1 Laboratory Performance in Neonatal Bilirubin Testing Using Commutable Specimens A Progress Report on a College of American Pathologists Study Stanley F. Lo, PhD; Bernadine Jendrzejczak, BA; Basil T. Doumas, PhD Context. In 2003 the Chemistry Resource Committee of the College of American Pathologists introduced a commutable specimen in the Neonatal Bilirubin Surveys. This specimen was intended to help evaluate all bilirubin methods. Objective. To evaluate the effect of commutable specimens on the performance of selected clinical analyzers in measuring neonatal bilirubin from 2003 through Design. A human serum based specimen enriched with unconjugated bilirubin in human serum has been included since 2003 in the Neonatal Bilirubin Surveys. The bilirubin values of these specimens were determined by the reference method and used to evaluate results reported by various chemistry analyzers. Results. Coefficients of variation for College of American Pathologists All Data ranged from 4.9% to 6.2% for the Neonatal Bilirubin Survey. However, coefficients of variation for the 4 major instrument groups (Dimension, Olympus, Synchron, and Vitros), which report 65% of all results, varied from 2% to 3%. College of American Pathologists All Data mean bilirubin values were within 0.46 mg/dl (7.8 mol/l) of the reference method mean in 2003; in subsequent years these differences became larger, peaking at 1.87 mg/dl (32 mol/l) in Conclusions. The large systematic error of bilirubin measurements is due primarily to failure of instrument manufacturers to produce reliable bilirubin calibrators. Primary calibrators should consist of human serum enriched with unconjugated bilirubin. Bilirubin values must be assigned by the reference method, the performance and robustness of which are reported in this article. Secondary calibrators distributed to users must be traceable to primary calibrators. (Arch Pathol Lab Med. 2008;132: ) In 2003, the Chemistry Resource Committee of the College of American Pathologists (CAP), acting on a report regarding inaccuracies in the measurement of bilirubin in clinical laboratories, 1 introduced a new specimen in the Neonatal Bilirubin Surveys (NB-Surveys). This specimen, human serum enriched with unconjugated bilirubin, was expected to be commutable (free of matrix effects) because it closely resembles specimens drawn from healthy neonates. Properly calibrated clinical analyzers were expected to be accurate, providing total bilirubin (TBIL) values close to those obtained by the reference method. The noncommutable or conventional NB-Survey specimens consisted of unconjugated bilirubin and ditaurobilirubin in bovine serum. 1 A guideline issued by the American Academy of Pediatrics recommended a blood exchange transfusion for infants older than 48 hours when bilirubin levels near 25 Accepted for publication April 2, From the Reference Standards Laboratory, Department of Pathology and Laboratory Medicine, Children s Hospital of Wisconsin, Milwaukee (Drs Lo and Doumas and Ms Jendrzejczak); and the Department of Pathology, Medical College of Wisconsin, Milwaukee (Drs Lo and Doumas). The authors have no relevant financial interest in the products or companies described in this article. Reprints: Stanley F. Lo, PhD, Reference Standards Laboratory, Children s Hospital of Wisconsin, 9000 W Wisconsin Ave, Wauwatosa, WI ( slo@mcw.edu). mg/dl (428 mol/l) cannot be reduced by phototherapy. 2 Because this is considered a critical bilirubin value, the CAP Chemistry Resource Committee decided to adjust the bilirubin level in the human serum based specimen near 25 mg/dl (428 mol/l). Results from the 2003 human serum based specimen (NB-02) were very encouraging. The grand mean value for CAP All Data was only 2.4% higher than the reference method mean and the coefficient of variation (CV) % of 5.7. This communication is a progress report on the performance of laboratories participating in the NB-Surveys for the years 2004 through The practice of including the special, commutable NB-Survey specimen in the Chemistry Survey (C-Survey), as specimen C-96 or C-97, continued until the end of year MATERIALS AND METHODS The preparation and distribution to participating laboratories of the human serum based specimens has been described previously. 3 All NB-Survey and C-Survey specimens (from 2003 through 2006) were analyzed for TBIL by the reference method 4 at the Reference Standards Laboratory of the Children s Hospital of Wisconsin, Milwaukee. All other bilirubin results shown in this article have been reported in the participant summary report. Instruments specifically considered in this report include the BR2 (Advanced Instruments Inc, Norwood, Mass), Advia (Siemens Medical Solutions Diagnostics, Tarrytown, NY), Aeroset and Architect (Abbott Diagnostics, Abbott Park, Ill), Cobas Integra, Arch Pathol Lab Med Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing Lo et al 1781

2 Table 1. College of American Pathologists (CAP) All Data and Reference Method Mean Bilirubin Values and Coefficient of Variation (CV)% for the Neonatal Bilirubin Survey and Chemistry Survey Specimens Specimen CAP All Data Mean Bilirubin Value, mg/dl CV% Reference Method Value, No. of mg/dl Laboratories 2003 NB C NB C NB C NB C NB Roche Modular/Hitachi (Roche Diagnostic Systems, Branchburg, NJ), Dimension (Dade Behring, Newark, Del), Olympus (Olympus America Inc, Melville, NY), Synchron (Beckman Coulter Inc, Fullerton, Calif), Unistat (Leica Microsystems Inc, Buffalo, NY), and Vitros (Ortho-Clinical Diagnostics, Raritan, NJ). According to the participant summary report, methods based on the coupling of bilirubin with a diazo compound are used in the Advia, Aeroset, Architect, Cobas Integra, Dimension, Roche Modular/Hitachi, Olympus, Synchron, and Vitros (TBIL slide). Methods based on the oxidation of bilirubin are used in the Advia (vanadate oxidation) and Synchron (bilirubin oxidase). Methods based on direct spectrophotometry are used in the Aeroset, Architect, Advanced BR2, Unistat, and Vitros neonatal bilirubin (BuBc) slide, which measures the sum of unconjugated bilirubin and conjugated bilirubins (bilirubin monoglucuronide and diglucuronide). According to Abbott, both the Aeroset and Architect methods use a diazo method (diazotized 2,4-dichloroaniline), but for the NB-Survey users prefer direct spectrophotometry. RESULTS Table 1 shows CAP All Data mean bilirubin values and CV% for the NB-Survey and C-Survey specimens consisting of unconjugated bilirubin in pooled human sera sent to participating laboratories during ; also shown are values obtained by the reference method. For each year from 2003 through 2006, one commutable sample was prepared and sent out as a specimen in the NB-Survey and C-Survey; thus, in 2003, NB-02 was identical to C-96; in 2004, NB-06 was identical to C-97; in 2005, NB-05 was identical to C-96; and in 2006, NB-01, C-97, and NB-11 were identical. Shown in Figure 1 are differences between mean values of CAP All Data and those of the reference method for NB-Survey and C-Survey specimens The best overall laboratory performance was achieved in The CAP All Data mean value for NB- 02 was identical to that for C-96 and the difference, 0.46 mg/dl (7.9 mol/l) or 2.4%, between the CAP All Data mean and that of the reference method was the smallest in 4 years. 3 In 2006, the commutable specimen was mailed to participants on 3 occasions, in April and November with the NB-Survey (specimens NB-01 and NB-11, respectively) and in June with the C-Survey (specimen C-97). The unusually large difference, 0.64 mg/dl (10.9 mol/l), be- Figure 1. College of American Pathologists (CAP) Neonatal Bilirubin and Chemistry Surveys Differences between CAP All Data and reference method mean values in milligrams per deciliter (1 mg/ dl bilirubin 17.1 mol/l bilirubin). tween the NB-01 and NB-11 CAP All Data bilirubin mean values (Figure 2) is due to large shifts in the mean values reported by the Cobas Integra, Dimension, Roche Modular/Hitachi, and Vitros, which constitute 60% of the participating laboratories. Discrepancies between NB-01 and C-97 (Figure 2) are primarily due to the large percentage of Vitros participants in the C-Survey. Figure 3 shows differences between bilirubin mean values for selected field methods and the reference method. With 3 exceptions, Advanced BR2, Aeroset ( ), and Architect (2006), mean bilirubin values for all other instrument groups are higher than those of the reference method. Note that there were no data from the Abbott instruments in 2005, and both Abbott instruments use the same diazo and direct spectrophotometry methods. The positive bias ranges from less than 1 mg/dl to 4 mg/dl (17.1 to 68.4 mol/l). The Synchron group consistently shows a small positive bias. The negative bias of the BR2 has decreased over time and is smallest in Reference methods are considered the gold standard against which field methods are compared and evaluated. It is important, therefore, to establish the long-term performance characteristics of the reference method for serum TBIL. From September 2003 to April 2007, the Reference Laboratory of the Children s Hospital of Wisconsin prepared 11 bilirubin standard solutions by adding National Institute of Standards and Technology Bilirubin, Standard Reference Material 916, to pooled human sera to a concentration of 20 mg/dl (342 mol/l). Following preparation, these solutions were analyzed by the reference method and the molar absorptivity ( ) of the alkaline azopigment was compared with the values established in 2 round robin studies with participation from national and international laboratories. 5,6 Collaborating laboratories in the 1988 round robin study were from the United States (3) and The Netherlands (2); in the 1998 round robin study, laboratories were from the United States (3), Germany (2), and The Netherlands (1). Data from the 2 round robins and from the Children s Hospital of Wisconsin Reference Laboratory, shown in Table 2, indicate that the established molar absorptivity of the alkaline azopigment, because of its reproducibility and narrow range, can and should be used to assess the accuracy of bilirubin calibrators consisting of unconjugated bilirubin in human sera. The 20-mg/dL (342- mol/l) stock solutions are dis Arch Pathol Lab Med Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing Lo et al

3 Figure 2. College of American Pathologists Neonatal Bilirubin Surveys Shown are differences between total bilirubin mean values (milligrams per deciliter) for specimens NB-01 and NB-11 (yellow bars) and between specimens NB-01 and C-97 (red bars) (1 mg/dl bilirubin 17.1 mol/l bilirubin). Figure 3. College of American Pathologists (CAP) Neonatal Bilirubin Surveys Shown are differences between total bilirubin mean values (milligrams per deciliter) for selected field and reference methods. The numbers on the horizontal axis identify different instruments: 1 indicates CAP All Data; 2, Advanced BR2; 3, Advia; 4, Aeroset (2003 and 2004)/Architect (2006); 5, Cobas Integra; 6, Cobas Integra, DZ Salt; 7, Dimension; 8, Hitachi, Roche Modular; 9, Olympus; 10, Synchron Diazo; 11, Synchron BO; 12, Unistat; 13, Vitros 250/350; 14, Vitros 950; and 15, Vitros 5,1 FS (1 mg/dl bilirubin 17.1 mol/l bilirubin). pensed in airtight plastic tubes and kept at 70 C until use. Each time the reference method is used, the stock solution is analyzed to verify that its value is within the accepted limits (mean value 3 SD). Next, the stock solution and 3 dilutions (2.5 mg/dl [42.7 mol/l], 5.0 mg/ dl [85.5 mol/l], and 10 mg/dl [171 mol/l]) are analyzed to construct a calibration curve, the regression equation of which is used to calculate the results of the specimens analyzed. Shown in Table 3 are molar absorptivities, slopes, and intercepts of calibration curves ob- Arch Pathol Lab Med Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing Lo et al 1783

4 Table 2. Molar Absorptivity of Bilirubin Alkaline Azopigment Results From 2 Round Robins and the Reference Laboratory of Children s Hospital of Wisconsin (CHW)* n (L mol 1 cm 1 ) SD Round Robin 1988 (5 labs) Round Robin 1998 (6 labs) Reference Lab CHW * n indicates number of determinations;, molar absorptivity at 598 nm. Corrected for purity (98.3%) of bilirubin. Table 3. Long-term Calibration Stability and Precision of the Reference Method for Total Bilirubin Calibration Curve (September 15, 2003 April 23, 2007) (L mol 1 cm 1 )* Slope Intercept Mean SD N Min Max * indicates molar absorptivity. One standard deviation for a single measurement. tained during a 3.5-year period, and precision data on controls prepared in our laboratory. The data in Tables 2 and 3 demonstrate the impressive stability and robustness of the reference method in US and European laboratories. In 2003, the CAP All Data mean values for NB-02 and C-96 were identical and within 2.4% of the reference value (Figure 1). In our opinion this was due to the use of calibrators with accurately assigned bilirubin values. In subsequent years accuracy deteriorated. Results for 2006 are strange; there is good agreement for CAP All Data means between specimens C-97 and NB-11, but the mean value for NB-01 is greater than NB-11 by 0.64 mg/dl (10.9 mol/l). The lower mean bilirubin value for NB-11 was due to the downward shift of 4 instrument group mean values (Cobas Integra, Dimension, Hitachi/Roche Modular, and Vitros, which represent 59% of the reported results) (Figure 2). The close agreement between the mean values for C-97 and NB-11, mg/dl (383.7 mol/l) and mg/dl (381.7 mol/l), indicates a small, if any, loss of bilirubin, in the 6-month period (June November). The most logical explanation for the difference between NB-01 and NB-11 would be a change in calibrators or reassignment of bilirubin values in the calibrators used by these instruments. Consistent with this explanation, Dade Behring and Roche Diagnostics were among the diagnostic companies that lowered the assigned values to their bilirubin calibrators (Diane Stille, Dade Behring Inc, Newark, Del, written communication, May 7, 2006, and Melanie Swartzentuber, Roche Diagnostics Corporation, Indianapolis, Ind, written communication, August 18, 2006). In addition, the Roche line of instruments revealed dramatic improvement from 2005 to 2006 (Figure 3) and is consistent with an earlier change to lower assigned bilirubin calibrator values (Melanie Swartzentuber, Roche Diagnostics Corporation, written communication, October 5, 2005). The large difference between C-97 and NB-01 mean bilirubin values for the Vitros indicates a lapse in the calibration of the TBIL and neonatal bilirubin (Bu Bc) methods; because NB-01 contained only unconjugated bilirubin, one would expect the TBIL and neonatal bilirubin valuestobeveryclose. Precision Coefficients of variation for CAP All Data increased in 2004 and 2005, but in 2006 returned to the level of 2003 (Table 1). Lack of improvement is attributed to the large differences in the mean bilirubin values reported by the various instrument groups (Figure 3). It should be noted that the CVs for the major instrument groups (Dimension, Olympus, Synchron, and Vitros), which report about 65% of all results, vary from 2% to 3%. Accuracy The goal of limiting the total error to 10% of the reference method value 3 can be realized only by improving accuracy. Since 2003 the gap between the reference values and the CAP All Data has progressively increased (Table 1); this applies to most instrument groups with notable exceptions of Synchron, and lately Advanced BR2 and Cobas Integra. To state simply: Calibrators used by most manufacturers are inaccurate, that is, the assigned values are much higher than the actual bilirubin concentrations. Because for most major instrument groups precision is quite good (CVs, 2% 3%), further improvement in accuracy would depend on the use of accurate calibrators. Instruments from the same manufacturer using methods based on different principles exhibit the widest variation in results. Examples are as follows: (1) By direct spectrophotometry (neonatal bilirubin method), differences between the reference method results and those reported by the Vitros are 2.74 mg/dl (49.9 mol/l) (NB- 01) and 2.06 mg/dl (35.2 mol/l) (NB-11), respectively, but only 0.36 mg/dl (6.1 mol/l) by the TBIL diazo method; the reference method and the mean TBIL Vitros values for C-97 were mg/dl (360 mol/l) and mg/dl (366 mol/l), respectively. (2) Synchron values obtained by the diazo method differ by less than 1.0 mg/ dl from those of the reference method while those of the bilirubin oxidase method differ (except for 2003) from 1.97 mg/dl (33.7 mol/l) to 2.96 mg/dl (50.6 mol/l) (Figure 3). (3) According to Abbott package inserts, 2 methods are available for TBIL for both the Aeroset and Architect analyzers. One is a diazo method (diazotized 2,4-dichloroaniline) and the other a direct spectrophotometry. However, because these are open platform instruments, the users may have chosen a variety of other methods (listed in the participant summary report as JG w/blank, Spectrophot w/o blank or w/blank, DZ Salt, oxidation, vanadate oxidation ), which yield very variable results. It is important to emphasize that calibrators containing ditaurobilirubin or made in a protein base other than human serum may not be suitable for calibrating all bilirubin methods, 1 that is, assigned values to calibrators may have to be different for methods based on different principles. A calibrator suitable for all methods is one made with human serum enriched with unconjugated bilirubin. This is because it has the same composition as specimens obtained from healthy neonates. COMMENT Because of the importance of accuracy in the measurement of bilirubin in neonates, we proposed setting a goal 1784 Arch Pathol Lab Med Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing Lo et al

5 for limiting the total error (inaccuracy plus imprecision) to 10% of the reference method value at bilirubin concentrations greater than 10 mg/dl (171 mol/l). 3 In 2003, the first year of the introduction of the commutable survey specimen consisting of human serum enriched with unconjugated bilirubin, that goal appeared to be within reach for at least 80% of the reported results. That was presumably due to better bilirubin calibrators used in that year as judged by the small difference between the CAP All Data and reference method mean value. In 2006, for specimen NB-01, only the Advanced BR2 and the Synchron CX4/5CE met the upper limit of the 10% goal; all instruments met the lower limit except the Architect. The 20% error allowed by the Clinical Laboratory Improvement Amendments of 1988 implies that acceptable values for a specimen having bilirubin concentration of 25 mg/dl (428 mol/l) would be from 20 mg/ dl (342 mol/l) to 30 mg/dl (512 mol/l); this very large window should be unacceptable to pediatricians and neonatologists. To improve bilirubin assays, manufacturers of clinical analyzers and reagent kits will need to improve the quality of bilirubin calibrators. A bilirubin reference method and reference material, as well as an established molar absorptivity for the alkaline azopigment, can be used to assess the accuracy of calibrators made with unconjugated bilirubin in human serum. These calibrators may then be used to assign values to secondary calibrators for field methods. References 1. Lo SF, Doumas BT, Ashwood ER. Performance of bilirubin determinations in US laboratories revisited. Clin Chem. 2004;50: American Academy of Pediatrics, Provisional Committee for Quality Improvement and Subcommittee on Hyperbilirubinemia. Practice parameter: management of hyperbilirubinemia in the healthy term newborn. Pediatrics. 1994;94: Lo SF, Doumas BT, Ashwood ER. Bilirubin proficiency testing using specimens containing unconjugated bilirubin and human serum: results of a College of American Pathologists study. Arch Pathol Lab Med. 2004;128: Doumas BT, Poon PKC, Perry BW, et al. Candidate reference method for determination of bilirubin in serum: development and validation. Clin Chem. 1985;31: Doumas BT, Perry BW, McComb RB, et al. Molar absorptivities of bilirubin (NIST SRM 916a) and its neutral and alkaline azopigments. Clin Chem. 1990;36: National Institute of Standards and Technology. Certificate of Analysis. Standard Reference Material 916a, Bilirubin. Gaithersburg, Md: National Institute of Standards and Technology; Arch Pathol Lab Med Vol 132, November 2008 Laboratory Performance in Neonatal Bilirubin Testing Lo et al 1785

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory

More information

How to Verify Performance Specifications

How to Verify Performance Specifications How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated

More information

Calcium. Table 1: Difference between method means in percent

Calcium. Table 1: Difference between method means in percent Calcium Measurement of total calcium is widely used for both the diagnosis and the monitoring of a range of conditions related to the bones, heart, nerves, and kidneys. Total calcium measurements include

More information

4.2 Bias, Standards and Standardization

4.2 Bias, Standards and Standardization 4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard

More information

D-Facts. Vitamin D Competitive Evaluation

D-Facts. Vitamin D Competitive Evaluation D-Facts Vitamin D Competitive Evaluation Abstract 101 patient samples collected across sites in Germany were used to compare the performance of the Siemens ADVIA Centaur Vitamin D Total (VitD), Abbott

More information

UNIT: Total and Direct Bilirubin

UNIT: Total and Direct Bilirubin UNIT: Total and Direct Bilirubin 13bili.wpd Task Determination of total and direct bilirubin. Objectives Upon completion of this exercise, the student will be able to: 1. Explain formation, excretion,

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

Validation of measurement procedures

Validation of measurement procedures Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Creatine Kinase REF (200 tests/cartridge) 442635 REF (400 tests/cartridge) 476836 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date

More information

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

How To Test For Creatinine

How To Test For Creatinine Creatinine is measured amperometrically. Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed

More information

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz

More information

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Introduction: The determination of protein concentration is frequently required in biochemical work. Several methods

More information

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare

More information

Biochemistry Validation Form

Biochemistry Validation Form Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August

More information

Sex Hormone Testing by Mass Spectrometry

Sex Hormone Testing by Mass Spectrometry Sex Hormone Testing by Mass Spectrometry Robert L. Fitzgerald, PhD, DABCC Professor of Pathology University of California-San Diego San Diego, CA, 92161 rfitzgerald@ucsd.edu Learning Objectives After this

More information

TRI-LEVEL CARDIAC CONTROL (CRD CONTROL 1, 2, 3)

TRI-LEVEL CARDIAC CONTROL (CRD CONTROL 1, 2, 3) TRI-LEVEL CARDIAC CONTROL (CRD CONTROL 1, 2, 3) CAT NO. CQ3259 LOT NOS. 3190CK, 3191CK, 3192CK SIZE: 3 x 2ml EXPIRY: 2015-02 INTENDED USE This product is intended for in vitro diagnostic use in the quality

More information

James O. Westgard, PhD, 1,2 and Sten A. Westgard, MS 2. Abstract

James O. Westgard, PhD, 1,2 and Sten A. Westgard, MS 2. Abstract Clinical Chemistry / ANALYTIC QUALITY TODAY The Quality of Laboratory Testing Today An Assessment of σ Metrics for Analytic Quality Using Performance Data From Proficiency Testing Surveys and the CLIA

More information

Mass Spectrometry Signal Calibration for Protein Quantitation

Mass Spectrometry Signal Calibration for Protein Quantitation Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

Lab 2. Spectrophotometric Measurement of Glucose

Lab 2. Spectrophotometric Measurement of Glucose Lab 2 Spectrophotometric Measurement of Glucose Objectives 1. Learn how to use a spectrophotometer. 2. Produce a glucose standard curve. 3. Perform a glucose assay. Safety Precautions Glucose Color Reagent

More information

Introduction to Error Analysis

Introduction to Error Analysis UNIVERSITÄT BASEL DEPARTEMENT CHEMIE Introduction to Error Analysis Physikalisch Chemisches Praktikum Dr. Nico Bruns, Dr. Katarzyna Kita, Dr. Corinne Vebert 2012 1. Why is error analysis important? First

More information

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i

More information

2 Spectrophotometry and the Analysis of Riboflavin

2 Spectrophotometry and the Analysis of Riboflavin 2 Spectrophotometry and the Analysis of Riboflavin Objectives: A) To become familiar with operating the Platereader; B) to learn how to use the Platereader in determining the absorption spectrum of a compound

More information

Measuring Protein Concentration through Absorption Spectrophotometry

Measuring Protein Concentration through Absorption Spectrophotometry Measuring Protein Concentration through Absorption Spectrophotometry In this lab exercise you will learn how to homogenize a tissue to extract the protein, and then how to use a protein assay reagent to

More information

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry General Medicine Jiřina Crkovská (translated by Jan Pláteník) 2010/2011 1 Spectrophotometry is one of the most widely used instrumental

More information

The Problem of the Use of different Units for the same Analyte

The Problem of the Use of different Units for the same Analyte The Problem of the Use of different Units for the same Analyte EQALM Symposium 2009 Ulla Tiikkainen ulla.tiikkainen @labquality.fi Comparability of results Measurement traceability to the reference material

More information

2. Contact Person: Garo Mimaryan, MS., RAC 7 Technical Regulatory Affairs Specialist III 3. Phone Number: (914)-524-3270

2. Contact Person: Garo Mimaryan, MS., RAC 7 Technical Regulatory Affairs Specialist III 3. Phone Number: (914)-524-3270 SIEM EN S Abbreviated 510(k) Premnarket Notification IMMULITE 2000 PSA Calibration Verification Material 6. 510(k) Summary as Required by 21 CFR 807.92 K( - is-1 3 (v A. 5 10(k) Number: B. Purpose for

More information

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K092353 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY B. Purpose for Submission: This is a new 510k application for a new indication for the MONOLISA Anti-HAV IgM EIA

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

NITRIC OXIDE and NITROGEN DIOXIDE 6014

NITRIC OXIDE and NITROGEN DIOXIDE 6014 NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm

More information

Evaluation of Interferences in Rate and Fixed-Time Nephelometric Assays of Specific Serum Proteins

Evaluation of Interferences in Rate and Fixed-Time Nephelometric Assays of Specific Serum Proteins Clinical Chemistry 45:1 62 67 (1999) Enzymes and Protein Markers Evaluation of Interferences in Rate and Fixed-Time Nephelometric Assays of Specific Serum Proteins Xavier Bossuyt * and Norbert Blanckaert

More information

Table. Analytical characteristics of commercial cardiac troponin I and T assays declared by the manufacturer.

Table. Analytical characteristics of commercial cardiac troponin I and T assays declared by the manufacturer. Table. Analytical characteristics of commercial cardiac troponin I and T assays declared by the manufacturer. Commercially available assays - Company/ platform(s)/ assay LoB a LoD b 99 th % %CV at 99 th

More information

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

General and statistical principles for certification of RM ISO Guide 35 and Guide 34 General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in

More information

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2 Attachment #2 510 (k) Summary 510(K) SUMMARY FEB - 12008 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

More information

18.6.1 Terms concerned with internal quality control procedures

18.6.1 Terms concerned with internal quality control procedures 18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable

More information

USE OF REFERENCE MATERIALS IN THE LABORATORY

USE OF REFERENCE MATERIALS IN THE LABORATORY USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently

More information

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Alexander Belenky, PhD Director -Product Development Samsung - Nexus-Dx Edward Brennan, PhD

More information

Six Sigma Metric Analysis for Analytical Testing Processes

Six Sigma Metric Analysis for Analytical Testing Processes Six Sigma Metric Analysis for Analytical Testing Processes Sten Westgard, MS, Westgard QC Introduction Laboratories seek objective assessment and comparison of analytical methods and instrumentation performance.

More information

Chemistry 111 Lab: Intro to Spectrophotometry Page E-1

Chemistry 111 Lab: Intro to Spectrophotometry Page E-1 Chemistry 111 Lab: Intro to Spectrophotometry Page E-1 SPECTROPHOTOMETRY Absorption Measurements & their Application to Quantitative Analysis study of the interaction of light (or other electromagnetic

More information

Basic Lessons in Laboratory Quality Control

Basic Lessons in Laboratory Quality Control Bio-Rad Laboratories QC education Basic Lessons in Laboratory Quality Control QC Workbook Basic Lessons in Laboratory Quality Control Written by Greg Cooper, CLS, MHA Manager of Clinical Standards and

More information

Sample Usage Protocol Jaundice Meter JM-105

Sample Usage Protocol Jaundice Meter JM-105 D-94164-2013 Sample Usage Protocol Jaundice Meter JM-105 This Sample Usage Protocol is provided as a template for creating a facility or department-specific protocol for using the Dräger Jaundice Meter

More information

Application of Sigma Metrics for the Assessment of Quality Assurance in Clinical Biochemistry Laboratory in India: A Pilot Study

Application of Sigma Metrics for the Assessment of Quality Assurance in Clinical Biochemistry Laboratory in India: A Pilot Study Ind J Clin Biochem (Apr-June 2011) 26(2):131 135 DOI 10.1007/s12291-010-0083-1 ORIGINAL ARTICLE Application of Sigma Metrics for the Assessment of Quality Assurance in Clinical Biochemistry Laboratory

More information

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical

More information

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination

More information

The sensitive marker for glomerular filtration rate (GFR) Estimation of GFR from Serum Cystatin C:

The sensitive marker for glomerular filtration rate (GFR) Estimation of GFR from Serum Cystatin C: The sensitive marker for glomerular filtration rate (GFR) Estimation of GFR from Serum Cystatin C: The good correlation allows close estimation of GFR Cystatin C GFR GFR in serum estimated* measured* n

More information

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK PROFICIENCY TESTING

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK PROFICIENCY TESTING FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK PROFICIENCY TESTING PRINCIPLE Laboratories may participate, by mandate or by choice, in external proficiency testing programs.

More information

2. Simple Linear Regression

2. Simple Linear Regression Research methods - II 3 2. Simple Linear Regression Simple linear regression is a technique in parametric statistics that is commonly used for analyzing mean response of a variable Y which changes according

More information

Med TechNet Online Services East Amherst, NY November 1997

Med TechNet Online Services East Amherst, NY November 1997 Med TechNet Online Services East Amherst, NY November 1997 DECISION MAKING IN THE CLINICAL LABORATORY: A QUANTITATIVE AND STATISTICAL APPROACH FOR METHODS EVALUATION Associate Professor Department of Clinical

More information

AP Physics 1 and 2 Lab Investigations

AP Physics 1 and 2 Lab Investigations AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks

More information

Transcutaneous Bilirubinometer

Transcutaneous Bilirubinometer Transcutaneous Bilirubinometer Neonatal jaundice occurs in nearly 70% of term and 80% of preterm babies. Management of jaundiced neonates often requires measurement of total serum bilirubin (TSB). Total

More information

TEST METHOD VERIFICATION AND VALIDATION

TEST METHOD VERIFICATION AND VALIDATION 1 TEST METHOD VERIFICATION AND VALIDATION SWACM 2014 MICHAEL LOEFFELHOLZ, PH.D., ABMM DEPT. PATHOLOGY UNIV TEXAS MEDICAL BRANCH GALVESTON, TX 2 OBJECTIVES Define validation and verification Describe components

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR CliniCAl laboratories 2nd Edition Effective 1 April 2010 International Patient Safety Goals (IPSG) Goals The following is a list of all goals.

More information

In vitro diagnostic reagent, calibrator and control material stability

In vitro diagnostic reagent, calibrator and control material stability In vitro diagnostic reagent, calibrator and control material stability Pauline Armstrong Randox Laboratories, Ltd. Crumlin The stability of in vitro medical diagnostic devices whether they are reagents,

More information

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness

More information

Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry

Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Why? Chemists are concerned with mass relationships in chemical reactions, usually run on a macroscopic scale (grams, kilograms, etc.). To deal with

More information

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine

More information

Laboratory Procedure Manual

Laboratory Procedure Manual Laboratory Procedure Manual Analyte: Matrix: Method: Creatine kinase (CPK) Serum Beckman UniCel DxC800 Synchron Method No.: Revised: as performed by: Collaborative Laboratory Services, L.L.C Contact: Important

More information

Chem 131A: Absorbance of Riboflavin

Chem 131A: Absorbance of Riboflavin Chem 131A: Absorbance of Riboflavin Purpose: The purpose of this experiment is to: 1) Familiarize the student with the use of the HP 8452 diode array spectrophotometer, 2) examine the limitations of the

More information

Assay Qualification Template for Host Cell Protein ELISA

Assay Qualification Template for Host Cell Protein ELISA Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to

More information

Blood Glucose Monitoring: The Facts about Accuracy

Blood Glucose Monitoring: The Facts about Accuracy Blood Glucose Monitoring: The Facts about Accuracy Summary The purpose of this document is to provide the facts about the accuracy of self-monitoring of blood glucose (SMBG) systems. Accuracy can be defined

More information

Chapter 5. Risk and Return. Copyright 2009 Pearson Prentice Hall. All rights reserved.

Chapter 5. Risk and Return. Copyright 2009 Pearson Prentice Hall. All rights reserved. Chapter 5 Risk and Return Learning Goals 1. Understand the meaning and fundamentals of risk, return, and risk aversion. 2. Describe procedures for assessing and measuring the risk of a single asset. 3.

More information

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Note: there is a second document that goes with this one! 2046 - Absorbance Spectrophotometry - Calibration Curve Procedure. The second document

More information

An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and

An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and CoaguChek systems Cardiovascular disease: the #1 killer in

More information

ph Alkalinity of Water

ph Alkalinity of Water ph Alkalinity of Water DOC316.52.93085 Based on ISO standard 9963-1:1994 ph-metric Titration 0.4 to 20 mmol/l of Total Alkalinity 1. Introduction Alkalinity of water is its acid-neutralizing capacity.

More information

Qualification Study CHO 360-HCP ELISA (Type A to D)

Qualification Study CHO 360-HCP ELISA (Type A to D) Short Report Qualification Study CHO 360-HCP ELISA (Type A to D) Presented by: http://www.biogenes.de Version 01 Issue date: 27.11.2013 Version 01 Page 1 of 8 Table of Content TABLE OF CONTENT... 2 1 INTRODUCTION...

More information

IMMUNOASSAY CONTROL - LEVEL 1 (IA CONTROL 1)

IMMUNOASSAY CONTROL - LEVEL 1 (IA CONTROL 1) 0843 IMMUNOASSAY CONTROL - LEVEL 1 (IA CONTROL 1) Cat. No. IA2638 / IA2633 Lot No. 978EC Size: 10 x 5 ml / 12 x 5 ml Expiry: 2013-10 INTENDED USE This product is intended for in vitro diagnostic use in

More information

Instrument/Method Validation. Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi

Instrument/Method Validation. Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi Instrument/Method Validation Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi Instrument/Method Validation QUESTION AND ANSWER PRESENTATION Instrument Validation What is instrument validation?

More information

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014 SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014 This experiment explores quality assurance practices which are commonly

More information

INTERNATIONAL OLIVE COUNCIL

INTERNATIONAL OLIVE COUNCIL INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 19/Rev. 3 February 2015 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: iooc@internationaloliveoil.org

More information

Laboratory Math II: Solutions and Dilutions

Laboratory Math II: Solutions and Dilutions Slide 1 Laboratory Math II: Solutions and Dilutions Philip Ryan, PhD Post-Doctoral Fellow National Cancer Institute, NIH Welcome to the National Institutes of Health, Office of Intramural Training & Education

More information

Assay Development and Method Validation Essentials

Assay Development and Method Validation Essentials Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System. CHOLESTEROL 7D62-20 30-3126/R3 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This

More information

A comparison of blood glucose meters in Australia

A comparison of blood glucose meters in Australia Diabetes Research and Clinical Practice 71 (2006) 113 118 www.elsevier.com/locate/diabres A comparison of blood glucose meters in Australia Matthew Cohen *, Erin Boyle, Carol Delaney, Jonathan Shaw International

More information

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard

More information

ALKALINE PHOSPHATASE

ALKALINE PHOSPHATASE ALKALINE PHOSPHATASE 7D55-20 and 7D55-30 30-3947/R7 ALKALINE PHOSPHATASE This package insert contains information to run the Alkaline Phosphatase assay on the ARCHITECT c Systems and the AEROSET System.

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff

Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Document issued on: April 25, 2013 The draft of this document was issued on January 5, 2009. For questions regarding

More information

Chemistry 2351: Inorganic Chemistry I Laboratory Manual

Chemistry 2351: Inorganic Chemistry I Laboratory Manual Spectroscopic Determination of a Complex Ion's Stoichiometry by Job's Method ABSTRACT This experiment is adapted from Angelici's classic experiment, but uses Fe(H 2 O) 6-n (SCN) n 3-n as the complex ion.

More information

Quality Assurance/Quality Control in Acid Deposition Monitoring

Quality Assurance/Quality Control in Acid Deposition Monitoring Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial

More information

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution

More information

Colorimetry Extinction coefficient (ε) Lambda max (λ max ) Qualitative vs. quantitative analysis

Colorimetry Extinction coefficient (ε) Lambda max (λ max ) Qualitative vs. quantitative analysis Lab Week 2 - Spectrophotometry Purpose: Introduce students to the use of spectrophotometry for qualitative (what is it) and quantitative (how much is there of it) analysis of biological samples and molecules.

More information

Colorimetric Determination of Iron in Vitamin Tablets

Colorimetric Determination of Iron in Vitamin Tablets Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered

More information

UPLC/MSMS in the analysis of physiological steroids. May 2012. Anders Feldthus. Anders_Feldthus@Waters.com. 2012 Waters Corporation 1

UPLC/MSMS in the analysis of physiological steroids. May 2012. Anders Feldthus. Anders_Feldthus@Waters.com. 2012 Waters Corporation 1 UPLC/MSMS in the analysis of physiological steroids Anders Feldthus Anders_Feldthus@Waters.com May 2012 2012 Waters Corporation 1 Clinical Chemistry Mass Spectrometry Citations (1955-2006) 70 Number of

More information

Spring Force Constant Determination as a Learning Tool for Graphing and Modeling

Spring Force Constant Determination as a Learning Tool for Graphing and Modeling NCSU PHYSICS 205 SECTION 11 LAB II 9 FEBRUARY 2002 Spring Force Constant Determination as a Learning Tool for Graphing and Modeling Newton, I. 1*, Galilei, G. 1, & Einstein, A. 1 (1. PY205_011 Group 4C;

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number

More information

Standard Test Method for Analysis of Barium, Calcium, Magnesium, and Zinc in Unused Lubricating Oils by Atomic Absorption Spectrometry 1

Standard Test Method for Analysis of Barium, Calcium, Magnesium, and Zinc in Unused Lubricating Oils by Atomic Absorption Spectrometry 1 Designation: D 4628 02 An American National Standard Standard Test Method for Analysis of Barium, Calcium, Magnesium, and Zinc in Unused Lubricating Oils by Atomic Absorption Spectrometry 1 This standard

More information

OLIVÉR BÁNHIDI 1. Introduction

OLIVÉR BÁNHIDI 1. Introduction Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY

More information

Intact PTH. system. Intact PTH 84-6434/R5. Key to symbols used 8K25

Intact PTH. system. Intact PTH 84-6434/R5. Key to symbols used 8K25 system en Intact PTH 8K25 84-6434/R5 Intact PTH Read Highlighted Changes Revised November, 2008 Customer Service United States: 1-877-4ABBOTT International: Call your Abbott Representative This package

More information

Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3

Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3 Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3 Prepared by Date Adopted Supersedes Procedure # Estelle Piwowar-Manning Quality Assessment vers 2 Sarah Dawson Paul Richardson

More information

Design of Experiments for Analytical Method Development and Validation

Design of Experiments for Analytical Method Development and Validation Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information