Analytical Method. Determination of metals in workplace air by inductively coupled plasma mass spectrometry (ICP-MS) ANALYTICAL METHOD 362
|
|
|
- Mercy Shaw
- 10 years ago
- Views:
Transcription
1 Analytical Method Determination of metals in workplace air by inductively coupled plasma mass spectrometry (ICP-MS) ANALYTICAL METHOD 362 Applicability This method applies to the determination of several metals and their compounds in workplace air. Standard See Regulation respecting occupational health and safety (ROHS). Sampling system MCE FILTER 0,8μm, 25 and 37mm. Recommended sampling flow rate and volume 180L at 1,5 L/min. Analytical technique Inductively. Minimum reported value (MRV) See Appendix 1. Working range See Appendix 1. Reliability See Appendix 1. Analytical uncertainty (CV A et CV E ) See Appendix 1.
2 Established in Québec since 1980, the Institut de recherche Robert-Sauvé en santé et en sécurité du travail (IRSST) is a scientific research organization known for the quality of its work and the expertise of its personnel. OUR RESEARCH is working for you! Mission To contribute, through research, to the prevention of industrial accidents and occupational diseases as well as to the rehabilitation of affected workers. To offer the laboratory services and expertise necessary for the activities of the public occupational health and safety prevention network. To disseminate knowledge, and to act as scientific benchmark and expert. Funded by the Commission de la santé et de la sécurité du travail, the IRSST has a board of directors made up of an equal number of employer and worker representatives. To find out more Visit our Web site for complete up-to-date information about the IRSST. All our publications can be downloaded at no charge. To obtain the latest information on the research carried out or funded by the IRSST, subscribe to Prévention au travail, the free magazine published jointly by the IRSST and the CSST. Subscription: Legal Deposit Bibliothèque et Archives nationales 2010 ISBN: (PDF) ISSN: IRSST Communications Division 505, De Maisonneuve Blvd West Montréal (Québec) H3A 3C2 Phone: Fax: [email protected] Institut de recherche Robert-Sauvé en santé et en sécurité du travail, 2010
3 Analytical Method Determination of metals in workplace air by inductively coupled plasma mass spectrometry (ICP-MS) ANALYTICAL METHOD 362 Disclaimer The IRSST makes no guarantee regarding the accuracy, reliability or completeness of the information contained in this document. In no case shall the IRSST be held responsible for any physical or psychological injury or material damage resulting from the use of this information. Note that the content of the documents is protected by Canadian intellectual property legislation. Analytical or calibration methods are the ones developed or chosen by the IRSST to carry out its different mandates. They may require the use of hazardous materials, operations or equipment. These methods do not mention all the safety problems related to their use. It is the user s responsibility to establish the appropriate health and safety practices. Use of the data included in these methods is at the user s own risk: the IRSST is not responsible for any errors or damage that may result from such use and application. The hyperlinks that appear in this document were validated at the time of publication. Technical person in charge of the method Pierre Larivière, M.Sc. chemist and Marie-France Bisson, technician, Laboratory Services and Expertise, IRSST Approval Pierre Larivière, M.Sc. chemist Laboratory Services and Expertise, IRSST Marie-Claude Barrette, M.Sc. chemist coordinator of the Quality Assurance Program, Laboratory Services and Expertise, IRSST Jacques Lesage, M.Sc. chemist director, Laboratory Services and Expertise, IRSST Authorization for publication Marie Larue, M.Sc. chemist president and CEO, Executive Office, IRSST This document is also available on the IRSST s Web site at the following address: This study was financed by the IRSST. The conclusions and recommendations are those of the authors.
4 IN CONFORMITY WITH THE IRSST S POLICIES The results of the research work published in this document have been peer-reviewed.
5 TABLE OF CONTENTS Preamble Working range Principle of the method Interferences Material Reagents Sampling Analytical protocol Analytical solutions Preparation of samples Verification of the ICP-MS performance criteria Calibration Quality control Analysis Calculations Performance parameters Limit of detection, limit of quantification and minimum reported value (MRV) Recovery Reliability Accuracy Measurement uncertainty References...16
6 Preamble Analytical method 362 Determination of metals in workplace air by inductively Page 1 The aim of the Act respecting occupational health and safety in Québec is to eliminate, at source, hazards to the health, safety and physical well-being of workers. Permissible exposure values (PEVs) for chemical substances have been established in Schedule I of the Regulation respecting occupational health and safety (ROHS). Section 44 of this regulation entitled "Methods" specifies that: " These dusts, gases, fumes, vapours and mists found in the workplace environment shall be sampled and analyzed to obtain an accuracy equivalent to that obtained by applying the methods described in the Sampling Guide for Air Contaminants published by the Institut de recherche Robert- Sauvé en santé et sécurité du travail du Québec..." To achieve these objectives, analytical methods for quantifying the workers' degree of exposure are developed and written to establish appropriate means of control. In order to help health and safety professionals in workplaces, the IRSST publishes, periodically revises, and disseminates the Sampling Guide for Air Contaminants in the Workplace and the Laboratory Services and Expertise Department publishes contaminant analysis methods. These methods must be used in combination with the following regulatory and normative references: Act respecting occupational health and safety. R.S.Q., chapter S-2.1. Éditeur officiel du Québec, (August 1, 2007). _2_1%2FS2_1_A.htm Regulation respecting occupational health and safety. S-2.1, r.19.01, O.C Éditeur officiel du Québec (July 25, 2007). _2_1%2FS2_1R19_01.htm Sampling Guide for Air Contaminants in the Workplace. Operations Division, IRSST, T-015 Technical Guide, Montréal, Québec, (mars 2005). NIOSH, National Institute for Occupational Safety and Health. ISO Guide 30, Terms and definitions used in connection with reference materials, 2 nd edition, ISO, International Vocabulary of Basic and General Terms in Metrology, 2 nd edition, American Industrial Hygiene Association (AIHA), organization that certifies the IRSST laboratory in the field of workplace chemical and microbiological contaminant analysis.
7 1. WORKING RANGE Analytical method 362 Determination of metals in workplace air by inductively Page 2 This method applies to the determination of different metals present in particles suspended in the air, by means of inductively. The tests for method development were performed on a PerkinElmer Elan DRC II ICP-MS. The calibration range for each of the metals is within the limits presented in the table below. Elements Isotope used Lower limit Upper limit (µg/l) (mg/m³) 1 (µg/l) (mg/m³) 1 Aluminium 27 Al Cadmium 114 Cd Chromium 52 Cr Cobalt 59 Co Copper 63Cu Iron 57 Fe Magnesium 26 Mg Manganese 55 Mn Nickel 60Ni Lead Pb, 207 Pb & Pb Vanadium 51V Zinc 66 Zn Concentration for a final volume of 10 ml and a sampling volume of 180 L 2 The signals of the three isotopes are added, due to the natural variation in individual abundances The document also presents the method's procedure, the performance controls, and proposes a routine analysis sequence. 2. PRINCIPLE OF THE METHOD In accordance with the parameters described in the Sampling Guide for Air Contaminants in the Workplace, particles suspended in the air containing different metals and their compounds are collected by passing a volume of air through a mixed cellulose ester (MCE) filter. The collected filter and sample are then treated in such a manner as to dissolve the different metals and their compounds. The sample solution is then analyzed by ICP-MS. Since ICP-MS is a relative analysis technique, the proportionality factor between the response and the concentration of the element is determined by analyzing standard solutions. 3. INTERFERENCES There are two types of interferences, spectral and non-spectral. The most common spectral interferences are polyatomic (or molecular) interferences produced by the combination of two or more atomic ions having the same nominal mass/charge ratio as the isotope of interest.
8 Page 3 These ions are generally formed in the plasma or the interface system from vector gases or constituents of the sample. Isobaric interferences are caused by isotopes of different elements that form singly or doubly charged ions with the same nominal mass/charge ratio. Non-spectral interferences, generally called matrix effects, can originate from three different processes or locations: the nebulization process, the plasma, or the zone of the interface and the lens. This type of interference also includes plugging of the nebulizer, the injector tube of the torch and the cones (sampler and skimmer) due to high concentrations of dissolved salts. The different strategies used in this analytical method to reduce or eliminate interferences are: Application of mathematical correction equations; Use of a reaction cell; Reduction of the water content of the plasma in order to reduce the level of all types of interferences due to oxides. This reduction is achieved by lowering the temperature of the spray chamber by 2 C to 5 C by Peltier effect; Addition of an internal standard to the calibration solutions and to the samples as well as matrix matching of the calibration and samples solutions; and Optimization of the analytical conditions, in order to maximize the ratio of the analytical signal to the interference signal (plasma power, flow rate of sampling vector gas, sample collection rate, distance between the torch and the top of the sampling cone, rinsing time between two solutions, etc.) 4. MATERIAL NOTE 1 - It is recommended that PP (polypropylene) or PE (polyethylene) plasticware be used. NOTE 2 All reusable plasticware must be carefully washed before use and decontaminated by soaking in dilute nitric acid (for example: 10% HNO3) for at least 12 hours and then rinsed several times with demineralized water. NOTE 3 All disposable plasticware must previously be subjected to a control to determine the presence of the different elements to be determined.
9 Page 4 Felt marker; Single-use impervious gloves (powder-free), for example made of nitrile, to prevent any possibility of hand contamination and contact with corrosive substances; Digestion beakers, 50 ml, made of Teflon, for dissolving the samples; NOTE 4 Beakers are previously decontaminated by reflux heating with concentrated HNO3 at 120 C for 2 hours. Non-metallic tweezers, for removing the filters from the cassettes; Distributors for the acids; Stirring hot plate, thermostatically controlled, that can maintain a surface temperature of at least 200 C; NOTE 5 The surface temperature may vary considerably, depending on the position on the hot plate. It may therefore be useful to characterize the hot plate's performances before it is used. Wash bottle; Piston stroke volumetric pipettes with disposable tips, for preparing calibration solutions, calibration control solutions, and for diluting sample solutions; Volumetric flasks of appropriate volumes with corresponding stoppers; Volumetric test tubes, single use, compatible with the autosampler; Dilution test tubes, single use, compatible with the autosampler; Syringes, polypropylene, 10 ml, with filter incorporating a membrane of porosity less than or equal to 0.8μm, single use); NOTE 6 Since the digestion method does not always allow the sample to be completely dissolved, the undissolved material can be eliminated from the sample solution by filtration. Storage bottles with corresponding tops; Inductively coupled plasma mass spectrometer (PerkinElmer Elan DRC II), quadripole type equipped with a reaction cell, platinum cones, all interfaced to a computerized system. Data acquisition and processing is done by ELAN software from PerkinElmer; Three-channel peristaltic pump; Black/black PVC tube (2 bridges, calibrated at 0.32 ml/min) for the carrier solution; Orange/green PVC tube (2 bridges, calibrated at 0.1 ml/min) for the internal standard; T mixing connection for the sample and internal standard; Injection valve with 2 ml sampling loop; Cyclonic spray chamber made of quartz cooled by Peltier effect with nebulizer made of Teflon PFA; Injection tube, sapphire, 1.8 mm with quartz torch. ESI auto sampler with SC-2 Fast rinse station equipped with an 0.8 mm inert sampling probe; Argon and ammonium hydroxide of suitable purity; MCE membranes, 37 mm, for quality controls on filter.
10 Page 5 5. REAGENTS NOTE 7 The concentration of the elements to be determined that are present in the reagents, the water and the acids must ideally be negligible compared to the lowest concentration to be determined. NOTE 8 Use commercial solutions of certified concentration and guaranteeing traceability to national standards. NOTE 9 Commercial solutions used for preparing calibration control solutions and control samples on filter must ideally come from a different manufacturer than the one used for the calibration standards, or from a different lot or, ultimately, from a different intermediate solution. Water of resistivity equal to or greater than 18 MΩ-cm (at 25 C) and final filtration on 0.1µ for all sample preparations and dilutions (CAS ); Nitric acid (HNO3) concentrated, for example from Fisher, TraceMetal grade (CAS No ); WARNING Concentrated nitric acid is corrosive and an oxidant, and nitric acid vapours are irritants. Avoid all exposure by skin or eye contact, or by vapour inhalation. Use personal protective equipment (appropriate gloves, face shield or safety goggles, etc.) for any work with concentrated or dilute nitric acid. Perchloric acid (HClO4) concentrated, for example from J.T. Baker, InstraAnalyzed grade (CAS No ); WARNING Perchloric acid is corrosive and an oxidizer and its fumes are irritants. Avoid all exposure by skin and eye contact, or by inhalation of the fumes. Use appropriate personal protective equipment (including appropriate gloves, face shield or safety goggles, etc.) for any work with concentrated or dilute perchloric acid, and any dissolving with perchloric acid must be done under a hood with a purifier specially designed for use with perchloric acid. Hydrochloric acid (HCl) concentrated, for example from Fisher, Optima grade (CAS No ); WARNING Concentrated hydrochloric acid is corrosive, and hydrochloric acid vapours are irritants. Avoid all exposure by skin or eye contact, or by inhalation of the vapours. Use appropriate personal protective equipment (for example gloves, face shield or safety goggles, etc.) for any work with concentrated or dilute hydrochloric acid. Handle open containers of concentrated hydrochloric acid in a hood. The vapour pressure of hydrochloric acid is high; as a result, pay attention to the pressure created in capped vials when preparing hydrochloric acid/water mixtures. Hydrogen peroxide (H2O2), for example from EMD, Supra quality (CAS No ); WARNING Hydrogen peroxide is corrosive and an oxidant. Avoid all exposure by skin or eye contact. Use personal protective equipment (including appropriate gloves, face shield or safety goggles, etc.) for any work with hydrogen peroxide. Multi-element commercial solution for instrument and mass optimization (for example: SPEX Tuning Solution, Claritas quality at 10 mg/l). This solution contains Ba, Be, Ce, Co, In, Li, Mg, Pb, Rh, Tl, U and Y. Multi-element commercial solution for the internal standard (for example: SPEX Internal Standard Solution, Claritas quality at 10 mg/l). This solution contains Li 6, Sc, Ge, Y, In, Tb and Bi.
11 Page 6 Multi-element commercial solution "custom made" for the preparation of standard solutions (for example: SPEX CertiPrep, Claritas quality of corresponding concentrations as needed for each element). This solution contains Al, Cd, Co, Cr, Cu, Fe, Mg, Mn, Ni, Pb, V and Zn; Multi-element commercial solution "custom made" for the preparation of calibration control solutions and control samples on filter (for example: SCP Science, PlasmaCal quality of corresponding concentrations as needed for each element). This solution contains Al, Cd, Co, Cr, Cu, Fe, Mg, Mn, Ni, Pb, V and Zn; Triton X-100, for example from Sigma, Ultra quality (CAS No ). 6. SAMPLING The different elements and their compounds suspended in the air are collected using a cassette containing a mixed cellulose ester (MCE) membrane with a porosity of 0.8µm, and a sampling pump with a previously adjusted flow rate. For each series of samples, a control cassette containing a membrane from the same lot must be planned for. This control must be treated in the same way as the cassettes used for sampling for all aspects relating to storage and transport to the sampling point, but must not be used for sampling. The samples are stored at ambient temperature. The recommended sampling parameters are described below: Flow rate: 1.5 L/min Volume: 180 L These parameters take into account the exposure standard, the sensitivity of the analytical method, and the capacity of the sampling system; however, the sampling volume can be modified to adapt to environmental constraints. When high concentrations or particles suspended in the air are expected, the sampling period should be shorter in order to avoid overloading the filter. The sampling period must also be sufficiently long for determining the different elements at significant industrial hygiene levels. To calculate the minimum sampling time necessary for the sampled amount to exceed the minimum reported value for the analytical method for an estimated time weighted average exposure value, the following equation is used: T min MRV / VE x TWAx Q [1] where Tmin is the minimum sampling time, in minutes (min) MRV is the minimum reported value for the analytical method, in µg VE is the estimated weighted average exposure value, as a decimal fraction (for example, 0.1 for 10% of the standard) TWA is the time-weighted exposure value, in mg/m³ Q is the recommended flow rate for sampling, in L/min
12 Page 7 7. ANALYTICAL PROTOCOL 7.1 Analytical solutions NOTE 10 The stability of the solutions depends on several factors, for example the concentration of the element, the test matrix, the type of storage container, and the storage conditions. Prepare fresh solutions every day or keep them for a maximum time determined from the results of stability experiments. 1% HNO3 solution (v/v); 3% HNO3 solution (v/v); Mixed solution of concentrated HNO3 and HClO4 [4:1]; Solution for optimizing the instrument at 1 µg/l in 1% HNO3 (v/v); Solution for mass calibration at 10 µg/l in 1% HNO3 (v/v); Internal standard solution at 200 µg/l in 1% HNO3 (v/v); Intermediate concentration working solutions for the preparation of calibration solutions; Prepare, if necessary, one or more solutions at intermediate concentrations by introducing precisely by pipette the necessary volume of the multi-element stock solution into a graduated flask of suitable volume containing a quantity of 1% HNO3 (v/v). Fill to approximately 90% of the volume of the flask with 1% HNO3 (v/v), cap and then shake to mix. Allow to cool at ambient temperature, fill to the mark with this same acid solution e.g. 1% HNO3 (v/v), cap and mix well. Multi-element calibration solutions in 1% HNO3 (v/v); From the multi-element working solutions at intermediate concentrations prepared for calibration, prepare a series of calibration solutions covering the range of concentrations of each of the elements in the analytical method by introducing precisely by pipette the appropriate volume into a graduated flask of suitable volume containing a quantity of 1% HNO3 (v/v). Fill to approximately 90% of the volume of the flask with 1% HNO3 (v/v), cap and then shake to mix. Allow to cool at ambient temperature, fill to the mark with this same acid solution e.g. 1% HNO3 (v/v), cap and mix well. It is recommended that at least three calibration solutions be prepared. Also prepare a calibration blank solution. Intermediate concentration working solutions for the preparation of calibration control and control on filter solutions; Prepare, if necessary, one or more intermediate concentration solutions by introducing precisely by pipette the necessary volume of the multi-element stock solution into a graduated flask of suitable volume containing a quantity of 1% HNO3 (v/v). The multi-element stock solution must ideally come from a different manufacturer than the one used for preparing the calibration solutions, or from a different lot from the same manufacturer or, ultimately, should be prepared separately. Fill to approximately 90% of the volume of the flask with 1% HNO3 (v/v), cap and then shake to mix. Allow to cool at ambient temperature, fill to the mark with this same acid solution e.g. 1% HNO3 (v/v), cap and mix well.
13 Page 8 Multi-element calibration control and control on filter solutions in 1% HNO3 (v/v); Two calibration control solutions and one filter spiking solution are necessary (see section 7.5). From the multi-element working solutions at intermediate concentrations, prepare the solutions by introducing precisely by pipette the necessary volume into a graduated flask of suitable volume containing a quantity of 1% HNO3 (v/v). Fill to approximately 90% of the volume of the flask with 1% HNO3 (v/v), cap and shake to mix. Allow to cool at ambient temperature, fill to the mark with this same acid solution e.g. 1% HNO3 (v/v), cap and mix well. Stock solution of 1% Triton X-100 (v/v) in 1% HNO3 (v/v); Rinsing solution of 0.001% Triton X-100 (v/v) in 3% HNO3 (v/v). 7.2 Preparation of samples NOTE 11 Samples must be prepared under a hood. NOTE 12 Single use nitrile gloves should be worn during preparation of the samples, in order to avoid any possibility of contamination by hands and to protect them against contact with toxic and corrosive substances. NOTE 13 Note in the laboratory records any relevant remark about the state of the sample and/or observed during solubilization of the sample. Open the cassettes and transfer each filter, using a pair of clean tweezers, into a digestion beaker numbered with the number of the corresponding sample. Follow the same procedure for the field blanks and quality controls on filter; NOTE 14 If there is a loss of dust during transfer into the digestion container, an appropriate note is added to the laboratory records in order to put this observation in writing, and the client is informed about it in the analytical report. Rinse with 6 ml (3 x 2 ml) of the mixture of concentrated HNO3 and HClO4 (4:1), the inside surfaces of the cassette (inlet section), to carry any particles adhering to them into the corresponding beaker; Let the beakers stand for a minimum of 30 minutes; Place the beakers on a hot plate at 150 C, and then evaporate by shaking until approximately 1 ml of acid remains; NOTE 15 If the solution darkens as it evaporates, slowly add concentrated HNO3 dropwise until the solution becomes colourless or slightly yellowish. Attention: splashes can occur if the concentrated HNO3 is poured too quickly, leading to losses of sample. Remove the beakers from the hot plate and allow the solutions to cool; Slowly add 1 ml of H2O2 to each beaker and let stand for 5 minutes; Place the beakers again on the plate and evaporate until approximately 1 ml remains; Remove the beakers from the hot plate and allow the solutions to cool well; Slowly add 5 ml of concentrated HCl by rinsing the inside of each beaker; NOTE 16 This step produces a release of irritant vapours.
14 Page 9 Place the beakers again on the hot plate at 150 C and then evaporate to dryness by shaking; NOTE 17 Do not allow the residues to carbonize. Remove the beakers from the hot plate and allow them to cool; Carefully rinse the walls of each beaker with several ml of 1% HNO3 v/v; Place the beakers again on the plate and heat for a few minutes while shaking to redissolve the residue; Transfer the contents to a graduated test tube. Rinse the beaker 3 times with a few ml of 1% HNO3 by transferring quantitatively after each rinse; Place the plugs in the test tubes and allow to cool at ambient temperature; Fill the test tubes to 10 ml with 1% HNO3, plug and then store for later analysis; NOTE 18 The digestion method does not always allow the sample to completely dissolve. The remaining undigested material is then eliminated from the sample solution by filtration. An appropriate note is added to the laboratory records in order to put this observation in writing, and the client is informed about it in the analytical report. 7.3 Verification of the ICP-MS performance criteria Before activating the plasma, certain parts of the instrument should be visually checked: excessive deposits on the cones, cleanliness of the torch, and mainly the introduction tube and integrity of the tubing peristaltic pump. The cones should also be changed and/or cleaned when the deposits are considered excessive, when the sensitivity is insufficient, or when the level of polyatomic ions and/or doubly charge ions is too high. The manufacturer's recommendations for turning on the instrument must be followed, for example, regarding compliance with the required laboratory environmental conditions, requirements relating to electrical power, to cooling water, to exhaust ventilation, warm-up time, etc. Once the visual checks have been done, activate the plasma and allow the instrument to warm up for a minimum of 30 minutes before using it. The calibration blank solution should be aspirated into the plasma during the warm-up period in order to avoid changing the conditions of the plasma during the analysis. Parameterize the instrument in accordance with the manufacturer's specifications for performance verification and begin the application. The performance results must be equal to or better than the manufacturer's; otherwise, the different parameters of the ICP-MS must be adjusted and the verification restarted. The instrument must be adjusted to reduce interferences (oxide formation, formation of doubly charged ions, for example) to a minimum and to obtain sufficient sensitivity.
15 Page 10 The instrument's performances are checked daily, before sample analysis, by aspiration of a prescribed tuning solution of the elements. Mass calibration and alignment of the two detectors (linear response) must be done as often as required by the manufacturer or according to the instrument's performance history. 7.4 Calibration The instrument is calibrated by measuring, in increasing order of concentration, the intensity of the different standards of known concentrations. Once the standards have been analyzed, the data acquisition software produces a calibration curve, of type mx+b, of the concentration of the element as the ordinate and the signal as the abscissa. Typically, one calibration blank solution and three standard solutions are analyzed to produce the calibration curve. Calibration with more than five solutions is rarely necessary. It is recommended that the signal originating from the calibration blank solution be subtracted from the other calibration solutions. The correlation coefficient (r), which translates the adjustment of the estimates of the linear regression equation, should be greater than If r is less than this value, the calibration standards should be reanalyzed or redone to reach this value. It may be possible to eliminate one calibration point and to treat the data again to obtain a correlation coefficient > 0.995, but one should comply with the minimum number of 4 calibration points (the calibration blank solution and three standard solutions). All the calibration solutions, control solutions, internal standard solution and samples are sent to the nebulizer by a peristaltic pump with several channels. The internal standard solution is added and mixed with all the solutions before analysis using a T connection and the peristaltic pump. The mixture is then sent to the nebulizer. The sample/ internal standard mixture ratio is 5 to 1. Before and after each analysis, the system tubing is rinsed with a mixed solution of 0.001% Triton X-100 and 3% HNO3 (v/v). The instrument's main analytical parameters are indicated in the tables below.
16 Page 11 PARAMETERS VALUE Plasma power (Watts) 1200 Argon flow rate (L/min) 15 Flow rate of auxiliary gas (L/min) 1.2 Gas flow to nebulizer (L/min) 0.9 Acquisition mode Peak hopping Sweeps (per reading) 10 Readings (per replicate) 1 Replicates 3 Mode of analysis Standard and DRC Calibration External calibration Detector mode Pulse Isotope Correction Internal equations standard Mode Flow (ml) Rpa 1 Rpq 27 Al Sc Standard Cd *Sn118 In Standard Cr Ge Standard Co In DRC-NH Cu Ge Standard Fe Ge Standard Mg Sc Standard Mn Ge Standard Ni In DRC-NH Pb 1*Pb *Pb 207 Bi Standard V In DRC-NH Zn *Ni 60 Ge Standard The Rpa value can vary in relation to system optimization. It is important to limit the count number of the last curve point to 2,000,000 cps so as not to saturate the detector in pulse mode. 7.5 Quality control Several types of controls are analyzed during the analytical sequence in order to check the calibration and the variation in sensitivity. Each control must meet the criteria established for this control, and any exceedence or deviation must be documented, and the appropriate actions planned by the quality system must be undertaken.
17 Page 12 The commercial solution used for preparing the calibration control solutions, minimum reported value and quality control on filter must ideally come from a different manufacturer than the one used for preparing the calibration solutions, or from a different lot from the same manufacturer or, ultimately, from a different intermediate solution from the same manufacturer. Initial calibration verification (ICV) solution. The calibration controls are solutions that are analyzed just after instrument calibration. Two solutions are recommended and their concentrations are approximately 25% (ICV-L) and 75% (ICV-H) of the concentration of the highest standard. Initial calibration blank verification solution (IBV). The IBV control is the same solution as the calibration blank solution. It is analyzed after the two (2) ICV controls and the results must be comparable to the results obtained in the analysis of the calibration blank solution. Minimum reported value (MRV). The MRV control is a control solution analyzed after the initial calibration controls. This control verifies the instrument's response at the analytical method's minimum reported value. Calibration control solution (CCV). The CCV control is the same solution as the calibration solution representing approximately the concentration at the midpoint of the calibration curve. The CCV is analyzed sequentially every 10 samples. Continuing calibration blank solution (CCB). The CCB control is the same solution as the calibration blank solution and is analyzed after each CCV. The results of the CCBs must be comparable to the results of the calibration blank solution. Quality control on filter (QC). These control samples are filters spiked with the different elements to be determined. They are subjected to all of the preparation and analytical steps as a sample. Two control samples are digested and analyzed as samples in the analytical sequence. The results are used to estimate the accuracy of the analytical method and the replicability (precision) between the two controls. The two filters are spiked with 50 µl of the control solution for filters. The mass (µg) of the different elements deposited on the filter represents approximately 50% of the value of the final calibration point. Laboratory reagent blank (LRB). The LRB control is a sample solution that contains only the reagents. It is subjected to all the preparation and analytical steps as a sample. It is used to verify whether the preparation steps result in contamination of the samples. Three (3) reagent blanks are digested per analytical sequence.
18 Page Analysis Once the calibration curve has been established, the reagent blanks, controls and samples are successively analyzed, and the concentration of the different elements present is determined by comparison with the calibration curve. The concentration of the different elements determined in the sample must fall within the analytical method's working range. If the concentration in the sample is higher than the highest concentration in the working range, the sample is appropriately diluted to match the matrix and analyzed again while taking the dilution factor into account in the calculations. Furthermore, if the concentration of an element greatly exceeds the concentration of the most concentrated sample, it becomes necessary to verify the elimination of the different elements in the tubing (memory effect) by analyzing the calibration blank solution. The result must be similar to the result obtained for the initial calibration blank solution. A longer rinsing time generally corrects this situation. When the internal standard's signal varies by more than ± 25% for a sample, it is important to verify whether dilution of this sample significantly affects the previous obtained result.
19 Page 14 The analyses are performed according to the following proposed sequence: Sample Type of sample 1% HNO3 Rinsing of the injection loop (3X) S0 Calibration blank solution S1 Standard #1 S2 Standard #2 S3 Standard #3 S4 Standard #4 S5 Standard #5 ICV-L (at 20% of S5) Initial calibration verification solution ICV-H (at 80% of S5) Initial calibration verification solution IBV Initial blank verification MRV Control solution at the MRV QC1 Digested spiked filter QC2 Digested spiked filter LRB #1 Reagent blank #1 LRB #2 Reagent blank #2 LRB #3 Reagent blank #3 5 samples CCV Continuing calibration verification CCB Continuing calibration blank 10 samples CCV and CCB Continuing calibration controls S0 to S5 Calibration blank and standards #1 to #5 ICV-L, ICV-H, IBV, VMR Initial calibration controls 10 samples CCV and CCB Continuing calibration controls 10 samples CCV and CCB Continuing calibration controls S0 to S5 Calibration blank and standards #1 to #5 ICV-L, ICV-H, IBV, VMR Initial calibration controls 10 samples CCV and CCB Continuing calibration controls 10 samples CCV and CCB Continuing calibration controls Etc. MRV Control solution at the MRV CCV and CCB Continuing calibration controls 1% HNO3 Rinsing of the injection loop (3X)
20 Page CALCULATIONS Calculation of the concentration of the different elements in the air at ambient conditions is done using the following equation: C = [(C1 x V1 x F) (C0 x V0)] / Ve Where C is the calculated concentration of the element in the sample, in mg/m³ C1 is the concentration of the element in the sample, in µg/l V1 is the final volume of the sample, in litres (L) F is the dilution factor (F=1 for no dilution) C0 is the average concentration of the element in the digestion blanks 1 (reagent blanks and/or filters), in µg/l V0 is the final volume of the sample, in litres (L) Ve is the sampling volume, in litres (L) 1 The results of the digestion blanks are results originating from laboratory reagent blank or filter blanks obtained by following the same analytical process as the sample. The results obtained for the samples are not corrected for the results of the field blanks. The results of the field blanks are reported as total mass (µg). 9. PERFORMANCE PARAMETERS 9.1 Limit of detection, limit of quantification and minimum reported value (MRV) The method detection limit (MDL) and quantification limit (MQL) were evaluated (n=10) for a final volume of 10 ml and are reported in Appendix 1. The minimum reported value (MRV) is the minimum amount of a contaminant that is quantified in the IRSST laboratory. The MRV is greater than or equal to the MDL. It takes into account one or more of the following aspects: linearity of the method under the experimental conditions used, the recovery efficiency, and the relevance of the determination at low concentration levels. 9.2 Recovery The dissolution procedure was evaluated by subjecting a series of spiked membranes (n=28, 4 concentration levels, 7 membranes per level) to the entire analytical procedure with a soluble form of the different elements in the method. Appendix 1 contains the different recovery percentages obtained for each of these elements. 9.3 Reliability The replicability and repeatability values were determined in the laboratory from the 44 spiked membranes subjected to the entire analytical procedure (4 concentration levels, 11 membranes per level). These values represent the precision of the method.
21 Page Accuracy The accuracy of the analytical method was evaluated (n=10) for each of the elements in the analytical method (see Appendix 1). Accuracy is also verified with each series of analyses by means of a membrane spiked with different metals. The results obtained are compiled in the context of the intra-laboratory quality control. 9.5 Measurement uncertainty Analytical measurement uncertainty was calculated with the results obtained from 28 spiked membranes (4 concentration levels, 7 membranes per level) subjected to the entire analytical procedure. This value corresponds to the analytical method's coefficient of variation (CVA) (see Appendix 1). 10. REFERENCES Also, the expanded measurement uncertainty (CVE) for the entire determination and sampling was calculated by taking into account a CV estimated at 5% for the sampling, and a probability threshold of 95%. 1. IRSST, Sampling Guide for Air Contaminants in the Workplace. T-06 Guide technique, Direction des opérations, Montréal, Québec, (Février 2005) ISO :2000, Workplace air - Determination of metals and metalloids in airborne particulate matter by inductively coupled plasma atomic emission spectrometry - Part 1: Sampling 3. ISO : 2001, Workplace air Determination of metals and metalloids in airborne particulate matter by inductively coupled plasma atomic emission spectrometry Part 2: Sample preparation. 4. ISO : 2004, Workplace air Determination of metals and metalloids in airborne particulate matter by inductively coupled plasma atomic emission spectrometry Part 3: Analysis. 5 ISO : 2004, Water quality - Application of inductively coupled plasma mass spectrometry (ICP-MS) - Part 1: General guidelines 5. ISO : 2003, Water quality Application of inductively coupled plasma mass spectrometry (ICP-MS) Part 2: Determination of 62 elements. 6. ASTM D , Standard test method for determination of elements in airborne particulate matter by inductively coupled plasma mass spectrometry (ICP-MS) 7. Institut National de Recherche et de Sécurité (INRS), Métrologie des polluants Évaluation de l exposition professionnelle - Méthodes de prélèvement et d analyse de l air: Fiche 003, Métaux - Métalloïdes, INRS, Paris. 8. Centre d expertise en Analyse Environnementale du Québec, Détermination des métaux : méthode par spectrométrie de masse à source ionisante au plasma d argon. MA. 200 Mét 1.1, Ministère de l Environnement du Québec, 2004, 37 p. 9. EPA, Test Method for Evaluating Solid Waste, Physical/Chemical Methods, Method EPA, Test Method for Evaluating Solid Waste, Physical/Chemical Methods, Method 3050.
22 Page 17 Appendix 1 Performance parameters MDL MQL MRV 1 Replicability Repeatability Accuracy Recovery CVa CVE µg/l µg/l µg % % % % % % Al Cd Co Cr Cu Fe Mg Mn Ni Pb V Zn Minimum reported value, it takes into account the final volume of 10 ml and the 1:10 dilution of the samples during the analysis.
Analytical Method. Endotoxin analysis ANALYTICAL METHOD 332
Analytical Method Endotoxin analysis ANALYTICAL METHOD 332 Applicability This method is used for analyzing endotoxins Standard(s) 1 No standard. Sampling system Glass fibre filters with three-section cassettes
HS 1003 Part 2 HS 1003 Heavy Metals Test
HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:
Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry
Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
How To Use The Nexion 300
APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of
Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
NITRIC OXIDE and NITROGEN DIOXIDE 6014
NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm
Work Assignment 5-03. Prepared for: Office of Air Quality Planning and Standards U.S. Environmental Protection Agency Research Triangle Park, NC
ERG No.: 0143.04.005 EPA Contract No.: 68-D-00-264 Standard Operating Procedure for the Determination of Metals In Ambient Particulate Matter Analyzed By Inductively Coupled Plasma/Mass Spectrometry (ICP/MS)
Sample Analysis Design Isotope Dilution
Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published
Laboratory Procedure Manual
Laboratory Procedure Manual Analyte: Matrix: Method: Method No.: Lead Dust Wipes Modification of the Automated AAII-25 Colorimetric method Nitric and Hydrochloric Acid Digestion followed by Atomic Absorption
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics,
Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist
Application of ICP-MS as a Tool for Environmental Monitoring Bill Spence Environmental Market Specialist Introduction Knowledge of the concentration of metallic species in the environment has been important
USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls
Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required
MILESTONE CLEAN CHEMISTRY LINE DUOPUR TRACECLEAN ULTRA-TRACE INSERTS. A Breakthrough in Reduction and Control of the Analytical Blank MILESTONE
MILESTONE CLEAN CHEMISTRY LINE A Breakthrough in Reduction and Control of the Analytical Blank DUOPUR TRACECLEAN ULTRA-TRACE INSERTS MILESTONE MILESTONE CLEAN CHEMISTRY LINE An innovative and complete
Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
METHOD 200.8 DETERMINATION OF TRACE ELEMENTS IN WATERS AND WASTES BY INDUCTIVELY COUPLED PLASMA - MASS SPECTROMETRY. Revision 5.
METHOD 200.8 DETERMINATION OF TRACE ELEMENTS IN WATERS AND WASTES BY INDUCTIVELY COUPLED PLASMA - MASS SPECTROMETRY Revision 5.4 EMMC Version S.E. Long (Technology Applications Inc.), T.D. Martin, and
All the prepared formulations were subjected for following. evaluation parameters and obtained results were showed in Tables 6.3 &
105 6.1 CHARACTERIZATION OF TABLETS All the prepared formulations were subjected for following evaluation parameters and obtained results were showed in Tables 6.3 & 6.4. 6.1.1 Description (Size, Shape,
VALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol
Page 1 van 36 pages VALIDATION REPORT Investigation of the major interferences for proposed Horizontal ICP AES method Identification code Validation in accordance with Reference method Scope Reference
Target Mole Lab. Mole Relationships and the Balanced Equation. For each student group Hydrochloric acid solution, HCl, 3 M, 30 ml
elearning 2009 Introduction Target Mole Lab Mole Relationships and the Balanced Equation Publication No. A common chemical reaction used in chemistry class is zinc and hydrochloric In this lab, students
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
Elemental Scientific. Stability. Elemental Scientific
Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dilution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field ([email protected]) Brief The prepfast inline
METHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY
METHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY 1.0 SCOPE AND APPLICATION 1.1 Inductively coupled plasma-atomic emission spectrometry (ICP-AES) determines trace elements, including
Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
Coating and Extraction of Honeycomb Denuders
QSM Approval: Coating and Extraction of Honeycomb Denuders 1 INTRODUCTION 1.1 The following procedures are used for the coating of honeycomb denuders with citric acid and sodium carbonate solutions, and
Elemental Analyses by ICP-AES
Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014 This experiment explores quality assurance practices which are commonly
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows
Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for
Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy
Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy In support of the toy industry, many laboratories are being called upon to analyze a wide variety of
Estimation of Alcohol Content in Wine by Dichromate Oxidation followed by Redox Titration
Sirromet Wines Pty Ltd 850-938 Mount Cotton Rd Mount Cotton Queensland Australia 4165 www.sirromet.com Courtesy of Jessica Ferguson Assistant Winemaker & Chemist Downloaded from seniorchem.com/eei.html
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
Colorimetric Determination of Iron in Vitamin Tablets
Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered
OLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
The Empirical Formula of a Compound
The Empirical Formula of a Compound Lab #5 Introduction A look at the mass relationships in chemistry reveals little order or sense. The ratio of the masses of the elements in a compound, while constant,
Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT
Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.
General Chemistry I (FC, 09-10) Lab #3: The Empirical Formula of a Compound. Introduction
General Chemistry I (FC, 09-10) Introduction A look at the mass relationships in chemistry reveals little order or sense. The ratio of the masses of the elements in a compound, while constant, does not
EXPERIMENT 12: Empirical Formula of a Compound
EXPERIMENT 12: Empirical Formula of a Compound INTRODUCTION Chemical formulas indicate the composition of compounds. A formula that gives only the simplest ratio of the relative number of atoms in a compound
Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media
Enhanced Organic Precursor Removals Using Aged Media Page 1 Enhanced Organic Precursor Removals Using Aged Media O b j e c t i v e s The main goals of this project was to evaluate the dissolved organic
Solutions and Dilutions
Learning Objectives Students should be able to: Content Design a procedure for making a particular solution and assess the advantages of different approaches. Choose the appropriate glassware to ensure
Experiment 8 - Double Displacement Reactions
Experiment 8 - Double Displacement Reactions A double displacement reaction involves two ionic compounds that are dissolved in water. In a double displacement reaction, it appears as though the ions are
EXPERIMENT 7 Reaction Stoichiometry and Percent Yield
EXPERIMENT 7 Reaction Stoichiometry and Percent Yield INTRODUCTION Stoichiometry calculations are about calculating the amounts of substances that react and form in a chemical reaction. The word stoichiometry
Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)
Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Grace Analytical Lab 536 South Clark Street 10th Floor Chicago, IL 60605 April 15, 1994 Revision 2 Standard Operating Procedure
Determination of the Mass Percentage of Copper in a Penny. Introduction
Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny
ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND
#3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric
APPLICATION OF ICP-MS TECHNOLOGY FOR TRACE METALS ANALYSIS
MR. TELLIARD: Our next speaker is Paula Hogg. Paula is currently a lab manager at Hampton Roads Sanitation District s Central Environmental Laboratory. We would like to also thank Hampton Roads for sending
METHOD 200.7 DETERMINATION OF METALS AND TRACE ELEMENTS IN WATER AND WASTES BY INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY
METHOD 200.7 DETERMINATION OF METALS AND TRACE ELEMENTS IN WATER AND WASTES BY INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY Revision 4.4 EMMC Version USEPA-ICP Users Group (Edited by T.D. Martin
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration.
81 experiment5 LECTURE AND LAB SKILLS EMPHASIZED Synthesizing an organic substance. Understanding and applying the concept of limiting reagents. Determining percent yield. Learning how to perform a vacuum
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:
Determination of Calcium and Magnesium in Brine
Application Note Determination of Calcium and Magnesium in Brine INTRODUCTION To prevent membrane poisoning, new membrane technology in chlor-alkali cells requires feed brine that is relatively free of
105 Adopted: 27.07.95
105 Adopted: 27.07.95 OECD GUIDELINE FOR THE TESTING OF CHEMICALS Adopted by the Council on 27 th July 1995 Water Solubility INTRODUCTION 1. This guideline is a revised version of the original Guideline
LACHAT METHOD NUMBER 10-204-00-5-D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately 3-500 g CN/L 2-500 g CN/L. 09e
1 SIDE BY SIDE COMPARISON TABLE FOR THE DETERMINATION OF CYANIDE BY UV DIGESTION AND AMPEROMETRIC DETECTION FIA METHOD - 10-204-00-5-D (ASTM D7511-09 e ) TOPIC Method ASTM D7511-09e LACHAT METHOD NUMBER
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility
Use of Absorption Mechanisms to Decrease Heavy Metal Mobility by Jianmin Wang Honglan Shi Joe G. Burken NUTC R304 A National University Transportation Center at Missouri University of Science and Technology
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of total nitrogen in food and crude protein calculation (Kjeldahl method) Responsible person: Assoc.Prof. Ing.Kateřina Riddellová,
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method
ICP - Mass Spectrometry APPLICATION NOTE. Authors. Introduction. Lee Davidowski Ewa Pruszkowski PerkinElmer, Inc. Shelton, CT USA
APPLICATION NOTE ICP - Mass Spectrometry Authors Lee Davidowski Ewa Pruszkowski PerkinElmer, Inc. Shelton, CT USA Benefits of the NexION 300X/350X ICP-MS Coupled with the prepfast In-line Auto-dilution/calibrationSystem
Safety Data Sheet Aluminium profiles Apply for alloys Sapa EN AW 6060, 6063, 6005, 6005A, 6082, yellow chromated, with or without powder coating.
Apply for alloys Sapa EN AW 6060, 6063, 6005, 6005A, 6082, yellow chromated, with or without powder coating. 1. Identification of the Substance/Preparation and Company Product name: (Alloys, Sapa EN AW
Mechanical Systems Competency 1.20
Competency 1.20 Mechanical systems personnel shall demonstrate a working level knowledge of the safety and health fundamentals of mechanical systems and/or components. 1. Supporting Knowledge and Skills
Determination of elements by ICP-AES and ICP-MS
September 2003 HORIZONTAL - 19 Determination of elements by ICP-AES and ICP-MS Henk J van de Wiel National Institute of Public Health and the Environment (RIVM) Bilthoven, The Netherlands 1 HORIZONTAL
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical
Exposure and Control of Ten High Priority MSHA Contaminants in Metal Mines. Presented by: Shannon E. Newton, MPH, CIH
Exposure and Control of Ten High Priority MSHA Contaminants in Metal Mines Presented by: Shannon E. Newton, MPH, CIH Topics What are these hazards? How can they harm you? Why do we care? How do we tell
Universal Data Acquisition (UDA)
Universal Data Acquisition (UDA) I C P - O P T I C A L E M I S S I O N P R O D U C T N O T E Introduction Historically, Inductively Coupled Plasma (ICP) spectroscopy has been used for multiple analyte
How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use.
BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use. USP proposes the revision and development of a suite of plastic packaging system standards in the current issue of PF. General test chapter
Quantitation of Drugs in Dried Bloodstains. Thomas Meyer Anna Gomenyuk Nadiah Lester
Quantitation of Drugs in Dried Bloodstains Thomas Meyer Anna Gomenyuk Nadiah Lester Objective To find the quantitative value of drugs in dried bloodstains Currently Cloth With Dried Bloodstain Containing
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid
Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid Introduction Many metals react with acids to form hydrogen gas. In this experiment, you will use the reactions
Wipe Analysis to Determine Metal Contamination on Critical Surfaces
By Albert Dato, Ph.D., Warren York, Jennifer Jew, Laarni Huerta, Brice Norton, and Michael Coste On-wafer metallic contamination is detrimental to the fabrication and performance of semiconductor devices.
Chromium Speciation in Drinking Water by LC-ICP-MS
APPLICATION NOTE LC-ICP-MS Authors: Helmut Ernstberger Ken Neubauer PerkinElmer, Inc. Shelton, CT Chromium Speciation in Drinking Water by LC-ICP-MS Introduction With the growing concern of pollutants
Standard Operation Procedure. Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry
Standard Operation Procedure Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
OXIDATION-REDUCTION TITRATIONS-Permanganometry
Experiment No. Date OXIDATION-REDUCTION TITRATIONS-Permanganometry INTRODUCTION Potassium permanganate, KMnO 4, is probably the most widely used of all volumetric oxidizing agents. It is a powerful oxidant
Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate
Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in
W1 WORKSHOP ON STOICHIOMETRY
INTRODUCTION W1 WORKSHOP ON STOICHIOMETRY These notes and exercises are designed to introduce you to the basic concepts required to understand a chemical formula or equation. Relative atomic masses of
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
IB Chemistry. DP Chemistry Review
DP Chemistry Review Topic 1: Quantitative chemistry 1.1 The mole concept and Avogadro s constant Assessment statement Apply the mole concept to substances. Determine the number of particles and the amount
hij GCSE Additional Science Chemistry 2 Foundation Tier Chemistry 2F SPECIMEN MARK SCHEME Version 1.0
hij GCSE Additional Science Chemistry 2 Foundation Tier Chemistry 2F SPECIMEN MARK SCHEME Version.0 Copyright 20 AQA and its licensors. All rights reserved. The Assessment and Qualifications Alliance (AQA)
SCP SCIENCE www.scpscience.com. Certified Reference Materials. Materials
Certified Reference 47 EnviroMAT and AgroMAT Certifi ed Reference (CRMs) are designed to complement existing Performance Evaluation Programs in environmental and agricultural analysis. EnviroMAT CRMs allow
NANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For
NANYANG TECHNOLOGICAL UNIVERSITY School of Civil and Environmental Engineering CV2701: Laboratory 2A Laboratory Manual For Experiment Lab 2A-6(ENV) Wastewater Quality Analysis Location : Environment Laboratory,
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
MATERIAL SAFETY DATA SHEET SECTION 1 CHEMICAL PRODUCT AND COMPANY INFORMATION SECTION 2 COMPOSITION/INFORMATION ON INGREDIENTS
SECTION 1 CHEMICAL PRODUCT AND COMPANY INFORMATION Chemical Name: Brass and Copper Alloys Product Use: Manufacturer Information: OMNISOURCE CORPORATION Telephone: (260)422-5541 Safety Department 1610 North
Simulation of the determination of lead azide content in waste water from explosives manufacture
Simulation of the determination of lead azide content in waste water from explosives manufacture Lead azide ranks in the category of intensive explosives, which may, even in an insignificant amount, initiate
Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.
Conductivity of strong and weak electrolytes TEC Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law. Principle It is possible to
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
Experiment 7: Titration of an Antacid
1 Experiment 7: Titration of an Antacid Objective: In this experiment, you will standardize a solution of base using the analytical technique known as titration. Using this standardized solution, you will
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
Certificate of Analysis
The Japan Society for Analytical Chemistry Certificate of Analysis Certified Reference Material JSAC 0602-3 Plastics (tip form) for Chemical Analysis of Hazardous Metals This certified reference material
Elemental Analysis Manual for Food and Related Products
Elemental Analysis Manual for Food and Related Products 4.7 Inductively Coupled Plasma-Mass Spectrometric Determination of Arsenic, Cadmium, Chromium, Lead, Mercury, and Other Elements in Food Using Microwave
EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion
EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1 Introduction This lab is designed for you to discover the properties of electrochemical cells. It requires little previous knowledge of electrochemical
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Why? Chemists are concerned with mass relationships in chemical reactions, usually run on a macroscopic scale (grams, kilograms, etc.). To deal with
Holcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants
Holcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants Version 2004-01 / August 2004 (HGRS-JW-04-25s) HGRS-CTS/MT J. Waltisberg SUMMARY ( ) shall not