FLUOROMETER PERFORMANCE EVALUATION REPORT 9/2002
|
|
- Clifford Hoover
- 7 years ago
- Views:
Transcription
1 FLUOROMETER PERFORMANCE EVALUATION REPORT 9/22 ABSTRACT A series of performance evaluation tests were conducted at three tests sites to investigate the overall performance of the Turner Designs Submersible Fluorometer and the chlorophyll probe. Performance was determined through results designed to determine minimum detection limit of chlorophyll a, dynamic range, accuracy of readings to actual chlorophyll a, ambient light rejection, and response time. The results indicate that the fluorometer outperforms the sensor but the performance of both sensors is adequate for most environmental conditions. A performance limitation of the is the response time. Using a moving average calculation, it responds relatively slowly to changing concentrations and would not be effective for vertical profiling applications. In some extreme conditions, such as chlorophyll concentrations less than 1µg/L and high ambient light and/or high concentrations of interfering compounds, the may also not provide accurate measures of algal biomass. INTRODUCTION A series of performance evaluation tests were conducted to investigate the overall performance of the Turner Designs Submersible Fluorometer and the chlorophyll probe. Tests were conducted at three sites; Moss Landing Marine Laboratory, Turner Designs Testing Laboratory and the USGS Menlo Park facility. Performance was evaluated through designing tests to investigate the following areas; minimum detection limit, instrument range, response time, light rejection and accuracy of fluorometer reading to actual chlorophyll a concentration. Due to different specification standards, comparing instruments on published specifications is not always straightforward. Below is the published performance data for each instrument. Through our testing we have attempted to compare the instruments in a meaningful and informative way for the customer. We have compared instruments identically and have attempted to present the data in as an objective manner as possible. In most cases the raw fluorescence signal was recorded in order to eliminate scaling issues related to calibration. The raw fluorescence was recorded as %FS which indicates the % of a sample s signal in relation to 1% full scale of the particular instrument. Calibrated data is referred to as in vivo fluorescence or. Published Performance Specifications Range -4µg/L 4 orders of magnitude Minimum Detection Limit N/A.2 µg/l Resolution.1 µg/l,.1%fs 12 bit Max Sampling Rate NA 5Hz Depth -61m -6m Instrument Overview: While both instruments are chlorophyll fluorometers there are several key distinctions between the two units. Firstly, the physical size and appearance is significantly different. The fluorometer is designed to operate as a stand-alone unit and can also be interfaced to multi-parameter systems. Being a stand-alone unit, it incorporates many additional features that are not required in the probe. The following is a list of unique features; 6m depth capability, temperature compensation, internal data logging, autoranging, turbidity option, digital and analog output, configurable analog (-5V) limits, and calibration software. The presence of these features has resulted in a much different looking sensor. However, when interfaced to a multi-parameter system, the operates in a very similar manner to the probe. The has been designed to interface directly onto multi-parameter sondes. The probe is also now available on an optical sonde that only offers fluorescence, conductivity, 1 FLUOROMETER COMPARISON REPORT
2 temperature and depth. The probe provides an analog signal to the sonde which integrates the data from all of the sonde sensors into one digital data stream. The provides both digital and analog data continuously. MATERIALS AND METHODS All tests were conducted under controlled settings in a laboratory. In most cases, monocultures on algae were used as the test media (see table below for algal culture details). In some cases natural algal samples were used. Chlorophyll a extractions were conducted on all samples using the Non-acidification fluorometric method (Welschmeyer, 1997) and the Acidification technique (E.P.A. Method 445.) in the case of the pheophytin testing. Algal samples used: Genus Algal group 1 Accessory Pigment Dunaliella sp. chlorophyte chlorophyll b Thallassiosira sp. diatom chlorophyll c Phaeodactylum sp. diatom chlorophyll c San Francisco Bay Delta natural assemblage RESULTS MINIMUM DETECTION LIMIT The Minimum Detection Limit is the lowest concentration of algae in clean culture media than can is distinguishable from a blank culture media solution. Three detection limit tests were run by mixing a dilution series of algal cultures. The readings were allowed to stabilize and at least 1 datapoints were recorded per sample. Dunaliella Phaeodactylum Chl A Chl A Thallassiosira Chl A Table 1: Dunaliella and Phaeodactylum test data from Moss Landing Marine Laboratory (MLML). Data from Thallassiosira test from Turner Designs. The Chloro column is extracted chlorophyll a data. The and column is the in vivo fluorescence data. From the data shown it is clear the difference in detection limit between algal groups. The data supports the fact that the was unable to distinguish some algal cultures at concentrations of 1µg/L chlorophyll a or less while the was capable of distinguishing all samples. The lowest concentration tested was.34µg/l of Thallassiosira culture. Because each species of algae has varying in vivo fluorescence ():extracted chlorophyll, the limits of detection can vary between species. The demonstrated the ability to detect concentration as low as.79µg/l of Thallassiosira culture while not detecting other cultures that were above 1µg/L. 2 FLUOROMETER COMPARISON REPORT
3 Minimum Detection Limit Test #1 Minimum Detection Limit Test # R 2 =.9737 R 2 = chlorophyll a (ug/l) Graph 1 & 2: MDL test data from Moss Landing Marine Lab showing that the not differentiating between 2.5µg/L and 1.1µg/L Dunaliella samples. MDL Test#2 shows data from tests with Thallassiosira cultures indicate that the did not distinguish the blank and the first two culture samples. RANGE Both instruments provide an analog -5V output signal while the has a digital output as well. The -5V ranges are factory set for a given fluorescence range. The comes standard with a -1µg/L analog range while the comes standard with a -4µg/L range. As the analog range increases, the resolution of the data decreases and it is therefore advantageous to set the -5V limits to the environments in which you work. The digital outputs from the Sonde and the were used for all tests. Both instruments demonstrated good linearity for typical chlorophyll a ranges found in natural environments. The only difference of note for chlorophyll measurement is the ability to customize the -5V range on the. This abiltiy allows customers working in low chlorophyll environments to decrease the -5V range and therefore improve the resolution of the analog data R 2 =.9319 R 2 = chlorophyll a (ug/l) Linearity Test R 2 =.997 R 2 = chlorophyll a (ug/l) Graph 3: Data taken with calibrated values at MLML using Dunaliella cultures. 3 FLUOROMETER COMPARISON REPORT
4 ACCURACY In order to test the accuracy of measuring chlorophyll a fluorescence, the effects of various interfering compounds were investigated. In an ideal situation, an in situ fluorometer would measure chlorophyll a fluorescence only since chlorophyll a is the only photosynthetic pigment that all algae and photosynthetic bacteria have in common. By measuring chlorophyll a only you will obtain the best estimate of total, living algal biomass. Compounds that can interfere with fluorescent signals optimized to detect chlorophyll a include accessory pigments such as chlorophyll b and c, degraded chlorophyll a, pheophytin a, and dissolved organic matter. All of these interfering substances have fluorescence spectra that can overlap with chlorophyll a given the appropriate concentrations and fluorometer optics. When discussing an instrument s susceptibility to interfering compounds, the performance is determined by the optics. The optics in a fluorometer encompass three components; a light source, optical filters and a light detector. Both the and the instrument use solid-state components, LED light source and photodiode light detector, which makes them equivalent in this category except that the utilizes two LED lamps while the uses one. The difference between the optics lies in the optical filters, specifically the emission filters. The emission filter is placed in front of the light detector. It s purpose is to block all light except for the wavelengths of interest, in this case chlorophyll a fluorescence which peaks at 685nm. The more specific the filter is to the chlorophyll peak, the more accurate the fluorometer will be at measuring total algal biomass. The uses long-pass emission filters which allows in all light >6nm. This means that any material other than chlorophyll a that is excited by the blue LED light source (peak at 46nm) and fluorescing above 6nm will be detected. Likely candidates for this include pheophytin a, chlorophyll b, chlorophyll c and CDOM. The, on the other hand, utilizes a band-pass filter that allows in light from 66-7nm only, peaking at 685nm. Therefore, the fluorescence from compounds other than chlorophyll a are much less likely to be detected. The benefit of using a long-pass filter is the low material cost and higher transmission. The downside is greater susceptibility to interfering compounds. The tests were effective in demonstrating the effect of optical components in the pheophytin testing. One of the most likely interference s is pheophytin a. Pheophytin can accumulate in water column at density gradients, close to the bottom, and be increased due to active zooplankton feeding. A test was run to compare the fluorescence signal from a healthy algal culture and fluorescence from the same culture after exposed to a weak acid that converted much of the chlorophyll in the cells to pheophytin a. 1mL of 1N HCl was added to 35mL of Thallasisosira culture. The culture was stirred for 1 minute and then the fluorescence was read using both instruments. The results were as follows: Before After Before After FS% Ratio FLUOROMETER COMPARISON REPORT
5 The Difference in the ratios indicates a difference in the fluorometers ability to detect pheophytin fluorescence. The ratio of 2.76 indicates that it is less sensitive to pheophytin fluorescence than the sensor that had a ratio of Regression Comprison R TD USGS-Thal TD-Thal TD-Thal2 MLML-Pheo MLML-Pheo MLML-Dun MLML-Dun MLML-Dun TD Fresh USGS-Delta Graph 4: The bar graph displays the correlations (r 2 ) obtained from the minimum detection and linearity tests conducted at all three testing sites. In most cases the data correlated with actual chlorophyll a more closely. All tests returned an r 2 of greater than.85 or better. The samples market TD Fresh and USGS-Delta are natural water samples with an assemblage of algal species. All other samples were moncultures of algae. AMBIENT LIGHT REJECTION Ambient light rejection refers to the ability of the fluorometer to be unaffected by sunlight or other ambient light sources. Light rejection is typically performed by flashing the LED light source. When the light source is on, the instrument detects the fluorescence and any ambient light, when the light source is off, the instrument detects the ambient light. The instrument compares the two to determine the fluorescence. Ambient light rejection test results indicate that there is a difference in how the two instruments correct ambient light. The displayed no significant difference in fluorescence signal while the showed a decrease in signal as ambient light increased and became more variable Ambient Light Rejection Test Scufa no light light at 9 degrees light at 18 degrees Position of Light Source % Full Scale Ambient Light Rejection Test #2 Lit Lab Dark Lab Light 12" Direct Light 12" Variable Graph 5 & 6: Graph 5 displays data from Moss Landing Marine Lab. The instruments were placed in a solution of rhodamine WT dye. A halogen light source was used to test ambient light. Graph 6 displays data from tests run at Turner Designs with a halogen light source. The units were not placed in a solution. 5 FLUOROMETER COMPARISON REPORT
6 RESPONSE TIME The graph below shows the response time of the two instruments as they were alternately placed into samples of natural water containing algae and blank deionized water. The readings were taken at one-second intervals. The difference in the graphs displays a significant difference between the two instruments. The system uses a moving average as opposed to the realtime sampling of the. The moving average allows the system to smooth the data as well as enables it to improve the limit of detection. The, designed for high-speed vertical profiling applications, samples at a rate of 1Hz and outputs data at 1Hz or 5Hz. The spikes at the beginning of the readings of the algal cultures is due to removing the instrument from the blank water and then placing it into an algal culture. The change from air to liquid creates some bubbles that cause spikes in the readings. The sensor is less sensitive to these changes due to the slow moving average (~1sec) of the data which smoothes the data but does not supply instantaneous data. RESPONSE TIME TEST IN VIVO FLUORESCENCE :55:55 13:56:38 13:57:22 13:58:5 13:58:48 13:59:31 14::14 14::58 14:1:41 TIME Graph 7 CONCLUSIONS The results of the performance tests indicate that both instruments are effective at estimating chlorophyll a concentrations in typical environmental conditions. Although accuracy tests favored the fluorometer, most correlations between the in vivo fluorescence and the actual chlorophyll concentration exceeded.9 for both instruments. It is in the presence of interfering compounds where the data from the two units may differ significantly. Results from testing with degraded chlorophyll, pheophytin, support this conclusion, with the not differentiating between chlorophyll and pheophytin as well as the. Other extreme environmental conditions could limit the effectiveness of the probe. Conditions such as high ambient light,from shallow waters with reflective sediment, and low productivity waters could be problematic. Also, due to the slow response time of the probe, vertical profiling would be a challenge. On the other hand, the probe seems well suited for deployed use in coastal or freshwater environments where chlorophyll levels exceed 1µg/L and ambient light and interfering compounds are not a significant issue. The performed well in all categories. It is a more versatile instrument because of ability to be used in self-contained mode. 6 FLUOROMETER COMPARISON REPORT
7 REFERENCES: Arar, E.J. and G.B. Collins In vitro determination of chlorophyll a and pheophytin a in marine and freshwater algae by fluorescence. E.P.A. Method 445., Revision 1.2 Welschmeyer, N.A Fluorometeric analysis of chlorophyll a in the presence of chlorophyll b and pheopigments. Limnol. Oceanogr. 39: TESTING FACILITIES: MOSS LANDING MARINE LABORATORY, MOSS LANDING, CA. Tests run under the supervision of Dr. Nick Welschmeyer. TURNER DESIGNS TESTING FACILITY, SUNNYVALE, CA. UNITED STATES GEOLOGICIAL SURVEY, MENLO PARK, CA Report written by Robert Ellison, Turner Designs, Inc., 22 7 FLUOROMETER COMPARISON REPORT
An Introduction to Algae Measurements Using In Vivo Fluorescence
An Introduction to Algae Measurements Using In Vivo Fluorescence Submersible fluorescence sensors enable real-time field estimates of phytoplankton that can be directly correlated to standard laboratory
More informationHow to measure Chlorophyll-a
World Bank & Government of The Netherlands funded Training module # WQ - 40 How to measure Chlorophyll-a New Delhi, March 2000 CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas, New Delhi 11 00 16
More informationFundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
More informationSpectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
More informationALMOFRONT 2 cruise in Alboran sea : Chlorophyll fluorescence calibration
Vol. 3 : 6-11, 2010 Journal of Oceanography, Research and Data ALMOFRONT 2 cruise in Alboran sea : Chlorophyll fluorescence calibration CUTTELOD Annabelle 1,2 and CLAUSTRE Hervé 1,2 1 UPMC, Univ. Paris
More informationDetermining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy
Determining the Free Chlorine Content of Swimming Pool Water Computer 33 Physicians in the nineteenth century used chlorine water as a disinfectant. Upon the discovery that certain diseases were transmitted
More informationDSM155 Digital LED Light Meter User s Manual DSM155
DSM155 Digital LED Light Meter User s Manual DSM155 CONTENTS TABLE OF CONTENTS......3 LIMITED WARRANTY AND LIMITATION OF LIABILITY......3 Out of Box....3 Accessories....3 Safety Information....3 Certification....5
More informationPhosphorus, Reactive (Orthophosphate)
Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from
More informationEnvironmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
More informationEPA New England Regional Laboratory. EPA Project Team: Liz McCarthy. Matt Destino Tim Bridges Dick Siscanaw
EPA s River Real Time Monitoring Buoy in the Charles and Mystic Watersheds Access to the data has been through a password protected site http://www.ysieconet.com/ Tom Faber EPA New England Regional Laboratory
More informationNederland België / Belgique
Cronus Brochure 2 Introduction Cronus: The first spectrometer and colorimeter in one The Cronus is world s first spectrocolorimeter combining a high en VIS spectrometer with a high speed XYZ colorimeter.
More informationBACTERIAL ENUMERATION
BACTERIAL ENUMERATION In the study of microbiology, there are numerous occasions when it is necessary to either estimate or determine the number of bacterial cells in a broth culture or liquid medium.
More informationThermo Scientific NanoDrop Products. Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules
Thermo Scientific NanoDrop Products Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules Thermo Scientific NanoDrop instruments have pre-programmed methods
More informationBeckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab
Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Room 213 Beckman Coulter DTX 880 Multimode Detector Information The Beckman Coulter DTX 880 Multimode Detector is
More informationMethod 445.0. In Vitro Determination of Chlorophyll a and Pheophytin a in Marine and Freshwater Algae by Fluorescence
Method 445.0 In Vitro Determination of Chlorophyll a and Pheophytin a in Marine and Freshwater Algae by Fluorescence Elizabeth J. Arar and Gary B. Collins Revision 1.2 September 1997 National Exposure
More informationUVC LEDs for. Environmental Monitoring
UVC LEDs for Environmental Monitoring Environmental monitoring relies extensively on molecular spectroscopy for tracking air quality, water and wastewater quality, and detecting hazardous substances. Advances
More informationApplication Note: Absorbance
Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light
More informationUsing the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
More informationExperiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
More informationReal-Time Monitoring Buoy in Mystic River Watershed
Real-Time Monitoring Buoy in Mystic River Watershed Access to the data has been through a password protected site http://www.ysieconet.com/ Project Assistance: Liz McCarthy Laura Senefeld Matt Destino
More informationUV-Vis spectrophotometers
UV-Vis spectrophotometers an Advanced Digital Technology EXPERIENCE SECOMAM is proud to provide the new fourthgeneration UVIKON ultraviolet-visible spectrophotometers.with more than five decades of experience
More informationUSEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls
Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required
More informationDetermining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
More informationAbout Coffee and Refractometers 2008-2010 Voice Systems Technology, Inc. (VST)
About Coffee and Refractometers 2008-200 Voice Systems Technology, Inc. (VST) www.mojotogo.us.0 Coffee and Refractive Index Refractive Index measurements have been used for process control in the food
More informationFerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows
Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for
More informationA Beer s Law Experiment
A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining
More informationSpectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
More informationThe Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
More informationChapter 28: High-Performance Liquid Chromatography (HPLC)
Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion
More informationAgilent Cary 60 UV-Vis
Agilent Cary 60 UV-Vis Efficient. Accurate. Flexible. Specifications Introduction The Agilent Cary 60 UV-Vis spectrophotometer is efficient, accurate and flexible, and is designed to meet both current
More informationA Complete Solution for Method Linearity in HPLC and UHPLC
Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,
More informationExperiment 5. Lasers and laser mode structure
Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,
More informationPesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
More informationLZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506
Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking
More informationLesson 5: Water Conductivity and Total Dissolved Solids Water Quality Sampling
Lesson 5: Water Conductivity and Total Dissolved Solids Water Quality Sampling Time Frame: Two 45-50 minute class periods Grade Level: 8 th 12 th Grade Overview: There are a wide variety of inorganic substances
More informationAustin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law
Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine
More informationTechnical Note. Roche Applied Science. No. LC 19/2004. Color Compensation
Roche Applied Science Technical Note No. LC 19/2004 Purpose of this Note Color The LightCycler System is able to simultaneously detect and analyze more than one color in each capillary. Due to overlap
More informationThe photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
More informationFiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy
Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Introduction The availability and applications of fiber optic sampling in UV/Vis
More informationOperating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso
Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso SevenEasy Conductivity Italiano Español Français Deutsch English Contents Contents Page 1. Introduction...
More informationGraphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
More informationAnalytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
More informationCadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
More informationQuaternary Ammonium Compounds DOC316.53.01128
Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower
More informationEXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare
More informationOnline Infrared Flue Gas Analyzer Gasboard 3000
Online Infrared Flue Gas Analyzer Gasboard 3000 O 2 CO CO2 SO 2 NO NO x Use of measurement methods Proprietary Infrared NDIR Detectors - Micro-flow in ppm range: CO, NO, SO2, CO2 - Dual beam in % volume
More informationChlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information
Chlorine, Total DOC316.53.01028 DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials Scope and application: For testing higher levels of total (free plus combined) chlorine in drinking water,
More informationDrop Counter Sensor Product Number: ENDRP-AD100
imagine explore learn Drop Counter Sensor Product Number: ENDRP-AD100 Overview The Drop Counter sensor is an optical sensor that accurately records the number of drops of titrant added during a titration.
More informationFeatures Rich Detectors for Every Laboratory and Application
SofTA ELSD Features Rich Detectors for Every Laboratory and Application Model 2300 New modern design. Similar performance to Model 1300 Model 1300 Designed for UPLC with narrow peak widths and high sensitivity.
More informationQuantification of Reactive Phosphorus in Lake Mendota Sediments
Quantification of Reactive Phosphorus in Lake Mendota Sediments By: Jeremy, Advisor Dr. Mike Penn Most of the phosphorus (P) that enters lakes from external inputs is attached to soil or sediment. This
More informationConversion Between Analog and Digital Signals
ELET 3156 DL - Laboratory #6 Conversion Between Analog and Digital Signals There is no pre-lab work required for this experiment. However, be sure to read through the assignment completely prior to starting
More informationATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
More informationAppendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
More informationAN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz
AN1200.04 Application Note: FCC Regulations for ISM Band Devices: Copyright Semtech 2006 1 of 15 www.semtech.com 1 Table of Contents 1 Table of Contents...2 1.1 Index of Figures...2 1.2 Index of Tables...2
More informationData Communications Competence Center
Importance of Cable Balance For Improving Noise Susceptibility Data Communications Competence Center DCCC03101702 July 11, 2007 Summary In a study of category 5e and category 6 UTP cables, a strong correlation
More informationLaboratory 5: Properties of Enzymes
Laboratory 5: Properties of Enzymes Technical Objectives 1. Accurately measure and transfer solutions with pipettes 2. Use a Spectrophotometer to study enzyme action. 3. Properly graph a set of data. Knowledge
More informationStrategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide
More informationApplication Requirement
Hazardous Gas Monitors Sensor Selection Overview Safety Monitoring (LEL) Toxic Limit Detection (PEL) Leak Detection Personal Safety Application Requirement Exposure Assessment (TWA) Ambient Air Quality
More informationThe Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium constant, K
More informationREAL-TIME DUST AND AEROSOL MONITORING
REAL-TIME DUST AND AEROSOL MONITORING THE DUSTTRAK II AND DRX AEROSOL MONITORS UNDERSTANDING, ACCELERATED REAL-TIME DUST MONITORING. ANY ENVIRONMENT. ANY APPLICATION. The DustTrak II and DRX Aerosol Monitors
More informationStep-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
More informationthe importance of ph in water purification the importance of ph in water purification Water Purification Systems UK
the importance of ph in water purification Water Purification Systems UK What is ph? The ph scale has been in use for over a century, ever since it was first defined in 1909 by the Danish chemist, Søren
More informationUser Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
More informationLEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS
LEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS March 2010 Lattice Semiconductor 5555 Northeast Moore Ct. Hillsboro, Oregon 97124 USA Telephone: (503) 268-8000 www.latticesemi.com
More informationSection 3. Sensor to ADC Design Example
Section 3 Sensor to ADC Design Example 3-1 This section describes the design of a sensor to ADC system. The sensor measures temperature, and the measurement is interfaced into an ADC selected by the systems
More informationKatharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service
Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?
More informationBeer's Law: Colorimetry of Copper(II) Solutions
Exercise 11 Page 1 Illinois Central College CHEMISTRY 130 Name: Beer's Law: Colorimetry of Copper(II) Solutions Objectives In this experiment, we will use Beer's Law to determine the unknown concentrations
More informationShimadzu UV-VIS User s Guide
Shimadzu UV-VIS User s Guide 1) Push the F4 button on the UV-VIS instrument keypad. This will enable PC control. Push the F4 Button 2) Log into the UV-VIS software with your username and password. 3) After
More informationZeiss 780 Training Notes
Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn
More informationICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
More informationThe Measurement of Sensitivity in Fluorescence Spectroscopy
The Measurement of Sensitivity in Fluorescence Spectroscopy Among instrumental techniques, fluorescence spectroscopy is recognized as one of the more sensitive. In fluorescence, the intensity of the emission
More informationDirect ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
More informationDuct Humidity Transmitter
SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers
More informationDDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS
DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display
More informationReversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
More informationPART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
More informationInterference Identification Guide. Table of Contents
Interference Identification Guide This document is a guide to help IT professionals optimize the performance of wireless networks by using spectrum analysis tools to identify sources of wireless interference.
More informationFIBRE-OPTICS POWER METER INSTRUCTION MANUAL
FIBRE-OPTICS POWER METER INSTRUCTION MANUAL FIBRE-OPTICS POWER METER ELLMAX ELECTRONICS LTD. Unit 29, Leyton Business Centre, Etloe Road, Leyton, London E10 7BT, England Tel: 020-8539 0136 Fax: 020-8539
More informationOptical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry
Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry Models UVS-1, MoniSpec-UV & MoniSpec-UVd Model UVS-1 general information
More information2 Spectrophotometry and the Analysis of Riboflavin
2 Spectrophotometry and the Analysis of Riboflavin Objectives: A) To become familiar with operating the Platereader; B) to learn how to use the Platereader in determining the absorption spectrum of a compound
More informationGas Custody Transfer Calibration
Gas Custody Transfer Calibration Using multi variable temperature / pressure calibrators for flowmeter calibration 2013 Introduction Gas custody transfer flow computers require special calibration to perform
More informationMETHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine
More informationHITACHI INVERTER SJ/L100/300 SERIES PID CONTROL USERS GUIDE
HITACHI INVERTER SJ/L1/3 SERIES PID CONTROL USERS GUIDE After reading this manual, keep it for future reference Hitachi America, Ltd. HAL1PID CONTENTS 1. OVERVIEW 3 2. PID CONTROL ON SJ1/L1 INVERTERS 3
More informationDigital image processing
746A27 Remote Sensing and GIS Lecture 4 Digital image processing Chandan Roy Guest Lecturer Department of Computer and Information Science Linköping University Digital Image Processing Most of the common
More informationEXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS
1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides
More informationAt the bottom of the screen you will also see some of our sister brands, all of which can also plug in to the Storm data logger.
This presentation will discuss how the challenges of water quality monitoring and realtime data collection can be addressed with new technology. We ll focus on the EXO sonde platform from YSI and the Storm
More informationIntroduction to flow cytometry
Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression
More informationNJWMC Continuous Monitoring Inventory & DEP s Continuous Monitoring
NJWMC Continuous Monitoring Inventory & DEP s Continuous Monitoring Leslie McGeorge NJ Department of Environmental Protection Water Monitoring and Standards Presented at: NJ Water Monitoring Council Meeting
More informationEvaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories
Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,
More informationLab #11: Determination of a Chemical Equilibrium Constant
Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a
More informationHand-held thermometer Model CTH7000
Calibration technology Hand-held thermometer Model CTH7000 WIKA data sheet CT 55.50 Applications Precision thermometer for very accurate temperature measurements in a range of -200... +962 C Reference
More informationSeries DFMT Digital Paddlewheel Flowmeter
Series DFMT Digital Paddlewheel Flowmeter Bulletin F-DFMT Specifications - Installation Operating Instructions -/8 [00] -/ [8] -/8 [00] A The Series DFMT Digital Paddlewheel Flow Transmitter provides instantaneous
More informationph Alkalinity of Water
ph Alkalinity of Water DOC316.52.93085 Based on ISO standard 9963-1:1994 ph-metric Titration 0.4 to 20 mmol/l of Total Alkalinity 1. Introduction Alkalinity of water is its acid-neutralizing capacity.
More informationThe Sievers 900 Series TOC Analyzers
GE Water & Process Technologies Analytical Instruments The Sievers 900 Series TOC Analyzers imagination at work Delivering Next-Generation Ease of Use, Productivity, and Reliability The Sievers* 900 Series
More informationAdvanced Solutions for Gas Monitoring
Continuous Emissions Monitoring Stack Testing Process Monitoring Quality Control Engine Exhaust Gas Monitoring Workplace Air Quality Monitoring Combustion Research Emergency Rescue Services Advanced Solutions
More informationVentostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Measurement & Control. Features:
GE Measurement & Control Ventostat Wall Mount Telaire Wall Mount, Humidity and Temperature Transmitters Features: Patented, Absorption Infrared Gas sensing engine provides high accuracy in a compact low
More informationNeutralizing an Acid and a Base
Balancing Act Teacher Information Objectives In this activity, students neutralize a base with an acid. Students determine the point of neutralization of an acid mixed with a base while they: Recognize
More informationCommon Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC
Common Problems with Online Water Quality Analyzers Greg Macy Director AquatiPro LLC Objectives Diagnosing Colorimetric Analyzers Identifying Common Problems with Sodium Analyzers Troubleshooting ph Analyzers
More informationMeasuring Laser Power and Energy Output
Measuring Laser Power and Energy Output Introduction The most fundamental method of checking the performance of a laser is to measure its power or energy output. Laser output directly affects a laser s
More informationCharacterizing Digital Cameras with the Photon Transfer Curve
Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently
More information