CCM.M-K3.1 comparison / 50 kg Mass
|
|
|
- Emery Phillips
- 9 years ago
- Views:
Transcription
1 CCM.M-K3.1 comparison / 50 kg Mass The 7 th of April 2010 G. MANN - T. MADEC - P.A. MEURY Laboratoire Commun de Métrologie LNE-CNAM Abstract This comparison is a subsequent to the CCM.M-K3 comparison performed in and published in 2005 [1]. This comparison is between the CEM (Spain) as participating laboratory and the LNE (France) as pilot laboratory. The aim of this comparison was to compare the results obtained by each laboratory when calibrating the same two 50 kg stainless steel mass standards, one provided by CEM and one by LNE. 1. Introduction The purpose of this document is to define the organisation of the subsequent comparison and to provide instructions on the transport of the transfer standards and the reporting of the measurement results. It is defined by applying the guidelines for CIPM key comparisons (appendix F to the MRA ). 2. Organisation Two travelling 50 kg standards circulated between the pilot and the participant of the comparison, one belongs to LCM and the other one belongs to CEM. The LCM standard was used to perform the loop 1. It was calibrated by LCM before and after the CEM calibration. The CEM standard was used to perform the loop 2. It was calibrated by the CEM before and after the LCM calibration. 2.1 Participating laboratories Pilot Laboratory Country Laboratoire Commun de Métrologie LNE-CNAM LCM France Participant Laboratory Country Centro Español de Metrología CEM Spain CCM.M-K3.1 comparison draft B 7 April 2010 page 1/10
2 2.2 Comparison scheme Two travelling 50 kg standards circulated among the participants. The travel of the standards was set up in a circular form with two loops: - Loop 1, using the first travelling standard (January 2009 to July 2009) from France to Spain, then return to France, - Loop 2, using the second travelling standard (January 2009 to November 2009) from Spain to France, then return to Spain. Comparison schedule: Dates January 09 January 09 February May 09 June 09 July 09 November 09 Activity By each laboratory, calibration of one s own travelling standard transfer CEM travelling standard from Spain to France then transfer LCM travelling standard from France to Spain By each laboratory, calibration of travelling standard of the other laboratory transfer LCM travelling standard from Spain to France then transfer CEM travelling standard from France to Spain New calibration of one s own travelling standard by LCM New calibration of one s own travelling standard by CEM Table 1 - Comparison schedule CCM.M-K3.1 comparison draft B 7 April 2010 page 2/10
3 2.3 Characteristics of the Mass standards Loop 1: travelling standard 6 The LCM travelling standard, used in loop 1, is in stainless steel X 18 M 25 W and was manufactured by the ZWIEBEL company. It is cylinder with a circular fork groove. This standard is called six comma (6 ) and identified by a serial number engraved on the top plane face. A handling fork was supplied with this standard. Parameter Value expanded uncertainty (95% coverage) Density at 20 C kg.m kg.m -3 Magnetic susceptibility 3.30 x x10-3 Height mm 0.20 mm Diameter mm 0.20 mm Height of centre of gravity above base mm 0.40 mm Table 2: Characteristics of the LCM standard Picture 1 - the LCM travelling standard (n 6 ) CCM.M-K3.1 comparison draft B 7 April 2010 page 3/10
4 Loop 2 : travelling standard MP9 The CEM travelling standard, used in loop 2, is identified by a serial number engraved on the top plane face: number MP9. It has the following features: Parameter Value expanded uncertainty (95% coverage) Density at 20 C kg.m -3 1,5 kg.m -3 Height mm 0.2 mm Diameter mm 0.2 mm Height of centre of gravity above base mm 2.0 mm Table 3 : Characteristics of the CEM standard Picture 2 - the CEM travelling standard (n MP9) CCM.M-K3.1 comparison draft B 7 April 2010 page 4/10
5 2.4 Travelling conditions Staff from CEM was in charged of standards transportation between institutes. The travelling standards were examined on receipt and any scratches or other marks, if any, were recorded on the Travelling standard visual inspection form. This form was sent by fax or to LCM within 24 hours after the inspection. 2.5 Measurement Cleaning No cleaning is applied to the standards. If necessary a light brushing was made with a soft-haired sable brush Ambient conditions The measurements were made under ambient conditions of air. The parameters contributing to air density were recorded for each weighing and the air density was calculated using the CIPM 2007 formula Weighing procedure The laboratory applied its own weighing procedure. 3. Results of the comparison 3.1 Stability of the travelling mass standards Each laboratory compared its one s own travelling standards before and after calibration made by the other laboratory. The results of these comparisons are given in tables 4a and 4b below. travelling standard 6 LCM travelling standard MP9 CEM Date value Date value Jan mg Jan mg July mg Nov mg Table 4a: Deviations d of the travelling standards from the nominal value CCM.M-K3.1 comparison draft B 7 April 2010 page 5/10
6 travelling standard 6 travelling standard MP9 d u( d) d u( d) mg 0.92 mg mg 0.84 mg Table 4b: drift d of the travelling standards during the comparison and standard uncertainty u( d) of this drift These results show that the mass of the two travelling standards did not change significantly during the comparison. 3.2 Results reported by the CEM Table 5 shows the results and combined uncertainties as given by the CEM. Reported results of the participant (m: mass of the travelling standard - m 0 ; nominal value of the standard) are as follows: u (m-m 0 ) standard uncertainty reported. Loop 1 Date Laboratory m-m 0 u(m-m 0 ) Travelling Standard 6 May 09 CEM mg 3.96 mg Loop 2 Date Laboratory m-m 0 u c Travelling Standard MP9 January 09 November 09 CEM mg 3.96 mg CEM mg 3.96 mg Table 5 - Results reported by the CEM CCM.M-K3.1 comparison draft B 7 April 2010 page 6/10
7 3.3 Reference value coming from CCM.M-K3 comparison The two travelling standards (n 6 and n MP9) were calibrated each by two mass standards (n 9 and n 7) used in the previous CCM.M-K3 comparison [1]. According to this comparison, the reference value mc ref and its expanded uncertainty are as follows: Reference CCM.M-K3 reference standard Value Expanded uncertainty (95% coverage) n mg 2.10 mg n mg 2.10 mg Table 6 - Reference values coming from CCM.M-K3 comparison 3.4 Reference value of the travelling standards Results of the LCM, coming from the calibration, are given in table 7. The reference value of n 6 travelling standard is the mean value of calibration before and after CEM (see table 4a). Reference Travelling standard Value Expanded uncertainty (95% coverage) Calibrated vs n mg 3.0 mg n 9 and n 7 n MP mg 3.0 mg Table 7 - Reference value of the travelling standards CCM.M-K3.1 comparison draft B 7 April 2010 page 7/10
8 Graph 1 Calibration results of 50 kg n 6 Graph 2 Calibration results of 50 kg MP9 3.5 Degree of equivalence of the CEM The degree of equivalence deq A of the CEM laboratory (noted A) is equal to the difference mc A mc ref between the participant s corrected value and the reference value. For the 6 travelling standard, mc ref is the mean value of the calibration before and after CEM. For the MP9 travelling standard, mc A is the mean value of the calibration before and after LCM. The uncertainty of the degree of equivalence u(deq A ) takes into account the uncertainty components coming from the result given by the laboratory u(m A ), the CCM.M-K3.1 comparison draft B 7 April 2010 page 8/10
9 correction applied to this result u d, the drift of the travelling standard d (with the assumption of a rectangular distribution) and the uncertainty u(mc ref ) of the reference value : u 2 2 d deq u m u d A A 12 2 u 2 mc ref (1) Table below gives the degree of equivalence deq A of each laboratory with the assigned expanded uncertainty U(deq) (95% coverage). (mg) reference value calibration by n 7 and n 9 from CCM.M-K3 comparison deq A U(deq A ) MET mg 8.70 mg MP mg 8.40 mg Table 8 - Degree of equivalence of the CEM Graph 3 degree of equivalence of the CEM CCM.M-K3.1 comparison draft B 7 April 2010 page 9/10
10 4. Mass Comparator used by participant Laboratory Manufacturer Type Resolution Standard deviation (1) Degree of freedom CEM Mettler Toledo AX ,1 mg 0.4 mg (1) Standard deviation of repeatability or reproducibility of the result of one comparison process Table 9 - Mass Comparator used by CEM 5. References [1] CCM.M-K3: Metrologia, 2005, 42, Tech. Suppl., CCM.M-K3.1 comparison draft B 7 April 2010 page 10/10
Calibration Certificate
P.O. Box 87-0087 Phone:856-686-1600 Fax:856-686-1601 www.troemner.com e-mail: [email protected] SECTION 1: Page 1 of 7 Pages SECTION 2: APPROVED SIGNATORY Joseph Moran SECTION 3: PERSON PERFORMING
Estimates of Uncertainty of the Calibration of Balances
Estimates of Uncertainty of the Calibration of Balances Chart shows the uncertainty (estimates) of the calibration of representative electronic balances. Chart. Uncertainty of calibration (estimates) Minimum
User checks and maintenance of laboratory balances
Technical Note 13 March 2014 Issued: July 1995 Amended and reissued: July 2005, July 2007, October 2007, May 2010, March 2013, March 2014 User checks and maintenance of laboratory balances Copyright National
OIML R 111-1 RECOMMENDATION. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION , E 2. Weights of classes E 1 , F 2 , M 1 2 , M 1
INTERNATIONAL RECOMMENDATION OIML R 111-1 Edition 2004 (E) Weights of classes E 1, E 2, F 1, F 2, M 1, M 1 2, M 2, M 2 3 and M 3 Part 1: Metrological and technical requirements Poids des classes E 1, E
EURAMET 1197 SUPPLEMENTARY BILATERAL COMPARISON OF HYDRAULIC GAUGE PRESSURE STANDARDS UP TO 50 MPa
Final Report of EURAMET 1197 SUPPLEMENTARY BILATERAL COMPARISON OF HYDRAULIC GAUGE PRESSURE STANDARDS UP TO 50 MPa Yasin DURGUT 1 Nenad PETROVSKI 2 Vanco KACARSKI 2 (1) National Metrology Institute (UME),
Weight Calibration ComparatorPac TM Simple Calibration Efficient Workflow Reliable Results
Weight Calibration ComparatorPac TM Simple Calibration Efficient Workflow Reliable Results Safe and Simple Weight Calibration with ComparatorPac TM ComparatorPac Weight Calibration for Traceable Results
Gravimetric determination of pipette errors
Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with
S Davidson, M Perkin, M Buckley
No. 71 Measurement Good Practice Guide The Measurement of Mass and Weight S Davidson, M Perkin, M Buckley The National Physical Laboratory is operated on behalf of the DTI by NPL Management Limited, a
Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 7 Contact: Mr S C Sparks Solent Industrial Estate Tel: +44 (0)1489 790296 Hedge End Fax: +44 (0)1489 790294 Southampton E-Mail: [email protected]
A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS
Presented 2006 May 16-18 2006 ISFFM Queretaro, Mexico A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS P. Delajoud, M. Bair, C. Rombouts, M. Girard DH Instruments,
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
Guidelines on the Calibration of Static Torque Measuring Devices
European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET cg-14 Version.0 (03/011) Previously EA-10/14 Calibration Guide EURAMET cg-14
Key comparison CCM.M-K1 MEASURAND : Mass NOMINAL VALUE : 1 kg. x (rep) i :
Key comparison CCM.M-K1 MEASURAND : Mass NOMINAL VALUE : 1 kg x (rep) i : u i : x i : reported result obtained as the difference between the measurement of the mass standards carried out by laboratory
Certificate of Calibration No 131 00835
Schweizerische Confederation Confederazione Confederaziun Eidgeno5senschaft suisse svizra Federal Department of Justice alld Police FDJp Federal Office of Metrology METAS Certificate of Calibration No
Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance
INL/CON-10-18643 PREPRINT Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance 2010 NCSL International Workshop and Symposium Michael Stears July 2010 This is
Report of the Spectral Irradiance Comparison EURAMET.PR-K1.a.1 between MIKES (Finland) and NIMT (Thailand)
Report of the Spectral Irradiance Comparison EURAMET.PR-K1.a.1 between MIKES (Finland) and NIMT (Thailand) M. Ojanen 1, M. Shpak 1, P. Kärhä 1, R. Leecharoen 2, and E. Ikonen 1,3 1 Helsinki University
Training Material for Balances (1) Prepared by R&D Division 1. Marketing Department A&D Company, Limited
Training Material for Balances (1) Prepared by R&D Division 1 Marketing Department A&D Company, Limited 1 Contents Things you should know before using a balance --------------------------- P.3 1. Convection
Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK
code Location code Customers Sites locations: 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK Calibration Centre Bolkiah Garrison BB3510 Negara Brunei Darussalam Contact: Mr Lim Tiong Thai Tel: +673-2-386475
Testing Your Laboratory Balance
Testing Your Laboratory Balance If you use a laboratory balance in your day-to-day work, you may have wondered whether it is accurate. Here are four tests that you can use to confirm your instrument's
How To Use A Corona Mdi 110
CORONA MDI 110 APPLICATION Domestic water meter FEATURES 4 Modular multi-jet domestic meter with inductive scanning (without magnetic influence) 4 System capability, with standard applicable pulse output
Activity P13: Buoyant Force (Force Sensor)
Activity P13: Buoyant Force (Force Sensor) Equipment Needed Qty Equipment Needed Qty Economy Force Sensor (CI-6746) 1 Mass and Hanger Set (ME-9348) 1 Base and Support Rod (ME-9355) 1 Ruler, metric 1 Beaker,
Determining Measurement Uncertainty for Dimensional Measurements
Determining Measurement Uncertainty for Dimensional Measurements The purpose of any measurement activity is to determine or quantify the size, location or amount of an object, substance or physical parameter
FLOW MEASUREMENT 2001 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER
FLOW MEASUREMENT 200 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER Dr D G Stewart, NEL Dr M Reader-Harris, NEL Dr R J W Peters, McCrometer Inc INTRODUCTION The V-Cone
Activity P13: Buoyant Force (Force Sensor)
July 21 Buoyant Force 1 Activity P13: Buoyant Force (Force Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Archimedes Principle P13 Buoyant Force.DS P18 Buoyant Force P18_BUOY.SWS
Case study Realistic and best uncertainties with humidity instruments
Case study Realistic and best uncertainties with humidity instruments Domen Hudoklin September 2014 Brdo, Slovenia Humidity calibrations Dewpoint primary - against dewpoint generator secondary - against
WEIGHTS AND MASS STANDARDS
WEIGHTS AND MASS STANDARDS IN ACCORDANCE WITH THE OIML R111: 2004 MADE BY WEEGTECHNIEK HOLLAND B.V. Patroonsweg 23-27 3892 DA Zeewolde The netherlands Tel. +31 (0)36-5222030 Fax. +31 (0)36-5222060 E-mail:
Oilfield Services Measurement Solutions
Oilfield Services Measurement Solutions Superior flow and density measurement Best-in-class measurement How can I streamline my operations for efficiency without compromising environmental, safety and
Using Equotip Hardness Test Blocks Abstract 1. Introduction 2. Verification of Leeb hardness testers according to standards
Using Equotip Hardness Test Blocks Heinz-Horst Pollok, consultant for material testing and instrument electronics (Cologne, Germany) Dr. Ralph T. Mennicke, product manager for material testing, Proceq
The Analytical Balance
Chemistry 119: Experiment 1 The Analytical Balance Operation of the Single-Pan Analytical Balance Receive instruction from your teaching assistant concerning the proper operation of the Sartorius BP 210S
LAB14 EDITION 4 November 2006
LAB14 EDITION 4 November 2006 Calibration of Weighing Machines CONTENTS SECTION PAGE 1 Introduction 2 2 Calibration of Weights 2 3 Considerations in the use of weighing machines 4 4 In-house calibration
Measurement and Calibration
Adapted from: H. A. Neidig and J. N. Spencer Modular Laboratory Program in Chemistry Thompson Learning;, University of Pittsburgh Chemistry 0110 Laboratory Manual, 1998. Purpose To gain an understanding
Specific volume and absorption properties
Accepted 1980 Fluff Specific volume and absorption properties 1 Scope This SCAN-test Method describes the preparation of test pieces of fluff and a procedure for their use in determining the specific volume
Estimating Weighing Uncertainty From Balance Data Sheet Specifications
Estimating Weighing Uncertainty From Balance Data Sheet Specifications Sources Of Measurement Deviations And Uncertainties Determination Of The Combined Measurement Bias Estimation Of The Combined Measurement
THE DETERMINATION OF THE MAXIMUM DRY DENSITY AND OPTIMUM MOISTURE CONTENT OF MATERIALS USING THE VIBRATORY HAMMER COMPACTION
THE DETERMINATION OF THE MAXIMUM DRY DENSITY AND OPTIMUM MOISTURE CONTENT OF MATERIALS USING THE VIBRATORY HAMMER COMPACTION 1. SCOPE The maximum dry density and optimum moisture content, as defined below,
Classic Balances NewClassic ML Precise results Compact and portable Affordable for all budgets
Classic Balances NewClassic ML Precise results Compact and portable Affordable for all budgets A Smart Choice Swiss Weighing Top Performance A Smart Choice for all Budgets Precise results, comfortable
Chapter 16. Mensuration of Cylinder
335 Chapter 16 16.1 Cylinder: A solid surface generated by a line moving parallel to a fixed line, while its end describes a closed figure in a plane is called a cylinder. A cylinder is the limiting case
INVESTIGATION OF FALLING BALL VISCOMETRY AND ITS ACCURACY GROUP R1 Evelyn Chou, Julia Glaser, Bella Goyal, Sherri Wykosky
INVESTIGATION OF FALLING BALL VISCOMETRY AND ITS ACCURACY GROUP R1 Evelyn Chou, Julia Glaser, Bella Goyal, Sherri Wykosky ABSTRACT: A falling ball viscometer and its associated equations were studied in
Balance Accuracy. FlexiWeigh
FlexiWeigh The balance is an integral part of the FlexiWeigh workstation. The role and limitations of the balance are important to understand in this environment. This technical note describes the different
Vandal-proof keypad series
Model No. LBKP35900 / LBKP35900E (Encryption available) Interface: PS2, USB available. Dimensions: 100mm x 91.5mm No. of keys: 16 keys, high quality stainless steel G.W.: 0.75kg. All languages version
GENERAL SCIENCE LABORATORY 1110L Lab Experiment 5 THE SPRING CONSTANT
GENERAL SCIENCE LABORATORY 1110L Lab Experiment 5 THE SPRING CONSTANT Objective: To determine the spring constant of a spiral spring Apparatus: Pendulum clamp, aluminum pole, large clamp, assorted masses,
Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids
Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Dr. J. M. Meyers Dr. D. G. Fletcher Dr. Y. Dubief 1. Introduction Last lab you investigated flow loss in a pipe due to the roughness
Address for Correspondence
International Journal of Advanced Engineering Technology E-ISSN 0976-3945 Research Paper DEVELOPMENT OF LOW COST SHAKE TABLES AND INSTRUMENTATION SETUP FOR EARTHQUAKE ENGINEERING LABORATORY C. S. Sanghvi
Torque Key Comparison CCM.T-K1.3
Torque Key Comparison CCM.T-K1.3 Measurand Torque: 0 N m, 500 N m, 1000 N m Final Report March 9, 2015 Pilot Institute: Contact Person: E-mail: Co-author: Korea Research Institute of Standards and Science,
The entire document shall be read and understood before proceeding with a test. ISTA 3B 2013 - Page 1 of 35
Packaged-Products for Less-Than-Truckload (LTL) Shipment ISTA 3 Series General Simulation Performance Test PROCEDURE VERSION DATE Last TECHNICAL Change: NOVEMBER 2012 Last EDITORIAL Change: JANUARY 2013
Physics 40 Lab 1: Tests of Newton s Second Law
Physics 40 Lab 1: Tests of Newton s Second Law January 28 th, 2008, Section 2 Lynda Williams Lab Partners: Madonna, Hilary Clinton & Angie Jolie Abstract Our primary objective was to test the validity
Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Old Bracknell Lane West Bracknell Berkshire RG12 7AH Contact: Mr M Trotter Tel: +44 (0)1344 459314 Fax: +44 (0)1344 465556 E-Mail: [email protected]
Measuring equipment in a laboratory.
Measuring equipment in a laboratory. Laboratory and metrology are two words that practically do not function separately. They are integrally connected to measuring equipment for tests and measurements.
S.3.6. BULK DENSITY AND TAPPED DENSITY OF POWDERS. Final text for addition to The International Pharmacopoeia
March 2012 S.3.6. BULK DENSITY AND TAPPED DENSITY OF POWDERS Final text for addition to The International Pharmacopoeia This monograph was adopted at the Forty-sixth WHO Expert Committee on Specifications
Three Methods for Calculating the Buoyant Force Gleue: Physics
Three Methods for Calculating the Buoyant Force Gleue: Physics Name Hr. The Buoyant Force (F b ) is the apparent loss of weight for an object submerged in a fluid. For example if you have an object immersed
Pipette Calibration. Accurate Pipettes Day by Day. Ensure Operational Reliability With Service XXL. www.mt.com For more information
Ensure Operational Reliability With Service XXL Qualify your balances and work safely from day one. METTLER TOLEDO offers customized solutions for working in regulated environments. EQ-Pac Conformity and
Experiment #4 Sugar in Soft Drinks and Fruit Juices. Laboratory Overview CHEM 1361. August 2010
Experiment #4 Sugar in Soft Drinks and Fruit Juices Laboratory Overview CHEM 1361 August 2010 Gary S. Buckley, Ph.D. Department of Physical Sciences Cameron University Learning Objectives Relate density
CHAPTER 15 FORCE, MASS AND ACCELERATION
CHAPTER 5 FORCE, MASS AND ACCELERATION EXERCISE 83, Page 9. A car initially at rest accelerates uniformly to a speed of 55 km/h in 4 s. Determine the accelerating force required if the mass of the car
DENSITY MEASURING SYSTEMS
SHIBAYAMA SCIENTIFIC CO., LTD. DENSITY MEASURING SYSTEMS Density Gradient Tube Method Direct Reading Type A / Type B Comply with the standards below: JIS K-0061-1992, K-7112 1980 ASTM D1505 Type A /Left
GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards
March 2003 GMP 11 Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards 1. Introduction 1.1. Purpose Measurement processes are dynamic systems and often
Chapter 11. h = 5m. = mgh + 1 2 mv 2 + 1 2 Iω 2. E f. = E i. v = 4 3 g(h h) = 4 3 9.8m / s2 (8m 5m) = 6.26m / s. ω = v r = 6.
Chapter 11 11.7 A solid cylinder of radius 10cm and mass 1kg starts from rest and rolls without slipping a distance of 6m down a house roof that is inclined at 30 degrees (a) What is the angular speed
www.kem-kueppers.com [email protected] Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters
www.kem-kueppers.com [email protected] Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters Description The Tricor Mass Flow Meters measure simultaneously mass
CHAPTER 6 WEAR TESTING MEASUREMENT
84 CHAPTER 6 WEAR TESTING MEASUREMENT Wear is a process of removal of material from one or both of two solid surfaces in solid state contact. As the wear is a surface removal phenomenon and occurs mostly
Negative gauge pressure comparison
J6/2009 Negative gauge pressure comparison Range -95 kpa to + 95 kpa EURAMET Project 1131 Markku Rantanen, Sari Saxholm, Aykurt Altintas, Richard Pavis and Guliko Peterson Centre for Metrology and Accreditation
ANSWER KEY. Work and Machines
Chapter Project Worksheet 1 1. inclined plane, wedge, screw, lever, wheel and axle, pulley 2. pulley 3. lever 4. inclined plane 5. Answers will vary: top, side, or bottom 6. Answers will vary; only one
Eco Green Labs India Pvt. Ltd., Ground Floor, Plot No. 100, New Sidco Industrial Estate, Srinagar, Hosur, Tamil Nadu
Last Amended on - Page 1 of 5 I. DIMENSION $ 1. PLAIN PLUG S Upto Ø 100 mm 1.1 m Using ULM > Ø100 mm to Ø 300 mm 2.4 m 2. PLAIN RING S Upto Ø 100 mm 1.7 m Using ULM >Ø100 mm to Ø 300 mm 3.6 m 3. THREAD
Physics Lab Report Guidelines
Physics Lab Report Guidelines Summary The following is an outline of the requirements for a physics lab report. A. Experimental Description 1. Provide a statement of the physical theory or principle observed
Establishment of the Gas Flow Calibration Facility at CSIR-NPL, India
Establishment of the Gas Flow Calibration Facility at CSIR-PL, India Speaker: Shiv Kumar Jaiswal Fluid Flow Measurement Standards CSIR-ational Physical Laboratory (PLI) Dr. K.S. Krishnan Road ew Delhi
P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features
P3000 Series Pneumatic Deadweight Testers P3000 Technical Data Features Pressure ranges from Vacuum through 2 000 psi (140 bar) Accuracy better than 0.015 % of reading. (Increased accuracy option of 0.008
Model F822 thru F834 Mulsifyre Directional Spray Nozzles, Open, High Velocity General Description
Worldwide Contacts www.tyco-fire.com Model F thru F3 Mulsifyre Directional Spray Nozzles, Open, High Velocity General Description The Mulsifyre Nozzles are open (nonautomatic) nozzles and they are designed
PHYSICAL QUANTITIES AND UNITS
1 PHYSICAL QUANTITIES AND UNITS Introduction Physics is the study of matter, its motion and the interaction between matter. Physics involves analysis of physical quantities, the interaction between them
Test Report issued under the responsibility of:
Test Report issued under the responsibility of: TEST REPORT EHPA-DACH Testing Regulation Supplemental requirements for granting the international quality label for heat pumps Testing of Air/Water Heat
Intercomparison in gauge pressure 0 25 MPa
J4/2010 Intercomparison in gauge pressure 0 25 MPa Markku Rantanen and Sari Saxholm Centre for Metrology and Accreditation Espoo 2010 Publication J4/2010 Intercomparison in gauge pressure 0-25 MPa Markku
Test equipment for temperature and its reduction
Data sheet 90.2721 Page 1/6 Test equipment for temperature and its reduction Pt100 precision RTD temperature probe Calibrated temperature measuring chain with precision display device Calibrations as service
(This report is endorsed) Industrivej 20, 9900 Frederikshavn, Danmark
Spectrum Laboratories Ltd is accredited by International Accreditation New Zealand (formerly Telarc). The tests reported herein have been performed in accordance with the terms of our accreditation. This
Heating oven with mechanical control
ED series 53 Drying ovens / Heating ovens Heating oven with mechanical control The strength of the BINDER ED series heating oven: routine drying and sterilization applications up to 300 C (572 F). Thanks
Physics 41, Winter 1998 Lab 1 - The Current Balance. Theory
Physics 41, Winter 1998 Lab 1 - The Current Balance Theory Consider a point at a perpendicular distance d from a long straight wire carrying a current I as shown in figure 1. If the wire is very long compared
Florida Method of Test for TESTING OF GROUND TIRE RUBBER Designation: FM 5-559
Florida Method of Test for TESTING OF GROUND TIRE RUBBER Designation: FM 5-559 1 SCOPE 1.1 This method is used to determine the physical requirements of ground tire rubber for use in asphalt rubber using
Check-list for auditors. Audit visit at candidate registered laboratories
Check-list for auditors Audit visit at candidate registered laboratories 1 Contents Introduction Laboratory documentation required by the standard(s), available for inspection by the auditor Items to be
Welcome to Physics 40!
Welcome to Physics 40! Physics for Scientists and Engineers Lab 1: Introduction to Measurement SI Quantities & Units In mechanics, three basic quantities are used Length, Mass, Time Will also use derived
AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL
FUEL CONSUMPTION MEASUREMENT AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL Description The combination of AVL Fuel Mass Flow Meter and AVL Fuel Temperature Control is a high precise and continuous
The Determination of Uncertainties in Charpy Impact Testing
Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials Code of Practice No. 06 The Determination of Uncertainties in Charpy Impact Testing M.A. Lont
Freezers - Ultrafreezers. Platilab & Irilab Series
Freezers - Ultrafreezers Platilab & Irilab Series PLATILAB: top security and performances with reliable temperature control. IRILAB: combining the high reliability of PLATILAB with the possibility to store
MATLAB AS A PROTOTYPING TOOL FOR HYDRONIC NETWORKS BALANCING
MATLAB AS A PROTOTYPING TOOL FOR HYDRONIC NETWORKS BALANCING J. Pekař, P. Trnka, V. Havlena* Abstract The objective of this note is to describe the prototyping stage of development of a system that is
NIST Calibration Services for Liquid Volume NIST Special Publication 250-72
NIST Calibration Services for Liquid olume NIST Special Publication 250-72 ern E. Bean, Pedro I. Espina, John. D. Wright, John F. Houser, Sherry D. Sheckels, and Aaron N. Johnson March 24, 2006 Fluid Metrology
Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control Solutions. Benefits.
GE Measurement & Control Solutions Sentinel LCT Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement Benefits Extremely reliable and highly accurate flowmeter, designed specifically
Installation Detailed installation instructions are provided which should be strictly followed.
Specification System 6 The Metal Technology Thermally-Broken Vertical Slide Window has been designed in such a way that its'attractive and clean lines will enhance all types of residential and commercial
Kinetic Friction. Experiment #13
Kinetic Friction Experiment #13 Joe Solution E01234567 Partner- Jane Answers PHY 221 Lab Instructor- Nathaniel Franklin Wednesday, 11 AM-1 PM Lecture Instructor Dr. Jacobs Abstract The purpose of this
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation
NITP 6.1 to 6.4 National Instrument Test Procedures for Non-automatic Weighing Instruments
NITP 6.1 to 6.4 National Instrument Test Procedures for Non-automatic Weighing Instruments Commonwealth of Australia 2012 NSC V 1 First edition 2000 Second edition October 2002 Third edition February 2004
Apr 17, 2000 LAB MANUAL 1811.0
Apr 17, 2000 LAB MANUAL 1811.0 1811 BULK SPECIFIC GRAVITY (GMB) AND DENSITY OF COMPACTED BITUMINOUS SPECIMENS USING PARAFFIN OR PARAFILM ASTM Designation D 1188 (MN/DOT Modified) 1811.1 SCOPE This test
Operating Principle of the Sample Positioning Flaps
Test WEIGHT, THICKNESS, DIAMETER (or LENGTH), and HARDNESS of regular tablets, odd shaped tablets (caplets, oblongs etc.) and other solid samples fully automatically. Use the WHT 3ME for the fully automated
E X P E R I M E N T 8
E X P E R I M E N T 8 Torque, Equilibrium & Center of Gravity Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 8:
Physics 121 Sample Common Exam 3 NOTE: ANSWERS ARE ON PAGE 6. Instructions: 1. In the formula F = qvxb:
Physics 121 Sample Common Exam 3 NOTE: ANSWERS ARE ON PAGE 6 Signature Name (Print): 4 Digit ID: Section: Instructions: Answer all questions 24 multiple choice questions. You may need to do some calculation.
INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS. Published: April 1999
INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS Published: April 1999 CONTENTS 1 Introduction 2 Test Procedures General 3 Standards and Specific Test Procedures
Rotational Motion: Moment of Inertia
Experiment 8 Rotational Motion: Moment of Inertia 8.1 Objectives Familiarize yourself with the concept of moment of inertia, I, which plays the same role in the description of the rotation of a rigid body
R&S ZCxxx Millimeter-Wave Converters Specifications
ZCxxx_dat-sw_en_3607-1471-22_v0200_cover.indd 1 Data Sheet 02.00 Test & Measurement R&S ZCxxx Millimeter-Wave Converters Specifications 21.07.2015 15:09:16 CONTENTS Definitions... 3 General information...
Sound Power Measurement
Sound Power Measurement A sound source will radiate different sound powers in different environments, especially at low frequencies when the wavelength is comparable to the size of the room 1. Fortunately
METHOD A10 (a) THE DETERMINATION OF THE IN-PLACE DRY DENSITY OF SOIL OR GRAVEL BY THE SAND REPLACEMENT METHOD
METHOD A10 (a) THE DETERMINATION OF THE IN-PLACE DRY DENSITY OF SOIL OR GRAVEL BY THE SAND REPLACEMENT METHOD 1 SCOPE The in-place dry density of compacted soil or gravel, as defined below, is determined
Calibration Service Selection Guide
Calibration Service Selection Guide Certification of Accreditation available for download at www.ricelake.com/accreditation. Answer Questions to determine the calibration service you need Need accredited*
(Equation 1) to determine the cord s characteristics. Hooke s Law represents the
Using Hooke s Law to Solve for Length of Bungee Cord Needed for Egg Drop Introduction This experiment is the second part of a three- part experiment. The first two lead up to the final in which we aim
75 Ω Transmission System
NRAO NTC-DSL Laboratory Report Dynamic Spectroscopy Laboratory Report Series Report 03 September, 2006 75 Ω Transmission System Chaitali R. Parashare Department of Electrical and Computer Engineering,
HAMBURG WHEEL-TRACKING TEST
Test Procedure for HAMBURG WHEEL-TRACKING TEST TxDOT Designation: Tex-242-F Effective Date: September 2014 1. SCOPE 1.1 This test method determines the premature failure susceptibility of bituminous mixtures
Installation information. METTLER TOLEDO MultiRange DW150 / DW150T wall scale
Installation information METTLER TOLEDO MultiRange DW150 / DW150T wall scale DW150 / DW150T wall scale Contents Contents Page 1. Installation... 2 1.1 Preparatory work... 2 1.2 Installing the wall plates
www.kem-kueppers.com [email protected]
www.kem-kueppers.com [email protected] Calibration of flow meters according to DIN EN ISO/IEC 17025:2005 Introduction... 2 Calibration laboratory... 3 Description of the calibration process... 3 Calibration
