FLUOROMETER PERFORMANCE EVALUATION REPORT 9/2002

Size: px
Start display at page:

Download "FLUOROMETER PERFORMANCE EVALUATION REPORT 9/2002"

Transcription

1 FLUOROMETER PERFORMANCE EVALUATION REPORT 9/22 ABSTRACT A series of performance evaluation tests were conducted at three tests sites to investigate the overall performance of the Turner Designs Submersible Fluorometer and the chlorophyll probe. Performance was determined through results designed to determine minimum detection limit of chlorophyll a, dynamic range, accuracy of readings to actual chlorophyll a, ambient light rejection, and response time. The results indicate that the fluorometer outperforms the sensor but the performance of both sensors is adequate for most environmental conditions. A performance limitation of the is the response time. Using a moving average calculation, it responds relatively slowly to changing concentrations and would not be effective for vertical profiling applications. In some extreme conditions, such as chlorophyll concentrations less than 1µg/L and high ambient light and/or high concentrations of interfering compounds, the may also not provide accurate measures of algal biomass. INTRODUCTION A series of performance evaluation tests were conducted to investigate the overall performance of the Turner Designs Submersible Fluorometer and the chlorophyll probe. Tests were conducted at three sites; Moss Landing Marine Laboratory, Turner Designs Testing Laboratory and the USGS Menlo Park facility. Performance was evaluated through designing tests to investigate the following areas; minimum detection limit, instrument range, response time, light rejection and accuracy of fluorometer reading to actual chlorophyll a concentration. Due to different specification standards, comparing instruments on published specifications is not always straightforward. Below is the published performance data for each instrument. Through our testing we have attempted to compare the instruments in a meaningful and informative way for the customer. We have compared instruments identically and have attempted to present the data in as an objective manner as possible. In most cases the raw fluorescence signal was recorded in order to eliminate scaling issues related to calibration. The raw fluorescence was recorded as %FS which indicates the % of a sample s signal in relation to 1% full scale of the particular instrument. Calibrated data is referred to as in vivo fluorescence or. Published Performance Specifications Range -4µg/L 4 orders of magnitude Minimum Detection Limit N/A.2 µg/l Resolution.1 µg/l,.1%fs 12 bit Max Sampling Rate NA 5Hz Depth -61m -6m Instrument Overview: While both instruments are chlorophyll fluorometers there are several key distinctions between the two units. Firstly, the physical size and appearance is significantly different. The fluorometer is designed to operate as a stand-alone unit and can also be interfaced to multi-parameter systems. Being a stand-alone unit, it incorporates many additional features that are not required in the probe. The following is a list of unique features; 6m depth capability, temperature compensation, internal data logging, autoranging, turbidity option, digital and analog output, configurable analog (-5V) limits, and calibration software. The presence of these features has resulted in a much different looking sensor. However, when interfaced to a multi-parameter system, the operates in a very similar manner to the probe. The has been designed to interface directly onto multi-parameter sondes. The probe is also now available on an optical sonde that only offers fluorescence, conductivity, 1 FLUOROMETER COMPARISON REPORT

2 temperature and depth. The probe provides an analog signal to the sonde which integrates the data from all of the sonde sensors into one digital data stream. The provides both digital and analog data continuously. MATERIALS AND METHODS All tests were conducted under controlled settings in a laboratory. In most cases, monocultures on algae were used as the test media (see table below for algal culture details). In some cases natural algal samples were used. Chlorophyll a extractions were conducted on all samples using the Non-acidification fluorometric method (Welschmeyer, 1997) and the Acidification technique (E.P.A. Method 445.) in the case of the pheophytin testing. Algal samples used: Genus Algal group 1 Accessory Pigment Dunaliella sp. chlorophyte chlorophyll b Thallassiosira sp. diatom chlorophyll c Phaeodactylum sp. diatom chlorophyll c San Francisco Bay Delta natural assemblage RESULTS MINIMUM DETECTION LIMIT The Minimum Detection Limit is the lowest concentration of algae in clean culture media than can is distinguishable from a blank culture media solution. Three detection limit tests were run by mixing a dilution series of algal cultures. The readings were allowed to stabilize and at least 1 datapoints were recorded per sample. Dunaliella Phaeodactylum Chl A Chl A Thallassiosira Chl A Table 1: Dunaliella and Phaeodactylum test data from Moss Landing Marine Laboratory (MLML). Data from Thallassiosira test from Turner Designs. The Chloro column is extracted chlorophyll a data. The and column is the in vivo fluorescence data. From the data shown it is clear the difference in detection limit between algal groups. The data supports the fact that the was unable to distinguish some algal cultures at concentrations of 1µg/L chlorophyll a or less while the was capable of distinguishing all samples. The lowest concentration tested was.34µg/l of Thallassiosira culture. Because each species of algae has varying in vivo fluorescence ():extracted chlorophyll, the limits of detection can vary between species. The demonstrated the ability to detect concentration as low as.79µg/l of Thallassiosira culture while not detecting other cultures that were above 1µg/L. 2 FLUOROMETER COMPARISON REPORT

3 Minimum Detection Limit Test #1 Minimum Detection Limit Test # R 2 =.9737 R 2 = chlorophyll a (ug/l) Graph 1 & 2: MDL test data from Moss Landing Marine Lab showing that the not differentiating between 2.5µg/L and 1.1µg/L Dunaliella samples. MDL Test#2 shows data from tests with Thallassiosira cultures indicate that the did not distinguish the blank and the first two culture samples. RANGE Both instruments provide an analog -5V output signal while the has a digital output as well. The -5V ranges are factory set for a given fluorescence range. The comes standard with a -1µg/L analog range while the comes standard with a -4µg/L range. As the analog range increases, the resolution of the data decreases and it is therefore advantageous to set the -5V limits to the environments in which you work. The digital outputs from the Sonde and the were used for all tests. Both instruments demonstrated good linearity for typical chlorophyll a ranges found in natural environments. The only difference of note for chlorophyll measurement is the ability to customize the -5V range on the. This abiltiy allows customers working in low chlorophyll environments to decrease the -5V range and therefore improve the resolution of the analog data R 2 =.9319 R 2 = chlorophyll a (ug/l) Linearity Test R 2 =.997 R 2 = chlorophyll a (ug/l) Graph 3: Data taken with calibrated values at MLML using Dunaliella cultures. 3 FLUOROMETER COMPARISON REPORT

4 ACCURACY In order to test the accuracy of measuring chlorophyll a fluorescence, the effects of various interfering compounds were investigated. In an ideal situation, an in situ fluorometer would measure chlorophyll a fluorescence only since chlorophyll a is the only photosynthetic pigment that all algae and photosynthetic bacteria have in common. By measuring chlorophyll a only you will obtain the best estimate of total, living algal biomass. Compounds that can interfere with fluorescent signals optimized to detect chlorophyll a include accessory pigments such as chlorophyll b and c, degraded chlorophyll a, pheophytin a, and dissolved organic matter. All of these interfering substances have fluorescence spectra that can overlap with chlorophyll a given the appropriate concentrations and fluorometer optics. When discussing an instrument s susceptibility to interfering compounds, the performance is determined by the optics. The optics in a fluorometer encompass three components; a light source, optical filters and a light detector. Both the and the instrument use solid-state components, LED light source and photodiode light detector, which makes them equivalent in this category except that the utilizes two LED lamps while the uses one. The difference between the optics lies in the optical filters, specifically the emission filters. The emission filter is placed in front of the light detector. It s purpose is to block all light except for the wavelengths of interest, in this case chlorophyll a fluorescence which peaks at 685nm. The more specific the filter is to the chlorophyll peak, the more accurate the fluorometer will be at measuring total algal biomass. The uses long-pass emission filters which allows in all light >6nm. This means that any material other than chlorophyll a that is excited by the blue LED light source (peak at 46nm) and fluorescing above 6nm will be detected. Likely candidates for this include pheophytin a, chlorophyll b, chlorophyll c and CDOM. The, on the other hand, utilizes a band-pass filter that allows in light from 66-7nm only, peaking at 685nm. Therefore, the fluorescence from compounds other than chlorophyll a are much less likely to be detected. The benefit of using a long-pass filter is the low material cost and higher transmission. The downside is greater susceptibility to interfering compounds. The tests were effective in demonstrating the effect of optical components in the pheophytin testing. One of the most likely interference s is pheophytin a. Pheophytin can accumulate in water column at density gradients, close to the bottom, and be increased due to active zooplankton feeding. A test was run to compare the fluorescence signal from a healthy algal culture and fluorescence from the same culture after exposed to a weak acid that converted much of the chlorophyll in the cells to pheophytin a. 1mL of 1N HCl was added to 35mL of Thallasisosira culture. The culture was stirred for 1 minute and then the fluorescence was read using both instruments. The results were as follows: Before After Before After FS% Ratio FLUOROMETER COMPARISON REPORT

5 The Difference in the ratios indicates a difference in the fluorometers ability to detect pheophytin fluorescence. The ratio of 2.76 indicates that it is less sensitive to pheophytin fluorescence than the sensor that had a ratio of Regression Comprison R TD USGS-Thal TD-Thal TD-Thal2 MLML-Pheo MLML-Pheo MLML-Dun MLML-Dun MLML-Dun TD Fresh USGS-Delta Graph 4: The bar graph displays the correlations (r 2 ) obtained from the minimum detection and linearity tests conducted at all three testing sites. In most cases the data correlated with actual chlorophyll a more closely. All tests returned an r 2 of greater than.85 or better. The samples market TD Fresh and USGS-Delta are natural water samples with an assemblage of algal species. All other samples were moncultures of algae. AMBIENT LIGHT REJECTION Ambient light rejection refers to the ability of the fluorometer to be unaffected by sunlight or other ambient light sources. Light rejection is typically performed by flashing the LED light source. When the light source is on, the instrument detects the fluorescence and any ambient light, when the light source is off, the instrument detects the ambient light. The instrument compares the two to determine the fluorescence. Ambient light rejection test results indicate that there is a difference in how the two instruments correct ambient light. The displayed no significant difference in fluorescence signal while the showed a decrease in signal as ambient light increased and became more variable Ambient Light Rejection Test Scufa no light light at 9 degrees light at 18 degrees Position of Light Source % Full Scale Ambient Light Rejection Test #2 Lit Lab Dark Lab Light 12" Direct Light 12" Variable Graph 5 & 6: Graph 5 displays data from Moss Landing Marine Lab. The instruments were placed in a solution of rhodamine WT dye. A halogen light source was used to test ambient light. Graph 6 displays data from tests run at Turner Designs with a halogen light source. The units were not placed in a solution. 5 FLUOROMETER COMPARISON REPORT

6 RESPONSE TIME The graph below shows the response time of the two instruments as they were alternately placed into samples of natural water containing algae and blank deionized water. The readings were taken at one-second intervals. The difference in the graphs displays a significant difference between the two instruments. The system uses a moving average as opposed to the realtime sampling of the. The moving average allows the system to smooth the data as well as enables it to improve the limit of detection. The, designed for high-speed vertical profiling applications, samples at a rate of 1Hz and outputs data at 1Hz or 5Hz. The spikes at the beginning of the readings of the algal cultures is due to removing the instrument from the blank water and then placing it into an algal culture. The change from air to liquid creates some bubbles that cause spikes in the readings. The sensor is less sensitive to these changes due to the slow moving average (~1sec) of the data which smoothes the data but does not supply instantaneous data. RESPONSE TIME TEST IN VIVO FLUORESCENCE :55:55 13:56:38 13:57:22 13:58:5 13:58:48 13:59:31 14::14 14::58 14:1:41 TIME Graph 7 CONCLUSIONS The results of the performance tests indicate that both instruments are effective at estimating chlorophyll a concentrations in typical environmental conditions. Although accuracy tests favored the fluorometer, most correlations between the in vivo fluorescence and the actual chlorophyll concentration exceeded.9 for both instruments. It is in the presence of interfering compounds where the data from the two units may differ significantly. Results from testing with degraded chlorophyll, pheophytin, support this conclusion, with the not differentiating between chlorophyll and pheophytin as well as the. Other extreme environmental conditions could limit the effectiveness of the probe. Conditions such as high ambient light,from shallow waters with reflective sediment, and low productivity waters could be problematic. Also, due to the slow response time of the probe, vertical profiling would be a challenge. On the other hand, the probe seems well suited for deployed use in coastal or freshwater environments where chlorophyll levels exceed 1µg/L and ambient light and interfering compounds are not a significant issue. The performed well in all categories. It is a more versatile instrument because of ability to be used in self-contained mode. 6 FLUOROMETER COMPARISON REPORT

7 REFERENCES: Arar, E.J. and G.B. Collins In vitro determination of chlorophyll a and pheophytin a in marine and freshwater algae by fluorescence. E.P.A. Method 445., Revision 1.2 Welschmeyer, N.A Fluorometeric analysis of chlorophyll a in the presence of chlorophyll b and pheopigments. Limnol. Oceanogr. 39: TESTING FACILITIES: MOSS LANDING MARINE LABORATORY, MOSS LANDING, CA. Tests run under the supervision of Dr. Nick Welschmeyer. TURNER DESIGNS TESTING FACILITY, SUNNYVALE, CA. UNITED STATES GEOLOGICIAL SURVEY, MENLO PARK, CA Report written by Robert Ellison, Turner Designs, Inc., 22 7 FLUOROMETER COMPARISON REPORT

An Introduction to Algae Measurements Using In Vivo Fluorescence

An Introduction to Algae Measurements Using In Vivo Fluorescence An Introduction to Algae Measurements Using In Vivo Fluorescence Submersible fluorescence sensors enable real-time field estimates of phytoplankton that can be directly correlated to standard laboratory

More information

How to measure Chlorophyll-a

How to measure Chlorophyll-a World Bank & Government of The Netherlands funded Training module # WQ - 40 How to measure Chlorophyll-a New Delhi, March 2000 CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas, New Delhi 11 00 16

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

ALMOFRONT 2 cruise in Alboran sea : Chlorophyll fluorescence calibration

ALMOFRONT 2 cruise in Alboran sea : Chlorophyll fluorescence calibration Vol. 3 : 6-11, 2010 Journal of Oceanography, Research and Data ALMOFRONT 2 cruise in Alboran sea : Chlorophyll fluorescence calibration CUTTELOD Annabelle 1,2 and CLAUSTRE Hervé 1,2 1 UPMC, Univ. Paris

More information

Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy

Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy Determining the Free Chlorine Content of Swimming Pool Water Computer 33 Physicians in the nineteenth century used chlorine water as a disinfectant. Upon the discovery that certain diseases were transmitted

More information

DSM155 Digital LED Light Meter User s Manual DSM155

DSM155 Digital LED Light Meter User s Manual DSM155 DSM155 Digital LED Light Meter User s Manual DSM155 CONTENTS TABLE OF CONTENTS......3 LIMITED WARRANTY AND LIMITATION OF LIABILITY......3 Out of Box....3 Accessories....3 Safety Information....3 Certification....5

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from

More information

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters

More information

EPA New England Regional Laboratory. EPA Project Team: Liz McCarthy. Matt Destino Tim Bridges Dick Siscanaw

EPA New England Regional Laboratory. EPA Project Team: Liz McCarthy. Matt Destino Tim Bridges Dick Siscanaw EPA s River Real Time Monitoring Buoy in the Charles and Mystic Watersheds Access to the data has been through a password protected site http://www.ysieconet.com/ Tom Faber EPA New England Regional Laboratory

More information

Nederland België / Belgique

Nederland België / Belgique Cronus Brochure 2 Introduction Cronus: The first spectrometer and colorimeter in one The Cronus is world s first spectrocolorimeter combining a high en VIS spectrometer with a high speed XYZ colorimeter.

More information

BACTERIAL ENUMERATION

BACTERIAL ENUMERATION BACTERIAL ENUMERATION In the study of microbiology, there are numerous occasions when it is necessary to either estimate or determine the number of bacterial cells in a broth culture or liquid medium.

More information

Thermo Scientific NanoDrop Products. Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules

Thermo Scientific NanoDrop Products. Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules Thermo Scientific NanoDrop Products Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules Thermo Scientific NanoDrop instruments have pre-programmed methods

More information

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Room 213 Beckman Coulter DTX 880 Multimode Detector Information The Beckman Coulter DTX 880 Multimode Detector is

More information

Method 445.0. In Vitro Determination of Chlorophyll a and Pheophytin a in Marine and Freshwater Algae by Fluorescence

Method 445.0. In Vitro Determination of Chlorophyll a and Pheophytin a in Marine and Freshwater Algae by Fluorescence Method 445.0 In Vitro Determination of Chlorophyll a and Pheophytin a in Marine and Freshwater Algae by Fluorescence Elizabeth J. Arar and Gary B. Collins Revision 1.2 September 1997 National Exposure

More information

UVC LEDs for. Environmental Monitoring

UVC LEDs for. Environmental Monitoring UVC LEDs for Environmental Monitoring Environmental monitoring relies extensively on molecular spectroscopy for tracking air quality, water and wastewater quality, and detecting hazardous substances. Advances

More information

Application Note: Absorbance

Application Note: Absorbance Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light

More information

Using the Spectrophotometer

Using the Spectrophotometer Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to

More information

Experiment #5: Qualitative Absorption Spectroscopy

Experiment #5: Qualitative Absorption Spectroscopy Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions

More information

Real-Time Monitoring Buoy in Mystic River Watershed

Real-Time Monitoring Buoy in Mystic River Watershed Real-Time Monitoring Buoy in Mystic River Watershed Access to the data has been through a password protected site http://www.ysieconet.com/ Project Assistance: Liz McCarthy Laura Senefeld Matt Destino

More information

UV-Vis spectrophotometers

UV-Vis spectrophotometers UV-Vis spectrophotometers an Advanced Digital Technology EXPERIENCE SECOMAM is proud to provide the new fourthgeneration UVIKON ultraviolet-visible spectrophotometers.with more than five decades of experience

More information

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required

More information

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,

More information

About Coffee and Refractometers 2008-2010 Voice Systems Technology, Inc. (VST)

About Coffee and Refractometers 2008-2010 Voice Systems Technology, Inc. (VST) About Coffee and Refractometers 2008-200 Voice Systems Technology, Inc. (VST) www.mojotogo.us.0 Coffee and Refractive Index Refractive Index measurements have been used for process control in the food

More information

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for

More information

A Beer s Law Experiment

A Beer s Law Experiment A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium

More information

Chapter 28: High-Performance Liquid Chromatography (HPLC)

Chapter 28: High-Performance Liquid Chromatography (HPLC) Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion

More information

Agilent Cary 60 UV-Vis

Agilent Cary 60 UV-Vis Agilent Cary 60 UV-Vis Efficient. Accurate. Flexible. Specifications Introduction The Agilent Cary 60 UV-Vis spectrophotometer is efficient, accurate and flexible, and is designed to meet both current

More information

A Complete Solution for Method Linearity in HPLC and UHPLC

A Complete Solution for Method Linearity in HPLC and UHPLC Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506 Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking

More information

Lesson 5: Water Conductivity and Total Dissolved Solids Water Quality Sampling

Lesson 5: Water Conductivity and Total Dissolved Solids Water Quality Sampling Lesson 5: Water Conductivity and Total Dissolved Solids Water Quality Sampling Time Frame: Two 45-50 minute class periods Grade Level: 8 th 12 th Grade Overview: There are a wide variety of inorganic substances

More information

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine

More information

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation Roche Applied Science Technical Note No. LC 19/2004 Purpose of this Note Color The LightCycler System is able to simultaneously detect and analyze more than one color in each capillary. Due to overlap

More information

The photoionization detector (PID) utilizes ultraviolet

The photoionization detector (PID) utilizes ultraviolet Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).

More information

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Introduction The availability and applications of fiber optic sampling in UV/Vis

More information

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso SevenEasy Conductivity Italiano Español Français Deutsch English Contents Contents Page 1. Introduction...

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Quaternary Ammonium Compounds DOC316.53.01128

Quaternary Ammonium Compounds DOC316.53.01128 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower

More information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare

More information

Online Infrared Flue Gas Analyzer Gasboard 3000

Online Infrared Flue Gas Analyzer Gasboard 3000 Online Infrared Flue Gas Analyzer Gasboard 3000 O 2 CO CO2 SO 2 NO NO x Use of measurement methods Proprietary Infrared NDIR Detectors - Micro-flow in ppm range: CO, NO, SO2, CO2 - Dual beam in % volume

More information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information Chlorine, Total DOC316.53.01028 DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials Scope and application: For testing higher levels of total (free plus combined) chlorine in drinking water,

More information

Drop Counter Sensor Product Number: ENDRP-AD100

Drop Counter Sensor Product Number: ENDRP-AD100 imagine explore learn Drop Counter Sensor Product Number: ENDRP-AD100 Overview The Drop Counter sensor is an optical sensor that accurately records the number of drops of titrant added during a titration.

More information

Features Rich Detectors for Every Laboratory and Application

Features Rich Detectors for Every Laboratory and Application SofTA ELSD Features Rich Detectors for Every Laboratory and Application Model 2300 New modern design. Similar performance to Model 1300 Model 1300 Designed for UPLC with narrow peak widths and high sensitivity.

More information

Quantification of Reactive Phosphorus in Lake Mendota Sediments

Quantification of Reactive Phosphorus in Lake Mendota Sediments Quantification of Reactive Phosphorus in Lake Mendota Sediments By: Jeremy, Advisor Dr. Mike Penn Most of the phosphorus (P) that enters lakes from external inputs is attached to soil or sediment. This

More information

Conversion Between Analog and Digital Signals

Conversion Between Analog and Digital Signals ELET 3156 DL - Laboratory #6 Conversion Between Analog and Digital Signals There is no pre-lab work required for this experiment. However, be sure to read through the assignment completely prior to starting

More information

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS 1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz

AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz AN1200.04 Application Note: FCC Regulations for ISM Band Devices: Copyright Semtech 2006 1 of 15 www.semtech.com 1 Table of Contents 1 Table of Contents...2 1.1 Index of Figures...2 1.2 Index of Tables...2

More information

Data Communications Competence Center

Data Communications Competence Center Importance of Cable Balance For Improving Noise Susceptibility Data Communications Competence Center DCCC03101702 July 11, 2007 Summary In a study of category 5e and category 6 UTP cables, a strong correlation

More information

Laboratory 5: Properties of Enzymes

Laboratory 5: Properties of Enzymes Laboratory 5: Properties of Enzymes Technical Objectives 1. Accurately measure and transfer solutions with pipettes 2. Use a Spectrophotometer to study enzyme action. 3. Properly graph a set of data. Knowledge

More information

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview. Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide

More information

Application Requirement

Application Requirement Hazardous Gas Monitors Sensor Selection Overview Safety Monitoring (LEL) Toxic Limit Detection (PEL) Leak Detection Personal Safety Application Requirement Exposure Assessment (TWA) Ambient Air Quality

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium constant, K

More information

REAL-TIME DUST AND AEROSOL MONITORING

REAL-TIME DUST AND AEROSOL MONITORING REAL-TIME DUST AND AEROSOL MONITORING THE DUSTTRAK II AND DRX AEROSOL MONITORS UNDERSTANDING, ACCELERATED REAL-TIME DUST MONITORING. ANY ENVIRONMENT. ANY APPLICATION. The DustTrak II and DRX Aerosol Monitors

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

the importance of ph in water purification the importance of ph in water purification Water Purification Systems UK

the importance of ph in water purification the importance of ph in water purification Water Purification Systems UK the importance of ph in water purification Water Purification Systems UK What is ph? The ph scale has been in use for over a century, ever since it was first defined in 1909 by the Danish chemist, Søren

More information

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510 User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

LEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS

LEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS LEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS March 2010 Lattice Semiconductor 5555 Northeast Moore Ct. Hillsboro, Oregon 97124 USA Telephone: (503) 268-8000 www.latticesemi.com

More information

Section 3. Sensor to ADC Design Example

Section 3. Sensor to ADC Design Example Section 3 Sensor to ADC Design Example 3-1 This section describes the design of a sensor to ADC system. The sensor measures temperature, and the measurement is interfaced into an ADC selected by the systems

More information

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?

More information

Beer's Law: Colorimetry of Copper(II) Solutions

Beer's Law: Colorimetry of Copper(II) Solutions Exercise 11 Page 1 Illinois Central College CHEMISTRY 130 Name: Beer's Law: Colorimetry of Copper(II) Solutions Objectives In this experiment, we will use Beer's Law to determine the unknown concentrations

More information

Shimadzu UV-VIS User s Guide

Shimadzu UV-VIS User s Guide Shimadzu UV-VIS User s Guide 1) Push the F4 button on the UV-VIS instrument keypad. This will enable PC control. Push the F4 Button 2) Log into the UV-VIS software with your username and password. 3) After

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

The Measurement of Sensitivity in Fluorescence Spectroscopy

The Measurement of Sensitivity in Fluorescence Spectroscopy The Measurement of Sensitivity in Fluorescence Spectroscopy Among instrumental techniques, fluorescence spectroscopy is recognized as one of the more sensitive. In fluorescence, the intensity of the emission

More information

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure

More information

Duct Humidity Transmitter

Duct Humidity Transmitter SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers

More information

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display

More information

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination

More information

PART 1 - INTRODUCTION...

PART 1 - INTRODUCTION... Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART

More information

Interference Identification Guide. Table of Contents

Interference Identification Guide. Table of Contents Interference Identification Guide This document is a guide to help IT professionals optimize the performance of wireless networks by using spectrum analysis tools to identify sources of wireless interference.

More information

FIBRE-OPTICS POWER METER INSTRUCTION MANUAL

FIBRE-OPTICS POWER METER INSTRUCTION MANUAL FIBRE-OPTICS POWER METER INSTRUCTION MANUAL FIBRE-OPTICS POWER METER ELLMAX ELECTRONICS LTD. Unit 29, Leyton Business Centre, Etloe Road, Leyton, London E10 7BT, England Tel: 020-8539 0136 Fax: 020-8539

More information

Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry

Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry Models UVS-1, MoniSpec-UV & MoniSpec-UVd Model UVS-1 general information

More information

2 Spectrophotometry and the Analysis of Riboflavin

2 Spectrophotometry and the Analysis of Riboflavin 2 Spectrophotometry and the Analysis of Riboflavin Objectives: A) To become familiar with operating the Platereader; B) to learn how to use the Platereader in determining the absorption spectrum of a compound

More information

Gas Custody Transfer Calibration

Gas Custody Transfer Calibration Gas Custody Transfer Calibration Using multi variable temperature / pressure calibrators for flowmeter calibration 2013 Introduction Gas custody transfer flow computers require special calibration to perform

More information

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine

More information

HITACHI INVERTER SJ/L100/300 SERIES PID CONTROL USERS GUIDE

HITACHI INVERTER SJ/L100/300 SERIES PID CONTROL USERS GUIDE HITACHI INVERTER SJ/L1/3 SERIES PID CONTROL USERS GUIDE After reading this manual, keep it for future reference Hitachi America, Ltd. HAL1PID CONTENTS 1. OVERVIEW 3 2. PID CONTROL ON SJ1/L1 INVERTERS 3

More information

Digital image processing

Digital image processing 746A27 Remote Sensing and GIS Lecture 4 Digital image processing Chandan Roy Guest Lecturer Department of Computer and Information Science Linköping University Digital Image Processing Most of the common

More information

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

At the bottom of the screen you will also see some of our sister brands, all of which can also plug in to the Storm data logger.

At the bottom of the screen you will also see some of our sister brands, all of which can also plug in to the Storm data logger. This presentation will discuss how the challenges of water quality monitoring and realtime data collection can be addressed with new technology. We ll focus on the EXO sonde platform from YSI and the Storm

More information

Introduction to flow cytometry

Introduction to flow cytometry Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression

More information

NJWMC Continuous Monitoring Inventory & DEP s Continuous Monitoring

NJWMC Continuous Monitoring Inventory & DEP s Continuous Monitoring NJWMC Continuous Monitoring Inventory & DEP s Continuous Monitoring Leslie McGeorge NJ Department of Environmental Protection Water Monitoring and Standards Presented at: NJ Water Monitoring Council Meeting

More information

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,

More information

Lab #11: Determination of a Chemical Equilibrium Constant

Lab #11: Determination of a Chemical Equilibrium Constant Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a

More information

Hand-held thermometer Model CTH7000

Hand-held thermometer Model CTH7000 Calibration technology Hand-held thermometer Model CTH7000 WIKA data sheet CT 55.50 Applications Precision thermometer for very accurate temperature measurements in a range of -200... +962 C Reference

More information

Series DFMT Digital Paddlewheel Flowmeter

Series DFMT Digital Paddlewheel Flowmeter Series DFMT Digital Paddlewheel Flowmeter Bulletin F-DFMT Specifications - Installation Operating Instructions -/8 [00] -/ [8] -/8 [00] A The Series DFMT Digital Paddlewheel Flow Transmitter provides instantaneous

More information

ph Alkalinity of Water

ph Alkalinity of Water ph Alkalinity of Water DOC316.52.93085 Based on ISO standard 9963-1:1994 ph-metric Titration 0.4 to 20 mmol/l of Total Alkalinity 1. Introduction Alkalinity of water is its acid-neutralizing capacity.

More information

The Sievers 900 Series TOC Analyzers

The Sievers 900 Series TOC Analyzers GE Water & Process Technologies Analytical Instruments The Sievers 900 Series TOC Analyzers imagination at work Delivering Next-Generation Ease of Use, Productivity, and Reliability The Sievers* 900 Series

More information

Advanced Solutions for Gas Monitoring

Advanced Solutions for Gas Monitoring Continuous Emissions Monitoring Stack Testing Process Monitoring Quality Control Engine Exhaust Gas Monitoring Workplace Air Quality Monitoring Combustion Research Emergency Rescue Services Advanced Solutions

More information

Ventostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Measurement & Control. Features:

Ventostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Measurement & Control. Features: GE Measurement & Control Ventostat Wall Mount Telaire Wall Mount, Humidity and Temperature Transmitters Features: Patented, Absorption Infrared Gas sensing engine provides high accuracy in a compact low

More information

Neutralizing an Acid and a Base

Neutralizing an Acid and a Base Balancing Act Teacher Information Objectives In this activity, students neutralize a base with an acid. Students determine the point of neutralization of an acid mixed with a base while they: Recognize

More information

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC Common Problems with Online Water Quality Analyzers Greg Macy Director AquatiPro LLC Objectives Diagnosing Colorimetric Analyzers Identifying Common Problems with Sodium Analyzers Troubleshooting ph Analyzers

More information

Measuring Laser Power and Energy Output

Measuring Laser Power and Energy Output Measuring Laser Power and Energy Output Introduction The most fundamental method of checking the performance of a laser is to measure its power or energy output. Laser output directly affects a laser s

More information

Characterizing Digital Cameras with the Photon Transfer Curve

Characterizing Digital Cameras with the Photon Transfer Curve Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently

More information