Evaluation of headspace Solid Phase Micro-extraction method for analysis of phosphine residues in wheat
|
|
|
- Della Mitchell
- 10 years ago
- Views:
Transcription
1 Evaluation of headspace Solid Phase Micro-extraction method for analysis of phosphine residues in wheat Ren, Y.L. 1,2,3 *#, Padovan, B. 4 1 Cooperative Research Centre for National Plant Biosecurity, Australia, [email protected] 2 School of Biological Science and Biotechnology, Murdoch University, South Street, Murdoch, WA, 6150 Australia 3 Department of Agriculture and Food, Western Australia, 3 Baron Hay Court South Perth WA 6151, Australia 4 CSIRO Entomology, GPO Box 1700, Canberra ACT 2601, Australia * Corresponding author # Presenting author DOI: /jka Abstract This new method utilizes headspace-solid-phase micro extraction (HS-SPME) for pre-concentration of PH 3. Phosphine was determined with gas chromatography/pulsed flame photometric detector (PFPD). Spiked samples were used for calculation of phosphine residue in grain. Four types of fibres (100μm- PDMS, 85µm-CAR/PDMS, 75µm-CAR/PDMS and 65μm-PDMS/DVB) were tested. The bipolar fibres (CAR/PDMS and PDMS/DVB) can extract PH 3, but the non-polar fibre (PDMS) did not. Larger size fibres extracted PH 3 more efficiently than the smaller size fibres (e.g., 85 µm > 75 µm > 65 µm). The 85µm CAR/PDMS fibre was used to optimize the different parameters that affect the SPME extraction efficiency of PH 3. In the validation study, 50 grams of wheat in a 250 ml glass flask and capped with an open-top screw cap and PTFE/Silicon septa were spiked at 0.02 ng PH 3 /g of wheat. The flask was then heated to 45 C in an oil bath for 45 min, after which time the 85 µm CAR/PDMS fibre was exposed for 20 min and then exposed in the heated injection port of a GC/PFPD and desorbed for 2 min. Under conditions of the validation study, the limit of detection (LOD) or level of quantification (LOQ) was in the range of ng PH 3 /g of wheat. Keywords: Fumigant, Phosphine, Residue, SPME, HS-SPME 1. Introduction The world grain industry relies heavily on chemicals for grain treatment. Currently, phosphine (PH 3 ) is the only fumigant available to treat bulk grains and oil seeds (more than 85% grains are treated/re-treated with PH 3 ) in each of the linkages from on-farm storage to central storage (Collins et al., 2000; Collins et al., 2002; Ren and Mahon, 2007). However, a restrictive Codex Maximum Residue Limit (MRL) of 0.1 mg/kg, and lower limits set by some purchasers, are challenging the use of PH 3 for treatment of grain. Phosphine residues can remain in the grain for several months. Multi-fumigation with PH 3 can also cause accumulation of residues in the grain exceeding the MRLs. It is also important to note that levels of PH 3 residues in grain, producing low aerial concentrations of PH 3, can also facilitate resistance. Therefore, it is necessary to understand PH 3 residues in bulk grain to establish better procedures for multi-fumigation. This will guide industry in the conduct of good PH 3 fumigation practice to minimise PH 3 residue in fumigated grain. These include application methods and choice of the right periods of exposure and airing. Typical procedures for analysis of PH 3 residues include: 1) removal from the commodity matrix by either purge and trap techniques or by solvent extraction (Miyahara and Saito, 1994), and 2) and then analysis by gas chromatography (GC) or gas chromatography-mass spectra (GC-MS). However, purge and trap methods are not suitable for highly volatile fumigants such as PH 3 as they are unable to trap all fumigant (Ren and Desmarchelier, 1998). Solvent extraction has the problem of solvent interference and is timeconsuming. None of the above methods are suitable for analysis of PH 3 residues in grains, particularly at very low levels. Microwave irradiation is being increasingly used in the digestion of samples (Ren, 2001; Desmarchelier et al., 1998), and, in recent work (Chemfate 2004), excellent recoveries and precision have been obtained from microwave extractions of PH 3 from wheat and hay (Ren and Mahon, 2007). However, care is required in selecting the appropriate power setting and the safety implications of heating sealed flasks in microwave ovens is of concern, and water vapour generated from microwave irradiation can significantly interfere with GC results (Ren and Mahon, 2007). Julius-Kühn-Archiv, 425,
2 Solid-phase microextraction (SPME) in combination with head space (SH-SPME) analysis by gas chromatography is a convenient alternative method for volatiles. Solid-phase micro extraction is a simple, sensitive, and solvent-free technique that has become popular in a wide range of applications (Penalver et al., 2001). It has been used to study pollutant gases and volatile degradation products (Lattuai-Derieux, et al., 2004). For example residues of methyl isothiocyante (MITC) were analysed by HS-SPME with the 85 µm Polyacrylate (PA) fibre (Ren et al., 2008a). Using the HS-SPME method, MITC residues from wheat can be successfully analysed at levels below 0.1 ng/g, compared with purge and trap methods where the limit of detection is 10 ng/g (Ren et al., 2008a,b). This paper describes a new method for analysis of PH 3 residues in wheat which utilises headspace-solid-phase micro extraction (HS- SPME). It includes selection of the SPME fibre and factors that affect the extraction efficacy of PH Materials and methods 2.1. Reagents and apparatus One litre Erlenmeyer flasks (Bibby Sterilin, Staffordshire, Cat. No. FE 1 L/3) were used for fumigation of wheat and preparation of standards. The measured volume of each Erlenmeyer flask and inlet system was calculated from the weight of water required to fill the container and was used for calculations. Bottles of 250 ml (Alltech Cat. No. 9535) were used for the microwave extraction. Each bottle was fitted with a Mininert valve equipped with septa (Alltech Cat. No ). A 100 μl air tight syringe with valve (SGE, Melbourne, Australia; Cat. No ) was used for GOW-MAC gas density balance injection and transfer of PH 3 from source to fumigation chambers and flasks. The SPME fibres used were coated with 85 µm Carboxen/Polydimethylsiloxane (CAR/PDMS) fibre (Sigma-Aldrich Australia, Cat U), 75 µm Carboxen/Polydimethylsiloxane (CAR/PDMS) fibre (Sigma-Aldrich Australia, Cat U), 65μm Polydimethylsiloxane/Divinylbenzene (PDMS/DVB) fibre (Sigma-Aldrich Australia, Cat U) and 100 μm Polydimethylsiloxane (PDMS) fibre (Sigma- Aldrich Australia, Cat ), respectively. All fibres were conditioned at 270 C for 1 h prior to use in accordance with manufacturers recommendations. Phosphine (85.0% PH 3 and 15.0% air and CO 2 ) was laboratory prepared by the FAO method (FAO, 1975). The purity of PH 3 was determined using a GOW-MAC gas density balance (GOW-MAC Instrument Co., Madison, N.J.) after separation of the gases on a 1 m 5 mm i.d. Porapak Q 100/120 mesh (Alltech Associates, Cat. No. 2702) column at 105 C with a carrier flow (N 2 ) of 150 ml/min. The reference gas used was tetrafluoroethane (> 99.9%), which was supplied by ACTROL Ltd, Australia. Phosphine (PH 3 ) was determined on a Varian CP-3800 (Varian Instruments, Sunnyvale, CA), equipped with a pulsed flame photometric detector (PFPD) with phosphorus filter. Separation was achieved on a 30 m 0.53 mm ID, AT-Q column (Alltech Associates, Cat. No , Part No ) at 125 C and carrier flow (N 2 ) of 5.0 ml/min at 5.0 psi. Injector and detector temperatures were 200 C. Injection volumes of gases were 40 µl. A minimum interval of 5 min was kept between injections, in order to elute interfering chemicals Wheat sample and fumigation of wheat Wheat used was Australian standard white wheat, 10.9% moisture content, w/w wet basis. The moisture content of the wheat was measured by oven drying at 105 C for 2 hours (ISO 712 International Standard 1998). In order to ensure the moisture result is precise, 500 kg wheat sample was quartered equally using a Boerner grain divider. One quarter sample (125 g) was quartered equally again and then one quarter sample of about 30 g was collected and ground with a laboratory sealed metal grinder for measurement of moisture content. Four replicates samples were dried at 105 C in mechanical convection oven ( DK 62 American Scientific Products Columbus, OH, USA) and the loss of weight was used to calculate the moisture content of the sample. Wheat samples (860 g) were fumigated in an Erlenmeyer flask (1 L) equipped with a lid fitted a septum injection system for 7 days at concentrations of 0.5 mg (PH 3 )/L. After 7 days exposure, the flask was opened and aired for 7 days in a fume hood to obtain samples containing residual fumigant. 408 Julius-Kühn-Archiv, 425, 2010
3 2.3. Preparation of diluted PH 3 gas and spiking standards Diluted PH 3 gases were prepared by first removing the same volume of air as the known volume (0.5 ml) of concentrated PH 3 to be injected into an Erlenmeyer flask (1 L) containing six glass beads (2-3 mm OD). For validation study, the limit of detection (LOD) or level of quantification (LOQ) was investigated using spiked standards. Spiked wheat samples at and 0.02 and 0.5 ng PH 3 /g of wheat were prepared by adding appropriate volumes of PH 3 into sealed flasks (250 ml) containing wheat (50 g) 3 min before oil bath heating. Triplicate samples were used and each sample was injected duplicates (n=6). The dosages and required volumes for PH 3 concentrations, calibrated to the current laboratory temperature and pressure, were calculated from Eq. 1 Where: 4 T C V V f = (1 ) ( ) Eq P M N V is volume of fumigation container (L); P is pressure (mm Hg); T is temperature ( C); C is the intended concentration of methyl bromide (mg/l); V f is dosage volume of fumigant (ml); M is molecule weight of fumigant; N is purity of gas (%) Selection of fibre and HS-SPME extraction time Four different types of fibres (100 μm-pdms, 85 µm-car/pdms, 75 µm-car/pdms and 65 μm- PDMS/DVB) were tested at 20 min extraction time. Diluted PH 3 gas was injected into a 250 ml flask fitted with a sample port to obtain 0.3, 0.1, 0.05 and 0.01 ppm of PH 3, respectively. Headspace samplings were carried out on all four types of fibres. The needle was carefully inserted into the headspace of the flask and the fibre exposed into the headspace for 20 min. At the end of the defined extraction time, the fibre was withdrawn from the headspace into the needle. The fibre holder was removed from the extraction flask and inserted into the injection port. The fibre was extended into a GC-PFPD inlet where sample components were desorbed at 200 C for 5 min to clean it between extractions. For evaluation of HS-SPME extraction time on the efficacy of PH 3 extraction, the SPME equilibration time was determined by exposing the fibres (85 µm-car/pdms, 75 µm-car/pdms and 65 μm- PDMS/DVB) to 0.3 ppm of PH 3 for 1, 5, 10, 15, 20 and 30 min, respectively. Finally the fibre was retracted and then exposed in the heated injection port (200 C) of a GC-PFPD and desorbed for 2 min. Results are the mean of duplicate samples and injections for each sample (n=4) Evaluation of GC Injector temperature for desorption of PH 3 from fibre (85µm CAR/PDMS) Efficient thermal desorption of the analysis in a GC injection port is dependent on the type of fibres and on the GC injector temperature. The 85 µm CAR/PDMS fibre was used to optimize the injector temperature. It was inserted into the headspace of the flask containing 0.1 ppm of PH 3 and exposed for 20 min. The fibre was carefully injected into the GC at different injector temperatures (100,150, 200 and 250 C) and desorbed for 5 min. Results are the mean of duplicate samples and injections for each sample (n=4) Evaluation of wheat treatment and HS-SPME extraction temperature for extraction efficacy of PH 3 A fumigated wheat sample of 50 g in a 250 ml flask fitted with a sample port was immersed in a heated oil bath at 30 C for 30 min, after which time the fibre was inserted into the headspace of the flask and exposed for 20 min. Equivalent flasks were heated a 35, 40, 45 and 50 C. The fibre was injected (200 C) into a GC-PFPD and desorbed for 5 min. Spiked wheat samples were treated in exactly the same manner. Levels of PH 3 residue were determined against spiked standards. The peak areas were calibrated periodically using a fortified standard, and the data presented are the mean of duplicate samples and injections for each sample (n=4). Julius-Kühn-Archiv, 425,
4 2.7. Statistical analysis The GC readings of fumigant concentration were averaged respectively. The average variation of GC readings and PH 3 concentration and standard deviations (SD) between the duplicate treatments and injections were analysed with Microsoft Excel. 3. Result and discussion 3.1. Effect of fibre types and HS-SPME extraction time on extraction efficacy of PH 3 The amounts of PH 3 absorbed by each of four different SPME fibres are shown in Figure 1 for each of four PH 3 concentrations. These four fibres showed different efficiency and selectivity in extracting PH 3 in the headspace. The bipolar fibres (CAR/PDMS and PDMS/DVB) which dispersed solid adsorbents in polymer extracted PH 3, but the nonpolar fibres (PDMS) extracted no PH 3. The larger size of polymer fibre (e.g., 85 µm) extracted PH 3 more efficiently than the smaller size of polymer fibre (e.g., 75 µm and 65 µm). Figure1 Fibre types (100 µm PDMS, 85 µm CAR/PDMS, 75 µm CAR/PDMS and 65 µm PDMS/DVB) affect the extraction efficacy of PH 3 after 20 minutes extraction. The effect of time of extraction of PH 3 is shown in Figure 2. The responses of GC peak areas progressively increased with increasing the extraction time, but no increase in the response occurred after 20 min. That is, the amount of PH 3 absorbed by the tested fibres increased over a period of 20 min and then attained equilibrium. After 30 min extraction, the amount of absorption was 85 µm CAR/PDMS > 75 µm CAR/PDMS > 65 µm PDMS/DVB. The 85 µm CAR/PDMS fibre was used in this study to optimize the different parameters that affect the SPME extraction efficiency of PH Julius-Kühn-Archiv, 425, 2010
5 Figure 2 Time of extraction with different type of fibres (85 µm CAR/PDMS, 75 µm CAR/PDMS and 65 µm PDMS/DVB) affects the extraction efficacy of PH 3 (0.3 ppm) at 1, 5, 10, 15, 20 and 30 min Effect of GC Injector temperature for desorption of PH 3 from fibre (85 µm CAR/PDMS) The effects of GC injector or desorption temperature were evaluated by varying the temperature from 100 to 250ºC. The injector temperature profile obtained is shown in Figure 3. The amount of desorbed PH 3 increased with increasing the injector temperature and then reached a maximum at 200ºC. The GC used had split/splitless capillary injectors. It was suitable for direct introduction of the fibre into the injection port. In this study, the CG injector operated at 200ºC of optimal desorption temperature and did not cause peak broadening and tailing (Fig. 4) GC readings (m Vols) GC injector temperature (ºC) Figure 3 The GC injector temperature affects desorption of PH 3 from fibre (85 µm CAR/PDMS) at 0.1 ppm of PH 3 after 20 minutes extraction. Julius-Kühn-Archiv, 425,
6 Figure 4 GC spectra with fibre (85 um CAR/PDMS) after 20 min extraction at levels of PH 3 at 0.5 ng/g (A) and 0.02 ng/g (B) Effect of wheat treatment and HS-SPME extraction temperature on extraction efficacy of PH 3 The effects of temperature for treatment of wheat sample and extraction efficacy were evaluated by varying the temperature from 30 to 50ºC. The temperature profile for PH 3 released from fumigated wheat into the headspace and absorption of PH 3 on the fibre (85 µm CAR/PDMS) obtained is shown in Figure GC readings (m Vols) Extraction temperature (ºC) Figure 5 Temperature of extraction with fibre (85 µm CAR/PDMS) for 20 minutes extraction affects the extraction efficacy of PH 3. The amount of absorbed PH 3 increased with increasing the injector temperature and then reached a maximum at 45ºC. As the temperature increases, the liquid and gas phases equilibrate faster and drive more PH 3 partitioning into the headspace, but increasing fibre temperature causes a competitive effect of 412 Julius-Kühn-Archiv, 425, 2010
7 desorption from the fibre, ultimately limiting its analytical sensitivity of PH 3. This is consistent with the results of Vás and Vékey (2004). Therefore, we used the low temperature (45ºC) and long term (30 min) extraction for treatment of wheat samples. The limit of detection (LOD) or level of quantification (LOQ) was estimated based on a signal to noise ratio shown in Figure 5 (B) in the range of ng/g. Although the determination of PH 3 in grain is well established in the scientific literature, our results have demonstrated that the combination of SPME with headspace sampling is very effective for determining trace levels of PH 3. Acknowledgements The authors would like to acknowledge the support the Australian Government s Cooperative Research Centres Program and Jim Desmarchelier for helpful contributions. References Chemfate, Environmental Fate Database. Syracuse Research Corporation, Syracuse, NY. USA. accessed June 2004www.syrres.com/esc/chemfate.htm. Collins, P.J., Daglish, G.J., Pavic, H., Lambkin, T.M., Kopittke, R Combating strong resistance to phosphine in stored grain pests in Australia. In: Wright, E.J., Banks, H.J., Highley, E. (Eds), Stored Grain in Australia. Proceedings of the Australian Postharvest Technical Conference, 1-4 August 2000, Adelaide, Australia, pp Collins, P.J., Emery, R.N., Wallbank, B.E Two decades of monitoring and managing phosphine resistance in Australia. In: Credland, P.F.A., Armitage, D.M., Bell, C.H., Cogan, P.M., Highley, E. (Eds), Proceedings of the Eighth, International Working Conference on Stored-product Protection, July 2002, York, UK, CAB International, Wallingford, UK, pp Desmarchelier, J. M., Allen, S. E., Ren, Y. L., Modifications of a method for determining multifumigant residues. Journal of the Association of Official Analytical Chemists, International 81, FAO, Recommended methods for the detection and measurement of resistance of agricultural pests to pesticides. Tentative method for adults of some major pest species of stored cereals, with methyl bromide and phosphine. FAO Method NO. 16, 1975, Plant Protection Bulletin No. 23, Vás, G., Vékey, K., Solid-phase microextraction: a powerful sample preparation tool prior to mass spectrometric analysis. Journal of Mass Spectrometry, 2004, 39, Miyahara, M., Saito, Y., Determination of bromide ions in food by unsuppressed ion chromatography with ultraviolet detection after microwave digestion in a sealed PTHE vessel. Journal of Agricultural and Food Chemistry 42, Penalver, A., Pocurull, E., Aguilar, C., Borrull, F., Marce, R. M., Comparison of different fibres in the solid phase microextraction of phthalate esters from water samples. Journal of Chromatography A 922, Ren, Y. L., Comparison of solvent extraction and microwave extraction for release of dimethyl sulfide from cereals and canola. Journal of Agricultural and food Chemistry 49, Ren, Y.L., Desmarchelier, J.M., Release of Fumigant Residue from Grain by Microwave Irradiation. Journal of the Association of Official Analytical Chemists, International 81, Ren, Y.L., Mahon, D., Evaluation of microwave irradiation for analysis of carbonyl sulfide, carbon disulfide, cyanogen, ethyl formate, methyl bromide, sulfuryl fluoride, propylene oxide, and phosphine in hay. Journal of Agricultural and Food Chemistry 55, Ren, Y.L., van Emmerik, T., Mahon, D., Lee, B.H., Padovan, B., 2008a. Ethyl formate plus methyl isothiocyanate is a potential liquid fumigant for stored grains. In: Guo, D.L., Navarro, S., Jian, Y., Cheng, T., Zuxun, J., Yue, L., Yang, L., Haipeng, W. (Eds), Proceedings of the 8th International Conference on Controlled Atmospheres and Fumigation in Stored Products, September 2008, Chengdu, China, Sichuan Publishing House of Science and Technology, Chengdu, China, pp Ren, Y.L,, Lee, B.H., Mahon, D.A., Xin, N., Head, M., Reid, R., 2008b. Fumigation of wheat using liquid ethyl formate plus methyl isothiocyanate in 50 tonne farm bins. Journal of Economic Entomology 101, Julius-Kühn-Archiv, 425,
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D.
Gas Chromatograph Mass Spectrometer No. GCMS-1603 Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS Shilpi Chopra, Ph.D. Introduction Organophosphorus (OP) pesticides are a class
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
SUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
POLYVINYL ALCOHOL. SYNONYMS Vinyl alcohol polymer, PVOH, INS No. 1203 DEFINITION DESCRIPTION FUNCTIONAL USES CHARACTERISTICS
POLYVINYL ALCOHOL Prepared at the 68 th JECFA (2007) and published in FAO JECFA Monographs 4 (2007), superseding specifications prepared at the 63 rd JECFA (2004) and published in the Combined Compendium
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*
Method 552.1 Revision 1.0 Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* UCT Products: EUQAX156 (quaternary amine with Cl - counter ion, 6 ml cartridge)** CLTTP050 (CLEAN-THRU
ß-CYCLODEXTRIN SYNONYMS
ß-CYCLODEXTRIN Prepared at the 44th JECFA (1995), published in FNP 52 Add 3 (1995) superseding specifications prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications
With proper calibration, the upper limit may be extended as needed.
While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the
Gas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
Expectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
Split / Splitless Injection for Capillary GC
Split / Splitless Injection for Capillary GC Speakers John V Hinshaw CHROMacademy GC Dean GC Connections editor Moderator Tony Taylor CHROMacademy Trainer Technical Director, Crawford Scientific Dave Walsh
Aromatic plants: from extraction to analysis
Aromatic plants: from extraction to analysis Thierry TALOU Agro-industrial Chemistry Laboratory National Polytechnic Institute of Toulouse ENSIACET 4 allée Emile Monso 31432 Toulouse Olfactory system Olfactory
Ethyl formate fumigation as a replacement for methyl bromide fumigation in dried vine fruit.
Ethyl formate fumigation as a replacement for methyl bromide fumigation in dried vine fruit. Dried Fruit Research and Development Council, Project CSH 64D C.R. Tarr 1, R. Reuss 2, P.R.Clingelefer 1, P.
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
Purification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory
Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate
Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
Automation of Solid Phase Extraction and Column Chromatographic Cleanup
Automation of Solid Phase Extraction and Column Chromatographic Cleanup Haibin Wan Ltd. Problems with Manual Operation 1. Unstable flow rate affects reproducibility and throughput. 2. Not convenient for
Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup
Application Note 25570306 Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup Keywords Gel Permeation Chromatography GPC AutoPrep 2000 Pesticides RapidVap N2 System
GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits
Issue Implementation Page 1 of 7 GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits Scope and Application This method will determine the level of vanillin and ethyl vanillin in distilled
Dissolved and precipitated oxalate
Accepted 2005 Process liquors from bleach plants Dissolved and precipitated oxalate Using Ion Chromatography 0 Introduction In bleach plants of pulp mills with a high degree of system closure, there is
Chapter 28: High-Performance Liquid Chromatography (HPLC)
Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion
ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC)
ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC) 1 Fast Analysis of C1 to C8 hydrocarbons for mud logging applications within 2 minutes using a temperature programmable Micro GC Fusion NEN GAS2015 Debbie
Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles
THERMAL DESORPTION Introduction and Principles Thermal Desorption Thermal desorption is a simple extension of the technique of gas chromatography (GC) It involves the use of heat and a flow of inert gas
105 Adopted: 27.07.95
105 Adopted: 27.07.95 OECD GUIDELINE FOR THE TESTING OF CHEMICALS Adopted by the Council on 27 th July 1995 Water Solubility INTRODUCTION 1. This guideline is a revised version of the original Guideline
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
Analyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water
A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water Sjaak de Koning 1, Mark van Lieshout 2, Hans-Gerd Janssen 2 and Wil van Egmond 1 1 ATAS, De Run 4334, PO Box
MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process
- Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
Version 1.0 September, 2008. Jean W. Munch and Paul E. Grimmett
EPA/600/R-08/101 METHOD 522 DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/ MASS SPECTROMETRY (GC/MS) WITH SELECTED ION MONITORING (SIM) Version 1.0
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2
Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall
Analysis of Liquid Samples on the Agilent GC-MS
Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps
Application Note: ANTV10 Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps Abstract In order to regulate product safety and quality
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine
Air Sampling for Gases and Vapors. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike License. Your use of this material constitutes acceptance of that license and the conditions of use of materials on this
CARBON DIOXIDE CONCENTRATION IN HERMETIC STORAGE OF SOYBEAN (GLYCINE MAX) IN SMALL GLASS JARS
Ochandio D, Bartosik R, Yommi A, Cardoso L (2012) Carbon dioxide concentration in hermetic storage of soybean (Glycine max) in small glass jars. In: Navarro S, Banks HJ, Jayas DS, Bell CH, Noyes RT, Ferizli
HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.
HEXANES Prepared at the 51st JECFA (1998), published in FNP 52 Add 6 (1998) superseding specifications prepared at the 14th JECFA (1970), published in NMRS 48B (1971) and in FNP 52 (1992). ADI "limited
Enhancing Utility and Throughput of SPME using Advanced Automation Strategies
Supelco Innovation Seminars November 11 th, 2010 Atlanta, GA Enhancing Utility and Throughput of SPME using Advanced Automation Strategies Presented by Tom Tobien, Ph.D. Analytical Chemist, LEAP Technologies,
ORGANIC SAMPLE PREPARATION
ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,
CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA 91342 ABSTRACT
CRYOCOOLER CONTAMINATION STUDY S.W.K. Yuan, D.T.Kuo, and A.S. Loc EI Technologies, Cryocooler Group Sylmar, CA 91342 ASTRACT Eliminating or reduction of contamination is essential for the success of cryocoolers
Vitamin C quantification using reversed-phase ion-pairing HPLC
Vitamin C quantification using reversed-phase ion-pairing HPLC Thomas Grindberg and Kristy Williams Department of Chemistry, Concordia College, 901 8 th St S, Moorhead, MN 56562 Abstract Vitamin C, an
SCION SQ Series GC-MS
SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative
EnAlgae SOP: Biogas analysis
Table of content 1. PURPOSE... 2 2. PRINCIPLE... 2 3. REQUIREMENTS... 3 3.1. EQUIPMENT AND MATERIALS... 3 3.2. REAGENTS... 3 3.3. PREPARATION OF MEASUREMENT... 3 4. PROCEDURE... 4 4.1. SAMPLE MEASUREMENT...
for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)
for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) IonSwift MAC-100 & MAC-200 Product Manual Page 1 of 16 PRODUCT MANUAL for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) MAC-100, 0.5 x 80mm (P/N 074702)
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase
α-cyclodextrin New specifications prepared at the 57th JECFA (2001) and published in FNP 52 Add 9 (2001). An ADI not specified was established at the 57th JECFA (2001). SYNONYMS α-schardinger dextrin,
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
Bologna 12-13 Maggio 2011; 1 MS Environmental Day
; 1 MS Environmental Day E. Salvadori, A. Falai, M. Cappelli, S. Dugheri, A. Bonacchi, G. Arcangeli, V. Cupelli, D. Santianni, D. Burrini ISPRA reports: 50% of monitoring points of surface water are contaminated
TANNIC ACID. SYNONYMS Tannins (food grade), gallotannic acid, INS No. 181 DEFINITION DESCRIPTION
TANNIC ACID Prepared at the 39th JECFA (1992), published in FNP Add 1 (1992) superseding specifications prepared at the 35th JECFA (1989), published in FNP 49 (1990) and in FNP 52 (1992). Metals and arsenic
A new concept for isotope ratio monitoring LC/MS. A Wide Range of Applications
A new concept for isotope ratio monitoring LC/MS A Wide Range of Applications Andreas Hilkert, Dieter Juchelka, Michael Krummen Thermo Electron (Bremen) GmbH Overview Introduction in Isotope Ratio Monitoring-LC/MS
Novel Method for Solid Phase Peptide Synthesis Using Microwave Energy
Novel Method for Solid Phase Peptide Synthesis Using Microwave Energy Jonathan M. Collins, Michael J. Collins, Rebecca C. Steorts CEM Corporation, Matthews, NC 28106-0200, U.S.A. Presented at American
MILESTONE H E L P I N G C H E M I S T S
MILESTONE H E L P I N G C H E M I S T S 8 good reasons to choose the Milestone DMA-80 1. No sample preparation The DMA-80 does not require any sample preparation or other wet chemistry prior the analysis.
PLOT Columns. Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative
PLOT Columns Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative PLOT Column Selection 93 Features & Benefits of Restek PLOT Columns Features
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic
Multi-Shot Pyrolyzer EGA/PY-3030D. Flexible Versatile Reproducible
Multi-Shot Pyrolyzer EGA/PY-3030D Flexible Versatile Reproducible Why Evolved Gas Analysis? Why Pyrolysis? Today, analytical pyrolysis encompasses much more than simple flash pyrolysis of polymeric materials.
(S&G 5th ed. Expt 27, 6th, 7th & 8th eds. Expt 26)
Chem 367-2/ Physical Adsorption 63 PHYSICAL ADSORPTION (S&G 5th ed. Expt 27, 6th, 7th & 8th eds. Expt 26) The sample used in this experiment needs to be conditioned for at least two hours before the experiment;
Gas Chromatograph GC Consumables. p/n: 220-94898-02
Gas Chromatograph GC Consumables p/n: 220-94898-02 Diag. # 1) Septum Nut 2) Needle Guide 3) Septum 4) Injection Port A Assy 5) O-ring for Glass Insert 6) Glass Insert 7) Gold Gasket 8) Capillary Adaptor
Determination of Calcium and Magnesium in Brine
Application Note Determination of Calcium and Magnesium in Brine INTRODUCTION To prevent membrane poisoning, new membrane technology in chlor-alkali cells requires feed brine that is relatively free of
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY. Andrew Tipler
AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY Andrew Tipler Chromatography Research and Technology Manager PerkinElmer, Inc. Table of Contents Introduction 3 Fundamental
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:
Identification of Arson Accelerants by Gas Chromatography
Identification of Arson Accelerants by Gas Chromatography Purpose Arson is an insidious crime that annually claims the lives of hundreds of Americans and costs billions of dollars. 1 Scientists and criminal
The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection
Application Note 9 The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection INTRODUCTION The accurate measurement of the amount of sugar in final molasses
Micro Cryo-Trap Model 971
Micro Cryo-Trap Model 971 for Liquid CO2 Cooling Scientific Instrument Services, Inc. 1027 Old York Road Ringoes, NJ 08551 Phone (908) 788-5550 Part #951001M revised 10/ /98 Notice The information in this
Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT
AppNote 6/2003 Analysis of Flavors using a Mass Spectral Based Chemical Sensor Vanessa R. Kinton Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA Kevin L. Goodner USDA, Citrus & Subtropical
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen
Development and Validation of Gas Chromatographic Analytical Method for Estimation of Residual Solvents in Excipients (Klucell, Triacetan, HPMC)
Research Article Development and Validation of Gas Chromatographic Analytical Method for Estimation of Residual Solvents in Excipients (Klucell, Triacetan, HPMC) Rohitkumar M.Yadav* and Kapil Patil Analytical
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
On-line Dissolved Gas Analysis
On-line Dissolved Gas Analysis accord. to IEC 567/ASTM 3612 Online-Monitoring of transformers Automatic on-line analysis of 11 gases The analysis is done in 2 steps with high selectivity and accuracy fully
SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322. [email protected].
SPE and HPLC Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322 [email protected] Solid-Phase Extraction- SPE Simple, fast and efficient sample preparation
23 The Thermal Conductivity Detector
23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing
6. ANALYTICAL METHODS
45 6.1 BIOLOGICAL MATERIALS No methods were located that are routinely used for the detection of BCEE in biological materials. Norpoth et al. (1986) reported a method for measuring thiodiglycolic acid
Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)
Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction
Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids
Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C
Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder
Lecture Chromo-3: Gas Chromatography CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder Outline Introduction Instrument overview Carrier gas Sample
This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA
An Automated Method for the Fractionation of Extractable Petroleum Hydrocarbons (EPH) from Water and Soil Extracts Using the Gilson GX-274 ASPEC System Keywords: Aliphatic Hydrocarbons, Aromatic Hydrocarbons,
