Miniaturized Two-Pressure Generator for Relative Humidity

Size: px
Start display at page:

Download "Miniaturized Two-Pressure Generator for Relative Humidity"

Transcription

1 Miniaturized To-Pressure Generator for Relative Humidity by Dr. Helmut Mitter The original publication is available at.springerlin.com DOI.7/s y ABSTRACT The concept and design of a miniaturized to-pressure humidity generator are presented. The generator is suitable for achieving relative humidity ranging from % to 95% ith uncertainties of under % at ambient temperature and can be used for the calibration of relative humidity instruments in the laboratory and on site. By virtue of the concept, the relative humidity achieved is traced to only pressure measurements. Further references are not necessary. A special adjustment algorithm enables standard industrial pressure sensors to be used. The to pressure sensors are synchronized by performing alignment at ambient pressure. The resulting correlation of the uncertainties leads to very small systematic errors in the humidity display. Each humidity generator can still be individually adjusted through comparison ith a chilled mirror hygrometer, the total uncertainty of the generator being largely determined by the uncertainty of this hygrometer. Any drift of the pressure sensors that occurs can be compensated for at any time by performing an alignment at ambient pressure ithout changing the individual adjustment of the generator. It can be demonstrated that the uncertainty of the displayed humidity remains practically unchanged over the course of a year by virtue of this alignment process. KEY WORDS humidity generator, miniaturized, relative humidity, traceable, to pressure generator

2 . INTRODUCTION With the maret for humidity meters for industrial and semi-industrial applications expanding rapidly, it is becoming increasingly useful to have calibration equipment that enables the measurand relative humidity to be traceable to international units. measuring chamber ith ducts for samples rel. humidity indication temperature indication samples indication selector sitch for devices under test rel. humidity adjustment (pressure regulator) Fig.. Miniaturized to-pressure generator HUMOR (E+E Eletroni, Austria). poer supply devices under test 4 VDC Various commercially available humidity generators are in principle suitable for providing this type of proof. Hoever, such systems are generally very complex and expensive and only suitable for stationary operation, rendering them unsuited to in-house quality control departments catering to a broad maret. Saturated, and also unsaturated, salt solutions are also still used for relative humidity calibration. In this case, hoever, it is very difficult, if not impossible, to demonstrate complete traceability to international units. Various miniaturized humidity generators are no also available on the maret. In most cases, these or on the principle of a mixed-gas generator ith an internal reference, ith either a dedicated chilled mirror hygrometer or a capacitive polymer sensor serving as the internal reference. In the first case, high system costs must be reconed ith. The second case provides a reference that to all intents and purposes exhibits the same stability and measuring uncertainty as most of the devices to be tested. A transportable, user-friendly humidity generator (Fig. ) employing the same oring principle as a to-pressure generator ill be presented. Basically, this device displays humidity using a fundamental technique analogous to, for instance, the reference humidity generator of NIST []. The traceability of the device is ensured by periodic calibration against a traceable chilled-mirror hygrometer. Beteen the periodic calibrations, the device operates ith high stability due to its fundamental basis. The generator ors at ambient temperature and therefore displays the relative humidity at a slightly elevated ambient temperature. The only operating resources required are a poer supply and compressed, oil-free air at. MPa. The aim of the paper is not to give a full uncertainty calculation of the generator but to demonstrate ho to minimize the influence of pressure sensor tolerance and drift on the performance of the generator by means of different adjustment procedures. v. / Modification rights reserved

3 . GENERATOR CONCEPT AND SETUP Compressed oil-free air is completely saturated ith ater vapour in a saturation chamber at a certain pressure depending on the requested relative humidity and fed to a pressure reduction unit via a heated line. The pressure of the gas is first regulated to 5 Pa above ambient pressure and then throttled to ambient pressure p a via a heated needle valve. The needle valve is used to set a gas flo of 3 l min -. This is largely independent of the pressure in the saturation chamber ([6], [7]), as the admission pressure of the needle valve is alays held stable at p a + 5 Pa. Finally, the humid gas is fed into a measuring chamber here the required relative humidity is achieved (Fig. ), and ultimately released into the open air via an exhaust line. By varying the pressure in the saturation chamber beteen approximately MPa and 5 Pa at an ambient pressure of p a = Pa, the relative humidity in the measuring chamber can be varied beteen % and 95%. The displayed relative humidity U in the measuring chamber is derived from the saturation vapour pressure in the saturator e (t,p ) [4] at temperature t and pressure p, the relationship beteen the pressures in the saturator and the measuring chamber, and the associated saturation vapour pressure in the measuring chamber at temperature t according to Eq. () and Eq. (). device under test measuring chamber saturation chamber (saturator) heated high-pressure line heated pressure regulator and needle valve for flo adjustment p measurement pressure supply (. MPa) ith pressure regulator (rel. humidity adjustment) p measurement lo-pressure line ith temp. conditioning Fig.. Principle scheme of the miniaturized to-pressure generator. v. / Modification rights reserved

4 U e ( t, p) p = e ( t, p ) p () e ( t, p) = e ( t) f ( t, p) () e (t) saturation vapour pressure in the pure phase in respect to ater [], [3] ƒ (t,p) enhancement factor of air [5] e (t,p) saturated vapour pressure of the actual system in the presence of air or nitrogen in respect to ater [4] The measuring chamber and saturation chamber are manufactured jointly from a solid, highly heat-conductive metal bloc (special Aluminium alloy), the measuring chamber being positioned in the saturation chamber (Fig. ). The generator has to be operated either ith distilled ater or deionised ater and the user is obliged to change the ater each month for purity reasons. Hoever, failing to change the ater for several months resulted in no measurable contamination effects on the achieved relative humidity. Before being discharged into the measuring chamber, the humid gas is also fed repeatedly through gas conduits inside the all of the saturator hich very effectively brings the temperature of the gas into equilibrium ith that of the saturator. These design features ensure that the saturator temperature t and measuring chamber temperature t are practically the same and Eqs. () and () can therefore be simplified: t = t = t U e ( t, p) p e( t) f ( t, p) p p f ( t, p) = = = e ( t, p ) p e ( t) f ( t, p ) p p f ( t, p ) (3a) Equality of temperatures is ensured by the construction of the chambers as described in Section and shon in Fig.. Also, the effects of changing environmental temperatures are shon in Fig. 7. Even if the ambient temperature differs by more then K from the chamber temperature, the effect on the achieved relative humidity is very small. Nevertheless, the user has to operate the generator under normal laboratory conditions ithout heat radiation sources. As the generator is operated in a restricted temperature range for an average temperature t, the relation ƒ(t,p )/ ƒ(t,p ) can be sufficiently approximated by a polynomial in p /p, ƒ(p /p ) f ( t, p) f ( t, p ) p f p (polynomial in p /p ) (3b) and the fundamental equation of a to-pressure humidity generator is obtained. U = p p p f p [%] (3) Apart from a small correction parameter that taes into account the real saturation behaviour of air, the displayed v. / Modification rights reserved

5 humidity depends only on the pressure relationship beteen the measuring chamber and the saturation chamber. The pressure measurement and its measuring uncertainty are therefore crucially significant. Although other device-specific properties such as temperature homogeneity and saturation behaviour play a role in calibrating a generator and can lead to either systematic display errors or increased uncertainty, this design means these properties ill not change significantly beteen the individual devices or over the course of time and can be corrected by adjusting the device. 3. ADJUSTMENT AND CALIBRATION CONCEPT A ey objective in the development of the humidity generator as to eep system costs lo. For the pressure measurement, industrial pressure transmitters ith a specified accuracy of.5 % of the maximum readout are used. A simple estimate shos immediately that the pressure sensor specification leads to an uncertainty of up to 6 % rh for the displayed relative humidity. In principle, this deviation could be corrected by adjusting the generator, e.g. ith reference to a chilled-mirror hygrometer. Hoever, similar problems are immediately encountered hen looing at the long-term stability of the pressure sensors, hich means that this procedure does not produce an acceptable result ithout employing additional measures. An adjustment concept is therefore implemented, in hich the tolerances and drifts of the sensors are largely cancelled out by means of softare-based adjustment of the to pressure sensors for p and p at ambient pressure, thereby ensuring the humidity generator achieves a high degree of accuracy and stability. The individual steps in the adjustment concept - basic adjustment - adjustment - alignment are presented in the sections that follo and an uncertainty estimate is performed for the humidity generator, both in its ne state and after one year. All steps of the adjustment concept can be repeated at any time, but usually for the user there is only the need to do step 3 (alignment). 3.. Basic adjustment For pressure measurement, industrial sensors ith a specified tolerance of.5 % of the measuring range are used, the specification being interpreted as a rectangularly distributed uncertainty. The pressure sensors are described by the folloing model: Saturation chamber p : Measuring range MPa p = a + a p (4) a = u(a ) = 5.8 Pa a = u(a ) = 5.8E-3 (=) (=) v. / Modification rights reserved

6 Measuring chamber p : Measuring range. MPa p = b + p (5) b = u(b ) =.5 Pa (=) p = p a p a...ambient pressure p, p measured values in saturation chamber and measuring chamber, respectively The uncertainties of the pressure measurement are derived from the models of the pressure sensors, Eqs. (4) and (5), using: / u( p ) = u ( a ) + p u ( a (6) [ ] ) u ( p) = u( b ) (7) As the aim of the paper is to point out the effects of pressure sensor tolerance and drift and ho to eliminate them by a special adjustment procedure, here the measurement uncertainties of p and p are neglected. Besides, the contribution of the measurement uncertainties of p and p is very small compared ith the effects of pressure sensor tolerance (~. %rh compared ith 6 %rh) If the pressure sensors are used ithout any additional measures, the uncertainty of the displayed relative humidity can be determined as follos from Eq. (3):.8 uncertainty contribution pressure sensors delta u, % rh measurement uncertainty using a chilled mirror hygrometer (=) rel.humidity u, % rh Fig. 3. Deviation of indicated relative humidity of 3 generators from a certified chilled-mirror hygrometer after basic adjustment. Uncertainty contributions of pressure sensors after basic adjustment, including the contribution of pressure measurement uncertainty and the measurement uncertainty using a chilled-mirror hygrometer of about. % at 5 % and.6 % at 75 %, respectively (all =), are also indicated v. / Modification rights reserved

7 u( U hich leads to the uncertainty mentioned above of ~ 6 % rh at 95 %rh (=) (the small correction made by the enhancement factors has been omitted here as it does not mae any significant contribution to the total uncertainty). In the basic adjustment, the generator is depressurized, both pressure sensors are read at ambient pressure pa, and the difference (p a ) = p (p a ) p (p a ) calculated. Using the difference, a corrected value p is calculated for the saturator pressure and a revised model equation is obtained for the corrected pressure p / p ) = u( p) + ( ) u p p p (8) p = a + a p p ) + ( pa) = b + a ( p a p a (9) and the uncertainty of p is u ( p ) = u ( b ) + ( p pa) u ( a) () Basically, the offset of the pressure measurement p at ambient pressure p a is corrected to the same value as p and the to pressure measurements synchronized at ambient pressure. The readings from the to pressure sensors thereby become correlated, ith total correlation at ambient pressure. The uncertainty of the displayed humidity, taing into account the correlation, is derived as follos from Eqs. (3), (5) and (9): [( p p ) u ( a )] p p p u ( U ) = u ( b ) a p p p p () It can easily be demonstrated that the uncertainty of the displayed relative humidity u(u ) for p = p = p a ( U = %) becomes exactly zero (total correlation). In practice, the uncertainty of the pressure measurements includes the standard deviation from random variations, the limited resolution of the electronics, and the sensor stability must also be taen into account, hich results in a further small residual uncertainty of approximately. % rh at the adjustment point. Figure 3 shos measurements of the displayed humidity compared to a certified chilled-mirror hygrometer after the basic adjustment. The uncertainty calculated as per Eq. () resulting from the pressure sensor tolerance is also shon, including the contribution from the pressure measurement uncertainty of about. %. Most measuring values lie ithin the uncertainty estimate of Eq. (). Hoever, if the measuring uncertainty of the chilled mirror hygrometer is also tae into account, all measurements and the uncertainty estimate resulting from the pressure sensor contribution are compatible after the basic adjustment. 3.. Adjustment of the humidity generator After the basic adjustment, the humidity generator is checed against a certified chilled-mirror hygrometer and the deviations are stored as a correction table. Figure 4 shos the setup for the calibration of the generator. The output of the gas stream into the measuring chamber is closed and the gas stream is discharged via a special insulated pipe. A Pt reference thermometer is placed ithin the pipe. The gas stream is then routed via a heated line and split by a T-section, ith one part (typically.5 l min - ) being sent to the chilled mirror hygrometer hile the rest is released into the open air via a needle valve controlling the gas stream split. Located behind v. / Modification rights reserved

8 needle valve flo meter chilled-mirror hygrometer (DPH) heated exhaust line Pt closed standard exhaust Fig. 4. Setup for verifying a generator (principle scheme) using a chilled-mirror hygrometer. the chilled-mirror hygrometer is a flometer to control the gas stream through the chilled-mirror hygrometer (DPH). When calculating the relative humidity from temperature and de-point temperature, the folloing ey uncertainty contributions are taen into account: DPH calibration uncertainty Stability of DPH in the calibration interval De-point stability during the measurement Pressure loss in the measuring line Pt calibration uncertainty Pt self heating Pt heat conduction Stability of the Pt in the calibration interval Temperature stability during calibration Spatial temperature inhomogeneity in the saturation and measurement chambers Resolution of the humidity generator When adjusted in this ay, the generator exhibits deviations from the chilled-mirror reference ithin the specified tolerance indicated in Fig. 5. The total uncertainty of the generator is calculated from the specification limits (rectangular distribution) and the uncertainty of the calibration using the chilled-mirror hygrometer. The spatial v. / Modification rights reserved

9 u(u ), % ( = ) total uncertainty HUMOR accuracy (specification) uncertainty of reference measurement rel.humidity u, % Fig. 5. Specification, calibration uncertainty, and total uncertainty (=) of a ne generator. distribution of the relative humidity ithin the measuring chamber is largely determined by the temperature inhomogeneity of. C (difference beteen maximum and minimum values, rectangularly distributed) in the measuring chamber, varies from.3 to.33 % rh depending on the displayed humidity, and is included in the uncertainty calculation Alignment By means of the adjustment outlined in Section 3., all systematic deviations of the generator from an ideal generator ere corrected, ithin the measuring uncertainty of a DPH. In particular, the systematic readout errors of the pressure sensors ere also corrected. During operation, hoever, a generator drift must be reconed ith. This derives almost exclusively from drift of the pressure sensors. For the pressure sensors, the manufacturer specifies a maximum drift per year of.5 Pa (p ) or Pa (p ). For the annual drift, manufacturer s specifications (rectangular distributed) are doubled, resulting in the folloing models for the additional uncertainty of the pressure measurement after one year, based on Eqs. (4) and (5): p = a' + a'' p () a = u(a ) =.3 Pa (=) a = u(a ) =.3E-3 (=) p = b ' + p a (3) b = u(b ) =.5 Pa (=) Exactly as ith the basic adjustment in Section 3., here too the generator is depressurized and the pressure sensors are once again synchronized at ambient pressure. Unlie the basic adjustment, hoever, the generator correction values as per Section 3. v. / Modification rights reserved

10 u(u ), % ( = ) total uncertainty (ne) total uncertainty after year drift contribution pressure sensors change in total uncertainty after year rel.humidity u, % Fig. 6. Total uncertainty of a ne generator and after one year. Also, the drift contributions of the pressure sensors and the change in total uncertainty after one year are indicated (=). are retained in the alignment, i.e. only the change occurring ithin one year is examined. The additional uncertainty contribution after a year is then derived as follos, according to Eq. (): [( p p ) u ( a ')] p p p u ( U ) = u ( b ') a p p p p (4) Figure 6 shos the uncertainty contribution resulting from the drift after one year. An additional contribution resulting from the resolution of the electronics has also been taen into account. The total uncertainty hen ne and after one year, and the increase in uncertainty after one year, have also been entered. The generator can be aligned by the user at any time and as often as required. An alignment is particularly recommended henever the humidity readout deviates significantly i.e. more than.3 % rh from % rh hen the generator is depressurized. 4. MEASUREMENTS For stable humidity display, the generator requires only a room ith normal climatic conditions and no direct sources of thermal radiation. In order to demonstrate the stability of the humidity display ith fluctuating temperatures, a generator as stabilized at 75 % and exposed to changing temperature conditions in a temperature cabinet, hile at the same time the displayed humidity as monitored ith a chilled-mirror hygrometer (Fig. 7). At different ambient temperatures of C, 5 C, and 4 C, the largest change in the achieved humidity is less than. % rh. During the temperature change from C to 5 C and from 5 C to 4 C, the humidity display v. / Modification rights reserved

11 deviation of achieved rel.humidity delta.u, % time, min Fig. 7. Influence of the ambient temperature t a. Different ambient temperatures of C, 5 C, and 4 C lead to a change in achieved relative humidity of less than. %. ta = 5 C --> 4 C ta = C --> 5 C temperature difference measuring chamber to ambient temp., K deviation from adjusted value, % rh delta generator delta sample adjusted value adjusted rel.humidity u, % -. time, hour Fig. 8. Response of generated humidity and of a device under test after selecting a ne value. is incorrect by less than %, even though there is a difference of up to C beteen the temperature of the generator s measuring chamber and the ambient temperature. Figure 8 shos the response of the humidity display over time, the deviation of a humidity meter aaiting calibration from the displayed humidity, and the deviation of the generator from the target humidity. The target humidity is stable to ithin. % rh after a fe minutes; the setting time is primarily determined by the speed at hich a ne target pressure can be set in the saturator. The setting response of the device under test is largely determined by its properties. The generator itself maes practically no contribution to the stabilization time. The humidity generator can be traced to international units (SI units) by means of its calibration against a certified chilled-mirror hygrometer. Because of its high stability and small uncertainty, it is used as a humidity reference device in various calibration laboratories. v. / Modification rights reserved

12 . deviation from reference, % rh NMI t = 5 C NMI t = C HUMOR t = 4 C -. relative humidity u, % Fig. 9. Comparison of calibration measurements using - the described miniaturized to-pressure generator (HUMOR ) - the national Austrian reference generator (NMI) The uncertainties of each generator (=) are indicated. Figure 9 compares the calibration of a humidity meter based on a capacitive polymer sensor carried out using the Austrian national standard generator at temperatures ( and 5 C) ith a calibration using the described HUMOR miniaturized to-pressure generator as a reference device of the Austrian Calibration Service (ÖKD). The accredited uncertainties (=) are shon for each case. Within the combined measuring uncertainties, no differences in the calibrations can be identified. References. S. Hasegaa, J.W. Little, J. Res. Nat. Bur. Stand. 8A, 8 (977). A. Wexler, J. Res. Nat. Bur. Stand -A. Physics and Chemistry 8A, 775 (976) 3. D. Sonntag, Z. Meteorol. 7, 34 (99) 4. L.P. Harrison, in Humidity and Moisture, Vol. 3, Fundamentals and Standards, A.Wexler, W.A.Wildhac, (Eds.) (Reinhold Publishing, Ne Yor 965), pp L. Greenspan, J. Res. Nat. Bur. Stand -A. Physics and Chemistry 8A, 4 (976) 6. Accorded patent, US B 7. Accorded patent, EP B v. / Modification rights reserved

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration Relative Humidity Measurement and Calibration The widest range of humidity products from a pioneer of RH sensor development The best in Relative Humidity technology to optimize your process Whether you

More information

Rense Product Range. Customized Humidity Solutions. Rense Product Range

Rense Product Range. Customized Humidity Solutions. Rense Product Range Customized Humidity Solutions Rense Product Range Background to Michell Instruments Michell Instruments is the international leader in the field of moisture and humidity measurement solutions. With over

More information

Case study Realistic and best uncertainties with humidity instruments

Case study Realistic and best uncertainties with humidity instruments Case study Realistic and best uncertainties with humidity instruments Domen Hudoklin September 2014 Brdo, Slovenia Humidity calibrations Dewpoint primary - against dewpoint generator secondary - against

More information

REDUCING RISK OF HAND-ARM VIBRATION INJURY FROM HAND-HELD POWER TOOLS INTRODUCTION

REDUCING RISK OF HAND-ARM VIBRATION INJURY FROM HAND-HELD POWER TOOLS INTRODUCTION Health and Safety Executive Information Document HSE 246/31 REDUCING RISK OF HAND-ARM VIBRATION INJURY FROM HAND-HELD POWER TOOLS INTRODUCTION 1 This document contains internal guidance hich has been made

More information

NIST Calibration Services for Liquid Volume NIST Special Publication 250-72

NIST Calibration Services for Liquid Volume NIST Special Publication 250-72 NIST Calibration Services for Liquid olume NIST Special Publication 250-72 ern E. Bean, Pedro I. Espina, John. D. Wright, John F. Houser, Sherry D. Sheckels, and Aaron N. Johnson March 24, 2006 Fluid Metrology

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

CRITERIUM FOR FUNCTION DEFININING OF FINAL TIME SHARING OF THE BASIC CLARK S FLOW PRECEDENCE DIAGRAMMING (PDM) STRUCTURE

CRITERIUM FOR FUNCTION DEFININING OF FINAL TIME SHARING OF THE BASIC CLARK S FLOW PRECEDENCE DIAGRAMMING (PDM) STRUCTURE st Logistics International Conference Belgrade, Serbia 8-30 November 03 CRITERIUM FOR FUNCTION DEFININING OF FINAL TIME SHARING OF THE BASIC CLARK S FLOW PRECEDENCE DIAGRAMMING (PDM STRUCTURE Branko Davidović

More information

Relative Humidity Calibration Kit

Relative Humidity Calibration Kit Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration

More information

Infrared Thermometer Calibration A Complete Guide

Infrared Thermometer Calibration A Complete Guide Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate

More information

Dew-Point Measurement Solutions

Dew-Point Measurement Solutions Dew-Point Measurement Solutions www.michell.com Dew-Point Measurement Solutions Special Features No downtime Michell Instruments offers a unique sensor exchange programme, which means that you will never

More information

How is Dew Point Measured Reliably?

How is Dew Point Measured Reliably? How is Dew Point Measured Reliably? Air compression increases vapor pressure and thus dew point. = water vapor Vapor pressure Frost point Saturation - 0 C + Dew point Monitoring with alarming and real

More information

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS XVII IMEKO World Congress Metrology in the 3rd Millennium June 22 27, 2003, Dubrovnik, Croatia LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS Tanfer Yandayan TUBITAK-Ulusal

More information

A MEASUREMENT ASSURANCE PROGRAM FOR CALIBRATION BY THE TRANSFER METHOD CS800

A MEASUREMENT ASSURANCE PROGRAM FOR CALIBRATION BY THE TRANSFER METHOD CS800 A EAUREENT AURANCE PROGRA FOR CALIBRATION BY THE TRANFER ETHOD C800 JEFF PEERY DEC. 6, 011 This document outlines the easurement Assurance Program (AP for eaetrics C800 flo measurement system a liquid

More information

The photoionization detector (PID) utilizes ultraviolet

The photoionization detector (PID) utilizes ultraviolet Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).

More information

Environment-Laboratory Ambient Conditions

Environment-Laboratory Ambient Conditions ISSUE 3.0: MADE MINOR CLARIFICATION CHANGES. FOR CLIENT LABS Purpose To assure laboratory conditions are stable during the performance of testing and data collection. Why this requirement is important?

More information

Online Infrared Flue Gas Analyzer Gasboard 3000

Online Infrared Flue Gas Analyzer Gasboard 3000 Online Infrared Flue Gas Analyzer Gasboard 3000 O 2 CO CO2 SO 2 NO NO x Use of measurement methods Proprietary Infrared NDIR Detectors - Micro-flow in ppm range: CO, NO, SO2, CO2 - Dual beam in % volume

More information

Traceable Moisture Measurements by Means of Residual Water Detection

Traceable Moisture Measurements by Means of Residual Water Detection 1/24 Traceable Moisture Measurements by Means of Residual Water Detection Peter Østergaard Danish Technological Institute peo@dti.dk 2/24 What you will learn from this presentation Why SI traceability

More information

PC BASED PID TEMPERATURE CONTROLLER

PC BASED PID TEMPERATURE CONTROLLER PC BASED PID TEMPERATURE CONTROLLER R. Nisha * and K.N. Madhusoodanan Dept. of Instrumentation, Cochin University of Science and Technology, Cochin 22, India ABSTRACT: A simple and versatile PC based Programmable

More information

DM Series CONTENTS. 1. Features 3. 2. Applications 3. 3. General 4. 4. Calibration 5. 5. Dimensions 6. 6. Specifications 7. 7. Calibration procedure 7

DM Series CONTENTS. 1. Features 3. 2. Applications 3. 3. General 4. 4. Calibration 5. 5. Dimensions 6. 6. Specifications 7. 7. Calibration procedure 7 DM Series CONTENTS 1. Features 3 2. Applications 3 3. General 4 4. Calibration 5 DM-509-(M/T/S)X-(01/02) & DM-509-(H3/H5)-(01/02) 5. Dimensions 6 6. Specifications 7 7. Calibration procedure 7 The DM Series

More information

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration

More information

Topic 4: Introduction to Labour Market, Aggregate Supply and AD-AS model

Topic 4: Introduction to Labour Market, Aggregate Supply and AD-AS model Topic 4: Introduction to Labour Market, Aggregate Supply and AD-AS model 1. In order to model the labour market at a microeconomic level, e simplify greatly by assuming that all jobs are the same in terms

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK code Location code Customers Sites locations: 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK Calibration Centre Bolkiah Garrison BB3510 Negara Brunei Darussalam Contact: Mr Lim Tiong Thai Tel: +673-2-386475

More information

Conductivity Sensor Calibrations to Meet Water Industry Requirements

Conductivity Sensor Calibrations to Meet Water Industry Requirements Conductivity Sensor Calibrations to Meet Water Industry Requirements Victor M. Braga Technical Service/Training Manager Mettler-Toledo Thornton Inc. 36 Middlesex Turnpike Bedford, Massachusetts 01730 Phone

More information

Pressure Dew Point Probe 0636.9840 / 0636.9841 Application Information

Pressure Dew Point Probe 0636.9840 / 0636.9841 Application Information Pressure Dew Point Probe 0636.9840 / 0636.9841 Application Information en Contents Instrument description...3 User instructions...4 Safety instructions...5 Folding ferrite...5 Operating instructions...6

More information

On-line Dissolved Gas Analysis

On-line Dissolved Gas Analysis On-line Dissolved Gas Analysis accord. to IEC 567/ASTM 3612 Online-Monitoring of transformers Automatic on-line analysis of 11 gases The analysis is done in 2 steps with high selectivity and accuracy fully

More information

Guidelines on the Calibration of Static Torque Measuring Devices

Guidelines on the Calibration of Static Torque Measuring Devices European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET cg-14 Version.0 (03/011) Previously EA-10/14 Calibration Guide EURAMET cg-14

More information

Calibration of Dallas sensors

Calibration of Dallas sensors Calibration of Dallas sensors Mariusz Sapinski INFN Sezione di Roma1 Roma, Italy (April 2006) 1. Objectives The goal of this work is to perform a calibration of Dallas sensors. Nine Dallas sensors are

More information

THE STATISTICAL TREATMENT OF EXPERIMENTAL DATA 1

THE STATISTICAL TREATMENT OF EXPERIMENTAL DATA 1 THE STATISTICAL TREATMET OF EXPERIMETAL DATA Introduction The subject of statistical data analysis is regarded as crucial by most scientists, since error-free measurement is impossible in virtually all

More information

User checks and maintenance of laboratory balances

User checks and maintenance of laboratory balances Technical Note 13 March 2014 Issued: July 1995 Amended and reissued: July 2005, July 2007, October 2007, May 2010, March 2013, March 2014 User checks and maintenance of laboratory balances Copyright National

More information

The Research on Demand Forecasting of Supply Chain Based on ICCELMAN

The Research on Demand Forecasting of Supply Chain Based on ICCELMAN 505 A publication of CHEMICAL ENGINEERING TRANSACTIONS VOL. 46, 2015 Guest Editors: Peiyu Ren, Yancang Li, Huiping Song Copyright 2015, AIDIC Servizi S.r.l., ISBN 978-88-95608-37-2; ISSN 2283-9216 The

More information

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and

More information

16. Industrial Platinum Resistance Thermometers

16. Industrial Platinum Resistance Thermometers 129 16. Industrial Platinum Resistance Thermometers Industrial platinum resistance thermometers (IPRTs) are constructed from platinum of a lower quality than is used for SPRTs; typically, values of W(H

More information

Leveraging Multipath Routing and Traffic Grooming for an Efficient Load Balancing in Optical Networks

Leveraging Multipath Routing and Traffic Grooming for an Efficient Load Balancing in Optical Networks Leveraging ultipath Routing and Traffic Grooming for an Efficient Load Balancing in Optical Netorks Juliana de Santi, André C. Drummond* and Nelson L. S. da Fonseca University of Campinas, Brazil Email:

More information

NRC Publications Archive Archives des publications du CNRC

NRC Publications Archive Archives des publications du CNRC NRC Publications Archive Archives des publications du CNRC Digital 3D Imaging and Modeling: A Metrological Approach Beraldin, Jean-Angelo Publisher s version / Version de l'éditeur: Time Compression Technologies

More information

USE OF REFERENCE MATERIALS IN THE LABORATORY

USE OF REFERENCE MATERIALS IN THE LABORATORY USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently

More information

COUNTERBALANCE VALVES

COUNTERBALANCE VALVES COUNTERBALANCE VALVES Introduction They are modulating valves which allow free flow into the actuator and then block the reverse flow until they feel a pilot pressure inversely proportional to the load

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Old Bracknell Lane West Bracknell Berkshire RG12 7AH Contact: Mr M Trotter Tel: +44 (0)1344 459314 Fax: +44 (0)1344 465556 E-Mail: mtrotter@bis.fm

More information

HM1500LF Relative Humidity Module

HM1500LF Relative Humidity Module Small size product Product free from Lead, Cr(6+), Cd and Hg Humidity calibrated within +/-2% @55%RH Typical 1 to 3.6 Volt DC output for 0 to 100% RH at 5Vdc supply Ratiometric to voltage supply from 4.75Vdc

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

Estimating Weighing Uncertainty From Balance Data Sheet Specifications

Estimating Weighing Uncertainty From Balance Data Sheet Specifications Estimating Weighing Uncertainty From Balance Data Sheet Specifications Sources Of Measurement Deviations And Uncertainties Determination Of The Combined Measurement Bias Estimation Of The Combined Measurement

More information

THE HUMIDITY/MOISTURE HANDBOOK

THE HUMIDITY/MOISTURE HANDBOOK THE HUMIDITY/MOISTURE HANDBOOK Table of Contents Introduction... 3 Relative Humidity... 3 Partial Pressure... 4 Saturation Pressure (Ps)... 5 Other Absolute Moisture Scales... 8 % Moisture by Volume (%M

More information

Degrees of Freedom In HVAC Controls Design BY JOHN VARLEY, P.E., HBDP, MEMBER ASHRAE

Degrees of Freedom In HVAC Controls Design BY JOHN VARLEY, P.E., HBDP, MEMBER ASHRAE This article was published in ASHRAE Journal, June 05. Copyright 05 ASHRAE. Posted at www.ashrae.org. This article may not be copied and/or distributed electronically or in paper form without permission

More information

Calibration. How can you get lower costs for the calibration? What is traceability? What is measurement uncertainty? What is calibration?

Calibration. How can you get lower costs for the calibration? What is traceability? What is measurement uncertainty? What is calibration? Calibration Tvinnargatan 25 SE-507 30 Brämhult SWEDEN Phone: +46 33 22 56 30 Fax: +46 33 13 88 71 info@kalibrera.se www.kalibrera.se In May 2011 Elastocon purchased the company Mätcentrum in Eskilstuna,

More information

Calibrating DC Shunts: Techniques and Uncertainties

Calibrating DC Shunts: Techniques and Uncertainties Calibrating DC Shunts: Techniques and Uncertainties Jay Klevens Ohm-Labs, Inc. 611 E. Carson St., Pittsburgh, PA 15203 (412) 431-0640 jklevens@ohm-labs.com Abstract Accurate electrical current measurement

More information

COMBIMASS. Technical Data COMBIMASS eco-bio +

COMBIMASS. Technical Data COMBIMASS eco-bio + COMBIMASS Technical Data THE SYSTEM COMBIMASS The field transmitters of the COMBIMASS eco series are suitable for gas flow measurement and cover a wide range of different applications. The instruments

More information

Ti-Pure TITANIUM DIOXIDE. DETERMINATION OF BROOKFIELD VISCOSITY, TiO 2 SLURRY METHOD: T4400.535.01.WP

Ti-Pure TITANIUM DIOXIDE. DETERMINATION OF BROOKFIELD VISCOSITY, TiO 2 SLURRY METHOD: T4400.535.01.WP Ti-Pure TITANIUM DIOXIDE DETERMINATION OF BROOKFIELD VISCOSITY, TiO 2 SLURRY METHOD: T4400.535.01.WP 1 TITANIUM DIOXIDE (Ti-Pure ) Determination of Viscosity by Brookfield Viscometer I. Principle Viscosity

More information

Not All are Equal Many questions have arisen since the widespread availability of wideband air-fuel meters.

Not All are Equal Many questions have arisen since the widespread availability of wideband air-fuel meters. By Mike Kojima [Note: This is extracted from, and owned by, www.fordmuscle.com. For complete text, and to view discussions, please visit: http://www.fordmuscle.com/archives/2007/06/widebandshootout/index.php

More information

CALIBRATION PRINCIPLES

CALIBRATION PRINCIPLES 1 CALIBRATION PRINCIPLES After completing this chapter, you should be able to: Define key terms relating to calibration and interpret the meaning of each. Understand traceability requirements and how they

More information

Initial Experiments of a Novel Liquid Desiccant Dehumidifier for Industrial and Comfort Air Conditioning Systems

Initial Experiments of a Novel Liquid Desiccant Dehumidifier for Industrial and Comfort Air Conditioning Systems Abstract Initial Experiments of a Novel Liquid Desiccant Dehumidifier for Industrial and Comfort Air Conditioning Systems M. Jaradat, R. Heinzen, U. Jordan, K. Vajen Kassel University (Germany), Institute

More information

Research on the Air Conditioning Water Heater System

Research on the Air Conditioning Water Heater System Purdue University Purdue e-pubs International Refrigeration and Air Conditioning Conference School of Mechanical Engineering 28 Research on the Air Conditioning Water Heater System Fei Liu Gree Electric

More information

How To Use A Ds340 (Dsp)

How To Use A Ds340 (Dsp) Ideal for monitoring dust levels in the exhaust gas of industrial combustion or air filtration processes. Innovative Dynamic Detection Principle (DDP) measurement technique Immune to gradual reductions

More information

DUST EMISSION MONITORING SYSTEM

DUST EMISSION MONITORING SYSTEM Applications Measurement of dust concentration in dry or moist, steamsaturated and corrosive stack gases Monitoring of municipal and hazardous waste and sewage sludge incinerators Monitoring of power plants

More information

AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002

AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002 AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002 First Edition MONITORING AND LABORATORY DIVISION August

More information

Calibrating Differential Pressure Sensors

Calibrating Differential Pressure Sensors Calibrating Differential Pressure Sensors Additel Corporation 22865 Savi Ranch Parkway, Ste F Yorba Linda, CA 92887 Introduction Differential pressure sensors are common in the process industry and cover

More information

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter Thermal Mass meter and Temperature Transmitter Description The MTI10 insertion thermal mass flowmeter and the MTL10 in-line thermal mass flowmeter provide accurate mass flow measurement of clean, dry gases

More information

COLLATERAL RESPONDING UNDER A DRL SCHEDULE'

COLLATERAL RESPONDING UNDER A DRL SCHEDULE' JOURNAL OF THE EXPERMENTAL ANALYSS OF BEHAVOR RENFORCEMENT CONTNGENCES MANTANNG COLLATERAL RESPONDNG UNDER A DRL SCHEDULE' D. E. MCMLLAN2 DOWNSTATE MEDCAL CENTER STATE UNVERSTY OF NEW YORK 1959, 12, 413-422

More information

The Virtual Spring Mass System

The Virtual Spring Mass System The Virtual Spring Mass Syste J. S. Freudenberg EECS 6 Ebedded Control Systes Huan Coputer Interaction A force feedbac syste, such as the haptic heel used in the EECS 6 lab, is capable of exhibiting a

More information

THE EFFECTS OF TEMPERATURE VARIATION ON SINGLE-LENS-REFLEX DIGITAL CAMERA CALIBRATION PARAMETERS

THE EFFECTS OF TEMPERATURE VARIATION ON SINGLE-LENS-REFLEX DIGITAL CAMERA CALIBRATION PARAMETERS Coission V Symposium, Newcastle upon Tyne, UK. 2010 THE EFFECTS OF TEMPERATURE VARIATION ON SINGLE-LENS-REFLEX DIGITAL CAMERA CALIBRATION PARAMETERS M. J. Smith *, E. Cope Faculty of Engineering, The University

More information

Custody Transfer Measurement. with the V-Cone Flowmeter

Custody Transfer Measurement. with the V-Cone Flowmeter Custody Transfer Measurement with the V-Cone Flowmeter Stephen A. Ifft McCrometer Inc. Hemet, California, USA Abstract This paper will discuss the approval of the McCrometer V-Cone flowmeter for custody

More information

Duct Humidity Transmitter

Duct Humidity Transmitter SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers

More information

White Paper. Detecting Hydrogen Sulfide Gas and Understanding its Danger in the Field. Introduction

White Paper. Detecting Hydrogen Sulfide Gas and Understanding its Danger in the Field. Introduction White Paper PN FGD_TWP_H2S_Hydrogen_Sulfide_Gas_Understand_Danger January 2015 Detecting Hydrogen Sulfide Gas and Understanding its Danger in the Field Introduction Many oil fields, especially mature ones,

More information

M&V of Hot Water Boiler Plant. Boban Ratkovich, P. Eng, CEM, BESA, LEED AP President CES Engineering Ltd

M&V of Hot Water Boiler Plant. Boban Ratkovich, P. Eng, CEM, BESA, LEED AP President CES Engineering Ltd M&V of Hot Water Boiler Plant Boban Ratkovich, P. Eng, CEM, BESA, LEED AP President CES Engineering Ltd AIA Quality Assurance Learning Objectives 1. Case study examples of procedures and obstacles in measuring

More information

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon

More information

Using Thermocouple Sensors Connecting Grounded and Floating Thermocouples

Using Thermocouple Sensors Connecting Grounded and Floating Thermocouples Connecting Grounded and Floating Thermocouples For best performance, Thermocouple sensors should be floating. This will ensure that no noise currents can flow in the sensor leads and that no common-mode

More information

Applying Pressure Independent Control Valves in H.V.A.C. Systems. A Presentation to: Orange Empire ASHRAE Santa Ana Nov. 17, 2009

Applying Pressure Independent Control Valves in H.V.A.C. Systems. A Presentation to: Orange Empire ASHRAE Santa Ana Nov. 17, 2009 Applying Pressure Independent Control Valves in H.V.A.C. Systems A Presentation to: Orange Empire ASHRAE Santa Ana Nov. 17, 2009 1 Introduction I know, as building design consultants, so much of your effort

More information

T0118 T2118 T3118. Instruction Manual

T0118 T2118 T3118. Instruction Manual Programmable indoor transmitter of temperature T0118 Programmable indoor transmitter of atmospheric pressure T2118 Programmable indoor transmitter of temperature, relative humidity and other derived humidity

More information

15. Liquid-in-Glass Thermometry

15. Liquid-in-Glass Thermometry 123 15. Liquid-in-Glass Thermometry Even though the liquid-in-glass thermometer is used much less frequently today than formerly, it is still a very commonly used device. Although it is normally an instrument

More information

the importance of ph in water purification the importance of ph in water purification Water Purification Systems UK

the importance of ph in water purification the importance of ph in water purification Water Purification Systems UK the importance of ph in water purification Water Purification Systems UK What is ph? The ph scale has been in use for over a century, ever since it was first defined in 1909 by the Danish chemist, Søren

More information

The NEW Advanced Range of Portable Temperature Calibrators PROVISIONAL. Portable Dry Blocks, Liquid Baths and Multi-functional Heat Sources

The NEW Advanced Range of Portable Temperature Calibrators PROVISIONAL. Portable Dry Blocks, Liquid Baths and Multi-functional Heat Sources PROVISIONAL The Source for Calibration Professionals Portable Dry Blocks, Liquid Baths and Multi-functional Heat Sources The NEW Advanced Range of Portable Temperature Calibrators Advanced Range Isotech

More information

Welcome to chapter 10. The following chapter is called Equipment maintenance and monitoring. The author is Ronny Janssens.

Welcome to chapter 10. The following chapter is called Equipment maintenance and monitoring. The author is Ronny Janssens. Welcome to chapter 10. The following chapter is called Equipment maintenance and monitoring. The author is Ronny Janssens. 1 After finishing this chapter, the student should be familiar with - the requirements

More information

Operating principle and construction of zirconium dioxide oxygen sensors of the XYA series

Operating principle and construction of zirconium dioxide oxygen sensors of the XYA series 1. PHYSICAL BACKGROUND 1.1 Partial pressure Definition: The partial pressure is defined as the pressure of a single gas component in a mixture of gases. It corresponds to the total pressure which the single

More information

A Demonstration Plant of a Liquid Desiccant Air Conditioning Unit for Drying Applications

A Demonstration Plant of a Liquid Desiccant Air Conditioning Unit for Drying Applications A Demonstration Plant of a Liquid Desiccant Air Conditioning Unit for Drying Applications Mustafa Jaradat, Klaus Vajen, Ulrike Jordan Institut für Thermische Energietechnik, Universität Kassel, 34125 Kassel,

More information

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications P.O. Box 26, FIN-00421 Helsinki, FINLAND Tel: +358 9 894 91 Fax: +358 9 8949 2485 Email: industrialsales@vaisala.com www.vaisala.com The Vaisala

More information

Solar air collectors for industry and larger halls S Ø. Efficient dehumidification and air heating for free...

Solar air collectors for industry and larger halls S Ø. Efficient dehumidification and air heating for free... Solar air collectors for industry and larger halls N S Ø Efficient dehumidification and air heating for free... This booklet contains information about SolarVenti Industrial air solar system. The system

More information

Instruction Manual. T3111 T3111L T3111P T3111Ex

Instruction Manual. T3111 T3111L T3111P T3111Ex www.cometsystem.cz Instruction Manual T3111 T3111L T3111P T3111Ex Programmable transmitter of temperature, relative humidity and other calculated humidity values with external probe and 4-20 ma outputs

More information

TEMPERATURE SENSOR USER MANUAL

TEMPERATURE SENSOR USER MANUAL TEMPERATURE SENSOR USER MANUAL 1 Temperature sensor user manual Copyright 2012 Unisense A/S Version October 2012 TEMPERATURE SENSOR USER MANUAL UNISENSE A/S TABLE OF CONTENTS WARRANTY AND LIABILITY....8

More information

Electricity. Confirming Coulomb s law. LD Physics Leaflets P3.1.2.2. 0909-Wie. Electrostatics Coulomb s law

Electricity. Confirming Coulomb s law. LD Physics Leaflets P3.1.2.2. 0909-Wie. Electrostatics Coulomb s law Electricity Electrostatics Coulomb s law LD Physics Leaflets Confirming Coulomb s law P3... Measuring with the force sensor and newton meter Objects of the experiments Measuring the force between two charged

More information

indoor air quality and energy saving TECHNICAL DATA HRE-TOP EC VENTILATION UNIT WITH HEAT RECOVERY FOR COMMERCIAL AND INDUSTRIAL BUILDINGS

indoor air quality and energy saving TECHNICAL DATA HRE-TOP EC VENTILATION UNIT WITH HEAT RECOVERY FOR COMMERCIAL AND INDUSTRIAL BUILDINGS indoor air quality and energy saving TECHNICAL DATA HRE-TOP EC VENTILATION UNIT WITH HEAT RECOVERY FOR COMMERCIAL AND INDUSTRIAL BUILDINGS HRETOP-EC Non-residential ventilator unit ith dual flo and high

More information

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

General and statistical principles for certification of RM ISO Guide 35 and Guide 34 General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in

More information

STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H)

STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H) STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H) by Joseph F. Greenbeck, Jr. And Courtney W. Taylor INTRODUCTION The following procedure is to be used for the maintenance and calibration

More information

Closed Loop Pressure Control for the Extrusion Process

Closed Loop Pressure Control for the Extrusion Process Closed Loop Pressure Control for the Extrusion Process By John Pacini Updated by Douglas Joy Extrusion is a continuous process and successful economic production depends on maintaining stable output and

More information

18.6.1 Terms concerned with internal quality control procedures

18.6.1 Terms concerned with internal quality control procedures 18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable

More information

New Trends in the Field of Automobile Air Conditioning

New Trends in the Field of Automobile Air Conditioning New Trends in the Field of Automobile Air Conditioning E. Janotkova and M. Pavelek Department of Thermomechanics and Environmental Engineering Brno University of Technology, 61669 Brno, Czech Republic

More information

DEGASSED CATION CONDUCTIVITY MEASUREMENT

DEGASSED CATION CONDUCTIVITY MEASUREMENT (Presented at EPRI's 8th International Conference on Cycle Chemistry in Fossil and Combined Cycle Plants with Heat Recovery Steam Generators - June 20-23, 2006 Calgary, Alberta Canada) DEGASSED CATION

More information

User Manual IC-485AI 2002-09-27

User Manual IC-485AI 2002-09-27 User Manual IC-485AI Note: This equipment has been tested and found to comply ith the limits for a Class A digital device pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable

More information

Temperature Calibration; Depths of Immersion

Temperature Calibration; Depths of Immersion Temperature Calibration; epths of Immersion Author: John P. Tavener Company Isothermal Technology Limited, Pine Grove, Southport, England Abstract Of all the sources of errors and uncertainties in thermal

More information

Energy and Flow Measurement for Hydronic Systems

Energy and Flow Measurement for Hydronic Systems Energy and Flow Measurement for Hydronic Systems Presented By: George Szyszko Applications Consultant MCR for ONICON Incorporated Why Measure Thermal Energy? To control something, you must first measure

More information

Features, Limitations & Potential Errors

Features, Limitations & Potential Errors MINIMIZING USER ERRORS IN ph MEASUREMENTS Features, Limitations & Potential Errors of the Tools of ph Measurements Advanced Sensor Technologies, Inc. 603 North Poplar Street Orange CA 92868-1011 USA General

More information

LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS

LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS 1. Define an accumulator and explain its function A hydraulic accumulator is a device that stores the potential energy of an incompressible fluid

More information

15 th International Conference on Experimental Mechanics

15 th International Conference on Experimental Mechanics PAPER REF: 3762 LABORATORY TESTS OF COMPRESSIVE CREEP IN CONCRETE SAMPLES: DEVELOPMENT OF SOLUTIONS FOR PROBLEMS OF ACTUATION, MEASUREMENT, DATA ACQUISITION AND LOAD BALANCING João Palma (*), Paulo Morais,

More information

Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD)

Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD) Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD) Ulrich Breuel 1, Barbara Werner 2, Petra Spitzer 3, Hans D. Jensen 4 1,2 Zentrum für Messen und

More information

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Various Technics of Liquids and Solids Level Measurements. (Part 3) (Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,

More information

Fundamentals of Mass Flow Control

Fundamentals of Mass Flow Control Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of

More information

CHEMICAL ENGINEERING LABORATORY 3, CE 427 DRYING OF SOLIDS

CHEMICAL ENGINEERING LABORATORY 3, CE 427 DRYING OF SOLIDS Revised 10/00 CHEMICAL ENGINEERING LABORATORY 3, CE 427 DRYING OF SOLIDS Introduction Material covered here pertains to Chapter 24 of M,S&H. Please read relevant sections of this chapter. Drying involves

More information

MEASURING MACHINES. Pratt & Whitney METROLOGY LABORATORY. Measurement Systems, Inc.

MEASURING MACHINES. Pratt & Whitney METROLOGY LABORATORY. Measurement Systems, Inc. METROLOGY LABORATORY Pratt & Whitney Measurement s, Inc. METROLOGY LABORATORY The Standard of Accuracy Pratt & Whitney Metrology Laboratory Machines are the standard of accuracy to which all other gages

More information

Application Requirement

Application Requirement Hazardous Gas Monitors Sensor Selection Overview Safety Monitoring (LEL) Toxic Limit Detection (PEL) Leak Detection Personal Safety Application Requirement Exposure Assessment (TWA) Ambient Air Quality

More information

8001782 Owner s Manual

8001782 Owner s Manual 8001782 Digital Infrared Thermometer Owner s Manual Introduction This instrument is a portable, easy to use compact-size digital thermometer with laser sighting designed for one hand operation. The meter

More information

OPTIMIZING WEB SERVER'S DATA TRANSFER WITH HOTLINKS

OPTIMIZING WEB SERVER'S DATA TRANSFER WITH HOTLINKS OPTIMIZING WEB SERVER'S DATA TRANSFER WIT OTLINKS Evangelos Kranakis School of Computer Science, Carleton University Ottaa,ON. K1S 5B6 Canada kranakis@scs.carleton.ca Danny Krizanc Department of Mathematics,

More information

Establishment of the Gas Flow Calibration Facility at CSIR-NPL, India

Establishment of the Gas Flow Calibration Facility at CSIR-NPL, India Establishment of the Gas Flow Calibration Facility at CSIR-PL, India Speaker: Shiv Kumar Jaiswal Fluid Flow Measurement Standards CSIR-ational Physical Laboratory (PLI) Dr. K.S. Krishnan Road ew Delhi

More information

Accurate Air Flow Measurement in Electronics Cooling

Accurate Air Flow Measurement in Electronics Cooling Accurate Air Flow Measurement in Electronics Cooling Joachim Preiss, Raouf Ismail Cambridge AccuSense, Inc. E-mail: info@degreec.com Air is the most commonly used medium to remove heat from electronics

More information