Proton dose calculation algorithms and configuration data

Size: px
Start display at page:

Download "Proton dose calculation algorithms and configuration data"

Transcription

1 Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems

2 Agenda Broad beam algorithms Concept of pencil beam dose calculation Beam configuration data for Eclipse Proton CT calibration VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 2

3 Agenda Broad beam algorithms Concept of pencil beam dose calculation Beam configuration data for Eclipse Proton CT calibration VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 3

4 Broad beam algorithm - Concept Find the range and the modulation of the beam required to cover the target Find the depth and the distance to the aperture for each point of interest (POI) Use look-up tables for the dose in depth and apply a correction for the lateral penumbra VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 4

5 Broad beam algorithm - Concept Distance to aperture Depth POI Range Modulation VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 5

6 Broad beam algorithm - Applications First proton dose calculation algorithms Used today only for ocular applications Very fast algorithm Loss in accuracy is small due to homogeneous tissues VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 6

7 Agenda Broad beam algorithms Concept of pencil beam dose calculation Beam configuration data for Eclipse Proton CT calibration VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 7

8 Eclipse pencil beam algorithm - Concept Principle Fluence e Φ(x,y,z) Dose Beamlet kernel B(x,y,WeR) Convolution of 3D undisturbed proton fluence in air with a beamlet in water. In practice Superposition of inhomogeneity - corrected beamlets and multiplication with fluence at calculation position. VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 8

9 Fluence of a double scattering beam line 1st and 2nd scatterer Homogeneous fluence distribution Eff. Source size and position Absorber, Modulator Broadening of penumbra Fluence Compensator Broadening of penumbra Inhomogeneity of fluence Block Transversal shape of beam (Block scattering component) VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 9

10 Example of fluence distribution VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 10

11 Fluence of a scanning beam line Phase space of single beam spot (Divergence, beam size at reference position) Fluence Scan pattern and weights of individual beam spots VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 11

12 Example of scanning fluence distribution Fluence of single pencil beam z Total fluence x Scan pattern and weights y x VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 12

13 Beamlet in water Definition: Dose deposited by a proton beam, which hits a water surface without lateral extension and without angular divergence or confusion. Characterization: Depth dose distribution Transversal distribution Independant of technique! VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 13

14 Model of beamlet in water Model is based on analytical functions fitted to Monte Carlo calculated depth dose distributions Depth dose and scattering distributions are modeled for primary and secondary protons separately Measured depth dose curves are used to extract beam-line specific parameters for the model VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 14

15 Example of calculated depth dose distribution VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 15

16 Transversal distribution Beamlet transversal distribution at each position in depth is calculated from the sigma of one or several Gaussians. Transversal distribution of primary protons = sum of two Gaussians VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 16

17 Inhomogeneity correction Each beamlet is corrected for density variations before it is multiplied with the proton fluence in air VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 17

18 Total SOBP dose Contributions of all layers are added in a final step. VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 18

19 Agenda Broad beam algorithms Concept of pencil beam dose calculation Beam configuration data for Eclipse Proton CT calibration VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 19

20 Measurements the input for the dose model Fluence calculation Open field cross profiles -> Virtual SAD Fluence along Z axis -> Effective SAD Half beam block -> Lateral penumbra of fluence Cross profiles of spot fluence -> Phase space Beamlet calculation Depth dose curve -> Depth dose parameters VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 20

21 Virtual SAD Calculation The virtual SAD is the geometric SAD It is found by a linear fit to the field sizes measured at different positions along the beam axis Field Size Vitual SAD Block Position VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 21

22 Effective SAD Calculation The effective SAD is the dosimetric SAD It is found by a 1/r 2 fit to the measurements of the fluence along the beam direction Fluence in air Measurement Effective SAD VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 22

23 Effective Source Size Calculation It is assumed that the penumbra shape can be modelled by the effective source size concept introduced by Hong et al. The proton source is assumed to be located at the position of the nominal source and to have a Gaussian distribution. Proton distribution at source position Half-beam block Error function shaped penumbra Nominal SAD VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 23

24 Phase Space Parameters Phase space parameters are extracted from cross profile measurements of the spot fluence in air. The following formula is fitted to the spot size plotted as a function of the distance from isocenter: A C f z = + B z Parabola with A, B, C parameters 2 ( ) z Nominal SAD Spot fluence VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 24

25 Preprocessing of measured depth dose Measured depth dose gets shifted in depth by water equivalent thickness of the beamline Measured depth dose is corrected for intensity loss due to 1/r 2 effect Dose [Gy/MU] Measurement Shifted by NeT SAD corrected Depth in water [cm] VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 25

26 Calculation of Depth Dose Parameters Resulting depth dose is used for a least square fit of analytic depth dose formula with free parameters Energy Spectral distribution Σ Normalization ~Energy Normalization MeV/cm Gy/MU Fraction of secondary protons (optional) Spectrum Σ VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 26

27 Parametrization vs. LUTs Parametrization Requires an analytical model for measured data. Failure of the model leads to bad dose calculation results Allows interpolation (extrapolation) for situations other than the measured one (e.g. intermediate energy) Smoothes or averages results from measurements LUT (Look up table) Measured curve can have an arbitrary shape Difficult to interpolate, almost impossible to extrapolate Needs some processing of measurements to smooth uncertainties VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 27

28 Parametrization vs. LUTs Example 1 Depth dose Extrapolation of dose calculation and better model for coarse measurements Interpolation of calibration factor for different energies VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 28

29 Parametrization vs. LUTs Example 2 A-Parameter (= 2*spot size at isocenter^2) for a scanning beam Note: User must know, whether observed behavior of spot size is due to measurement uncertainties or other effects. VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 29

30 Agenda Broad beam algorithms Concept of pencil beam dose calculation Beam configuration data for Eclipse Proton CT calibration VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 30

31 Why CT calibration? CT scanners measure Hounsfield units defined as μ HU =1000 water -μ μ water material Hounsfield units are a measure for the attenuation of an X-ray beam Hounsfield units cannot be used directly to calculate the behavior of proton or even high energy photon beams VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 31

32 Definitions Scaled HU: HU sc = HU Absolute stopping power (SP) : -de/dx [MeV/cm] Absolute mass SP : -(1/ρ)dE/dx [MeV/(g/cm 2 )] Relative SP : Ratio between abs. SP for material and water Relative mass SP : Ratio between abs. mass SP for material and water CT calibration: LUT for conversion between HU and relative SP VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 32

33 Importance of accurate CT calibration A few percent error in the CT calibration causes a few percent shift of the dose in depth Due to the high distal gradient of the dose, this may lead to 100% difference in dose! Note: The calibration curve is sensitive to the X-ray spectrum of the CT scanner. It is important, that each scanner is calibrated individually. VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 33

34 The stoichiometric CT calibration method Scan tissue equivalent samples of well known composition individually in the center of a water phantom Obtain parameters for HU sc calculation by fitting results to HU-formula Calculate (HU sc,sp rel ) pairs for real tissue compositions Define calculation curve to follow calculated (HU sc,sp rel ) pairs for real tissue compositions VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 34

35 Properties of the stoichiometric CT calibration method Scanner specific due to the use of reference measurements Uses a simplified, empirical theory for the calculation of Hounsfield units Yields good results for commonly used scanner energies of approx. 80 kev and biological tissue VARIAN Medical Systems

36 Calculation of (HU sc, SP rel ) reference data pairs Calculation of HU sc and SP rel pair based on literature data for tissue composition { } ph 3.62 coh 1.86 KN HU sc = Cmat K + K + K Don t use for MV CT! Z ~ Z ~ K xy parameterize the response of the scanner. They are determined by fitting the above formula to HU sc - values measured on samples of well known chemical composition SP rel calculated with Bethe-Bloch formula VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 36

37 Plot of calculated (HU sc,sp rel ) pairs and linear fits 1.8 Small body calibration Rel. Stopping Power Organs and muscle Adipose Breast Bone marrow Cartilage Bone Fit to soft tissue Fit to bone Fit to adipose Calibration curve Hounsfield Units Rel. Stopping Power Organs and muscle Adipose Breast Bone marrow Cartilage Fit to soft tissue Fit to bone Fit to adipose Calibration curve Hounsfield Units VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 37

38 Some references for further reading Schneider U, Pedroni E, Lomax A 1996 The calibration of CT Hounsfield units for radiotherapy treatment planning Phys. Med. Biol Schaffner B, Pedroni E 1998 The precision of proton range calculations in proton radiotherapy treatment planning: experimental verification of the relation between CT-HU and proton stopping power Phys. Med. Biol Kanematsu N et al A CT calibration method based on the polybinary tissue model for radiotherapy treatment planning Phys. Med. Biol Schneider U et al Patient specific optimization of the relation between CT-Hounsfield units and proton stopping power with proton radiography Med. Phys. 32(1) VARIAN Medical Systems Barbara Schaffner, PTCOG 46 Slide 38

39 Thank you for your interest and attention VARIAN Medical Systems

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning Karin M. Andersson,

More information

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu

More information

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually

More information

Introduction to the Monte Carlo method

Introduction to the Monte Carlo method Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw

More information

Proton tracking for medical imaging and dosimetry

Proton tracking for medical imaging and dosimetry Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4

More information

Use of 3D Printers in Proton Therapy

Use of 3D Printers in Proton Therapy Use of 3D Printers in Proton Therapy Nicholas Remmes, PhD Mayo Clinic, Rochester, MN Department of Radiation Oncology Remmes.Nicholas@mayo.edu 2015 MFMER slide-1 Disclosures/Confessions My 3D Printing

More information

Manual for simulation of EB processing. Software ModeRTL

Manual for simulation of EB processing. Software ModeRTL 1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended

More information

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

Variance reduction techniques used in BEAMnrc

Variance reduction techniques used in BEAMnrc Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,

More information

Challenges in small field MV photon dosimetry

Challenges in small field MV photon dosimetry Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland maria.aspradakis@ksl.ch Why are

More information

Proton Therapy for Head & Neck Cancers

Proton Therapy for Head & Neck Cancers Proton Therapy for Head & Neck Cancers Robert S Malyapa, MD, PhD and William M Mendenhall, MD University of Florida Proton Therapy Institute, Jacksonville, USA Carmen Ares, MD and Ralf Schneider, MD Paul

More information

Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues

Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues Journal of mathematics and computer science 9 (2014), 41-45 Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues S. A. Mahdipour 1, H. R. Abdi Roknabadi

More information

Clinical Photon Beams

Clinical Photon Beams Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips

More information

Therapy with protons and ion beams

Therapy with protons and ion beams Therapy with protons and ion beams Biomedical Physics Lecture WiSe 2012/13 many slides from J. Willkens (TUM) Key quesions: How does ion cancer therapy work? What is the physics behind it? How can physics

More information

The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning.

The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning. The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning. KAROLINA KOKUREWICZ S U P E R V I S O R S : D I N O J A R O S Z Y N S K I, G I U S E P P E S C

More information

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department

More information

Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS

Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS W. STRYDOM Department of Medical Physics, Medical University of Southern Africa, Pretoria, South Africa W. PARKER, M. OLIVARES Department of Medical

More information

THE DOSIMETRIC EFFECTS OF

THE DOSIMETRIC EFFECTS OF THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,

More information

Introduction to X-Ray Powder Diffraction Data Analysis

Introduction to X-Ray Powder Diffraction Data Analysis Introduction to X-Ray Powder Diffraction Data Analysis Center for Materials Science and Engineering at MIT http://prism.mit.edu/xray An X-ray diffraction pattern is a plot of the intensity of X-rays scattered

More information

Quality Assurance for Treatment Planning Systems

Quality Assurance for Treatment Planning Systems Quality Assurance for Treatment Planning Systems PTCOG Educational Session May 19 th 2010 Oliver Jäkel Heidelberg Ion Beam Therapy Center and German Cancer Research Center, Heidelberg Introduction Outline

More information

QA of intensity-modulated beams using dynamic MLC log files

QA of intensity-modulated beams using dynamic MLC log files 36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,

More information

Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy

Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.176-180 (2011) ARTICLE Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy Fada

More information

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50

More information

2 Absorbing Solar Energy

2 Absorbing Solar Energy 2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could

More information

Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods

Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods Short report Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods A. Chaparian 1*,M.A. Oghabian 1, V. Changizi

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

AP Physics 1 and 2 Lab Investigations

AP Physics 1 and 2 Lab Investigations AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada

More information

THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode

THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode Swedish Laser-Medical Society The dose is the most important parameter in laser phototherapy. At a first glance, the dose seem very

More information

Status and perspective of emission imaging techniques for ion beam therapy in Lyon

Status and perspective of emission imaging techniques for ion beam therapy in Lyon 14th International Conference on Nuclear Reaction Mechanisms, Varenna 18th June 2015 Status and perspective of emission imaging techniques for ion beam therapy in Lyon on behalf of the CAS-PHABIO group

More information

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac

More information

Marcus Fager Master of Science Thesis University of Pennsylvania, 11 June 2013. Supervisors:

Marcus Fager Master of Science Thesis University of Pennsylvania, 11 June 2013. Supervisors: R A D I O B I O L O G I C A L P L A N O P T I M I Z A T I O N I N P R O T O N T H E R A P Y F O R P R O S T A T E T U M O R S U S I NG A P A T C H E D I N T E G R A T E D E D G E [ P I E ] T E C H N I

More information

Secondary Neutrons in Proton and Ion Therapy

Secondary Neutrons in Proton and Ion Therapy Secondary Neutrons in Proton and Ion Therapy L. Stolarczyk Institute of Nuclear Physics PAN, Poland on behalf of WG9 EURADOS Acknowledgments EURADOS Workig Group 9 Roger Harrison Jean Marc Bordy Carles

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry

Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry John Keightley NSUF 18 May 2010 Acknowledgement: Several discussions with Stefaan Pommé, IRMM, EC-JRC, Geel Some diagrams

More information

Monte Carlo Simulations in Proton Dosimetry with Geant4

Monte Carlo Simulations in Proton Dosimetry with Geant4 Monte Carlo Simulations in Proton Dosimetry with Geant4 Zdenek Moravek, Ludwig Bogner Klinik und Poliklinik für Strahlentherapie Universität Regensburg Objectives of the Study what particles and how much

More information

Constrained Tetrahedral Mesh Generation of Human Organs on Segmented Volume *

Constrained Tetrahedral Mesh Generation of Human Organs on Segmented Volume * Constrained Tetrahedral Mesh Generation of Human Organs on Segmented Volume * Xiaosong Yang 1, Pheng Ann Heng 2, Zesheng Tang 3 1 Department of Computer Science and Technology, Tsinghua University, Beijing

More information

Overview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota

Overview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota Creation of 3D printed phantoms for clinical radiation therapy Eric Ehler, PhD Assistant Professor University of Minnesota ehler 046@umn.edu Overview Background of 3D Printing Practical Information Current

More information

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation

More information

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100) Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know

More information

Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center

Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster Jesse Garant, JG&A Metrology Center Traditional Metrology and Inspection Tactile Devices (Touch Probe)

More information

ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE

ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE Gevaldo Lisboa de Almeida, Carlos Feu Alvim Agência Brasileiro-Argentina de

More information

Atomic and Nuclear Physics Laboratory (Physics 4780)

Atomic and Nuclear Physics Laboratory (Physics 4780) Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *

More information

Appendix A. An Overview of Monte Carlo N-Particle Software

Appendix A. An Overview of Monte Carlo N-Particle Software Appendix A. An Overview of Monte Carlo N-Particle Software A.1 MCNP Input File The input to MCNP is an ASCII file containing command lines called "cards". The cards provide a description of the situation

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

Scripps Proton Therapy Center: Configuration and Implementation

Scripps Proton Therapy Center: Configuration and Implementation Scripps Proton Therapy Center: Configuration and Implementation Anthony Mascia Proton Symposium 2013 Annual AAPM Meeting Indianapolis, Indiana USA Facility Configuration Scripps Proton Therapy Center ***All

More information

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Performance evaluation and quality assurance of Varian enhanced dynamic wedges JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department

More information

The accurate calibration of all detectors is crucial for the subsequent data

The accurate calibration of all detectors is crucial for the subsequent data Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved

More information

Measuring Line Edge Roughness: Fluctuations in Uncertainty

Measuring Line Edge Roughness: Fluctuations in Uncertainty Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as

More information

Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY

Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY W. PARKER, H. PATROCINIO Department of Medical Physics, McGill University Health Centre, Montreal, Quebec, Canada 7.1. INTRODUCTION

More information

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons SLAC-PUB-7722 January 9 degrees Bremsstrahlung Source Term Produced in Thick Targets by 5 MeV to GeV Electrons X. S. Mao et al. Presented at the Ninth International Conference on Radiation Shielding, Tsukuba,

More information

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY N. GEERAERT a,b,c, R. KLAUSZ a, S. MULLER a, I. BLOCH c, H. BOSMANS b a GE Healthcare, Buc, France b Department

More information

World-first Proton Pencil Beam Scanning System with FDA Clearance

World-first Proton Pencil Beam Scanning System with FDA Clearance Hitachi Review Vol. 58 (29), No.5 225 World-first Proton Pencil Beam Scanning System with FDA Clearance Completion of Proton Therapy System for MDACC Koji Matsuda Hiroyuki Itami Daishun Chiba Kazuyoshi

More information

Fundamentals of Cone-Beam CT Imaging

Fundamentals of Cone-Beam CT Imaging Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation

More information

Copyright 2007 Casa Software Ltd. www.casaxps.com. ToF Mass Calibration

Copyright 2007 Casa Software Ltd. www.casaxps.com. ToF Mass Calibration ToF Mass Calibration Essentially, the relationship between the mass m of an ion and the time taken for the ion of a given charge to travel a fixed distance is quadratic in the flight time t. For an ideal

More information

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review

More information

Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021

Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021 Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021 Introduction Prostate is one of the treatment sites that is well suited for IMRT. For

More information

Calorimetry in particle physics experiments

Calorimetry in particle physics experiments Calorimetry in particle physics experiments Unit n. 8 Calibration techniques Roberta Arcidiacono Lecture overview Introduction Hardware Calibration Test Beam Calibration In-situ Calibration (EM calorimeters)

More information

Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy

Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy This set of 117 slides is based on Chapter 11 authored by M.D.C. Evans of the IAEA publication (ISBN 92-0-107304-6):

More information

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures

More information

Proton Radiotherapy. Cynthia Keppel Scientific and Technical Director Hampton University Proton Therapy Institute. Lead Virginia.

Proton Radiotherapy. Cynthia Keppel Scientific and Technical Director Hampton University Proton Therapy Institute. Lead Virginia. Proton Radiotherapy Cynthia Keppel Scientific and Technical Director Hampton University Proton Therapy Institute Lead Virginia g October 16, 2009 About Proton Therapy Cancer is the (2nd) largest cause

More information

Jitter Measurements in Serial Data Signals

Jitter Measurements in Serial Data Signals Jitter Measurements in Serial Data Signals Michael Schnecker, Product Manager LeCroy Corporation Introduction The increasing speed of serial data transmission systems places greater importance on measuring

More information

The Science behind Proton Beam Therapy

The Science behind Proton Beam Therapy The Science behind Proton Beam Therapy Anthony Zietman MD Shipley Professor of Radiation Oncology Massachusetts General Hospital Harvard Medical School Principles underlying Radiotherapy Radiation related

More information

Evolution of Head and Neck Treatment Using Protons. Mayank Amin, M.Sc,CMD

Evolution of Head and Neck Treatment Using Protons. Mayank Amin, M.Sc,CMD Evolution of Head and Neck Treatment Using Protons Mayank Amin, M.Sc,CMD Facility Layout Gantry Room 3 Fixed Beams Room 4 HEBT Gantry Room 2 Gantry Room 1 Synchrotron Linac Treatment Planning Imaging Area

More information

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Standard Test Method for Classification of Film Systems for Industrial Radiography 1 Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately

More information

Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer

Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer Evaluation of complexity and deliverability of IMRT- treatment plans Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer June 11 Abstract Background: Intensity modulated beams are used

More information

Facility Start Up, Training, and Operational Issues AT PTC H or Protons A Positive Marathon Inelastic Collision Experience

Facility Start Up, Training, and Operational Issues AT PTC H or Protons A Positive Marathon Inelastic Collision Experience Facility Start Up, Training, and Operational Issues AT PTC H or Protons A Positive Marathon Inelastic Collision Experience Presented from a Physics Bias Michael Gillin, Ronald Zhu, Narayan Sahoo And Many

More information

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,

More information

Advanced variance reduction techniques applied to Monte Carlo simulation of linacs

Advanced variance reduction techniques applied to Monte Carlo simulation of linacs MAESTRO Advanced variance reduction techniques applied to Monte Carlo simulation of linacs Llorenç Brualla, Francesc Salvat, Eric Franchisseur, Salvador García-Pareja, Antonio Lallena Institut Gustave

More information

1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002

1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical

More information

Testing thermo-acoustic sound generation in water with proton and laser beams

Testing thermo-acoustic sound generation in water with proton and laser beams International ARENA Workshop DESY, Zeuthen 17th 19th of May 25 Testing thermo-acoustic sound generation in water with proton and laser beams Kay Graf Universität Erlangen-Nürnberg Physikalisches Institut

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

FTIR Instrumentation

FTIR Instrumentation FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation

More information

Development of on line monitor detectors used for clinical routine in proton and ion therapy

Development of on line monitor detectors used for clinical routine in proton and ion therapy Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary

More information

Chapter 18: The Structure of the Atom

Chapter 18: The Structure of the Atom Chapter 18: The Structure of the Atom 1. For most elements, an atom has A. no neutrons in the nucleus. B. more protons than electrons. C. less neutrons than electrons. D. just as many electrons as protons.

More information

Current Status and Future Direction of Proton Beam Therapy

Current Status and Future Direction of Proton Beam Therapy Current Status and Future Direction of Proton Beam Therapy National Cancer Center Hospital East Division of Radiation Oncology and Particle Therapy Tetsuo Akimoto Comparison of status of particle therapy

More information

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine

More information

Classification of Fingerprints. Sarat C. Dass Department of Statistics & Probability

Classification of Fingerprints. Sarat C. Dass Department of Statistics & Probability Classification of Fingerprints Sarat C. Dass Department of Statistics & Probability Fingerprint Classification Fingerprint classification is a coarse level partitioning of a fingerprint database into smaller

More information

Bildverarbeitung und Mustererkennung Image Processing and Pattern Recognition

Bildverarbeitung und Mustererkennung Image Processing and Pattern Recognition Bildverarbeitung und Mustererkennung Image Processing and Pattern Recognition 1. Image Pre-Processing - Pixel Brightness Transformation - Geometric Transformation - Image Denoising 1 1. Image Pre-Processing

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters

Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters C-A/AP/#122 November 23 Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters J. Beebe-Wang, P. Vaska, F.A. Dilmanian, S.G. Peggs and D.J. Schlyer "This paper

More information

Positron Range Modeling for Statistical PET Image Reconstruction

Positron Range Modeling for Statistical PET Image Reconstruction Positron Range Modeling for Statistical PET Image Reconstruction Bing Bai, Ananya Ruangma, Richard Laforest, Yuan-Chuan Tai and Richard M. Leahy Signal and Image Processing Institute, University of Southern

More information

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY 243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution

More information

Application Report. Propeller Blade Inspection Station

Application Report. Propeller Blade Inspection Station 34935 SE Douglas Street, Suite, Snoqualmie, WA 9874 Ph: 425-396-5577 Fax: 425-396-586 Application Report Propeller Blade Inspection Station Prepared By Kyle Johnston, Ph. D. Metron Systems Inc.5.5 3 2.5

More information

Algebra 1 Course Title

Algebra 1 Course Title Algebra 1 Course Title Course- wide 1. What patterns and methods are being used? Course- wide 1. Students will be adept at solving and graphing linear and quadratic equations 2. Students will be adept

More information

ACCELERATORS AND MEDICAL PHYSICS 2

ACCELERATORS AND MEDICAL PHYSICS 2 ACCELERATORS AND MEDICAL PHYSICS 2 Ugo Amaldi University of Milano Bicocca and TERA Foundation EPFL 2-28.10.10 - U. Amaldi 1 The icone of radiation therapy Radiation beam in matter EPFL 2-28.10.10 - U.

More information

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,

More information

GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2

GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2 GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures

More information

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der

More information

DESCRIPTIVE STATISTICS. The purpose of statistics is to condense raw data to make it easier to answer specific questions; test hypotheses.

DESCRIPTIVE STATISTICS. The purpose of statistics is to condense raw data to make it easier to answer specific questions; test hypotheses. DESCRIPTIVE STATISTICS The purpose of statistics is to condense raw data to make it easier to answer specific questions; test hypotheses. DESCRIPTIVE VS. INFERENTIAL STATISTICS Descriptive To organize,

More information

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,

More information

COMPARISON OF FOUR DATA ANALYSIS SOFTWARE FOR COMBINED X-RAY REFLECTIVITY AND GRAZING INCIDENCE X-RAY FLUORESCENCE MEASUREMENTS

COMPARISON OF FOUR DATA ANALYSIS SOFTWARE FOR COMBINED X-RAY REFLECTIVITY AND GRAZING INCIDENCE X-RAY FLUORESCENCE MEASUREMENTS COMPARISON OF FOUR DATA ANALYSIS SOFTWARE FOR COMBINED X-RAY REFLECTIVITY AND GRAZING INCIDENCE X-RAY FLUORESCENCE MEASUREMENTS Bérenger Caby (1), Fabio Brigidi (2), Dieter Ingerle (3), Blanka Detlefs

More information

How To Improve Your Ct Image Quality

How To Improve Your Ct Image Quality Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Potential Benefits of Proton Therapy in Clinic

Potential Benefits of Proton Therapy in Clinic Potential Benefits of Proton Therapy in Clinic Simone Pieplenbosch Proton therapy offers great potential in cancer treatment compared with the conventional photon therapy. The advantages of the unique

More information

1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range

1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range 1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range The radiation from orthovoltage units is referred to as x-rays, generated by bombarding

More information

TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS

TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS AAPM REPORT NO. 85 TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS Report of Task Group No. 65 of the Radiation Therapy Committee of the American Association of Physicists in Medicine Members

More information

Acknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers

Acknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers H. Helen Liu, PhD Department of Radiation Physics, U.T. MD Anderson Cancer Center, Houston, TX AAPM, Seattle, 2005 Acknowledgement

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information