MALDI-TOF/TOF MS Protocols

Size: px
Start display at page:

Download "MALDI-TOF/TOF MS Protocols"

Transcription

1 MALDI-TOF/TOF MS Protocols Objectives: 1. To be familiar with the basic operation of the MALDI-TOF-TOF mass spectrometer. 2. To use MALDI-TOF-TOF mass spectrometer for peptide mass fingerprinting, peptide sequencing, and in source decay sequencing of intact proteins. Protocols included: 1. Obtaining a peptide mass fingerprint with MALDI-TOF 2. ISD and T 3 sequencing 3. In-gel tryptic digest 4. Sample spotting with the dried droplet method 5. Ultraflex III Operation for Obtaining a PMF Introduction: MALDI refers to the ion source, i.e., the Matrix Assists in Laser Desorption/Ionization of the analyte. The matrix, is co-crystallized with the sample, it must absorb light maximally at 337 nm (wavelength of the laser), and must participate in proton transfer with the analyte. TOF refers to the mass analyzer. The mass analyzer separates ions based on their Time Of Flight. Flight time is a function of the mass to charge ratio (m/z) of the analyte. TOF-TOF means that this is a tandem mass spectrometer, so an ion can be selected for, fragmented, and mass analyzed again. Tandem mass spectrometry, MS/MS, can provide additional information for structural elucidation of the analyte in question. The Ultraflex III is a MALDI-TOF-TOF mass spectrometer that will be used in this laboratory (Figure 1). A suite of software is used with this instrument. Acquisition of mass spectra is done in Flex Control, spectrum processing and annotation in Flex Analysis, and additional data processing and data searching is done in BioTools and Sequence Editor (Bruker). 1

2 Figure 1: Schematic of the Ultraflex III MALDI-TOF-TOF. ( Bruker training presentation) A typical peptide mass fingerprint (PMF) work flow would involve digesting a target protein with trypsin, obtaining a mass spectrum of the peptides, and using those m/z values to search the databases for a match. This method relies solely on the peptide map obtained in the mass spectrum and not on the amino acid sequence. If cross-species identification is used during the data search, the sample must have sequence identity for a subset of the protein. For proteins with some post translational modification, a PMF can help with identifying those regions, but too many post translational modifications may result in no identification. Reviews are available (Gilany, et al. 2010; Henzel, et al. 2003). A significant PMF result requires a mass tolerance of approximately 1 Da and in a MALDI-TOF PMF experiment, a mass accuracy of better than 20 ppm (0.02 Da for an m/z of 1000) can be obtained. The mass spectrum is acquired as follows. The cocrystallized matrix-analyte is hit with the laser. Irradiation of the matrix and analyte causes a plume to form in which proton transfer occurs between the matrix and analyte, resulting in ionization. After a delay, the ions are accelerated by an electric field towards the TOF mass analyzer where ions of differing m/z are separated based on their flight time. This flight time is converted to mass values, which are often in the form of isotope patterns, due to the high resolving power of the instrument. This allows for determination of the exact mass and structural information. Resolution can be improved in several ways. Each way involves decreasing the velocity distribution of ions having the same m/z. First, the matrix-analyte crystals should be a flat layer. This in effect allows ions to have an equal starting point. The ion source is equipped with pulsed ion extraction, or delayed extraction of ions. After the plume of ions is formed, ion source plate 1 (IS/1) and ion source plate 2 (IS/2) are kept at the same voltage for a short period of time, allowing slower ions to catch up to the faster ions. After this delay, the voltage of IS/2 is dropped and ions are accelerated towards the detector. Adjustment of the delay time, lens voltage, and digitizing rate of the detector can improve the sensitivity and resolution of some analytes. 2

3 Figure 2: Pulsed ion extraction. (Bruker training presentation.) Operating in linear mode has the advantage of increasing sensitivity and is often used for molecular weight determination of proteins. Operating in reflector mode, while decreasing the sensitivity, drastically improves the resolution. Instead of hitting the linear detector, the ions are reflected in an ion mirror. This increases the flight time, and focuses the ions. The results of a PMF can be strengthened or confirmed through peptide sequencing. Tandem mass spectrometry can be performed in two ways with the Ultraflex III: laser induced dissociation (LID) for peptide mass fingerprinting, or collision induced dissociation (CID) for de novo sequencing because the high energy collisions with Argon gas produce fragment ions used in distinguishing between isobaric residues. The laser power is increased relative to MS so that more precursor ions per shot are obtained. The initial accelerating voltage is low relative to MS, allowing for a long flight time (10-20 μs). During this time, ion families are formed through metastable decay of parent ions. The ion family of interest passes through a voltage gate, the timed ion selector, while other ion families are deflected. The ion family is post-accelerated and focused in the lift cell. The post-lift metastable suppressor acts like the timed ion selector and deflects parent ions while fragment spectra are acquired. After focusing by the ion reflector, ions are detected by the reflector detector (Suckau et al. 2003). Peptide mass fingerprinting and identification by peptide sequencing are considered as bottom-up approaches. A top-down approach is available, where intact proteins are identified based on the sequencing of their N- and C-termini. ISD sequencing is performed as a MS experiment in positive reflector mode, using a method optimized for peptides. Instrument parameters are adjusted so that fragmentation of the termini can occur in the plume before ion extraction. The acceleration voltage is 25 kv, pulsed ion extraction time set at 200 ns laser power and the detector gain are increased. Unambiguous sequence assignment is attributed to the predictable fragmentation along the peptide backbone. Fragments from the N- terminus are predominantly c-ions, and fragments from the C- terminus are y- ions. ISD sequencing cannot be used for ions below 1000 m/z since this region of the mass spectrum is dominated by matrix peaks. T 3 sequencing, i.e., MS/MS of an ISD fragment ion, is performed for sequencing of the 3

4 terminal residues (Suckau and Reseman, 2003). Figure 3 shows a schematic of this approach. Processing methods perform peak picking, baseline subtraction, and smoothing operations. During peak picking, signal to noise ratio, relative intensity threshold, minimum intensity threshold, and maximum number of peaks are all taken into account. Three peak picking algorithms are available: SNAP, centroid, and sum. SNAP (Sophisticated Numerical Annotation Procedure) searches for isotope patterns in the spectrum and performs baseline correction and noise determination. The result is that the monoisotopic peak is identified for each isotope pattern. Identifying the monoisotopic peak is necessary for obtaining a PMF. are faster peak picking algorithms. Centroid and Sum calculations do not take the isotope pattern into consideration for peak picking. Centroid finds peaks based on first and second order derivative calculations. Centroid is used for proteins and the average peak, as opposed to the monoisotopic peak is calculated. The Sum algorithm is used in applications where speed is a key factor; peaks are picked based on a pseudo-derivative calculation. For Centroid and Sum methods, baseline subtraction and smoothing operations can be adjusted by the user as needed. Figure 3: ISD fragmentation provides the amino acid sequence near the termini. Each ISD fragment contains the actual terminus so one of these ions can be used to obtain the sequence of the terminus (Suckau and Reseman, 2003). The Mascot search engine (Matrix Science) is used to search available databases during protein identification experiments. The first step in PMF searching is to define the search. Figure 4 shows a search dialog box. Close attention should be paid to taxonomy, database, variable modifications, and mass tolerance. These search parameters can drastically affect the search results. Points on some of the search parameters are listed below. 4

5 Database: Some databases (e.g., SwissProt) are highly annotated and contain few errors, but are limited on the number of entries. Others (e.g. expressed sequence tag (EST) databases) have many more entries but contain errors, redundancies, and entries are not annotated. Variable vs. global modifications: Global refers to the known chemistry of the sample while variable refers to possible modifications. Missed cleavages: This refers to the completeness of the digest. For example, the tryptic peptide, GHMNIRFR, contains 1 missed cleavage. Mass tolerance: ppm is generally used for a MALDI-TOF PMF. If the mass tolerance is too restricted, the search may not result in a significant assignment. Data file or query data: This is automatically entered if going through BioTools. If searching from the Matrix Science website, it is entered manually. Mascot search results are given as a histogram of the top scores. Hits in the green region have a higher chance (5% or more) of being a false positive and hits outside of this region (more than 70) are considered significant. The score for the false positive likelihood is a based on the search conditions and the database used. Figure 4: Mascot search box for peptide mass fingerprinting. 5

6 The format of the results can be adjusted with the Format As dropdown box. The results can be viewed as groups of homologs, in an index, coverage of a specific protein, or as an error plot. The error is shown as the root mean square (RMS) error, a measure of the average mass error of the dataset. Figure 5: A histogram of the Mascot search results. Protocols 1. Obtaining a peptide mass fingerprint with MALDI-TOF Sample spotting: Spot the sample with α-cyano-4-hydroxycinnamic acid (HCCA) using the dried droplet method. The peptide digest should be at a concentration of 1 pmol/µl. Spot Peptide Mix II calibration standard. Acquisition method: RP_Proteomics_HPC is the FlexControl method. External calibration is performed with peptide calibration mix II. Data processing and analysis: The Protein Mass Fingerprint Flex Analysis Mass Spectrometry (PMF.FAMS) method, which uses the SNAP peak picking algorithm, is used for spectrum processing. Mascot search is performed from BioTools. Peptide sequencing with MALDI-TOF-TOF Sample spotting: Spot the sample with HCCA using the dried droplet method. The peptide digest should be at a concentration of 1 pmol/µl. No calibration standard is necessary. (Calibration is performed for the LIFT method approximately once a year with a different method than what is described here). Acquisition: A current LIFT method is used in FlexControl. Data Processing and analysis: SNAP_full_process method in FlexAnalysis is used for processing. The sequence can be annotated manually in Flex Analysis or a Mascot search can be performed in BioTools. 2. ISD and T 3 sequencing 6

7 Sample spotting: Use the dried droplet method to spot the pure, intact protein ( pmol/µl). 2,5-Dihydroxybenzoic acid (DHB) is recommended. ISD acquisition: In Flex Control, a peptide method is used. The low mass deflection set to 900 Da, detection gain to 20X, and pulsed ion extraction set to 200 ns. ISD data processing and analysis: The spectrum is processed in Flex Analysis with the PMF_FAMS method, and then it is sent to BioTools In the Analysis pull down menu, select Permanently assign as an ISD type 2. Switch to MS/MS analysis and choose Top Down 3. Click on Find Tags 4. Settings can be changed, but default settings are usually a good starting point. Click Create Sequence Tags 5. Sequence tags are listed in order of highest to lowest ranking. They are ranked based on Σ (peak area/number of peaks). The top five are automatically highlighted. Click Copy/MS BLAST, then Open MS BLAST, and paste into the provided field. Click Submit Query. 6. Press Open Hit in SE, and Sequence Editor will open with the sequence. 7. Truncate the sequence to send it to BioTools. ISD is used to sequence the termini of the protein, so use the first 100 of each terminus. Send these sequences to BioTools. 8. In BioTools, change the annotation parameters to ISD_extended. Work in the MSMS Fragments tab to check the mapping of the theoretical mass fragments to those in the experimental mass spectrum. T 3 sequencing acquisition: In Flex Control, use a LIFT method. Define an ISD fragment ion as the parent, detection gain may need to be increased and several thousands of shots are likely to be required since sensitivity is low for this method. Data processing and analysis: In Flex Analysis, process the spectrum with SNAP_full_process. The sequence can be annotated manually in Flex Analysis or a Mascot search can be performed in BioTools using Amide (C-term) as the variable modification. 3. In-gel tryptic digest Materials 55 mm iodoacetamide: Dissolve 10.2 mg of IAA to a final volume of 1.0 ml with 100 mm ammonium bicarbonate. 100 mm ammonium bicarbonate: Dissolve g of NH 4 HCO 3 in 50 ml with HPLC grade H 2 O. Dehydration solution (HPLC grade acetonitrile) Peptide extraction solution (50% acetonitrile, 5% formic acid) Protein sample embedded in gel plug 10 mm dithiothreitol (DTT): Dissolve 1.54 mg of DTT to a final volume of 1.0 ml with 100 mm ammonium bicarbonate. Trypsin solution (Promega sequencing grade modified trypsin (Cat. # V5111): One vial contains 20 μg of trypsin. Dissolve in 200 μl of reconstitution buffer (50 mm

8 acetic acid) provided with trypsin. Aliquot the trypsin into 50 μl aliquots and freeze at -80 C. Trypsin is inactive at ph 4.0 or below. Scale this so that the trypsin solution is only re-frozen less than 5 times. Procedure 1. Cut protein band from the gel as close to the stained band as possible. Place in an eppendorf tube and remove any excess liquid. 2. For gels stained with Coomassie Blue, destain the plug by dispensing 50 μl of 50 mm ammonium bicarbonate and 50 μl of acetonitrile. Incubate for 10 minutes at 37 C. Aspirate the solution and discard. Repeat until gel plug is clear. 3. Dehydrate the plug in 50 μl acetonitrile for 5 minutes at 37 C. Aspirate the solution, discard, and incubate the gel plug for 10 minutes at 37 C. It should turn opaque-white. 4. Reduce the protein in 50 μl of 10 mm DTT, for 20 minutes at 37 C. Aspirate the solution and discard. 5. Alkylate the protein in 30 μl of 55 mm iodoacetamide for 20 minutes at 37 C. Aspirate the solution and discard. 6. Dehydrate the gel plug (see step 3). 7. Activate the trypsin solution. Add 160 μl of 50 mm ammonium bicarbonate to a 50 μl aliquot of trypsin (aliquot is 100 μg/ml). Thus, 15 μl will contain 300 ng of trypsin. (The addition of ammonium bicarbonate activates the trypsin, which is maximally active in the ph range 7-9). 8. Add 15 μl of the active trypsin solution to the gel plug and incubate at room temperature for 10 minutes to allow the gel plugs to swell. 9. Add 15 μl of 50 mm ammonium bicarbonate and incubate for 4-16 hours at 37 C. 10. Add 40 μl of peptide extraction solution to halt digestion and extract peptide. Sonicate 2 times for 5 minutes each. Combine extracts in a tube. 11. Evaporate solution under SpeedVac for 1 h at RT. 12. Add 50 μl of 5 % formic acid to dissolve the peptide residue. Or add 0.1% triflouroacetic acid (TFA). 4. Sample spotting with the dried droplet method Materials Matrix (20 g/l HCCA (α-cyano-4-hydroxycinnamic acid), DHB (2,5-dihydroxybenzoic acid), or SA (sinnapinic acid) in acetonitrile: 0.1% trifluoroacetic acid (70:30 v/v)) Peptide/protein sample in 0.1% trifluoroacetic acid Calibration standard (PepmixII, ProtMixI or ProtMixII stored at -20 C in aliquots in 0.1% TFA) Ground Steel Target Plate Procedure 1. Dissolve HCCA. Sonicate for 5 minutes, then centrifuge for 5 minutes. 2. Apply 0.5 μl of the sample solution to a target spot. Overlay with 0.5 μl of the binary matrix and mix. Prepare, as in step 3, a calibration spot near the sample. 3. Allow samples to air dry or dry them under a gentle stream of N 2 gas. 8

9 5. Ultraflex III Operation for Obtaining a PMF 1. Turn on computer. 2. Double-Click Flex Control icon to open acquisition program. 3. Insert the target plate/holder into the instrument. First, press the bottom of the door, open the latch and gently place the target into the carrier. The correct orientation is for the 1 column and A row to be in the top left corner with the sample side facing towards the entrance to the MS room. 4. In Flex Control, you are prompted to select a Method. Choose the most recent RP_Proteomics_HPC method. (Reflector detector, Positive polarity) 5. Single-Click the Inject/Eject icon to insert the target. 6. Go to the Details tab. Record the number of laser shots taken so far in the instrument log book. 7. When the target is inserted, Single-Click on the spot containing the calibration standard. The spot will now be shown in the camera view. The cross-hairs designating where the laser will shoot will be in the center of the spot. 8. Go to the Calibration tab. From the dropdown list, select the calibrant Peptide Calibration Standard II Monoisotopic. 9. Decrease the laser intensity to about 5%. With the cursor in the vicinity of the camera viewer, the wheel can be used to change the laser intensity. Click Start and slowly increase the laser intensity until the calibration standard peaks can be seen. Once this laser intensity has been identified, increase laser a little past this threshold. Do not increase if the baseline drifts upward. This is a sign that laser intensity is too high. Excessive noise and low signal intensity is a sign that laser intensity is too low. 10. After the spectrum has been acquired, click Add. Several points of acquisition can be added to the sum buffer. To get rid of an acquisition point that has been added, click Undo. To get rid of the summed acquisitions, click Clear Sum. 11. Go to the Calibration tab. Click Automatic Assign. The mass tolerance is set to 500 ppm and uses a quadratic function by default. The automatic assign function will select the monoisotopic peaks in the spectrum that correspond to the expected m/z values. For it to be a good calibration, the standard deviation should be within +/- 5 ppm of the expected value. 12. For the calibration to take effect, click Assign. Then go to File à Save Method As and save the method under a new name, in your folder. This personal method can be saved over each time you calibrate. The default method cannot be saved over. The default method should be used each time you calibrate. 13. Click on the spot containing the analyte of interest using the Target Map Icon. Decrease the laser intensity again to about 5%. Click Start and slowly increase the laser intensity until you see the analyte signal. Click Add and collect about 400 laser shots in total. 14. Click Save As. You will be prompted to write any relevant notes, where the file should be saved to, and under what name. Put a tick mark in the box for Open in Flex Analysis. In the dropdown box for Processing Method choose PMF.FAMS. Click Save. 15. The window for Flex Analysis will automatically open. The summed spectra that were acquired will be processed and annotated within the mass range of m/z. 9

10 16. To eliminate background peaks, click on MassList/Filter Background Peaks. Choose the MassControlList appropriate to the sample. For example, trypsin autoproteolysis or contaminants like keratins. 17. Press the icon for Biotools. The processed spectrum will be sent to the Biotools program. 18. Press the icon MS to open the MASCOT search window. 19. Enter all relevant information into the MS Search Dialogue. 20. Click Start and wait for the MASCOT results. 21. To export the top hit from the MASCOT Search Results window, click Get Hits. Close the window. The results will be displayed in BioTools. 6. References Gilany, K., Moens, L., & Dewilde, S. (2010). Mass spectrometry-based proteomics in the life sciences: a review. Journal Of Paramedical Sciences, 1(1). Henzel, W. J., Watanabe, C., & Stults, J. T. (2003). Protein identification: the origins of peptide mass fingerprinting. Journal of the American Society for Mass Spectrometry, 14(9), Link, A. J. and LaBaer, J. (2009). Proteomics: A cold Spring Harbor Laboratory Manual. Cold Spring Harbor, New York: Cold Spring Harbor Laboratory Press. Suckau, D., Resemann, A., Schuerenberg, M., Hufnagel, P., Franzen, J., and Holle, A., A novel MALDI LIFT-TOF/TOF mass spectrometer for proteomics. Analytical and Bioanalytical Chemistry, 376(7): , August Suckau, D. and Resemann, A., T 3 -Sequencing: Targeted Characterization of the N- and C-termini of Undigested Proteins by Mass Spectrometry. Analytical Chemistry, 75(21): , November In-house protocol for protein analyses by LC-IT-MS Tutorials for BioTools, Version 3.1 (January 2007) 10

Application Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008

Application Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008 Bruker Daltonics Application Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008 In the ABRF-PRG study 2008 [*] the ability to characterize

More information

Choices, choices, choices... Which sequence database? Which modifications? What mass tolerance?

Choices, choices, choices... Which sequence database? Which modifications? What mass tolerance? Optimization 1 Choices, choices, choices... Which sequence database? Which modifications? What mass tolerance? Where to begin? 2 Sequence Databases Swiss-prot MSDB, NCBI nr dbest Species specific ORFS

More information

Tutorial for Proteomics Data Submission. Katalin F. Medzihradszky Robert J. Chalkley UCSF

Tutorial for Proteomics Data Submission. Katalin F. Medzihradszky Robert J. Chalkley UCSF Tutorial for Proteomics Data Submission Katalin F. Medzihradszky Robert J. Chalkley UCSF Why Have Guidelines? Large-scale proteomics studies create huge amounts of data. It is impossible/impractical to

More information

Master course KEMM03 Principles of Mass Spectrometric Protein Characterization. Exam

Master course KEMM03 Principles of Mass Spectrometric Protein Characterization. Exam Exam Master course KEMM03 Principles of Mass Spectrometric Protein Characterization 2010-10-29 kl 08.15-13.00 Use a new paper for answering each question! Write your name on each paper! Aids: Mini calculator,

More information

LABORATÓRIUMI GYAKORLAT SILLABUSZ SYLLABUS OF A PRACTICAL DEMONSTRATION. financed by the program

LABORATÓRIUMI GYAKORLAT SILLABUSZ SYLLABUS OF A PRACTICAL DEMONSTRATION. financed by the program TÁMOP-4.1.1.C-13/1/KONV-2014-0001 projekt Az élettudományi-klinikai felsőoktatás gyakorlatorientált és hallgatóbarát korszerűsítése a vidéki képzőhelyek nemzetközi versenyképességének erősítésére program

More information

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM It s just what you expect from the industry leader. The AB SCIEX 4800 Plus MALDI TOF/TOF

More information

An In-Gel Digestion Protocol

An In-Gel Digestion Protocol An In-Gel Digestion Protocol This protocol describes the digestion of a protein present in an SDS-PAGE gel band with trypsin. The band can be taken from either a 1D or 2D electrophoresis gel. Reagents

More information

ProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry

ProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry ProteinScape Proteomics Data Analysis & Management Innovation with Integrity Mass Spectrometry ProteinScape a Virtual Environment for Successful Proteomics To overcome the growing complexity of proteomics

More information

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests AB SCIEX TOF/TOF 5800 System with DynamicExit Algorithm and ProteinPilot Software for Robust Protein

More information

MASCOT Search Results Interpretation

MASCOT Search Results Interpretation The Mascot protein identification program (Matrix Science, Ltd.) uses statistical methods to assess the validity of a match. MS/MS data is not ideal. That is, there are unassignable peaks (noise) and usually

More information

Standard Mixture. TOF/TOF Calibration Mixture. Calibration Mixture 1 (Cal Mix 1, 1:10) Calibration Mixture 1 (Cal Mix 1, 1:100)

Standard Mixture. TOF/TOF Calibration Mixture. Calibration Mixture 1 (Cal Mix 1, 1:10) Calibration Mixture 1 (Cal Mix 1, 1:100) Mass Standards Kit for Calibration of AB SCIEX TOF/TOF Instruments Protocol 1 Product Description The Mass Standards Kit includes reagents needed to test instrument function, optimize instrument parameters,

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests Application Note: 386 Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests Rosa Viner, Terry Zhang, Scott Peterman, and Vlad Zabrouskov, Thermo Fisher Scientific, San Jose, CA,

More information

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start. Q-TOF User s Booklet Enter your username and password to login. After you login, the data acquisition program will automatically start. ~ 1 ~ This data acquisition window will come up after the program

More information

Aiping Lu. Key Laboratory of System Biology Chinese Academic Society APLV@sibs.ac.cn

Aiping Lu. Key Laboratory of System Biology Chinese Academic Society APLV@sibs.ac.cn Aiping Lu Key Laboratory of System Biology Chinese Academic Society APLV@sibs.ac.cn Proteome and Proteomics PROTEin complement expressed by genome Marc Wilkins Electrophoresis. 1995. 16(7):1090-4. proteomics

More information

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics

More information

TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature

TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature ProteoMass Peptide & Protein MALDI-MS Calibration Kit Catalog Number MSCAL1 Store at Room Temperature TECHNICAL BULLETIN Description This kit provides a range of standard peptides and proteins for the

More information

ProteinPilot Report for ProteinPilot Software

ProteinPilot Report for ProteinPilot Software ProteinPilot Report for ProteinPilot Software Detailed Analysis of Protein Identification / Quantitation Results Automatically Sean L Seymour, Christie Hunter SCIEX, USA Pow erful mass spectrometers like

More information

SpikeTides TM Peptides for relative and absolute quantification in SRM and MRM Assays

SpikeTides TM Peptides for relative and absolute quantification in SRM and MRM Assays Protocol SpikeTides TM Peptides for relative and absolute quantification in SRM and MRM Assays Contact us: InfoLine: +49-30-6392-7878 Order per fax: +49-30-6392-7888 or e-mail: www: peptide@jpt.com www.jpt.com

More information

flexanalysis User Manual

flexanalysis User Manual flexanalysis User Manual Version 3.0 (September 2006) Copyright Copyright 2006 Bruker Daltonik GmbH This document is published by Bruker Daltonik Bremen. All rights reserved. Reproduction, adaptation or

More information

Laboration 1. Identifiering av proteiner med Mass Spektrometri. Klinisk Kemisk Diagnostik

Laboration 1. Identifiering av proteiner med Mass Spektrometri. Klinisk Kemisk Diagnostik Laboration 1 Identifiering av proteiner med Mass Spektrometri Klinisk Kemisk Diagnostik Sven Kjellström 2014 kjellstrom.sven@gmail.com 0702-935060 Laboration 1 Klinisk Kemisk Diagnostik Identifiering av

More information

How Mascot Integra helps run a Core Lab

How Mascot Integra helps run a Core Lab How Mascot Integra helps run a Core Lab 1 Areas where a database can help a core lab Project, experiment and sample tracking Flexibility in experiment design Role based security Automation Custom results

More information

ProSightPC 3.0 Quick Start Guide

ProSightPC 3.0 Quick Start Guide ProSightPC 3.0 Quick Start Guide The Thermo ProSightPC 3.0 application is the only proteomics software suite that effectively supports high-mass-accuracy MS/MS experiments performed on LTQ FT and LTQ Orbitrap

More information

Mass Spectrometry Based Proteomics

Mass Spectrometry Based Proteomics Mass Spectrometry Based Proteomics Proteomics Shared Research Oregon Health & Science University Portland, Oregon This document is designed to give a brief overview of Mass Spectrometry Based Proteomics

More information

Error Tolerant Searching of Uninterpreted MS/MS Data

Error Tolerant Searching of Uninterpreted MS/MS Data Error Tolerant Searching of Uninterpreted MS/MS Data 1 In any search of a large LC-MS/MS dataset 2 There are always a number of spectra which get poor scores, or even no match at all. 3 Sometimes, this

More information

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,

More information

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080 Training on the ZQ Open access MS Questions? Contact Dr. Allis Chien allis@stanford.edu 650-723-0710 0710 STANFORD UNIVERSITY MASS SPECTROMETRY STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD

More information

Supplementary Materials and Methods (Metabolomics analysis)

Supplementary Materials and Methods (Metabolomics analysis) Supplementary Materials and Methods (Metabolomics analysis) Metabolite extraction and mass spectrometry analysis. 12 Vldlr-/- knock out and 12 wild type (WT) retinas were separately frozen at -80 ºC in

More information

Background Information

Background Information 1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software

More information

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data Automated Data Interpretation for Glycan Characterization Jenny Albanese 1, Matthias Glueckmann 2 and Christof Lenz 2 1 AB

More information

Proteomics in Practice

Proteomics in Practice Reiner Westermeier, Torn Naven Hans-Rudolf Höpker Proteomics in Practice A Guide to Successful Experimental Design 2008 Wiley-VCH Verlag- Weinheim 978-3-527-31941-1 Preface Foreword XI XIII Abbreviations,

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

Monoclonal Antibody Fragment Separation and Characterization Using Size Exclusion Chromatography Coupled with Mass Spectrometry

Monoclonal Antibody Fragment Separation and Characterization Using Size Exclusion Chromatography Coupled with Mass Spectrometry Monoclonal ntibody Fragment Separation and haracterization Using Size Exclusion hromatography oupled with Mass Spectrometry uthors Haiying hen Katherine McLaughlin Sepax Technologies, Inc. 5 Innovation

More information

Tutorial 9: SWATH data analysis in Skyline

Tutorial 9: SWATH data analysis in Skyline Tutorial 9: SWATH data analysis in Skyline In this tutorial we will learn how to perform targeted post-acquisition analysis for protein identification and quantitation using a data-independent dataset

More information

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,

More information

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers

More information

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11. QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition

More information

ProteinChip Energy Absorbing Molecules (EAM)

ProteinChip Energy Absorbing Molecules (EAM) ProteinChip Energy Absorbing Molecules (EAM) Instruction Manual Catalog #C30-00001, #C30-00002, #C30-00003, #C30-00004 For technical support, call your local Bio-Rad office, or in the US, call 1-800-4BIORAD

More information

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview

More information

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method Introduction During the last decade, the complexity of samples

More information

Mass Frontier 7.0 Quick Start Guide

Mass Frontier 7.0 Quick Start Guide Mass Frontier 7.0 Quick Start Guide The topics in this guide briefly step you through key features of the Mass Frontier application. Editing a Structure Working with Spectral Trees Building a Library Predicting

More information

泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics

泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics 泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics 2014 Training Course Wei-Hung Chang ( 張 瑋 宏 ) ABRC, Academia

More information

Global and Discovery Proteomics Lecture Agenda

Global and Discovery Proteomics Lecture Agenda Global and Discovery Proteomics Christine A. Jelinek, Ph.D. Johns Hopkins University School of Medicine Department of Pharmacology and Molecular Sciences Middle Atlantic Mass Spectrometry Laboratory Global

More information

Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)

Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III) Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III) Purpose: Verify the synthesis of the synthesis product, Cr(acac) 3. The presence of any urea containing byproducts will also be determined.

More information

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract

More information

Micromass LCT User s Guide

Micromass LCT User s Guide Micromass LCT User s Guide 1) Log on to MassLynx with your username & password. 2) After you have logged in, the MassLynx software will automatically run. 3) After MassLynx has come up, open your project

More information

MS Amanda Standalone User Manual

MS Amanda Standalone User Manual MS Amanda Standalone User Manual February 2014 Viktoria Dorfer, Peter Pichler, Stephan Winkler, and Karl Mechtler 1. Installation of MS Amanda Standalone In order to install MS Amanda Standalone please

More information

[ Care and Use Manual ]

[ Care and Use Manual ] PREP Calibration Mix DIOS Low i. Introduction Pre-packaged PREP Calibration Mixtures eliminate the need to purchase and store large quantities of the component calibration reagents, simplifying sample

More information

Data mining with Mascot Integra ASMS 2005

Data mining with Mascot Integra ASMS 2005 Data mining with Mascot Integra 1 What is Mascot Integra? Fully functional out-the-box solution for proteomics workflow and data management Support for all the major mass-spectrometry data systems Powered

More information

Agilent G2721AA/G2733AA Spectrum Mill MS Proteomics Workbench

Agilent G2721AA/G2733AA Spectrum Mill MS Proteomics Workbench Agilent G2721AA/G2733AA Spectrum Mill MS Proteomics Workbench Application Guide Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any

More information

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,

More information

Protein Prospector and Ways of Calculating Expectation Values

Protein Prospector and Ways of Calculating Expectation Values Protein Prospector and Ways of Calculating Expectation Values 1/16 Aenoch J. Lynn; Robert J. Chalkley; Peter R. Baker; Mark R. Segal; and Alma L. Burlingame University of California, San Francisco, San

More information

Principles and Applications of Proteomics

Principles and Applications of Proteomics Principles and Applications of Proteomics Why Proteomics? 2-DE Overview Sample preparation 1 st & 2 nd dimension seperation Data Analysis Sample preparation for Mass Spectrometry Mass Spectrometry MALDI-TOF,

More information

13C NMR Spectroscopy

13C NMR Spectroscopy 13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number

More information

Application of Structure-Based LC/MS Database Management for Forensic Analysis

Application of Structure-Based LC/MS Database Management for Forensic Analysis Cozette M. Cuppett and Michael P. Balogh Waters Corporation, Milford, MA USA Antony Williams, Vitaly Lashin and Ilya Troisky Advanced Chemistry Development, Toronto, ntario, Canada verview Application

More information

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10

More information

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines

More information

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Preprocessing, Management, and Analysis of Mass Spectrometry Proteomics Data

Preprocessing, Management, and Analysis of Mass Spectrometry Proteomics Data Preprocessing, Management, and Analysis of Mass Spectrometry Proteomics Data M. Cannataro, P. H. Guzzi, T. Mazza, and P. Veltri Università Magna Græcia di Catanzaro, Italy 1 Introduction Mass Spectrometry

More information

Introduction to Database Searching using MASCOT

Introduction to Database Searching using MASCOT Introduction to Database Searching using MASCOT 1 Three ways to use mass spectrometry data for protein identification 1.Peptide Mass Fingerprint A set of peptide molecular masses from an enzyme digest

More information

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data Automated Data Interpretation for Glycan Characterization Jenny Albanese1, Matthias Glueckmann2 and Christof Lenz2 1Applied

More information

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can

More information

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate

More information

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative

More information

Quick Start Guide Chromeleon 7.2

Quick Start Guide Chromeleon 7.2 Chromeleon 7.2 7229.0004 Revision 1.0 July 2013 Table of Contents 1 Introduction... 1 1.1 About this Document... 1 1.2 Other Documentation... 2 2 Using Chromeleon... 3 2.1 Overview... 3 2.2 Starting Chromeleon...

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental

More information

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics Ilan Beer Haifa Research Lab Dec 10, 2002 Pep-Miner s Location in the Life Sciences World The post-genome era - the age of proteome

More information

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes Investigating Biological Variation of Liver Enzymes in Human Hepatocytes MS/MS ALL with SWATH Acquisition on the TripleTOF Systems Xu Wang 1, Hui Zhang 2, Christie Hunter 1 1 AB SCIEX, USA, 2 Pfizer, USA

More information

Wizard DNA Clean-Up System INSTRUCTIONS FOR USE OF PRODUCT A7280.

Wizard DNA Clean-Up System INSTRUCTIONS FOR USE OF PRODUCT A7280. Technical Bulletin Wizard DNA Clean-Up System INSTRUCTIONS FOR USE OF PRODUCT A7280. PRINTED IN USA. Revised 4/06 AF9TB141 0406TB141 Wizard DNA Clean-Up System All technical literature is available on

More information

Analysis of Liquid Samples on the Agilent GC-MS

Analysis of Liquid Samples on the Agilent GC-MS Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >

More information

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

Mass Spectra Alignments and their Significance

Mass Spectra Alignments and their Significance Mass Spectra Alignments and their Significance Sebastian Böcker 1, Hans-Michael altenbach 2 1 Technische Fakultät, Universität Bielefeld 2 NRW Int l Graduate School in Bioinformatics and Genome Research,

More information

Shimadzu UV-VIS User s Guide

Shimadzu UV-VIS User s Guide Shimadzu UV-VIS User s Guide 1) Push the F4 button on the UV-VIS instrument keypad. This will enable PC control. Push the F4 Button 2) Log into the UV-VIS software with your username and password. 3) After

More information

Separation of Peptides from Enzymatic Digestion on Different Acclaim Columns: A Comparative Study

Separation of Peptides from Enzymatic Digestion on Different Acclaim Columns: A Comparative Study Application Update Now sold under the Thermo Scientific brand Separation of Peptides from Enzymatic Digestion on Different claim Columns: A Comparative Study INTRODUCTION Separation of peptides which can

More information

Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis

Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis B. Apicella*, M. Passaro**, X. Wang***, N. Spinelli**** mariadellarcopassaro@gmail.com

More information

Functional Data Analysis of MALDI TOF Protein Spectra

Functional Data Analysis of MALDI TOF Protein Spectra Functional Data Analysis of MALDI TOF Protein Spectra Dean Billheimer dean.billheimer@vanderbilt.edu. Department of Biostatistics Vanderbilt University Vanderbilt Ingram Cancer Center FDA for MALDI TOF

More information

2D gel Protocol. 2. Determining Protein Concentration of cell lysates

2D gel Protocol. 2. Determining Protein Concentration of cell lysates 2D gel Protocol 1. Lysis and Protein Extraction from cells Prepare cell lysates with Trizol extraction by following Kathleen Lyons s protocol: AfCS Procedure Protocol PP00000155, Version 1, 05/12/03 (Ref.1).

More information

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Mark Belmont and Alexander N. Semyonov, Thermo Fisher Scientific, Austin, TX, USA Ming Gu, Cerno Bioscience,

More information

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.

More information

Protein Hit1, a novel box C/D snornp assembly factor, controls cellular concentration of protein Rsa1p by direct interaction

Protein Hit1, a novel box C/D snornp assembly factor, controls cellular concentration of protein Rsa1p by direct interaction Supplementary data Protein Hit1, a novel box C/D snornp assembly factor, controls cellular concentration of protein Rsa1p by direct interaction Benjamin Rothé, Jean-Michel Saliou, Marc Quinternet, Régis

More information

Challenges in Computational Analysis of Mass Spectrometry Data for Proteomics

Challenges in Computational Analysis of Mass Spectrometry Data for Proteomics Ma B. Challenges in computational analysis of mass spectrometry data for proteomics. SCIENCE AND TECHNOLOGY 25(1): 1 Jan. 2010 JOURNAL OF COMPUTER Challenges in Computational Analysis of Mass Spectrometry

More information

TOF FUNDAMENTALS TUTORIAL

TOF FUNDAMENTALS TUTORIAL TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] info@rmjordan.com [e-mail] This

More information

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system

More information

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that

More information

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2 Uscn Life Science Inc. Wuhan Website: www.uscnk.com Phone: +86 27 84259552 Fax: +86 27 84259551 E-mail: uscnk@uscnk.com ELISA Kit for Human Prostaglandin E1(PG-E1) Instruction manual Cat. No.: E0904Hu

More information

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Room 213 Beckman Coulter DTX 880 Multimode Detector Information The Beckman Coulter DTX 880 Multimode Detector is

More information

Mascot Search Results FAQ

Mascot Search Results FAQ Mascot Search Results FAQ 1 We had a presentation with this same title at our 2005 user meeting. So much has changed in the last 6 years that it seemed like a good idea to re-visit the topic. Just about

More information

Workshop IIc. Manual interpretation of MS/MS spectra. Ebbing de Jong. Center for Mass Spectrometry and Proteomics Phone (612)625-2280 (612)625-2279

Workshop IIc. Manual interpretation of MS/MS spectra. Ebbing de Jong. Center for Mass Spectrometry and Proteomics Phone (612)625-2280 (612)625-2279 Workshop IIc Manual interpretation of MS/MS spectra Ebbing de Jong Why MS/MS spectra? The information contained in an MS spectrum (m/z, isotope spacing and therefore z ) is not enough to tell us the amino

More information

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry Thermo Scientific Pierce Controls and Standards for Mass Spectrometry calibration tools for mass spectrometry Ensure confidence in instrument performance with Thermo Scientific Pierce Calibration Solutions

More information

Thermo Scientific Mass Spectrometric Immunoassay (MSIA)

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) INSTRUCTIONS FOR USE OF PRODUCTS Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Demonstration Protocol for Affinity Purification and Analysis of Human Insulin from Human Plasma Utilizing MSIA

More information

Mass Spectrometry Signal Calibration for Protein Quantitation

Mass Spectrometry Signal Calibration for Protein Quantitation Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of

More information

Peak intensity Trypsin. Protein Sequence Before trypsin digestion After digestion with trypsin for 24 h digestion

Peak intensity Trypsin. Protein Sequence Before trypsin digestion After digestion with trypsin for 24 h digestion Supplemental table 1. The completeness of trypsin digestion of the signature peptides with native flanking sequence. The signature peptides with native flanking sequence were digested with trypsin as described

More information

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in

More information

BD FACSComp Software Tutorial

BD FACSComp Software Tutorial BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS

More information

Two-Dimensional Gel Electrophoresis (2-DGE)

Two-Dimensional Gel Electrophoresis (2-DGE) - Introduction - Sample preparation - First dimension: Isoelectric focusing - Second dimension: SDS-PAGE - Detection of protein spots: staining - Imaging analysis & 2D Gel databases - Spot handling: excision,

More information

Proteomic data analysis for Orbitrap datasets using Resources available at MSI. September 28 th 2011 Pratik Jagtap

Proteomic data analysis for Orbitrap datasets using Resources available at MSI. September 28 th 2011 Pratik Jagtap Proteomic data analysis for Orbitrap datasets using Resources available at MSI. September 28 th 2011 Pratik Jagtap The Minnesota http://www.mass.msi.umn.edu/ Proteomics workflow Trypsin Protein Peptides

More information

Mascot Integra: Data management for Proteomics ASMS 2004

Mascot Integra: Data management for Proteomics ASMS 2004 Mascot Integra: Data management for Proteomics 1 Mascot Integra: Data management for proteomics What is Mascot Integra? What Mascot Integra isn t Instrument integration in Mascot Integra Designing and

More information