Establishment of the Gas Flow Calibration Facility at CSIR-NPL, India

Size: px
Start display at page:

Download "Establishment of the Gas Flow Calibration Facility at CSIR-NPL, India"

Transcription

1 Establishment of the Gas Flow Calibration Facility at CSIR-PL, India Speaker: Shiv Kumar Jaiswal Fluid Flow Measurement Standards CSIR-ational Physical Laboratory (PLI) Dr. K.S. Krishnan Road ew Delhi India Phone: ; Fax: Authors: Shiv Kumar Jaiswal, I. S. Taak and Chatar Singh Abstract Recently Gas Flow Calibration System (GFCS) has been established in the Fluid Flow Measurement Standard group of PLI in the flow range from 10 sccm (standard cubic centimeter per minute) to 1000 slm (standard litre per minute) at 0 ºC and kpa. The expanded measurement uncertainty of the system in the above flow range is ±0.2 % at k=2. This GFCS is suitable for calibration of different types of flowmeters i.e. mass flow controllers, mass flow meters, rotameters, totalizer type meters, digital flow calibrators, compact provers, etc used in many research, development and industrial applications mainly process controls, environmental monitoring, pharmaceuticals and drugs manufacturing, health assessment etc. and also in many ABL accredited laboratories. Present paper describes the details of the GFCS architecture and calibration of a rotameter (i.e. volumetric flow rate device) as a case study. The uncertainty has been estimated as per ISO GUM document guidelines [1]. The standards used for calibration are traceable to appropriate ational standards. Index Terms Gas Flow Calibration System, uncertainty, national metrology institute (MI), air compressor, laminar flow element, sonic nozzle, volume flow rate, mass flow rate. 1. Introduction The CSIR-ational Physical Laboratory, India (PLI) is the national metrology institute (MI) of India and it is custodian of all physical parameters except ionizing radiation. The Fluid Flow Measurement Standard group of CSIR-ational Physical Laboratory, India (PLI) has water meter testing and calibration facilities. The group has testing facility up to 50 mm and calibration facility up to 200 mm. Presently, the group is providing testing services of the water flow meters as per IS 779, IS 6784 and ISO 4064 documentary standards to the various user organizations traceable to the ational Standard. The upgradation of old calibration facility using state-ofthe-art instrumentation and control has been planned to make it comparable to advanced MIs and provide traceable measurements to Indian users. As a follow up of the upgradation, a new prototype automatic water flow calibration standard (i.e. primary standard of water flow) of size D100 was designed and developed indigenously. The expanded uncertainty of this new system at 2000 kg collection mass is 0.03% at k=2. Keeping in view the requirement of in-house users such as Chemical Metrology group, Radio and Atmospheric Science Division, Solar Energy

2 group, etc. and external users such as Pollution Control Boards, pharmaceutical industries, petrochemical industries, environmental monitoring equipment manufacturers, R&D laboratories, ABL accredited laboratories, gas flow meters (mass flow controllers, mass flow meters, rotameters, digital flow calibrators) manufacturers, aerospace industries, etc., the GFCS has been established recently at Fluid Flow Laboratory which is based on laminar flow elements and sonic nozzles. The flow range of laminar flow element is from 10 sccm to 50 slm with operating pressure of 100 kpa absolute to 525 kpa absolute. The flow range of sonic nozzle (also called critical flow nozzle) is from 10 slm to 1000 slm with operating pressure of 50 kpa absolute to 525 kpa absolute. The expanded uncertainty of the GFCS is ±0.2 % at k=2 in the flow range of 10 sccm to 1000 slm. This system is suitable for calibration of different types of flowmeters such as mass flow controllers, mass flow meters, rotameters, totalizer type meters, digital flow calibrators, compact provers, etc. The GFCS has capabilities to supports multi-gas calibration such as nitrogen, air, carbon dioxide, helium etc. In the present paper, the details of the system have been described highlighting calibration results of a rotameter as a case study. 2. Details of Architecture and Methodology of the GFCS The system is essentially consisted of laminar flow elements, sonic nozzles, high precision flow display device named as mass flow terminal, molstic for mounting laminar flow elements, sonic nozzles and device under calibrations along with associated instrumentations such as barometric pressure monitor and digital thermometer (PT-100 sensor with indicator). There are four numbers of laminar flow elements (i.e. 100 sccm, 1 slm, 10 slm and 30 slm) which cover the flow range of 10 sccm to 50 slm with operating pressure of 100 kpa absolute to 525 kpa absolute. The lower flow of 0.8 sccm is also achievable with increased uncertainty. There are two numbers of sonic nozzles having flow coefficients of 200 sccm/kpa and 2 slm/kpa which cover the flow range of 10 slm to 1000 slm with operating pressure of 50 kpa absolute to 525 kpa absolute which uses vacuum pump for realizing the flow below 30% of full scale (FS) rating of sonic nozzle. The laminar flow elements and sonic nozzles are known as molblocs in Fluke* terminology and represented as molbloc-l and molboc-s respectively. Each molbloc has two numbers of built-in PT-100 sensors and upstream and downstream pressure taps. It has one EEPROM chip which stores information like flow range, calibration coefficients, different types of calibration, etc. The output of laminar and sonic nozzle is read by mass flow terminal. The mass flow terminal is a high precision display device which consists of two nos. of high precision absolute pressure sensors (reference pressure transducers) for reading the upstream and downstream pressure of laminar flow elements and sonic nozzles. The terminal has one number of PT-100 sensor indicator for gas temperature measurement. The pressure, temperature and IST gas reference property data are used to realize the flow. The mass flow terminal is essentially a flow computer which displays the flow in different units such as sccm, slm, vlm, uccm, pccm, mg/s, etc. Due to use of absolute pressure sensors, the accuracy of differential pressure measurement has increased resulting into improvement of flow accuracy. The molbox terminal has features for leak testing in the flow line and over pressure protection system [2-3]. Figure 1 shows the photographs of the GFCS which also includes air/ gas supply source. * Commercial equipment and materials are identified in order to adequately specify certain procedures. In no case does such identification imply recommendation or endorsement by PL-India nor does it imply that the material or equipment identified are necessarily the best available for the purpose.

3 (a) Photograph showing system architecture and calibration set up of a rotameter (b) Air Receiver of 1000 L capacity (c) Air compressor system and gas cylinders Figure 1. Photograph of Gas Flow Calibration System

4 2.1. Realization of Flow in Laminar Flow Element The mass flow rate of a compressible fluid in laminar flow element is realized using Poiseulle s Law [3] as follows: where, q m = mass flow rate (kg/s) P 1 = upstream absolute pressure (Pa) P 2 = downstream absolute pressure (Pa) P = (P 1 + P 2 )/2= average of upstream and downstream pressures (Pa) T = absolute temperature of gas (K) = gas density under P, T conditions (kg/m 3 ) P,T q m P 1 P 2 P, T P, T 6 L 3 R h R= flow passage radius (m) h = gap between piston and cylinder (m) η = dynamic gas viscosity under P, T conditions (Pa.s) L = length of laminar flow (m) (1) The gas density P,T is calculated by the following equation: where, ρ = standard gas density P = standard pressure ( kpa) T = standard temperature ( K) Z = gas compressibility factor under standard conditions = gas compressibility factor under P, T conditions Z P, T P, T P T P Z T Z P, T (2) By defining: C D R h 3 6 L and which becomes C G (i.e. experimentally determined geometrical constant) by calibration, the equation (1) becomes (3) q m P P 1 P 2 T Z P, T P, T T P Z C G (4)

5 Thus, for laminar flow element, the flow is proportional to the differential pressure (P 1 -P 2 ), created across the laminar flow gap Realization of Flow in Sonic ozzle The sonic nozzles also known as critical flow venturies (CFV), measure the flow using a geometrically specified flow path where the flow rate is the maximum based on the upstream conditions of the gas. As long as the ratio of downstream pressure to upstream pressure is low enough, the flow is choked, i.e. the speed of the gas is limited to approximate local sonic velocity. The ratio of downstream pressure to upstream pressure is called back pressure ratio (BPR). The sonic nozzle is under choked condition if the value of BPR is less than 0.5. Once this condition is met, the mass flow rate of a compressible fluid in sonic nozzle is realized using the following formula [3]: q m A * C C R R M T 0 P 0 (5) where q m = mass flow rate (kg/s) A* = throat area (m 2 ) C = discharge coefficient C R = critical flow coefficient P 0 = absolute stagnation pressure (Pa) R = universal gas constant (J/kg-mole-K) M = molecular mass (kg/kg-mole) T 0 = Absolute stagnation temperature (K) In a sonic nozzle, the flow is increased or decreased by changing the upstream pressure to the nozzle using either a variable pressure regulator or a variable restriction. 3. Calibration Set-up and Method Figure 2 shows the schematic diagram of the calibration of a rotameters using molbloc. The flow can be also controlled by using a mass flow controller or flow control knob of rotameters, if the rotameter has provision for it. In calibration, the comparison method is used where volume flow rate read by standard meter (i.e. molbloc and mass flow terminal) is compared against volume flow rate indicated by meter under calibration (MUC). The volume flow rates are referenced to standard temperature (25 ºC) and standard barometric pressure ( kpa) conditions in our laboratory. The percentage error is estimated from the following formula: (Q vmuc Q vstd ) Percentage Error (% e) = (6) Q vstd

6 where, Q vmuc = the flow indicated by the MUC under normal temperature T and normal pressure P conditions Q vmuc = the flow indicated by the MUC corrected to standard temperature T S and standard Pressure P S conditions Q vstd = the flow measured by reference standard under standard temperature T S and standard pressure P S conditions Q vmuc is calculated from the following equation: P Ts Q vmuc = Q vmuc (7) P S T Flow In Pressure Regulator Molbloc eedle Valve Rotameter Flow Out Molbox1+ Terminal Barometer Thermometer Legend: Pressure cable Temperature & EEPROM cable Figure 2. Schematic diagram of a rotameter calibration using molboc. 4. Uncertainty Analysis 4.1. Mathematical Model The mathematical model in calibration of rotameter is as follows: e = [Q vmuc + δq A + δq R + δq ZD ] [Q vstd + δq vstd ] (8) P Ts and Q vmuc = Q vmuc (9) P S T where; e = error of measurement (absolute unit) Q vmuc = volume flow from the meter under calibration (MUC) corrected to standard pressure and temperature condition (absolute unit) δq A = error due repeatability in measurement of MUC (absolute unit)

7 δq R = error contribution due to resolution of MUC (absolute unit) δq ZD = error contribution due to zero drift of MUC (absolute unit) Q vstd = volume flow from the reference standard (absolute unit) δq vstd = error contribution from the reference standard (absolute unit) Q vmuc = volume flow indicated by the MUC under normal temperature T and normal pressure P (absolute unit) 4.2. Uncertainty Equation Assuming all the contributory factors as uncorrelated quantities, the combined uncertainty u c in volume flow rate is given by; u c 2 = (c 1 u 1 ) 2 +(c 2 u 2 ) 2 +(c 3 u 3 ) 2 + (c 4 u 4 ) 2 +(c 5 u 5 ) 2 + (c 6 u 6 ) 2 (10) where, u 1 (δq A ) = uncertainty due to repeatability in measurement of MUC u 2 (δq R ) = uncertainty due to resolution of MUC u 3 (δq ZD ) = uncertainty due to zero drift of MUC u 4 (δq STD ) = uncertainty from calibration certificate of the reference standard u 5 (P) = uncertainty due to barometric pressure monitor u 6 (T) = uncertainty due to thermometer (i.e. PT-100 sensor with indicator) The sensitivity coefficients of various contributions as c 1, c 2, c 3, c 4, c 5 and c 6 are estimated using; c 1 = e/ δq A =1, c 2 = e/ δq R =1, c 3 = e/ δq ZD =1, c 4 = e/ δq STD =1; c 5 = Q vmuc / P = Ts Q vmuc / (Ps T); c 6 = Q vmuc / T = -(P Ts Q vmuc )/ (Ps T 2 ) Since there are product components apart from additive components for uncertainty, therefore, it is easier to estimate the uncertainty in relative form using below given equation [4] and sensitivity coefficients are 1: [u c (e)/e] 2 = [u 1 (δq A )/Q VMUC ] 2 + [u 2 (δq R )/Q VMUC ] 2 + [u 3 (δq ZD )/Q VMUC ] 2 + [u 4 (δq VSTD )/Q VSTD ] 2 + [u 5 (P)/P] 2 + [u 6 (T)/T] 2 (11) The rotameter of 5 to 55 L/min flow range was calibrated using laminar flow standard. Table 1 shows the data of calibration of rotameter at a particular flow rate (i.e. 50 L/min). Type A uncertainty was determined from those data. Table 2 shows the uncertainty budget of rotameter calibration at 50 L/min. Depending on MUC and customer requirement, the % error of reading or FS may be reported.

8 Observation o. Table 1. Data of rotameter calibration at 50 L/min. MUC Flow C Ref. Flow C Error (MUC-STD) C FS Error % Average FS error= % Standard deviation (σ) = % Type A uncertainty = σ/ n = % Quantity X i Table 2. Uncertainty budget of rotameter calibration at 50 L/min. Estimate x i Limits ±Δ x i (%) Probability distribution Type A/B Repeatability of MUC (δq A ) 50 L/min - ormal, Type A Resolution of 1 L/min Rectangular, MUC (δq R ) Zero drift of 0 L/min 0 Rectangular, MUC (δq ZD ) Standard ormal, (δq STD ) L/min Barometer 100 kpa 0.01 ormal, Thermometer 25 C 0.04 ormal, Error (e) FS % Combined uncertainty u c (e) Expanded uncertainty U Relative standard uncertainty ±u i (%) Sensitivity coefficient c i Relative uncertainty contribution u i (y)= c i. u i (%) Degree of freedom ν i At coverage factor k= % (FS) Results and Future work In the present paper, the main features of GFCS have been reported. The calibration of rotameter using GFCS has also been presented along with detailed estimation of measurement uncertainty. In a typical calibration of rotameter, the uncertainty of measurement is found to be ±1.077 % at k=2. Similarly, if some improved artifact such as digital flow calibrators which have much better resolution and uncertainty is chosen, the estimated uncertainty would have been much improved.

9 In case of present rotameter calibration, resolution of the MUC is a dominant factor but digital gauges have better resolution. Our GFCS is a secondary standard but its uncertainty is comparable to primary standards like piston provers and bell provers. The advantages with piston provers and bell provers are that their traceability can be established in the fundamental units of length and time due to being primary device but same is not applicable to laminar flow elements and sonic nozzles. The laminar flow elements and sonic nozzles are made traceable by calibrating them against primary standards such as piston prover, bell prover, gravimetric system and PVTt standards. The uncertainties of gravimetric systems and PVTt standards are better than piston prover and bell prover. Keeping in view the better uncertainty of PVTt standard, it is being planned to develop PVTt standard at PLI in technical collaboration with advanced MIs like IST, USA and CMS, Taiwan so that the traceability of our gas flow calibration system may be established at in-house. Acknowledgement: The authors are thankful to Dr. A. Sen Gupta, Director, CSIR-ational Physical Laboratory, ew Delhi and Dr. V.. Ojha, Head, Apex Level Standards and Industrial Metrology (ALSIM) division for their constant encouragement. References: 1. JCGM 100, Evaluation of measurement data Guide to the expression of uncertainty in measurement, Fluke User manual for Molbox1+ Terminal and associated literature for molblocs. 3. Michael Bair, Technical ote 2011T06A: Uncertainty analysis for flow measured by molboc-l and molbloc-s mass flow transfer standards, 17 Feb 2003, Revised 18 May D. Kruh, Assessment of uncertainty in calibration of a gas mass flowmeter, Accred Qual Assur, vol. 5, Springer-Verlag, , 2000.

Flow System Calibration in Medical and Bio-medical Production Processes.

Flow System Calibration in Medical and Bio-medical Production Processes. Flow System Calibration in Medical and Bio-medical Production Processes. Speaker / Author: Mark Finch Fluke Calibration Norfolk, Norwich, NR6 6JB, UK e-mail: mark.finch@fluke.com Phone: + 44 (0)1908 678

More information

The Fundamentals of Gas Flow Calibration

The Fundamentals of Gas Flow Calibration The Fundamentals of Gas Flow Calibration Application Note Introduction Understanding the fundamentals of gas flow calibration is essential for evaluating calibration systems, estimating the magnitude of

More information

A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS

A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS Presented 2006 May 16-18 2006 ISFFM Queretaro, Mexico A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS P. Delajoud, M. Bair, C. Rombouts, M. Girard DH Instruments,

More information

Gas Flowmeter Calibrations with the Working Gas Flow Standard

Gas Flowmeter Calibrations with the Working Gas Flow Standard Gas Flowmeter Calibrations with the Working Gas Flow Standard NIST Special Publication 50-80 John D. Wright, Jean-Philippe Kayl, Aaron N. Johnson, and Gina M. Kline November 3, 009 Fluid Metrology Group

More information

RITTER Multiple Sonic Nozzle Calibration System

RITTER Multiple Sonic Nozzle Calibration System RITTER Multiple Sonic Nozzle Calibration System »Wouldn t it be great to eliminate doubtful measurement results by using proven measurement technology? Providing the most precise results ensures and increases

More information

Grant Agreement No. 228296 SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

Grant Agreement No. 228296 SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme Grant Agreement No. 228296 SFERA Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME Capacities Specific Programme Research Infrastructures Integrating Activity - Combination of

More information

What is the "Best" Transfer Standard for Gas Flow?

What is the Best Transfer Standard for Gas Flow? What is the "Best" Transfer Standard for Gas Flow? Flow Systems 220 Bunyan Ave. Berthoud, CO 80513 USA Phone: 800-466-3569 Fax: 970-532-0748 www.flowsystemsinc.com WHAT IS THE BEST TRANSFER STANDARD FOR

More information

CO 2 41.2 MPa (abs) 20 C

CO 2 41.2 MPa (abs) 20 C comp_02 A CO 2 cartridge is used to propel a small rocket cart. Compressed CO 2, stored at a pressure of 41.2 MPa (abs) and a temperature of 20 C, is expanded through a smoothly contoured converging nozzle

More information

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers Flow and Flow rate. Laminar and Turbulent flow Laminar flow: smooth, orderly and regular Mechanical sensors have inertia, which can integrate out small variations due to turbulence Turbulent flow: chaotic

More information

Meriam Instrument LFE Data Sheet

Meriam Instrument LFE Data Sheet 1. AIR FLOW IN SCFM AT STANDARD CONDITIONS OF 29.92 AND 70 F. SCFM = Standard cubic feet per minute. ACFM = Actual cubic feet per minute CFM = Cubic feet per minute. When the flowing gas is air and the

More information

AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory *

AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory * AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory * Paper prepared for the 4 th International Symposium on Fluid Flow

More information

Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article

Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article F e a t u r e A r t i c l e Feature Article Development of a High Accuracy, Fast Response Mass Flow Module Utilizing Pressure Measurement with a Laminar Flow Element (Resistive Element) Criterion D200

More information

191: Calibration and Standards in Flow Measurement. Richard Paton National Engineering Laboratory, Scotland, UK 1 GENERAL PRINCIPLES

191: Calibration and Standards in Flow Measurement. Richard Paton National Engineering Laboratory, Scotland, UK 1 GENERAL PRINCIPLES 191: Calibration and Standards in Measurement Richard Paton National Engineering Laboratory, Scotland, UK 1 General Principles 1 2 Gravimetric Calibration of Liquid meters 2 3 Volumetric Calibration of

More information

Typical Aluminum GFM Mass Flow Meter

Typical Aluminum GFM Mass Flow Meter MASS FLOW METERS Design Features n Rigid metallic construction. n Maximum pressure of 1000 psig (70 bars). n Leak integrity 1 x 10-9 of helium. n NIST traceable certification. n Built-in tiltable LCD readout.

More information

GOOD PRACTICE GUIDE THE CALIBRATION OF FLOW METERS

GOOD PRACTICE GUIDE THE CALIBRATION OF FLOW METERS GOOD PRACTICE GUIDE THE CALIBRATION OF FLOW METERS www.tuvnel.com The calibration of flow meters This good practice guide covers the general principles of calibration when applied to the flow meters or

More information

FLOW MEASUREMENT 2001 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER

FLOW MEASUREMENT 2001 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER FLOW MEASUREMENT 200 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER Dr D G Stewart, NEL Dr M Reader-Harris, NEL Dr R J W Peters, McCrometer Inc INTRODUCTION The V-Cone

More information

T U R B I N E G A S M E T E R

T U R B I N E G A S M E T E R TURBINE GAS METER TURBINE GAS METER CGT 1 2 3 4 5 6 7 Design and function page 2 General technical data page 3 Measurement outputs page 4 Dimensions and weights page 5 Performance page 7 Pressure loss

More information

A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW. 1998 ASME Fluids Engineering Division Summer Meeting

A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW. 1998 ASME Fluids Engineering Division Summer Meeting TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW Proceedings of FEDSM 98: June -5, 998, Washington, DC FEDSM98 49 ABSTRACT The pressure

More information

Precision Mass Flow Metering For CVD Applications.

Precision Mass Flow Metering For CVD Applications. Precision Mass Flow Metering For CVD Applications. Ir. H.J. Boer Research Department of Bronkhorst High-Tech B.V. Nijverheidsstraat 1A, 7261 AK Ruurlo The Netherlands. Tel: +31 (0)573 458800 Fax: +31 (0)573

More information

DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING

DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING Toshiharu Kagawa 1, Yukako Saisu 2, Riki Nishimura 3 and Chongho Youn 4 ABSTRACT In this paper, we developed a new laminar flow

More information

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. ASGMT / Averaging Pitot Tube Flow Measurement Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. Averaging Pitot Tube Meters Introduction

More information

Natural Gas Flow Calibration Service (NGFCS)

Natural Gas Flow Calibration Service (NGFCS) NIST Measurement Services: NIST Special Publication 1081 Natural Gas Flow Calibration Service (NGFCS) Aaron N. Johnson U. S. Department of Commerce Technology Administration National Institute of Standards

More information

Swissmetro travels at high speeds through a tunnel at low pressure. It will therefore undergo friction that can be due to:

Swissmetro travels at high speeds through a tunnel at low pressure. It will therefore undergo friction that can be due to: I. OBJECTIVE OF THE EXPERIMENT. Swissmetro travels at high speeds through a tunnel at low pressure. It will therefore undergo friction that can be due to: 1) Viscosity of gas (cf. "Viscosity of gas" experiment)

More information

User s Manual Laminar Flow Elements

User s Manual Laminar Flow Elements 501:440-13 User s Manual Laminar Flow Elements 10920 Madison Avenue Cleveland, Ohio 44102 Tel: 216.281.1100 email:meriam@meriam.com Web:www.meriam.com 1 TABLE OF CONTENTS Subject Page Introduction... 3

More information

Technical & Sizing Bulletin for Meriam LFEs. Laminar Flow Elements

Technical & Sizing Bulletin for Meriam LFEs. Laminar Flow Elements Technical & Sizing Bulletin for Meriam LFEs Laminar Flow Elements What they are... Meriam Laminar Flow Elements are gas volume rate of flow differential producers operating on capillary flow principles.

More information

FLUID FLOW Introduction General Description

FLUID FLOW Introduction General Description FLUID FLOW Introduction Fluid flow is an important part of many processes, including transporting materials from one point to another, mixing of materials, and chemical reactions. In this experiment, you

More information

Flow Calibration System

Flow Calibration System Gas Flow Standards Flow Calibration System INTRODUCTION... TABLE OF CONTENTS INTRODUCTION 1 GENERAL FEATURES 3 PUTTING TOGETHER 8 A molbloc/molbox FLOW CALIBRATION SYSTEM molbox FLOW 10 TERMINALS molbloc

More information

Gas Custody Transfer Calibration

Gas Custody Transfer Calibration Gas Custody Transfer Calibration Using multi variable temperature / pressure calibrators for flowmeter calibration 2013 Introduction Gas custody transfer flow computers require special calibration to perform

More information

Valve Sizing. Te chnic al Bulletin. Flow Calculation Principles. Scope. Sizing Valves. Safe Product Selection. www.swagelok.com

Valve Sizing. Te chnic al Bulletin. Flow Calculation Principles. Scope. Sizing Valves. Safe Product Selection. www.swagelok.com www.swagelok.com Valve Sizing Te chnic al Bulletin Scope Valve size often is described by the nominal size of the end connections, but a more important measure is the flow that the valve can provide. And

More information

Calibrating Differential Pressure Sensors

Calibrating Differential Pressure Sensors Calibrating Differential Pressure Sensors Additel Corporation 22865 Savi Ranch Parkway, Ste F Yorba Linda, CA 92887 Introduction Differential pressure sensors are common in the process industry and cover

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

MODEL TEG-F20. Specification Sheet. TEG-F20 Specification Sheet Ver. 2011-0310. Copyright(C) 2011 Tsukasa Sokken Co., Ltd.

MODEL TEG-F20. Specification Sheet. TEG-F20 Specification Sheet Ver. 2011-0310. Copyright(C) 2011 Tsukasa Sokken Co., Ltd. Direct EGR Flow Meter MODEL TEG-F20 Specification Sheet TEG-F20 Specification Sheet Ver. 2011-0310 Copyright(C) 2011 Tsukasa Sokken Co., Ltd. 1. Overview Sokken model TEG series is an EGR Flow meter to

More information

Custody Transfer Measurement. with the V-Cone Flowmeter

Custody Transfer Measurement. with the V-Cone Flowmeter Custody Transfer Measurement with the V-Cone Flowmeter Stephen A. Ifft McCrometer Inc. Hemet, California, USA Abstract This paper will discuss the approval of the McCrometer V-Cone flowmeter for custody

More information

In semiconductor applications, the required mass flows

In semiconductor applications, the required mass flows Model-Based Multi-GasMulti-Range Mass Flow Controllers With Single Gas Calibration and Tuning JOOST LÖTTERS A model has been developed that accurately predicts the behavior of the flow sensor combined

More information

Chapter 10. Flow Rate. Flow Rate. Flow Measurements. The velocity of the flow is described at any

Chapter 10. Flow Rate. Flow Rate. Flow Measurements. The velocity of the flow is described at any Chapter 10 Flow Measurements Material from Theory and Design for Mechanical Measurements; Figliola, Third Edition Flow Rate Flow rate can be expressed in terms of volume flow rate (volume/time) or mass

More information

TELEDYNE HASTINGS INSTRUMENTS

TELEDYNE HASTINGS INSTRUMENTS TELEDYNE HASTINGS INSTRUMENTS MODELS Model HFM-200 with LFE (Laminar Flow Element) FEATURES Range 25 to 15,000 slm (N 2 Equivalent) Accuracy in Nitrogen ±1% of Full scale Modular Construction NIST Traceable

More information

Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF

Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF Table of Contents Section 1 - General Introduction 1.1 1 Principles Of Operation 1.2 1 Reshaping The Velocity Profile

More information

Sheet 5:Chapter 5 5 1C Name four physical quantities that are conserved and two quantities that are not conserved during a process.

Sheet 5:Chapter 5 5 1C Name four physical quantities that are conserved and two quantities that are not conserved during a process. Thermo 1 (MEP 261) Thermodynamics An Engineering Approach Yunus A. Cengel & Michael A. Boles 7 th Edition, McGraw-Hill Companies, ISBN-978-0-07-352932-5, 2008 Sheet 5:Chapter 5 5 1C Name four physical

More information

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Dr. J. M. Meyers Dr. D. G. Fletcher Dr. Y. Dubief 1. Introduction Last lab you investigated flow loss in a pipe due to the roughness

More information

Advances in Thermal Dispersion Mass Flow Meter Accuracy

Advances in Thermal Dispersion Mass Flow Meter Accuracy Advances in Thermal Dispersion Mass Flow Meter Accuracy By Dan McQueen President Fluid Components International (FCI) Advances in Thermal Dispersion Mass Flow Meter Accuracy By Dan McQueen President Fluid

More information

Experiment # 3: Pipe Flow

Experiment # 3: Pipe Flow ME 05 Mechanical Engineering Lab Page ME 05 Mechanical Engineering Laboratory Spring Quarter 00 Experiment # 3: Pipe Flow Objectives: a) Calibrate a pressure transducer and two different flowmeters (paddlewheel

More information

ME 538: Experiment # 1 Flow measurement and particle generation instruments

ME 538: Experiment # 1 Flow measurement and particle generation instruments ME 538: Experiment # 1 Flow measurement and particle generation instruments Suresh Dhaniyala Mechanical and Aeronautical Engineering Clarkson University Flow rate measurement Various types of flowmeters

More information

BASIC UNDERSTANDING OF FLOW CALCULATIONS AND ESTIMATES MAKES SIZING VALVES SIMPLER

BASIC UNDERSTANDING OF FLOW CALCULATIONS AND ESTIMATES MAKES SIZING VALVES SIMPLER BASIC UNDERSTANDING OF FLOW CALCULATIONS AND ESTIMATES MAKES SIZING VALVES SIMPLER Valve size often is described by the nominal size of the end connections but a more important measure is the flow that

More information

Experimental Evaluation of the Discharge Coefficient of a Centre-Pivot Roof Window

Experimental Evaluation of the Discharge Coefficient of a Centre-Pivot Roof Window Experimental Evaluation of the Discharge Coefficient of a Centre-Pivot Roof Window Ahsan Iqbal #1, Alireza Afshari #2, Per Heiselberg *3, Anders Høj **4 # Energy and Environment, Danish Building Research

More information

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control Solutions. Benefits.

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control Solutions. Benefits. GE Measurement & Control Solutions Sentinel LCT Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement Benefits Extremely reliable and highly accurate flowmeter, designed specifically

More information

Testing methods applicable to refrigeration components and systems

Testing methods applicable to refrigeration components and systems Testing methods applicable to refrigeration components and systems Sylvain Quoilin (1)*, Cristian Cuevas (2), Vladut Teodorese (1), Vincent Lemort (1), Jules Hannay (1) and Jean Lebrun (1) (1) University

More information

Source Sampling Manual Volume II

Source Sampling Manual Volume II Source Sampling Manual Volume II December, 1980 Revisions: May, 1981 January, 1992 April, 2015 Operations Division 811 SW 6 th Avenue Portland, OR 97204 Phone: 503-229-5696 800-452-4011 Fax: 503-229-5850

More information

NIST Calibration Services for Liquid Volume NIST Special Publication 250-72

NIST Calibration Services for Liquid Volume NIST Special Publication 250-72 NIST Calibration Services for Liquid olume NIST Special Publication 250-72 ern E. Bean, Pedro I. Espina, John. D. Wright, John F. Houser, Sherry D. Sheckels, and Aaron N. Johnson March 24, 2006 Fluid Metrology

More information

PORTER MASS FLOW MEASUREMENT AND CONTROL SYSTEMS

PORTER MASS FLOW MEASUREMENT AND CONTROL SYSTEMS PORTER MASS MEASUREMENT AND CONTROL SYSTEMS PORTER: THE STANDARD OF EXCELLENCE IN MASS Porter Mass Flow products reflect nearly three decades of experience in the design and manufacture of precision instruments

More information

KALIBRERINGSBEVIS. 25 temperature loggers Ecoguard model TS-020000, identification se table of results.

KALIBRERINGSBEVIS. 25 temperature loggers Ecoguard model TS-020000, identification se table of results. utfärdat av riksmätplats 01 issued by a Swedish National Laboratory Contact person Lars-Erik Josefson 2014-03-27 1 (3) MTv4F006969-K01 Measurement Technology +46 10 516 54 56 larserik.josefson@sp.se EcoGuard

More information

Flow Measurement in Pipes and Ducts. Flow Measurement in Pipes and Ducts, Course #503. Presented by:

Flow Measurement in Pipes and Ducts. Flow Measurement in Pipes and Ducts, Course #503. Presented by: Flow Measurement in Pipes and Ducts, Course #503 Presented by: PDH Enterprises, LLC PO Box 942 Morrisville, NC 27560 www.pdhsite.com This course is about measurement of the flow rate of a fluid flowing

More information

Refrigeration and Airconditioning Prof. M. Ramgopal Department of Mechanical Engineering Indian Institute of Technology, Kharagpur

Refrigeration and Airconditioning Prof. M. Ramgopal Department of Mechanical Engineering Indian Institute of Technology, Kharagpur Refrigeration and Airconditioning Prof. M. Ramgopal Department of Mechanical Engineering Indian Institute of Technology, Kharagpur Lecture No. # 22 Refrigeration System Components: Compressor (Continued)

More information

The Unique Accelabar Flow Meter

The Unique Accelabar Flow Meter The Unique Accelabar Flow Meter The Accelabar is a new and unique flow meter that combines two differential pressure technologies to produce operating ranges never before attainable in a single flow meter.

More information

Barometric Effects on Transducer Data and Groundwater Levels in Monitoring Wells D.A. Wardwell, October 2007

Barometric Effects on Transducer Data and Groundwater Levels in Monitoring Wells D.A. Wardwell, October 2007 Barometric Effects on Transducer Data and Groundwater Levels in Monitoring Wells D.A. Wardwell, October 2007 Barometric Effects on Transducer Data Barometric Fluctuations can Severely Alter Water Level

More information

Carbon Dioxide and an Argon + Nitrogen Mixture. Measurement of C p /C v for Argon, Nitrogen, Stephen Lucas 05/11/10

Carbon Dioxide and an Argon + Nitrogen Mixture. Measurement of C p /C v for Argon, Nitrogen, Stephen Lucas 05/11/10 Carbon Dioxide and an Argon + Nitrogen Mixture Measurement of C p /C v for Argon, Nitrogen, Stephen Lucas 05/11/10 Measurement of C p /C v for Argon, Nitrogen, Carbon Dioxide and an Argon + Nitrogen Mixture

More information

A MEASUREMENT ASSURANCE PROGRAM FOR CALIBRATION BY THE TRANSFER METHOD CS800

A MEASUREMENT ASSURANCE PROGRAM FOR CALIBRATION BY THE TRANSFER METHOD CS800 A EAUREENT AURANCE PROGRA FOR CALIBRATION BY THE TRANFER ETHOD C800 JEFF PEERY DEC. 6, 011 This document outlines the easurement Assurance Program (AP for eaetrics C800 flo measurement system a liquid

More information

Fluid Flow Instrumentation

Fluid Flow Instrumentation Fluid Flow Instrumentation In the physical world, mechanical engineers are frequently required to monitor or control the flow of various fluids through pipes, ducts and assorted vessels. This fluid can

More information

International Journal of Latest Research in Science and Technology Volume 4, Issue 2: Page No.161-166, March-April 2015

International Journal of Latest Research in Science and Technology Volume 4, Issue 2: Page No.161-166, March-April 2015 International Journal of Latest Research in Science and Technology Volume 4, Issue 2: Page No.161-166, March-April 2015 http://www.mnkjournals.com/ijlrst.htm ISSN (Online):2278-5299 EXPERIMENTAL STUDY

More information

STACK EMISSIONS MONITORING REPORT

STACK EMISSIONS MONITORING REPORT STACK EMISSIONS MONITORING REPORT Unit 5 Crown Industrial Estate Kenwood Road Stockport SK5 6PH Tel: 0161 443 0980 Fax: 0161 443 0989 Your contact at ESG Andrew Palliser Business Manager Tel: 0161 443

More information

The Viscosity of Fluids

The Viscosity of Fluids Experiment #11 The Viscosity of Fluids References: 1. Your first year physics textbook. 2. D. Tabor, Gases, Liquids and Solids: and Other States of Matter (Cambridge Press, 1991). 3. J.R. Van Wazer et

More information

PORTABLE FLOW METER KITS TDS-100H

PORTABLE FLOW METER KITS TDS-100H Transit Time Flow Meters TT-TDS-100H-HS ECEFast offers amazing prices on Transit Time Flow Meter portable kits. We are buying from the manufacturer who supplies 60% of the Chinese market and the majority

More information

Fundamentals of Mass Flow Control

Fundamentals of Mass Flow Control Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of

More information

Millikan Oil Drop Experiment Matthew Norton, Jurasits Christopher, Heyduck William, Nick Chumbley. Norton 0

Millikan Oil Drop Experiment Matthew Norton, Jurasits Christopher, Heyduck William, Nick Chumbley. Norton 0 Millikan Oil Drop Experiment Matthew Norton, Jurasits Christopher, Heyduck William, Nick Chumbley Norton 0 Norton 1 Abstract The charge of an electron can be experimentally measured by observing an oil

More information

Fiscal Measurement Natural Gas

Fiscal Measurement Natural Gas White Paper FSG-WP-0012 February 2014 Fiscal Measurement Natural Gas Fiscal Measurement How much? Fiscal Measurement must not be confused with Custody Transfer; in fact, fiscal measurement is a more general

More information

A basic introduction to steam

A basic introduction to steam A basic introduction to steam FOR HOT, COLD, MOIST AND DRY, FOUR CHAMPIONS FIERCE. STRIVE HERE FOR MASTERY Milton 1666 Steam Wonderful Steam Very high heat content Recyclable Clean, non toxic Biodegradable

More information

Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD)

Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD) Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD) Ulrich Breuel 1, Barbara Werner 2, Petra Spitzer 3, Hans D. Jensen 4 1,2 Zentrum für Messen und

More information

Experiment 3 Pipe Friction

Experiment 3 Pipe Friction EML 316L Experiment 3 Pipe Friction Laboratory Manual Mechanical and Materials Engineering Department College of Engineering FLORIDA INTERNATIONAL UNIVERSITY Nomenclature Symbol Description Unit A cross-sectional

More information

Mass Flow Meters & Controllers

Mass Flow Meters & Controllers HFM-D-3A & 305A Mass Flow Meters HFC-D-3A & 307A Mass Flow Controllers FEATURES Range 0-25 slm to 0-2500 slm (N 2 Equivalent) Excellent Accuracy ±(0.5% of Reading + 0.2% of Full Scale) 0-5 VDC or 4-20

More information

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features P3000 Series Pneumatic Deadweight Testers P3000 Technical Data Features Pressure ranges from Vacuum through 2 000 psi (140 bar) Accuracy better than 0.015 % of reading. (Increased accuracy option of 0.008

More information

MATLAB AS A PROTOTYPING TOOL FOR HYDRONIC NETWORKS BALANCING

MATLAB AS A PROTOTYPING TOOL FOR HYDRONIC NETWORKS BALANCING MATLAB AS A PROTOTYPING TOOL FOR HYDRONIC NETWORKS BALANCING J. Pekař, P. Trnka, V. Havlena* Abstract The objective of this note is to describe the prototyping stage of development of a system that is

More information

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions The Versatile Differential Pressure Transmitter By David Gunn Honeywell Process Solutions The Versatile Differential Pressure Transmitter 2 Table of Contents Abstract... 3 Pressure Fundamentals... 3 Applications...

More information

CE 204 FLUID MECHANICS

CE 204 FLUID MECHANICS CE 204 FLUID MECHANICS Onur AKAY Assistant Professor Okan University Department of Civil Engineering Akfırat Campus 34959 Tuzla-Istanbul/TURKEY Phone: +90-216-677-1630 ext.1974 Fax: +90-216-677-1486 E-mail:

More information

Enhancement of heat transfer of solar air heater roughened with circular transverse RIB

Enhancement of heat transfer of solar air heater roughened with circular transverse RIB Enhancement of heat transfer of solar air heater roughened with circular transverse RIB Gurpreet Singh 1, Dr. G. S. Sidhu 2 Lala Lajpat Rai Institute of Engineering and Technology, Moga Punjab, India 1,2

More information

PRODUCT REVIEW. Flowmeters and Controllers. Vacuum Instrumentation

PRODUCT REVIEW. Flowmeters and Controllers. Vacuum Instrumentation PRODUCT REVIEW Flowmeters and Controllers Vacuum Instrumentation TELEDYNE HASTINGS VACUUM Instrumentation Thermocouple Gauges ANALOG METERS AND CONTROLLERS - VT Series Three Standard Ranges: 0 to 100 mtorr,

More information

IN LINE FLOW-METER CALIBRATION ADVANCE METHODS AND SOLUTIONS

IN LINE FLOW-METER CALIBRATION ADVANCE METHODS AND SOLUTIONS ATC, Inc. 4037 GUION LANE INDIANPOLIS, IN 46268 PHONE: (317) 328-8492 FAX: (317) 328-2686 www.atcinc.net IN LINE FLOW-METER CALIBRATION ADVANCE METHODS AND SOLUTIONS AUTHORS: Hemi Sagi William G. Roberts

More information

Calculation of flow rate from differential pressure devices orifice plates

Calculation of flow rate from differential pressure devices orifice plates Calculation of flow rate from differential pressure devices orifice plates Peter Lau EMATEM - Sommerschule Kloster Seeon August -, 008 Contents Introduction - Why using dp-devices when other principles

More information

The PACE5000 Modular Pressure Controller/Indicator

The PACE5000 Modular Pressure Controller/Indicator GE Sensing The PACE5000 Modular Pressure Controller/Indicator First in a new generation of modular, high precision Druck pressure controllers/indicators, designed for ATE and test bench applications. The

More information

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors Tom Wise on behalf of Daya Bay collaboration Day Bay is a reactor based experiment dedicated to measurement of the last

More information

Ravi Kumar Singh*, K. B. Sahu**, Thakur Debasis Mishra***

Ravi Kumar Singh*, K. B. Sahu**, Thakur Debasis Mishra*** Ravi Kumar Singh, K. B. Sahu, Thakur Debasis Mishra / International Journal of Engineering Research and Applications (IJERA) ISSN: 48-96 www.ijera.com Vol. 3, Issue 3, May-Jun 3, pp.766-77 Analysis of

More information

HOW TO SIZE A PRESSURE REGULATOR

HOW TO SIZE A PRESSURE REGULATOR CHOOSING THE CORRECT REGULATOR HOW TO SIZE A PRESSURE REGULATOR In order to choose the correct regulator the following information is needed: - Function: Pressure reducing or backpressure control? - Pressure:

More information

HIGH PRESSURE TECHNOLOGY HYDRAULICS PNEUMATICS TESTING EQUIPMENT

HIGH PRESSURE TECHNOLOGY HYDRAULICS PNEUMATICS TESTING EQUIPMENT HIGH PRESSURE TECHNOLOGY HYDRAULICS PNEUMATICS TESTING EQUIPMENT MAXIMATOR GmbH 2 MAXIMATOR has an extensive know-how concerning concept, development, construction and manufacturing of test benches and

More information

Correcting a Coriolis Meter for Two Phase Oil & Gas Flow

Correcting a Coriolis Meter for Two Phase Oil & Gas Flow Correcting a Coriolis Meter for Two Phase Oil & Gas Flow Chris Mills, NEL 1 INTRODUCTION In the oil & gas industry, the requirement for flow measurement of two phase oil and gas flow is prevalent. This

More information

Micro Motion 3098 Gas Specific Gravity Meter

Micro Motion 3098 Gas Specific Gravity Meter Product Data Sheet PS-001161, Rev. D April 2013 Micro Motion 3098 Gas Specific Gravity Meter Micro Motion density and concentration meters are built to tackle the most demanding process and fiscal applications.

More information

Model 111 Mass Flow Instruments

Model 111 Mass Flow Instruments 111 Mass Instruments Porter Mass products reflect over four decades of experience in the design and manufacture of precision instruments for the measurement and control of gas flow. They incorporate design

More information

Theory overview of flow measurement using differential pressure devices based on ISO-5167 standard.

Theory overview of flow measurement using differential pressure devices based on ISO-5167 standard. Theory overview of flow measurement using differential pressure devices based on ISO-567 standard. rian FL40 flow computer description. Flow Cad software users manual. Technical note, Differential pressure

More information

www.kem-kueppers.com info@kem-kueppers.com

www.kem-kueppers.com info@kem-kueppers.com www.kem-kueppers.com info@kem-kueppers.com Calibration of flow meters according to DIN EN ISO/IEC 17025:2005 Introduction... 2 Calibration laboratory... 3 Description of the calibration process... 3 Calibration

More information

1. Fluids Mechanics and Fluid Properties. 1.1 Objectives of this section. 1.2 Fluids

1. Fluids Mechanics and Fluid Properties. 1.1 Objectives of this section. 1.2 Fluids 1. Fluids Mechanics and Fluid Properties What is fluid mechanics? As its name suggests it is the branch of applied mechanics concerned with the statics and dynamics of fluids - both liquids and gases.

More information

APPENDIX A CONTROL VALVE TESTING PROCEDURES AND EQUATIONS FOR LIQUID FLOWS

APPENDIX A CONTROL VALVE TESTING PROCEDURES AND EQUATIONS FOR LIQUID FLOWS APPENDIX A CONTROL VALVE TESTING PROCEDURES AND EQUATIONS FOR LIQUID FLOWS Section A.1. Flow Coefficients Definition The flow coefficient or pressure loss coefficient is used to relate the pressure loss

More information

= 800 kg/m 3 (note that old units cancel out) 4.184 J 1000 g = 4184 J/kg o C

= 800 kg/m 3 (note that old units cancel out) 4.184 J 1000 g = 4184 J/kg o C Units and Dimensions Basic properties such as length, mass, time and temperature that can be measured are called dimensions. Any quantity that can be measured has a value and a unit associated with it.

More information

Turbine Flowmeters for Gas Applications

Turbine Flowmeters for Gas Applications Turbine Flowmeters for Gas Applications Description Flow Technology s FT Series turbine flowmeters utilize a proven flow measurement technology to provide exceptionally reliable digital outputs. Because

More information

8. ENERGY PERFORMANCE ASSESSMENT OF COMPRESSORS 8.1 Introduction

8. ENERGY PERFORMANCE ASSESSMENT OF COMPRESSORS 8.1 Introduction 8. ENERGY PERFORMANCE ASSESSMENT OF COMPRESSORS 8.1 Introduction The compressed air system is not only an energy intensive utility but also one o the least energy eicient. Over a period o time, both perormance

More information

2 1/2 Pipe. 40 = height. the gauge pressure inside the vessel from the gauge pressure at the nozzle inlet as shown:

2 1/2 Pipe. 40 = height. the gauge pressure inside the vessel from the gauge pressure at the nozzle inlet as shown: 116eering. Engineering. Engineering. Engineering. Engineerin Engineering Information SPECIFYING SPRAY NOZZLES Spray nozzles have three basic functions: meter flow distribute liquid break up a liquid stream

More information

Druck DPI 142/150. GE Infrastructure Sensing. Pressure Indicator Calibration manual - K382

Druck DPI 142/150. GE Infrastructure Sensing. Pressure Indicator Calibration manual - K382 GE Infrastructure Sensing 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Druck DPI 142/150 Pressure Indicator Calibration manual - K382 Introduction This technical manual

More information

ngenieros Ingenieros Asociados De Control S.L.

ngenieros Ingenieros Asociados De Control S.L. ngenieros sociados de ontrol 1 Company founded in 1985 to serve commercial and technical assistance in the fields of instrumentation, control and installation for all types of industries. We offer top

More information

XI / PHYSICS FLUIDS IN MOTION 11/PA

XI / PHYSICS FLUIDS IN MOTION 11/PA Viscosity It is the property of a liquid due to which it flows in the form of layers and each layer opposes the motion of its adjacent layer. Cause of viscosity Consider two neighboring liquid layers A

More information

Turbulence Modeling in CFD Simulation of Intake Manifold for a 4 Cylinder Engine

Turbulence Modeling in CFD Simulation of Intake Manifold for a 4 Cylinder Engine HEFAT2012 9 th International Conference on Heat Transfer, Fluid Mechanics and Thermodynamics 16 18 July 2012 Malta Turbulence Modeling in CFD Simulation of Intake Manifold for a 4 Cylinder Engine Dr MK

More information

COUNTERBALANCE VALVES

COUNTERBALANCE VALVES COUNTERBALANCE VALVES Introduction They are modulating valves which allow free flow into the actuator and then block the reverse flow until they feel a pilot pressure inversely proportional to the load

More information

WEEK 1. Engineering Calculations Processes Process Variables

WEEK 1. Engineering Calculations Processes Process Variables WEEK 1 Engineering Calculations Processes Process Variables 2.1 Units and Dimensions Units and dimensions are important in science and engineering A measured quantity has a numerical value and a unit (ex:

More information

Energy and Flow Measurement for Hydronic Systems

Energy and Flow Measurement for Hydronic Systems Energy and Flow Measurement for Hydronic Systems Presented By: George Szyszko Applications Consultant MCR for ONICON Incorporated Why Measure Thermal Energy? To control something, you must first measure

More information