HOW DO I OBTAIN MOTOR UNIT DATA?

Size: px
Start display at page:

Download "HOW DO I OBTAIN MOTOR UNIT DATA?"

Transcription

1 HOW DO I OBTAIN MOTOR UNIT DATA? In order to obtain motor unit data, the user performs an experiment to collect the data and then uses the demg software algorithm to decompose the data into the constituent MUAPTs. The experiment consists of a subject tracking a trapezoidal force profile presented on a screen while semg signals are collected from a demg sensor placed on the muscle. The force profile is normalized to the subject s maximal voluntary contraction (MVC) level. The demg System provides a simple procedure to design the force profile and perform an experiment to collect semg signals for decomposition. The steps are outlined below. 1) Design the Force Profile a. Establish the characteristics of the force profile. b. Establish the modality of the force feedback. 2) Locate and Apply the demg Sensor a. Determine preferred location for the demg sensor. b. Prepare skin and apply sensor. 3) Collect and Decompose the semg signal a. Establish the MVC level for normalization of force. b. Perform a signal quality check to assure that the collected semg signal can be decomposed. c. Collect the semg signal during a force profile task. d. Decompose the semg signal. Step 1: Design the Force Profile a) Force Profile The demg software (v2.0) is designed to decompose semg signals from isometric contractions having trapezoidal force profiles defined by the parameters in the figure below.

2 Quiescent Region Up Ramp Constant-Force Region Down Ramp Quiescent Region Duration of Contraction Length of semg signal file Recommended Parameters Quiescent Region Up Ramp Constant-Force Region Down Ramp Maximum Duration of force profile At least 3 s duration required 5%MVC/s < slope < 20%MVC/s From >0% to 80% MVC, minimum duration 10 s -5%MVC/s < slope < -20%MVC/s < 45 s Detailed instructions for generating and customizing the force profile may be found in the demg User s Guide. All detected semg signals require a quiescent region of at least 3 s at the beginning and at the end of the signal. The 3 s epoch at the beginning is required for the algorithms to establish the magnitude of the baseline noise and to ascertain the presence of line noise contamination, that is, 50 or 60 Hz. The 3s epoch at the end is needed to ensure that the contraction has ended and no motor units are firing. The recommended slope of the up/down ramp region is designed to provide sufficient information for the decomposition algorithm to progressively identify the recruitment and derecruitment of motor units. The duration of the constant contraction region should be at least 10 seconds long to provide a sufficient number of motor unit firings for proper decomposition.

3 If the length of the semg signal file exceeds 45 s, the decomposition is likely to fail because the PC s virtual memory space is insufficient to store the intermediate results produced by the decomposition program. b) Select RMS or Force Feedback Force feedback is the preferred mode because the signal is relatively smooth. Using force feedback requires the following: 1) a force sensor (load cell) to measure the applied force, and 2) an apparatus to constrain the muscle to produce an isometric contraction. The apparatus should not touch or interfere with the sensor to avoid inducing movement artifacts in the semg signal. This approach is problematic in cases such the muscles of the face, neck, or back, where the generated muscle force cannot be easily measured. Alternatively, the root-mean-squared (RMS) value of the collected semg signal may also be used as a substitute to estimate the level of the contraction. However, due to the inherent variability the RMS signal (even at constant force), it is more difficult for the subject to smoothly track along the trajectory. Using RMS feedback requires only an apparatus to constrain the muscle to produce an isometric contraction. Examples of the experimental setup for using force sensor feedback and semg RMS feedback are show below. Force sensor feedback Secure the joint on which the tested muscle acts in a device so as to immobilize the joint. Place the force sensor against the body segment on which the force acts. The software will display the force output on a computer screen that also presents the force profile to be tracked. See figure for additional clarity. Force Sensor Biceps Force Sensor FDI F 90 F

4 semg Root Mean Squared (RMS) feedback This mode can be used to perform a tracking task when no force sensor is available. The joint on which the tested muscle acts is secured in a device so as to immobilize the joint, which performs a contraction against a fixed resistance. The RMS of the semg signal (either from the demg sensor or a DE-2.1 parallel-bar sensor) may be used for feedback and is displayed to the subject on the computer screen along with the tracking profile. See figure for additional clarity. Fixed Resistance Biceps 90 F

5 Step 2: Locate and Apply the demg Sensor After the subject has been positioned in the apparatus, the location of the demg sensor is determined and the site prepared so that the sensor can be applied to the skin. The following sections describe where to place the sensor on the surface of the muscle and how to prepare the skin before affixing the sensor. a) Preferred Sensor Location Locate the demg sensor in the middle of the muscle surface area. It is well known that the characteristics of the semg signal vary in amplitude and in frequency components along the surface of the muscle (Basmajian and De Luca 1985 ; Roy et al ; De Luca 1997 ). The MUAPs have greater amplitude in the middle of the muscle where the diameter of the muscles fibers is generally the greatest. The amplitude decreases when the sensor is placed near the tendon zones. The innervation zones should be avoided because in these regions the action potential propagates in opposing directions, rather than across the demg sensor s detection pins. It follows that the middle of the muscle is the preferred location. A study by Zaheer et al. (2012) supports this logic and presents data showing that the motor unit decomposition yield is greatest from the recommended location. The study also recommends avoiding locations where large amounts of adipose tissue are present beneath the skin. This problem is likely to be encountered in regions such as the Quadriceps and the Biceps Femoris where there may be excessive (> 1.5 cm) fatty tissue between the surface of the muscle and the sensor. The additional tissue attenuates the signal amplitude.

6 b) Skin Preparation and Sensor Application The demg sensor is specifically designed to work with the demg algorithms. It should not be treated as an ordinary semg signal sensor. The small electrode contacts (pins) of the demg sensor, by their nature, generate more electrical skin/interface noise than sensors with larger electrodes. Their performance is affected by small quantities of debris and oils on the surface of the skin. It is important that the skin where the sensor is to be attached should be clean and moist. Consequently, it is necessary to prepare the skin surface prior to attaching the demg sensor. The quality of the recorded data is only as good as the quality of the skin preparation. For Best Practice adhere the following steps: Shave site to remove hair. To remove skin oils, swab the identified sensor location site twice with a commonly available alcohol prep pad saturated with 70% isopropyl alcohol and allow 30 s to dry. Remove the surface dead layers of the skin by applying hypo-allergenic tape (3M TM Transpore or equivalent) to the site, then peel back to remove contaminants. Repeat 3-5 times. Allow 1-2 minutes for the skin to re-hydrate. Moisten the tip of the 5 contact pins on the bottom of the sensor by dabbing with a wet alcohol prep pad, making sure not to wet the surface in-between the contacts. Next, use a fresh alcohol prep pad to lightly dab the sensor application site so that the skin is also moist. Do not soak the skin, as it will cause the semg signal to be attenuated. Immediately apply the sensor to the skin while it is still moist. Use hypo-allergenic tape (3M TM Transpore or equivalent) to firmly hold the sensor in place. The sensor pins should push into the skin, but not break the skin. If the skin is Dry Without using the alcohol pads, perform the tape peels as described in the normal skin preparation. Gently wash the sensor application site with a mild solution of soap and water and wipe dry. Use a fresh alcohol prep pad to pre-moisten the sensor pins, also dab the sensor application site with the pad so that the skin is moist, and immediately apply the sensor to the skin. Re-application of demg Sensor When re-applying the demg sensor to address baseline noise or line interference, use a fresh alcohol prep pad to re-moisten the sensor pins. Next, dab the sensor application

7 site with the pad so that the skin is moist, and then immediately re-apply the sensor while the skin is still moist. Over-wetting will greatly attenuate the semg signal. Ensure that the bottom surface of the sensor is dry before re-moistening the pins. Excessive, repeated rubbing with alcohol wipes will overly dry the skin resulting in a noisy electrical interface. The Reference Electrode Use a large-area (~20cm 2 ) reference electrode such as Dermatrode ( or similar reference electrode. Before application clean the site with an alcohol wipe. Place reference electrode on top of a bony region. CAUTION Use care when routing the demg sensor, input module, and reference electrode cables to avoid interference with power line noise. Use of devices containing RF transmitters (including cell phones, and laptops with wireless internet) may cause interference with the sensitive signal recording equipment. Be sure to remove these sources from the recording environment. Line interference from nearby wires connected to power sources may interfere with the proper working of the system. Step 3: Collect and Decompose the Data After the tracking profile has been defined and the sensor has been correctly positioned on the muscle, semg signals can be acquired. The demg System provides a simple graphical interface to help the operator perform the sequence of tasks required for data acquisition. A series of icons representing each type of task are displayed on the screen in the order in which they are to be performed. The operator can click on an icon to execute that task, or the program can automatically advance to the next task when the current task is completed. A screenshot showing the tasks required for acquiring the semg signal in a typical experiment using force feedback is shown below:

8 First, the MVC task is performed to normalize the force feedback. Then the Signal Quality Check task is performed to ensure that the semg signals are suitable for decomposition. Finally, the Profile Tracking task is performed to acquire the semg and force signals. Clicking on the Analyze task icon transfers the resultant data to a workspace for viewing and decomposing the signal. Detailed instructions for creating each of these tasks may be found in the demg User s Guide. Each of the tasks performed during data acquisition is described in the following sections: a) Establish MVC The software automatically calculates the MVC value from either the force or the semg RMS signal, depending on which feedback modality was chosen. The MVC is determined by instructing the subject to contract as forcefully as possible for about 3 s. We suggest that at least three attempts be performed, each spaced at least 3 min apart to reduce the effect of muscle fatigue. The program automatically selects the highest reading of the 3 attempts as the MVC value, and uses this value to normalize the tracking profile. The following plot shows an example of a successful attempt at collecting the MVC value. MVC trial 1 trial 2 trial 3 Force or RMS 3 minute rest 3 minute rest

9 b) Verify Signal Quality Note: This is a PATENT PENDING procedure. It consists of Proprietary Material. The signal to noise ratio (SNR) of the semg signal is the critical factor that determines the ability of the decomposition algorithm to decompose the collected signal as well as the accuracy of the decomposition. The signal amplitude is influenced by: the level of muscle contraction, the location of the sensor on the surface area of the muscle, and the amount of fatty tissue between the muscle and the skin. The factors that influence SNR are: 1) the amplitude of the semg signal, 2) the level of baseline noise, 3) the level of 50/60 Hz line interference, and 4) signal clipping. The signal quality factors are automatically measured by the program and are provided to the user as a Pass/Fail display to indicate if all signal conditions for a successful decomposition have been met. If not, the contraction must be repeated. The program provides suggested remedies for correcting failure conditions which should be performed before repeating the contraction. A description of each measurement and acceptable values are provided below. Signal to Noise Ratio (SNR) The SNR value must be > 2 in order to perform decomposition. Refer to Preferred sensor Location link for locating the sensor for best SNR. The SNR is defined by the following equation: semg signal amplitude (RMS) SNR = Baseline signal amplitude (RMS) The following data plots acquired from one of the sensor channels show the baseline noise region and the signal region where the muscle is contracting. The upper plot is the amplitude of the semg signal and the lower plot is its RMS value.

10 +500 µv µvrms 0 0 Baseline signal amplitude (RMS) time (s) semg signal amplitude (RMS) 20 Baseline Noise Must be < 4.8 µv RMS The baseline noise level reflects the stability of the skin-sensor interface. The baseline noise component of the signal should be minimized to ensure the detection of low-threshold motor units. With proper skin preparation the typical baseline noise value is 3 µv RMS. The following is a plot of the baseline noise acquired from one of the sensor channels with appropriate skin preparation. +10 µv 0-10 Baseline Noise.05s Line Interference Must be < 1.0 (normalized magnitude of Power Spectrum at 50/60 Hz) Interference from power lines and other sources may adversely affect motor unit yield and accuracy. There may be excessive line interference if the demg sensor and reference electrode

11 are not properly applied. Use care when routing the sensor, input module, and reference electrode cables to avoid interference with power line noise. The following is a plot of the baseline noise acquired from one of the sensor channels showing an additional 60Hz line interference component with an unacceptable level of contamination. In this situation the demg sensor should be re-applied. Baseline Noise + 60 Hz line interference +10 µv s Signal Clipping (saturation) No signal clipping should be present in the acquired data. demg sensor detachment, reference electrode detachment, or excessive signal amplitude, may cause saturation of the amplifier output commonly referred to as Clipping. If this occurs, check the demg sensor, reference electrode, and cable. Re-position the demg sensor to reduce signal level, or reduce the peak contraction level to a lower percentage of the MVC level. The following is a plot of the semg signal acquired from one of the sensor channels showing regions of the signal where the excessive signal amplitude is clipped as indicated by the arrows. Signal Saturation Clipping +5 mv s

12 Performing a Signal Quality Check The Signal Quality Check is automatically performed and the Pass/Fail results are displayed at the completion of every tracking task to ensure that the acquired semg signal is suitable for decomposition. The Signal Quality Check may also be configured as a separate task, typically performed at the outset of the experiment in order to evaluate if the location site and skin interface of the applied sensor are suitable for decomposition analysis. The following figure shows a sample display of a successful Signal Quality Check result, expanded to see its details. The values indicate a stable, low noise, skin preparation with suitably high SNR value. The operator is instructed that the signal check is successful and that it is OK to proceed.

13 A description and usages of the Signal Quality Check features may be found in the demg User s Guide. c) Force Profile Task The tracking profile is presented to the subject on a computer screen along with the force feedback signal. The subject is instructed to contract the muscle so that their force feedback signal matches the tracking profile, while the operator monitors the display to ensure that the subject s force follows the force profile. It is recommended that the subject practice the tracking task several times to become familiar with producing a smooth force that closely follows the force profile. The screen shot below shows a 30% MVC experiment using force feedback with the target force profile in red and the subject s force trace in blue. The subject was able to smoothly follow the force profile with only small deviations from the target. 30 Force (% MVC) Time (s) At the end of the task, the demg acquisition software automatically performs an additional signal quality check on the four channels of semg data to ensure that they meet the criteria for decomposition analysis, and suggest remedies to any problems encountered.

14 d) Analysis The analyze function transfers the collected signals into the EMGworks Analysis software where the user creates a workspace to view and decompose the collected data file into its constituent MUAPs). Viewing the collected data The four semg signals and force signal collected from the tracking task can be previewed using the EMGworks analysis software plotting functions. An example of the force trace and four channel semg signals recorded from the tracking task is shown in the figure below. Force (N) 10 0 Ch 1 semg signal Amplitude (+/- 2mV) Ch 2 Ch 3 Ch semg signal Amplitude (+/- 1mV) Ch 1 Ch 2 Ch 3 Ch Time (s) 5.55

15 Amplitude (mv) The yellow highlighted regions in expanded segment of data show the occasional occurrences of individual MUAPs whose wave shapes are uncontaminated by superposition. These MUAP s can be identified by simple visual observation of their different wave shape signatures in the four channels. However, the remaining areas of the signal are typically more complex, consisting of numerous superpositions of MUAPs. The demg sensor is designed to provide this more complex signal that is representative of a large population of MUAPs, unlike the design of indwelling needle sensors that only capture a limited number of MUAPs. The demg sensor also retains the ability to discriminate between numerous MUAP wave shapes that would be lost when using sensors with electrodes having much larger surface area and 1 to 2 cm inter-electrode spacing. A comparison with between the demg sensor signal and the iemg sensor signal is provided in the following figure showing their difference in complexity. Surface EMG Signal Indwelling EMG Signal ms ms Decomposing the collected data The demg algorithms simultaneously analyze the four semg signal detected by the sensor. All four channels are needed for the decomposition algorithms to work effectively. Each MUAP has a different electrical signature in each of the four channels; this property assists the decomposition algorithms in identifying the contributory MUAPs by providing four representations of the same physiological event. The automatic decomposition algorithms use this additional information to resolve the complex signal superpositions of MUAPTs, and determine the firing times for the individual motor units. Decomposition of the data is initiated from EMGworks Analysis, and the results of the decomposition process are then made available in the workspace for further exploration.

Author: Dr. Society of Electrophysio. Reference: Electrodes. should include: electrode shape size use. direction.

Author: Dr. Society of Electrophysio. Reference: Electrodes. should include: electrode shape size use. direction. Standards for Reportin ng EMG Data Author: Dr. Roberto Merletti, Politecnico di Torino, Italy The Standards for Reporting EMG Data, written by Dr. Robertoo Merletti, are endorsed by the International Society

More information

PART 1 - INTRODUCTION...

PART 1 - INTRODUCTION... Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART

More information

Laboratory Guide. Anatomy and Physiology

Laboratory Guide. Anatomy and Physiology Laboratory Guide Anatomy and Physiology TBME04, Fall 2010 Name: Passed: Last updated 2010-08-13 Department of Biomedical Engineering Linköpings Universitet Introduction This laboratory session is intended

More information

SURFACE ELECTROMYOGRAPHY: DETECTION AND RECORDING

SURFACE ELECTROMYOGRAPHY: DETECTION AND RECORDING SURFACE ELECTROMYOGRAPHY: DETECTION AND RECORDING Carlo J. De Luca 2002 by DelSys Incorporated. All rights reserved. CONTENTS GENERAL CONCERNS... 2 CHARACTERISTICS OF THE EMG SIGNAL... 2 CHARACTERISTICS

More information

Lab 1: The Digital Oscilloscope

Lab 1: The Digital Oscilloscope PHYSICS 220 Physical Electronics Lab 1: The Digital Oscilloscope Object: To become familiar with the oscilloscope, a ubiquitous instrument for observing and measuring electronic signals. Apparatus: Tektronix

More information

HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3

HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3 3 How to Use the LIFEPAK 1000 Defibrillator HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3 This section provides an overview of information and instructions for using the LIFEPAK 1000 defibrillator. Modes

More information

ARTICLE IN PRESS. Journal of Biomechanics

ARTICLE IN PRESS. Journal of Biomechanics Journal of Biomechanics 4 () 57 579 Contents lists available at ScienceDirect Journal of Biomechanics journal homepage: www.elsevier.com/locate/jbiomech www.jbiomech.com Filtering the surface EMG signal:

More information

Interference in the ECG and its elimination

Interference in the ECG and its elimination Interference in the ECG and its elimination General points The signal voltages in ECG recording are known to be very small and are in the millivolt range, i.e. they have amplitudes of only a few thousands

More information

Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS

Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS Operation Manual Read Before Using LUMI-TENS-INS-LAB-RevA08 TABLE OF CONTENTS INTRODUCTION TO TENS INDICATIONS AND CONTRAINDICATIONS WARNINGS

More information

Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle

Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle Objectives Students will record a six-lead ECG from a resting subject and determine the QRS axis of the

More information

EEG Cap Instructions

EEG Cap Instructions EEG Cap Instructions Supplies You will need to have the following supplies at hand before applying the cap, all of which can be found in the daily preparation supplies cabinet: NuPrep mildly abrasive skin

More information

Trigno/Vicon System Integration

Trigno/Vicon System Integration Delsys and Vicon Analog Integration Motion capture systems will often have the ability to sample analog data channels as a convenient means for synchronizing external data streams with motion capture data.

More information

Spectrum and Power Measurements Using the E6474A Wireless Network Optimization Platform

Spectrum and Power Measurements Using the E6474A Wireless Network Optimization Platform Application Note Spectrum and Power Measurements Using the E6474A Wireless Network Optimization Platform By: Richard Komar Introduction With the rapid development of wireless technologies, it has become

More information

Bagnoli-2 EMG System. User s Guide

Bagnoli-2 EMG System. User s Guide Bagnoli-2 EMG System User s Guide Bagnoli TM 2-Channel Handheld EMG System User s Guide August 2009 Edition PM-B06 Copyright 2009 by Delsys Incorporated Specifications and procedures outlined in this document

More information

FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER

FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER 2014 Amplifier - 1 FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER The objectives of this experiment are: To understand the concept of HI-FI audio equipment To generate a frequency response curve for an audio

More information

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS Blood Pressure Project 9 Blood pressure is a measure of the fluid pressure within the circulatory system. This pressure is required to ensure the delivery of oxygen and nutrients to, and the removal of

More information

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.

More information

AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz

AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz 1 Sine wave with a DC offset f = frequency in Hz A = DC offset voltage (average voltage) B = Sine amplitude Vpp = 2B Vmax = A +

More information

BODY FAT & HYDRATION % MONITOR SCALE

BODY FAT & HYDRATION % MONITOR SCALE EVALUATION BODY FAT & HYDRATION % MONITOR SCALE INSTRUCTION MANUAL MODEL 12-1190 ver 7/08 BODY FAT & HYDRATION % MONITOR SCALE Introduction and Precaution: MODEL: 12-1190 1. Congratulations for purchasing

More information

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2006 Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil OBJECTIVES 1. To learn how to visualize magnetic field lines

More information

Exercise, the Electrocardiogram, and Peripheral Circulation

Exercise, the Electrocardiogram, and Peripheral Circulation iworx Physiology Lab Experiment Experiment HH-3 Exercise, the Electrocardiogram, and Peripheral Circulation Note: The lab presented here is intended for evaluation purposes only. iworx users should refer

More information

Email: tjohn@mail.nplindia.ernet.in

Email: tjohn@mail.nplindia.ernet.in USE OF VIRTUAL INSTRUMENTS IN RADIO AND ATMOSPHERIC EXPERIMENTS P.N. VIJAYAKUMAR, THOMAS JOHN AND S.C. GARG RADIO AND ATMOSPHERIC SCIENCE DIVISION, NATIONAL PHYSICAL LABORATORY, NEW DELHI 110012, INDIA

More information

Building a Simulink model for real-time analysis V1.15.00. Copyright g.tec medical engineering GmbH

Building a Simulink model for real-time analysis V1.15.00. Copyright g.tec medical engineering GmbH g.tec medical engineering GmbH Sierningstrasse 14, A-4521 Schiedlberg Austria - Europe Tel.: (43)-7251-22240-0 Fax: (43)-7251-22240-39 office@gtec.at, http://www.gtec.at Building a Simulink model for real-time

More information

Use and Application of Output Limiting Amplifiers (HFA1115, HFA1130, HFA1135)

Use and Application of Output Limiting Amplifiers (HFA1115, HFA1130, HFA1135) Use and Application of Output Limiting Amplifiers (HFA111, HFA110, HFA11) Application Note November 1996 AN96 Introduction Amplifiers with internal voltage clamps, also known as limiting amplifiers, have

More information

Inductors in AC Circuits

Inductors in AC Circuits Inductors in AC Circuits Name Section Resistors, inductors, and capacitors all have the effect of modifying the size of the current in an AC circuit and the time at which the current reaches its maximum

More information

The Sonometer The Resonant String and Timbre Change after plucking

The Sonometer The Resonant String and Timbre Change after plucking The Sonometer The Resonant String and Timbre Change after plucking EQUIPMENT Pasco sonometers (pick up 5 from teaching lab) and 5 kits to go with them BK Precision function generators and Tenma oscilloscopes

More information

T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p

T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p Data Transmission Concepts and terminology Transmission terminology Transmission from transmitter to receiver goes over some transmission medium using electromagnetic waves Guided media. Waves are guided

More information

Department of Electrical and Computer Engineering Ben-Gurion University of the Negev. LAB 1 - Introduction to USRP

Department of Electrical and Computer Engineering Ben-Gurion University of the Negev. LAB 1 - Introduction to USRP Department of Electrical and Computer Engineering Ben-Gurion University of the Negev LAB 1 - Introduction to USRP - 1-1 Introduction In this lab you will use software reconfigurable RF hardware from National

More information

Standards for surface electromyography: the European project "Surface EMG for non-invasive assessment of muscles (SENIAM)

Standards for surface electromyography: the European project Surface EMG for non-invasive assessment of muscles (SENIAM) Standards for surface electromyography: the European project "Surface EMG for non-invasive assessment of muscles (SENIAM) D.F. Stegeman 1,3, H.J. Hermens 2 1 Institute of Neurology, Department of Clinical

More information

Logging of RF Power Measurements

Logging of RF Power Measurements Logging of RF Power Measurements By Orwill Hawkins Logging of measurement data is critical for effective trend, drift and Exploring the use of RF event analysis of various processes. For RF power measurements,

More information

Current Probes. User Manual

Current Probes. User Manual Current Probes User Manual ETS-Lindgren L.P. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall

More information

Data Analysis Methods: Net Station 4.1 By Peter Molfese

Data Analysis Methods: Net Station 4.1 By Peter Molfese Data Analysis Methods: Net Station 4.1 By Peter Molfese Preparing Data for Statistics (preprocessing): 1. Rename your files to correct any typos or formatting issues. a. The General format for naming files

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

Transcutaneous Electrical Nerve Stimulation Device. Operation Manual Read Before Using

Transcutaneous Electrical Nerve Stimulation Device. Operation Manual Read Before Using Transcutaneous Electrical Nerve Stimulation Device Operation Manual Read Before Using TABLE OF CONTENTS INTRODUCTION TO TENS INDICATIONS AND CONTRAINDICATIONS WARNINGS PRECAUTIONS/ADVERSE REACTIONS ABOUT

More information

Activity 4.2.3: EKG. Introduction. Equipment. Procedure

Activity 4.2.3: EKG. Introduction. Equipment. Procedure Activity 4.2.3: EKG The following is used with permission of Vernier Software and Technology. This activity is based on the experiment Analyzing the Heart with EKG from the book Human Physiology with Vernier,

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 2011 Interference - 1 INTERFERENCE OF SOUND WAVES The objectives of this experiment are: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves. To observe interference phenomena

More information

RLC Series Resonance

RLC Series Resonance RLC Series Resonance 11EM Object: The purpose of this laboratory activity is to study resonance in a resistor-inductor-capacitor (RLC) circuit by examining the current through the circuit as a function

More information

Introduction to Data Acquisition

Introduction to Data Acquisition Introduction to Data Acquisition Overview This tutorial is part of the National Instruments Measurement Fundamentals series. Each tutorial in this series, will teach you a specific topic of common measurement

More information

FREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5

FREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5 Physics 161 FREE FALL Introduction This experiment is designed to study the motion of an object that is accelerated by the force of gravity. It also serves as an introduction to the data analysis capabilities

More information

Manual Analysis Software AFD 1201

Manual Analysis Software AFD 1201 AFD 1200 - AcoustiTube Manual Analysis Software AFD 1201 Measurement of Transmission loss acc. to Song and Bolton 1 Table of Contents Introduction - Analysis Software AFD 1201... 3 AFD 1200 - AcoustiTube

More information

NEULOG HEART RATE AND PULSE LOGGER SENSOR GUIDE

NEULOG HEART RATE AND PULSE LOGGER SENSOR GUIDE NeuLog heart rate and pulse logger sensor NUL-208 The NeuLog pulse sensor can be used for any science experiment which requires a human heart rate or pulse measurement such as in the fields of Biology,

More information

Table of Contents. Introduction. Radio Frequency Control. Initial setup Setting the time. Setting the temperature

Table of Contents. Introduction. Radio Frequency Control. Initial setup Setting the time. Setting the temperature LHZ USER GUIDE LHZ User Guide LHZ would like to thank you for purchasing your LHZ radiator. You have purchased a high quality, German designed and manufactured heating device, which will give you soft

More information

CONVECTION CURRENTS AND ANOMALOUS BEHAVIOUR OF WATER

CONVECTION CURRENTS AND ANOMALOUS BEHAVIOUR OF WATER CONVECTION CURRENTS AND ANOMALOUS BEHAVIOUR OF WATER Objective: To compare the thermal behaviour of water with that of other liquids, specifically alcohol and edible oil. To point out the anomaly of water

More information

Lab 3: Introduction to Data Acquisition Cards

Lab 3: Introduction to Data Acquisition Cards Lab 3: Introduction to Data Acquisition Cards INTRODUCTION: In this lab, you will be building a VI to display the input measured on a channel. However, within your own VI you will use LabVIEW supplied

More information

WIRELESS INSTRUMENTATION TECHNOLOGY

WIRELESS INSTRUMENTATION TECHNOLOGY BS&B WIRELESS, L.L.C. BS&B WIRELESS, L.L.C. WIRELESS INSTRUMENTATION TECHNOLOGY Printed February 2004 BS&B WIRELESS, L.L.C. 7422-B East 46th Place, Tulsa, OK74145 Phone: 918-622-5950 Fax: 918-665-3904

More information

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON. WHY DIFFERENTIAL? Voltage, The Difference Whether aware of it or not, a person using an oscilloscope to make any voltage measurement is actually making a differential voltage measurement. By definition,

More information

Sterile Dressing Change with Tegaderm CHG for Central Venous Catheter (CVC)

Sterile Dressing Change with Tegaderm CHG for Central Venous Catheter (CVC) Sterile Dressing Change with Tegaderm CHG for Central Venous Catheter (CVC) The dressing protects your catheter site. It also helps prevent infection at the site. Keep your dressing clean and dry at all

More information

R Series Quick Reference Guide

R Series Quick Reference Guide R Series Quick Reference Guide 2010 ZOLL Medical Corporation. All rights reserved. CPR Index, OneStep, R Series, Real CPR Help, See-Thru CPR, statpadz, and ZOLL are trademarks or registered trademarks

More information

Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies

Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies Soonwook Hong, Ph. D. Michael Zuercher Martinson Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies 1. Introduction PV inverters use semiconductor devices to transform the

More information

PR 50 Frymaster Universal Holding Cabinet (UHC)

PR 50 Frymaster Universal Holding Cabinet (UHC) PREP EQUIPMENT PR 50 Frymaster Universal Holding Cabinet (UHC) Models UHCP2, UHCP4, UHCPN4 Daily maintenance tasks PR 50 D1 Clean UHC Monthly maintenance tasks PR 50 M1 Calibrate UHC Annual maintenance

More information

iworx Sample Lab Experiment HN-2: Stretch Receptors and Reflexes

iworx Sample Lab Experiment HN-2: Stretch Receptors and Reflexes Experiment HN-2: Stretch Receptors and Reflexes Exercise 1: Achilles Tendon Reflex Aim: To determine conduction time from tendon tap to response of the gastrocnemius muscle in the Achilles tendon reflex

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

Features of Corona 2.4GHz transmitting module and receiver. The new lightweight, advanced technology unit has the following features:

Features of Corona 2.4GHz transmitting module and receiver. The new lightweight, advanced technology unit has the following features: Instruction Manual Corona 2.4GHz RF module & Receiver Thank you for purchasing our Corona 2.4Ghz RF transmitter module & receiver. In order to fully utilize the performance potential of this system, we

More information

Introduction to the TI-Nspire CX

Introduction to the TI-Nspire CX Introduction to the TI-Nspire CX Activity Overview: In this activity, you will become familiar with the layout of the TI-Nspire CX. Step 1: Locate the Touchpad. The Touchpad is used to navigate the cursor

More information

CurveMaker v2.1 DYNAFS programmable ignition software

CurveMaker v2.1 DYNAFS programmable ignition software CurveMaker v2.1 DYNAFS programmable ignition software Dynatek 164 S Valencia St. Glendora CA 91741 phone (626)963-1669 fax (626)963-7399 Contents 1) Installation...1 2) Overview...1 3) Programming a Curve...4

More information

Sound Pressure Measurement

Sound Pressure Measurement Objectives: Sound Pressure Measurement 1. Become familiar with hardware and techniques to measure sound pressure 2. Measure the sound level of various sizes of fan modules 3. Calculate the signal-to-noise

More information

Dash 18X / Dash 18 Data Acquisition Recorder

Dash 18X / Dash 18 Data Acquisition Recorder 75 Dash 18X / Dash 18 Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 3.1 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH

More information

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURE STANDARD OPERATING PROCEDURE DATA ACQUISITION USING THE C1 MEMOCONE SYSTEM REVISION DATE: OKT 21, 2011 Prepared by Per Ljunggren TABLE OF CONTENTS STANDARD OPERATING PROCEDURE...3 DATA ACQUISITION USING

More information

RGB for ZX Spectrum 128, +2, +2A, +3

RGB for ZX Spectrum 128, +2, +2A, +3 RGB for ZX Spectrum 128, +2, +2A, +3 Introduction... 2 Video Circuitry... 3 Audio Circuitry... 8 Lead Wiring... 9 Testing The Lead... 11 Spectrum +2A/+3 RGB Differences... 12 Circuitry Calculations...

More information

User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction

User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 407780 Product Page Integrating Sound Level Datalogger

More information

Lastest Development in Partial Discharge Testing Koh Yong Kwee James, Leong Weng Hoe Hoestar Group

Lastest Development in Partial Discharge Testing Koh Yong Kwee James, Leong Weng Hoe Hoestar Group Lastest Development in Partial Discharge Testing Koh Yong Kwee James, Leong Weng Hoe Hoestar Group INTRODUCTION Failure of High Voltage insulation is the No 1 cause of High voltage system failures with

More information

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz Author: Don LaFontaine Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz Abstract Making accurate voltage and current noise measurements on op amps in

More information

Experiment MP-1: Electromyogram (EMG) Activity and Muscle Strength

Experiment MP-1: Electromyogram (EMG) Activity and Muscle Strength Experiment MP-1: Electromyogram (EMG) Activity and Muscle Strength Calibrating the Hand Dynamometer 1. Collect 5 textbooks or use the barbell weight. Weigh the stack of books on the bathroom scale. Record

More information

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2009 Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil OBJECTIVES 1. To learn how to visualize magnetic field lines

More information

Tektronix RSA306 USB Spectrum Analyzer

Tektronix RSA306 USB Spectrum Analyzer Tektronix RSA306 USB Spectrum Analyzer Simple Demos The Demo of the RSA306 is easy. Even you do not have signal generators, devices under test, or demo boards, using the whip antenna provided in box, you

More information

5. Tutorial. Starting FlashCut CNC

5. Tutorial. Starting FlashCut CNC FlashCut CNC Section 5 Tutorial 259 5. Tutorial Starting FlashCut CNC To start FlashCut CNC, click on the Start button, select Programs, select FlashCut CNC 4, then select the FlashCut CNC 4 icon. A dialog

More information

Transmitter Interface Program

Transmitter Interface Program Transmitter Interface Program Operational Manual Version 3.0.4 1 Overview The transmitter interface software allows you to adjust configuration settings of your Max solid state transmitters. The following

More information

24 Wide Screen LED Monitor with HDMI

24 Wide Screen LED Monitor with HDMI 24 Wide Screen LED Monitor with HDMI USER MANUAL L24LE11B_MANUAL_EN_R1 Please read this manual carefully before using, and keep it for future reference. TABLE OF CONTENTS GETTING STARTED Monitor Stand

More information

Maximizing Throughput and Coverage for Wi Fi and Cellular

Maximizing Throughput and Coverage for Wi Fi and Cellular Maximizing Throughput and Coverage for Wi Fi and Cellular A White Paper Prepared by Sebastian Rowson, Ph.D. Chief Scientist, Ethertronics, Inc. www.ethertronics.com March 2012 Introduction Ask consumers

More information

Wireless Alarm System. Panic Button. User s Manual. Choice ALERT. Control all sensors, accessories and remote transmitters from one location

Wireless Alarm System. Panic Button. User s Manual. Choice ALERT. Control all sensors, accessories and remote transmitters from one location Wireless Alarm System Panic Button User s Manual Choice ALERT Control all sensors, accessories and remote transmitters from one location Table of Contents Important Safeguards 4 Introduction 5 Activating

More information

DSAM Digital Quality Index (DQI) A New Technique for Assessing Downstream Digital Services

DSAM Digital Quality Index (DQI) A New Technique for Assessing Downstream Digital Services Application Note DSAM Digital Quality Index (DQI) A New Technique for Assessing Downstream Digital Services Overview As cable operators move to digital simulcast and all digital networks, the majority

More information

Improving ECG Quality

Improving ECG Quality Improving ECG Quality Application Note Introduction Good ECG signal quality it is something every clinician strives to secure every time they monitor a patient s heart. This application note will look

More information

Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide)

Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide) Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide) OVERVIEW Students will measure a sound wave by placing the Ward s DataHub microphone near one tuning fork A440 (f=440hz). Then

More information

Classic EEG (ERPs)/ Advanced EEG. Quentin Noirhomme

Classic EEG (ERPs)/ Advanced EEG. Quentin Noirhomme Classic EEG (ERPs)/ Advanced EEG Quentin Noirhomme Outline Origins of MEEG Event related potentials Time frequency decomposition i Source reconstruction Before to start EEGlab Fieldtrip (included in spm)

More information

Pulse Width Modulation (PWM) LED Dimmer Circuit. Using a 555 Timer Chip

Pulse Width Modulation (PWM) LED Dimmer Circuit. Using a 555 Timer Chip Pulse Width Modulation (PWM) LED Dimmer Circuit Using a 555 Timer Chip Goals of Experiment Demonstrate the operation of a simple PWM circuit that can be used to adjust the intensity of a green LED by varying

More information

Impedance 50 (75 connectors via adapters)

Impedance 50 (75 connectors via adapters) VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:

More information

Heart Rate Data Collection Software Application User Guide

Heart Rate Data Collection Software Application User Guide Y O R K B I O F E E D B A C K w w w. y o r k - b i o f e e d b a c k. c o. u k G l y n B l a c k e t t Heart Rate Data Collection Software Application User Guide Table of Contents 1 Introduction...1 2

More information

HEART RATE + CALORIE MONITOR

HEART RATE + CALORIE MONITOR S7 S7 HEART RATE + CALORIE MONITOR SB4065IS01SPORB.indd 1 TABLE OF CONTENTS INTRODUCTION 1 HOW TO USE THIS MANUAL 1 ACTIVATING THE WATCH FROM POWER DOWN MODE 1 WARNINGS & CAUTIONS 2 CARE & MAINTENANCE

More information

Adding Heart to Your Technology

Adding Heart to Your Technology RMCM-01 Heart Rate Receiver Component Product code #: 39025074 KEY FEATURES High Filtering Unit Designed to work well on constant noise fields SMD component: To be installed as a standard component to

More information

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of the blood 7 Procedure to test glucose 7 Procedure to

More information

ELECTRON SPIN RESONANCE Last Revised: July 2007

ELECTRON SPIN RESONANCE Last Revised: July 2007 QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron

More information

Data Analysis 1. Click the 2-Cursor icon (Figure HH-7-L1) so that two blue vertical lines appear over the recording window.

Data Analysis 1. Click the 2-Cursor icon (Figure HH-7-L1) so that two blue vertical lines appear over the recording window. Experiment HH-7: Electrocardiograms from the Six Chest Leads Exercise 1: ECG from the Chest Leads of a Resting Subject Aim: To record electrocardiograms from the six chest leads and determine the transition

More information

BrainMaster tm System Type 2E Module & BMT Software for Windows tm. Helpful Hints

BrainMaster tm System Type 2E Module & BMT Software for Windows tm. Helpful Hints . BrainMaster tm System Type 2E Module & BMT Software for Windows tm Helpful Hints 1995-2004 BrainMaster Technologies, Inc., All Rights Reserved BrainMaster and From the Decade of the Brain are registered

More information

Product Introduction MyoMuscle. Telemyo DTS-System. Telemetry system for EMG and Biomechanical Sensors. www.noraxon.com

Product Introduction MyoMuscle. Telemyo DTS-System. Telemetry system for EMG and Biomechanical Sensors. www.noraxon.com Product Introduction MyoMuscle Telemyo DTS-System Telemetry system for EMG and Biomechanical Sensors DTS Device Family 3 major device categories can be used for the Direct Transmission System technology

More information

Application Note, Rev.1.0, September 2008 TLE8366. Application Information. Automotive Power

Application Note, Rev.1.0, September 2008 TLE8366. Application Information. Automotive Power Application Note, Rev.1.0, September 2008 TLE8366 Automotive Power Table of Contents 1 Abstract...3 2 Introduction...3 3 Dimensioning the Output and Input Filter...4 3.1 Theory...4 3.2 Output Filter Capacitor(s)

More information

HIGH RELIABILITY POWER SUPPLY TESTING

HIGH RELIABILITY POWER SUPPLY TESTING HIGH RELIABILITY POWER SUPPLY TESTING Dale Cigoy Keithley Instruments, Inc. The reliability of a power supply must match or exceed the rest of the system in which it is installed. Generally, this requires

More information

K-Type Thermocouple Sensor User s Guide

K-Type Thermocouple Sensor User s Guide K-Type Thermocouple Sensor User s Guide 1 TABLE OF CONTENTS: 1 INTRODUCTION... 2 2 TYPICAL APPLICATION:... 2 3 INSTALLATION RULES:... 2 3.1 Connecting the sensor to M1/MD4 data logger:... 2 3.2 Connecting

More information

For all turfgrass areas. Now you can test soil reaction at different levels in the soil profile. Check Soil ph on:

For all turfgrass areas. Now you can test soil reaction at different levels in the soil profile. Check Soil ph on: Turf-Tec ph Meter Test soil ph in the mat, thatch and above and below the root zone. ph can be checked, 0" to 4" inch level right on the turfgrass area. Simply insert probe to desired depth, allow to stand

More information

Module Connectors and Cable Specifications

Module Connectors and Cable Specifications Module Connectors, page 1 Cables and Adapters, page 2 Cleaning the Fiber-Optic Connectors, page 6 Module Connectors RJ-45 Connector The RJ-45 connector is used to connect a Category 3, Category 5, Category

More information

Current Loop Tuning Procedure. Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) General Procedure AN-015

Current Loop Tuning Procedure. Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) General Procedure AN-015 Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) The standard tuning values used in ADVANCED Motion Controls drives are conservative and work well in over 90%

More information

PowlVac Vacuum Integrity Tester

PowlVac Vacuum Integrity Tester Instructions IB-60025B PowlVac Vacuum Integrity Tester POWELL ELECTRICAL MANUFACTURING COMPANY 8550 MOSLEY DRIVE HOUSTON, TEXAS 77075 USA PHONE (713) 944-6900 FAX (713) 947-4453 www.powellelectric.com

More information

Measuring Temperature withthermistors a Tutorial David Potter

Measuring Temperature withthermistors a Tutorial David Potter NATIONAL INSTRUMENTS The Software is the Instrument Application Note 065 Measuring Temperature withthermistors a Tutorial David Potter Introduction Thermistors are thermally sensitive resistors used in

More information

Wireless Medical Telemetry Laboratory

Wireless Medical Telemetry Laboratory Wireless Medical Telemetry Laboratory 0 Introduction The development of wireless medical telemetry has become an increasingly popular application in recent years. As the elderly population continues to

More information

Event recording (or logging) with a Fluke 287/289 Digital Multimeter

Event recording (or logging) with a Fluke 287/289 Digital Multimeter Event recording (or logging) with a Fluke 287/289 Digital Multimeter Our thanks to Fluke Corp for allowing us to reprint the following article. One of the major features of the Fluke 280 Series digital

More information

USB 3.0* Radio Frequency Interference Impact on 2.4 GHz Wireless Devices

USB 3.0* Radio Frequency Interference Impact on 2.4 GHz Wireless Devices USB 3.0* Radio Frequency Interference Impact on 2.4 GHz Wireless Devices White Paper April 2012 Document: 327216-001 INFORMATION IN THIS DOCUMENT IS PROVIDED IN CONNECTION WITH INTEL PRODUCTS. NO LICENSE,

More information

EECC694 - Shaaban. Transmission Channel

EECC694 - Shaaban. Transmission Channel The Physical Layer: Data Transmission Basics Encode data as energy at the data (information) source and transmit the encoded energy using transmitter hardware: Possible Energy Forms: Electrical, light,

More information

LogicLab s.r.l Via della valle 67 20048 Carate Brianza (MI) Tel. +39 0362 805287 Fax. +39 0362 1914102 www.logiclab.it

LogicLab s.r.l Via della valle 67 20048 Carate Brianza (MI) Tel. +39 0362 805287 Fax. +39 0362 1914102 www.logiclab.it Test system DV1512A2NCI Single Current Generator Maximum current 15 Arms in continuity and 30 Arms per 1 s Single Voltage generator 150 Vrms Programmable phase between Current and Voltage Programmable

More information

SmartDiagnostics Application Note Wireless Interference

SmartDiagnostics Application Note Wireless Interference SmartDiagnostics Application Note Wireless Interference Publication Date: May 27, 2015 KCF Technologies, Inc. Background The SmartDiagnostics wireless network is an easy to install, end-to-end machine

More information

SYMBOL MC9060 HAND HELD TERMINAL USER MANUAL

SYMBOL MC9060 HAND HELD TERMINAL USER MANUAL SYMBOL MC9060 HAND HELD TERMINAL USER MANUAL CORPORATE VERSION FEBRUARY 2007 TABLE OF CONTENTS Welcome... 3 Getting Started... 3 Equipment Setup... 3 Battery Installation... 4 About the Stylus... 5 Aligning

More information

Pre Site Survey Questionnaire

Pre Site Survey Questionnaire Pre Site Survey Questionnaire Version 1.15 Page 1 Contents List 1 DOCUMENT VERSION HISTORY...3 2 INTRODUCTION...4 3 PURPOSE & SCOPE...5 4 SITE DETAILS...6 4.1 End User Details... 6 4.2 Contact Details...

More information