manual Solar Fountain Mobile Phone Charger Copyrights 2008 Victron Energy B.V. All Rights Reserved This publication or parts thereof may not be reproduced in any form, by any method, for any purpose.
MAKING MODERN LIVING POSSIBLE. AK-SC255 On-Site Installation Guide DANFOSS ELECTRONIC CONTROLS & SENSORS
MAKING MODERN LIVING POSSIBLE AK-SC255 On-Site Installation Guide DANFOSS ELECTRONIC CONTROLS & SENSORS How to Use This Guide Read this Guide completely as you install and start up your new AK-SC 255 controller.
PERF10 Rubidium Atomic Clock
Owner s Manual Audio Stanford Research Systems Revision 1.0 February, 2011 2 Certification Stanford Research Systems certifies that this product met its published specifications at the time of shipment.
2. Terminal arrangement. Default (PV display) (SV display) Communication protocol selection Selects the Communication protocol. Modbus ASCII mode:
COMMUNICATION INSTRUCTION MANUAL TEMPERATURE CONTROLLER KT4, KT8 and KT9 No.KTC1E6 2009.05 To prevent accidents arising from the misuse of this controller, please ensure the operator receives this manual.
T7560A,B,C Digital Wall Module
T7560A,B,C Digital Wall Module HONEYWELL EXCEL 5000 OPEN SYSTEM BEFORE INSTALLATION All wiring must comply with local electrical codes and ordinances or as specified on installation wiring diagrams. Digital
USB-500/600 Series Low-Cost Data Loggers and Accessories
Low-Cost Data Loggers and Accessories Features Stand-alone, remote data loggers and portable logger assistant Measure temperature, humidity, voltage, current, or event/state change 1 or 2 channels Low
Temperature Controller CD401/CD901/CD701/CD501/CD100/CD101 Instruction Manual
Ⅰ. Warn Warning of connection Temperature Controller CD401/CD901/CD701/CD501/CD100/CD101 Instruction Manual If the controller fails to operate or error occurs, the system will bring fault for it, please
Cross-beam scanning system to detect slim objects. 100 mm 3.937 in
891 Object Area Sensor General terms and conditions... F-17 Related Information Glossary of terms... P.1359~ Sensor selection guide...p.831~ General precautions... P.1405 PHOTO PHOTO Conforming to EMC
Instruction Manual Copy Adapter "CPAD-C1A" "CP Fuji Electric Systems Co., Ltd. INR-SI47-0886a-EU Rev 052010
Instruction n Manual Copy Adapter "CPAD-C1A" Thank you for purchasing our Copy Adapter CPAD-C1A. This product is designed to copy function codes and their data, which are stored in the FRENIC-Mini series
EPGM1 HARDWIRED ZONE & PGM OUTPUT EXPANSION MODULE
EPGM1 HARDWIRED ZONE & PGM OUTPUT EXPANSION MODULE User manual v1.5 Compatible with: ESIM264 v7.09.02 and up. ESIM364 all versions. Main features: Hardwired zone expansion adding 16 additional zones per
J TB/TW Temperature limiter, temperature monitor
JUMO GmbH & Co. KG Delivery address:mackenrodtstraße 14, 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 e-mail: [email protected] Internet: www.jumo.net
MCR1900 Media Converter 19-Slot Chassis
MCR1900 Media Converter 19-Slot Chassis Installation Guide Part #5500304-11 Copyright Statement This document must not be reproduced in any way whatsoever, either printed or electronically, without the
Input signal Maximum Range Accuracy. Thermocouple E -50 to 700 C (-58 to 1832 F) ±1 C
F4 Process Controller Installation and Operation Guide SAFETY ALERTS The symbols below are used on the equipment and throughout this document to draw the user s attention to important operational and safety
RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer
RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer EN FR IT DE ES CONTENTS / EN Title CONTENTS Page 1. SAFETY INFORMATION...1 2. INTRODUCTION...2 3. SPECIFICATIONS...3 4. PARTS & CONTROLS...4
No.1. Digital Multimeter RIGOL TECHNOLOGIES, INC.
No.1 DM368 6½ digits Digital Multimeter Real 6 ½ digits readings resolution Up to 1 K rdgs/s of measurement speed and 512 K rdgs Nonvolatile Memory TrueRMS AC Voltage Current measuring Builtin 1 groups
RGC-IR Remote Gas Calibrator for IR400
Remote Gas Calibrator for IR400 The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically authorized in writing by
USER MANUAL V5.0 ST100
GPS Vehicle Tracker USER MANUAL V5.0 ST100 Updated on 15 September 2009-1 - Contents 1 Product Overview 3 2 For Your Safety 3 3 ST100 Parameters 3 4 Getting Started 4 4.1 Hardware and Accessories 4 4.2
USER S MANUAL HSC-24A
AIRREX AIR CONDITIONER USER S MANUAL HSC-24A Thank you for purchasing an AIRREX AIR CONDITIONER. BEFORE operation please read this user s manual carefully. Keep this manual readily available. It is ESSENTIAL
CAUTION OPC-LM1-IL. Option Card for Encoder of Line Driver Output. Instruction Manual
Instruction Manual OPC-LM1-IL Option Card for Encoder of Line Driver Output CAUTION Deliver this instruction manual without fail to those who actually operate the equipment. Read this operation manual
EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM
EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM INSTALLATION MANUAL Protec Fire Detection PLC, Protec House, Churchill Way, Nelson, Lancashire, BB9 6RT. Telephone: +44 (0) 1282 717171 Fax: +44 (0) 1282 717273
How To Use A Cdm250 Digital Multimeter
User Manual CDM250 Digital Multimeter 070-6736-03 Copyright Tektronix, Inc. 1987. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this
ADM1TE 5/30A DIN rail single phase two wire energy meter
ADMTE 5/30A DIN rail single phase two wire energy meter. Safety instruction.2 Foreword.3 Performance criteria.4 Specifications.5 Basic errors.6 Description.7 Dimensions.8 Installation.9 Operating.0 Troubleshooting.
Model 1756 Test Lead Kit
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com Model 1756 Test Lead Kit Gerneral Purpose Test Lead Information Description These test leads allow you
e-4 AWT07MLED 7 Q TFT LCD MONITOR (LED Backlighted) USER MANUAL
Thank you for purchasing our product. Please read this User s Manual before using the product. Change without Notice AWT07MLED 7 Q TFT LCD MONITOR (LED Backlighted) USER MANUAL e-4 SAFETY PRECAUTIONS Federal
Doc.No. NDP 192U-02. Electric Pump Controller CE-124P. Instruction Manual
Doc.No. NDP 192U-02 Electric Pump Controller CE-124P Instruction Manual 1. Parts Name and Function 1) POWER LED Indicate that CE-124P is powered up. 2) START/STOP KEY (SELECTOR KEY) Used to start and stop
User's Manual. Metapace L-1. Label Printer Rev. 1.00
User's Manual Label Printer Rev. 1.00 Table of Contents Manual Information & Usage Precautions 3 1. Content Confirmation 6 2. Product Part Names 7 3. Installation & Usage 9 3-1 Power Connection 9 3-2 Interface
Users Manual. Temperature Calibrator
724 Temperature Calibrator Users Manual February 2000 Rev.1, 8/03 2000-2003 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies. LIMITED WARRANTY & LIMITATION