International Comparison of Surface Roughness and Step Height (Depth) Standards, SIML-S2 (SIM 4.8)
|
|
|
- Warren Casey
- 10 years ago
- Views:
Transcription
1 International Comparison of Surface Roughness and Step Height (Depth) Standards, SIML-S2 (SIM 4.8) K. Doytchinov, Institute for National Measurement Standards, National Research Council (NRC), Ottawa, Canada F. Kornblit, Instituto Nacional de Technologia Industrial (INTI), Buenos Aires, Argentina C. Colin Castellanos, Centro Nacional de Metrologia (CENAM), Queretaro, Mexico J. C. V. Oliveira, National Institute of Metrology Standardization and Industrial Quality (INMETRO), Duque de Cacias, Brazil T.B. Renegar and T.V. Vorburger, National Institute of Standards and Technology (NIST), Gaithersburg, USA Abstract: Calibration services of five countries from the Sistema Interamericano de Metrología (SIM) region are compared through measurements of surface roughness and step height standards. A surface roughness standard with a nominal roughness average (Ra) value of 0.2 µm, a surface roughness standard with a nominal Ra value of 3 µm and a nominal spatial wavelength of 99 µm, and three step height standards with nominal values of 2.55 µm, 0.38 µm and 0.03 µm are compared. Special attention is paid to the influence of the long wavelength cut-off ratio of the measurements. Results are reported for Ra, maximum height of profile Rz, mean width of profile elements RSm, and step height d, depending on the sample measured. The initial reported results show that the laboratories agree on all of the measurements within their stated and published uncertainties. Observations are then discussed about the definition of Rz, the effect of instrument noise on Rz, the different step height parameters d and Pt, differences between the laboratories in reporting Type A statistical uncertainties, the method for calculating the uncertainty of the reference value, and the importance of accounting for correlations between the reference value and individual lab values when calculating the degrees of equivalence. After corrections and reanalysis the laboratories still agree well considering their stated uncertainties. 1. Introduction Under an international Mutual Recognition Arrangement (MRA) [1], the metrological equivalence of national measurement standards and of calibration certificates issued by national metrology institutes is established by a set of key comparisons chosen and organized by the Consultative Committees of the Committee International des Poids et Mesures (CIPM) and regional comparisons by the regional metrology organizations in collaboration with the Consultative Committees. Five national measurement institutes (s) from the SIM (Sistema Interamericano de Metrología) region have carried out a surface roughness regional comparison, with the National Research Council (NRC) as the pilot laboratory. Participating laboratories as shown above were: NRC, NIST, INMETRO, INTI, and CENAM. Results of this international comparison are intended for inclusion in the agreement for establishing the metrological equivalence between the s. The measurements were carried out according to the schedule below: Laboratory Country Date NRC Canada October 2000 NIST USA December 2000 INMETRO Brazil February 2001 INTI Argentina March 2001 CENAM Mexico April 2001 Participants used the equipment and procedures normally used when calibrating customer artifacts. 1
2 2. Description of the Standards The following physical standards were used in the study: Step height specimen, Model 112/964, serial number manufactured by Taylor Hobson Pneumo Ltd. * (UK) with three patches with 2.55 µm, 0.38 µm and µm nominal step height, respectively (shown in Fig. 1). Figure 1: Step height specimen, serial number Surface roughness specimen, serial number Special 3144, having nominal 0.20 µm roughness average (Ra), manufactured by Taylor Hobson Pneumo Ltd. (UK), shown in Fig. 2. Both standards belong to NRC and were supplied in special cases as shown in Fig. 3 below. Figure 2: Surface roughness specimen serial number Special 3144 * Certain equipment and materials are described in this article in order to specify clearly the experimental procedures used. In no case does this imply a recommendation by NRC, INTI, CENAM, INMETRO, or NIST that the equipment or materials are necessarily the best available for the purpose. 2
3 Figure 3: NRC step height and roughness specimens in their cases NIST Standard Reference Material (SRM) 2073a Prototype, Serial No (Fig. 4), calibrated for Ra and surface spatial wavelength D, a parameter nearly identical to the standard parameter RSm, the mean width of profile elements. The standard is intended for calibration checks of stylus instruments used to measure both Ra and RSm. The SRM is a steel block of nominal Knoop hardness 500, which has been coated by the electroless nickel deposition process. A sinusoidal roughness profile has been machined onto the top surface of the specimen. The surface profile is highly sinusoidal as shown in Fig. 5. This standard belongs to NIST. Figure 4: Standard Reference Material (SRM) 2073a Prototype, Serial No Figure 5: Representative surface profile trace of SRM 2073a. Dimensions are approximate. 3
4 3. Measuring equipment. The instruments used in each laboratory to measure the surfaces are described below. Organization Description Traceability NRC Form Talysurf 120L with a laser pick-up. Nominal stylus tip radius 2 µm Calibration ball, 22 mm radius. Special laser interferometer set-up for generating step heights. Both traceable to SI unit of length NIST Roughness: Federal Surfanalyzer 2000 with LVDT stylus pickup and nominal 2 µm tip. Laser interferometer system for calibration of X-displacement. Step height: Taylor Hobson Talystep with LVDT stylus pickup and nominal 2 µm tip. through NRC. Z-Calibration: Both instruments calibrated by measurement of an interferometrically calibrated step height standard, of similar height to the Z-scale being used. Step standards traceable to the SI unit of length. X-Calibration (Federal 2000): Displacement measured directly with HeNe Laser Interferometer. INMETRO INTI CENAM PERTHOMETER "S8P" (MAHR/PERTHEN) with the "PRK" reference unit and a "pick-up" type "RFHTB- 50" with a nominal stylus radius of 5 µm. Federal 5000 Surfanalyzer, different stylus tips with nominal radii between 2 µm and 5 µm. Form Talysurf 120 mm series 2 with inductive pick-up. Nominal stylus tip radius - 2 µm. Standards traceable to SI unit of length through PTB Height standard, H=5,76 µm, traceable to SI unit of length through PTB Calibration ball, 12,5 mm radius traceable to SI unit of length through CENAM 4. Measurement Conditions Although the participants were asked to use the equipment and procedures normally used when calibrating clients standards, it was agreed that the cut-off filter for the Ra measurements should be the ISO Gaussian type [2] for which the specifications are the long-wavelength cut-off λc, the short-wavelength cut-off λs, and the resulting cut-off ratio between them λc/λs. The participants agreed to use a λc of 0.8 mm for which the default λs value is 2.5 µm, resulting in a cut-off ratio of 300:1, as specified in the ISO standard [3]. However, many companies in industry use instruments with a roughness cut-off ratio of 100:1. These companies are using the services of the calibration labs to get their standards calibrated. If a calibration standard is calibrated using a cut-off ratio of 300:1 and then is used to calibrate an instrument with a 100:1 capability, there is a danger of a bias being introduced in the traceability chain because roughness measured with a 300:1 cut-off ratio should be larger than roughness measured with a 100:1 cutoff ratio. That is because roughness is basically a noise property and its measured value depends on the bandwidth of the measuring instrument. To estimate the magnitude of the problem, the pilot laboratory specifically asked the participants to evaluate the same raw roughness data using both 100:1 and 300:1 roughness cut-off ratio. The goal was to show if the differences in the results would be significant. Measurement of Ra and Rz (ISO) on the Roughness Specimen, Serial number SPECIAL
5 NRC, INTI, CENAM, and INMETRO performed 20 measurements on this roughness patch according to the pattern of Fig. 6. NIST followed its customary calibration procedure of nine measurements as shown schematically in Fig. 7. All labs used a sufficient traversing length so that after filtering, the evaluation length amounted to five cut-offs. The cut-off filter was the ISO Gaussian type, discussed in ISO Standard [2] and ASME B46.1 [4] with a cut-off length of 0.8 mm. CENAM and NRC evaluated their raw data using both 100:1 and 300:1 cut-off ratios. INTI and NIST evaluated the data using the 300:1 cut-off ratio only. The measuring equipment of INMETRO did not have a λs filter built in. NRC, CENAM, NIST, and INTI used styli with a nominal tip radius of 2 µm while INMETRO used a stylus with a nominal tip radius of 5 µm. Because the tip radius acts as a mechanical filter that attenuates small wavelengths of the surface, the participants decided to put the INMETRO results in the 100:1 cut-off ratio category. Figure 6: Schematic diagram of 20 positions used by NRC, INTI, INMETRO, and CENAM for measurement of Ra and Rz (ISO) on the Roughness Specimen, Serial number SPECIAL 3144 Both the roughness average Ra and the maximum height of profile Rz were calculated from the profile data. The calculation of Rz can be ambiguous. Rz is presently defined in ISO Standard 4287 (1997) [5] as the sum of the largest profile peak height Zp and the largest profile valley depth Zv within a sampling length. The method of calculation is illustrated in Fig. 8. When this Rz (ISO-1997) is evaluated by averaging over five sampling lengths as was the case here, the result is identical to the earlier parameter Rz (DIN) [6]. Therefore, software for Rz(DIN) may be used to calculate Rz(ISO-1997). The value for both is larger than that derived from a previous parameter Rz(ISO ), now withdrawn, which was defined as the average of the heights of the five highest peaks and depths of the five lowest valleys within the sampling length. Measurement of Ra and RSm on the 3 µm Roughness Specimen SRM 2073a The scans were performed according to Fig. 7. A total of nine measurements were performed on this roughness patch using a sufficient traversing length so that after filtering, the evaluation length amounts to 5 cut-offs. The cut-off filter was the ISO Gaussian type, discussed in ISO Standard [2] and ASME B46.1 [4] with a cut-off length of 0.8 mm. CENAM and NRC evaluated the same raw data using both 100:1 and 300:1 cut-off ratios. INTI and NIST evaluated the data using a 300:1 cut-off ratio only. The styli were the same as those used for the SPECIAL 3144 specimen. It is known that different manufacturers have different approaches when calculating the RSm values. Some algorithms count all of the peaks while others count only the peaks above a pre-defined threshold. The participants were advised to reduce these thresholds to a minimum if the software allowed. 5
6 Figure 7: Measurement positions for a prototype of NIST roughness specimen SRM 2073a. All dimensions are in mm. A similar pattern with smaller dimensions was used by NIST for the SPECIAL 3144 specimen. Figure 8: Schematic diagram, reprinted from ISO 4287, Ref. 5, of the method of calculation of the maximum height of profile parameter Rz Avoidance of flaws SRM 2073a has been manufactured with very few visible flaws. The participants were advised to avoid measuring paths on SRM 2073a that intersect flaws visible to the naked eye. In addition and more specifically, even single flaws in surface profiles, such as those shown in Fig. 9, should be avoided because they can change the measured RSm value significantly (by 2.5 % if the surface profile contains 40 peaks and valleys). That is because the customary algorithms for calculating RSm rely on counting the crossings of the mean line (shown dashed here) by the surface profile. Single flaws such as these do not change the Ra values as significantly as they can change RSm. 6
7 Figure 9: Avoidance of flaws in surface measurements Measurement of the step heights on the step height specimen, serial number The step height samples contain three grooves. Only the depth of the middle one was measured. The scans were performed in a direction perpendicular to the groove. The scanned data were to be levelled using a least squares (LS) best-fit line on the two portions marked A and B in Fig. 10 below. The step height was defined in accordance with ISO standard [7] as the average depth (d) from the best-fitted LS line to the middle third of the bottom of the groove (C in Fig. 10). A total of 9 scans over the central 2 mm of each step were performed. NRC, INTI, CENAM, and NIST calculated step heights in this way. However, INMETRO calculated the peak-valley height Pt [5] from the measured step profiles. 5. Initial Results and Observations Figure 10: Assessment of the step height samples (Ref. 7) Results for Ra, Rz, RSm, d, and Pt obtained by each laboratory are shown in Figs along with their estimated uncertainties (k=2) calculated according to the GUM [8]. The sources of uncertainty include both Type A and Type B components. The Type A components were calculated from statistical methods. The Type B components were calculated by non-statistical methods such as estimates based on physical models. The principal source of type A uncertainty is variation in results measured at different positions across the specimen. Other components of uncertainty may be calculated by Type A or Type B methods depending on the data or models used to derive the estimates. For the z-parameters, Ra, Rz, and d, major sources of uncertainty include 7
8 The form error and surface finish of a master artifact, such as a step height or sphere, used to calibrate the z-scale of the stylus instrument, Instrumental variations during measurements taken to calibrate the z-scale of the instrument. These can arise from instrument noise occurring during the measurement of a master calibration artifact and lead to variations in the z-scale of the instrument from day-to-day, Uncertainty in the calibrated value, such as step height or radius, of the master artifact, Uncertainty in the horizontal resolution of the instrument, due primarily to uncertainty in the short-wavelength cutoff, Instrumental noise, which tends to increase measured roughness values. Uncertainty due to nonlinearity of the instrument transducer, Uncertainty (for measurement of d) due to the mechanical or electrical ringing that occurs when the stylus encounters the trailing edge of a step. In addition to the Type A variation of measured values, Type B sources of uncertainty for RSm include Uncertainty in the calibration of the x-axis displacement. Variation due to the choice of algorithm to calculate the spacings of profile irregularities. According to a protocol agreed upon by the five laboratories before the measurements began, the reference value (x ref ) for each measured quantity was the arithmetic mean of the values x i from the individual s, given by x = x = (1/ n), (1) ref and the standard uncertainty u(x ref ) of the reference value was taken as the standard deviation of the mean of the set of reported x i values, that is, i x i u ( x ) ref = ( xi x ref ) n( n 1) 2. (2) In order to evaluate the degree of equivalence between each and the mean x ref, the E n statistic was used. It was calculated as E n x x i ref = (3) ui + u ( xref, ) where u i is the standard uncertainty for the measured value x i. The formula of Eqn. 3 for degree of equivalence does not account for the correlation between x ref and x i.. This correlation is especially significant because of the small number of contributions, four at most, for each average. Therefore, although permitted by the MRA, the above formula for equivalence is not strictly correct. This subject is addressed later in Section 6, when we describe the reanalysis of the results. Another point is that the method of Eqn. 2 for calculating the reference value uncertainty emphasizes the consistency between the results of the different laboratories, rather than the overall uncertainty of the results. The same bias could be present in all of the results for a particular parameter, and the uncertainty of the reference value for that parameter as calculated here would not account for that bias. Such Type B sources of bias should be accounted for in the quoted uncertainties u i of the individual laboratory values x i. These sources of bias include stylus tip radius and uncertainties in the filter characteristics of the instruments. From Eqn. 1, because the input quantities for the reference value x ref are the individual laboratory values x i, the uncertainty of the reference value should be calculated from the uncertainties of the individual laboratory values. We take this approach for the re-analysis shown later below. 8
9 µm Ra, Bw: 300:1 and 100: Ra, µm Ra, 300:1 Ra, 100:1 300:1 Mean 100:1 Mean 0.18 NRC NIST INMETRO INTI CENAM Ra, µm CR, 300:1 CR, 100:1 U, µm En Ra, µm U, µm En NRC NIST N/A N/A N/A INMETRO N/A N/A N/A INTI N/A N/A N/A CENAM Std. Dev Ref. Value U(X ref ), k= Figure 11: Initial results for roughness specimen SPECIAL 3144 Ra with band ratios of 300:1 and 100:1. 9
10 Rz ISO, 300:1 and 100:1 Original results 0.52 Rz ISO, µm Rz, 300:1 0.2 Rz, 100:1 Mean 300:1 Mean 100:1 NRC NIST INMETRO INTI CENAM CR, 300:1 CR, 100:1 Rz, µm U, µm En, 300:1 Rz, µm U, µm En, 100:1 NRC NIST N/A N/A N/A INMETRO N/A N/A N/A INTI N/A N/A N/A CENAM Std. Dev Ref. Value U(X ref ), k= Figure 12: Initial results for roughness specimen SPECIAL 3144 Rz with cut-off ratios (CR) of 300:1 and 100:1. 10
11 3.0 µm Ra, 100:1 Measured Ra Value, µm Mean s 2.98 NRC INMETRO CENAM 3.0 µm Ra, 300:1 Measured Ra, µm s Mean 3.02 NRC NIST INTI CENAM Sm 99 µm, SMA 2073a 99.7 Measured Sm, µm MNIs Mean 98.5 NRC NIST INTI CENAM Ra RSm CR, 300:1 CR, 100:1 CR, 300:1 Ra, µm U, µm En Ra, µm U, µm En RSm, µm U, µm En NRC NIST N/A N/A INMETRO N/A N/A N/A N/A N/A N/A INTI N/A N/A CENAM Std. Dev Ref. Value U(X ref ), k = Figure 13: Initial results for roughness specimen SRM 2073a - Ra and RSm 11
12 2.55 µm Step 2.60 Measured Step Value, µm Pt s Mean 2.50 NRC NIST INMETRO INTI CENAM 0.38 µm Step Measured Step Value, µm Pt NRC NIST INMETRO INTI CENAM s Mean 0.03 µm Step Measured Step Value, µm s Mean NRC NIST INTI CENAM d, µm d = 2.55 µm d = 0.38 µm d = 0.03 µm U, µm En d, µm U, µm En d, µm U, µm En NRC NIST INMETRO (Pt ) N/A N/A N/A INTI CENAM Std. Dev Ref. Value U(X ref ), k= Figure 14: Initial results for step height specimen, Serial Number
13 Overall, the initial agreement between the five laboratories was very good, as indicated by the generally small values of E n. A satisfactory degree of agreement is indicated by E n < 1. The average E n value for all results here was 0.24 with a standard deviation of The maximum E n value for all measurements was This suggests that at least some of the uncertainty budgets may have been on the conservative side. For individual parameters we observed the following. The Ra measurements of the specimen SPECIAL 3144 (See Fig.11) have excellent agreement. The parameter E n (max) is equal to For the Rz measurement of the same specimen, the results (Fig. 12) are somewhat more varied. Explanations for this behavior are explored below. For the measurement of Ra of the SRM 2073a Sinusoidal Prototype (See Fig. 13), the results are in very good agreement with E n (max) = The step height measurements (See Fig. 14) are also in good agreement with E n (max) = This comparison was also used to study the effect of roughness cut-off ratio (bandwidth) [2]. Two of the laboratories (NRC and CENAM) measured Ra and Rz with a cut-off ratio (λc/λs ) of both 300:1 and 100:1. The differences between results for both Ra and Rz at the two cut-off ratios are significant for the rectangular profile specimen with nominal Ra of 0.2 µm, which has a significant fraction of short spatial wavelengths on its surface (See Fig. 11 and 12). Clearly, the shortest spatial wavelength components of this surface are more attenuated by the filter with cut-off ratio of 100:1, whose short cut-off value is 8 µm, than by the filter with 300:1 cut-off ratio. For the sinusoidal specimen SRM 2073a, which has a fundamental spatial wavelength of 100 µm and only a very small fraction of spatial wavelengths below 8 µm, the differences in results using the different bandwidth instruments are not significant (See Fig. 13). For the measurement of RSm, the quoted uncertainties vary widely. This is because this parameter is not as commonly used as Ra. Hence, the measurement algorithms and the uncertainty budgets for this parameter are likely to vary widely. Best practices for RSm are not as firmly established as they are for other parameters, such as Ra and Rz. NIST s low uncertainty is based on direct use of laser interferometry to measure the lateral displacement of the stylus. This is not common practice for stylus type measurements of RSm. Also the method of calculation of RSm [5] can vary significantly depending on what discrimination criteria are applied to ignoring peaks that are small or close together. By contrast, the definition for a similar parameter Sm defined in the 1995 revision of the U.S. national standard [4] does not apply such criteria and counts all of the profile peaks and valleys. However, these issues are not significant in the case of a sinusoidal specimen with 100 µm RSm, because very short spatial wavelengths are not a significant component of the surface roughness. In spite of the lack of a common practice, the agreement between the laboratories is well within the quoted uncertainties. For step height measurements, the well known algorithm from ISO (2000) [7] (Fig. 10) was used by NRC, NIST, CENAM, and INTI. INMETRO measured a different quantity, known as Pt, the height difference between the highest and lowest points on the profile. For a leveled, flat profile, the value of Pt should be very close to the step height itself. All results agree with the reference value within the quoted uncertainties. 6. Further Observations and Reanalysis After the initial submission and analysis of the results, a number of differences between labs were recognized. We therefore applied a number of corrections to the analysis and recalculated some of the parameters and uncertainty values. The changes took place in several categories as described below. 6.1 Definition of Rz 13
14 During the evaluation process the participants noticed that both NRC and CENAM had Rz (ISO) values smaller than Rz (DIN) if the same raw data is used. Further investigation revealed that the software algorithm of both instruments was based on the older ISO standard. Both laboratories used their raw data and recalculated the Rz (ISO) values using the Rz (DIN) algorithm, which is identical to the new Rz (ISO) algorithm when the evaluation length consists of five sampling lengths, as was the case here. 6.2 Instrument Noise and Rz Because Rz is an extreme value parameter, its value is more sensitive to noise than averaged parameters such as Ra. Hence, instrument noise is a more important cause of variation for Rz measurements than for Ra measurements. Because the instruments likely had different noise levels, it is important to consider the effects carefully. During a summary coordination meeting held at NRC, the participants discussed the measurement uncertainties. It was noted that NRC s expanded uncertainty for Rz (51 nm) was very large in comparison with the other labs. The main reason for this large uncertainty was that NRC estimated the effect of the instrument noise by measuring Rz on an optical flat using the same system and filter conditions that were used to measure the 3144 specimen. This experiment revealed significant noise and therefore influenced the NRC uncertainty estimate. Because it is a range parameter, the Rz parameter is extremely sensitive to noise. Furthermore, the noise can only increase the Rz values. The Rz-noise, which is clearly evident during a smooth-surface measurement, is also present during an artifact measurement and introduces a significant systematic increase into measured Rz results, effectively causing a bias. The participants agreed to conduct the same experiment as NRC, and the results in most of the cases have shown significant instrument noise. It was proposed that each participant correct its Rz result and uncertainty by the following to account for the Rz noise bias and its large uncertainty: Rz Correction = - (0.5 Rz-noise) + (0.5 Rz-noise/ 3) (k=1). (4) The uncertainty conservatively assumes a rectangular distribution for the correction because the size of the correction would depend on the shape of the surface profile and is difficult to calculate with precision. In fact, we expect the correction to be larger for a periodic surface profile such as the SPECIAL 3144 specimen than for a random surface profile. 6.3 NIST Type A Uncertainty NRC, INTI, INMETRO, and CENAM developed their uncertainty budgets with the principal Type A component calculated as one standard deviation of the mean of the set of measured values. This is conventional practice for laboratory comparisons. NIST, on the other hand, initially calculated its uncertainty budgets with the variation due to surface position calculated as one standard deviation of the measurements, according to its practice for measurement calibrations for customers. NIST uses this method because NIST assumes that a customer may calibrate its instrument with the NIST calibrated standard with as little as a single measurement. The method of calculation for this uncertainty component should have been agreed to beforehand but was not. For these measurements the difference in the total uncertainty for the two methods is insignificant for five of NIST s seven submitted results because the instrument uncertainty is a more dominant component in the uncertainty budget than the variation in the measured results. However, for the measurements of Rz and the 0.03 µm step height, the uncertainty is dominated by the variation in measurements and is therefore significantly smaller when calculated using the standard deviation of the mean. Because this method of uncertainty calculation can affect the calculated degree of equivalence between NIST and other laboratories for these two quantities, NIST recalculated its uncertainty budgets for all seven measured parameters using the standard deviation of the mean to represent the variation in measurements, thus putting its statistics in line with others. 6.4 INMETRO Pt Value 14
15 It is expected that the Pt value calculated by INMETRO should be very close to the d parameter for step height for quality step height surfaces, such as the 2.55 µm step and the µm step measured by INMETRO here, which have very little roughness or flatness deviation. This is shown by the good agreement between the INMETRO results for Pt and the results of other laboratories for d in Fig. 14. However, because the parameters are different, it is better to compute the reference value for d only from the four measured results for d. Therefore, we reanalyzed the data in this way. Even though the average value changes slightly, the revised tables for d below still show good agreement between the INMETRO values and the other laboratories. 6.5 Summary Statistics The summaries shown in Section 5 were done according to an agreed upon protocol before any measurements were taken. After the initial submission of results, discussion among the participants, and reviews by colleagues, we decided that the summary statistics should be changed in three ways: 1) Computing the uncertainty of the reference value u(x ref ) from the deviations of the individual lab values with respect to x ref (as shown by Eqn. 3) does not take into account the estimated uncertainty u i of each laboratory s results. The uncertainty of x ref for each computed parameter should be calculated from the law of propagation of uncertainties associated with the input quantities in Eqn. 1. The propagation of uncertainty from the inputs x i to the output x ref yields a standard uncertainty u(x ref ) given by the quadratic sum u n xref ) = u 2 i n i= 1 (. (5) The uncertainty of the reference value for each quantity was therefore recalculated in this way. 2) It was pointed out that the standard method [9] to express degree of equivalence is with the pair of values d i, U(d i ), where d i is the difference between the lab value and reference value d i = xi - x ref (6) and U(d i ) is the uncertainty of that difference. Therefore, the tables of final results were reformatted to show those values explicitly. 3) Taking into account the relatively small number of lab results in the comparison (for all the calculated parameters the number n of s was less than or equal to 5), the dependence of the reference value x ref for any quantity with respect to a single result x i is relatively high. So, there may be a significant correlation between x i and x ref (on the order of 1/ n). Thus the square root of the denominator in Eqn. (3) may be significantly greater than the standard uncertainty of the numerator. It then follows that the parameter E n as first calculated could be an underestimation of the normalized difference between x i and x ref,. Therefore, the variance of the difference u 2 (d i ) was recalculated to account for the correlation between x i and x ref. This quantity is given by [9] n ui u ( d i ) = var( xi ) + var( xref ) 2 cov( xi, xref ) = ui + u 2 i 2 n i= 1 n 2 = ui + u ( xref ), (7) n where var is the variance of a quantity and cov is the covariance of two quantities. Alternatively, exclusive statistics can be used [10,11] to account for the correlations. In this case, different reference values would be obtained for each lab. 15
16 With this recalculation of u 2 (d i ) we avoid the statistical correlations between each single result and the reference value, except those arising from the possible physical correlations among the original results x 1, x n, given by crossed traceability chains or other causes. The tables in Figs. 15 to 20 show the expanded uncertainty U(d i ) equal to 2 u(d i ). These tables also show the recalculated equivalence quotients En given by E n = x x ) /[2 u( d )]. (8) ( i ref i 6.6 Reanalysis Summary Figures 16 to 20 show the results with all changes applied. Overall the agreement between the laboratories is still excellent, and only modest differences were found between E n in the initial results and E n in the final results. The maximum value for E n was 0.72 from the initial results, and one value for E n reaches a value of 0.98 for the final results. The modest E n values shown in Figs suggest that the participants most likely assigned uncertainty values that were too conservative. For 39 measurements, we can expect that we will typically see about 2 results with E n >1, whereas here we do not see any. After the corrections and reanalysis, the Rz results of Fig. 16 all have smaller values than those of Fig. 12 and the uncertainties are larger on the average. The difference between the Rz reference values before and after correction is µm for the 300:1 cut-off ratio and µm for the 100:1 cut-off ratio. Applying the noise correction alone produces very significant changes in the values and uncertainties for Rz. If the noise correction were not applied, the average value of Rz would be higher by µm for the 300:1data and µm for the 100:1 data. These differences are significant with respect to the quoted uncertainties. Because the participants used instruments based on different principles (laser, inductive) and from different manufacturers it seems clear that the surface metrology community is facing a common issue of calculation of Rz in the presence of noise. It is important that the proposal in Eqn. 4 for handling the noise-related component of uncertainty be discussed and that a uniform approach for reckoning noise effects for Rz measurement be adopted. We note that in recent comparisons of measured surface roughness parameters conducted by both the Asia-Pacific Metrology Programme [12] and the European Metrology Programme (EUROMET) [13], there were significant differences among the laboratory results for the averaged Rz parameter [12] (also known as R tm and Ry5) and for the Rmax parameter [12,13]. The parameter Rmax, defined in DIN 4768 [6], is the largest Rz value in a set of five calculated over an evaluation length consisting of five sampling lengths. Perhaps, those differences could be due to the presence of different amounts of noise in the measured profiles and to different procedures to account for noise effects. 16
17 mm Ra, CR: 300:1 and 100:1 After reanalysis 0.23 Ra, m Ra, 300:1 Ra, 100:1 300:1 Mean 100:1 Mean NRC NIST INMETRO INTI CENAM Special 3144 Ra, CR,300:1 Ra, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO INTI CENAM x ref std. unc U, k = Special 3144 Ra, CR,100:1 Ra, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO INTI CENAM x ref std. unc U,k = Figure 15: Results for Ra measurements of roughness specimen SPECIAL 3144 after reanalysis 17
18 0.52 Rz ISO, CR 300:1 and 100:1 After reanalysis Rz, mm Rz, 300:1 Mean, 300:1 Rz, 100:1 Mean 100: NRC NIST INMETRO INTI CENAM Special 3144 Rz, CR, 300:1 Rz, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO INTI CENAM RV std. unc U,k = Special 3144 Rz ISO, CR, 100:1 Rz, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO INTI CENAM x ref std. unc U, k = Figure 16. Comparison of surface roughness Rz results from five SIM laboratories after reanalysis. 18
19 Sm 99 µm, SMA 2073a After reanalysis Measured Sm, m MNIs Mean NRC NIST INTI CENAM SRM 2073a Ra, CR - 300:1 Ra, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO INTI CENAM RV std. unc U,k = SRM 2073a Ra, CR - 100:1 Ra, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO INTI CENAM x ref std. unc U, k = SRM 2073a RSm, CR - 300:1 RSm, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO INTI CENAM x ref std. unc U,k = Figure 17: Comparison of surface roughness Ra and RSm results from five SIM laboratories after reanalysis. The graphs for Ra are essentially unchanged from those of Fig. 13 and are not repeated here. 19
20 µm Step After reanalysis Measured Step Value, µm Pt s Mean 2.5 NRC NIST INMETRO INTI CENAM 18235, d =2.55 µm d, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO (Pt) INTI CENAM x ref std. unc U,k = Figure 18: Comparison of step height (d) and Pt results from five SIM laboratories after reanalysis for the 2.55 µm. 20
21 0.38 µm Step After reanalysis Measured Step Value, µm Pt s Mean NRC NIST INMETRO INTI CENAM 18235, d =0.38 µm d, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO (Pt) INTI CENAM x ref std. unc U,k = Figure 19: Comparison of step height (d) and Pt results from five SIM laboratories after reanalysis for the 0.38 µm step. 21
22 Measured Step Value, µm µm Step After reanalysis s Mean NRC NIST INTI CENAM 18235, d =0.03 µm d, µm U, µm di, µm U (di ), µm En NRC NIST INMETRO INTI CENAM x ref std. unc U,k = Figure 20: Comparison of step height (d) and Pt results from four SIM laboratories after reanalysis for the 0.03 µm step. 22
23 7.0 Final Observation: Bandwidth Limits In future, the choice of the bandwidth limit at small spatial wavelengths for such comparisons will require further consideration. The short wavelength limit usually depends either on the λs filter or on the stylus tip size or perhaps both if those sizes are comparable. For the Ra and Rz results for 300:1 cut-off ratio, NRC, INTI, CENAM, and NIST all used styli with nominal radii of 2 µm and λs of 2.5 µm. However, the stylus used at NIST was measured using the method of tracing over a razor blade [14, 7], and the measured result for the stylus radius was 5.0 µm µm (k=1). The styli used by the other laboratories were not measured, only specified. The comparison of NIST Ra results in Fig. 11 with those of NRC, INTI, and CENAM suggest that the measured 5 µm stylus tip size does not significantly attenuate measured roughness values by comparison with the results for the nominal 2 µm stylus tip sizes as much as the change from a 2.5 µm λs cut-off to a 8 µm λs cut-off. This observation is further supported by the following results from INTI for the 0.2 µm Ra specimen. For a nominal stylus tip radius of 5 µm, the Ra measured by INTI was µm with a standard deviation of 0.001µm, essentially the same as the average result µm shown in Fig. 11 for nominal 2 µm tip radius. The question here is: what was the effective short wavelength cut-off of the INMETRO roughness measurements? The INMETRO data were placed in the 100:1 category and compared with the 100:1 NRC and CENAM data because the nominal stylus radius was 5 µm, higher than the nominal radius of the others. The INMETRO results agreed well with those of NRC and CENAM. However, INMETRO s nominal radius was the same as NIST s measured radius value. So a case could be made that INMETRO s results should be tabulated with the 300:1 results. Perhaps in future comparisons, the stylus radii should be evaluated from measurements or else the short wavelength limit of the measured profiles should be clearly limited by Gaussian filtering. 8.0 Acknowledgement The participants acknowledge the support from NRC piloting this comparison and hosting an evaluation meeting and the support from SIM for some travel expenses. The NIST authors are grateful to A. Hornikova for the careful inspection of these results and suggestions concerning the analysis and to C.D. Foreman for taking surface profile measurements contributing to this report. 9.0 References [1] Mutual Recognition of National Measurement Standards and of Calibration and Measurement Certificates Issued by National Metrology Institutes (International Committee for Weights and Measures, 1999), available at [2] ISO 11562:1996 Geometrical Product Specifications (GPS) Surface texture: Profile method Metrological characteristics of phase correct filters (International Organization for Standardization, Geneva, Switzerland, 1996). [3] ISO 3274:1996 Geometrical Product Specifications (GPS) Surface texture: Profile method Nominal characteristics of contact (stylus) instruments (International Organization for Standardization, Geneva, Switzerland, 1996). [4] ASME B Surface Texture (American Society of Mechanical Engineers, New York, 2003) [5] ISO 4287:1997 Geometrical Product Specifications (GPS) Surface texture: Profile method Terms, definitions and surface texture parameters (International Organization for Standardization, Geneva, Switzerland, 1997). [6] Deutsche Normen DIN 4768, Part 1 Determination of Surface Roughness Values R a, R z, R max with Electric Stylus Instruments, Basic Data (Beuth Verlag GmbH, Berlin, 1974). 23
24 [7] :2000 Geometrical Product Specifications (GPS) Surface texture: Profile method; Measurement standards Part 1: Material measures (International Organization for Standardization, Geneva, Switzerland, 1999). [8] Guide to the Expression of Uncertainty in Measurement (GUM), International Organization for Standardization, Geneva, 1995). [9] M.G. Cox, The Evaluation of Key Comparison Data, Metrologia, 2002, 39, [10] A.G. Steele, B.M. Wood, R.J. Douglas, Exclusive statistics: simple treatment of the unavoidable correlations from key comparison reference values, Metrologia, 2001, 38, No. 6, [11] J.E. Decker, J. Altshuler, H. Beladie, I. Malinovsky, E. Prieto, J. Stoup, A. Titov, M. Villesid, J.R. Pekelsky, Report of SIM 4.2 Regional Comparison, Stage One: Calibration of Gauge Blocks by Optical Interferometry (in press). [12] W.J. Giardini, C. Calderon, F. Cheng-yen, K.K. Hau, C.-S. Kang, S.M. Lee, F. Lesha, A.M. MeKawi, R. Monteros, C. Quan, V. Roonwal, N. Sakano, R.P. Singhal, S.Y. Wong, S.Z.A. Zahwi, International Comparison of Surface-texture Parameters Organized by the Asia-Pacific Metrology Programme, Metrologia, 2002, 39, [13] EUROMET Supplementary Comparison, Surface Texture, Project No. 600, Technical Report (Physikalisch-Technische Bundesanstalt, Braunschweig, Germany, March 23, 2001). [14] J.-F. Song, T.V. Vorburger, Stylus Profiling at High Resolution and Low Force, Applied Optics, 1991, 30,
Perthometer. Surface Texture Parameters New Standards DIN EN ISO / ASME
Perthometer. Surface Texture Parameters New Standards DIN EN ISO / ASME MAHR GMBH, Germany - EDITION 09/01/99 Contents Real surface... Geometrical Product Specification Definition Profile filter P t Profile
Length, Finland, MIKES (VTT Technical Research Centre of Finland Ltd, Centre for Metrology / Mittatekniikan keskus)
absolute mise en pratigue: mise en pratigue: absolute absolute Level or Range 633 633 nm 0.04 fm 2 95% No 1 474 474 THz 24 khz 2 95% No 1 532 532 nm 0.08 fm 2 95% No 50 563 563 THz 0.08 MHz 2 95% No 51
Concept for an Algorithm Testing and Evaluation Program at NIST
Concept for an Algorithm Testing and Evaluation Program at NIST 1 Introduction Cathleen Diaz Factory Automation Systems Division National Institute of Standards and Technology Gaithersburg, MD 20899 A
Optical Methods of Surface Measurement
Optical Methods of Surface Measurement Ted Vorburger, Guest Researcher National Institute of Standards and Technology (NIST) Measurement Science and Standards in Forensic Firearms Analysis 2012 NIST, Gaithersburg,
Report of the Spectral Irradiance Comparison EURAMET.PR-K1.a.1 between MIKES (Finland) and NIMT (Thailand)
Report of the Spectral Irradiance Comparison EURAMET.PR-K1.a.1 between MIKES (Finland) and NIMT (Thailand) M. Ojanen 1, M. Shpak 1, P. Kärhä 1, R. Leecharoen 2, and E. Ikonen 1,3 1 Helsinki University
DETECTION OF SUBSURFACE DAMAGE IN OPTICAL TRANSPARENT MATERIALSS USING SHORT COHERENCE TOMOGRAPHY. Rainer Boerret, Dominik Wiedemann, Andreas Kelm
URN (Paper): urn:nbn:de:gbv:ilm1-2014iwk-199:0 58 th ILMENAU SCIENTIFIC COLLOQUIUM Technische Universität Ilmenau, 08 12 September 2014 URN: urn:nbn:de:gbv:ilm1-2014iwk:3 DETECTION OF SUBSURFACE DAMAGE
LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS
XVII IMEKO World Congress Metrology in the 3rd Millennium June 22 27, 2003, Dubrovnik, Croatia LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS Tanfer Yandayan TUBITAK-Ulusal
A NEW ALGORITHM FOR CLOCK WEIGHTS FOR THE SIM TIME SCALE
Simposio de Metrología 202 8-2 de Octubre, 202 A NEW ALGORITHM FOR CLOCK WEIGHTS FOR THE SIM TIME SCALE 2 J. M. López R, M.A. Lombardi, N. Diaz, E. de Carlos L. División de Tiempo y Frecuencia, Centro
Calibration of AFM with virtual standards; robust, versatile and accurate. Richard Koops VSL Dutch Metrology Institute Delft
Calibration of AFM with virtual standards; robust, versatile and accurate Richard Koops VSL Dutch Metrology Institute Delft 19-11-2015 VSL Dutch Metrology Institute VSL is the national metrology institute
CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS
CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS E. Batzies 1, M. Kreutzer 1, D. Leucht 2, V. Welker 2, O. Zirn 1 1 Mechatronics Research
Stage Two: Calibration of Gauge Blocks by Mechanical Comparison
Report on SIM.4.2 Regional Comparison Stage Two: Calibration of Gauge Blocks by Mechanical Comparison Final Report August 14, 2002 J. E. Decker 1, J. Alschuler, J. Castillo Candanedo, E. De la cruz, E.
TIME AND FREQUENCY METROLOGY WORKING GROUP Working to support time and frequency metrology throughout the Americas
J. Mauricio Lopez R. a, Michael A. Lombardi b, Andrew N. Novick b, Jean-Simon Boulanger c, Raymond Pelletier c, Carlos Donado M. d, and Ricardo Carvalho e a Centro Nacional de Metrología (CENAM), Querétaro,
A guide to Surface Texture Parameters
www.taylor-hobson.com Taylor Hobson SA 6 avenue de Norvège, Hightec 4 91953 Courtaboeuf Cedex, France Tel: +33 160 92 14 14 Fax: +33 160 92 10 20 e-mail: [email protected] Taylor Hobson Germany
TR Series Surface Roughness Tester for workshop and laboratory
TR Series Surface Roughness Tester for workshop and laboratory Portable, rugged and instant operation on metallic and ceramic surfaces Surface roughness measurement the basics Application Surface roughness
Measuring large areas by white light interferometry at the nanopositioning and nanomeasuring machine (NPMM)
Image Processing, Image Analysis and Computer Vision Measuring large areas by white light interferometry at the nanopositioning and nanomeasuring machine (NPMM) Authors: Daniel Kapusi 1 Torsten Machleidt
The Determination of Uncertainties in Charpy Impact Testing
Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials Code of Practice No. 06 The Determination of Uncertainties in Charpy Impact Testing M.A. Lont
Guidelines on the Calibration of Static Torque Measuring Devices
European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET cg-14 Version.0 (03/011) Previously EA-10/14 Calibration Guide EURAMET cg-14
5-Axis Test-Piece Influence of Machining Position
5-Axis Test-Piece Influence of Machining Position Michael Gebhardt, Wolfgang Knapp, Konrad Wegener Institute of Machine Tools and Manufacturing (IWF), Swiss Federal Institute of Technology (ETH), Zurich,
Application Report: Running µshape TM on a VF-20 Interferometer
: Running µshape TM on a VF-20 Interferometer General This report describes how a fiber interferometer from Arden Photonics Ltd was used together with the µshape TM Generic software package. The VF-20
SCS Directory Accreditation number: SCS 0115
International standard: ISO/IEC 17025:2005 Swiss standard: SN EN ISO/IEC 17025:2005 METRON SA Calibration Laboratory Via Luserte Sud 7 CH-6572 Quartino Head: Angelo Capone Responsible for MS: Angelo Capone
Surface Roughness Testing
Surface Roughness Testing M-1 Roughness Parameters Commonly Used in Short Mean roughness Ra (ISO 4287, DIN 4768) The mean roughness Ra matches the arithmetical mean of the absolute values related to the
Solving Simultaneous Equations and Matrices
Solving Simultaneous Equations and Matrices The following represents a systematic investigation for the steps used to solve two simultaneous linear equations in two unknowns. The motivation for considering
The Layman's Guide to ANSI, CEN, and ISO Bar Code Print Quality Documents
The Layman's Guide to ANSI, CEN, and ISO Bar Code Print Quality Documents AIM, Inc. 634 Alpha Drive Pittsburgh, PA 15238-2802 Phone: +1 412 963 8588 Fax: +1 412 963 8753 Web: www.aimglobal.org This Guideline
COMPUTATIONAL ACCURACY ANALYSIS OF A COORDINATE MEASURING MACHINE UNDER STATIC LOAD
COMPUTATIONAL ACCURACY ANALYSIS OF A COORDINATE MEASURING MACHINE UNDER STATIC LOAD Andre R. Sousa 1 ; Daniela A. Bento 2 CEFET/SC Federal Center of Technological Education Santa Catarina Av. Mauro Ramos,
INTRODUCTION TO ERRORS AND ERROR ANALYSIS
INTRODUCTION TO ERRORS AND ERROR ANALYSIS To many students and to the public in general, an error is something they have done wrong. However, in science, the word error means the uncertainty which accompanies
Reflection and Refraction
Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,
PROVE, the next generation registration metrology tool, status report
PROVE, the next generation registration metrology tool, status report Dirk Beyer a, Patricia Gabella b, Greg Hughes b, Gerd Klose c, Norbert Rosenkranz a a Carl Zeiss SMS GmbH (Germany) Carl-Zeiss-Promenade
PCB Component Placement Inspection
Executive Summary PCB Component Placement Inspection Optimet s ConoProbe Mark10 HD with a 50 mm focal length lens was used to inspect PCB component placement. The PCB board inspected contained both individual
digital quality control fail pass at your fingertips Smart Metrology Solutions.
digital quality control fail pass µphase smartgage The clever innovation in digital metrology by FISBA, which lets you smarten up your quality management and economize your production processes, yet spares
COMPARATIVE EVALUATION OF UNCERTAINTY TR- ANSFORMATION FOR MEASURING COMPLEX-VALU- ED QUANTITIES
Progress In Electromagnetics Research Letters, Vol. 39, 141 149, 2013 COMPARATIVE EVALUATION OF UNCERTAINTY TR- ANSFORMATION FOR MEASURING COMPLEX-VALU- ED QUANTITIES Yu Song Meng * and Yueyan Shan National
DEUTSCHE NORM June 1998. Forms and dimensions of undercuts
DEUTSCHE NORM June 1998 Forms dimensions of s { 509 ICS 01.100.20; 25.020 Descriptors: Undercuts, types, dimensions, technical drawings. Supersedes August 1966 edition. Freistiche Formen, Maße In keeping
Good Practice Guide No. 37. The Measurement of Surface Texture using Stylus Instruments. Richard K Leach. Issue 2
Good Practice Guide No. 37 The Measurement of Surface Texture using Stylus Instruments Richard K Leach Issue 2 Measurement Good Practice Guide No. 37 The Measurement of Surface Texture using Stylus Instruments
USE OF FIBRE OPTICS INTERNATIONAL STANDARDS FOR CALIBRATION LABORATORY ACCREDITATION INTERNATIONAL ELECTROTECHNICAL COMMISSION
USE OF FIBRE OPTICS INTERNATIONAL STANDARDS FOR CALIBRATION LABORATORY ACCREDITATION INTERNATIONAL ELECTROTECHNICAL COMMISSION USE OF FIBRE OPTICS INTERNATIONAL STANDARDS FOR CALIBRATION LABORATORY ACCREDITATION
Multiple Optimization Using the JMP Statistical Software Kodak Research Conference May 9, 2005
Multiple Optimization Using the JMP Statistical Software Kodak Research Conference May 9, 2005 Philip J. Ramsey, Ph.D., Mia L. Stephens, MS, Marie Gaudard, Ph.D. North Haven Group, http://www.northhavengroup.com/
Angle metrology at the PTB: current status and developments
Angle metrology at the PTB: current status and developments M. Krause, A. Just, R. D. Geckeler, H. Bosse Physikalisch-Technische Bundesanstalt, Braunschweig und Berlin Bundesallee 100, D-38116 Braunschweig,
ENGINEERING METROLOGY
ENGINEERING METROLOGY ACADEMIC YEAR 92-93, SEMESTER ONE COORDINATE MEASURING MACHINES OPTICAL MEASUREMENT SYSTEMS; DEPARTMENT OF MECHANICAL ENGINEERING ISFAHAN UNIVERSITY OF TECHNOLOGY Coordinate Measuring
EDUCATIONAL SOFTWARE FOR METROLOGY USING JAVA APPLETS
EDUCATIONAL SOFTWARE FOR METROLOGY USING JAVA APPLETS Galván C 1, Hernández I. 1 1 Centro Nacional de Metrología (CENAM), Querétaro, Mexico, [email protected] Abstract: This paper presents some java applets
Injection moulding and modelling on a micro scale
Injection moulding and modelling on a micro scale Technology Update Injection moulding and welding of plastics 11 November 2014 Research Projects (National / European) Micro/Nano/Multimaterial Manufacturing
National Performance Evaluation Facility for LADARs
National Performance Evaluation Facility for LADARs Kamel S. Saidi (presenter) Geraldine S. Cheok William C. Stone The National Institute of Standards and Technology Construction Metrology and Automation
Determination of g using a spring
INTRODUCTION UNIVERSITY OF SURREY DEPARTMENT OF PHYSICS Level 1 Laboratory: Introduction Experiment Determination of g using a spring This experiment is designed to get you confident in using the quantitative
Module 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
Estimating Uncertainties in Testing
Estimating Uncertainties in Testing An Intermediate Guide to Estimating and Reporting Uncertainty of Measurement in Testing Keith Birch British Measurement and Testing Association At least two numbers
UNCERTAINTY AND CONFIDENCE IN MEASUREMENT
Ελληνικό Στατιστικό Ινστιτούτο Πρακτικά 8 ου Πανελληνίου Συνεδρίου Στατιστικής (2005) σελ.44-449 UNCERTAINTY AND CONFIDENCE IN MEASUREMENT Ioannis Kechagioglou Dept. of Applied Informatics, University
MANUAL TR-110 TR-110
MANUAL TR-110 TR-110 CONTENTS 1. GENERAL INTRODUCTION 2 2. WORK PRINCIPLE 3 3. STANDARD DELIVERY 4 4. NAME OF EACH PART 5 5. OPERATION 6 5.1 Preparation before operation 6 5.2 Switch on, Switch off and
Sound Power Measurement
Sound Power Measurement A sound source will radiate different sound powers in different environments, especially at low frequencies when the wavelength is comparable to the size of the room 1. Fortunately
The Viscosity of Fluids
Experiment #11 The Viscosity of Fluids References: 1. Your first year physics textbook. 2. D. Tabor, Gases, Liquids and Solids: and Other States of Matter (Cambridge Press, 1991). 3. J.R. Van Wazer et
Plastic Film Texture Measurement With 3D Profilometry
Plastic Film Texture Measurement With 3D Profilometry Prepared by Jorge Ramirez 6 Morgan, Ste156, Irvine CA 92618 P: 949.461.9292 F: 949.461.9232 nanovea.com Today's standard for tomorrow's materials.
Torque Key Comparison CCM.T-K1.3
Torque Key Comparison CCM.T-K1.3 Measurand Torque: 0 N m, 500 N m, 1000 N m Final Report March 9, 2015 Pilot Institute: Contact Person: E-mail: Co-author: Korea Research Institute of Standards and Science,
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
Metrology of silicon photovoltaic cells using coherence correlation interferometry
Loughborough University Institutional Repository Metrology of silicon photovoltaic cells using coherence correlation interferometry This item was submitted to Loughborough University's Institutional Repository
Application Report. Propeller Blade Inspection Station
34935 SE Douglas Street, Suite, Snoqualmie, WA 9874 Ph: 425-396-5577 Fax: 425-396-586 Application Report Propeller Blade Inspection Station Prepared By Kyle Johnston, Ph. D. Metron Systems Inc.5.5 3 2.5
LOCATION DEPENDENCY OF POSITIONING ERROR IN A 3-AXES CNC MILLING MACHINE
th International & 26 th All India Manufacturing Technology, Design and Research Conference (AIMTDR 214) December 12 th 14 th, 214, IIT Guwahati, Assam, India LOCATION DEPENDENCY OF POSITIONING ERROR IN
CCM.M-K3.1 comparison / 50 kg Mass
CCM.M-K3.1 comparison / 50 kg Mass The 7 th of April 2010 G. MANN - T. MADEC - P.A. MEURY Laboratoire Commun de Métrologie LNE-CNAM Abstract This comparison is a subsequent to the CCM.M-K3 comparison performed
G104 - Guide for Estimation of Measurement Uncertainty In Testing. December 2014
Page 1 of 31 G104 - Guide for Estimation of Measurement Uncertainty In Testing December 2014 2014 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without
Material data sheet. EOS StainlessSteel GP1 for EOSINT M 270. Description, application
EOS StainlessSteel GP1 for EOSINT M 270 A number of different materials are available for use with EOSINT M systems, offering a broad range of e-manufacturing applications. EOS StainlessSteel GP1 is a
DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements
DMS 680 Universal length measuring system gauge inspection for»iso 9000«requirements Universal length measuring system DMS 680 Wide range of applications For periodic inspection of gauges, reference gauges,
The Influence of Tool Tolerances on the Gear Quality of a Gear Manufactured by an Indexable Insert Hob
technical The Influence of Tool Tolerances on the Gear Quality of a Gear Manufactured by an Indexable Insert Hob Mattias Svahn, Lars Vedmar and Carin Andersson Recently, a new type of hob with carbide
Some additional standards and specifications that directly or indirectly affect surface finish requirements include:
Surface Finish Charts Surface Finish Affects Performance The surface finish of process vessels, piping and related components can be a critical factor in their performance, maintenance costs, and service
Physics Lab Report Guidelines
Physics Lab Report Guidelines Summary The following is an outline of the requirements for a physics lab report. A. Experimental Description 1. Provide a statement of the physical theory or principle observed
CNC-STEP. "LaserProbe4500" 3D laser scanning system Instruction manual
LaserProbe4500 CNC-STEP "LaserProbe4500" 3D laser scanning system Instruction manual 2 Hylewicz CNC-Technik Siemensstrasse 13-15 D-47608 Geldern Fon.: +49 (0) 2831 133236 E-Mail: [email protected] Website:
Sensor/Raft Metrology Status
Sensor/Raft Metrology Status Peter Z. Takacs Instrumentation Division Brookhaven National Laboratory 17 Oct 2005 1 Sensor/Raft metrology - current status Laboratory space preparation near completion. Optical
Insertion Losses of Fiber Optical Connectors
Insertion Losses of Fiber Optical Connectors Martin Strasser, Fiber Optics, HUBER+SUHNER AG, Switzerland H+S Technical Series HUBER+SUHNER Excellence in Connectivity Solutions Table of contents 1 Origins
Certificate Standard Reference Material 1866b
National Institute of Standards & Technology Certificate Standard Reference Material 1866b Common Commercial Asbestos This standard reference material (SRM) is comprised of three commercial-grade asbestos
Using Equotip Hardness Test Blocks Abstract 1. Introduction 2. Verification of Leeb hardness testers according to standards
Using Equotip Hardness Test Blocks Heinz-Horst Pollok, consultant for material testing and instrument electronics (Cologne, Germany) Dr. Ralph T. Mennicke, product manager for material testing, Proceq
Quality Regulations for Profiles
Quality Regulations for Profiles September 2010 Quality Regulations for Profiles Page 2/42 Contents Preface... 4 1. Terms and Definitions... 5 1.1 Third parties... 5 1.2 Independent laboratories... 5 1.3
Efficiency, Dispersion and Straylight Performance Tests of Immersed Gratings for High Resolution Spectroscopy in the Near Infra-red
Changing the economics of space Efficiency, Dispersion and Straylight Performance Tests of Immersed Gratings for High Resolution Spectroscopy in the Near Infra-red J. Fernandez-Saldivar 1, F. Culfaz 1,
Self-Mixing Laser Diode Vibrometer with Wide Dynamic Range
Self-Mixing Laser Diode Vibrometer with Wide Dynamic Range G. Giuliani,, S. Donati, L. Monti -, Italy Outline Conventional Laser vibrometry (LDV) Self-mixing interferometry Self-mixing vibrometer Principle:
SJ-201P/M PORTABLE SURFACE ROUGHNESS TESTER
CATALOG No. E4152-178 PORTABLE SURFACE ROUGHNESS TESTER Smart tool in the workshop PORTABLE SURFACE ROUGHNESS TESTER Large characters are displayed on the large easy-to-view LCD. Portable for easy measurement
* This work is an official contribution of the National Institute of Standards and Technology and
Variability in the Geometric Accuracy of Additively Manufactured Test Parts A.L. Cooke and J.A. Soons National Institute of Standards and Technology * Gaithersburg, MD, USA Abstract This paper describes
SPINDLE ERROR MOVEMENTS MEASUREMENT ALGORITHM AND A NEW METHOD OF RESULTS ANALYSIS 1. INTRODUCTION
Journal of Machine Engineering, Vol. 15, No.1, 2015 machine tool accuracy, metrology, spindle error motions Krzysztof JEMIELNIAK 1* Jaroslaw CHRZANOWSKI 1 SPINDLE ERROR MOVEMENTS MEASUREMENT ALGORITHM
Optical Quality Control for Industry: Applicable in Laboratory up to Inline-Inspection. Dr. Josef Frohn NanoFocus AG Oberhausen, Ettlingen
Optical Quality Control for Industry: Applicable in Laboratory up to Inline-Inspection Dr. Josef Frohn NanoFocus AG Oberhausen, Ettlingen 1 NanoFocus AG founded in 1994 optical surface inspection: development
8.1 HPC for improved efficiency on standard machine tools by using new fluid-driven spindles
8.1 HPC for improved efficiency on standard machine tools by using new fluid-driven spindles A. Schubert 1, O. Harpaz 2, B. Books 2, U. Eckert 1, R. Wertheim 1 1 Fraunhofer IWU, Reichenhainer Str. 88,
Pipeline External Corrosion Analysis Using a 3D Laser Scanner
Pipeline Technology Conference 2013 Pipeline External Corrosion Analysis Using a 3D Laser Scanner Pierre-Hugues ALLARD, Charles MONY Creaform, www.creaform3d.com 5825 rue St-Georges, Lévis (QC), Canada,
Factoring Polynomials
UNIT 11 Factoring Polynomials You can use polynomials to describe framing for art. 396 Unit 11 factoring polynomials A polynomial is an expression that has variables that represent numbers. A number can
Grazing incidence wavefront sensing and verification of X-ray optics performance
Grazing incidence wavefront sensing and verification of X-ray optics performance Timo T. Saha, Scott Rohrbach, and William W. Zhang, NASA Goddard Space Flight Center, Greenbelt, Md 20771 Evaluation of
7.2 Quadratic Equations
476 CHAPTER 7 Graphs, Equations, and Inequalities 7. Quadratic Equations Now Work the Are You Prepared? problems on page 48. OBJECTIVES 1 Solve Quadratic Equations by Factoring (p. 476) Solve Quadratic
Fatigue Life Estimates Using Goodman Diagrams
Fatigue Life Estimates Using Goodman Diagrams by Robert Stone The purpose of this paper is to review the proper methods by which spring manufacturers should estimate the fatigue life of a helical compression
Keywords: Torque, calibration, uncertainty, torque tools, torque testers. Contents
Measurement uncertainty in torque calibration Luca G. Bochese, ASQ CQE, CRE (Boch - Milano), [email protected], www.boch.net Dr. A. Bochese ( Feanor - Tallinn), [email protected], www.feanor.com Keywords: Torque,
Measuring Line Edge Roughness: Fluctuations in Uncertainty
Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as
A Fuzzy System Approach of Feed Rate Determination for CNC Milling
A Fuzzy System Approach of Determination for CNC Milling Zhibin Miao Department of Mechanical and Electrical Engineering Heilongjiang Institute of Technology Harbin, China e-mail:[email protected]
CNC Applications. Tool Radius Compensation for Machining Centers
CNC Applications Tool Radius Compensation for Machining Centers Why Cutter Diameter Compensation? When machining finished surfaces with the side of a milling cutter (generally called profiling), the accuracy
Experiment (13): Flow channel
Introduction: An open channel is a duct in which the liquid flows with a free surface exposed to atmospheric pressure. Along the length of the duct, the pressure at the surface is therefore constant and
Machine Tool Inspection & Analyzer solutions
Machine Tool Inspection & Analyzer solutions Introduction IBS Precision Engineering machine tool inspection series offers a range of products to provide peace of mind and control to machine operators by
Prototyping to Production
White Paper Prototyping to Production Konrad Goffin David Montgomery Cicely Rathmell INTRODUCTION CVI Laser Optics quick turnaround prototype services smooth the transition from prototype to production.
Shrinkage Modeling of Injection Molded Precision Optics
Shrinkage Modeling of Injection Molded Precision Optics Jay Udayasankaran*, Nelson E. Claytor, Donald M. Combs, Oscar M. Lechuga, John J. Mader, and Rajesh Mittal Fresnel Technologies, Inc., 101 W. Morningside
APPENDIX N. Data Validation Using Data Descriptors
APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection
TEST REPORT. Rendered to: HOMELAND VINYL PRODUCTS, INC. For:
TEST REPORT Rendered to: HOMELAND VINYL PRODUCTS, INC. For: PRODUCT: Gorilla Column PVC Porch Post System with PVC Base / Cap Plates, PVC Bushings and 2-3/8 in Diameter and 3-1/2 in Diameter Galvanized
Betriebsanleitung Operating Instructions Guide de l utilisateur Manuale di istruzioni Manual de instrucciones MarSurf PS1
Betriebsanleitung Operating Instructions Guide de l utilisateur Manuale di istruzioni Manual de instrucciones MarSurf PS1 Mahr GmbH Carl-Mahr-Straße 1 D-37073 Göttingen Telefon +49 551 7073-0 Fax +49 551
Experimental Evaluation of the Discharge Coefficient of a Centre-Pivot Roof Window
Experimental Evaluation of the Discharge Coefficient of a Centre-Pivot Roof Window Ahsan Iqbal #1, Alireza Afshari #2, Per Heiselberg *3, Anders Høj **4 # Energy and Environment, Danish Building Research
AP Physics 1 and 2 Lab Investigations
AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks
P R E A M B L E. Facilitated workshop problems for class discussion (1.5 hours)
INSURANCE SCAM OPTICS - LABORATORY INVESTIGATION P R E A M B L E The original form of the problem is an Experimental Group Research Project, undertaken by students organised into small groups working as
Negative gauge pressure comparison
J6/2009 Negative gauge pressure comparison Range -95 kpa to + 95 kpa EURAMET Project 1131 Markku Rantanen, Sari Saxholm, Aykurt Altintas, Richard Pavis and Guliko Peterson Centre for Metrology and Accreditation
DISPLAYING SMALL SURFACE FEATURES WITH A FORCE FEEDBACK DEVICE IN A DENTAL TRAINING SIMULATOR
PROCEEDINGS of the HUMAN FACTORS AND ERGONOMICS SOCIETY 49th ANNUAL MEETING 2005 2235 DISPLAYING SMALL SURFACE FEATURES WITH A FORCE FEEDBACK DEVICE IN A DENTAL TRAINING SIMULATOR Geb W. Thomas and Li
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity
Determining Measurement Uncertainty for Dimensional Measurements
Determining Measurement Uncertainty for Dimensional Measurements The purpose of any measurement activity is to determine or quantify the size, location or amount of an object, substance or physical parameter