1 A P P L I C A T I O N N O T E Bacterial Detection and Live/Dead Discrimination by Flow Cytometry M I C R O B I A L C Y T O M E T R Y BD FACSCalibur, BD FACSort Flow Cytometers Introduction Accurate determination of live, dead, and total bacteria is important in many microbiology applications. Traditionally, viability in bacteria is synonymous with the ability to form colonies on solid growth medium and to proliferate in liquid nutrient broths. These traditional, culture-based tests are time-consuming and can work poorly with slow-growing or viable, but non-culturable organisms. They do not provide real-time results or timely information that is needed in applications such as industrial manufacturing. Flow cytometry, a technique first applied to eukaryotic cells, has been adapted to the analysis of viability, metabolic state, and antigenic markers of bacteria. 1-4 In particular, flow cytometry can be readily applied to the enumeration of viable bacteria in a sample. Live cells have intact membranes and are impermeable to dyes such as propidium iodide (PI), which only leaks into cells with compromised membranes. Thiazole orange* (TO) is a permeant dye and enters all cells, live and dead, to varying degrees. With gram-negative organisms, depletion of the lipopolysaccharide layer with EDTA greatly facilitates TO uptake. Thus a combination of these two dyes provides a rapid and reliable method for discriminating live and dead bacteria. If enumeration of the bacteria is important, BD Liquid Counting Beads (BD Biosciences, San Jose, CA), a flowcytometry bead standard, can be used to accurately quantify the number of live, dead, and total bacteria in a sample. * US Patent Nos. 4,883,867 and 4,957,870 BD BIOSCIENCES Immunocytometry Systems BD Biosciences Clontech Discovery Labware Immunocytometry Systems Pharmingen
2 Beads R2 Materials Bacteria R1 For cultured bacteria, dilute to an approximate concentration range of 5 x 10 5 to 9 x 10 6 bacteria/ml in staining buffer. Figure 1A E.coli R3 For other sample types, such as pharmaceutical, food or environmental samples, at least 100 organisms per ml are required to be detected using flow cytometry. If necessary, samples can be brought into this range by an initial concentration step. We recommend that you optimize your sample preparation using samples spiked with known concentrations of your target organism(s) before attempting to analyze unknowns. To prepare killed bacteria, mix 0.5 ml of culture before dilution with 0.5 ml of SPOR-KLENZ (Steris Corporation, St. Louis, MO, Catalog No ) disinfectant for 5 minutes. R6 R4 Figure 1B Dead R5 Injured Live Reagents 1. Thiazole orange solution, 17 µm (BD Biosciences Catalog No or ), or equivalent at 8.1 µg/ml (FW 476.6) in DMSO 2. Propidium iodide solution, 1.9 mm (BD Biosciences Catalog No or ), or equivalent, at 1.3 mg/ml (FW 668.4) in water 3. BD Liquid Counting Beads (BD Biosciences Catalog No ), for counting 4. Staining buffer: Phosphate-buffered saline, 1 mm EDTA, 0.2% Pluronic F-68 (BASF Corporation, Mount Olive, NJ, Catalog No ), 0.1% sodium azide, ph 7.4. Tween-20 at 0.01% can be substituted for Pluronic F-68. The staining buffer should be passed through a 0.22-µm filter and used within 2 weeks. Figure 1C Figure 1 Gating strategy for bacterial populations, using a sample of E. coli stained with TO and PI in a staining tube A. FSC vs SSC dot plot with the bacterial population on-scale, with region R1 set liberally around the target population, region R2 set around the beads B. The FL2 vs SSC dot plot with region R3 set around the stained bacteria C. FL1 vs FL3 dot plot gated on (R1 OR R2) AND R3, with regions set around the live, injured, and dead bacterial populations Equipment 1. Disposable 12 x 75-mm capped BD Falcon polystyrene test tubes (BD Labware Catalog No ), or equivalent 2. Vortex mixer 3. Micropipettor with tips (Pipetman, Rainin Instrument Company, Emeryville, CA, or equivalent) 4. BD FACS brand flow cytometer (BD FACSCalibur flow cytometer or equivalent) equipped with 488-nm laser excitation Procedure Staining 1. Label 12 x 75-mm polystyrene tubes. 2. Vortex bacterial suspension or sample and dilute at least 1:10 in staining buffer. 3. Add 200 µl of bacterial suspension, diluted as above in staining buffer. 4. Add 5.0 µl of each dye solution to the tubes. The final staining concentrations are 420 nm for TO and 48 µm for PI. 5. Vortex and incubate for 5 minutes at room temperature. 2
3 6. Reverse pipet 50 µl of BD Liquid Counting Beads into the staining tube if you want to determine the concentration of live, dead and total bacteria. 7. Analyze on a BD FACS brand flow cytometer (BD FACSCalibur flow cytometer or equivalent). 8. Dispose of stained samples and extra dye solution according to local regulations. Flow Cytometer Setup 1. Use BD CaliBRITE 3 beads (BD Biosciences Catalog No ) and the appropriate software, such as BD FACSComp or BD AutoCOMP software, for setting the photomultiplier tube (PMT) voltages and the fluorescent compensations, and for checking instrument sensitivity prior to use. Figure 2 SSC histogram of a sample containing E. coli, with the threshold set at the point where the noise level is just starting to increase (arrow) 2. Initial instrument settings should be as follows: Threshold SSC FSC E01, logarithmic amplification SSC 375 V, logarithmic amplification FL1 600 V, logarithmic amplification FL3 800 V, logarithmic amplification Compensation none used 3. Actual settings can vary with the application and should be optimized as follows: Set threshold on side scatter (SSC), and adjust PMT voltages and threshold levels using an unstained sample of diluted bacteria. The bacterial population should be positioned so that it is entirely on scale on an FSC vs SSC plot (Figure 1A). Individual FSC and SSC histograms should be checked to be sure that the bell-shaped populations are not cut off on the display (Figures 2 and 3). If the entire population is not present, adjust PMT values to position the peak on the histograms and decrease the threshold until the entire population is visible. As the voltage is further increased, the background noise should become evident on the lower end of the histogram. The balance of PMT voltage and threshold should allow the entire peak to be observed with at least a portion of the valley between the bacteria and the noise. Actual peak shapes and resolution from noise will vary with bacterial morphology and sample matrix. Figure 3 FSC histogram of a sample containing E. coli 4. Set FL1 and FL3 PMT voltages to place the unstained population in the lower left quadrant of an FL1 vs FL3 plot (Figure 4). Data Acquisition and Analysis 1. Acquire prepared samples on a BD FACS brand flow cytometer using an SSC threshold. Acquire data with BD CellQuest Pro or BD LYSYS II software, in Acquisition-to-Analysis mode. Set up an FSC vs SSC plot with a live gate around the bacterial population R1 (Figure 1A). If BD Liquid Counting Beads are used, set a region R2 around the beads on the FSC vs SSC plot. Set another region R3 around the stained bacterial population in the FL2 vs SSC dot plot (Figure 1B) and an FL1 vs FL3 plot, gated on combined parameters FSC, SSC, and FL2 (FL1 vs FL3 * US Patent No. 5,232,828 Figure 4 FL1 vs FL3 dot plot of unstained E. coli gated on the bacterial region in FSC vs SSC, with the unstained bacteria placed in the first decade on FL1 and FL3 3
4 gated on [R1 OR R2] AND R3) to display the stain results (Figure 1C). 2. Acquire a total of 10,000 events. 3. In Analysis mode, draw rectilinear regions around the live, dead, and injured populations (Figure 1C). 4. If you used BD Liquid Counting Beads and want to determine the absolute count, use the following equation: # of events in region containing cell population # of events in bead population # of beads per test* X test volume X dilution factor = Concentration of bacterial population Controls Use an unstained bacterial sample to confirm that PMT voltages are set appropriately (Figure 4). Dilute, stain, and acquire an aliquot of culture media or sample matrix, diluted the same as a bacterial sample, to confirm that assay background is low. Use a mixture of live and killed bacteria to confirm that stained live, injured, and dead bacterial populations are sufficiently resolved. Results and Discussion Live/Dead Discrimination of Bacteria The TO and PI dye combination provides significant resolution between live and dead cell populations (Figure 1C). An intermediate or injured population can often be observed between the live and dead populations. These designations were confirmed by index sorting. Only live and injured populations produced colonies on plates (data not shown). Figure 5 shows staining patterns for five different bacteria. Dead cells were produced using SPOR-KLENZ treatment. Approximately equal amounts of live and dead cells were mixed in the samples. Stain results are shown for Escherichia coli (ATCC Catalog No ), Bacillus globigii (ATCC Catalog No ), Salmonella choleraesuis (ATCC Catalog No ), Pseudomonas aeruginosa (ATCC Catalog No ), and Staphylococcus epidermidis (ATCC Catalog No ). Resulting populations are similar, but there are organism-specific differences. In particular, two of the gram negative bacteria, E. coli and S. cholerasuis, demonstrated less efficient uptake of TO than the other bacteria. The live and dead populations do not line up in a rectilinear fashion and analysis is best accomplished by drawing regions in the appropriate areas. Comparison of Flow vs Plate count Figure 6 shows the relationship between the concentration of viable B. globigii determined by flow cytometry and plate counts of the same bacterium. A serial dilution was made on a B. globigii culture grown in BD Trypticase Soy Broth (BD Diagnostic Systems, Sparks, MD, Catalog No ) and the suspensions were analyzed either by flow cytometry using TO, PI, and BD Liquid Counting Beads or by plating with appropriate dilutions on Trypticase Soy Agar (BD Diagnostic Systems Catalog No ). When concentrations were corrected for dilution and plotted, flow cytometric results were linear with plate count results down to the range of 0.1 to 1 bacterium per microliter. Sensitivity of detection by flow cytometry will vary with the noise present in a sample matrix and should be evaluated for each experimental situation. * This value is found on the BD Liquid Counting Beads label and might vary from lot to lot. 4
5 Figure 5 Representative FL1 vs FL3 dot plots of several bacteria stained with TO Bacillus globigii 47.7% Dead Pseudomonas aeruginosa 31.1% and PI, gated on a combined parameters FSC, SSC, and FL2 (( R1 OR R2) AND R3), 3.1% Injured 1.0% with samples containing about equal amounts of live and killed bacteria 49.2% Live 67.9% Stphylococcus epidermidis Salmonella choleraesuis Escherichia coli 34.3% 42.0% 31.3% 1.9% 5.4% 2.6% 63.8% 52.6% 66.1% Hints TO fluoresces primarily in FL1 and FL2; PI fluoresces primarily in FL3. Therefore, the best discrimination of live and dead populations is on an FL1 vs FL3 plot. This method can be applied with a variety of buffer systems, but optimal resolution requires some surfactant to be present in the staining buffer. There will be differences between bacteria in their abilities to take up TO and PI. The LPS on gram negative bacteria can interfere with the uptake of TO and other permeant dyes. Interference can be largely overcome by inclusion of 1 mm EDTA in the staining buffer, which has been reported to remove the LPS from the bacteria. In practice, staining protocols must be adjusted to the bacteria being analyzed. Thiazole orange is hydrophobic. Stock solutions should be maintained in DMSO or alcohol. TO concentration will decrease over time in aqueous solution due to adsorption to surfaces. Figure 6 Correlation between the concentration of B. globigii determined by flow cytometry and by plating on BD Trypticase Soy Agar TO staining is adequate for analysis at 2 to 5 minutes but requires at least 15 minutes to achieve maximum intensity. PI stains very quickly, while TO enters the cells more slowly. 5
6 BD Biosciences publishes this method as a service to investigators. Detailed support for non flow cytometric aspects of this procedure might not be available from BD Biosciences. Setting FSC and SSC on logarithmic amplification assures that a wide range of bacterial sizes can appear on-scale and helps present recognizable populations for gating. An event rate of 1000 events per second minimizes the chance of coincidence and improves population resolution. High event rates can be corrected either by dilution or by decreasing the instrument flow rate. At least 1000 bead events should be collected to provide reliable concentration data. If the population of interest cannot be adequately resolved using an SSC threshold alone, a secondary threshold on FL1 can be used. An FL1 threshold alone might not be adequate due to the large amount of small fluorescent debris that can be present in a stained bacterial sample. As the concentration of bacteria decreases, background noise will become progressively more prominent. If high background counts are observed on an instrument, the staining buffer and sheath fluid should be checked for particles. An instrument cleaning cycle and a drain/fill cycle can also reduce noise. Samples can be checked by fluorescence microscopy to confirm that target organisms are stained. References 1. Davey HM, Kell DB. Flow cytometry and cell sorting of heterogeneous microbial populations: the importance of single cell analyses. Microbiological Reviews. 1996;60: Nebe-von-Caron G, Stephens PJ, Hewitt CJ, Powell JR, Badley RA. Analysis of bacterial function by multi-colour fluorescence flow cytometry and single cell sorting. Journal of Microbiological Methods. 2000;42: Shapiro HN. Microbial analysis at the single-cell level: tasks and techniques. Journal of Microbiological Methods. 2000;42: Nebe-von-Caron G, Stephens PJ, Badley RA. Bacterial detection and differentiation by cytometry and fluorescent probes. Proceedings of the Royal Microbiological Society. 1999;34: Contributors Rana Alsharif, William Godfrey BD Biosciences, San Jose, CA BD Biosciences Asia Pacific Tel (65) Fax (65) Brazil Tel (55) Fax (55) Canada Tel (888) Fax (905) For Research Use Only. Not for use in diagnostic procedures. BD, BD Falcon, BD Trypticase, BD FACS, BD FACSCalibur, BD FACSComp, BD AutoCOMP, BD LYSYS, and BD CellQuest software are trademarks of Becton, Dickinson and Company. Pipetman is a registered trademark of Rainin Instrument Company, Inc. SPOR-KLENZ is a registered trademark of Steris Corporation. Pluronic is a registered trademark of BASF Corporation Becton, Dickinson and Company Clontech Discovery Labware Immunocytometry Systems PharMingen Europe Tel (32) Fax (32) Japan Tel Mexico Tel (52) Fax (52) United States Tel (877) Fax (408) recycled FM /2002 BD Biosciences Clontech Discovery Labware Immunocytometry Systems Pharmingen
Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction
Application Note 10 Measurement of Cell Recovery After Sorting with a Catcher-Tube-Based Cell Sorter Introduction In many experiments using sorted cells, it is important to be able to count the number
123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.
Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.
No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer
APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes
APPLICATION INFORMATION
DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
BD FACSComp Software Tutorial
BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS
Islet Viability Assessment by Single Cell Flow Cytometry
Islet Viability Assessment by Single Cell Flow Cytometry Page 1 of 8 Purpose: To comprehensively assess the viability of the islet cell preparation prior to transplantation. Tissue Samples: A sample containing
Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer
Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
ArC Amine Reactive Compensation Bead Kit
ArC Amine Reactive Compensation Bead Kit Catalog no. A1346 Table 1. Contents and storage information. Material Amount Composition Storage Stability ArC reactive beads (Component A) ArC negative beads (Component
BD Trucount Tubes IVD
02/2015 23-3483-07 IVD BD Trucount Tubes For determining absolute counts of leucocytes in blood 25 Tubes Catalog No. 340334 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company.
Flow Cytometry Analysis of the Activity of Disinfecting Agents Tested with Staphylococcus aureus and Escherichia coli
Polish Journal of Microbiology 2005, Vol. 54, No 1, 21 26 Flow Cytometry Analysis of the Activity of Disinfecting Agents Tested with Staphylococcus aureus and Escherichia coli A. WO NIAK-KOSEK and J. KAWIAK*
Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T
Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep
COMPENSATION MIT Flow Cytometry Core Facility
COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,
Introduction to Flow Cytometry
Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
EdU Flow Cytometry Kit. User Manual
User Manual Ordering information: (for detailed kit content see Table 2) EdU Flow Cytometry Kits for 50 assays: Product number EdU Used fluorescent dye BCK-FC488-50 10 mg 6-FAM Azide BCK-FC555-50 10 mg
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service
Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?
Introduction to flow cytometry
Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression
Flow cytometric Annexin V/Propidium Iodide measurement in A549 cells
Project: VIGO Flow cytometric Annexin V/Propidium Iodide measurement in A549 cells Detecting apoptosis and necrosis in A549 cells AUTHORED BY: DATE: Cordula Hirsch 20.01.2014 REVIEWED BY: DATE: Harald
Cell Cycle Tutorial. Contents
Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6
Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility
10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa
CyAn : 11 Parameter Desktop Flow Cytometer
CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on
BACTERIAL ENUMERATION
BACTERIAL ENUMERATION In the study of microbiology, there are numerous occasions when it is necessary to either estimate or determine the number of bacterial cells in a broth culture or liquid medium.
Immunophenotyping peripheral blood cells
IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune
SmartFlare RNA Detection Probes: Principles, protocols and troubleshooting
Technical Guide SmartFlare RNA Detection Probes: Principles, protocols and troubleshooting Principles of SmartFlare technology RNA detection traditionally requires transfection, laborious sample prep,
Absolute counting using flow cytometry. Bright Count Microspheres
Absolute counting using flow cytometry Single-platform system for absolute counts I PRODUCT MANUAL [REF] IQP-391 [REF] IQP-392 s 50 tests s 100 tests [IVD] In Vitro Diagnostic medical device IQ Products
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done
Cell Cycle Phase Determination Kit
Cell Cycle Phase Determination Kit Item No. 10009349 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 3 Safety
Test Method for the Continuous Reduction of Bacterial Contamination on Copper Alloy Surfaces
Test Method for the Continuous Reduction of Bacterial Contamination on Copper Alloy Surfaces Test Organisms: Staphylococcus aureus (ATCC 6538) Enterobacter aerogenes (ATCC 13048) Pseudomonas aeruginosa
Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit
2 Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit Presented by Ellen Meinelt, MS MLS(ASCP) CM, Technical Applications Specialist and Calin Yuan, Product Course Developer,
Annexin V-EGFP Apoptosis Detection Kit
ab14153 Annexin V-EGFP Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of apoptosis in various samples This product is for research use only and is not intended
Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer
Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer Purpose This procedure gives instruction on minimal residual disease (MRD) detection in patients with
BD FC Beads 8-Color Kit for BD OneFlow Assays
BD FC Beads 8-Color Kit for BD OneFlow Assays 7/2014 23-15785-00 IVD 5 tests per kit Catalog No. 658621 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2014 BD Becton,
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline
Analyzer Experiment setup guide for LSRII and Canto s
Analyzer Experiment setup guide for LSRII and Canto s 1. Check the instrument configuration on our website to determine the most appropriate cytometer for your experiment, https://depts.washington.edu/flowlab/instrumentation.html,
Becton-Dickinson FACSCalibur Flow Cytometer Instructions
Start-up 1. Turn on flow cytometer by pushing green button on right side of instrument, toward the back. Give it 30 minutes to warm up before collecting data. 2. Turn on monitor. 3. Turn on computer. If
PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.
PROTOCOL Immunostaining for Flow Cytometry 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 Rev.0 Background The combination of single cell analysis using flow cytometry and the specificity of antibody-based
How to operate the BD FACSCanto flow cytometer
How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.
Introduction to Flow Cytometry
Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: [email protected] BDBiosciences Application Examples
CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles
CyFlow SL Microbiology Detection and Analysis of Microorganisms and Small Particles 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS
FOR IN-VITRO DIAGNOSTIC USE INVITROGEN CAL-LYSE TM Lysing Solution WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS CAL-LYSE TM CATALOG No. GAS-010 250 tests 25 ml CATALOG No. GAS-010S-100
Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806)
Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-86) The Hoechst purification was established for murine hematopoietic stem cells (HSC)
FACS-CALIBUR. USER-Guide
FACS Calibur User guide Version 2.1-03/2006 AXS 1 FACS-CALIBUR Version 2.1 USER-Guide Note: This guide does note compensate for a personal introduction from an operator before first usage Draft mainly
Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835
Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that
BD StemFlow Human Pluripotent Stem Cell Sorting and Analysis Kit
BD StemFlow Human Pluripotent Stem Cell Sorting and Analysis Kit Instruction Manual Catalog No. 560461 ii Human Pluripotent Stem Cell Sorting and Analysis Kit 2009, Becton, Dickinson and Company. All rights
Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and
Using BD FACSDiva CST To Evaluate Cytometer Performance, Create Custom Assay Settings and Implement Cross-Instrument and Cross-Site Standardization of Assays PART 2 Alan M. Stall Director, Advanced Cytometry
LIVE/DEAD Fixable Dead Cell Stain Kits
USER GUIDE LIVE/DEAD Fixable Dead Cell Stain Kits Pub. No. MAN0002416 (MP34955) Rev. A.0 Table 1. Contents and storage Material Amount Storage Stability Individual Kits: Blue, violet, aqua, yellow-, green,
Annexin V-FITC Apoptosis Detection Kit
ab14085 Annexin V-FITC Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of Apoptosis in living cells (adherent and suspension). This product is for research
Below is a list of things you should be aware of before you schedule your sort.
Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will
protocol handbook 3D cell culture mimsys G hydrogel
handbook 3D cell culture mimsys G hydrogel supporting real discovery handbook Index 01 Cell encapsulation in hydrogels 02 Cell viability by MTS assay 03 Live/Dead assay to assess cell viability 04 Fluorescent
Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers
An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers BD Biosciences, San Jose, CA Contents 2 Introduction to compensation 5 Tips for setting up compensation 10 Tools for easy,
Standard Operating Procedure
1.0 Purpose: 1.1 The characterisation of of main leukocyte subsets in peripheral blood cells from mice by flow cytometry. Reliable values of frequencies of leukocyte clusters are very much dependent on
Introduction to Flow Cytometry:
Introduction to Flow Cytometry: A Learning Guide Manual Part Number: 11-11032-01 April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 1-800-448-2347 Introduction to Flow Cytometry: A Learning
Annexin V-FITC Apoptosis Detection Kit
ab14085 Annexin V-FITC Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of Apoptosis in living cells (adherent and suspension). This product is for research
Flow Cytometry. flow cytometer DNA apoptosis ph
Flow Cytometry flow cytometer DNA apoptosis ph flow cytometry flow cytometer (a) (b) cell counting instrument (c) 1960 ink-jet technology 17 19 1940 1950 fluorescence microscopy fluorescent dye DNA polyclonal
LAB 4. Cultivation of Bacteria INTRODUCTION
LAB 4. Cultivation of Bacteria Protocols for use of cultivation of bacteria, use of general growth, enriched, selective and differential media, plate pouring, determination of temperature range for growth
Science is hard. Flow cytometry should be easy.
Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening
Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1
78 Irma Lee Circle, Unit 0, Lake Forest, Illinois 00 SPHERO TM Technical Note STN- Rev B 008 DETERMINING PMT LINEARITY IN FLOW CYTOMETERS USING THE SPHERO TM PMT QUALITY CONTROL EXCEL TEMPLATE Introduction
UCHC Guide to FACS DiVa. LSR II instruments
UCHC Guide to FACS DiVa LSR II instruments Start Up: Turn on the instrument using the Green power button. Fluidics pump power, LSR II-B & C LSR II-B and LSRII-C-> Turn on the fluidics pump and ensure that
INSIDE THE BLACK BOX
FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben
Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...
Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and
Notes on titering antibodies
Home Events This Week Bulletin Boards Lab Personnel Lab Policies Protocols MolBio Tetramers Proteins Flow Cytometry Immunoassays Instruments Computers Lab Publications Scientiific Literature Science on
Enumeration and Viability of Nucleated Cells from Bone Marrow, Cord Blood, and Mobilized Peripheral Blood
Application Note Primary Cell Analysis Enumeration and Viability of Nucleated Cells from Bone Marrow, Cord Blood, and Mobilized Peripheral Blood Cell Viability Introduction: Bone Marrow, Cord Blood, Stem
Microbiology BIOL 275 DILUTIONS
DILUTIONS Occasionally a solution is too concentrated to be used as is. For example, when one is performing manual blood counts, the blood contains too many cells to be counted as such. Or when performing
Western Blot Analysis with Cell Samples Grown in Channel-µ-Slides
Western Blot Analysis with Cell Samples Grown in Channel-µ-Slides Polyacrylamide gel electrophoresis (PAGE) and subsequent analyses are common tools in biochemistry and molecular biology. This Application
MTT Cell Proliferation Assay
ATCC 30-1010K Store at 4 C This product is intended for laboratory research purposes only. It is not intended for use in humans, animals or for diagnostics. Introduction Measurement of cell viability and
Fighting the Battles: Conducting a Clinical Assay
Fighting the Battles: Conducting a Clinical Assay 6 Vocabulary: In Vitro: studies in biology that are conducted using components of an organism that have been isolated from their usual biological surroundings
use. 3,5 Apoptosis Viability Titration Assays
Introduction Monitoring cell viability can provide critical insights into the effectiveness of biological protocols and the stability of biological systems including: the efficacy of cell harvesting procedures,
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor
Evaluation of Microbial Growth and Survival on Construction materials treated with Anabec NewBuild 30
Evaluation of Microbial Growth and Survival on Construction materials treated with Anabec NewBuild 30 Absar Alum, Ph.D. Department of Civil and Environmental Engineering Arizona State University Tempe,
FACSCanto RUO Special Order QUICK REFERENCE GUIDE
FACSCanto RUO Special Order QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW
GETTING STARTED. 6. Click the New Specimen icon.
BD FACSDiva Software Basic Experiment Guide Please note, this guide is not intended to take the place of in-person training but is to be used as a reference as needed after completing training with Unified
Detecting Intracellular Cytokines in Activated Monocytes
BD BIOSCIENCES LIFE SCIENCE RESEARCH I M M U N E F U N C T I O N FastImmune Cytokine System A P P L I C A T I O N N O T E Detecting Intracellular Cytokines in Activated Monocytes Scope Originating in bone
Institut für Experimentelle Immunologie und Bildgebung
Page 1 von 6 Starting Up the System Start up the computer, start Windows and log in as IMCES_user start BD FACSDiva software with your Groupname Turn on the cytometer main power Wait a moment and Check
Quantifying Bacterial Concentration using a Calibrated Growth Curve
BTEC 4200 Lab 2. Quantifying Bacterial Concentration using a Calibrated Growth Curve Background and References Bacterial concentration can be measured by several methods, all of which you have studied
1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8
Table of Contents 1. Instrument Layout pg 2 2. Summit Overview pg 3 3. Data Storage and Data Files pgs 6-7 4. Saving Data pg 7 5. Start Up pg 8 6. Choosing and Naming Parameters pg 9 7. Creating Histograms
UCHC Guide to CellQuest Pro. FACS Calibur instruments
UCHC Guide to CellQuest Pro FACS Calibur instruments Startup: Turn on Cytometer using green switch on the side of the cytometer Cytometer Power Switch Within a minute of turning on the cytometer; turn
Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose
Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities:
Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract
arch 212 Standardizing Application Setup Across ultiple Flow Cytometers Using BD FACSDiva Version 6 Ellen einelt, ervi Reunanen, ark Edinger, aria Jaimes, Alan Stall, Dennis Sasaki, Joe Trotter Contents
Application Note. Cell Cycle and DNA Content Analysis Using the BD Cycletest Assay on the BD FACSVerse System. Summary
Cell Cycle and DNA Content Analysis Using the BD Cycletest Assay on the BD FACSVerse System Yibing Wang, Catherine McIntyre, and Dev Mittar BD Biosciences, San Jose Contents 1 Summary 2 Introduction 3
ENUMERATION OF MICROORGANISMS. To learn the different techniques used to count the number of microorganisms in a sample.
ENUMERATION OF MICROORGANISMS I. OBJECTIVES To learn the different techniques used to count the number of microorganisms in a sample. To be able to differentiate between different enumeration techniques
Transformation Protocol
To make Glycerol Stocks of Plasmids ** To be done in the hood and use RNase/DNase free tips** 1. In a 10 ml sterile tube add 3 ml autoclaved LB broth and 1.5 ul antibiotic (@ 100 ug/ul) or 3 ul antibiotic
FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences
FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data Monisha Sundarrajan, PhD BD Biosciences 23 13795 00 Agenda Introduction Overview of BD CBA bead based immunoassays FCAP
These particles have something in common
These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative
BD FACSCalibur. The flow cytometer for your routine cell analysis needs
BD FACSCalibur The flow cytometer for your routine cell analysis needs A system with a rich application basis and a modular approach that continues to meet evolving needs of cell analysis worldwide. The
After completing this module, you will be able to: Describe the BD FACScan system and how it works.
BD FACScan System After completing this module, you will be able to: Describe the BD FACScan system and how it works. Perform instrument startup and shutdown procedures. Describe instrument maintenance
LSR II Users Guide Stanford Shared FACS Facility 06/10/13
LSR II Users Guide Stanford Shared FACS Facility 06/10/13 A configuration sheet is located on the Website in the Instrument Overview area Do I need to run beads when I collect data using Diva Digital?
FITC Annexin V/Dead Cell Apoptosis Kit with FITC annexin V and PI, for Flow Cytometry
FITC Annexin V/Dead Cell Apoptosis Kit with FITC annexin V and PI, for Flow Cytometry Catalog no. V13242 Table 1. Contents and storage information. Material Amount Composition Storage* Stability FITC annexin
FACSCalibur System User s Guide
11-10823-02 Rev. A FACSCalibur System User s Guide 02-61760-02 August, 1996 Becton Dickinson Immunocytometry Systems 2350 Qume Drive San Jose, CA 95131-1807 Ordering Information (800) 223-8226 Customer
MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:
INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals
Cell Counting Using the BD Accuri C6 Flow Cytometer
Cell Counting Using the BD Accuri C6 Flow Cytometer Koen Verbrugghe, PhD Marketing Applications Specialist BD Biosciences 23-14763- The BD Accuri C6 Flow Cytometer System An affordable, full-featured,
Lab 10: Bacterial Transformation, part 2, DNA plasmid preps, Determining DNA Concentration and Purity
Lab 10: Bacterial Transformation, part 2, DNA plasmid preps, Determining DNA Concentration and Purity Today you analyze the results of your bacterial transformation from last week and determine the efficiency
CFSE Cell Division Assay Kit
CFSE Cell Division Assay Kit Item No. 10009853 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 4 Precautions
RPCI 004 v.002 Staining Procedure For all Directly Conjugated Reagents (Whole Blood Method)
Immune Tolerance Network RPCI 004 v.002 Staining Procedure For all Directly Conjugated Reagents (Whole Blood Method) Author: Paul Wallace, Director, RPCI Laboratory of Flow Cytometry Approved by: Paul
Boundary-breaking acoustic focusing cytometry
Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory