STRUCTURE OF CHEMICAL COMPOUNDS, METHODS OF ANALYSIS AND PROCESS CONTROL

Size: px
Start display at page:

Download "STRUCTURE OF CHEMICAL COMPOUNDS, METHODS OF ANALYSIS AND PROCESS CONTROL"

Transcription

1 Pharmaceutical Chemistry Journal Vol. 38, No. 4, 004 STRUCTURE OF CHEMICAL COMPOUNDS, METHODS OF ANALYSIS AND PROCESS CONTROL VALIDATION OF HPLC TECHNIQUES FOR PHARMACEUTICAL ANALYSIS N. A. Épshtein 1 Translated from Khimiko-Farmatsevticheskii Zhurnal, Vol. 38, No. 4, pp , April, 004. Original article submitted June 18, Akrikhin Chemico-Pharmaceutical Joint-Stock Company, Staraya Kupavna, Moscow Region, Russia. Here and below the terms accuracy and precision (repeatability, reproducibility) are treated in accordance with the State Standard GOST R ISO Previously, accuracy had the meaning of correctness, but now correctness is described in terms of trueness. In justifying the suitability of techniques related to the qualitative determination of substances, statistical processing of the experimental results is obligatory [8]. Validation (evaluation of suitability) of an analytical technique is a procedure aimed at obtaining experimentally justified evidence of the ability of this technique to give results characterized by the required accuracy and precision [1 7]. All analytical techniques used for the development of pharmaceuticals and for the determination of their quality characteristics have to be validated. In the case of using methods stipulated and described in the State Pharmacopoeia, it is not necessary to evaluate their suitability, provided that the analyses are conducted with strict observation of the text of each particular article. In most other cases, especially in cases of modification of the drug composition, the scheme of synthesis, or the analytical procedure, it is necessary to re-evaluate the suitability of the analytical techniques. This paper is aimed at (i) considering the peculiarities of validation of HPLC techniques for pharmaceutical analysis, (ii) critically assessing the main approaches to evaluation of the validation characteristics, and (iii) providing practical recommendations and criteria for finding correct solutions. The USSR State Pharmacopoeia (valid in the Russian Federation) introduced the section Statistical Analysis of Biological Test Results in 1968 (Xth Ed.) and the section Statistical Processing of Chemical Experimental Data in 1988 (XIth Ed.). These sections are devoted to problems involved in the metrological attestation of analytical techniques. In 1987, the United States Food and Drug Administration (FDA) issued practical guides on the main principles of validation [5] and on the presentation of samples and analytical data pertaining to the validation of methods [6]. In 1993, the International Conference on Harmonization (ICH) developed generalized recommendations on the validation of analytical procedures; these documents were published in 1994 and treated in more detail in 1995 [1 3]. In 1994, the US FDA Center for Drug Evaluation and Research (CDER) issued a guide on the validation of chromatographic methods [4]. These documents and some review papers and monographs [9 13] provided a basis for extensive implementation of the procedure of validation of analytical methods. The Internet offers the Laboratory Guide to Method Validation and Related Topics at /. In recent years, new guides have become available from CDER [15], Waters Company [16], and Labcompliance [17 19]. Special sections are devoted to these problems in national and international pharmacopoeias and guides on the validation of analytical procedures [7, 0 ]. Also available are computer program packages, such as ELISA Method Validation Templates (Waters) and LaChrom 000 Validation Manager (Merck) representing electronic tables with incorporated functions of statistical processing of the results of measurements and issuing validation certificates, and monographs on the related subjects [3 34]. Tables 1 and summarize the most recent recommendations concerning selection of the validation characteristics depending on the type of analytical procedures. A comparative analysis of these data shows that, according to the United States Pharmacopoeia (USP-6), methods of dissolution testing have to be validated only with respect to precision (repeatability, reproducibility), while the other characteristics may require validation, depending on the specific test nature. In contrast, according to the FDA CDER guidelines [15], procedures used for quantitative analysis and dissolution testing need the same validation characteristics X/04/ Plenum Publishing Corporation

2 Validation of HPLC Techniques for Pharmaceutical Analysis 13 Moreover, according to CDER [15], all methods of quantitative analysis have to be characterized with respect to robustness (see below). Robustness is not included in USP-6 [7] because this characteristic has to be studied at the stage of development of an analytical procedure, rather than in the course of validation. Since any analytical situation poses a multifactor problem, validation has to include at least testing of the analytical system as a whole under the conditions stipulated by the description. 3 The second task is evaluation of the stability (robustness) of the analytical system with respect to small variations of the main factors (for example, the ratio of the mobile phase components, ph, temperature, etc.). This problem is less important than the first one because it is usually solved at the stage of development of a given analytical procedure. For this reason, problems pertaining to robustness are only briefly mentioned in this review. Let us consider the main stages of validation of the HPLC techniques used in pharmacy. 1. TESTING THE ANALYTICAL PROCEDURE AS A UNIFIED SYSTEM UNDER THE CONDITIONS STIPULATED IN THE DESCRIPTION 3 According to this concept, instrumentation, electronics, analytical procedures, and analyzed samples constitute a unified analytical system, which can be considered as a whole [7]. 4 Sometimes, the stability of phases and solutions is considered within the framework of the problem of robustness. Specificity (Section 1.1). Precision (Section 1.).1 Repeatability (Section 1..1). Intermediate precision (Section 1..) 3. Linearity (Section 1.3) 4. Accuracy (Section 1.4) 5. Suitability range (Section 1.5) 6. Limit of detection (Section 1.6) 7. Limit of quantitation (Section 1.7) 8. Stability of solutions (Section 1.8.) 9. Robustness (stability) of HPLC procedures with respect to small variations in the main system factors (ratio of the mobile phase components, ph, temperature, etc.). Robustness is usually studied in the stage of development of the given HPLC tecnique (Section ) Recommended requirements to the repeatability of sample injections (Section 3.3).3 Reproducibility. Checked in special cases (Section 1..3) It is expedient to determine these validation characteristics in the course of proof of the analytical system accuracy (using statistical parameters determined for the calibration graph) (Sections 1.3; 1.4; 1.6.; and 1.7.) Criteria of suitability of a given chromatographic system (Section 3) Fig. 1. The general scheme of validation of HPLC-based analytical procedures. Figure 1 shows the general scheme of evaluation of the suitability of an analytical procedure, which takes into account specific features of HPLC. As can be seen, testing an analytical procedure as a whole in the general case allows the following validation characteristics to be determined [1 7]: (i) specificity; (ii) precision; (iii) linearity; (iv) accuracy; (v) suitability range; (vi) limit of detection; (vii) limit of quantitation; and (viii) stability of solutions. 4 Depending on a particular type of the analytical procedure (HPLC technique) only a part rather than all of the above characteristics may be required (see Tables 1 and ) Confirming the Specificity of a Given Analytical Procedure (Separating Power of a Chromatographic System) By the specificity of a system is meant its ability to detect a given substance unambiguously (reliably) in the presence of other components (including impurities) that may be present in the samples [1 4]. The proof of specificity depends on the task of a given procedure and on the availability of reference samples of the main impurities The specificity of procedures aimed at determination of the content of a parent substance, the parameters of solubility, and the homogeneity of dosage. In order to confirm the specificity of these procedures, it is usually required that peaks of the substances to be determined are sufficiently well resolved between themselves and from peaks of the main impurities, the system components (e.g., of the sample solvent), and the placebo. For this purpose, the separation of peaks is confirmed by a set of chromatograms, at least of (a) the test solution, (b) the reference parent substance solution, (c) the solvent (blank), (d) the placebo (for filled drugs), and (e) the solution used for the evaluation of suitability of the chromatographic system. The degree of peak separation is usually described in terms of the separation coefficient R s. Recommended values of this coefficient are given below in the Section 3.3. It is recommended to confirm the specificity by investigation of the purity of peaks of the parent substance to be determined [4, 7]. This test is usually performed using a diode matrix detector and a special program evaluating spectral homogeneity of the measured peak (e.g., Peak Purity Millennium, Waters). The principle of evaluation of the peak purity is as follows. The sample chromatograms are measured at various detector wavelengths (numbered n ). Each point of the peak is characterized by a spectrum, which is mathematically described by a vector in the n-dimensional space of the values of absorption (in absorption units, AU) at the preset n wavelengths (the length of this vector is proportional to the substance concentration in the solution studied). The difference between the spectra is evaluated by the angle between the corresponding vectors (called the spectral con-

3 14 N. A. Épshtein TABLE 1. Validation Characteristics according to th United States Pharmacopoeia (003) Characteristic I II Category Quantitative tests Limiting tests Accuracy Yes Yes MB MB No Precision Yes Yes No Yes No Specificity Yes Yes Yes MB Yes Limit of detection No No Yes MB No Limit of quantitation No Yes No MB No Linearity Yes Yes No MB No Suitability range Yes Yes MB MB No Notes: Yes = usually studied; No = usually not studied; MB = may be required (depending on a specific test nature). Category I includes analytical methods intended for determination of the content of the main component in parent substances or ready-to-use medicinal forms; category II includes methods of determination of impurities and decomposition products; category III includes methods of determination of the parameters of dissolution, drug release, etc.; category IV includes identification tests; the terms accuracy and precision (repeatability, reproducibility) are treated in accordance with the State Standard GOST R ISO III IV TABLE. Validation Characteristics Recommended for Various Tests by the United States FDA (CDER and CBER) [15] Chatacteristic Identification Type of analytical procedure Tests for impurities quantitative limiting Qualitative determination and dissolution tests Accuracy No Yes No Yes Repeatability No Yes No Yes Intermediate precision No Yes 1 No Yes 1 Specificity Yes Yes Yes Yes 4 Limit of detection No No 3 Yes No Limit of quantitation No Yes No No Linearity No Yes No Yes Suitability range No Yes No Yes Robustness No Yes No 3 Yes Notes: Yes = usually studied; No = usually not studied; 1 in cases where the reproducibility is studied, there is no need to specially determine the intermediate precision; insufficient specificity of a given analytical procedure can be compensated by introducing additional tests; 3 may be required in some cases (e.g., when the limit of detection is close to the rated limiting content of an impurity); 4 insufficient specificity of a given analytical procedure can be compensated by determining the content of impurities. trast angle ). If = 0, the spectra are considered similar (homogeneous). This implies that the value of absorption in one spectrum measured at a given wavelength can be obtained from the value in another spectrum measured at the same wavelength by multiplying by a certain constant factor. In order to evaluate the spectral homogeneity of a chromatographic peak, the spectral contrast angles are calculated for all points of this peak relative to the angle at the peak maximum and then the maximum value p (purity angle) is determined. Two spectra determined for the same substance can differ, for example, because of the influence of the baseline noise. In order to take this into account, a threshold spectral angle th is determined (i) by determining the maximum spectral angle between pints of the baseline with maximum noise levels or (ii) by taking six chromatograms of a standard sample solution, determining the maximum purity angle p for each chromatogram, and considering the maximum of these values as the threshold spectral angle th. This threshold angle characterizes the level below which the difference between two spectra can be considered insignificant. The obtained p values are compared to th. For p < th, the peak is considered spectrally homogeneous; otherwise the peak is influenced by the presence (i.e., additional absorption) of another substance. In practice, it is possible to use a simplified procedure, whereby the chromatograms of the same peak are measured at two or three wavelengths and the chromatograms are checked for the proportionality (see above) at all points. The latter method is less reliable, because the spectrum can be influenced by variations in the mobile phase composition (e.g., under gradient HPLC conditions), or by deviations from the Lambert Beer law at high levels of absorption. Therefore, negative results of evaluation of the spectral homogeneity of peaks in the gradient HPLC should be critically assessed and checked for optical densities not exceeding 1 AU. It should be noted that specificity is rarely checked using chromatomass spectroscopy (HPLC MS) because of the high cost of this procedure and the chemical and other analyses of the eluate fractions corresponding to the parent substance, because of tedious procedures The specificity of procedures aimed at determination of impurities. In evaluating the specificity of such procedures, it is necessary to differentiate between two cases. Case 1. The main impurities are known and available. In order to confirm the specificity of a procedure used for determining the impurities in a parent substance, it is necessary to show that (i) this procedure allows the peaks of the main products of decomposition of the parent substance to be detected and that (ii) the peaks of impurities are sufficiently well separated between themselves and from peaks of the parent substance and the system components (solvents). Developers of the technology of a parent substance have to prove additionally that (iii) the proposed procedure allows determining the main impurities related to the technological process and that (iv) all such impurities present at an amount of 0.1% are identified [7].

4 Validation of HPLC Techniques for Pharmaceutical Analysis 15 In order to confirm the specificity of a procedure used for determining impurities in parent substances, it is necessary to demonstrate that (i) this procedure allows the peaks of the main rated impurities to be detected and that (ii) the peaks of these impurities are sufficiently well separated between themselves and from peaks of the parent substance and the system components (solvents). The specificity is confirmed by a set of chromatograms, including at least those of (a) a model solution of the parent substance and the main impurities (prepared by adding these known impurities or their solutions to the parent compound or its mixture with the placebo), (b) the solvent, (c) the placebo (for filled drugs), (d) the test solution, and (e) the solution used for the evaluation of suitability of the chromatographic system. In addition, it is recommended to confirm the specificity of the given analytical procedure by data on the purity of the main peaks in the chromatograms of test solutions. Case. The main impurities are unknown (unidentified) or absent. In this case, it is expedient to use special experiments involving modification of the parent compound (sometimes called stress testing [, 15, 7]), whereby a solution of the given drug (or of the parent compound) is subjected to factors leading to its partial destruction with the formation of related identifiable compounds. The degree of decomposition can be readily determined from the decrease in the area of the main peak in the chromatogram of the final solution relative to that in the initial solution. The specificity of the given analytical procedure is judged by separation of the peaks of impurities between themselves and from the peak of the main component and by the peak purity of the parent compound. In particular, the specificity of HPLC procedures can be proved using the following methods of chemical modification. (i) Hydrolysis with 0.1 N solutions of HCl or NaOH at room temperature or at elevated temperatures. Example: a granulate was treated with 0.1 N HCl solution for 5 h at 60 C, after which the sample was extracted according to the proposed procedure and analyzed by HPLC. (ii) Oxidation with a 3% hydrogen peroxide solution, 0.05 M iodine solution, etc. Example: captopril was partly oxidized with 0.05 M iodine solution (in order to obtain captopril disulfide the main technological impurity []), extracted according to the proposed procedure, and analyzed by HPLC. (iii) Thermal decomposition by heating to C. Example: a granulate was kept for 7 days at 80 C. The resulting solid products of decomposition can be stored and used as control substances in the tests for suitability of a chromatographic system. (iv) Photochemical decomposition under illumination (e.g., UV irradiation). (v) Chemical addition reactions, for example, drug bromination at multiple bonds. A mixture of the initial substance and the products of its chemical modification can be used for preparing solutions used for the evaluation of suitability of the chromatographic system. The duration of action used for the chemical modification of drugs is selected taking into account the following factors. (i) The peaks of the products of drug modification are to be clearly distinguishable in the chromatogram. Therefore, the treatment duration must be sufficient to provide for a not less than 10% decrease in the main peak height (area), which is proportional to the drug content. (ii) The peaks of the products of drug modification have to be sufficiently well separated from the main peak, but the main peak height (area) must be comparable with that in the initial test solution. According to [7], it is recommended that the main peak intensity would decrease by no more than 30%. However, this requirement is not as critical, since the stronger decomposition of the parent compound can be compensated by adding it in the necessary amount. (iii) It is desired that the percentage content of the products of drug modification determined by the method of internal normalization would be close to the level of maximum permissible content of a single impurity. Further steps in the proof of specificity of the analytical procedure in the case under consideration are the same as in case 1. The validation characteristics additionally include chromatograms obtained in the course of experiments on the chemical modification of the parent compound. Data presentation. The proof of the specificity of an analytical procedure is presented in the form of a set of chromatograms (see above) with discussion of the obtained results. These data are supplemented by the results of calculation of (i) the separation coefficient R s for two peaks of the most closely spaced components, (ii) the column efficiency N, and (iii) the asymmetry parameters (tailing factors) of the main analytical peaks. It is necessary to make the following remark. Solving the problem of detection and separation of impurities by HPLC is still an art, with the results depending on the skill of developers. It is very important to select the optimum chromatographic columns and mobile phases. Moreover, it is known that, depending on the column loading, it is possible (a) to observe no additional peaks of impurities, (b) to find a certain number of such peaks, a part of which cannot be quantitatively characterized, or (c) to reveal a very large number of additional measurable peaks upon overloading the column with respect to the main drug component. Therefore, in developing and validating the methods of impurity determination (especially in the case of parent compounds), it is expedient to check for the possible presence of additional peaks (i.e., impurities) by chromatography of drug solutions with elevated concentrations providing overloading of the column with respect to the main component. However, this approach is usually inapplicable in the case of ready-to-use drugs containing large amounts of auxiliary substances.

5 16 N. A. Épshtein 1.. Confirming the Precision of a Given Analytical Procedure The precision (repeatability, reproducibility) of an analytical procedure characterizes the random scatter (variation) of the results relative to the mean value. It should be emphasized that, in order to obtain reliable results, the sample must have a homogeneous composition. The results have to be statistically treated. Variants leading to rough errors (e.g., using variation range R or the three sigma rule [8]) should be rejected. In evaluating the suitability of HPLC procedures (actually, in determining the metrological characteristics), it is necessary to use measuring vessels of class A or special calibrated measuring vessels and take all other possible measures to increase the reproducibility and accuracy of HPLC measurements [7]. The precision (repeatability, reproducibility) of an analytical procedure is evaluated in terms of the standard deviation (SD) of the relative standard deviation (percentage RSD) determined in a series of measurements and calculated by the formulas m SD ( X i X) ( m1 ), i1 SD RSD(%) 100. (1) X According to the ICH recommendations [4] on the validation of chromatographic procedures, the characteristics of precision are considered at three levels: repeatability, intermediate precision, and reproducibility (Table 3). In [4, 7] and in many other papers, the set of validation characteristics of HPLC procedures includes the injection repeatability. However, the author believes that this is incorrect: this parameter characterizes the quality of injector (e.g., syringe) rather than the suitability of a proposed procedure. TABLE 3. Main Precision Characteristics for the Validation of HPLC Procedures According to ICH [4] Precision characteristic Conditions of determination Repeatability (Rt ) Intermediate precision (Ip ) Reproducibility (Rp ) Injection repeatability Determined for the same sample preparation by the same analyst using the same instrument (chromatograph) during a short period of time Intra-assay precision Determined for the same sample preparation by different analysts using various instruments (chromatographs) during a prolonged period of time (not less than two days) The same, in various laboratories This characteristic is not included in the list of CDER [15] and USP-6 [7]. In HPLC validation, data on the injection repeatability should be provided as additional material required in cases of search for the bottleneck of a proposed method (dispersion analysis) Injection repeatability and intra-assay precision. In the general case, repeatability (R p ) characterizes the reproducibility of a given analytical procedure for the same sample preparation, as performed by the same analyst using the same instrument (chromatograph) during a relatively short period of time. For evaluating the repeatability in a given laboratory, the same analyst prepares samples of a model mixture or the same batch of a parent compound or a drug: (a) not less that nine samples of solutions covering the rated range of concentrations. For example: a homogenized powder of triturated tablets from the same batch is used to prepare drug solutions with concentrations equal to 50, 100, and 150% of the reference sample solution concentration according to the proposed procedure, or (b) not less than six samples of solutions in the region of concentrations close to the nominal value. For example: a homogenized powder of a parent compound or triturated tablets from the same batch is used to prepare six solutions with nominal concentration according to the proposed procedure. Note that each sample solution is (i) prepared independently of the other solutions and (ii) chromatographed at least three times. Each solution is characterized by the drug content X i (i =1,,N ), the average value X X i N, the standard deviation SD, the relative standard deviation RSD of particular measurements, and the confidence interval (for P = 95%) of the average value. It is required to show that the average results are statistically equivalent (e.g., in terms of the Student t-criterion) or, which is more convenient for the practical analysis, that RSD 1.0% for determination of parent compounds, RSD.0% for drugs, or RSD 10.0% for impurities [13, 7] Intermediate precision. This value characterizes the reproducibility of results obtained in the same laboratory by different analysts using various instruments (chromatographs) during a prolonged period of time (not less than two days) for the same homogeneous sample or a model drug mixture according to the proposed analytical procedure. Typically, not less than six solutions are prepared with concentrations close to the nominal value (see the preceding section). Each sample solution is (i) prepared independently of the other solutions and (ii) chromatographed at least three times. 5 For small values of the standard deviation (SD << 1), the t-criterion may give statistically significant differences even for close (almost identical) values of the compared average concentrations. This is related to the fact that this criterion is proportional to the ratio of systematic and random errors.

6 Validation of HPLC Techniques for Pharmaceutical Analysis 17 These solutions are characterized by the average drug content according to the results obtained by each of the analysts, X i and X j (i, j =1,,N ). These data are statistically processed and characterized by generalized average values X1 X i N and X X j N and the corresponding standard deviations (SD i, SD j ) and relative standard deviations (RSD i, RSD j ) of particular measurements. First, it is required to show that the proposed procedure of determination of the drug content and the impurity coincentraiton provides for the statistically equivalent standard deviations SDi and SD j of the results obtained by different analysts (in terms of the Fisher F-criterion). Then, it is necessary to demonstrate that the average results of these (for certainty, two) analysts are statistically reliably (P = 95%) identical in terms of the t-criterion calculated as t X 1 X X 1 X SD1 SD SD SD m m m 1 1, where X 1, X are the average results of analyses performed by analysts 1 and and SD 1,SD are the standard deviations in the particular series of m 1 and m parallel determinations (usually m 1 = m = m ). This t value is compared to tabulated values of the Student criterion t (P = 95%, f = m 1 + m ), where P = 95% is the confidence probability and f = m 1 + m is the number of degrees of freedom. If the calculated parameter t is lower than the tabulated value, the difference of average values can be considered as statistically insignificant with a 95% confidence probability. Otherwise, the average results differ to a greater extent than that admitted by random errors in both series [35]. In pactice, validation of the procedures of determination of the content of impurities is sometimes performed using a less strict method [13], by showing that the scatter (RSD) of the results of one analyst (characterized by a greater standard deviation (SD i or SD j ) relative to the average result of another analyst (with a lower SD i or SD j ) does not exceed a certain preset level, for example, so that RSD 10% for impurities with a rated content up to 1%, RSD 5% for an impurity content within 0.1 1%, and RSD 50% for an impurity level below 0.1%. It should be noted that, according to USP-6 [7], it is in most cases sufficient to determine only the repeatability for proper validation of an analytical procedure, while the intermediate precision and reproducibility characteristics should be determined for procedures included in the pharmacopoeial articles Reproducibility. This characteristic is determined by comparing the results obtained upon analysis of the same samples in different laboratories using a proposed analytical procedure. The necessary statistical methods are described in monograph [35, Chapter 8.4] and in the State Standard GOST R ISO It should be noted that for reliable evaluation of the statistical significance of the difference between the results obtained in such investigation, it is necessary that the round robin tests involve not less than five laborqtories [35]. In practice, however, the reproducibility is usually evaluated using two or three laboratories and characterized by less strict estimates. For example, validation of a procedure proposed for the quantitative determination of a parent compound is performed by demonstrating the statistical equivalence of the standard deviations SD i and SD j of the results obtained in different laboratories (in terms of the Fisher F-criterion). Then, it is demonstrated that the scatter (RSD) of the results of analyses in one laboratory (characterized by the maximum standard deviation (SD i or SD j ) relative to the average results of analyses in other laboratories (with lower SD i or SD j values) does not exceed a certain preset level [13]. The full-scale reproducibility of analytical procedures is rarely validated because (i) it is necessary to involve certified laboratories capable of reproducing the proposed procedure with high precision and (ii) this requires high organizational facilities and expenditures Confirming Linearity of the Response to Drug Concentration Linearity characterizes the ability of a proposed analytical procedure to give (within the suitability range) a response signal with the magnitude Y (e.g., peak height or area) directly proportional to the amount C (concentration) of a drug to be determined: Y = a + bc. According to ICH recommendations [3], the linearity in practice is first visually estimated from the linear appearance of the plot of Y versus C. Ifthe plot appears linear, this relation is studied by methods of regression analysis in terms of the linear equation Y = a + bc. For the analytical procedures for determining the content of a parent compound, CDER recommends establishing the criterion of linearity at a level of the correlation coefficient r not lower than [4]. However, even such a high level of correlation may be accompanied by significant deviations from linearity in the regions of high and low drug concentrations [13]. For this reason, ICH [3] recommends that the linearity be validated by a plot of the difference Y C showing deviations (residuals) of the calculated values y i = a + bc from the measured Y i values as the function of the concentration C i. The outbursts of the points (x i, y i ) relative to the regression model can be determined by calculating the parameter t using the formula [36] t SD 0 y Y i 1 ( Yi y) 1 N ( N 1) SD Here, y i and Y i are the calculated and experimentally measured values of the response, respectively; y = y i /N; N is the total number of experimental points (x i, y i ), and i y,

7 18 N. A. Épshtein ( yi Yi) SD 0 N ( yi y), SD y. N 1 The calculated t value is compared to tabulated values of the Student criterion t (P = 95%, f = N - ). If the calculated parameter is greater than the tabulated value, the given point can be considered as deviating from the adopted regression model with a 95% confidence probability. In practice, validation of the procedures of determination of the content of impurities with respect to linearity is sometimes performed proceeding from a correlation coefficient of r 0.98 [13]. According to our estimates, use of the calibration graphs with such correlation coefficients may lead to RSD values on the order of 0% and above. Therefore, it would be more correct to establish the criterion of linearity at least at the level of r The linearity should be validated based on the analysis of at least five solutions with various concentrations covering the entire suitability range of a proposed analytical procedure [35]. According to ICH recommendations [3], the linearity can be demonstrated directly by using the reference parent substance (dilutions of a standard solution) andor model artificial mixtures including components of the drug studied. The most adequate approach consists in taking thoroughly weighed aliquots of the drug components and preparing solutions according to the proposed procedure, since all operations of the analyst should correspond strictly to those stipulated in the description. In practice, however, an intermediate approach recommended by ICH [3] is frequently employed. According to this, solutions are partly prepared using weighed aliquots of the drug components and the other are obtained by diluting these stock solutions. It should be emphasized that the linearity of a proposed analytical procedure should be confirmed in the course of validation of the accuracy, which reduces expenditures and saves time. The usual procedures are as follows. (a) For the analysis of parent compounds, it is common practice to prepare a reference solution of the compound with a concentration at or above the upper limit of the expected concentration interval (suitability range of the proposed procedure). Then, a series of dilutions is prepared so as to cover the entire range. Each solution is studied in a series of two or three injections. (b) For the analysis of ready-to-use drugs, the linearity is frequently checked in the same way as for the parent compound (i.e., using solutions of the parent compound as described in (a)). However, it is incorrect to ignore the possibility that auxiliary components (placebo) may influence the results. Therefore, it is more correct to validate the linearity using model mixtures of the parent compound and placebo. (c) For the determination of impurities using the method of internal normalization of the peak areas or heights, it is possible to evaluate the linearity by preparing dilutions of the reference sample solution (with a concentration equivalent to the rated value) in the model impurity solution so as to obtain drug concentrations in the range from 0.05% (dilution by a factor of 000!) to.5%. Instead of making some dilutions, it is possible to use a proportional decrease in the volume of applied sample (which saves the mobile phase). Data presentation. The validation characteristics include (i) a regression equation of the S i = a + bc type (for example, S = C [mgml]), (ii) the correlation coefficient (e.g., r = ), and (iii) a plot visually confirming the linearity of relationship between S and C Confirming the Accuracy of a Given Analytical Procedure The accuracy characterizes the proximity of the experimental results, obtained using a proposed analytical procedure, to the true value in the entire suitability range of this procedure. The accuracy represents a combination of the random and systematic error. 6 In order to provide for accurate HPLC determinations, it is recommended to use standard solutions with concentrations close to within 10% of the test solution concentration. The accuracy of analytical procedures should be determined using homogeneous samples with exactly known concentrations of the compounds to be determined. For validation purposes a series of such solutions is prepared using the reference parent compound. According to ICH recommendations and USP-6 [1, 7, 15], the accuracy can be expressed both in the classical form, as the difference X between the average experimental value (X ) and the true value () with the corresponding confidence interval X, 7 ( X ) X, () and in an alternative (and more illustrative) form, in terms of the percentage recovery of the known amount of the compound to be determined, ( found content) R 100 %. (3) ( introduced content) The author believes that the content recovery testing should be preferred for evaluating the accuracy. This approach provides a more illustrative characteristic of the reliability of results obtained using a proposed analytical procedure and reveals the need for additional checks in the case of a significant systematic error (see below). On the other hand, the recovery defined by formula (3) incompletely characterizes the accuracy, since this quantity is also random and requires knowledge of the corresponding confidence interval. Thus, it is recommended to determine both the recovery R (%) and the confidence interval at a preset probability (P = 95%), representing the accuracy in a form analogous to expression (): R R. (4) Obviously, the proposed methods should not involve significant systematic errors. In the absence of systematic er- 6 Accordiong to the State Standard GOST R ISO the systematic error usually characterizes trueness. 7 The value of X characterizes random errors.

8 Validation of HPLC Techniques for Pharmaceutical Analysis 19 rors, the error is determined by the precision. Based on the permissible RSD values (see above), experimental experience, and analysis of the published data [13, 7], it is possible to draw the conclusion that there is no need to verify a proposed analytical procedure in the absence of a significant systematic error, provided that it is established that the recovery with allowance of the confidence interval does not fall outside the following limits: %, for the quantitative analysis of parent substances with a high rated content of the active component (98% and above); %, for the quantitative analysis of parent substances with a lower rated content of the active component (98% and below) and ready-to-use drugs; %, for the quantitative determination of impurities with a rated maximum content of up to 1%; 75 15%, for the quantitative determination of impurities with a rated maximum content from -0.1 to 1%; %, for the quantitative determination of impurities with a rated maximum content below 0.1% The procedure of accuracy evaluation. ICH recommends making three determinations (i.e., analyze three model mixtures) for three different concentrations. However, this approach does not allow the accuracy to be determined together with linearity and other validation characteristics. The author believes that the accuracy should be evaluated using not less than nine determinations at various concentrations covering the entire range of suitability of the proposed procedure. This provides for the possibility of determining this characteristic simultaneously with the calibration graph parameters and their statistical characteristics (for evaluating the linearity according to Section 1.3), the limit of quantitation (Section 1.7.), and the limit of detection (Section 1.6.). It should be emphasized that these determinations should include all stages of the proposed analytical procedure. For parent substances, the accuracy of analysis is usually determined by comparison to a reference sample. According to this, the reference sample (or a high-purity substance) is analyzed using the standard procedure and the results are compared to data in the certificate of the reference sample or to the results of analysis of the high-purity parent substance performed by an alternative method (e.g., titration) with known accuracy and precision. It is recommended that, for parent substances with a high rated content of the active component, the average recovery should be not less than % at each level [13]. For ready-to-use drugs, the accuracy is evaluated through the analysis of mixtures containing known amounts of the parent compounds and placebo; for the quantitative determination of impurities, this characteristic is determined by the analysis of such mixtures containing known amounts of these impurities. These analyses are performed using two principal methods. The method of recovery of a parent compound introduced into the placebo (matrix). This method, used for the analysis of drugs comprising mixtures of parent and auxiliary compounds, is based on determining the recovery of a known amount of the parent substance introduced into the placebo. The placebo is prepared separately and then introduced in a nominal (or proportional) amount into measuring flasks. Then, thoroughly weighed amounts of the parent compound or its concentrated solutions are added so that (upon filling the flasks to the marks) the sample concentrations would cover the entire expected suitability range of the proposed analytical procedure. For example, a parent compound can be introduced into a placebo solution at a level of 80, 100, and 10% of the nominal concentration indicated on the label (or the reference solution concentration). The method of standard additives. This method is generally analogous to that described above but is applied only when it is impossible to prepare a solution of placebo free from the parent compound or when this compound is present in the placebo in an unknown amount (e.g., in biological samples). In these cases, the reference sample of the parent compound is added at an amount of 50, 80, 100, 10, and 150% of its expected content in the analyzed solution. Using the proposed procedure, the amount of the parent compound is determined (found content) and compared to the known additive (introduced content). Alternatively, it is possible to compare the results of analyses performed using the proposed method and the data obtained for the same samples by validated alternative methods. For the validation of analytical procedures intended for the analysis of ready-to-use drugs, it is expedient to use the same reference substance for preparing both model mixtures and standard solutions. This eliminates errors related to the possible uncertainty of the composition indicated on the label of the reference sample and allows using commercial samples instead of special reference compounds. At a low concentration of the parent compound in the mixture (when it is impossible to add a thoroughly weighed amount of this compound), one may add a known amount of concentrated solution and then fill the measuring flask with a solvent stipulated by the proposed analytical procedure. For the quantitative determination of impurities, the accuracy of evaluation has certain peculiarities. In this case, the method of recovery of a parent compound introduced into the placebo and the method of standard additives have limited applicability because these validation procedures require large amounts of identified impurities. In this case, the accuracy is most frequently checked using the method described below. The method of internal normalization with or without response factors. According to this method, identified impurities are characterized by the response factors with respect to the parent compounds determined by the analytical procedure under consideration. These coefficients depend on the mobile phase composition and the analytical wavelength. For reproduction (or modification) of the analytical procedure, the detector wavelength is not changed, while the mobile phase composition can be corrected for the difference in the

9 0 N. A. Épshtein parameters of chromatographic columns. In the case of a considerable change in this composition (see Section 4), it is necessary to re-determine at least the values of the response factors. For determining unidentified and accidental impurities, these factors are conventionally taken equal to unity (assuming that the sensitivity for these impurities is the same as that for the parent compound). If the reference samples of impurities and decomposition products are unavailable, the validation has to be performed using alternative methods Testing for systematic error. The analytical procedure can be checked for the absence of systematic errors by one of the three methods considered below. Method based on the Student t-criterion without regression analysis [8]. Each sample (whose total number is N 5) with known values (introduced content) of the component to be determined is analyzed in m = 3 6 parallel determinations. The total data array is characterized by the dispersion (SD 0 ) and the Student criterion 8 x m t SD 0 This t value is compared to tabulated values of the Student criterion t (P, f = m 1). If the calculated parameter is greater than the tabulated value for P = 95% and f = m 1. t > t (P, f ), the results obtained using the proposed method can be considered as involving a systematic error. This error is calculated by the formula x 100 %. (5) The standard deviation SD 0 in a particular analysis is calculated using the set of all m parallel determinations performed for N (or g in the notation of [8]) samples. This allows the SD 0 value to be reduced and the sensitivity of determination of the systematic error to be increased with a simultaneous decrease in the confidence interval for the results of analysis, X = t SD 0 m, where t is the Student criterion for f = m 1 degrees of freedom. The algorithm of these calculations is as follows. Each k th sample is characterized by the deviation of the experimental value from average, di X i X, and the dispersion m SD k d i ( m 1). Then, the difference between the i1 maximum and minimum values of the dispersion SD k is checked to be insignificant in terms of the Fisher F-criterion. If this difference is actually insignificant, the SD 0 value is calculated as the sum of square deviations for all samples divided by the number of samples, 8 This t value is essentially the ratio of the systematic error x and the random error SD 0 m. SD 0 N SD k 1 N Finally, the Student t-criterion is calculated as k. t ( x m) SD 0 (6) and compared to tabulated values of the Student criterion t (P, f ) for f = N(m 1) degrees of freedom. It should be emphasized that, for small (practically insignificant) systematic error and small random error, the calculated t-criterion can be greater than the tabulated value. However, a small systematic error can be ignored in practice when the analytical problem does not require high accuracy of determination. This approach is also valid for other methods of evaluation of the accuracy of a proposed analytical procedure. Method of regression analysis with the Student t-criterion [35, 38]. This method seems to be the most effective, since it provides for the possibility of using the calibration graph for the accuracy evaluation and determination of some other validation characteristics (see the generalized scheme in Fig. 1). For simultaneous determination of the constant and variable systematic errors, not less than N = 5 samples with known values of the parent compound are studied and the relationship between the introduced content (m t ) and the found content (m f ) is processed by least squares in terms of the equation m f = a + bm i. Using these data, the parameters t a = a SD 0 and t b = 1 bsd 0 are calculated and compared to the critical (tabulated) values of the Student criterion t (P, f ) for the confidence probability P = 95% and f = N degrees of freedom. Using the results of regression analysis, it is possible to judge with 95% probability about the absence of a constant systematic error, provided that t a t (P, f = N ), and the absence of a linear variable systematic error, provided that t b t (P, f = N ). It is expedient to perform validation of the accuracy of an analytical procedure together with checking for the linearity of the system response (area or height of the peak) as a function of the concentration of the parent compound to be determined (in fact, linearity of the calibration graph in terms of the criteria described in Section 1.3) and with finding the limit of detection (Section 1.6.) and the limit of quantitation (Section 1.7.). Method of regression analysis with the Fisher F-criterion. This method is based on the assumption that a linear variable systematic error can be ignored. This assumption is justified because HPLC in pharmacy is used in a relatively narrow range of sample solution concentrations. The measurements are performed for not less than N 5 samples (model mixtures) with known values of the parent compound, after which the relationship between the introduced content (m t ) and found content (m f ) is processed

10 Validation of HPLC Techniques for Pharmaceutical Analysis 1 by least squares in terms of the relation m f = a + bm t. The absence of a constant systematic error is confirmed by the insignificance of the coefficient a [35] at a commonly accepted confidence probability level of P = 95%. For this purpose, the Fisher criterion calculated is as F( P, f1 N 1, f N ) SD 01( N 1) SD 0( N ), SD ( N ) 0 (7), where SD 01 and SD 0 are the standard deviations obtained for the above relations without the free term (m f = bm t ) and with the free term (m f = a + bm t ). If the calculated F value is smaller than the tabulated (critical) values F (P = 95%, f 1, f ), the free term a is in fact insignificant and the error is absent to within a 95% confidence probability Data presentation and evaluation of the systematic error. The final judgment about the accuracy of a proposed analytical procedure can be made upon validation of its specificity, precision (repeatability, reproducibility), and linearity. 9 It is necessary to indicate a particular method of normalization of the content of impurities (weight fractions, percentage of the area under the peak of the main component, etc.). The validation results are presented by data on the recovery of the amount of introduced parent compound with a confidence interval, R R, or the difference between the average experimental value X and the true value with the corresponding confidence interval: ( X ) X. The proposed procedure involves no significant systematic error, provided that the recovery with allowance of the confidence interval does not fall outside the following limits: %, for the quantitative analysis of parent substances with a high rated content of the active component (98% and above); %, for the quantitative analysis of parent substances with a lower rated content of the active component (98% and below) and ready-to-use drugs; %, for the quantitative determination of impurities with a rated maximum content of up to 1%; 75 15%, for the quantitative determination of impurities with a rated maximum content from 0.1 to 1%; %, for the quantitative determination of impurities with a rated maximum content below 0.1%. The numerical result of a particular analysis, X (or R), must contain the last significant digit in the same position as that in the numerical value of the error of determination, X (or R) [37]. The number of significant digits in the latter value is determined as follows. If the first significant digit in the error is 3, then X is expressed by a value with one significant digit; should the first significant digit in the error 9 It is expedient to confirm the linearity together with determining the accuracy. be < 3, then X is expressed by a value with two significant digits Validation of the Suitability Range of a Given Analytical Procedure The range of suitability of a given analytical procedure is the interval between minimum and maximum concentrations (amounts) of a compound to be determined in which (i) the linearity is observed, (ii) the characteristics of repeatability fall within permissible (preset) limits, and (iii) the accuracy is maintained at a sufficiently high level [1 3]. This interval must contain all values of the concentrations (amounts), which can be encountered in the course of routine analyses. The range of suitability of a given analytical procedure is expressed in the same units as the results of analyses. It should be noted that, in practice, it is not necessary to determine the maximum possible range of suitability for an HPLC procedure. If it were necessary, this range could be determined using threshold RSD values obtained in the course of validation of the linearity and precision. For example, RSD must not exceed 3% for HPLC procedures aimed at determination of the parent compounds and 10% for procedures of impurity determination [13]. In practice, it is sufficient to show that a given range of suitability covers with margin the rated limiting concentrations of the substances to be determined, as indicated in the corresponding pharmacopoeial articles. For this reason, it is recommended that the range of suitability of a given analytical procedure be not less than the following intervals [3, 7]. (i) For the quantitative determination of the main component concentration in parent compounds and ready-to-use drugs: from 80 to 10% of the nominal content (i.e., the concentrations of test solutions should range within 80 10% relative to the concentration of a reference sample solution used accordsing to the proposed procedure). (ii) For evaluation of the homogeneity of dosage: from 70 to 130% of the nominal content, provided that a wider interval is not required (in special cases such as aerosols). (iii) For dissolution tests: 0% (absolute percentage) of the rated value of drug release. For example, for monitoring the behavior of a drug with delayed release of a parent compound, for which the release is rated as 0% within the first hour and up to 90% within a 4-h period of time, the suitability range must extend from 0 to 110% of the nominal drug content. (iv) For the quantitative determination of impurities by the method of external standard: from 50 to 10% of the nominal content. It should be noted that, in view of the possibility of RSD values amounting up to 50% (Section 1..), it would be more correct to establish the upper limit at 150% of the nominal content. For impurities exhibiting a very high biological activity, toxicity, or unpredictable behavior, the limits of detection and quantitation must correspond to the level at which these impurities have to be controlled. 10 (v) In cases where the quantitative determination of a parent compound and the detection of impurities are per-

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination

More information

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 Simultaneous determination of Telmisartan

More information

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,

More information

SPECIFICATIONS AND CONTROL TESTS ON THE FINISHED PRODUCT

SPECIFICATIONS AND CONTROL TESTS ON THE FINISHED PRODUCT SPECIFICATIONS AND CONTROL TESTS ON THE FINISHED PRODUCT Guideline Title Specifications and Control Tests on the Finished Product Legislative basis Directive 75/318/EEC as amended Date of first adoption

More information

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE Rajgor VM,, 2014; Volume 3(3): 188-197 INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE ANALYTICAL METHOD DEVELOPEMENT AND VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF MEMANTINE HCL AND

More information

Implementing New USP Chapters for Analytical Method Validation

Implementing New USP Chapters for Analytical Method Validation Implementing New USP Chapters for Analytical Method Validation March 2010 Ludwig Huber Fax.: +49 7243 602 501 E-mail: Ludwig_Huber@labcompliance.com Today s Agenda Handling Method Changes vs. Adjustments

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 204 3R HANDLING AND USE OF REFERENCE STANDARDS IN THE OMCL NETWORK Full document title and reference Document type Legislative

More information

Design of Experiments for Analytical Method Development and Validation

Design of Experiments for Analytical Method Development and Validation Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

4.2 Bias, Standards and Standardization

4.2 Bias, Standards and Standardization 4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard

More information

Assay Development and Method Validation Essentials

Assay Development and Method Validation Essentials Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation

Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation IJPAR Vol.4 Issue 4 Oct Dec - 2015 Journal Home page: ISSN: 2320-2831 Research Article Open Access Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation V.

More information

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

General and statistical principles for certification of RM ISO Guide 35 and Guide 34 General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in

More information

IMPURITIES IN NEW DRUG PRODUCTS

IMPURITIES IN NEW DRUG PRODUCTS INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE IMPURITIES IN NEW DRUG PRODUCTS Q3B(R2) Current

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number

More information

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. ludwig_huber@labcompliance.com Overview GMP requirements for Quality Control laboratories

More information

Performance and advantages of qnmr measurements

Performance and advantages of qnmr measurements Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended

More information

Statistical estimation using confidence intervals

Statistical estimation using confidence intervals 0894PP_ch06 15/3/02 11:02 am Page 135 6 Statistical estimation using confidence intervals In Chapter 2, the concept of the central nature and variability of data and the methods by which these two phenomena

More information

Method Validation. Introduction. Página Web 1 de 16

Method Validation. Introduction. Página Web 1 de 16 Página Web 1 de 16 Method Validation Introduction Strategy for Validation of Methods Validation of Standard Methods Revalidation Parameters for Method Validation Selectivity/specificity Precision and Reproducibility

More information

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in

More information

Applying Statistics Recommended by Regulatory Documents

Applying Statistics Recommended by Regulatory Documents Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services steven@statisticaloutsourcingservices.com 301-325 325-31293129 About the Speaker Mr. Steven

More information

11.I In-process control Authors: Dr. Christian Gausepohl / Paolomi Mukherji / Update 07

11.I In-process control Authors: Dr. Christian Gausepohl / Paolomi Mukherji / Update 07 In-process control In-process control Authors: Dr. Christian Gausepohl / Paolomi Mukherji / Update 07 Here you will find answers to the following questions: What are the in-process control tasks? Where

More information

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection 52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis

More information

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax. Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative

More information

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS 1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Frank Tontala, Thermo Fisher Scientific, Germering, Germany Technical Note 708 Key Words Chromeleon Chromatography Data System,

More information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare

More information

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this

More information

Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant

Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant 2 USP Stimuli Article Lifecycle Management of Analytical Procedures: Method Development, Procedure

More information

The Effect of Temperature on Conductivity Measurement

The Effect of Temperature on Conductivity Measurement The Effect of on Measurement Authors: John J. Barron & Colin Ashton Technical Services Department, Reagecon Diagnostics Ltd, Shannon Free Zone, County Clare, Ireland Abstract The conductivity of all solutions

More information

ERDOSTEINE - MONOGRAPH.

ERDOSTEINE - MONOGRAPH. STRUCTURAL FORMULA (±1S-(2-[N-3-(2-oxotetrahydro thienyl)]acetamido)-thioglycolic acid) C 8 H 11 NO 4 S 2 M.W. = 249.307 DESCRIPTION Color : White to ivory white Appearance : Microcrystalline powder SOLUBILITY

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

APPENDIX N. Data Validation Using Data Descriptors

APPENDIX N. Data Validation Using Data Descriptors APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

To avoid potential inspection

To avoid potential inspection Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become

More information

Uses of Derivative Spectroscopy

Uses of Derivative Spectroscopy Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative

More information

COMMITTEE FOR PROPRIETARY MEDICINAL PRODUCTS (CPMP)

COMMITTEE FOR PROPRIETARY MEDICINAL PRODUCTS (CPMP) European Medicines Agency Inspections London, 17 December 2003 CPMP/QWP/122/02, rev 1 corr COMMITTEE FOR PROPRIETARY MEDICINAL PRODUCTS (CPMP) GUIDELINE ON STABILITY TESTING: STABILITY TESTING OF EXISTING

More information

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11. QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition

More information

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness

More information

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so

More information

High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage Form

High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage Form ISSN: 0973-4945; CODEN ECJHAO E- Chemistry http://www.e-journals.net Vol. 5, No.3, pp. 427-430, July 2008 High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage

More information

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas

More information

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND #3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric

More information

Chapter Test B. Chapter: Measurements and Calculations

Chapter Test B. Chapter: Measurements and Calculations Assessment Chapter Test B Chapter: Measurements and Calculations PART I In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1.

More information

Analytical Chemistry Lab Reports

Analytical Chemistry Lab Reports Analytical Chemistry Lab Reports Format and Calculations John Collins collinj2@gmail.com Measurement Analytical chemistry is entirely about measurement, what these measurements signify, and the understanding

More information

OLIVÉR BÁNHIDI 1. Introduction

OLIVÉR BÁNHIDI 1. Introduction Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

A Beer s Law Experiment

A Beer s Law Experiment A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining

More information

Chapter 1: Chemistry: Measurements and Methods

Chapter 1: Chemistry: Measurements and Methods Chapter 1: Chemistry: Measurements and Methods 1.1 The Discovery Process o Chemistry - The study of matter o Matter - Anything that has mass and occupies space, the stuff that things are made of. This

More information

Validating Methods using Waters Empower TM 2 Method. Validation. Manager

Validating Methods using Waters Empower TM 2 Method. Validation. Manager Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information

More information

Chemistry 111 Laboratory Experiment 7: Determination of Reaction Stoichiometry and Chemical Equilibrium

Chemistry 111 Laboratory Experiment 7: Determination of Reaction Stoichiometry and Chemical Equilibrium Chemistry 111 Laboratory Experiment 7: Determination of Reaction Stoichiometry and Chemical Equilibrium Introduction The word equilibrium suggests balance or stability. The fact that a chemical reaction

More information

Method Development and Validation for Particle Size and Shape Measurements

Method Development and Validation for Particle Size and Shape Measurements Method Development and Validation for Particle Size and Shape Measurements Ulf Willén Divisional Product Manager Analytical Imaging Systems Malvern Instruments Ltd, Malvern, UK. FDA guidance: when should

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry General Medicine Jiřina Crkovská (translated by Jan Pláteník) 2010/2011 1 Spectrophotometry is one of the most widely used instrumental

More information

A commitment to quality and continuous improvement

A commitment to quality and continuous improvement Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous

More information

Alignment and Preprocessing for Data Analysis

Alignment and Preprocessing for Data Analysis Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise

More information

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation

More information

Introduction to Fourier Transform Infrared Spectrometry

Introduction to Fourier Transform Infrared Spectrometry Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,

More information

AP Physics 1 and 2 Lab Investigations

AP Physics 1 and 2 Lab Investigations AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks

More information

Quality. Now Certified to ISO 9001:2008

Quality. Now Certified to ISO 9001:2008 Quality Now Certified to ISO 90012008 Quality Policy It is Peptides International's goal is to achieve complete customer satisfaction by addressing customer needs and delivering what we promise. The company

More information

Analytical Methods for Cleaning Validation

Analytical Methods for Cleaning Validation Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2011, 3 (6): 232-239 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4

More information

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:

More information

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93)

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93) INTRODUCTION HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93) Headache, sore muscles, arthritis pain... How do you spell relief?

More information

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample

More information

QbD Approach to Assay Development and Method Validation

QbD Approach to Assay Development and Method Validation QbD Approach to Assay Development and Method Validation Thomas A. Little Ph.D. 11/105/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8) Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:

More information

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL DOCUMENT OIML D 18 Edition 2008 (E) The use of certified reference materials in fields covered by metrological control exercised by national services of legal metrology. Basic principles

More information

Unit 31 A Hypothesis Test about Correlation and Slope in a Simple Linear Regression

Unit 31 A Hypothesis Test about Correlation and Slope in a Simple Linear Regression Unit 31 A Hypothesis Test about Correlation and Slope in a Simple Linear Regression Objectives: To perform a hypothesis test concerning the slope of a least squares line To recognize that testing for a

More information

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview. Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide

More information

ISO 17025. How to Meet. Requirements for Method Verification. Prepared by:

ISO 17025. How to Meet. Requirements for Method Verification. Prepared by: How to Meet ISO 17025 Requirements for Method Verification Prepared by: AOAC INTERNATIONAL 481 N. Frederick Ave, Suite 500 Gaithersburg, MD 20877, USA www.aoac.org The Scientific Association Dedicated

More information

Evaluating Current Practices in Shelf Life Estimation

Evaluating Current Practices in Shelf Life Estimation Definition of Evaluating Current Practices in Estimation PQRI Stability Working Group Pat Forenzo Novartis James Schwenke Applied Research Consultants, LLC From ICH Q1E An appropriate approach to retest

More information

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1 Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification

More information

GUIDELINE ON ACTIVE PHARMACEUTICAL INGREDIENT MASTER FILE (APIMF) PROCEDURE 1 (The APIMF procedure guideline does not apply to biological APIs.

GUIDELINE ON ACTIVE PHARMACEUTICAL INGREDIENT MASTER FILE (APIMF) PROCEDURE 1 (The APIMF procedure guideline does not apply to biological APIs. 15 January 2007 GUIDELINE ON ACTIVE PHARMACEUTICAL INGREDIENT MASTER FILE (APIMF) PROCEDURE 1 (The APIMF procedure guideline does not apply to biological APIs.) TABLE OF CONTENTS 1 INTRODUCTION... 2 2

More information

STABILITY TESTING OF ACTIVE SUBSTANCES AND PHARMACEUTICAL PRODUCTS

STABILITY TESTING OF ACTIVE SUBSTANCES AND PHARMACEUTICAL PRODUCTS RESTRICTED STABILITY TESTING OF ACTIVE SUBSTANCES AND PHARMACEUTICAL PRODUCTS Discussions are currently ongoing with the WHO Eastern Mediterranean Region towards a synergistic approach in developing a

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 33 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,

More information

GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION

GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION International Document OIML 1 st CD GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION International Organization of Legal Metrology Contents 01 Scope 2 12

More information

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based

More information

BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use.

BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use. BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use. USP proposes the revision and development of a suite of plastic packaging system standards in the current issue of PF. General test chapter

More information

In mathematics, there are four attainment targets: using and applying mathematics; number and algebra; shape, space and measures, and handling data.

In mathematics, there are four attainment targets: using and applying mathematics; number and algebra; shape, space and measures, and handling data. MATHEMATICS: THE LEVEL DESCRIPTIONS In mathematics, there are four attainment targets: using and applying mathematics; number and algebra; shape, space and measures, and handling data. Attainment target

More information

FT-NIR for Online Analysis in Polyol Production

FT-NIR for Online Analysis in Polyol Production Application Note: 51594 FT-NIR for Online Analysis in Polyol Production Key Words Acid Number Ethylene Oxide FT-NIR Hydroxyl Value Polyester Polyols Abstract Hydroxyl value and other related parameters

More information

Analytical Procedures and Methods Validation for Drugs and Biologics

Analytical Procedures and Methods Validation for Drugs and Biologics Analytical Procedures and Methods Validation for Drugs and Biologics Guidance for Industry U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research

More information

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) October 2006 RESTRICTED LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) DRAFT FOR DISCUSSION World Health Organization 2006 All rights reserved. This draft is intended for

More information

3.4 TEST FOR BACTERIAL ENDOTOXINS. Final text for revision of The International Pharmacopoeia

3.4 TEST FOR BACTERIAL ENDOTOXINS. Final text for revision of The International Pharmacopoeia Document QAS/11.5 FINAL July 01 3. TEST FOR BACTERIAL ENDOTOXINS Final text for revision of The International Pharmacopoeia This monograph was adopted at the Forty-sixth WHO Expert Committee on Specifications

More information

99.37, 99.38, 99.38, 99.39, 99.39, 99.39, 99.39, 99.40, 99.41, 99.42 cm

99.37, 99.38, 99.38, 99.39, 99.39, 99.39, 99.39, 99.40, 99.41, 99.42 cm Error Analysis and the Gaussian Distribution In experimental science theory lives or dies based on the results of experimental evidence and thus the analysis of this evidence is a critical part of the

More information