Verification/Validation of Non- FDA Approved Tests
|
|
- Susan Jones
- 7 years ago
- Views:
Transcription
1 Verification/Validation of Non- FDA Approved Tests Michael Loeffelholz, Ph.D. ABMM Professor, Dept. Pathology Director, Clinical Microbiology Laboratory University of Texas Medical Branch, Galveston 1
2 Faculty Disclosure The Association of Public Health Laboratories adheres to established standards regarding industry support of continuing education for healthcare professionals. The following disclosures of personal financial relationships with commercial interests within the last 12 months as relative to this presentation have been made by the speaker(s): Michael Loeffelholz: Biofire Diagnostics (advisory board) Pro-Lab Diagnostics (grant funding) 2
3 Outline Definitions Review of regulatory requirements and guidance documents Verification/validation examples 3
4 Verification and Validation Definitions CLIA, CLSI CAP Validation On going process; tests continue to perform as per specs (aka quality assurance) One time process; modified tests and LDTs Verification One time process to demonstrate that test system performs according to specifications; all tests One time process; unmodified tests 4
5 I Will Discuss the One-Time I will not discuss Process Reagent QC New lot/new shipment QC Establishing internal control in lieu of external controls New instrument qualification 5
6 What are Non-FDA Approved Tests? Modifications (off label use) of FDAcleared/approved tests (ModTs) Laboratory-developed tests (LDTs) 6
7 Test Modification Examples* Pre-analytical Change in specimen handling instructions Specimen type or source Analytical Change or elimination of procedural step Change in cutoff Different calibration material Post-analytical Qualitative results reported as quantitative 7 *Not a complete list of test modifications.
8 CLIA and CAP Essentially Do Not Distinguish Between ModT and LDT 8
9 CLIA ModTs and LDTs (b)(2) Each laboratory that modifies an FDAcleared or approved test system, or introduces a test system not subject to FDA clearance or approval (including methods developed in-house and standardized methods such as text book procedures) must, before reporting patient results, establish performance specifications 9
10 Establish versus Verify CLIA regulations refer to establishment of performance specifications, for ModT or LDTs Breaking new ground. No extensive clinical trial data Could argue that standardized methods (e.g. bacterial or viral culture) have extensive data Going outside of package insert Essentially creating own package insert 10
11 CLIA ModTs and LDTs (b)(2) The establishment of method performance specifications should provide evidence that the accuracy, precision, analytical sensitivity, and analytical specificity of the procedure is adequate to meet the clients needs as determined by the laboratory director... Accuracy and precision studies are required of FDAcleared tests Analytical sensitivity and analytical specificity are additional elements, required for ModT and LDT 11
12 CAP All Common Checklist COM Body fluid validation Testing specimens using methods intended for other specimen types validated by the laboratory for accuracy, precision, analytical sensitivity, analytical interferences, and reportable range. COM LDT reporting Reports contain a description of the method, and statement that the assay was developed by the laboratory COM LDT clinical claims validation All clinical claims made about (LDT) are validated. COM Analytic methodology changes If the laboratory changes methodology so that test results or their interpretations may be SIGNIFICANTLY different, the change is explained to clients. 12
13 CAP Microbiology Checklist Standards related to verification/validation are limited to Molecular subsection MIC.64770; sample types or uses collection devices other than those listed in package insert, the laboratory performs validation studies MIC new cut-off value has been validated. MIC modifies an FDA-cleared/approved assay, the modified procedure has been validated equivalent or superior performance. MIC validation adequate number samples for each type of specimen MIC LDT report. contain description of method, statement that assay was developed by the laboratory MIC ASR report. report includes disclaimer required by federal regulations. 13
14 CAP Microbiology Checklist Molecular subsection is not well outlined, to clearly distinguish requirements for unmodified tests versus ModTs and LDTs Assay validation section (MIC ) refers to analytical and diagnostic sensitivity and specificity, precision, linearity, reportable and reference ranges. However, All Common checklist: Analytical sensitivity (COM.40400). may consist of data from manufacturer Analytical specificity (COM.40450). For modified or LDTs. Result is confusion over what the laboratory must do to validate a new test 14
15 15
16 Cumitech 31A ModTs and LDTs accuracy (diagnostic tests) should be verified minimum of 50 (+) specimens 100 (-) specimens identification tests minimum of 200 isolates very small modifications may not require the extensive testing Cumitech 31A, Verification and Validation of Procedures in the Clinical Microbiology Laboratory ASM Press 16
17 So What Must I Do? Unmodified, FDA-cleared/approved tests Accuracy (i.e. diagnostic sensitivity and specificity) Reproducibility/precision Reportable range Reference range Others required for test performance (e.g. linearity for quantitative tests) ModTs and LDTs Above, plus Analytical sensitivity Analytical specificity/interfering substances 17
18 How Should I Do It? Accuracy Specimen panel with expected results From another method in your lab Specimens provided by another lab Tests should have expected concordance Reproducibility/precision Repeat testing (within run, between run, between operators) gives consistent results Reportable range Test both weak and strong positive specimens (in panel above) 18
19 How Should I Do It? Reference range Determine cutoff such that specimens from healthy/normal subjects are negative Analytical sensitivity Limit of detection Analytical specificity/interfering substances Factors that can give false positive results (other organisms, specimen matrix components) False negative results (specimen components) 19
20 Examples of ModT/LDT Validation/Verification 20
21 #1. Change from Culturette swabs to flocked swab for bacterial cultures Is validation/verification required? 21
22 #1. Swab Change for Bacterial Culture..regulatory agencies require the validation of newly incorporated test systems, the specimen collection swab, which is not a component of the test system, is neither a test nor a system. It is in fact more like a scalpel or urine cup, neither of which requires validation before use. Swabs for general specimen collection. CAP and CLIA laboratory surveyors have no grounds to suggest that validation of swabs is necessary. 22 Source: Miller, Campbell, Loeffelholz Changing swabs: to validate or not to validate? J Clin Micro. 51(11):3910
23 #2. Rapid virus antigen detection test (FDA-cleared). Kit provides specified dacron swabs. Change to flocked swab Is validation/verification required? 23
24 #2. Swab Change for FDA-Cleared Test a modification to the FDA clearance and would require a validation to ensure that (the test using) the new collection swab works at least as well as the swab recommended by the manufacturer. You are validating the performance of the virus antigen test (with this modification), NOT the swab 24 Source: Miller, Campbell, Loeffelholz Changing swabs: to validate or not to validate? J Clin Micro. 51(11):3910
25 #2. Swab Change for FDA-Cleared Test Suggested validation approach Parallel comparison of specimens collected with FDA-cleared swabs and flocked swab. Can include contrived specimens (accuracy) As this is a minor change (same specimen source), smaller number of specimens may be adequate Test several specimens in duplicate, and with different operators to demonstrate reproducibility Include both weak and strong positives in panel, to verify reportable range 25
26 #2. Swab Change for FDA-Cleared Test Again, this is a minor change (no change in sample source, patient population, or analytical changes) Reference range, analytical sensitivity and specificity unlikely to be affected 26
27 #3. Swab Change for Fungal Culture Flocked swabs (Eswab; Copan) are FDA-clear for aerobes, anaerobes and fastidious bacteria (package insert) Use for recovery of fungi would be outside package insert claims (modification) However, swabs aren t a test system. Regulations re. validation refer to test system Relevant issue is survival of fungi in liquid transport system. Science dictates that validation/verification of this characteristic be performed. Spike several yeast, molds into liquid transport medium Hold at conditions that simulate transport (time, temperature) Perform quantitative culture (T0, T24 h, T48 h) 27
28 #4. Specimen Change for FDA- Cleared Test A more rigorous validation than a swab change Test characteristics (esp. diagnostic accuracy, analytical sens) are unknown and must be established Assemble panel of positive and negative specimens 28
29 #4. Specimen Change for FDA- Cleared Test Validation approach Number of specimens Twenty is probably not sufficient Up to 50+ves and 100-ves Local regulatory agencies may specify # Appropriate matrix and source of positives Actual patient specimens Spiked (e.g. create rectal swabs; spike with Neisseria gonorrhoeae from culture plate or external QC material) Aliquot of specimens from lab that has validated specimen source 29
30 #4. Specimen Change for FDA- Cleared Test Validation approach Accuracy Patient specimens- second, reference method Spiked specimens- expected result Shared specimens- versus other lab Reportable range Include specimens with both low and high levels of analyte Reference (normal) range Negative patients, or non-spiked samples should test negative 30
31 #4. Specimen Change for FDA- Cleared Test Validation approach Reproducibility/precision Repeat several positive specimens (inter-, intra-run, inter-operator) Analytical sensitivity Test several replicates of different amounts of analyte Analytical specificity/interference If applicable, compare assay internal control (e.g. molecular test), new vs. FDA-approved specimen Review manufacturer s package insert data for relevance to evaluated sample type Spiking study if necessary (e.g. blood) 31
32 #5. Change in Specimen Processing Goal- validate unextracted or crude specimens in place of nucleic acid extraction procedure Saves time, $ Risk: decreased sensitivity (diagnostic and analytical); specificity (assay interference from crude specimens); reproducibility 32
33 #5. Change in Specimen Processing Validation approach Same specimens as FDA-cleared assay Accuracy Assemble panel of specimens previously tested by unmodified assay (extracted specimens) Reportable range Include both strong and weak +ves in above panel Reference (normal) range -ves in panel above 33
34 #5. Change in Specimen Processing Reproducibility/precision Retain several positive specimens for rpt testing Analytical sensitivity Serial dilutions of several positives Spike with different amounts of analyte Analytical specificity/interference If applicable, compare assay internal control (e.g. molecular test), new vs. FDA-approved specimen Spiking study if necessary (e.g. blood) 34
35 #6. Validation of LDT Real-time PCR for virus X Pre-validation Literature review Identify specimen types Determine intended use. This is important! (Screening? Diagnostic? Symptomatic? Asymptomatic?) Select primers, probes. Identify vendor [may be analyte specific reagents (ASRs)] Sample processing and PCR conditions optimized and determined 35
36 #6. Validation of LDT Rigorous validation is required establishing performance characteristics (for your laboratory) Assemble panel of specimens Up to 50+ves and 100-ves From your laboratory (accuracy vs. another method) Can be spiked (accuracy vs. expected result) From another laboratory (lab to lab accuracy) 36
37 #6. Validation of LDT Reference and reportable range Ensure that weak positives are reliably detected May need to adjust threshold (C T value) for cutoff Reproducibility As for modified tests Analytical sensitivity Test several replicates of different amounts of analyte Analytical specificity Monitor PCR internal control Test relevant microorganisms Likely found in specimen source Genetically related viruses Test substances that could interfere (e.g. blood) 37
38 Acknowledgements Richard Clark Michael Lewinski Bob Tibbets Susie Sharp Alice Weissfeld 38
39 39
TEST METHOD VERIFICATION AND VALIDATION
1 TEST METHOD VERIFICATION AND VALIDATION SWACM 2014 MICHAEL LOEFFELHOLZ, PH.D., ABMM DEPT. PATHOLOGY UNIV TEXAS MEDICAL BRANCH GALVESTON, TX 2 OBJECTIVES Define validation and verification Describe components
More informationMethod Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
More informationBest Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD
Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD Director, Clinical Chemistry Molecular Diagnostics / Immunology Department of Laboratory Medicine Lahey Clinic Medical
More informationIQCP GUIDELINES and TEMPLATE FOR GETTING STARTED
IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED Linda C. Bruno, M.A., MT(ASCP) Director, Microbiology and Molecular Labs ACL Laboratories, Rosemont, IL June 1, 2015 Disclosures - No disclosures 2 IQCP
More informationIndividualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016)
Topic: Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016) Click on the links below to be taken to a specific section of the FAQs. General
More informationMycoplasma Testing Products & Services. M-175 CELLshipper Mycoplasma Detection Kit (In-house sample preparation and slide fixation)
Cell Culture Testing M-1500 Real-Time PCR with Broth Enrichment A specific and sensitive method for the detection of mycoplasma using Real-Time PCR coupled with a pre-enrichment procedure to enhance method
More informationHow to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
More informationAssay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff
Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Document issued on: April 25, 2013 The draft of this document was issued on January 5, 2009. For questions regarding
More informationDevelopment and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO
Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i
More informationMODULE 2D ENVIRONMENTAL MICROBIOLOGICAL LABORATORY ACCREDITATION PROGRAM (EMLAP) ADDITIONAL REQUIREMENTS
MODULE 2D ENVIRONMENTAL MICROBIOLOGICAL LABORATORY ACCREDITATION PROGRAM (EMLAP) ADDITIONAL REQUIREMENTS 2D.1 SCOPE The AIHA- Laboratory Accreditation Programs (AIHA-LAP), LLC s Environmental Microbiological
More informationAssay Qualification Template for Host Cell Protein ELISA
Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to
More informationWhat Does Having a FDA Cleared Pregnancy Test Mean?
What Does Having a FDA Cleared Pregnancy Test Mean? Clinical Trials Transformation Initiative (CTTI) July 15-16, 2013 Denise N. Johnson-Lyles, Ph.D. Toxicology Branch Chief Division of Chemistry and Toxicology
More informationQuality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical
More informationEUROTUBO DELTALAB 6. SWABS
92 93 94 The swab, sterile or non sterile, is used for biological sample collection. Specially used for processing samples which, after being coloured, will be analysed by microscopy. Also suitable for
More informationCAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD
CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD TABLE OF CONTENTS Introduction Applicability General Topics Proficiency Testing
More informationIndividualized Quality Control Plan
NOVEMBER 2014 CLIA Individualized Quality Control Plan WHAT IS AN IQCP? OVERVIEW Discovering, Deciphering and Designing an Individualized Quality Control Plan (IQCP) IQCP provides a framework for customizing
More informationTransportation of Specimens for Neisseria gonorrhoeae Culture
TRANSPORTATION OF SPECIMENS FOR NEISSERIA GONORRHOEAE MARCH 2014 ation of Specimens for Neisseria gonorrhoeae Culture There is a critical need to ensure accurate antimicrobial resistance testing for Neisseria
More information2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests
2010 Laboratory Accreditation Program Audioconference Accreditation Requirements for Waived Vs. Non-waived Tests Objectives: After participating in this audioconference you will be able to: Define the
More informationMolecular diagnostics is now used for a wide range of applications, including:
Molecular Diagnostics: A Dynamic and Rapidly Broadening Market Molecular diagnostics is now used for a wide range of applications, including: Human clinical molecular diagnostic testing Veterinary molecular
More informationCenter for Clinical Standards and Quality/Survey & Certification Group
DEPARTMENT OF HEALTH & HUMAN SERVICES Centers for Medicare & Medicaid Services 7500 Security Boulevard, Mail Stop C2-21-16 Baltimore, Maryland 21244-1850 Center for Clinical Standards and Quality/Survey
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number: K092353 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY B. Purpose for Submission: This is a new 510k application for a new indication for the MONOLISA Anti-HAV IgM EIA
More informationValidation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
More informationEvaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories
Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,
More informationDocument Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency?
Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency? http://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/clia_compbrochure_508.pdf
More informationMethods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1
Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification
More informationOptimizing the Post-implementation Monitoring of a LC-MS/MS Method
Optimizing the Post-implementation Monitoring of a LC-MS/MS Method Julianne Cook Botelho, PhD Centers for Disease Control and Prevention Atlanta, GA AACC- Mass Spectrometry in the Clinical Lab: Best Practice
More informationClinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver
Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver When is a CLIA Certificate of Waiver Required? NOTE: Congress passed the Clinical Laboratory Improvement Amendments
More informationASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
More informationCalibration Verification
Calibration Verification INTRODUCTION On January 24, 2003, the Centers for Medicare and Medicaid Services (CMS) published updated laboratory regulations that became effective April 24, 2003. Included in
More informationThe Facts on Equine Drug Testing
The Facts on Equine Drug Testing CONTACT Travis Mays, MS Interim Section Head Analytical Chemistry, Drug Testing Laboratory Toxicology Laboratory 979.845.3414 tmays@ LIMITATIONS IN DRUG TESTING While advancements
More informationUrinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren
Urinalysis Compliance Tools POCC Webinar January 19, 2011 Dr. Susan Selgren Learning Objectives Be able to review and improve upon a laboratory plan for compliance including: Competency Documentation Proficiency
More informationGeneral. What is a laboratory test? What is a waived test?
General What is a laboratory test? What is a waived test? Laboratory tests are medical procedures that involve testing samples of blood, urine, or other tissues or substances from the body. Information
More informationCollege of American Pathologists Accreditation for Public Health Laboratories
College of American Pathologists Accreditation for Public Health Laboratories Romesh Gautom, PhD Director, Washington State Public Health Laboratory Garry McKee, PhD, MPH Director, Oklahoma State Public
More informationSutter Health Support Services Shared Laboratory Position Description. Incumbent: Laboratory Date: October 23, 2006 Written By: Michele Leonard
Sutter Health Support Services Shared Laboratory Position Description Position Title: Clinical Laboratory Scientist, Microbiologist Incumbent: Entity: SHSS Shared Laboratory Reports To: Director of Operations,
More informationQuality control in dental practice Peter Kivovics DMD, BDS, MDSc, PhD, Chief Dental Officer for Hungary
Quality control in dental practice Peter Kivovics DMD, BDS, MDSc, PhD, Chief Dental Officer for Hungary National Center for Healthcare Audit and Improvement Quality Assurance vs. Quality Control Quality
More informationHow To Use A Cpl Sample Collection System
Liquid Based Microbiology Liquid Based Microbiology Diagnostic Microbiology involves the collection of a diversity of samples from the patient to determine the etiology of diseases. Samples range from
More informationQuality Assurance Guidelines for Testing Using Rapid HIV Antibody Tests Waived Under the Clinical Laboratory Improvement Amendments of 1988
Quality Assurance Guidelines for Testing Using Rapid HIV Antibody Tests Waived Under the Clinical Laboratory Improvement Amendments of 1988 U.S. Department of Health and Human Services Centers for Disease
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE AND INSTRUMENT TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE AND INSTRUMENT TEMPLATE A. 510(k) Number: K041502 B. Purpose for Submission: To change the name of a previously cleared analyzer, and
More informationLAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009
LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009 Paul Bachner, MD, FCAP Professor & Chair Department of Pathology & Laboratory Medicine University of Kentucky
More informationQuality Requirements of a Point of Care Testing Service
Quality Requirements of a Point of Care Testing Service Adil I. Khan MSc, PhD Assistant Professor of Pathology Director of Point of Care Testing & Clinical Chemistry Laboratory, Temple University Episcopal
More informationMEDICAL DEVICE GUIDANCE
May 2014 MEDICAL DEVICE GUIDANCE GN-23: Guidance on Labelling for Medical Devices PREFACE This document is intended to provide general guidance. Although we have tried to ensure that the information contained
More informationMedical Laboratory Technology Program. Student Learning Outcomes & Course Descriptions with Learning Objectives
Medical Laboratory Technology Program Student Learning Outcomes & Course Descriptions with Learning Objectives Medical Laboratory Technology Student Learning Outcomes All Colorado Mesa University associate
More informationJOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition
JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR CliniCAl laboratories 2nd Edition Effective 1 April 2010 International Patient Safety Goals (IPSG) Goals The following is a list of all goals.
More informationFEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION CLINICAL BIOLOGY SECTION
IPH J. Wytsmanstreet 14 B-1050 Brussels FEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION CLINICAL BIOLOGY SECTION External Quality Assessment for Molecular
More information2009 LAP Audioconference Series. How to Prepare and Comply with Your Quality Management Plan
2009 LAP Audioconference Series How to Prepare and Comply with Your Quality Management Objectives: As a result of participating in this session, you will be able to: Explain the reasons why the QM is important
More informationthe central lab of choice for investigative sites and sponsors www.labconnectllc.com
the central lab of choice for investigative sites and sponsors www.labconnectllc.com Overview The LabConnect difference. LabConnect begins by developing a uniquely tailored laboratory services plan for
More informationAbout Our Products. Blood Products. Purified Infectious/Inactivated Agents. Native & Recombinant Viral Proteins. DNA Controls and Primers for PCR
About Our Products Purified Infectious/Inactivated Agents ABI produces a variety of specialized reagents, allowing researchers to choose the best preparations for their studies. Available reagents include
More informationRealStar HBV PCR Kit 1.0 11/2012
RealStar HBV PCR Kit 1.0 11/2012 RealStar HBV PCR Kit 1.0 For research use only! (RUO) Product No.: 201003 96 rxns INS-201000-GB-02 Store at -25 C... -15 C November 2012 altona Diagnostics GmbH Mörkenstraße
More informationReal time and Quantitative (RTAQ) PCR. so I have an outlier and I want to see if it really is changed
Real time and Quantitative (RTAQ) PCR or.. for this audience so I have an outlier and I want to see if it really is changed Nigel Walker, Ph.D. Laboratory of Computational Biology and Risk Analysis, Environmental
More informationContents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1
ISBN 1-56238-584-4 Volume 25 Number 27 ISSN 0273-3099 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition Robert J. McEnroe, PhD Mary F. Burritt, PhD Donald M. Powers, PhD Douglas
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k103555 B. Purpose for Submission: New submission C. Measurand: Platelet aggregation D. Type of Test: Platelet aggregometer
More informationStep-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
More informationAdmission to the Second Degree BSCLS Program. Prerequisite Course Requirements for Second Degree BSCLS
Admission to the Second Degree BSCLS Program This is a 12 month online, second degree tract in clinical laboratory science for students who have completed a four-year science degree from an accredited
More informationFOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
More informationDirect Testing Systems and Serology
Direct Testing Systems and Serology Rapid Manual Tests 6-2 Serology Diagnostics 6-6 BD Diagnostics Diagnostic Systems Catalog 2005/2006 6-1 Rapid Manual Tests Meningitis Test Systems 252360 Directigen
More informationHow Does a Doctor Test for AIDS?
Edvo-Kit #S-70 How Does a Doctor Test for AIDS? S-70 Experiment Objective: The Human Immunodefi ciency Virus (HIV) is an infectious agent that causes Acquired Immunodefi ciency Syndrome (AIDS) in humans.
More informationCLSI TRAINING COURSES CATALOG
CLSI TRAINING COURSES CATALOG A Laboratory Development Resource CLINICAL AND LABORATORY STANDARDS INSTITUTE TRAINING COURSES CATALOG The Clinical and Laboratory Standards Institute (CLSI) is known for
More informationThe Antimicrobial Susceptibility Testing Device Manufacturer: Development Process, Timelines, and Challenges
The Antimicrobial Susceptibility Testing Device Manufacturer: Development Process, Timelines, and Challenges Sheila Farnham, MT(ASCP) President Susceptibility Testing Manufacturers Association January
More informationHIV Rapid Test Kits USAID Approval and Technical Guidance June 17, 2013 Revision
HIV Rapid Test Kits USAID Approval and Technical Guidance Revision Part 1 A. Introduction 1. This document contains the requirements for USAID evaluation and approval of HIV rapid test kits and related
More informationINTERPRETATION INFORMATION SHEET
Creative Testing Solutions 2424 West Erie Dr. 2205 Highway 121 10100 Martin Luther King Jr. St. No. Tempe, AZ 85282 Bedford, TX 76021 St. Petersburg, FL 33716 INTERPRETATION INFORMATION SHEET Human Immunodeficiency
More informationPrinciples of Microbiological Testing: Methodological Concepts, Classes and Considerations
Principles of Microbiological Testing: Methodological Concepts, Classes and Considerations Russell S. Flowers Silliker Group Corp. Relating Microbiological Testing and Microbiological Criteria to Public
More informationContent Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
More informationMLX BCG Buccal Cell Genomic DNA Extraction Kit. Performance Characteristics
MLX BCG Buccal Cell Genomic DNA Extraction Kit Performance Characteristics Monolythix, Inc. 4720 Calle Carga Camarillo, CA 93012 Tel: (805) 484-8478 monolythix.com Page 2 of 9 MLX BCG Buccal Cell Genomic
More informationGuidance for Industry and FDA Staff Commercially Distributed Analyte Specific Reagents (ASRs): Frequently Asked Questions
Guidance for Industry and FDA Staff Commercially Distributed Analyte Specific Reagents (ASRs): Frequently Asked Questions Document issued on: September 14, 2007 The draft of this guidance document was
More informationIntegrating Automated Systems for Regulated Bioanalysis Farmen, R.H., Struwe, P. and Groeschl, M.
Integrating Automated Systems for Regulated Bioanalysis Farmen, R.H., Struwe, P. and Groeschl, M. EBF Meeting Nov 2014 Implementation Challenges Integration of robotic systems into existing bioanalytical
More informationGuidelines for TB Blood Testing. Minnesota Department of Health TB Prevention and Control Program June 2011
Guidelines for TB Blood Testing Minnesota Department of Health TB Prevention and Control Program June 2011 Outline Interferon-Gamma Release Assays aka TB blood tests 1. What are they? 2. What are the current
More informationURINE COLLECTION, PREPARATION AND HANDLING
URINE III-38 URINE COLLECTION, PREPARATION AND HANDLING I. SPECIMEN COLLECTION A. Introduction Laboratory tests requiring urine specimens involve a wide variety of procedures. A basic urinalysis is almost
More informationCHAPTER 13. Quality Control/Quality Assurance
CHAPTER 13 Quality Control/Quality Assurance Quality Control/Quality Assurance (QC/QA) can be defined as the set of planned and systematic activities focused on providing confidence that quality requirements
More informationInternational GMP Requirements for Quality Control Laboratories and Recomendations for Implementation
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. ludwig_huber@labcompliance.com Overview GMP requirements for Quality Control laboratories
More informationUSING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
More informationFDA Regulation of Genetic Tests. Steven Gutman, M.D. Office of In Vitro Diagnostics
FDA Regulation of Genetic Tests Steven Gutman, M.D. Office of In Vitro Diagnostics Medical Device Amendments of 1976 General controls Registration and listing Good manufacturing practices Adverse event
More informationExamples: collecting the right sample on the right patient; labeling the sample accurately;
Quality Assessment: Basics & EMR is called by several names: and are just two of them. Quality Assessment (QA) is an on-going program designed to monitor and evaluate the quality or accuracy of all laboratory
More informationBiopharmaceutical Process Evaluated for Viral Clearance
Authored by S. Steve Zhou, Ph.D. Microbac Laboratories, Inc., Microbiotest Division The purpose of Viral Clearance evaluation is to assess the capability of a manufacturing production process to inactivate
More informationDocument and Record Control in a Hospital Clinical Laboratory
Document and Record Control in a Hospital Clinical Laboratory Linda A. Chambers, MD Transfusion Service Medical Director Riverside Methodist Hospital and Grant Medical Center OhioHealth, Columbus, Ohio,
More informationIntroduction to Medical Microbiology
Introduction to Medical Microbiology Course Medical Microbiology Unit I Introduction to Microbiology Essential Question What is Medical Microbiology? TEKS 130.207(c) 2A, 3D Prior Student Learning n/a Estimated
More informationAlgorithm for detecting Zika virus (ZIKV) 1
Algorithm for detecting Zika virus (ZIKV) 1 This algorithm is addressed to laboratories with established capacity (molecular, antigenic and/or serological) to detect dengue (DENV), Zika (ZIKV) 2, and chikungunya
More informationBlood, Plasma, and Cellular Blood Components INTRODUCTION
Blood, Plasma, and Cellular Blood Components INTRODUCTION This chapter of the Guideline provides recommendations to Sponsors of Requests for Revision for new monographs for blood, plasma, and cellular
More informationEcology Quality Assurance Glossary
Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability
More informationXV.Quality Assurance in the Blood Bank
XV.Quality Assurance in the Blood Bank A. Overview 1. Goals 2. Terms B. Record Keeping a. Safe transfusion b. Careful adherence to SOPs by trained personnel c. Develop comprehensive guidelines to be in
More informationVALIDATION AND QUALIFICATION
VALIDATION AND QUALIFICATION Lizette Caballero, B.S., M.T. (ASCP) Laboratory Manager Cellular Therapy Laboratory Florida Hospital Cancer Institute Orlando, FL 1 Learning Objectives Validation of Equipment
More informationAN AUDIT QUESTIONNAIRE THAT EXAMINES SPECIFICALLY THE MANAGEMENT OF TECHNICAL ACTIVITIES CLAUSES IN ISO 15189
Indian Journal of Clinical Biochemistry, 2010 / 25 (1) 92-98 LABORATORY DOCUMENTATION AN AUDIT QUESTIONNAIRE THAT EXAMINES SPECIFICALLY THE MANAGEMENT OF TECHNICAL ACTIVITIES CLAUSES IN ISO 15189 T F Hartley
More informationProcedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3
Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3 Prepared by Date Adopted Supersedes Procedure # Estelle Piwowar-Manning Quality Assessment vers 2 Sarah Dawson Paul Richardson
More informationA commitment to quality and continuous improvement
Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous
More informationIs Your Lab s Competency Assessment a Competent Assessment?
Every number is a life. Is Your Lab s Competency Assessment a Competent Assessment? Presented by Jean Ball Inspections Team Lead College of American Pathologists www.cap.org Objectives Understand the reasons
More informationSureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)
WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory
More informationEZValidation Online Tool. The easy way to apply guidelines. The EZValidation Online Tool. The implementation of new tests
The implementation of new tests The EZValidation Online Tool Since 999, AcroMetrix has provided quality control requires validation or verification studies that meet is a comprehensive tool specifically
More informationGENETICS CRIME SCENE DNA COLLECTION AND PRESERVATION FOR HUMAN IDENTIFICATION. Innovating Together. copanflock.com
GENETICS CRIME SCENE DNA COLLECTION AND PRESERVATION FOR HUMAN IDENTIFICATION Innovating Together copanflock.com A FULLY INTEGRATED COMPANY COPAN S MISSION IS TO IMPROVE THE QUALITY OF THE PRE-ANALYTICAL
More informationChapter 2: Quality Assurance and Legal Issues
Objectives Chapter 2: Quality Assurance and Legal Issues 1. Define the key terms and abbreviations listed at the beginning of this chapter. 2. Identify national organizations, agencies, and regulations
More informationChecklist. Standard for Medical Laboratory
Checklist Standard for Medical Laboratory Name of hospital..name of Laboratory..... Name. Position / Title...... DD/MM/YY.Revision... 1. Organization and Management 1. Laboratory shall have the organizational
More informationQuality Assurance for the Clinical Laboratory
EP Evaluator Quality Assurance for the Clinical Laboratory Simplified process, powerful results While we (clinical labs) are but a small component of the total healthcare system, our data control a large
More informationGene Expression Assays
APPLICATION NOTE TaqMan Gene Expression Assays A mpl i fic ationef ficienc yof TaqMan Gene Expression Assays Assays tested extensively for qpcr efficiency Key factors that affect efficiency Efficiency
More informationDefinition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope
More informationGLOBAL FUND QUALITY ASSURANCE POLICY FOR DIAGNOSTICS PRODUCTS. (Issued on 14 December 2010, amended on 5 February 2014)
GLOBAL FUND QUALITY ASSURANCE POLICY FOR DIAGNOSTICS PRODUCTS (Issued on 14 December 2010, amended on 5 February 2014) BASIC PRINCIPLES 1. Grant funds provided by the Global Fund may only be used to procure
More informationStreptococcus pneumoniae IgG AB (13 Serotypes), MAID... 7
Volume 12 December 2011 Table of Contents Summary of Test Changes... 3 Test Changes... 6 Human Anti-Mouse AB (HAMA), ELISA... 6 Hepatitis E Antibody (IgG, IgM)... 6 Hepatitis E Antibody (IgG)... 6 Hepatitis
More informationAAV specific issues pertaining to vector shedding in gene therapy clinical trials. Samuel Wadsworth Genzyme Corporation
AAV specific issues pertaining to vector shedding in gene therapy clinical trials Samuel Wadsworth Genzyme Corporation Workshop objectives Assess the impact of vector design on shedding (studies) Review
More informationFactors Influencing LC/MS/MS Moving into Clinical and Research Laboratories
Factors Influencing LC/MS/MS Moving into Clinical and Research Laboratories Matthew Clabaugh, Market Development AACC Workshop St. Louis, MO September 17,18 2013 Factors Influencing LCMSMS Moving into
More informationIn vitro diagnostic reagent, calibrator and control material stability
In vitro diagnostic reagent, calibrator and control material stability Pauline Armstrong Randox Laboratories, Ltd. Crumlin The stability of in vitro medical diagnostic devices whether they are reagents,
More informationVaginal ph Test (ph Hydrion TM Paper 4.5-7.5)
University of California, San Francisco Department of Laboratory Medicine San Francisco General Hospital 1001 Potrero Avenue, San Francisco CA 94110 Clinical Laboratory Eberhard Fiebig, M.D., Director
More informationQuality Assurance (QA) & Quality Control (QC)
Quality Assurance (QA) & Quality Control (QC) In the laboratory Georges Ruta Water Quality Scientist City West Water 1 City West Water Water Supply & Sewerage Authority Melbourne Pop. >800,000 Area 580
More information