1 PosiTector 6000 Coating Thickness Gages Quick Guide v. 7.4
2 Table of Contents Introduction... 1 Quick Start... 1 Certification... 2 Probes... 4 Calibration, Verification and Adjustment... 7 Menu Operation... 9 Cal Settings Menu Zero Pt Adjust Pt Adjust N Lock Cal Lock Cal Memory Setup Menu Reset Gage Info Hi Res Mode Fast Mode Scan Mode Statistics Menu HiLo Alarm Memory Management Standard Models Menu Advanced Models Menu New PA New 90/ Annotate Display Summaries Screen Capture... 23
3 Accessing Stored Measurement Data PosiSoft Desktop Manager (PDM) PosiSoft.net PosiSoft PosiSoft Mobile Connect Menu Sync Now Auto Sync USB Drive Bluetooth Bluetooth Wireless Printer WiFi Updates Temperature Power Supply / Battery Indicator Rechargeable Batteries Additional Accessories Coating Thickness Standards Troubleshooting - General Probe attempts measurements in the air Thickness readings are inconsistent Gage powers up but will not take readings Gage will not print to Bluetooth wireless printer Technical Data Returning for Service Warranty... 40
4 Introduction The PosiTector 6000 hand-held, electronic Gage non-destructively measures the thickness of coatings on all metals, quickly and accurately. It consists of a Gage body (Standard or Advanced) and probe (Built-In or Separate). Principles of Operation F probes use the magnetic principle to measure the thickness of non-magnetic coatings on ferrous metals. N probes use the eddy current principle to measure the thickness of non-conductive coatings on non-ferrous metals. FN probes combine the full abilities of both the "F" and "N" probes. PosiTector 6000 Kit Contents - PosiTector 6000 Probe - Protective rubber holster with belt clip - Protective Cap for Probe (some probes only) - Protective Lens Shield - Wrist strap - 3 Alkaline AAA batteries - Quick Guide Instructions - Nylon carrying case with shoulder strap - USB Cable - Precision Plastic Shims - Certificate of Calibration traceable to NIST - Two (2) year warranty on Body and Probe Quick Start The PosiTector 6000 powers-up when the center navigation button is pressed. To preserve battery life, it powers-down after approximately 5 minutes of no activity. All settings are retained. 1. Remove the protective plastic cap (if supplied) from probe. 2. Power-up Gage by pressing the center navigation button. 3. Place the probe FLAT on the surface to be measured. HOLD STEADY. When a valid measurement is calculated, the Gage BEEPS twice and the measurement is displayed. 4. Lift probe AT LEAST 2 INCHES (5cm) from the surface between measurements - OR - leave probe on the surface in the same location for continuous measurements. 1
5 If memory is ON while continuous measurements are being taken, only the last value on the display (when the probe is lifted) is stored into memory. Scan (pg.15) stores ALL measurements into memory. Protective Cap Some PosiTector 6000 models are shipped with a protective plastic cap over the probe. Remove this cap prior to use. Replace it when the Gage is not in use to protect the probe. Wrist Strap 2 We recommend attaching and wearing the supplied wrist strap. 1 Plastic Lens Shield The LCD is covered with a thin plastic film for protection against fingerprints and other marks during shipment. This film, while usually removed before using the Gage, can be left in place to protect against paint overspray. Replacements can be purchased. Golden Rule Measure your uncoated part first! This quick zero-check determines if a Calibration Adjustment is needed for your substrate. (see pg. 10) Next, lay the included plastic shims onto a bare surface and measure them individually to ensure the Gage measures a known thickness within tolerance. Certification All probes include a Certificate of Calibration. For organizations with re-certification requirements, instruments may be returned at regular intervals for calibration. DeFelsko recommends that customers establish calibration intervals based upon their own experience and work environment. Based on our product knowledge, data and customer feedback, a one year calibration interval from either the date of calibration, date of purchase, or date of receipt is a typical starting point. Written calibration procedures are available from DeFelsko Corporation at no charge. 2
6 PosiTector Body USB Port Advanced LCD Display Standard LCD Display Separate Probe Style Probe Connector Built-in Probe Style Typical Screen Substrate Battery Indicator Advanced model Factory Calibration in Use Standard model Current Calibration Setting (Advanced Only) Current Measurement Unit of Measurement Button Functions- Normal Operation Adjust Down Access the Menu Adjust Up Restore brightness after dimming (Advanced only) 3
7 Magnetic & Eddy Current Theory of Operation The PosiTector 6000 probes calculate coating thickness measurements using either the magnetic principle (F probes - ferrous models) or the eddy current principle (N probes - non-ferrous models). Combination FN probes (ferrous/non-ferrous) use both principles. FN probes first attempt a measurement using the magnetic principle and will display a reading with the letter "F" if the coating is non-magnetic over steel. If the coating is non-conductive over metal, then the probe will automatically attempt a measurement using the eddy current principle and display a reading with the letter "N (see also N Lock pg. 12). Probes The PosiTector 6000 consists of a body and a probe. A wide selection of interchangeable probes are available. Each retain their own unique calibration information. Gage bodies (Standard or Advanced) accept ALL probes. See Gage Info (pg. 15) to find your body and probe type. More probe information is available at To disconnect a probe from a body, power-down the Gage and pull the plastic probe connector horizontally (in the direction of the arrow) away from the body. Reverse these steps to attach a new probe. When powered-up the PosiTector automatically determines what type of probe is attached and does a self-check. Coating thickness probes sense when they are near metal and immediately attempt a measurement followed by another every 2 seconds. They stop when removed from the vicinity of metal and power-down after 5 minutes of no activity. This continuous measurement feature is intended to allow careful probe placement on small or odd-shaped surfaces. Ignore all readings taken before the probe is properly placed. 4
8 Two Probe Styles: * Built-in - sensor protrudes from bottom of probe connector * Separate - sensor is attached to the probe connector by a 1.2 meter (3 foot) cable. (Extended cables lengths available.) Available PosiTector 6000 Probes: Regular - These constant-pressure, stainless steel probes are hermetically sealed to be totally waterproof - ideal for underwater use. Micro - ideal for measuring small parts and hard-to-reach areas. Available in 0, 45 or 90 degree angles. Thick - ideal for thick, protective coatings. Xtreme - ideal for measuring rough and hot surfaces up to 250 C (500 F). Probe features an Alumina wear face and braided cable. For complete details, see: Additional PosiTector Probes: The PosiTector accepts a number of probe types including magnetic and eddy current coating thickness, surface profile, environmental and ultrasonic wall thickness probes. Perform the latest software Updates (pg. 33) to ensure probe compatibility with your Gage. For the latest information on probe interchangeability see 5
9 Why is Measurement Important? Dry Film Thickness (DFT) is arguably the single most important measurement made during the application and inspection of protective coatings. Coatings are designed to perform their intended function when applied within a tight DFT range as specified by the manufacturer. Correct thickness ensures optimum product performance. Even the most basic specification will require DFT to be measured. Many physical and appearance properties of the finished coating are directly affected by the film thickness including the color, gloss, surface profile, adhesion, flexibility, impact resistance, and hardness of the coating. The fit of pieces assembled after coating can be affected when film thickness isn t within tolerance. Regular film thickness measurement helps control material costs, manage application efficiency, maintain finish quality and ensure compliance with contract specifications. Paint manufacturers recommend target ranges to achieve optimum performance characteristics and clients expect these parameters to be met. Precisely measuring finish thickness has other benefits, too. Whether to meet International Organization for Standardization (ISO), quality, or customer requirements for process control, companies need to verify coating quality to avoid wasting money reworking product. 6
10 Calibration, Verification & Adjustment PosiTector 6000 probes non-destructively measure the thickness of coatings on metals. Three steps ensure best accuracy 1. Calibration - typically done by the manufacturer. All probes include a Certificate of Calibration. 2. Verification - typically done by the user on known reference standards such as plastic shims or coated thickness standards. 3. Adjustment - to a known thickness Calibration Calibration is the high-level, controlled and documented process of measuring traceable calibration standards over the full operating range of the probe, and verifying that the results are within the stated accuracy of the probe. Calibrations are performed by the manufacturer, their authorized agent, or by an accredited calibration laboratory in a controlled environment using a documented process. PosiTector 6000 probes are shipped with a Certificate of Calibration showing traceability to a National Metrology Institution. For organizations with re-certification requirements, probes may be returned at regular intervals for calibration. DeFelsko recommends that customers establish calibration intervals based upon their own experience and work environment. Based on DeFelsko s product knowledge, data and customer feedback, a one year calibration interval from either the last date of calibration, date of purchase, or date of receipt is a typical starting point. Written Calibration Procedures are available online at no charge. Verification Verification is an accuracy check performed by the user on known reference standards. A successful verification requires the Gage to read within the combined accuracy of the probe and the reference standards. 7
11 A reference standard is a sample of known thickness(es) against which a user may verify probe accuracy. Reference standards may be plastic shims, coated thickness standards, or sample parts whose coating thickness has been determined using other means. Verify accuracy at the beginning and the end of each work shift. During the work shift, if the Gage is dropped or suspected of giving erroneous readings, its accuracy should be re-verified. In the event of physical damage, wear, high usage, or after an established calibration interval, the probe should be returned to the manufacturer for repair or calibration. Adjustment Adjustment, or Cal Adjustment, is the physical act of aligning the probe s thickness readings to match those of a known thickness sample (removal of bias) in order to improve the accuracy of the probe on a specific surface or in a specific portion of its measurement range. 1-point or 2-point Cal adjustments are possible. Probes are factory calibrated and perform an automatic self-check each time a measurement is taken. For many applications no further adjustment is necessary after a Reset (pg. 14). Just check ZERO on the uncoated substrate, then measure. However, sometimes probe readings can be influenced by changes in substrate shape, composition, and surface roughness or by measuring in a different location on the part. That is why Cal adjustments are made possible. The XX symbol disappears whenever a Cal adjustment is made. Where a Cal adjustment method has not been specified, use a 1-point method first. If measuring the included shims on your uncoated surface reveals inaccuracies, use the 2-point method. Factory Cal settings can be restored at any time by performing a Reset (pg. 14), creating a NEW Cal setting (See Cal Memory pg. 13), or by DELETING the adjustments made to the Cal 1 calibration setting (pg. 13). The XX symbol appears on the display whenever factory Cal settings are in use. With FN probes, calibration adjustments are made only to the F or N mode (stored independently under a particular Cal), whichever was measured last. 8
12 Memory Statistics Cal Settings Setup Connect Help Power Off Exit Menu Operation Gage functions are menu controlled. To access the Menu, power-up the Gage, then press the center navigation button. Below is a sample menu for a PosiTector 6000 Advanced Model: Up Center Down Current selection is displayed with darkened background Navigation Button Some buttons have a tick box to their right to indicate current status. An empty box indicates that feature is not active. Sync Now USB Drive Auto SYNC Bluetooth WiFi Powder Probes Updates Exit To navigate, use the Up and Down buttons to scroll and to SELECT. Select Exit to exit from any menu. NOTE: The center button is purposely recessed to help eliminate unintentional powering-up of the Gage. 9
13 1-point Calibration Adjustment Also known as an offset or correction value, there are 4 ways to perform this adjustment: (i) Simple Zero Calibration Adjustment Measure your uncoated part. If the Gage does not read "0" within the tolerance of the probe being used, lift the probe from the surface and adjust the display down (-) or up (+) until it reads "0". Measure and adjust until the average of a series of readings on the uncoated surface is within tolerance of "0". Zero Cal Settings Menu (ii) Average Zero Calibration Adjustment To establish 0 on a rough or curved surface a preferred method to (i) is to take several readings on the uncoated part and average the result. (a) Select Zero from the Cal Settings menu. (b) Press (+) to select the number of readings to be used to obtain an average, typically 3 to 10 readings. The greater the variation between readings, the more readings should be taken to obtain an average. (c) Repeatedly measure the uncoated part. The Gage will wait 2 seconds after placing the probe on the surface to allow the user to correctly position the probe on the surface. After the last measurement the Gage will calculate a Zero which represents the average of all the Zero readings taken. (iii) Simple Adjustment to a Known Thickness It is sometimes desirable to adjust the Gage to a known thickness, such as a shim, rather than adjusting it to zero. Measure the object. If the expected reading is not obtained (within tolerance), lift the probe from the surface and adjust the displayed reading down (-) or up (+) to the expected thickness. Hold the button down to increase the rate of adjustment. 10
14 1 Pt Adjust (iv) Average Adjustment to a Known Thickness On rough or curved surfaces a preferred method to (iii) is to take several readings on the known thickness and average the result. (a) Select 1 Pt Adjust from the Cal Settings menu. (b) Press (+) to select the number of readings to be used to obtain an average, typically 3 to 10 readings. The greater the variation between readings, the more readings should be taken to obtain an average. (c) Repeatedly measure the known thickness reference. The Gage will wait 2 seconds between readings to allow the user to correctly position the probe on the surface. After the last measurement the Gage will calculate and display the reading which represents the average of all the measurements taken. If the expected reading is not obtained (within tolerance) lift the probe from the surface and adjust the reading down (-) or up (+) to the expected thickness and press. 2 Pt Adjust 2-point Calibration Adjustment Preferred method for very unusual substrate materials, shapes or conditions. Provides greater accuracy within a limited, defined range. 1. Select 2 Pt Adjust from the Cal Settings menu. 2. Press (+) to select the number of readings to be used to obtain an average on the thinner item, typically 3 to 10 readings. The greater the variation between readings, the more readings should be taken to obtain an average. 3. Repeatedly measure the thinner item. The Gage will wait for 2 seconds on the surface to allow the user to correctly position the probe on the surface. After the last measurement the Gage will calculate and display a thickness value which represents the average of all the readings taken using the factory calibration settings. 11
15 4. Lift the probe from the surface and adjust the displayed reading down (-) or up (+) to the known thickness value of the thin item. Press to accept this value. 5. Repeat steps 2-4 for the thicker item. Cal Lock When selected, the icon will appear and all calibration settings are locked to prevent further user adjustments. Uncheck to make further adjustments. N Lock (FN ferrous/non-ferrous combination probes only) Select N Lock (Non-Ferrous Lock) when operating regularly on non-ferrous substrates. The icon will appear and the probe will only use the eddy current principle to shorten measurement time and extend battery life. N Lock is useful when measuring coatings on plated steel. Normally the probe measures both the coating + plating over the steel using the magnetic principle. N Lock makes the Gage measure the coating over the plating only. Select N Lock to measure over slightly magnetic substrates; i.e. clear-coat on gold over nickel-plated brass. Although the probe's magnet is used for the magnetic principle, it is also used in N Lock to magnetically saturate a slightly magnetic substrate and allow the eddy current principle to operate unhindered. CAUTION: With N Lock engaged it is possible to obtain a reading when measuring non-conductive coatings on steel. This is not recommended. 12
16 Cal Memory Calibration Memory (Advanced models only) It is often convenient to store a particular calibration adjustment before making another. Then, if you return to that part, the corresponding calibration setting can be restored. A setting is any calibration adjustment. The PosiTector 6000 always displays the current calibration setting (ex. Cal 3) in the upper right corner of the display. The setting called Cal 1 has unique features. It can be adjusted but never deleted, and is always made active with factory settings after a Reset (pg.14). New Creates a new calibration setting using the next available number (Maximum of 10). By default, these new Cal settings are initially created with the Gage s factory settings. This is indicated with the icon which appears at the bottom of the display. A warning message will prevent the creation of a new Cal Memory if a batch is open and has readings. Delete the batch first (pg. 21). Open Loads an existing setting. Use the Up or Down buttons to scroll until the desired setting appears, then press. A warning message will prevent the opening of a stored Cal setting if a batch is open and has readings. Create a new batch first or open a batch containing no readings (pg. 20). Delete Removes a setting completely from the list. That Cal number can be reused later with the New command. A setting cannot be deleted if readings have been stored into a batch using that calibration setting. Delete all readings in that batch first (pg. 21). Although Cal 1 cannot be deleted, the Delete function will return it to factory settings. View View stored Calibration Settings. 13
17 Reset Setup Menu Reset (soft reset) restores factory settings and returns the Gage to a known condition. The following occurs: - All batches, stored measurements, images, and batch names are erased. - The icon will appear on the display. - Menu settings are returned to the following: Memory = OFF Cal Lock = OFF High Res = OFF Bluetooth = OFF Statistics Mode = OFF Display = None Hi Lo Alarm = OFF N Lock = OFF Scan Mode = OFF Perform a more thorough Hard Reset by powering down the Gage, waiting several seconds, then simultaneously holding both the center and (+) buttons until the Reset symbol appears. This returns the Gage to a known out-of-the-box condition. It performs the same function as a menu Reset with addition of: - All calibration adjustments and Cal Memory are cleared and returned to the Gage s factory calibration settings. - Bluetooth Pairing info is cleared. - Menu settings are returned to the following: Units = microns Flip Display = Normal Auto Sync = OFF White on Black = OFF Fast Mode = OFF Language = English Battery Type = Alkaline Backlight = Normal Bluetooth Streaming = OFF USB Drive= ON NOTES: - Keep the Gage away from metal during a Reset. - Date, Time and WiFi are not affected by either Reset. 14
18 Gage Info Displays the model number & serial number, probe type & serial number, PosiSoft.net registration key, the amount of remaining memory for storage of readings, date and time, and software packages. For security purposes, the registration key is required to add the Gage to your free PosiSoft.net account. Hi Res Increases the displayed Gage resolution as follows: Resolution Range 0.01 mil mils 0.1 mil mils 0.1 um um 1.0 um over 1000 um NOTE: Gage accuracy is not affected. Fast Increases measurement speed for most probes. Useful for quick inspection or when measuring large areas with thick coatings where proper probe positioning is not critical. Swift up/down probe movement is required. Reduced accuracy may be noted. Scan (Advanced models only) In ordinary operation, the PosiTector 6000 takes thickness measurements every 2 seconds when placed onto a surface. Normally, the user raises the probe after it has taken a single measurement and relocates it to a second location. Scan mode modifies this behavior in two ways. First, measurements are taken much faster. Second, all measurements are recorded when memory is on. As the probe slides along the surface in scan mode, the PosiTector 6000 takes single measurements in quick succession. This may reduce the life of the probe. Recommended for smooth surfaces only. 15
19 Flip Display This option causes the display to rotate 180. Ideal for right-sideup viewing when the sensors are pointed away from the user. White on Black (Advanced models only) Inverts the LCD display to white on a black background to provide better readability in some surroundings. Backlight (Advanced models only) Selects display brightness (Sun, Normal or Night). All settings will dim slightly after a period of no activity to conserve battery life. Press the Down button to brighten the display. Set Clock All measurements are date and time stamped (24 hour format) when stored into memory. It is therefore important to keep both the date and time current using this menu option. Use the Up and Down buttons to scroll, and the (-) and (+) buttons to adjust a value. The Gage s data and time can also be viewed in Gage Info (pg. 15). Battery Type Selects the type of batteries used in the Gage from a choice of Alkaline, Lithium or NiMH (Nickel-metal hydride rechargeable). If NiMH is selected, the Gage will trickle charge the batteries while connected via USB to a PC or optional AC charger (Gage must be powered ON). The battery state indicator icon is calibrated for the selected battery type. No damage will occur if the wrong battery type is selected. DeFelsko recommends eneloop (NiMH) rechargeable batteries (pg. 35). Units Converts the display from inch to metric or vice versa. Stored measurements in memory are not converted. Switching units will turn off Statistics, HiLo Alarm and closes Memory. Language Converts displayed and printed words to the selected language. 16
20 Statistics When selected, a the display. # of measurements Statistics Icon Statistics Menu icon and statistical summary will appear on Mean (average) Standard Deviation Minimum measurement Last reading Remove the last measurement by pressing the (-) button. Press (+) to clear statistics. HiLo Alarm x Maximum measurement n6 x 49.8x microns Allows the Gage to visibly and audibly alert the user when measurements exceed user-specified limits. When HiLo Alarm is selected, the current Lo setting is displayed. Adjust down (-) or up (+). Alternatively, measure a coating with a thickness close to the required value and make final adjustments with the buttons. Select NEXT to accept this value. The current Hi setting is now displayed. Follow the same procedure to adjust the Hi setting. σ The icon will appear on the display. Measurements will be compared to your defined HiLo limits. The Gage beeps if results are within those limits. A single low tone will sound if the reading is below the Lo limit, and a high tone if it is above the Hi limit. Press (+) to clear HiLo readings. Clear Clears all on-screen Statistics and HiLo tabulations. 17
21 Memory Management The PosiTector 6000 has internal memory storage for recording measurement data. Stored measurements can be reviewed on-screen or accessed via computers, tablets and smart phones. Measurements are date and time-stamped. Button functions with Memory ON: Scroll through display modes (Advanced only) Delete last reading Access the Menu Create a new batch (Advanced only) Restore brightness after dimming (Advanced only) Standard models store up to 250 readings in one batch. The Memory Menu includes the following options... On: turns memory on and begins recording Off: stops recording (stored readings remain in memory) Clear: removes all readings from memory View: lists group statistics and all stored readings on the display. It will begin by showing statistics based on all readings in memory. Use the Up and Down buttons to scroll through all readings. Press to exit. Standard model display with Memory ON shows measurement statistics: Mean (average) # of readings in memory Standard Deviation Maximum measurement Minimum measurement Last reading 18
22 Advanced models store 100,000 readings in up to 1,000 batches. The Memory Menu includes the following options... New Batch Closes any currently opened batch and creates a new batch name using the lowest available number. For example, if only Batch 1 and Batch 3 exist, then Batch 2 would be created and made the current batch. The icon appears and basic statistics are displayed. Each measurement will be displayed and stored into this new batch. On screen statistics are immediately updated with each measurement. New batch names are date stamped at the time they are created. Shortcut: When a batch is open, press (+) to create a new batch current batch Mean (average) Maximum measurement # readings in current batch Standard Deviation Minimum measurement Last reading NOTES: Remove the last reading from the current open batch by pressing (-). Calibration adjustments cannot be made if readings are stored in memory. 19
23 New Sub-Batch (appears only if a batch is currently open) Creates a new sub-batch within the currently opened batch. Shortcut: When a sub-batch is open, create a new sub-batch by pressing (+) In the following example, B2s2 is a sub-batch of Batch 2. Sub-batching allows the user to group related batches so that statistics can be accumulated for them. Batch 2 contains the statistics for B2s1 and B2s2. current batch current sub-batch statistics (see pg. 17) B2 s= x 1.5 s3 n5 72.4x σ σ microns # of sub-batches in current batch # readings in current sub-batch current reading New PA2 Helps determine if film thickness over a large area conforms to user specified min/max levels. See New 90/10 Determines if a coating system complies with the IMO performance standard for protective coatings. See Open Selects a previously created batch or sub-batch name to open and make current. If it contains measurements, on-screen statistics will immediately reflect values calculated from this batch. The calibration setting (i.e. Cal 2) associated with this batch is also opened (pg. 13). NOTE: A solid triangle is displayed to the right of the batch name when sub-batches are present. Press to view sub-batches. This also applies to the Delete, View and Print options. Close Stops the recording process, closes the current batch, and removes batch information from the display.
24 Delete Removes a batch or sub-batch completely from memory. The name is deleted and all measurements are erased. Sub-batches can be deleted individually. To delete all related sub-batches, simply delete the top-level batch. View Scroll using the Up or Down buttons through information, statistical summary, and a list of each reading in the currently opened batch. Press to exit. Annotate (Advanced models only) Create meaningful batch names and enter notes directly on the Gage using a familiar onscreen QWERTY keyboard. Use the Gages navigation and (-)(+) buttons to enter annotations. Annotations can be synchronized with PosiSoft.net, PosiSoft Mobile and are included in PosiSoft USB Drive reports. Print Sends a statistical summary and individual measurements to the optional Bluetooth wireless printer. NOTE: To cancel printing, press and hold the (-) and (+) buttons simultaneously 21
25 Display (appears only if a batch is currently open) The following user selectable display options are available: Chart: A real-time chart of batch readings x-y Chart for Single Layer Application Image: Batch image Notes: Instructions, descriptions or notes (see Annotate, pg.21) None: Default screen shows statistics Shortcut: When a batch is open, press Up to scroll through the above display options. 22
26 NOTES: PosiSoft.net and PosiSoft Mobile are used to insert an Image and Notes into a batch. Remove the last reading from the current open batch by pressing (-). Calibration adjustments cannot be made if any measurements were taken with that setting and stored into a batch. If memory is ON while continuous measurements are being taken, only the last value on the display (when the probe is lifted) is stored into memory. Scan Mode stores ALL measurements into memory (pg. 15). Each batch can contain a maximum of 10,000 readings. Summaries (Advanced models only) Displays a summary of all stored batches including the name, probe type, number of readings and type. In the following example, Batch 3 (B3) indicates an F 6000 coating thickness probe was used to record a total of 6 readings in PA2 mode. S c r e e n C a p t u r e Press both (-)(+) buttons at any time to capture and save an image copy of the current display. The last 10 screen captures are stored in memory and can be accessed when connected to a computer (see PosiSoft USB Drive pg. 27). Sample Screen Captures 23
27 Accessing Stored Measurements Data PosiSoft solutions for viewing, analyzing and reporting data: PosiSoft USB Drive - connect the PosiTector to a PC/Mac using the supplied USB cable to access and print stored readings, graphs, photos, notes and screen captures. No software or internet connection required. USB Drive must be selected (See pg. 27). PosiSoft.net - a free web-based application offering secure centralized storage of PosiTector readings. Access your data from any web connected device. Go to: PosiSoft Desktop Software for downloading, viewing and printing your measurement data. PosiSoft Mobile (Advanced models only) - access readings, graphs, capture photos and update annotations through WiFi enabled devices, such as tablets, smart phones and computers. PosiSoft Desktop Manager (PDM): A small Windows based application which allows automatic twoway communication (synchronization) between the Gage and PosiSoft.net (internet connection required). When installed, PDM runs as a start-up application and resides in the Windows notification area (system tray) of a PC. PDM is available as a free download within your PosiSoft.net account or at Register your Gage on PosiSoft.net to take full advantage of your Gage s capabilities. See NOTE: PDM is not required for WiFi enabled gages (pg. 30). 24
28 PosiSoft.net: To enhance the operation of your Gage, all PosiTector users have access to the features provided by PosiSoft.net. It is a web-based application offering secure centralized management of readings. After (1) a user account has been created, (2) a Gage has been registered, and (3) the PosiSoft Desktop Manager is downloaded and installed onto a Windows PC (not required for WiFi connected gages), synchronization of measurement data can be performed either manually or automatically whenever the PosiTector is connected to a web enabled PC (USB cable or Bluetooth wireless technology) or WiFi network (Advanced gages only). Gage measurements stored in memory are uploaded; images and batch notes are synchronized. Uploaded data can be manipulated using a standard internet web browser from any location in the world - job site or head office. Reports and graphs with annotations and corporate logo can be generated. Data can be exported to XML or CSV (comma delimited) text files. Measurement data can be shared with authorized users via a secure login from any computer and most web enabled devices including smart phones. NOTE: A PosiSoft.net account is not required to Update (pg. 33) your PosiTector. Simply download and install PosiSoft Desktop Manager at PosiSoft 3.0: Desktop Software that can be installed onto your Windows PC and allows the user to download, view, print and analyze their measurement data. Reports and graphs with annotations and corporate logo can be generated. NOTE: PosiSoft Software is available for use with DeFelsko s complete line of electronic instruments and is available for download at 25
29 PosiSoft Mobile: Is a Gage-based software application featured in all PosiTector Advanced instruments (serial numbers 730,000+). PosiSoft Mobile allows users to: browse stored measurement data including notes, images, statistics and charts update batch names/notes using your mobile device's keyboard insert images directly into Gage batches using your mobile device's camera or image library remotely view the live display of a working PosiTector & more PosiSoft Mobile can be accessed from any WiFi enabled device using a standard web browser. Ideal for use with devices such as Windows Phone/Mobile, Blackberry (RIM), Android, Apple ios and more. In addition to browser based access, PosiSoft Mobile Manager is available for Apple ios and Android users. See 26
30 Sync Now Connect Menu Synchronizing (Sync) is the process whereby PosiTector stored measurement data uploads to PosiSoft.net (from both Standard and Advanced models) while images and batch notes are downloaded to the Gage (Advanced model only). This transfer occurs when the Gage is connected via USB, Bluetooth or directly through your WiFi router/hotspot via WiFi. It can be triggered either manually (USB, Bluetooth or WiFi) or automatically (USB only). When selected, the Gage immediately synchronizes stored measurement data via USB, Bluetooth or WiFi to PosiSoft.net (PosiSoft Desktop Manager (pg. 24) and an internet connection are required when using USB or Bluetooth.) Multiple gages can be synchronized simultaneously when connected using multiple connections (for example two gages connected via USB and one gage connected via WiFi or Bluetooth). Auto SYNC Allows the Gage to automatically synchronize stored measurements with PosiSoft.net when initially connected to the internet via a PC running PosiSoft Desktop Manager (pg. 24) or a local WiFi network. Additional measurements added to memory while connected are synchronized only when the USB cable is disconnected, then reconnected or when the Sync Now option is selected. USB Drive Universal Serial Bus (USB) is a specification for communication between devices and a host controller (usually a personal computer). USB has effectively replaced a variety of interfaces such as serial and parallel ports. The PosiTector uses a USB mass storage device class which provides a simple interface to retrieve data in a manner similar to USB flash drives, cameras or digital audio players. 27
31 When checked, any computer can view datasets stored in memory by navigating a virtual drive labeled PosiTector using the included USB cable. A formatted HTML report is viewed by selecting the "index.html" or START_HERE.html file found in the root directory. Optionally, text ".txt" files located in each batch folder provide access to measurement values. Stored datasets and graphs can be viewed or copied using universal PC/Mac web browsers or file explorers. When your PosiTector is first connected to your Windows PC via a USB cable, an enumeration process is started that installs device drivers without re-booting your computer. You may see several pop-up windows in the taskbar at the bottom right of your screen. Wait for the entire process to be completed before proceeding. NOTE: When connected, power is supplied through the USB cable. The batteries are not used and the body will not automatically power down. If rechargeable (NiMH) batteries are installed, the Gage will trickle charge the batteries. 28
32 Retrieving stored screen captures The last 10 screen captures (pg. 23) stored in memory can be accessed by navigating to the screen_capture directory within the PosiTector virtual drive. Bluetooth (Advanced models only) PosiTector Advanced models have Bluetooth functionality to: - Communicate with the PosiSoft Desktop Manager (PDM) in lieu of a USB cable. - Stream individual datasets to a computer or Bluetooth wireless printer as they are taken. - Print to the optional battery powered Bluetooth wireless printer. On: Turns Bluetooth functionality On. When selected, the icon will appear on the display. To deactivate Bluetooth, select Off. Pairing: The Gage and receiving device must be paired before stored or streamed datasets can be transmitted. For pairing instructions, see Info: Lists information about your current Bluetooth connection, including the currently paired device and MAC address. Stream: When checked, the Gage will stream datasets to the paired Bluetooth Device as they are taken. Datasets can be streamed as they are taken to the optional Bluetooth printer or third-party computer software. 29
33 Bluetooth Wireless Printer PosiTector Advanced models can output to the optional battery powered Bluetooth wireless printer one of two ways: Stream individual readings as they are taken. Print stored batch readings and summaries. Begin by entering the Connect > Bluetooth menu. Turn Bluetooth ON and Pair the PosiTector to the printer. See: Streaming: In the Connect > Bluetooth menu, select the Stream tick box. All readings will now be simultaneously displayed on the LCD and sent to the printer. Printing: In the Memory menu, select Print. WiFi (Advanced models only) Allows wireless communication with devices such as tablets, smart phones and computers connected to your local wireless network or portable mobile hot spot. See On/Off : Turns WiFi functionality On. When selected, the icon will appear on the display. To deactivate WiFi, select Off. Access Point : Connect your smart device/computer to a PosiTector Advanced body wirelessly without the need for a separate network. Wirelessly import readings into PosiSoft 3.0 Desktop Software (pg. 24) or access PosiSoft Mobile (pg. 26) in the field or whenever a WiFi network is not available or out-of-range. To enable, select Access Point from the Connect > WiFi menu. The Access Point icon will display in the upper left of the PosiTector display. 30
34 Securing your Access Point To ensure the PosiTector is only accessible to authorized devices, it is important that you enter a passphrase (password) for the Access Point. The default Passphrase is password. In the Connect > WiFi > Setup menu, select AP Passphrase. Press button to display on-screen keyboard. Enter a Passphrase for the Access Point. The Passphrase will be required for all devices connecting to the PosiTector s Access Point. The PosiTector is now visible to all WiFi enabled devices. Simply connect your devices WiFi to the new PosiTector Access point. All PosiTectors are uniquely identified by their respective gage body serial numbers. AP Channel Default Channel: 6 For most users the default channel will not have to be altered. If you are experiencing poor connection or are unable to connect, try another channel. In the Connect > WiFi > Setup menu, select AP Channel. Press the UP center navigation button to highlight the channel. Use the (-) or (+) buttons to change the channel. Press the DOWN navigation button and select OK. Press the button. Import readings into PosiSoft 3.0 Desktop Software (pg. 25) From the PosiSoft 3.0 menu, select Import > From Gage. All connected gages will display their unique serial number. Measurement data will be displayed upon completion of the import. The time it takes for import will depend on the number of Batches and measurements present within the gage. Imports generally take no longer than a few seconds. Batches are not erased from the gages memory during import. They remain on the gage until they have been deleted by the user. For more information regarding PosiSoft 3.0: 31
35 Accessing your PosiTector PosiSoft Mobile (pg. 26) For use on Apple ios and Android devices: PosiSoft Mobile Manager apps exist for Apple ios and Android users. Simply download the App from the Apple App Store or Google Play. Once your smart device/computer has been connected to the PosiTector Access Point, open the PosiSoft Mobile Manager app. Select your PosiTector from the list to open PosiSoft Mobile. For use with all WiFi enabled devices: Any WiFi enable device connected to the PosiTector Access Point can view the instrument s built-in PosiSoft Mobile web interface. Using your device s web browser, enter the IP address of your PosiTector to access the features of PosiSoft Mobile. In the Connect > WiFi menu, select Information to view the IP address associated with your PosiTector. Enter the displayed IP address into your devices web browser exactly as it appears. Example: :8080 For more information regarding PosiSoft Mobile: Networks: The Gage will allow the user to add a new network and will automatically check for available local networks. Available networks detected by the Gage are listed on the screen along with any networks that the Gage has previously been connected to that are not currently within connection range. Information: Gage displays information about the local WiFi network connection including... SSID: the network s name State: displays if the Gage is connected to the network or not IP Address: the network s IP Address. Users can enter this number into a web-browser of any WiFi enabled device that is connected to the same network in order to view the Gage s synced batches through PosiSoft Mobile (See pg. 26). 32
36 Setup: Allows user to setup a WiFi connection IP Settings: enter the IP information as follows... IP Type (DHCP or Static), IP Address, Gateway, Netmask, DNS1, DNS2. Server Enable: enables a connection between the network and the Gage. Gage Name: enter a name for the Gage (up to 14 Characters). WiFi Reset: erases all WiFi settings. Updates Determines if a software update is available for your Gage. If an update is available, a prompt will appear allowing the user to choose to perform the update at this time or not. To perform an update the Gage must be connected to an internet connected PC with PosiSoft Desktop Manager (pg. 24), or WiFi network (pg. 30). NOTE: Ensure that stored measurements are backed up to a PC or PosiSoft.net. The Gage may perform a Hard Reset (pg. 14) after completion of the update and ALL datasets in memory would be erased. Updates - Sample Screen Currently installing update 2 of 3 total 2 / 3 Progress bar! DO NOT unplug the Gage during the update operation. 33
37 Temperature Operating Range: +32 to +120 F (0 to +50 C) The PosiTector 6000 compensates automatically for temperature. Allow a few minutes for the probe to reach ambient temperature before measuring. Discard the first measurement taken in a notably different temperature condition. When measuring surfaces much hotter or colder than ambient, lift the probe at least 6 inches (15cm) and allow 1 second off the surface between measurements. Power Supply / Battery Indicator Power Source: 3 AAA alkaline, Lithium or optional Nickel-metal hydride (NiMH) rechargeable batteries. For best battery indicator results, ensure the appropriate Battery Type is selected in the Setup > Battery Type menu (pg. 16). The battery indicator displays a full bar with fresh alkaline or fully charged batteries installed. As the batteries weaken, the bar will be reduced. When the battery symbol is low the Gage can still be used, but the batteries should be changed or recharged at the earliest opportunity. The Gage will turn off automatically when batteries are very low, preceded by a Low Battery Warning on the display.! To retain all user settings and stored memory readings, only replace the batteries after the Gage has powered-down. Battery performance decreases at low temperatures. Rechargeable Batteries (available option) DeFelsko recommends the use of eneloop (NiMH) rechargeable batteries. They combine the advantages of rechargeable batteries and disposable (Alkaline) batteries. They discharge very slowly and can be stored for long periods without having self discharge concerns. (See 34
38 Additional Accessories Rechargeable Battery Pack - 4 AAA eneloop batteries The PosiTector can operate on rechargeable batteries and DeFelsko recommends the use of eneloop (NiMH) rechargeables. eneloop batteries combine the advantages of regular rechargeable batteries and disposable (Alkaline) batteries. They discharge very slowly and can be stored for long periods without having self discharge concerns. Eneloop batteries come pre-charged and ready to use immediately. See: AC Power Cable Kit An optional AC Power Cable Kit is available for continuous operation or battery charging through the PosiTector s built-in USB port (as shown on pg. 24). This kit supplies several alternate power solutions for your battery-operated PosiTector. They allow the gage to operate continuously without the need for batteries. Use the cable alone to connect a PosiTector to your PC's built-in USB port that acts as a continuous power source. Or connect the cable to the included power adaptor which plugs into any AC wall electrical outlet, 110 or 220V. A selection of electrical plugs is included which are capable of dealing with most country s outlets.the USB cable provided can also be used for Accessing Stored Measurement Data (See pg. 24). USB Cable A USB Cable is provided with every PosiTector. Bluetooth Printer This lightweight printer receives data from all PosiTector gages via Bluetooth wireless technology. It is battery-operated and prints readings and statistical summaries. 35
39 Nylon Case Convenient soft shell nylon case for carrying a PosiTector. Case has compartments for Certificates of Calibration, instruction manuals and other accessories. PosiTector Case Convenient hard shell plastic case for carrying a PosiTector and multiple probes - including any of the PosiTector Coating Thickness, PosiTector Ultrasonic Coating Thickness, DPM - Dew Point Meter, SPG - Surface Profile, RTR - Replica Tape Reader, or UTG - Ultrasonic Wall Thickness probes. Case has compartments for Certificates of Calibration, instruction manuals and other accessories. Protective Lens Shield One lens shield is included with every PosiTector instrument. Additional package of five (5) thin plastic lens shields are available and ideal for protecting the PosiTector display from paint and overspray. Coating Thickness Standards Certified Thickness Standards are used to verify the accuracy and operation of coating thickness gages and are an important component in fulfilling both ISO/QS-9000 and in-house quality control requirements. Contracts often specify that coating thickness measurements be taken by gages whose measurement accuracy is traceable to a National Metrology Institute such as NIST or PTB. Standards are typically purchased as an accessory to coating thickness gages. Many customers find it more practical to verify the accuracy of their own gages in-house, rather than utilize DeFelsko s calibration services. This is particularly true when many gages are in use and/or when accuracy verification is performed often. 36
40 There are 5 types of coating thickness standards available: Certified Coated Metal Plates are the best solution for verifying the calibration, accuracy and operation of most magnetic, eddy current or ultrasonic coating thickness gages including magnetic pull-off gages as well as many competitive models. They fulfill both ISO and in-house quality control requirements. At ±0.43µm (±0.017 mil) Certified Coated Metal Plates are the most accurate solution. Certified Polystyrene Blocks are for use with gages that measure over 1500 µm (60 mils). Certified Plastic Shims (foils) provide an economical alternative to coated metal plates. They have a reduced accuracy of ±2 µm (±0.08 mil). A coating thickness Gage measures shim thickness when this shim is placed over a smooth metal surface (zero plate). Non-Certified Plastic Shims (foils) provide a quick operational check of the Gage and they allow the user to perform practice measurements when placed over metal. They protect the probe from damage or premature wear when placed over hot or abrasive surfaces, and they can be placed on top of soft or tacky coating films to obtain thickness measurements without the Gage probe depressing the coating film. A set of 5 shims is included with every PosiTector Zero plates are uncoated steel or aluminum disks. A zero plate is included with most coated metal sets. The first Gage measurement is on a zero plate to verify that the Gage measures zero correctly. Electronic gages can usually be adjusted to 0 if necessary. A zero plate is also a convenient base onto which a plastic shim or polystyrene block can be placed for measurement. Zero plates are also used to provide a substrate where none exists. Large sheets of material such as paper, plastic, fabric and rubber can be measured with a coating thickness Gage by placing the material over the zero plate. This is useful when the measured item is too large to access with a micrometer or other device. For more information, see 37
41 Troubleshooting Most conditions can be cleared with a Hard Reset (see pg. 14). If not, visit our Service & Support page for information on sending your Gage to us for a no charge evaluation. See: Probe attempts a measurement even while held in the air: The probe may have been left near metal, including jewelry, during power up or at rest. Possibly a finger was held over the probe. Try measuring on an uncoated object. Otherwise turn the Gage off, and then on again. Finally try a Reset (pg. 14). Thickness readings are inconsistent: - Probe tip may be damaged, scratched or worn - Make sure you lift the probe well clear of the surface between measurements - Try measuring on a different surface with the included plastic shims - Hold the probe on the surface and allow it to take several measurements. If the second and subsequent measurements are consistent, you are not putting the probe onto the surface fast enough. - The substrate and/or coating surface is uneven, in which case inconsistent readings are to be expected. Take several measurements and average them to get a meaningful result. Gage powers up but will not take any readings: - Substrate may not be metal. F probes measure coatings on ferrous metals such as steel and iron, and N probes measure over non-ferrous metals such as aluminum. - Protective cap must be removed from probe (if outfitted with one). - An extreme calibration adjustment might have been made. Gage will not print to the Bluetooth wireless printer: - Ensure the Gage has been paired for use with the printer and the Gage menu option Connect > Bluetooth > Stream box is ticked. - Check printer batteries. Switch it off and then on again. 38
42 Measuring Range: varies depending probe model Technical Data Gage Size: 137 x 61 x 28 mm 5.4 x 2.4 x 1.1 Accuracy: varies depending on probe model Weight: (without batteries) 140 g 4.9 oz Resolution: varies depending on probe model Temperature Range: +32 to +120 F 0 to +50 C Conforms to: ISO 2178/2360/2808, ISO 19840, ASTMB499/D1186/D1400/D7091/E376/ G12, BS3900-C5, SSPC-PA2 and others This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. 39
43 Returning for Service Before returning the Gage for service: 1.Install new or newly recharged batteries in the proper alignment as shown within battery compartment. 2.Examine the probe tip for dirt or damage. The probe should move up and down freely. 3.Perform a Hard Reset (pg. 14) 4.Place a plastic shim onto bare metal (steel or non-steel, depending upon whether you have an "F" or "N" probe) and attempt a measurement. IMPORTANT: If you must return the Gage for service, please fill out and include the Service Form located at with the Gage. Be sure to also include the probe, your company name, company contact, telephone number and fax number or address. Limited Warranty, Sole Remedy and Limited Liability DeFelsko's sole warranty, remedy, and liability are the express limited warranty, remedy, and limited liability that are set forth on its website: DeFelsko Corporation USA 2015 All Rights Reserved This manual is copyrighted with all rights reserved and may not be reproduced or transmitted, in whole or part, by any means, without written permission from DeFelsko Corporation. DeFelsko and PosiTector are trademarks of DeFelsko Corporation registered in the U.S. and in other countries. Other brand or product names are trademarks or registered trademarks of their respective holders. Every effort has been made to ensure that the information in this manual is accurate. DeFelsko is not responsible for printing or clerical errors. 40
44 Simple. Durable. Accurate. DeFelsko Corporation USA 2015 All Rights Reserved This manual is copyrighted with all rights reserved and may not be reproduced or transmitted, in whole or part, by any means, without written permission from DeFelsko Corporation. DeFelsko, PosiTector and PosiSoft are trademarks of DeFelsko Corporation registered in the U.S. and in other countries. Other brand or product names are trademarks or registered trademarks of their respective holders. Every effort has been made to ensure that the information in this manual is accurate. DeFelsko is not responsible for printing or clerical errors.
PosiTector. Coating Thickness Gages. INSTRUCTION MANUAL v. 7.0 for ALL models
PosiTector 6000 Coating Thickness Gages INSTRUCTION MANUAL v. 7.0 for ALL models Introduction The PosiTector 6000 hand-held, electronic Gage nondestructively measures the thickness of coatings on all metals,
1. Check the Accessories
This Quick User Guide helps you get started with the IRIScan Book Executive 3 scanner. This scanner is supplied with the software applications Readiris Pro 14 (Windows and Mac ), IRIScan Direct and IRISCompressor
QUICK START GUIDE Bluetooth Cordless Hand Scanner (CHS)
QUICK START GUIDE Bluetooth Cordless Hand Scanner (CHS) 1D Imager Models CHS 7Ci, 7Di, 7DiRx LED Trigger button Power button (also for ios Keyboard Pop-up) Model shown: CHS 7Di This document pertains to
Getting Started. Table of Contents. Quick User Guide - English
Getting Started This Quick User Guide helps you get started with the IRIScan TM Anywhere Wifi. Please read this guide before operating this scanner and its software. All information is subject to change
BLU Vivo 4.3 User Manual
BLU Vivo 4.3 User Manual 1 Table of Contents Phone Safety Information... 3 Appearance and Key Functions... 4 Installation and Charging... 5 Phone Operation... 7 Communication... 10 Camera... 11 Gallery...
Using the T&D Thermo App with TR-7wf Data Loggers
Using the T&D Thermo App with TR-7wf Data Loggers T&D Thermo The T&D Thermo App from T&D Corporation can be used to accomplish a variety of functions with the TR-7wf Series of Wi-Fi enabled wireless Data
Getting to Know Xsight... 3. Home Screen and Everyday Use... 9. Settings... 11. PC Setup... 17. 2 www.oneforall.com
User Manual English English Getting to Know Xsight........................................ 3 Unpacking and Setting Up Xsight............................................................. 3 Using Xsight
Point of View SmartTV HDMI 210 dongle - Android 4.2. General notices for use... 2 Disclaimer... 2 Box Contents... 2
Table of Contents General notices for use... 2 Disclaimer... 2 Box Contents... 2 1.0 Product basics... 3 1.1 Connecting your device for the first time... 3 1.2 Connections... 4 1.3 Using the remote control
Nighthawk AC1900 WiF Range Extender
Nighthawk AC1900 WiF Range Extender Model EX7000 User Manual December 2014 202-11469-01 350 East Plumeria Drive San Jose, CA 95134 USA Nighthawk AC1900 WiF Range Extender Support Thank you for selecting
Windows 10.1 Tablet (UB-15MS10 and UB-15MS10SA) FAQ December 2014
Windows 10.1 Tablet (UB-15MS10 and UB-15MS10SA) FAQ December 2014 Navigation... 2 Question 1: What are charms used for?... 2 Question 2: How do I navigate while using a mouse or trackpad?... 3 Question
This guide describes features that are common to most models. Some features may not be available on your tablet.
User Guide Copyright 2014 Hewlett-Packard Development Company, L.P. Bluetooth is a trademark owned by its proprietor and used by Hewlett-Packard Company under license. SD Logo is a trademark of its proprietor.
User Guide. Temperature and Humidity Datalogger. Model 42280
User Guide Temperature and Humidity Datalogger Model 42280 Introduction Congratulations on your purchase of the Extech 42280 Thermometer and Relative Humidity Datalogger. The 42280 is a wall-mount, tripod
MAINTENANCE & TROUBLESHOOTING
MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens replace the projection lamp replace the batteries in the remote use the security lock feature troubleshoot the projector Cleaning
SNMP Web Management. User s Manual For SNMP Web Card/Box
SNMP Web Management User s Manual For SNMP Web Card/Box Management Software for Off-Grid Inverter Version: 1.2 Table of Contents 1. Overview... 1 1.1 Introduction... 1 1.2 Features... 1 1.3 Overlook...
4-Channel Thermometer / Datalogger
USER GUIDE 4-Channel Thermometer / Datalogger RTD and Thermocouple Inputs Model SDL200 Introduction Congratulations on your purchase of the Extech SDL200 Thermometer, an SD Logger Series meter. This meter
7 Mini Tablet User Guide
7 Mini Tablet User Guide MODEL NUMBER: Powered by SYTABBL7 Wireless Mobile Internet Tablet Getting to Know Your New Tablet: Front View: Back View: Side Angle View: MENU MENU Power Key ON/OFF, Back Button
How To Use An Easypad On A Pc Or Macbook Or Ipad (For Pc)
User Guide Congratulations on your new EasyPad 1000. Please follow the instructions in this manual carefully. Statement: If the final version of this product has any differences from the description of
Lenovo Miix 2 8. User Guide. Read the safety notices and important tips in the included manuals before using your computer.
Lenovo Miix 2 8 User Guide Read the safety notices and important tips in the included manuals before using your computer. Notes Before using the product, be sure to read Lenovo Safety and General Information
Quick Start Guide. Cisco SPA232D Mobility Enhanced ATA
Quick Start Guide Cisco SPA232D Mobility Enhanced ATA Package Contents Analog Telephone Adapter Ethernet Cable Phone Cable Power Adapter Quick Start Guide Product CD-ROM Welcome Thank you for choosing
AC1200 WiFi Range Extender
Model EX6200 User Manual February 2014 202-11298-01 350 East Plumeria Drive San Jose, CA 95134 USA Support Thank you for selecting NETGEAR products. After installing your device, locate the serial number
Point of View SmartTV-500 Center - Android 4.2. General notices for use...2 Disclaimer...2 Box Contents...2
Point of View SmartTV-500 Center - Android 4.2 English Table of Contents General notices for use...2 Disclaimer...2 Box Contents...2 1.0 Product basics...3 1.1 Buttons and connections... 3 1.2 Connecting
How to connect your D200 using Bluetooth. How to connect your D200 using GPRS (SIM Card)
D200 En User Guide Content 3 Introduction 3 Scope 3 Related Documentation 4 Internet Connectivity 4 Using D200 with Mobile Phone 5 Using D200 with wireless (Wi-Fi) router 6 Using D200 with GPRS (requires
User Manual. Product Model: MiTraveler 10C3. OS: Android 4.0
User Manual Product Model: MiTraveler 10C3 OS: Android 4.0 Tablet of Contents Specifications.1 What s inside the box 2 Tablet Parts 2 Getting started...2 Connection to Internet..6 Task Bar..8 Install and
This section will focus on basic operation of the interface including pan/tilt, video, audio, etc.
Catalogue Basic Operation... 2 For Internet Explorer... 2 For Other Non-IE Web Browsers... 5 Camera Settings... 6 System... 6 About... 6 PT Setting... 7 Backup and Restore Setup... 8 NTP Setting... 8 System
Table of Contents. Use. Troubleshooting. Setup. Welcome. 11 How to arm/disarm system/camera(s) 19 Sync Module setup issues. 3 Installing the Blink app
User Guide Table of Contents Welcome Setup Use Troubleshooting 2 What s in the box 2 What you need 3 Installing the Blink app 4 Setting up the Sync Module 5 Connecting the Sync Module to Your Wi-Fi 7 Adding
Contents Contents Contents Overview Charging your headset Wearing the headset Powering on the headset Connecting your headset for dictation
1 Contents Contents Contents 2 Overview 3 Charging your headset 4 Wearing the headset 4 Powering on the headset 4 Connecting your headset for dictation 5 The adapter 5 Adjusting your computer's audio settings
Specifications. What s Inside The box
Tablet of Contents Specifications.1 What s inside the box Tablet Parts Getting started... Connection to Internet..6 Task Bar..8 Install and Uninstall 9 Camera.10 HDMI.11 Trouble Shooting 11 Specifications
Technical Support. Package Contents. CENTRIA WNDR4700/WNDR4720 Installation Guide
Technical Support After installing your device, locate the serial number on the label of your product and use it to register your product at https://my.netgear.com. You must register your product before
Getting Started... 1. What s included... 1. Setting up Fitbit One on a computer... 2. Mac & PC Requirements... 2
User Manual Table of Contents Getting Started... 1 What s included... 1 Setting up Fitbit One on a computer... 2 Mac & PC Requirements... 2 Installing Fitbit Connect on a computer... 2 Installing Fitbit
Android Dongle. User Manual TL869
Android Dongle User Manual TL869 2 User Manual Android TV Dongle Before using the product, please read this manual thoroughly. The specifications and software in this manual are subject to change without
Start Here USB. (802.11) (Ethernet)
Start Here 1 USB 802.11 Ethernet USB cable users: Do not connect the USB cable until Section A. USB A USB Wireless (802.11) or wired (Ethernet) network users : you must follow the instructions in this
Point of View ProTab 3XXL IPS - Android 4.0 Tablet PC. Contents... 1 General notices for use... 2 Disclaimer... 2 Box Contents...
Point of View ProTab 3XXL IPS - Android 4.0 Tablet PC English Contents Contents... 1 General notices for use... 2 Disclaimer... 2 Box Contents... 2 1.0 Product basics... 3 1.1 Buttons and connections...
BigPond Wi-Fi 4G Advanced II AirCard 790S User Manual
AirCard 790S User Manual October 2014 202-11471-02 350 East Plumeria Drive San Jose, CA 95134 USA Support Contact your Internet service provider for technical support. Compliance For regulatory compliance
N750 WiFi DSL Modem Router Premium Edition
Support Thank you for purchasing this NETGEAR product. After installing your device, locate the serial number on the label of your product and use it to register your product at https://my.netgear.com.
INSTRUCTION MANUAL All-In-One GSM Home Alarm System SB-SP7200-GSM
INSTRUCTION MANUAL All-In-One GSM Home Alarm System SB-SP7200-GSM Revised: August 28, 2014 PRODUCT REFERENCE MOUNTING ACCESSORIES PIR / MOTION DETECTION UNIT MAIN UNIT POWER ADAPTER MOUNTING ACCESSORIES
Product Specification instalert Rapid Messenger Variable Message Sign
instalert 2 units to cover any application instalert 18 (ia18): 18 x 28 full matrix instalert 24: (ia24): 24 x 60 full matrix Size, Weight without battery ia18: 30 x 20 x 2.74, 29 lbs ia24: Folds to fits
User Manual. Thermo Scientific Orion
User Manual Thermo Scientific Orion Orion Star Com Software Program 68X637901 Revision A April 2013 Contents Chapter 1... 4 Introduction... 4 Star Com Functions... 5 Chapter 2... 6 Software Installation
ImagineWorldClient Client Management Software. User s Manual. (Revision-2)
ImagineWorldClient Client Management Software User s Manual (Revision-2) (888) 379-2666 US Toll Free (905) 336-9665 Phone (905) 336-9662 Fax www.videotransmitters.com 1 Contents 1. CMS SOFTWARE FEATURES...4
User Guide. for Card Reader and Payment Application
User Guide for Card Reader and Payment Application Introduction to Pogo> This document provides detailed information on how best to use the Pogo> application to benefit your business. The intention is
Video Baby Monitor System. User Guide
Video Baby Monitor System User Guide What s inside Welcome!... 2 Getting started... 3 Tour of the system... 7 Everyday use...13 Cameras... 14 Motion/Alerts... 18 Recording... 21 Playing Back Pictures and
Preparing Log-ic 360 USB Loggers for Use
Preparing Log-ic 360 USB Loggers for Use Preparing Log-ic 360 USB Loggers for Use Step 1) Install the latest version of the Log-ic Software. Step 2) Connect the Desktop Reader or plug in a USB logger.
CHAPTER 2: USING THE CAMERA WITH THE APP
TABLE OF CONTENTS OVERVIEW... 1 Front of your camera... 1 Back of your camera... 2 ACCESSORIES... 3 CHAPTER 1: Navigating the Mobile Application... 4 Device List: How to Use this Page... 4 My Messages:
KANGAROO MOBILE DESKTOP USER GUIDE
KANGAROO MOBILE DESKTOP USER GUIDE Copyright InFocus Corporation., 2015. All rights reserved. This document contains information proprietary to InFocus Corporation, hereinafter InFocus. Use or disclosure
This guide describes features that are common to most models. Some features may not be available on your tablet.
User Guide Copyright 2013 Hewlett-Packard Development Company, L.P. Bluetooth is a trademark owned by its proprietor and used by Hewlett-Packard Company under license. Google is a trademark of Google Inc.
Using AORUS Notebook for the First Time
V2.0 Congratulations on your purchase of the AORUS Notebook! This Manual will help you to get started with setting up your notebook. For more detailed information, please visit our website at http://www.aorus.com.
Networking. General networking. Networking overview. Common home network configurations. Wired network example. Wireless network examples
Networking General networking Networking overview A network is a collection of devices such as computers, printers, Ethernet hubs, wireless access points, and routers connected together for communication
Point of View Mobii 925 - Android 4.2 Tablet PC. General notices for use... 2 Disclaimer... 2 Box Contents... 2
Table of Contents General notices for use... 2 Disclaimer... 2 Box Contents... 2 1.0 Product basics... 3 1.1 Buttons and connections... 3 1.2 Start up and shut down... 3 2.0 Introduction to Google Android
SA-9600 Surface Area Software Manual
SA-9600 Surface Area Software Manual Version 4.0 Introduction The operation and data Presentation of the SA-9600 Surface Area analyzer is performed using a Microsoft Windows based software package. The
GSM Alarm System User Manual
GSM Alarm System User Manual For a better understanding of this product, please read this user manual thoroughly before using it. Quick Guider After getting this alarm system, you need to do the following
Thank you for purchasing the Mobile WiFi. This Mobile WiFi brings you a high speed wireless network connection.
Quick Start Thank you for purchasing the Mobile WiFi. This Mobile WiFi brings you a high speed wireless network connection. This document will help you understand your Mobile WiFi so you can start using
ibaby Monitor Model: M3s User Manual
ibaby Monitor Model: M3s User Manual Index 1 Introduction...3 1.1 The package contents...3 1.2 Product Features...3 1.3 Product Specifications...4 2 Appearance and Interface...5 2.1 Appearance...5 2.2
Tablet PC User Manual
Tablet PC User Manual Please read all instructions carefully before use to get the most out of your Time2Touch Tablet PC. The design and features are subject to change without notice. Contents Important
N300 WiFi Range Extender WN2000RPT User Manual
N300 WiFi Range Extender WN2000RPT User Manual December 2013 202-11333-01 350 East Plumeria Drive San Jose, CA 95134 USA Support Thank you for selecting NETGEAR products. After installing your device,
Experience Next-Generation WiFi
Do More Experience Next-Generation WiFi Your new WiFi extender supports the 802.11ac networking standard. This new standard offers better speed, improved reliability, and more range than older WiFi networking
Using GIGABYTE Notebook for the First Time
Congratulations on your purchase of the GIGABYTE Notebook. This manual will help you to get started with setting up your notebook. The final product configuration depends on the model at the point of your
User Manual. Product Model: MiTraveler 7D-4A. OS: Android 4.0
User Manual Product Model: MiTraveler 7D-4A OS: Android 4.0 Tablet of Contents Specifications.1 What s inside the box 2 Tablet Parts 2 Getting started...2 Connection to Internet..6 Task Bar..8 Install
Using GIGABYTE Notebook for the First Time
Congratulations on your purchase of the GIGABYTE Notebook. This manual will help you to get started with setting up your notebook. The final product configuration depends on the model at the point of your
Using GIGABYTE Notebook for the First Time
P34 V6.0 Congratulations on your purchase of the GIGABYTE Notebook. This manual will help you to get started with setting up your notebook. The final product configuration depends on the model at the point
COPY EOS 70D (W) Wi-Fi Function Instruction Manual INSTRUCTION MANUAL
EOS 70D (W) Wi-Fi Function Instruction Manual E INSTRUCTION MANUAL What You Can Do Using the Wireless LAN Functions This camera s wireless LAN functions let you perform a range of tasks wirelessly, from
Thank you for purchasing the Mobile WiFi. This Mobile WiFi brings you a high speed wireless network connection.
Quick Start Thank you for purchasing the Mobile WiFi. This Mobile WiFi brings you a high speed wireless network connection. This document will help you understand your Mobile WiFi so you can start using
Nokia Bluetooth Headset BH-607. Issue 2. 1
Nokia Bluetooth Headset BH-607 2 3 1 4 5 6 7 8 9 11 10 Issue 2. 1 12 13 14 DECLARATION OF CONFORMITY Hereby, NOKIA CORPORATION declares that this BH-607 product is in compliance with the essential requirements
PA500II Enterprise Mobile Computer
PA500II Enterprise Mobile Computer - PA500II - Quick Reference Guide 400891G Version 1.0 PA500II Product Introduction & Accessory Kit After opening the box, ensure the following accessories for the PA500II
EOS 750D (W) Wi-Fi/NFC Function Instruction Manual ENGLISH INSTRUCTION MANUAL
EOS 750D (W) Wi-Fi/NFC Function Instruction Manual ENGLISH INSTRUCTION MANUAL Introduction What You Can Do Using the Wireless LAN Functions This camera s wireless LAN functions let you perform a range
FAQs. Conserve package. Gateway... 2 Range Extender... 3 Smart Plug... 3 Thermostat... 4 Website... 7 App and Mobile Devices... 7
FAQs Conserve package Gateway... 2 Range Extender... 3 Smart Plug... 3 Thermostat... 4 Website... 7 App and Mobile Devices... 7 FAQs Gateway Can I have someone install my system for me? If you are concerned
Model 288B Charge Plate Graphing Software Operators Guide
Monroe Electronics, Inc. Model 288B Charge Plate Graphing Software Operators Guide P/N 0340175 288BGraph (80207) Software V2.01 100 Housel Ave PO Box 535 Lyndonville NY 14098 1-800-821-6001 585-765-2254
owner s manual EDGE 200 GPS-ENABLED CYCLING COMPUTER December 2012 190-01368-00_0B Printed in Taiwan
EDGE 200 GPS-ENABLED CYCLING COMPUTER owner s manual December 2012 190-01368-00_0B Printed in Taiwan Getting Started warning Always consult your physician before you begin or modify any exercise program.
User Manual I1010Q16DCZ
User Manual I1010Q16DCZ Explanation, Note and Warning *Explanation: important information, which can help you better use your tablet PC. *Note: indicates that not observing the content may result in hardware
Frequently Asked Questions: Cisco Jabber 9.x for Android
Frequently Asked Questions Frequently Asked Questions: Cisco Jabber 9.x for Android Frequently Asked Questions (FAQs) 2 Setup 2 Basics 4 Connectivity 8 Calls 9 Contacts and Directory Search 14 Voicemail
Multi Client (Integration)
Multi Client (Integration) Installation and User s Guide Ver 12.10.x *All contents of this document may change without prior notice. INDEX 1. Index 01 2. Introduction 02 3. Installation 03 4. Using Multi
Technical Support. Trademarks
Getting Started Technical Support Thank you for selecting NETGEAR products. After installing your device, locate the serial number on the label of your product and use it to register your product at https://my.netgear.com.
Mobility with Eye-Fi Scanning Guide
Mobility with Eye-Fi Scanning Guide Scan and Transfer Images Wirelessly with Eye-Fi This document is to be used in addition to the scanner s user guide located on the installation disc. The instructions
10 Magni Tablet User Guide
10 Magni Tablet User Guide MODEL NUMBER: Wireless Mobile Internet Tablet Powered by SYTAB10ST Top View Front View Bottom View Side View SYTAB10ST Getting to Know Your New Tablet: MINI Micro SD Micro SD
WAMLocal. Wireless Asset Monitoring - Local Food Safety Software. Software Installation and User Guide BA/WAM-L-F
Wireless Asset Monitoring - Local Food Safety Software BA/WAM-L-F Software Installation and User Guide System Overview The BAPI Wireless Asset Monitoring Local (WAM Local) Software receives temperature
GETTING TO KNOW YOUR TELSTRA PRE-PAID 3G WI-FI
GETTING TO KNOW YOUR TELSTRA PRE-PAID 3G WI-FI LET S GET THIS SHOW ON THE ROAD You must be excited about your brand new Telstra Pre-Paid 3G Wi-Fi. This guide will help you get connected as quickly and
Eclipse 180 User s Manual
Eclipse 180 User s Manual Congratulations on purchasing your Eclipse Personal Media Player! To get the most out of your purchase, please take the time to carefully read through this manual. Foreword If
1. Open the battery compartment as shown in the image.
This Quick User Guide helps you get started with the IRIScan Book 3 scanner. This scanner is supplied with the software applications Readiris Pro 12, IRIScan Direct and IRISCompressor. Corresponding Quick
VIRTUAL DESKTOP COMPANION SOFTWARE FOR MAC
VIRTUAL DESKTOP COMPANION SOFTWARE FOR MAC USER MANUAL 25206A ABOUT THIS MANUAL Thank you for getting the Boogie Board Sync 9.7 LCD ewriter. This manual will help you with the Sync Virtual Desktop Companion
User Guide for Windows 10
User Guide for Windows 10 System requirements E10684 First Edition July 2015 To facilitate a smoother transition from your previous operating system, read the system requirements below before upgrading
Monnit Wi-Fi Sensors. Quick Start Guide
Monnit Wi-Fi Sensors Quick Start Guide Inside the Box You should find the following items in the box: Monnit Wi-Fi (MoWi ) Sensors 2 x AA Batteries for Each MoWi Sensor Quick Start Guide Note: A MoWi USB
CX Series. Video Recording Server. Quick Start Guide CX784 / CX788 / CX7816. Version 1.05.00
CX Series Video Recording Server CX784 / CX788 / CX7816 Quick Start Guide Version 1.05.00 Contents 1.Introduction...1 1.1.Packages Contents...1 1.2.Hardware Features...2 1.3.Functional Features...4 1.4.Hard
Ultra Dashcam. with Smartphone app control INSTRUCTION MANUAL
Ultra Dashcam with Smartphone app control EN INSTRUCTION MANUAL 1 Getting to know the Dashcam GPS Receiver Slot 1 Camera Lens Micro HDMI Port 2 Speaker 1 For future application 2 Requires Micro HDMI to
ASUS WebStorage Client-based for Windows [Advanced] User Manual
ASUS WebStorage Client-based for Windows [Advanced] User Manual 1 Welcome to ASUS WebStorage, your personal cloud space Our function panel will help you better understand ASUS WebStorage services. The
Fiery E100 Color Server. Welcome
Fiery E100 Color Server Welcome 2011 Electronics For Imaging, Inc. The information in this publication is covered under Legal Notices for this product. 45098226 27 June 2011 WELCOME 3 WELCOME This Welcome
Online Monitoring User Guide
High Resolution Temperature Sensing Strip Online Monitoring User Guide 888.637.3282 www.nerdata.com Page 1 of 26 Contents The Aurora Online Monitoring System... 3 1. Creating an Account on the Aurora Online
Premium Design Phone with Smartphone Connect
Smartphone Connect App Guide Premium Design Phone with Smartphone Connect Model No. KX-PRW130 Thank you for purchasing a Panasonic product. This document explains how to use the Smartphone Connect app.
Amcrest 960H DVR Quick Start Guide
Amcrest 960H DVR Quick Start Guide Welcome Thank you for purchasing our Amcrest 960H DVR! This quick start guide will help you become familiar with our DVR in a very short time. Before installation and
Using GIGABYTE Notebook for the First Time
Congratulations on your purchase of the GIGABYTE Notebook. This manual will help you to get started with setting up your notebook. The final product configuration depends on the model at the point of your
Support. Package Contents. D6200 WiFi DSL Modem Router Installation Guide
Support Thank you for selecting NETGEAR products. After installing your device, locate the serial number on the label of your product and use it to register your product at https://my.netgear.com. You
Appearance and Buttons
User Manual Thank you for buying the GoXtreme WiFi action camera. Before start using this product, please read this manual carefully to maximize its performance and extend the life of the product. Appearance
RFID Logger Software User Manual Rev 1.0
RFID Logger Software User Manual Rev 1.0 Table of Contents 1. System Overview... 2 2. Installation... 2 3. USB RFID Interface Box... 3 4. Operation... 4 4.1. Start Up... 4 4.2. Menu overview... 4 4.2.1.
Leaf River Outdoor Products Vibrashine, Inc. P.O. Box 557 Taylorsville, MS 39168 (601) 785-9854 FAX (601) 785-9874 www.vibrashine.
MULTI-CARD PICTURE VIEWER Model PV-2.5 Leaf River Outdoor Products Vibrashine, Inc. P.O. Box 557 Taylorsville, MS 39168 (601) 785-9854 FAX (601) 785-9874 www.vibrashine.com 2 THANKS and WELCOME What does
Telstra Wi-Fi 4GX Advanced III AirCard 810S User Manual
AirCard 810S User Manual July 2015 202-11529-02 350 East Plumeria Drive San Jose, CA 95134 USA Support Contact Telstra at livechat.telstra.com for technical support. Compliance For regulatory compliance
FOR MORE INFORMATION. 125 8880 or from a non-telstra phone 13 2200 and say pre-paid telstra.com/ppmbb visit a telstra store or partner
FOR MORE INFORMATION 125 8880 or from a non-telstra phone 13 2200 and say pre-paid telstra.com/ppmbb visit a telstra store or partner getting to know your telstra pre-paid 4G WI-FI (760S) LET S GET THIS
MAINTENANCE & TROUBLESHOOTING
MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens clean the fan intake filter replace the projection lamp replace the batteries in the remote control use the Kensington lock feature
Nokia Wireless Keyboard (SU-8W) User Guide
Nokia Wireless Keyboard (SU-8W) User Guide DECLARATION OF CONFORMITY We, NOKIA CORPORATION declare under our sole responsibility that the product SU-8W is in conformity with the provisions of the following
A6210 WiFi USB Adapter 802.11ac USB 3.0 Dual Band User Manual
802.11ac USB 3.0 Dual Band User Manual August 2014 202-11373-01 350 East Plumeria Drive San Jose, CA 95134 USA Support Thank you for selecting NETGEAR products. After installing your device, locate the
Thank you for purchasing the Mobile Wi-Fi. This Mobile Wi-Fi brings you a high speed wireless network connection.
Thank you for purchasing the Mobile Wi-Fi. This Mobile Wi-Fi brings you a high speed wireless network connection. This document will help you understand your Mobile Wi-Fi so you can start using it right
Chapter 1 Installing the Gateway
Chapter 1 Installing the Gateway This chapter describes how to set up the wireless voice gateway on your Local Area Network (LAN), connect to the Internet, and perform basic configuration. For information
FreeAgent DockStar Network Adapter User Guide
FreeAgent DockStar Network Adapter User Guide FreeAgent DockStar Network Adapter User Guide 2010 Seagate Technology LLC. All rights reserved. Seagate, Seagate Technology, the Wave logo, and FreeAgent are
IP Camera (L series) User manual 2013-05 V1.1
Dear users, the configuration for this camera is professional, so please read the user manual carefully before using the camera. IP Camera (L series) User manual 2013-05 V1.1 Statement If the user manual