3.3: Atmospheric Pressure Chemical Ionisation (APcI)
|
|
|
- Janel York
- 9 years ago
- Views:
Transcription
1 OPDAC 3.3: Atmospheric Pressure Chemical Ionisation (APcI) Introduction The atmospheric pressure chemical ionisation (APcI) interface nebulises the LC column effluent in a heated vaporiser tube placed concentrically around the capillary. Once vaporised the plasma of eluent and sample components enters an API source where it encounters a cloud of electrons emitted from the tip of a corona electrode pin (Figure 3.3 (1)). The eluent molecules, which are in vast excess compared to the analyte, are ionised and act as the reagent gas in a chemical ionisation process that closely resembles that of medium pressure chemical ionisation (CI). Analyte molecules are ionised by the reagent gas plasma and are subsequently sampled into the nozzle-skimmer region of the mass spectrometer in a similar fashion as described for electrospray ionisation. Figure 3.3 (1): Schematic representation of an APcI source APcI interfacing for LC-MS is amongst the eldest approaches, initial work being carried out by Horning in However, due to the lack of commercially available instrumentation, the first wide-scale implementation of the technique took place in the early 1990 s.
2 Today, APcI approaches to analyte ionisation are amongst the most widely applied interface strategies. Due to the essential similarity between the electrospray and APcI interfaces most modern LC-API-MS systems are equipped for both ionisation modes with a minimum of operator interaction or equipment modification. Although APcI is not as widely implemented as electrospray ionisation, a growing number of applications is reported within the literature. The region of applicability for the APcI technique is shown in Figure 3.3 (2)) [1]. Figure 3.3 (2): Applicability of various modes of ionization in LC and GC -MS APcI is the method of choice for many drugs and metabolites and is probably the most widely used technique for high throughput target applications (such as BAPK and combinatorial techniques), particularly in the pharmaceutical industry. Figure 3.3 (4): Typical application of APcI in quantitative bio-analysis Suitable samples for APcI The sensitivity, ruggedness and reliability of APcI puts it ahead of electrospray ionisation techniques for many pharmaceutical applications. APcI appears to be much less sensitive to chemical interferences than electrospray-ms and the ionisation process associated with APcI is one of the most efficient, approaching 100% under ideal conditions (Figure 3.3 (5)). LC-MS for the Chromatographer Crawford Scientific,
3 Figure 3.3 (5): APcI interfaces are less prone to ion suppression than the ESI interface The only major limitation with APcI applies to all API techniques in that the ability to collect and detect all of the ions generated in the interface is limited, with only approximately 1% of ions formed being detected. Examples of analytes suitable for analysis via APcI include PAHs, PCBs, fatty acids, phthalates, and triglycerides. Samples that contain heteroatoms such as benzodiazepines and carbamates also work well. Figure 3.3 (6): Suitable sample types for APcI analysis LC-MS for the Chromatographer Crawford Scientific,
4 Avoid samples that are thermally labile as these may decompose or react in the heated nebuliser of the source. Samples that are typically charged in solution such as proteins, peptides, amino acids, or oligonucleotides should also be avoided, as sensitivity will be much reduced. These species also tend to be thermally labile and due to their high masses will be less volatile. Figure 3.3 (7): Unsuitable sample types for APcI analysis A summary of analyte types suitable for APcI-MS analysis may include: Small molecules that are moderately polar to non-polar Samples that contain heteroatoms Avoid samples that typically are charged in solution Avoid samples that are very thermally unstable or photosensitive APcI Interfacing Details Figure 3.3 (8): Schematic diagram of a typical APcI interface APcI interfaces usually consist of the following components (Figure 3.3 (8)); LC-MS for the Chromatographer Crawford Scientific,
5 An HPLC system delivering column eluent at flow rates between 0.5 and 2.0 ml/min. (i.e the majority of current conventional HPLC solvent delivery systems) A device for nebulising the column effluent and vaporisation of the droplets produced. An atmospheric-pressure ion source, containing a corona discharge electrode, and is capable of delivering the ions produced from the atmospheric pressure region into the high vacuum region of the mass spectrometer. In this respect, APcI sources differ from ESI sources ion two major respects; The inclusion of a heated region that extends beyond the nebulising capillary tip through which a gas is passed to rapidly vaporise the eluent and sample components The inclusion of a corona discharge pin to produce a cloud of electrons that are used to initiate the chemical ionisation process in the vapour phase. This section will therefore concentrate on the major differences between the APcI and the electrospray source which has already been discussed in detail Nebuliser Types In contrast to electrospray ionisation, APcI ionisation occurs in the gas phase as a result of strong heating of the nebulised eluent. Several nebulising and vaporising devices are available from a variety of manufacturers these are briefly described below; Figure 3.3 (9): APcI nebuliser design after Covet et. Al. The nebuliser first introduced by Covey et. al. [2] is used extensively in modern instrument designs and consists of a concentric pneumatic nebuliser and a quartz vaporiser tube extending beyond the tip of the nebuliser capillary (Figure 3.3. (9)). In this system flows of up to 1.5 ml/min. column eluent are nebulised with a inert gas (usually nitrogen) at pressures of approximately 0.8Mpa. LC-MS for the Chromatographer Crawford Scientific,
6 The aerosol created is swept through the heated nebuliser tube (usually o C), assisted by a make-up gas (nitrogen at 1-3 L/min., 0.5 MPa). This results in a vapour temperature of approximately 100 o C, which achieves almost complete nebulisation of the aerosol components. The corona discharge electrode is positioned, off axis, near the exit of the quartz vaporiser tube and this source type is available from Applied Biosystems (Foster City, CA, US. Pneumatic nebulisers, comprising three concentric tubes (i.e. a liquid carrying capillary in the center, a nebuliser gas tube and an auxiliary gas tube), are used extensively [3]. The nebuliser gas passes through a heater built into the tube and the pre-heated aerosol passes into a heated vaporiser region directly in-front of the nebuliser to complete the vaporisation process. The corona electrode is placed axially and near to or within the heated vaporiser region of the source. This source design has been used by manufacturers such as Micromass (Manchester, UK) and Thermo Finnigan (Manchester, UK). In general the flow rates of gas used as nebulising, auxiliary, make-up and counter current flows in the APcI interface vary widely between instrument manufacturers. The flow rates of gas used will also very depending upon chemical nature of the eluent system. However, it should be noted that the required nitrogen supply could be as high as 600 L/h in some systems with a line pressure of 0.7MPa this will almost always require the Nitrogen supply to be directly pumped from a liquid nitrogen pressurised vessel or from a nitrogen generating unit supplied with high-pressure compressed air supply Ion sampling and transfer in APcI interfaces Ultimately the products of the APcI interface are gas phase ion products of eluent and analyte components, very similar to the products of the electrospray interface with the addition of eluent reagent ions such as large water clusters and heavily hydrated gas phase acids and bases (Figure 3.3. (10)). Figure 3.3 (10): Formation of hydrated cluster ions in the APcI source Therefore, the hardware employed to de-cluster and transfer these gas phase ions to the high vacuum region of the mass spectrometer is essentially similar to that previously described for electrospray instruments. LC-MS for the Chromatographer Crawford Scientific,
7 The important aspects of the interface and ion transfer hardware are highlighted below by way of a summary and reminder of the material covered in previous sections; Figure 3.3 (11): Ion optics of a typical APcI source The atmospheric pressure region terminates in an ion sampling opening (sampling orifice) or a transfer capillary (heated), which acts as a nozzle at the low-pressure face. A region between the nozzle and a skimmer plate acts as a momentum jet separator and represents the first vacuum region of the instrument (Figure 3.3 (11)). Lenses may be employed in this region to promote the preconcentration of analyte ions relative to neutral and low molecular weight ionic species arising from the eluent. Figure 3.3 (12): Ion optics of a typical APcI source Some systems employ a second skimmer to allow a second differentially pumped region of the instrument. An ion focusing device such as an rf only, quadrupole, hexapole or octapole ion bridge used to pre-concentrate analyte species and direct the ion beam onto an axial path prior to the mass analysing device (Figure 3.3 (12)). This region is usually pumped by an oil diffusion or turbo-molecular pump. LC-MS for the Chromatographer Crawford Scientific,
8 Analyte ion declustering The effects of declustering analyte species require more attention with APcI, as many of the gas phase analyte ions produced are clustered with water molecules and reagent ions derived from the eluent system (Figure 3.3 (13)). Figure 3.3 (13): Cluster ion formation at the corona electrode in APcI Declustering of analyte ions may be achieved using one, or a combination, of the following approaches; Using a counter current gas flow at the nozzle plate, also known as a curtain gas. Figure 3.3 (14): Declustering using a curtain gas LC-MS for the Chromatographer Crawford Scientific,
9 Using a heated transfer capillary between the API region and the nozzleskimmer region. Figure 3.3 (15): Declustering using a heated transfer capillary into the nozzle-skimmer region Using a drift voltage between the nozzle and skimmer plates to promote intermolecular collisions between the analyte clusters and the background gas molecules known as collision induced dissociation. Figure 3.3 (16): Declustering using collision induced dissociation Large solvent clusters may also form that do not contain an analyte ion and it is important that these clusters are removed to prevent noisy baselines and obscuring of the analyte signal. Increasing the drying gas flow and temperature in the main interface can reduce the formation of the solvent clusters. However, too high a drying gas flow will help to eliminate solvent clusters but will also attenuate the signal from the analyte [4]. LC-MS for the Chromatographer Crawford Scientific,
10 Figure 3.3 (17): Drying gas flow and nebuliser temperatures should be optimized to reduce cluster ion formation APcI Ionisation Mechanisms Overview (1) APcI is based on chemical ionisation by ion-molecule or electron capture reactions carried out on gas phase products of the vaporised LC eluent at atmospheric pressure. APcI can occur in a number of ways; Proton transfer or charge exchange in the positive ion mode LC-MS for the Chromatographer Crawford Scientific,
11 Figure 3.3 (18): Ionisation mechanisms in APcI (+ve mode) Proton absraction or electron capture in the negative ion mode. Figure 3.3 (19): Ionisation mechanisms in APcI (-ve mode) Charge transfer with positive ions takes place between a reactant ion generated from a molecule possessing a high ionisation potential, and a gas phase molecule with a lower ionisation potential. With a careful choice of the reactant ion, charge transfer can be a selective ionisation method. Benzene molecular ions have been used in positive ion mode as the reactant for the selective ionisation of polychlorinated biphenyls and the O 2 - ion may be perfused into the API source to promote charge transfer reaction in negative ion mode. Electron capture occurs if thermalised electrons are present in the API source, where suitable sample molecules are ionised by resonance electron capture, producing molecular anions, or by dissociative electron capture, producing negative fragment ions. Suitable analyte molecules posses electronegative functional groups, such as conjugated carbonyl species, nitro groups or halogen atoms Overview (2) The majority of ion-molecule reactions in APcI are gas phase acid-base type reactions [5]. For positive ion operation the reactant ions are acids - protonated water, methanol and acetonitrile are all mild acids and the ammonium ion is a weak selective acid (Figure 3.3 (20)) [6]. For negative ion operation, the reactant ions are bases hydroxide ions are stong gas phase bases, CH 3 O - and NCCH 2 - are mild bases and CH 3 COO - and Cl - are weak bases. If reactant gases are mixed, protons are transferred to form the weakest possible reactant ion (Figure 3.3 (21)). If acetonitrile is added to water, H 3 O + transfers its proton to the acetonitrile gas phase ion. Similarly if ammonia is added to LC-MS for the Chromatographer Crawford Scientific,
12 methanol vapour, the initially formed protonated methanol transfers its proton to ammonia, giving NH 4 +. It is not possible to convert a weak reagent into a stronger one. Figure 3.3 (20): Gas phase acidity and basicity of various reactant species in APcI Figure 3.3 (21): Formation of weakest reagent ion in APcI Which ionization mechanism dominates depends upon the chemistry of the mobile phase and analyte as well as the ion polarity (either positive or negative). In general however, negative APcI does not yield the same dramatic increase in sensitivity found when operating in negative CI in the gas phase in gas chromatography-mass spectrometry (GC-MS) Reagent Gas Formation The chemical ionisation reactions are initiated by means of electrons emitted from a corona discharge electrode kept at 5-10kV with a discharge current of 1-5µA. The energetic electrons released from the tip of the corona electrode (or corona pin as it is more usually known), begin a series of reactions; N 2 + e - N e - N 2 + 2N 2 N N 2 LC-MS for the Chromatographer Crawford Scientific,
13 The major benefit to ionisation at atmospheric pressure is that all gas phase ions undergo significant collisions with surrounding gas phase molecules. This results in interactions of all surrounding gases (N 2, H 2, H 2 O and air) and a cascade reaction induced is shown (Figure 3.3 (22)) [7]. Figure 3.3 (22): Cascade reaction resulting in water cluster reagent ions The major resulting ions are N 2 +, O 2 +, H 2 O + and NO + and in the presence of only trace amounts of water, cascade reactions are initiated that ultimately result in production of water cluster ions, which predominate the reagent gas mass spectrum. Since these water cluster ions are the major source of reagent ions, the proton affinity of these clusters relative to the gas phase proton affinity of the analyte and eluent ions has a great affect on the sensitivity [8-9]. Figure 3.3 (23): Predicting reactivity in APcI based on proton affinity LC-MS for the Chromatographer Crawford Scientific,
14 relative to the main reagent ions In order to make some rudimentary predictions about the relative sensitivities of each analyte ion and any possible suppression effects attributable to eluent components such as volatile buffers, a consideration of relative proton affinity can be useful (Figure 3.3 (23)). The optimum response region represented on Figure 3.3 (23) above is for analytes with greater proton affinity than water but one must consider than proton affinity for ALL eluent components as the buffers and additives used may compete for protons with the analyte, potentially reducing sensitivity [10]. Figure 3.3 (24): Proton affinity values for selected reagent ions and organic species in APcI Effect of eluent additives on APcI Ionisation (1) Gas phase acid-base chemistry is of overriding importance in LC-APcI-MS where reactant gas mixtures are the rule rather than the exception. Additives in the LC eluent used to improve separation may entirely change the chemical ionisation process. When eluents are used with additives and volatile buffers the APcI reactant ion spectrum is dominated by solventderived reactant ions and their clusters with water, methanol etc. [11]. If ammonium acetate is added to the eluent, the reactant ions are ammonium and acetate ions, clustered with water, ammonia, acetic acid and other polar components in the eluent system. LC-MS for the Chromatographer Crawford Scientific,
15 Figure 3.3 (25): Eluent additive cluster ion spectrum in the APcI interface Where water is the main component in the eluent system, the most abundant positive reactant ion series is H 3 O + clustered with water molecules, as has been previously stated. It can be demonstrated that the gas-phase acidity of H 3 O + (H 2 O)n decreases with increasing n. The ionisation of weakly basic samples is less effective or even not possible if n is increased. A decrease of n with increase of proton transfer efficiency is effected at higher temperatures in the interface region around the corona discharge electrode [11]. Figure 3.3 (26): With increased source temperature water cluster ion gas-phase acidity increases ensuring efficient ionization of the analyte Effect of eluent additives on APcI Ionisation (2) Gas phase acid-base chemistry is of overriding importance in LC-APcI-MS where reactant gas mixtures are the rule rather than the exception. Figure 3.3 (27): Effect of solvents / eluent additives on instrument response in APcI LC-MS for the Chromatographer Crawford Scientific,
16 In general, because of the possibility of charge exchange reactions, methanol is a better choice in APcI-MS than acetonitrile. Using acetonitrile will tend to reduce the sensitivity of the analysis (Figure 3.3 (27)) [12]. Samples with lower proton affinities than ammonia or triethylamine will often loose a proton, become neutralised or form adducts in the gas phase when these eluent additive species are present and therefore their use should be carefully controlled (Figure 3.3 (27)). One very common problem with gas-phase acid base chemistry is signal quenching by acidic mobile-phase constituents in negative ion APcI. Addition of 10mmol/L formic acid or 10 to 100 mmol/l ammonium acetate to the LC eluent significantly decreased the response of selected model compounds in a study by Schafer and Dixon [13]. The extent of the quenching (or signal suppression) was directly correlated to gas phase acidities and basicities. The use of N-methyl morpholine in the eluent appears to be a better alternative in terms of signal quenching. In general positive ion APcI is preferred over negative ion mode whenever this is possible Other ionisation mechanisms in APcI A less well documented mechanism for APcI is derived via the triboelectric effect (Figure 3.3 (28)). As the mobile phase and analyte species exit the nebuliser, the sheer forces generated by the nebulising gas tear the liquid stream into droplets. The friction generated by this process generates an electric charge (the triboelectric effect) at the liquid gas interface. This charge can cause ion formation in some non-volatile analytes. Figure 3.3 (28): Other ionization processes possible in the APcI interface LC-MS for the Chromatographer Crawford Scientific,
17 Triboelectric APcI does not depend upon the electrons generated by the corona dicharge electrode and ions may be generated (with their associated signals in the mass spectrum), even when the corona discharge is very low or even turned off. This effect will also occur when the corona electrode is switched on. Triboelectric APcI is most common for analyte molecules that are moderately polar and/or non-volatile. Ion ejection is the mechanism that is responsible for most of the ion production in ESI-MS, but may also occur in APcI-MS (Figure 3.3 (28)). Some analyte molecules in the LC eluent will exist as ions in the solution before they reach the APcI ion source. After nebulisation, the droplets of eluent pass through the vaporiser tube or heated region and as they shrink due to solvent vaporisation the pre-formed analyte ion concentrate at the droplet surface. From this point the mechanism of gas phase ion production closely match that described in the previous sections dealing with ESI-MS. This mechanism is common with analytes that are highly polar or ionic in solution and may be promoted or suppressed by adjusting the ph of the eluent solution APcI Source Parameter Optimisation APcI Source Parameter Optimisation (1) In general APcI-MS interfaces are considered as extremely simple to operate. This is bourn out by the fact that hardly any optimisation of the interface parameters is reported in many of the APcI application papers. Choosing the optimum settings for the more important interface parameters such as the eluent flow rate, the solvent composition, the nebuliser and auxiliary gas flows and the vaporiser temperature, which are highly interrelated, is far less critical than for electrospray interfaces. Figure 3.3 (29): Effect of vapourising temperature LC-MS for the Chromatographer Crawford Scientific,
18 on instrument response in APcI Of all the parameters that may be optimised within the APcI, the vaporising probe temperature has perhaps the most profound effect on signal intensity (Figure 3.3 (29)). The case study shows Vitamin D3 and furosemide showing little or no signal at a probe temperature of 200 o C [14]. As the temperature is raised to 400 o C the signal due to these analytes has improved appreciably this may be explained due to the increasing efficiency of volatilisation of these two analytes, ensuring full conversion to the gas phase in the region around the corona electrode pin. It should be noted that this approach may not be suitable for thermo-labile analytes that may decompose in the heated zone that can reach temperatures in excess of 200 o C. A further interesting point of note is the APcI signal intensity of compounds such as penicillin, methylene blue, basic yellow 2, and basic violet 10, which are normally charged in solution. These compounds will be less volatile in typical reverse phase eluent systems and so volatilisation will be less effective due to the high intermolecular attractions between the charged analyte species and the polar eluent constituents. Where ph may not be used to ensure that the analyte is neutral in solution, the interface of choice will be ESI. Similarly, higher molecular weight species such as gentamicin and streptomycin will be less volatile and therefore will not be efficiently ionised in the gas phase at the corona electrode. Again the ionisation method of choice will be ESI. In general, higher molecular weight species such as peptides and proteins are not suitable for APcI analysis due to the extremely high temperatures required to vaporise these species in the region immediately after the nebuliser APcI Source Parameter Optimisation (2) As has already been discussed, any non-volatile buffer species within the LC eluent can act to trap the analyte species either through intermolecular attractions between the less volatile species and the analyte or by a physical trapping effect as the non-volatile particles agglomerate within the evaporating eluent droplets (Figure 3.3 (30)). To demonstrate typical operating conditions for the APcI interface and highlight the generic approach to optimisation an experiment performed by Garcia et.al [15] is outlined. LC-MS for the Chromatographer Crawford Scientific,
19 Figure 3.3 (30): Trapping of analyte species by non-volatile eluent additives leading to reduced instrument sensitivity.. Ten parameters were evaluated for their effect on the performance of the Finnigan MAT APcI system: Mobile phase (5 95% water) Flow rate (0.1 1 ml/min.) Acetic acid concentration (0.1% in water) Nebuliser gas pressure ( Mpa) Vaporiser temperature ( o C) Corona discharge current ( ma) Transfer capillary temperature ( o C) Capillary voltage (20-200V) Tube lens voltage (20-200V) Figure 3.3 (31): APcI source parameter optimization after Garcia et. Al. LC-MS for the Chromatographer Crawford Scientific,
20 Ibuprofen was chosen as the model compound with optimum performance found at a flow rate of 0.1 ml/min., nebuliser gas pressure of 0.42 Mpa, transfer capillary temperature of 225 o C, tube lens voltage 55V. These settings proved to be the optimum setting not only for ibuprofen but also for 40 other chemical and pharmaceutical compounds. A caveat to the above findings however must be that all analytes will need a small amount of optimisation for each new experimental set up although, in general, the optimisation required for APcI interfaces will be much less than for the electrospray interface. LC-MS for the Chromatographer Crawford Scientific,
21 3.3.6 References [1] W.M.A Niessen and A.P. Tinke, J. Chromatogr. A., 703, (1995), [2] T.R. Covey, E.D. Lee and J.D. Henion, Anal. Chem., 58, (1986), [3] D.R. Doerge, S. Bajic and S. Lowes, Rapid Commun. Mass Spectrom., 7, (1993), [4] E.C. Horning, D.I. Carroll, I. Dzidic, S.-N. Lin, R.N. Stillwell and J.-P. Thenot, J. Chromatogr., 142, (1977), [5] A.P. Bruins, Adv. Mass Spectrom., 10, (1986), [6] A.P. Bruins, Mass Spectrom. Reviews, 10, (1991), [7] M.L. Huertas, J. Fontan, Evolution Times of the Tropospheric Positive Ions, Atmospheric Environ., 9, (1875), 1018 as shown in The API book, PE Sciex, Toronto, Canada. [8] J. Sunner, G. Nichol and P. Kebarle, Anal. Chem., 60, (1988), [9] J. Sunner, M.G. Ikonomou, G. Nichol and P. Kebarle, Anal. Chem., 60, (1988), [10] Data reproduced from Agilent Technologies Publication number G , (1998), Palo Alto, CA, USA. [11] F. Hunt, Reagent gases for CI. Adv. Mass Spectrom., 6, (1974), [12] Data reproduced from Agilent Technologies 1100 LCMSD Solution Chemistry Handbook, Agilent Technologies, Palo Alto, Ca, USA, [13] W.H. Schaefer and F. Dixon, Jr, J. Am. Soc. Mass Spectrom., 7 (1996), [14] Data reproduced from Agilent Technologies 1100 LCMSD Solution Chemistry Handbook, Agilent Technologies, Palo Alto, Ca, USA, [15] D.M. Garcia, S. K. Huang and W.F. Stansbury, J. Am. Soc. Mass Spectrom., 7, (1996), 59. LC-MS for the Chromatographer Crawford Scientific,
Electrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION
X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
Separation of Amino Acids by Paper Chromatography
Separation of Amino Acids by Paper Chromatography Chromatography is a common technique for separating chemical substances. The prefix chroma, which suggests color, comes from the fact that some of the
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
CHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING
CHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING Essential Standard: STUDENTS WILL UNDERSTAND THAT THE PROPERTIES OF MATTER AND THEIR INTERACTIONS ARE A CONSEQUENCE OF THE STRUCTURE OF MATTER,
Basics of LC/MS. A Primer
Basics of LC/MS A Primer Contents Introduction..................................... 2 Instrumentation.................................. 4 Interfacing LC and MS................................. 4 Atmospheric
Name Lab #3: Solubility of Organic Compounds Objectives: Introduction: soluble insoluble partially soluble miscible immiscible
Lab #3: Solubility of rganic Compounds bjectives: - Understanding the relative solubility of organic compounds in various solvents. - Exploration of the effect of polar groups on a nonpolar hydrocarbon
Analysis of Vegetable Oils by High Performance Liquid Chromatography
Analysis of Vegetable Oils by High Performance Liquid Chromatography Using Evaporative Light Scattering Detection and Normal Phase Eluents Andrew J. Aubin, Cecilia B. Mazza, Donald A. Trinite, and Patricia
How To Use Gc-Ms
The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Brønsted-Lowry Acids and Bases
Brønsted-Lowry Acids and Bases 1 According to Brønsted and Lowry, an acid-base reaction is defined in terms of a proton transfer. By this definition, the reaction of Cl in water is: Cl(aq) + Cl (aq) +
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
Chapter 2 Polar Covalent Bonds; Acids and Bases
John E. McMurry http://www.cengage.com/chemistry/mcmurry Chapter 2 Polar Covalent Bonds; Acids and Bases Javier E. Horta, M.D., Ph.D. University of Massachusetts Lowell Polar Covalent Bonds: Electronegativity
Vacuum Evaporation Recap
Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,
13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
Ionization Methods in Organic Mass Spectrometry Contents:
Ionization Methods in Organic Mass Spectrometry Contents:. Introduction. Gas-Phase Ionization. Electron Ionization (EI). Chemical Ionization (CI). Desorption Chemical Ionization (DCI). Negative-ion chemical
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
Chem101: General Chemistry Lecture 9 Acids and Bases
: General Chemistry Lecture 9 Acids and Bases I. Introduction A. In chemistry, and particularly biochemistry, water is the most common solvent 1. In studying acids and bases we are going to see that water
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
Oasis HLB Cartridges and 96-Well Plates
CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
Matrix Effect in Bio-Analysis of Illicit Drugs with LC-MS/MS: Influence of Ionization Type, Sample Preparation, and Biofluid
Matrix Effect in Bio-Analysis of Illicit Drugs with LC-MS/MS: Influence of Ionization Type, Sample Preparation, and Biofluid Riet Dams and Marilyn A. Huestis Chemistry and Drug Metabolism, National Institute
The Theory of HPLC. Gradient HPLC
The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims
Chemistry Assessment Unit AS 1
Centre Number 71 Candidate Number ADVANCED SUBSIDIARY (AS) General Certificate of Education January 2011 Chemistry Assessment Unit AS 1 assessing Basic Concepts in Physical and Inorganic Chemistry [AC111]
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
Chemistry Diagnostic Questions
Chemistry Diagnostic Questions Answer these 40 multiple choice questions and then check your answers, located at the end of this document. If you correctly answered less than 25 questions, you need to
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
ph: Measurement and Uses
ph: Measurement and Uses One of the most important properties of aqueous solutions is the concentration of hydrogen ion. The concentration of H + (or H 3 O + ) affects the solubility of inorganic and organic
Hardware Manual. AB SCIEX API 4000 LC/MS/MS System
Hardware Manual AB SCIEX API 4000 LC/MS/MS System Part Number: 5005565 A April 2010 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
Since we will be dealing with aqueous acid and base solution, first we must examine the behavior of water.
Acids and Bases Know the definition of Arrhenius, Bronsted-Lowry, and Lewis acid and base. Autoionization of Water Since we will be dealing with aqueous acid and base solution, first we must examine the
Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites
Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Carmen T. Santasania and David S. Bell SUPELCO, 595 North Harrison Road, Bellefonte, PA 16823 T404045 GUS Abstract
Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide
chromatography Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide The following guide provides considerations, tips and general guidelines for developing SPE methods using the Thermo
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
Chapter 17. How are acids different from bases? Acid Physical properties. Base. Explaining the difference in properties of acids and bases
Chapter 17 Acids and Bases How are acids different from bases? Acid Physical properties Base Physical properties Tastes sour Tastes bitter Feels slippery or slimy Chemical properties Chemical properties
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Nuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass
Protons, neutrons and electrons Nuclear Structure particle relative charge relative mass proton 1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass Protons and neutrons make up
SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322. [email protected].
SPE and HPLC Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322 [email protected] Solid-Phase Extraction- SPE Simple, fast and efficient sample preparation
CHEMISTRY II FINAL EXAM REVIEW
Name Period CHEMISTRY II FINAL EXAM REVIEW Final Exam: approximately 75 multiple choice questions Ch 12: Stoichiometry Ch 5 & 6: Electron Configurations & Periodic Properties Ch 7 & 8: Bonding Ch 14: Gas
States of Matter CHAPTER 10 REVIEW SECTION 1. Name Date Class. Answer the following questions in the space provided.
CHAPTER 10 REVIEW States of Matter SECTION 1 SHORT ANSWER Answer the following questions in the space provided. 1. Identify whether the descriptions below describe an ideal gas or a real gas. ideal gas
Name Class Date. In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question.
Assessment Chapter Test A Chapter: States of Matter In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1. The kinetic-molecular
Chemistry. The student will be able to identify and apply basic safety procedures and identify basic equipment.
Chemistry UNIT I: Introduction to Chemistry The student will be able to describe what chemistry is and its scope. a. Define chemistry. b. Explain that chemistry overlaps many other areas of science. The
Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)
Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III) Purpose: Verify the synthesis of the synthesis product, Cr(acac) 3. The presence of any urea containing byproducts will also be determined.
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection LGC/R/2011/181 LGC Limited 2011 GC11/0005A A feasibility study of the determination of B-group
Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance
Thermo Scientific SOLA SPE cartridges and plates Technical Guide Join the revolution unparalleled performance Join the revolution next-generation SPE Thermo Scientific SOLA products revolutionize Solid
TOF FUNDAMENTALS TUTORIAL
TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] [email protected] [e-mail] This
INTI COLLEGE MALAYSIA A? LEVEL PROGRAMME CHM 111: CHEMISTRY MOCK EXAMINATION: DECEMBER 2000 SESSION. 37 74 20 40 60 80 m/e
CHM111(M)/Page 1 of 5 INTI COLLEGE MALAYSIA A? LEVEL PROGRAMME CHM 111: CHEMISTRY MOCK EXAMINATION: DECEMBER 2000 SESSION SECTION A Answer ALL EIGHT questions. (52 marks) 1. The following is the mass spectrum
IB Chemistry. DP Chemistry Review
DP Chemistry Review Topic 1: Quantitative chemistry 1.1 The mole concept and Avogadro s constant Assessment statement Apply the mole concept to substances. Determine the number of particles and the amount
Chem 1B Saddleback College Dr. White 1. Experiment 8 Titration Curve for a Monoprotic Acid
Chem 1B Saddleback College Dr. White 1 Experiment 8 Titration Curve for a Monoprotic Acid Objectives To learn the difference between titration curves involving a strong acid with a strong base and a weak
Chemistry B11 Chapter 6 Solutions and Colloids
Chemistry B11 Chapter 6 Solutions and Colloids Solutions: solutions have some properties: 1. The distribution of particles in a solution is uniform. Every part of the solution has exactly the same composition
The Properties of Water
1 Matter & Energy: Properties of Water, ph, Chemical Reactions EVPP 110 Lecture GMU Dr. Largen Fall 2003 2 The Properties of Water 3 Water - Its Properties and Its Role in the Fitness of Environment importance
A Novel Approach to Low Volume Sample Preparation
A Novel Approach to Low Volume Sample Preparation Tony Edge, Jon Bardsley, Ken Meadows, Luisa Pereira International Symposium on Chromatography Salzburg, Austria September 2014 1 The world leader in serving
The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC
The Essential CHROMacademy Guide Mobile Phase Optimization Strategies for Reversed Phase HPLC Webcast Notes Type your questions in the Submit Question box, located below the slide window You can enlarge
Keystone Exams: Chemistry Assessment Anchors and Eligible Content. Pennsylvania Department of Education www.education.state.pa.
Assessment Anchors and Pennsylvania Department of Education www.education.state.pa.us 2010 PENNSYLVANIA DEPARTMENT OF EDUCATION General Introduction to the Keystone Exam Assessment Anchors Introduction
Gas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
Purification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
Composition of the Atmosphere. Outline Atmospheric Composition Nitrogen and Oxygen Lightning Homework
Molecules of the Atmosphere The present atmosphere consists mainly of molecular nitrogen (N2) and molecular oxygen (O2) but it has dramatically changed in composition from the beginning of the solar system.
Getting the most from this book...4 About this book...5
Contents Getting the most from this book...4 About this book....5 Content Guidance Topic 1 Atomic structure and the periodic table...8 Topic 2 Bonding and structure...14 Topic 2A Bonding....14 Topic 2B
Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Theory
Mass Spectrometry Fundamental LC-MS Electrospray Ionisation Theory i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown
Molecular Spectroscopy
Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
hij GCSE Additional Science Chemistry 2 Higher Tier Chemistry 2H SPECIMEN MARK SCHEME Version 1.0
hij GCSE Additional Science Chemistry 2 Higher Tier Chemistry 2H SPECIMEN MARK SCHEME Version.0 Copyright 20 AQA and its licensors. All rights reserved. The Assessment and Qualifications Alliance (AQA)
Organic Functional Groups Chapter 7. Alcohols, Ethers and More
Organic Functional Groups Chapter 7 Alcohols, Ethers and More 1 What do you do when you are in Pain? What do you do when you are in a lot of pain? 2 Functional Groups A functional group is an atom, groups
Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.
Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine
Contents. Contributors. 1 Quality control and regulation 1 C.J. MOORES
Contents Contributors Preface xiii xiv 1 Quality control and regulation 1 C.J. MOORES 1.1 Introduction 1 1.2 The quality of medicines 2 1.2.1 The meaning of quality 2 1.2.2 Medicines are special 3 1.2.3
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
Guide to achieving reliable quantitative LC-MS measurements
Guide to achieving reliable quantitative LC-MS measurements Guide to achieving reliable quantitative LC-MS measurements RSC Analytical Methods Committee Editor Mike Sargent First Edition 2013 Contributors
AAHS-CHEMISTRY FINAL EXAM PREP-REVIEW GUIDE MAY-JUNE 2014 DR. GRAY CLASS OF 2016
AAHS-CHEMISTRY FINAL EXAM PREP-REVIEW GUIDE MAY-JUNE 2014 DR. GRAY CLASS OF 2016 UNIT I: (CHAPTER 1-Zumdahl text) The Nature of Science and Chemistry 1. Explain why knowledge of chemistry is central to
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
Write the acid-base equilibria connecting all components in the aqueous solution. Now list all of the species present.
Chapter 16 Acids and Bases Concept Check 16.1 Chemists in the seventeenth century discovered that the substance that gives red ants their irritating bite is an acid with the formula HCHO 2. They called
CHE334 Identification of an Unknown Compound By NMR/IR/MS
CHE334 Identification of an Unknown Compound By NMR/IR/MS Purpose The object of this experiment is to determine the structure of an unknown compound using IR, 1 H-NMR, 13 C-NMR and Mass spectroscopy. Infrared
Coating Technology: Evaporation Vs Sputtering
Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information
New drugs of abuse? A case study on Phenazepam & Methoxetamine
New drugs of abuse? A case study on Phenazepam & Methoxetamine Presenter: Nadia Wong Co authors: Dr Yao Yi Ju & Alex Low Xuan Kai Analytical Toxicology Laboratory Clinical & Forensic Toxicology Unit Applied
Chemistry 1050 Chapter 13 LIQUIDS AND SOLIDS 1. Exercises: 25, 27, 33, 39, 41, 43, 51, 53, 57, 61, 63, 67, 69, 71(a), 73, 75, 79
Chemistry 1050 Chapter 13 LIQUIDS AND SOLIDS 1 Text: Petrucci, Harwood, Herring 8 th Edition Suggest text problems Review questions: 1, 5!11, 13!17, 19!23 Exercises: 25, 27, 33, 39, 41, 43, 51, 53, 57,
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen
Acids and Bases: A Brief Review
Acids and : A Brief Review Acids: taste sour and cause dyes to change color. : taste bitter and feel soapy. Arrhenius: acids increase [H ] bases increase [OH ] in solution. Arrhenius: acid base salt water.
Chapter 5 Classification of Organic Compounds by Solubility
Chapter 5 Classification of Organic Compounds by Solubility Deductions based upon interpretation of simple solubility tests can be extremely useful in organic structure determination. Both solubility and
Chapter 14 - Acids and Bases
Chapter 14 - Acids and Bases 14.1 The Nature of Acids and Bases A. Arrhenius Model 1. Acids produce hydrogen ions in aqueous solutions 2. Bases produce hydroxide ions in aqueous solutions B. Bronsted-Lowry
Chapter 1: Moles and equations. Learning outcomes. you should be able to:
Chapter 1: Moles and equations 1 Learning outcomes you should be able to: define and use the terms: relative atomic mass, isotopic mass and formula mass based on the 12 C scale perform calculations, including
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
This value, called the ionic product of water, Kw, is related to the equilibrium constant of water
HYDROGEN ION CONCENTRATION - ph VALUES AND BUFFER SOLUTIONS 1. INTRODUCTION Water has a small but definite tendency to ionise. H 2 0 H + + OH - If there is nothing but water (pure water) then the concentration
ORGANIC CHEMISTRY I PRACTICE PROBLEMS FOR BRONSTED-LOWRY ACID-BASE CHEMISTRY
RGANIC CHEMISTRY I PRACTICE PRBLEMS FR BRNSTED-LWRY ACID-BASE CHEMISTRY 1. For each of the species below, identify the most acidic proton and provide the structure of the corresponding conjugate base.
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
The 30-Minute Guide to ICP-MS
Technical Note The 30-Minute Guide to ICP-MS ICP-Mass Spectrometry A Worthy Member of the Inorganic Analysis Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been gaining
Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation
Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Craig Aurand 1, David Bell 1, Anders Fridstrom 2, and Klaus Buckendahl 2 1 Supelco, Div.
Inferred ph in Steam Plant Water Chemistry Monitoring
Application Data Sheet ADS 4900-87/rev.B January 2009 Power Industry Inferred ph in Steam Plant Water Chemistry Monitoring INTRODUCTION Inferred ph means ph calculated from straight and cation conductivity.
Chapter 2 Polar Covalent Bonds: Acids and Bases
John E. McMurry www.cengage.com/chemistry/mcmurry Chapter 2 Polar Covalent Bonds: Acids and Bases Modified by Dr. Daniela R. Radu Why This Chapter? Description of basic ways chemists account for chemical
The content is based on the National Science Teachers Association (NSTA) standards and is aligned with state standards.
Literacy Advantage Physical Science Physical Science Literacy Advantage offers a tightly focused curriculum designed to address fundamental concepts such as the nature and structure of matter, the characteristics
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.
Chapter 8: Chemical Equations and Reactions
Chapter 8: Chemical Equations and Reactions I. Describing Chemical Reactions A. A chemical reaction is the process by which one or more substances are changed into one or more different substances. A chemical
Chemistry. CHEMISTRY SYLLABUS, ASSESSMENT and UNIT PLANNERS GENERAL AIMS. Students should be able to
i CHEMISTRY SYLLABUS, ASSESSMENT and UNIT PLANNERS GENERAL AIMS Students should be able to - apply and use knowledge and methods that are typical to chemistry - develop experimental and investigative skills,
Molecular Models in Biology
Molecular Models in Biology Objectives: After this lab a student will be able to: 1) Understand the properties of atoms that give rise to bonds. 2) Understand how and why atoms form ions. 3) Model covalent,