Pulse Oximetry. Principles of oximetry
|
|
|
- Sharleen Freeman
- 9 years ago
- Views:
Transcription
1 Pulse Oximetry The principal advantage of optical sensors for medical applications is their intrinsic safety since there is no electrical contact between the patient and the equipment. (An added bonus is that they are also less suspect to electromagnetic interference). This has given rise to a variety of optical techniques to monitor physiological parameters: for example, the technique of Laser Doppler velocimetry to measure red blood cell velocity. However, in this lecture course we will concentrate on the technique of pulse oximetry for the non invasive measurement of arterial oxygen saturation in the blood (although a second use of the technology will be discussed right at the end of the course). For patients at risk of respiratory failure, it is important to monitor the efficiency of gas exchange in the lungs, ie how well the arterial blood is oxygenated (as opposed to whether or not air is going in and out of the lungs). Preferably, such information should be available to clinicians of a continuous basis (rather than every few hours). Both of these requirements can be met non invasively 2 with the technology of pulse oximetry. The technique is now well established and is in regular clinical use during anaesthesia and intensive care (especially neonatal intensive care since many premature infants undergo some form of ventilator therapy). Pulse oximetry is also being used in the monitoring of pulmonary disease in adults and in the investigation of sleep disorders. Principles of oximetry It was discovered in the 1860 s that the coloured substance in blood, haemoglobin, was also its carrier of oxygen. (Haemoglobin is a protein which is bound to the red blood cells.) At the same time, it was noticed that the absorption of visible light by a haemoglobin solution varied with oxygenation. This is because the two common forms of the molecule, oxidised haemoglobin (HbO 2 ) and reduced haemoglobin (Hb), have significantly different optical spectra in the wavelength range from 500nm to 1000nm, as shown in Figure 33. Figure 33: Absorption spectra of Hb and HbO 2 (the isobestic point is the wavelength at which the absorption by the two forms of the molecule is the same) 2 ie using instrumentation which does not require surgery
2 The oxygen chemically combined with haemoglobin inside the red blood cells makes up nearly all of the oxygen present in the blood (there is also a very small amount which is dissolved in the plasma). Oxygen saturation, which is often referred to as SaO 2 or SpO 2, is defined as the ratio of oxyhaemoglobin (HbO 2 ) to the total concentration of haemoglobin present in the blood (ie oxyhaemoglobin + reduced haemoglobin): SaO 2 HbO 2 Total haemoglobin Arterial SaO 2 is a parameter measured with oximetry and is normally expressed as a percentage. Under normal physiological conditions arterial blood is 97% saturated, whilst venous blood is 75% saturated. It is possible to use the difference in absorption spectra of HbO 2 and Hb for the measurement of arterial oxygen saturation in vivo because the wavelength range between 600 nm and 1000 nm is also the range for which the is least attenuation of light by body tissues (tissue and pigmentation absorb blue, green and yellow light and water absorbs the longer infra red wavelength). I in I Figure 34 By measuring the light transmitted through the fingertip (or the earlobe) at two different wavelengths, one in the red and the other in the near infra red part of the spectrum, the oxygen saturation of the arterial blood in the finger (or ear). If we assume initially that the transmission of light through the arterial bed is influenced only by the relative concentrations of HbO 2 and Hb and their absorption coefficients at the two measurement wavelengths, then the light intensity will decrease logarithmically with path length according to the well known Beer Lambert law. If we consider, as shown in Figure 34, an artery of length l through which light, initially of intensity I in is passed this law tells us that: At wavelength λ 1, I 1 o1 I in1 10 At wavelength λ 2, I 2 o2 I in2 10 o! C r1 o! C r2 C r l C r l where " C o is the concentration of oxyhaemoglobin (HbO 2 ) " C r is the concentration of reduced haemoglobin (Hb) " α on is the absorption coefficient of HbO2 at wavelength λ n " α rn is the absorption coefficeint of Hb at wavelength λ n If we let # log R 10 I 1 I # in1 log 10 I 2 I in2 artery then it is simple enough to show: l
3 SaO 2 C o C o $ C r % r2 R& % r1 % r2& % o2 R& % r1& % o1 (1) A brief history of oximetry The first devices to measure oxygen saturation in human blood by transilluminating it with coloured light were built in the 1930 s. These devices were incapable of distinguishing between arterial and venous (and capillary) blood. An attempt to include venous and capillary blood was then made by using one of two methods: zeroing the oximeter by taking a bloodless reading from an earlobe compressed between two fingers or arterialising the blood by heating it to 43 C (this method was developed during the Second World War as part of a project to investigate the problem of the loss of consciousness by R.A.F. pilots during dog fights). However, the light transmitted through the ear (or finger) is not only attenuated by the arterial, venous and capillary blood but also by the skin (whose pigmentation and hence absorption properties will vary from person to person) and by other tissues such as muscle, bone, etc... In the early 1970 s, the Hewlett Packard company (the medical division of which is now known as Agilent) developed an instrument which attempted to circumvent these problems by measuring the transmision of light across the earlobe at more than two wavelengths. A multi component model of the ear was set, composed of m light absorbing substances (skin, tissues, Hb and HbO 2, etc...). The model further assumed that each light absorber acted independently of the others. A set of simultaneous equations were then written for the total absorbance of light by the m substances at each of the measurement wavelengths. Empirical calibration coefficients were derived from a number of studies on a sample of volunteers. However, the high cost of the instrumentation together with the need for measurements at eight different wavelengths meant that it never found regular clinical use. Principles of pulse oximetry It is the recent development of pulse oximetry which has led to oximetry being accepted as a useful non invasive technique for the measurement of arterial SaO 2. With pulse oximetry, only that part of the signal directly related to the inflow of arterial blood into the body segment is used for the calculation of oxygen saturation. The intensity of light transmitted across the fingertip, for example, varies as shown in Figure 35. A pulsatile signal, which varies in time with the heart beat, is superimposed on a d.c. level. (The amplitude of this cardiac synchronous pulsatile signal is approximately 1% of the d.c. level. Pulse oximetry assumes that the attenuation of light by the body segment can be split into the three independent components shown in Figure 35(a): arterial blood, venous blood and tissues.
4 (a) (b) (c) Figure 35: (a) Transmission of light through the finger when the attenuation of light is caused by arterial blood (A), venous blood (V) and tissues (T). (b) and (c) show typical pulsatile signals detected in the intensity of detected light when light is shone through a finger. If we assume that the increase in attenuation of light is caused only by the inflow of arterial blood into the fingertip, we can calculate the oxygen saturation of the arterial blood by subtracting the d.c. component of the attenuation from the total attenuation, leaving only the cardiac synchronous pulsatile component for the dual wavelength determination of oxygen saturation. It is a simple matter to show (as you will be asked to do in one of the questions on the Problem Sheet) that the general oximetry equation derived earlier is equally valid for pulse oximetry if R is now given by: log R 10 log 10 I dc! ac I dc! ac # I dc ' 1 # I dc ' 2 Figures 35(b) and 35(c) show typical cardiac synchronous pulsatile signals detected when red or NIR light is shone through a finger. Note that the d.c. content (or baseline) has been removed from these traces. It is clear from these graphs that there is quite a wide variation in the shape of the signals between people note how the trace in Figure 35(c) has a secondary peak for each heart beat. This phenomenon is quite common: the second peak is known as the dichrotic notch and can cause signal processing problems. (2)
5 Calibration of pulse oximeters The first pulse oximeters, which were manufactured in the early 1980 s, used equation (1) to compute the values of arterial SaO 2. However, the Beer Lambert law, on which this equation is based, does not take into account the multiple scattering of light by the red blood cells. Although oximetry is a differential technique, the effect of scattering is only partially compensated for since scattering is wavelength dependent. Equation (1) is therefore an over simplification: Figure 36 shows two relationships, one using the Beer Lambert law and the other based on empirical data, between the ratio R and the oxygen saturation of the patient. Consequently, instruments based on the Beer Lambert law tended to give erroneous estimates of the true value of oxygen saturation (especially for SaO 2 values below 85%). There have been few attempts since then to modify the theory in order to take scattering into account, and most pulse oximeters now use look up tables derived from from calibration studies on large numbers of healthy volunteers whose oxygen saturation is also measured invasively. Figure 36: Two relationships between the ratio R and the Oxygen Saturation of the patient. Design of pulse oximetry instrumentation A block diagram of the circuit for a pulse oximeter is shown in Figure 37. The main sections of this block diagram are now described. (
6 AGC L.P.F. 0.5 Hz AGC L.P.F. 0.5 Hz RED NIR LED Driver Red LED Driver Sample and Hold 2V B.P.F Hz Amplifier RAM Shared Photodiode Detector Current to Voltage Converter A/D Converter Microprocessor Sample and Hold 2V NEAR INFRA RED B.P.F Hz Amplifier Timing Circuits Figure 37: Block diagram of a pulse oximeter In order to build finger (or earlobe) probes which are small and unobtrusive, we need miniature light sources and detectors. Light emitting diodes (LEDs) which work in the red and Near Infra Red (NIR) part of the spectrum are readily available. However, the average power which can be obtained from standard LEDs is limited and a very sensitive detector (such as a photomultiplier tube) would be required to detect the small amount of light transmitted through the finger. This problem can be overcome by using special purpose LEDs which have been developed: red LEDs are now being manufactured with internal lensing systems to give high intensity outputs. Similarly, high current NIR LEDs are designed to be pulsed so that the peak power available from them can be increased without increasing the average power. This makes it possible to detect the light transmitted through the finger with a simple, compact, solid state photodetector such as a photodiode. If we pulse both light sources, we can then use a single photodetector in the finger probe, since silicon devices are responsive to light having visible and NIR wavelengths. We could, for example, use timing circuits to supply, say, 50 µs pulses to the red and NIR LED drivers at a repetition rate of 1 khz, as shown in Figure 38 (a frequency of 1 khz is suitable because such a frequency is well above the maximum frequency present 3 in the arterial pulse). 1 ms 50 µ s Red 500 µ s Figure 38: Timing signals for the LED drivers NIR 3 The maximum cardiac frequency is never going to ne more than a few Hz. )
7 In this mode of operation, high intensity light outputs can be obtained with the NIR LED with currents of up to 1A over a low duty cycle. The transmitted light detected by the photodiode is amplified and converted to a voltage using an op amp configured as a current to voltage converter. At this point in the circuit the signal is fed to two identical sections, one for each of the transmitted wavelengths. Since the light is pulsed, we need to use a sample and hold circuit to reconstitute the waveforms at each of the two wavelengths. The same timing circuits which were used to control the red and NIR LED drivers are also used to provide the the control pulses for the corresponding sample and hold circuits. The outputs from these circuits are then filtered with a band pass filter (with 0.5 Hz and 5 Hz cut off frequencies) in order to remove primarily the d.c. component but also high frequency noise. The resulting signals thus represent the cardiac synchronous information in the waveforms and these are further amplified before they are converted to digital format for subsequent analysis by the microprocessor. It can be seen from the block diagram in Figure 37 that the output from each sample and hold is also passed to a low pass filter. This is the first stage of an automatic gain control (AGC) circuit which adjusts the light intensity from the corresponding LED so that the d.c. level always remains at the same value (say 2V) whatever the thickness or skin characteristics of the patient s finger. There are two equally important reasons for deciding to use an AGC circuit: firstly, it means that the amplitude of the a.c. signal (which may vary between 0.1% and 2% of the total signal) is also within a pre defined range and this makes the amplifier which follows the band pass filter easier to design. Secondly, the d.c. component of he transmitted red and NIR signals can be set at the same value (2 V) in each case. Hence it can be eliminated from the formula used by the microprocessor to calculate the oxygen saturation. A new index, # R log I ac ' log I 1 ac ' is defined (compare with Equation (2)). In practice, it is not 2 even necessary to convert the a.c. signal amplitudes at the two wavelengths to their logarithmic equivalents: instead a look up table can be loaded into memory and this will contain the values of oxygen saturation corresponding to each value of the (Red pulse amplitude)/(nir pulse amplitude) ratio. We will now consider in turn each of the main circuits shown in the block diagram. Constant current source for driving LEDs A simple potential circuit for achieving this is shown in Figure 39(a) in which an op amp is combined with a bipolar transistor. In this circuit, the negative feedback forces v e = v in. Thus, I e = V in /R 1. Since the collector current is almost equal to the emitter current (I c is equal to I e + I b ), the LED current is therefore also given by I LED = V in /R 1. However, this current source is slightly imperfect because the small base current, I b, may vary with V ce. This arises because the op amp stabilises the emitter current whereas the load sees the collector current. *
8 +12v +12v R 1 v in n p n G S v e v in D R 1 (a) Figure 39: Two possible circuits for constant current LED driving By using a FET instead of a bipolar transistor, this problem can be avoided as shown in Figure 39(b). Since the FET draws no gate current, the output is sampled at the source resistance without error, eliminating the base current error of the bipolar transistor. Any departures from ideal behaviour are due to non linearities in the current sampling resistor and to errors in the op amp input circuit, such as offsets and drifts. Note, however, that the load current is limited by the I DS(on) of the MOSFET. If a bipolar power supply is available, the circuits of Figure 39 can be further simplified by omitting Vin and tying the non inverting input of the op amp to ground as shown in Figures 40(a) and 40(b) in both of which I LED = 12 V/R 1. (b) +
9 +12v +12v R 1 c b e v e G S D R 1 12v (a) (b) 12v Figure 40: Alternative circuits for constant current LED driving when a bipolar power supply is available Timing circuit In this application, the accuracy of the timing is not of paramount importance, hence the timing circuit can be built around the 555 timer integrated circuit. From the data sheet for this i.c., it can easily be worked out that the circuit given in Figure 41 can be configured, for example by setting C = 22 nf, R a = 56kΩ and R b = 3.3kΩ, to give a 50 µs pulse approximately every millisecond, as intended. R a R b C v cc 8 4 Reset Discharg Trigger Threshold n 3 O/P T 1 T 2 T R a$ R b C T R b C Figure 41: Generating the timing pulses for pulse oximetry Pulsing the light output from the LEDs The output from the LED can be pulsed by connecting an n channel enhancement mode MOSFET across it as shown in Figure 42. The pulses from the output pin of the 555 timer (pin 3) are taken to the gate of the transistor. The FET needs to be an enhancement mode MOSFET for it to be turned fully off and on by the gate pulses. The MOSFET chosen for this task should also be capable of handling the maximum current flowing through the LED.
10 +12v R 1 G S D D off on off on S n channel enhancement mode MOSFET 12v Figure 42: Pulsing the LED Receiver circuit light pulses from LED: The simplest solid state optical detector is the photodiode. Photodiode detectors normally operate with reverse bias applied to the p n junction (photoconductive mode). When light falls on the junction region of the photodiode, an electron hole pair is created; under the influence of the junction (or built in) field, the hole is swept towards the p material and the electron towards the n material. The resulting light current is seen as a large increase in the reverse current. 47 pf +V R L v o = I R L V V Figure 43: Photodiode current to voltage converter circuit For the purposes of signal amplification, the photocurrent must be transformed into a voltage with moderate output impedance; this is achieved with the circuit shown in Figure 43, the op amp being configured as a current to voltage converter. Because of the high junction resistance of the reverse biased photodiode, the op amp should be an FET type with a very high input impedance. Since the negative input of the op amp acts like a virtual earth, the output voltage of the circuit is v o = I R L. A very large feedback resistance may be used, values as high as several tens of MΩ being typical in practice.
11 Sample and hold circuit In the sample mode, the output of an ideal sample and hold circuit is equal to the input signal at that particular instant. When switched to the hold mode, the output should remain constant at that value of the input signal which existed at the instant of switching. A simple sample and hold circuit is shown in Figure 44. This circuit uses a FET switch which passes the signal through during the sample period and disconnects it during the hold period. Whatever signal was present at the time the FET is turned off is then held on the capacitor C. The choice of a value for C is a compromise between two conflicting requirements: " Leakage currents in the FET and the op amp cause the capacitor voltage to droop during the hold period according to the equation dv dt I l C where I l is the leakage current. Thus C should be as large as possible in order to minimise droop. " The resistance of the FET when turned on (typically tens of ohms) forms a low pass filter in combination with C and so C should be small if high speed signals are to be followed accurately. Ready built sample and hold circuits are also available as monolithic integrated circuits which simply require the connection of an external hold capacitor. +12v v in S D v o G C 12v +12 V sample 12 V hold Figure 44: Sample and hold circuit Automatic gain control circuit The output from the sample and hold circuit, as indicated in the general description of the block diagram, is fed to a band pass filter which extracts the pulsatile signal prior to its further amplification and analysis. The same output is also taken to a low pass filter with a cut off frequency of, say, 0.1 Hz, which extracts the d.c. value of the transmitted signal. There are then several ways of implementing the AGC function. One of the simplest ways is to feed the d.c. signal to one input of a differential amplifier whose other input is a constant, reference voltage (from a zener diode, for example). The difference between these two voltages is then used to generate the voltage v in in Figure 39 which sets the value of the LED current.
LM 358 Op Amp. If you have small signals and need a more useful reading we could amplify it using the op amp, this is commonly used in sensors.
LM 358 Op Amp S k i l l L e v e l : I n t e r m e d i a t e OVERVIEW The LM 358 is a duel single supply operational amplifier. As it is a single supply it eliminates the need for a duel power supply, thus
DIGITAL-TO-ANALOGUE AND ANALOGUE-TO-DIGITAL CONVERSION
DIGITAL-TO-ANALOGUE AND ANALOGUE-TO-DIGITAL CONVERSION Introduction The outputs from sensors and communications receivers are analogue signals that have continuously varying amplitudes. In many systems
Operational Amplifier - IC 741
Operational Amplifier - IC 741 Tabish December 2005 Aim: To study the working of an 741 operational amplifier by conducting the following experiments: (a) Input bias current measurement (b) Input offset
Basic Electronics Prof. Dr. Chitralekha Mahanta Department of Electronics and Communication Engineering Indian Institute of Technology, Guwahati
Basic Electronics Prof. Dr. Chitralekha Mahanta Department of Electronics and Communication Engineering Indian Institute of Technology, Guwahati Module: 2 Bipolar Junction Transistors Lecture-2 Transistor
Transistor Characteristics and Single Transistor Amplifier Sept. 8, 1997
Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 8, 1997 1 Purpose To measure and understand the common emitter transistor characteristic curves. To use the base current gain
Transistor Amplifiers
Physics 3330 Experiment #7 Fall 1999 Transistor Amplifiers Purpose The aim of this experiment is to develop a bipolar transistor amplifier with a voltage gain of minus 25. The amplifier must accept input
Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 13, 2006
Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 13, 2006 1 Purpose To measure and understand the common emitter transistor characteristic curves. To use the base current gain
Content Map For Career & Technology
Content Strand: Applied Academics CT-ET1-1 analysis of electronic A. Fractions and decimals B. Powers of 10 and engineering notation C. Formula based problem solutions D. Powers and roots E. Linear equations
AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE
AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE Thank you for purchasing your Amplified High Speed Fiber Photodetector. This user s guide will help answer any questions you may have regarding the
Bipolar Transistor Amplifiers
Physics 3330 Experiment #7 Fall 2005 Bipolar Transistor Amplifiers Purpose The aim of this experiment is to construct a bipolar transistor amplifier with a voltage gain of minus 25. The amplifier must
Common-Emitter Amplifier
Common-Emitter Amplifier A. Before We Start As the title of this lab says, this lab is about designing a Common-Emitter Amplifier, and this in this stage of the lab course is premature, in my opinion,
Programmable Single-/Dual-/Triple- Tone Gong SAE 800
Programmable Single-/Dual-/Triple- Tone Gong Preliminary Data SAE 800 Bipolar IC Features Supply voltage range 2.8 V to 18 V Few external components (no electrolytic capacitor) 1 tone, 2 tones, 3 tones
Application Note AN1
TAKING INVENTIVE STEPS IN INFRARED. MINIATURE INFRARED GAS SENSORS GOLD SERIES UK Patent App. No. 799A USA Patent App. No. 9/78,7 World Patents Pending SENSOR OVERVIEW Application Note AN The Dynament
LABORATORY 2 THE DIFFERENTIAL AMPLIFIER
LABORATORY 2 THE DIFFERENTIAL AMPLIFIER OBJECTIVES 1. To understand how to amplify weak (small) signals in the presence of noise. 1. To understand how a differential amplifier rejects noise and common
APPLICATION NOTES: Dimming InGaN LED
APPLICATION NOTES: Dimming InGaN LED Introduction: Indium gallium nitride (InGaN, In x Ga 1-x N) is a semiconductor material made of a mixture of gallium nitride (GaN) and indium nitride (InN). Indium
Inrush Current. Although the concepts stated are universal, this application note was written specifically for Interpoint products.
INTERPOINT Although the concepts stated are universal, this application note was written specifically for Interpoint products. In today s applications, high surge currents coming from the dc bus are a
Transistor Biasing. The basic function of transistor is to do amplification. Principles of Electronics
192 9 Principles of Electronics Transistor Biasing 91 Faithful Amplification 92 Transistor Biasing 93 Inherent Variations of Transistor Parameters 94 Stabilisation 95 Essentials of a Transistor Biasing
Chapter 19 Operational Amplifiers
Chapter 19 Operational Amplifiers The operational amplifier, or op-amp, is a basic building block of modern electronics. Op-amps date back to the early days of vacuum tubes, but they only became common
A Practical Guide to Free Energy Devices
A Practical Guide to Free Energy Devices Device Patent No 29: Last updated: 7th October 2008 Author: Patrick J. Kelly This is a slightly reworded copy of this patent application which shows a method of
Operational Amplifiers
Module 6 Amplifiers Operational Amplifiers The Ideal Amplifier What you ll learn in Module 6. Section 6.0. Introduction to Operational Amplifiers. Understand Concept of the Ideal Amplifier and the Need
MAS.836 HOW TO BIAS AN OP-AMP
MAS.836 HOW TO BIAS AN OP-AMP Op-Amp Circuits: Bias, in an electronic circuit, describes the steady state operating characteristics with no signal being applied. In an op-amp circuit, the operating characteristic
Power Supplies. 1.0 Power Supply Basics. www.learnabout-electronics.org. Module
Module 1 www.learnabout-electronics.org Power Supplies 1.0 Power Supply Basics What you ll learn in Module 1 Section 1.0 Power Supply Basics. Basic functions of a power supply. Safety aspects of working
Series AMLDL-Z Up to 1000mA LED Driver
FEATURES: Click on Series name for product info on aimtec.com Series Up to ma LED Driver Models Single output Model Input Voltage (V) Step Down DC/DC LED driver Operating Temperature range 4ºC to 85ºC
SIMPLE HEART RATE MONITOR FOR ANALOG ENTHUSIASTS
SIMPLE HEART RATE MONITOR FOR ANALOG ENTHUSIASTS Jelimo B Maswan, Abigail C Rice 6.101: Final Project Report Date: 5/15/2014 1 Project Motivation Heart Rate Monitors are quickly becoming ubiquitous in
Reading: HH Sections 4.11 4.13, 4.19 4.20 (pgs. 189-212, 222 224)
6 OP AMPS II 6 Op Amps II In the previous lab, you explored several applications of op amps. In this exercise, you will look at some of their limitations. You will also examine the op amp integrator and
Op Amp Circuit Collection
Op Amp Circuit Collection Note: National Semiconductor recommends replacing 2N2920 and 2N3728 matched pairs with LM394 in all application circuits. Section 1 Basic Circuits Inverting Amplifier Difference
Building the AMP Amplifier
Building the AMP Amplifier Introduction For about 80 years it has been possible to amplify voltage differences and to increase the associated power, first with vacuum tubes using electrons from a hot filament;
BJT Characteristics and Amplifiers
BJT Characteristics and Amplifiers Matthew Beckler [email protected] EE2002 Lab Section 003 April 2, 2006 Abstract As a basic component in amplifier design, the properties of the Bipolar Junction Transistor
Lab 5 Operational Amplifiers
Lab 5 Operational Amplifiers By: Gary A. Ybarra Christopher E. Cramer Duke University Department of Electrical and Computer Engineering Durham, NC. Purpose The purpose of this lab is to examine the properties
Diode Circuits. Operating in the Reverse Breakdown region. (Zener Diode)
Diode Circuits Operating in the Reverse Breakdown region. (Zener Diode) In may applications, operation in the reverse breakdown region is highly desirable. The reverse breakdown voltage is relatively insensitive
LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS
LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS Objective In this experiment you will study the i-v characteristics of an MOS transistor. You will use the MOSFET as a variable resistor and as a switch. BACKGROUND
The full wave rectifier consists of two diodes and a resister as shown in Figure
The Full-Wave Rectifier The full wave rectifier consists of two diodes and a resister as shown in Figure The transformer has a centre-tapped secondary winding. This secondary winding has a lead attached
Lecture - 4 Diode Rectifier Circuits
Basic Electronics (Module 1 Semiconductor Diodes) Dr. Chitralekha Mahanta Department of Electronics and Communication Engineering Indian Institute of Technology, Guwahati Lecture - 4 Diode Rectifier Circuits
Line Reactors and AC Drives
Line Reactors and AC Drives Rockwell Automation Mequon Wisconsin Quite often, line and load reactors are installed on AC drives without a solid understanding of why or what the positive and negative consequences
Signal Extraction Technology
Signal Extraction Technology Technical bulletin Introduction Masimo SET pulse oximetry is a new and fundamentally distinct method of acquiring, processing and reporting arterial oxygen saturation and pulse
11: AUDIO AMPLIFIER I. INTRODUCTION
11: AUDIO AMPLIFIER I. INTRODUCTION The properties of an amplifying circuit using an op-amp depend primarily on the characteristics of the feedback network rather than on those of the op-amp itself. A
ANADOLU UNIVERSITY DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING
ANADOLU UNIVERSITY DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING EEM 102 INTRODUCTION TO ELECTRICAL ENGINEERING EXPERIMENT 9: DIODES AND DC POWER SUPPLY OBJECTIVE: To observe how a diode functions
Programmable-Gain Transimpedance Amplifiers Maximize Dynamic Range in Spectroscopy Systems
Programmable-Gain Transimpedance Amplifiers Maximize Dynamic Range in Spectroscopy Systems PHOTODIODE VOLTAGE SHORT-CIRCUIT PHOTODIODE SHORT- CIRCUIT VOLTAGE 0mV DARK ark By Luis Orozco Introduction Precision
Amplified High Speed Fiber Photodetectors
Amplified High Speed Fiber Photodetectors User Guide (800)697-6782 [email protected] www.eotech.com Page 1 of 7 EOT AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE Thank you for purchasing your Amplified
Alphasense Application Note AAN 105-03 DESIGNING A POTENTIOSTATIC CIRCUIT
AAN 10503 DESIGNING A POTENTIOSTATIC CIRCUIT Introduction In a threeelectrode sensor, each electrode has a specific use: The working electrode responds to the target gas, either oxidising or reducing the
Study Guide for the Electronics Technician Pre-Employment Examination
Bay Area Rapid Transit District Study Guide for the Electronics Technician Pre-Employment Examination INTRODUCTION The Bay Area Rapid Transit (BART) District makes extensive use of electronics technology
Frequency Response of Filters
School of Engineering Department of Electrical and Computer Engineering 332:224 Principles of Electrical Engineering II Laboratory Experiment 2 Frequency Response of Filters 1 Introduction Objectives To
css Custom Silicon Solutions, Inc.
css Custom Silicon Solutions, Inc. CSS555(C) CSS555/ PART DESCRIPTION The CSS555 is a micro-power version of the popular 555 Timer IC. It is pin-for-pin compatible with the standard 555 timer and features
ARRL Morse Code Oscillator, How It Works By: Mark Spencer, WA8SME
The national association for AMATEUR RADIO ARRL Morse Code Oscillator, How It Works By: Mark Spencer, WA8SME This supplement is intended for use with the ARRL Morse Code Oscillator kit, sold separately.
6.101 Final Project Report Class G Audio Amplifier
6.101 Final Project Report Class G Audio Amplifier Mark Spatz 4/3/2014 1 1 Introduction For my final project, I designed and built a 150 Watt audio amplifier to replace the underpowered and unreliable
SINGLE-SUPPLY OPERATION OF OPERATIONAL AMPLIFIERS
SINGLE-SUPPLY OPERATION OF OPERATIONAL AMPLIFIERS One of the most common applications questions on operational amplifiers concerns operation from a single supply voltage. Can the model OPAxyz be operated
Physics 120 Lab 6: Field Effect Transistors - Ohmic region
Physics 120 Lab 6: Field Effect Transistors - Ohmic region The FET can be used in two extreme ways. One is as a voltage controlled resistance, in the so called "Ohmic" region, for which V DS < V GS - V
WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.
WHY DIFFERENTIAL? Voltage, The Difference Whether aware of it or not, a person using an oscilloscope to make any voltage measurement is actually making a differential voltage measurement. By definition,
Photodiode/Phototransistor Application Circuit < E V3 I P I V OUT E V2. Figure 1. Fundamental Circuit of Photodiode (Without Bias)
Application Note Photodiode/Phototransistor Application Circuit FUNDAMENTAL PHOTODIODE CIRCUITS Figures 1 and 2 show the fundamental photodiode circuits. The circuit shown in Figure 1 transforms a photocurrent
Chapter 12: The Operational Amplifier
Chapter 12: The Operational Amplifier 12.1: Introduction to Operational Amplifier (Op-Amp) Operational amplifiers (op-amps) are very high gain dc coupled amplifiers with differential inputs; they are used
Low Noise, Matched Dual PNP Transistor MAT03
a FEATURES Dual Matched PNP Transistor Low Offset Voltage: 100 V Max Low Noise: 1 nv/ Hz @ 1 khz Max High Gain: 100 Min High Gain Bandwidth: 190 MHz Typ Tight Gain Matching: 3% Max Excellent Logarithmic
Operating Manual Ver.1.1
Class B Amplifier (Push-Pull Emitter Follower) Operating Manual Ver.1.1 An ISO 9001 : 2000 company 94-101, Electronic Complex Pardesipura, Indore- 452010, India Tel : 91-731- 2570301/02, 4211100 Fax: 91-731-
Diode Applications. by Kenneth A. Kuhn Sept. 1, 2008. This note illustrates some common applications of diodes.
by Kenneth A. Kuhn Sept. 1, 2008 This note illustrates some common applications of diodes. Power supply applications A common application for diodes is converting AC to DC. Although half-wave rectification
Diodes and Transistors
Diodes What do we use diodes for? Diodes and Transistors protect circuits by limiting the voltage (clipping and clamping) turn AC into DC (voltage rectifier) voltage multipliers (e.g. double input voltage)
Module 11: Conducted Emissions
Module 11: Conducted Emissions 11.1 Overview The term conducted emissions refers to the mechanism that enables electromagnetic energy to be created in an electronic device and coupled to its AC power cord.
Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19
Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and
OPERATIONAL AMPLIFIER
MODULE3 OPERATIONAL AMPLIFIER Contents 1. INTRODUCTION... 3 2. Operational Amplifier Block Diagram... 3 3. Operational Amplifier Characteristics... 3 4. Operational Amplifier Package... 4 4.1 Op Amp Pins
Electronics. Discrete assembly of an operational amplifier as a transistor circuit. LD Physics Leaflets P4.2.1.1
Electronics Operational Amplifier Internal design of an operational amplifier LD Physics Leaflets Discrete assembly of an operational amplifier as a transistor circuit P4.2.1.1 Objects of the experiment
How To Calculate The Power Gain Of An Opamp
A. M. Niknejad University of California, Berkeley EE 100 / 42 Lecture 8 p. 1/23 EE 42/100 Lecture 8: Op-Amps ELECTRONICS Rev C 2/8/2012 (9:54 AM) Prof. Ali M. Niknejad University of California, Berkeley
NTE923 & NTE923D Integrated Circuit Precision Voltage Regulator
NTE923 & NTE923D Integrated Circuit Precision Voltage Regulator Description: The NTE923 and NTE923D are voltage regulators designed primarily for series regulator applications. By themselves, these devices
GenTech Practice Questions
GenTech Practice Questions Basic Electronics Test: This test will assess your knowledge of and ability to apply the principles of Basic Electronics. This test is comprised of 90 questions in the following
APPLICATION NOTE TESTING PV MICRO INVERTERS USING A FOUR QUADRANT CAPABLE PROGRAMMABLE AC POWER SOURCE FOR GRID SIMULATION. Abstract.
TESTING PV MICRO INVERTERS USING A FOUR QUADRANT CAPABLE PROGRAMMABLE AC POWER SOURCE FOR GRID SIMULATION Abstract This application note describes the four quadrant mode of operation of a linear AC Power
Lecture 18: Common Emitter Amplifier. Maximum Efficiency of Class A Amplifiers. Transformer Coupled Loads.
Whites, EE 3 Lecture 18 Page 1 of 10 Lecture 18: Common Emitter Amplifier. Maximum Efficiency of Class A Amplifiers. Transformer Coupled Loads. We discussed using transistors as switches in the last lecture.
Zero voltage drop synthetic rectifier
Zero voltage drop synthetic rectifier Vratislav Michal Brno University of Technology, Dpt of Theoretical and Experimental Electrical Engineering Kolejní 4/2904, 612 00 Brno Czech Republic [email protected],
MRF175GU MRF175GV The RF MOSFET Line 200/150W, 500MHz, 28V
Designed for broadband commercial and military applications using push pull circuits at frequencies to 500 MHz. The high power, high gain and broadband performance of these devices makes possible solid
Hands On ECG. Sean Hubber and Crystal Lu
Hands On ECG Sean Hubber and Crystal Lu The device. The black box contains the circuit and microcontroller, the mini tv is set on top, the bars on the sides are for holding it and reading hand voltage,
The basic cascode amplifier consists of an input common-emitter (CE) configuration driving an output common-base (CB), as shown above.
Cascode Amplifiers by Dennis L. Feucht Two-transistor combinations, such as the Darlington configuration, provide advantages over single-transistor amplifier stages. Another two-transistor combination
EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT
EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT 4 Understand single-phase alternating current (ac) theory Single phase AC
0.00 18 TO18 0.00 _ nothing special M 0.20 49.00 A UVA 24.00 18ISO90 TO18 6.00 Lens concentr. Lens D 0.50 49.00 B UVB 24.
Basic Information That guide assists you selecting the right ilicon Carbide (ic) based UV photodiode for your application. Basically this selection is between active area, spectral behaviour, packaging
3-Phase Synchronous PWM Controller IC Provides an Integrated Solution for Intel VRM 9.0 Design Guidelines
3-Phase Synchronous PWM Controller IC Provides an Integrated Solution for Intel VRM 9.0 Design Guidelines Odile Ronat International Rectifier The fundamental reason for the rapid change and growth in information
0.9V Boost Driver PR4403 for White LEDs in Solar Lamps
0.9 Boost Driver for White LEDs in Solar Lamps The is a single cell step-up converter for white LEDs operating from a single rechargeable cell of 1.2 supply voltage down to less than 0.9. An adjustable
High Voltage Power Supplies for Analytical Instrumentation
ABSTRACT High Voltage Power Supplies for Analytical Instrumentation by Cliff Scapellati Power supply requirements for Analytical Instrumentation are as varied as the applications themselves. Power supply
13. Diode Rectifiers, Filters, and Power Supplies
1 13. Diode Rectifiers, Filters, and Power Supplies Introduction A power supply takes Alternating Current or A.C. power from your electric utility (Con Edison) and converts the A.C. electrical current
A Real Time Analysis of PPG Signal for Measurement of SpO 2 and Pulse Rate
A Real Time Analysis of PPG Signal for Measurement of SpO 2 and Pulse Rate Sangeeta Bagha Silicon Institute of Technology Silicon Hills, Patia Bhubaneswar Laxmi Shaw Silicon Institute of Technology Silicon
Unit/Standard Number. High School Graduation Years 2010, 2011 and 2012
1 Secondary Task List 100 SAFETY 101 Demonstrate an understanding of State and School safety regulations. 102 Practice safety techniques for electronics work. 103 Demonstrate an understanding of proper
The front end of the receiver performs the frequency translation, channel selection and amplification of the signal.
Many receivers must be capable of handling a very wide range of signal powers at the input while still producing the correct output. This must be done in the presence of noise and interference which occasionally
GLOLAB Universal Telephone Hold
GLOLAB Universal Telephone Hold 1 UNIVERSAL HOLD CIRCUIT If you have touch tone telephone service, you can now put a call on hold from any phone in the house, even from cordless phones and phones without
FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER
2014 Amplifier - 1 FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER The objectives of this experiment are: To understand the concept of HI-FI audio equipment To generate a frequency response curve for an audio
Rectifier circuits & DC power supplies
Rectifier circuits & DC power supplies Goal: Generate the DC voltages needed for most electronics starting with the AC power that comes through the power line? 120 V RMS f = 60 Hz T = 1667 ms) = )sin How
THE FIBRE-OPTICS TRAINER MANUAL
THE FIBRE-OPTICS TRAINER MANUAL THE FIBRE-OPTICS TRAINER MANUAL THE FIBRE-OPTICS TRAINER CONTAINS: TRANSMITTER UNIT RECEIVER UNIT 5m. LENGTH OF TERMINATED OPTICAL CABLE INSTRUCTION MANUAL (THIS BOOK) CARRYING
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
5B5BBasic RC Oscillator Circuit
5B5BBasic RC Oscillator Circuit The RC Oscillator which is also called a Phase Shift Oscillator, produces a sine wave output signal using regenerative feedback from the resistor-capacitor combination.
IGBT Protection in AC or BLDC Motor Drives by Toshio Takahashi
International Rectifier 233 Kansas Street, El Segundo, CA 90245 USA IGBT Protection in AC or BLDC Motor Drives by Toshio Takahashi The new IR2137 IGBT Gate Driver IC integrates Ground Fault and Over-Current
Evaluating AC Current Sensor Options for Power Delivery Systems
Evaluating AC Current Sensor Options for Power Delivery Systems State-of-the-art isolated ac current sensors based on CMOS technology can increase efficiency, performance and reliability compared to legacy
Impedance 50 (75 connectors via adapters)
VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:
ENGR-4300 Electronic Instrumentation Quiz 4 Spring 2011 Name Section
ENGR-4300 Electronic Instrumentation Quiz 4 Spring 2011 Name Section Question I (20 points) Question II (20 points) Question III (20 points) Question IV (20 points) Question V (20 points) Total (100 points)
Zero Voltage Switch with Adjustable Ramp. R 2 (R sync ) 220 k (250 V~) Synchronization. Full wave logic Pulse amplifier. 58 k N
Zero Voltage Switch with Adjustable Ramp U217B/ U217B-FP Description The integrated circuit, U217B, is designed as a zerovoltage switch in bipolar technology. It is used to control resistive loads at mains
NTE2053 Integrated Circuit 8 Bit MPU Compatible A/D Converter
NTE2053 Integrated Circuit 8 Bit MPU Compatible A/D Converter Description: The NTE2053 is a CMOS 8 bit successive approximation Analog to Digital converter in a 20 Lead DIP type package which uses a differential
Tire pressure monitoring
Application Note AN601 Tire pressure monitoring 1 Purpose This document is intended to give hints on how to use the Intersema pressure sensors in a low cost tire pressure monitoring system (TPMS). 2 Introduction
PCM Encoding and Decoding:
PCM Encoding and Decoding: Aim: Introduction to PCM encoding and decoding. Introduction: PCM Encoding: The input to the PCM ENCODER module is an analog message. This must be constrained to a defined bandwidth
CONSTRUCTING A VARIABLE POWER SUPPLY UNIT
CONSTRUCTING A VARIABLE POWER SUPPLY UNIT Building a power supply is a good way to put into practice many of the ideas we have been studying about electrical power so far. Most often, power supplies are
Experiment 5. Lasers and laser mode structure
Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,
Operational Amplifier as mono stable multi vibrator
Page 1 of 5 Operational Amplifier as mono stable multi vibrator Aim :- To construct a monostable multivibrator using operational amplifier 741 and to determine the duration of the output pulse generated
AP-1 Application Note on Remote Control of UltraVolt HVPS
Basics Of UltraVolt HVPS Output Voltage Control Application Note on Remote Control of UltraVolt HVPS By varying the voltage at the Remote Adjust Input terminal (pin 6) between 0 and +5V, the UV highvoltage
INTRODUCTION FIGURE 1 1. Cosmic Rays. Gamma Rays. X-Rays. Ultraviolet Violet Blue Green Yellow Orange Red Infrared. Ultraviolet.
INTRODUCTION Fibre optics behave quite different to metal cables. The concept of information transmission is the same though. We need to take a "carrier" signal, identify a signal parameter we can modulate,
Pressure Transducer to ADC Application
Application Report SLOA05 October 2000 Pressure Transducer to ADC Application John Bishop ABSTRACT Advanced Analog Products/OpAmp Applications A range of bridgetype transducers can measure numerous process
OPERATIONAL AMPLIFIERS. o/p
OPERATIONAL AMPLIFIERS 1. If the input to the circuit of figure is a sine wave the output will be i/p o/p a. A half wave rectified sine wave b. A fullwave rectified sine wave c. A triangular wave d. A
Chapter 4. LLC Resonant Converter
Chapter 4 LLC Resonant Converter 4.1 Introduction In previous chapters, the trends and technical challenges for front end DC/DC converter were discussed. High power density, high efficiency and high power
ICL8038. Features. Precision Waveform Generator/Voltage Controlled Oscillator. Ordering Information. Pinout. Functional Diagram
Semiconductor IL0 September 99 File Number 4. Precision Waveform Generator/Voltage ontrolled Oscillator The IL0 waveform generator is a monolithic integrated circuit capable of producing high accuracy
Fig. 1 :Block diagram symbol of the operational amplifier. Characteristics ideal op-amp real op-amp
Experiment: General Description An operational amplifier (op-amp) is defined to be a high gain differential amplifier. When using the op-amp with other mainly passive elements, op-amp circuits with various