X-Ray Patient Doses A Strategy for Evaluation & Recording
|
|
- Irma Ann Thomas
- 7 years ago
- Views:
Transcription
1 X-Ray Patient Doses A Strategy for Evaluation & Recording Bogdan OLTEANU M.D. Valentin PARASCHIVA Phy. Codrut CHERESTES Phy. Margareta CHERESTES Phy. Emergency Children s Hospital Grigore Alexandrescu Bucharest, Romania
2 Objectives To be aware of what different dose units mean Evaluate the Thermo-Luminescent Detector (TLD) measurements Plan a strategy for patient-dose recording that fits daily practice, is physically correct and has a meaning in evaluating individual and populational risks.
3 Some Basic Knowlege ugy (TL D) mas POL YMOBIL E 125 kv TE L E DIAG NOS T 125 kv POL YMOBIL E 70 Kv TE L E DIAG NOS T 70 Kv POL YMOBIL E 40 kv TE L E DIAG NOS T 40 kv The dose varies linearly with mas but also varies strongly with KV and significantly with technical differencies between X-Ray units
4 ugy (TL D) Some Basic Knowlege Dos es by F iltering (Telediagnost, 50 mas, 1.5m) F ilter: +0 F ilter: +2Al Filter: +0,2C u+1al P ower (F ilter: +0,2C u+1al) P ower (F ilter: +2Al) y = 0,0443x 2,295 R 2 = 0,9989 y = 0,0021x 2,8754 R 2 = 0, U, kv The dose varies with KV by a power equation and varies strongly with technical differencies between X-Ray units: -beam filtering -generator type
5 How to measure patient doses at X-Rays? Effective Dose (Sievert <Sv>) Measure radiation absorbed in the body, balanced by risk-related coefficients Is proportional with risk (deaths because of late cancers) Adds together radiation effects on different organs Different various irradiations (Rg, Fluoro, CT, Scinti, PET) can be added in a period of time or lifelong Best tool, but complicated to measure TLDs within a plastic phantom or TLDs on the patient skin and use of math. phantoms
6 How to measure patient doses at X-Rays? Incident dose (Gray <Gy>) Measure incident radiation (that enters the body) Indirect estimation of the absorbed dose and related risk It s not additive (Chest Rg + Lumbar Rg + Head CT...) Not the best unit, but easy to measure Single TLD on the patient skin in the center of X-ray beam Therefore chosen as Reference for Plain X-Rays!!! Includes the back-scatter
7 How to measure patient doses at X-Rays? Incident dose (Gy) Easy to measure, Reproducible, Standardized Reference-Doses available for standard patients : 70 Kg adult, 5 year old child, baby Reference Doses available for common X-Rays: Chest, Skull, Spine, Pelvis, Abdomen,... Comparisons are possible: Between practices Between a practice and the reference-doses Can lead to evaluation of the Effective Dose <Sv> Not easy!!!
8 X-Ray Doses ugy (Log scale!) Reference doses Chest (newborn) Chest (baby) Chest (5 y child) Chest lateral (5 y child) Chest (adult 70 Kg) Pelvis (5 y child) Abdomen (5 y child) Skull lateral (5 y child) Skull (5 y child) Chest lateral (adult 70 Kg) Skull lateral (adult 70 Kg) X-Ray type Skull (adult 70 Kg) Abdomen (adult 70 Kg) Pelvis (adult 70 Kg) Lumbar spine (adult 70 Kg) Lumbar spine lat (adult 70 Kg) Lumbo/sacral lat (adult 70 Kg)
9 X-Ray Doses ugy (Log scale!) Reference doses Median / Achievable doses HF X-ray units Tw o-pulses monophasic unit Chest (newborn) Chest (baby) Chest (5 y child) Chest lateral (5 y child) Chest (adult 70 Kg) Pelvis (5 y child) Abdomen (5 y child) Skull lateral (5 y child) Skull (5 y child) Chest lateral (adult 70 Kg) Skull lateral (adult 70 Kg) X-Ray type Skull (adult 70 Kg) Abdomen (adult 70 Kg) Pelvis (adult 70 Kg) Lumbar spine (adult 70 Kg) Lumbar spine lat (adult 70 Kg) Lumbo/sacral lat (adult 70 Kg)
10 How to measure patient doses at X-Rays? Personal Film-Dosimeters (F.D.) -alternative to TLDs The measurable dose-range covers all common X-rays (Almost) invisible on radiographs (Previous work of the same workgroup) First film (low doses) Second film (high doses) ugy R 2 = 0,9694 ugy R 2 = 0, O.D ,5 1 1,5 2 O.D.
11 How to measure patient doses at X-Rays? X-Ray beam dose: Dose-Area Product <Gy x m 2 > Measure radiation that goes out from the X-Ray unit Indirect estimation of the absorbed dose and related risk It s not fully additive ( Barium swallow + Enema + Lumbar Rg + Head CT...) Not the best unit, but very easy to measure DAP-meter Almost impossible to be used for a realistic evaluation of the Effective Dose Reference for fluoroscopy and whole-procedure doses!!! Includes beam that runs outside the body contour (not suitable for the majority of plain X-Rays)
12 How to measure patient doses at X-Rays? We can calculate the Incident dose (Gray <Gy>) using the Dose-Area Product <Gy x m 2 >
13 How to measure patient doses at X-Rays? We can calculate the Incident dose (Gray <Gy>) using the Dose-Area Product <Gy x m 2 > Can we?...
14 How to measure patient doses at X-Rays? We can calculate the Incident dose (Gray <Gy>) using the Dose-Area Product <Gy x m 2 > Can we?... No, we can t!
15 How to measure patient doses at X-Rays? We can calculate the Incident dose (Gray <Gy>) using the Dose-Area Product <Gy x m 2 > Can we?... No, we can t! Why?
16 DAP METER, ugy y = 0,8117x R 2 = 0, TLD, ugy 1.a. The DAP-meter measures 20% less than TLDs! -Why?
17 Dosis TLD (ugy) / in the air y = 0,7903x R 2 = 0, Dosis TLD (ugy) / on water phantom 1.b. Because back-scatter adds to dose 20%! 20x20 cm field; 20 cm thick water phantom / paper box air phantom Measurements at various KV and mas values
18 1.c. The back-scatter is also highly dependent on: Beam size Body segment size (thickness) To be further evaluated
19 2. Errors in real measurements Measuring errors: ± 1 cm Focus-patient dist. < FFD (up to 30% less) Cassette / X-Ray beam concordance up to ±3 x ±3 cm (as accepted by technical standards) Light / X-Ray beam concordance ±2 x ±2 cm (as accepted by technical standards) Measuring the light area in the plane of the incident central ray is not feasible in practice 1 cm error around a 24x30 cm field Error 15%
20 Evaluation of TLD measurements A range of X-ray units and a 20 cm water phantom were used
21 Evaluation of TLD measurements 4 TLDs exposed together in various conditions Standard Deviations at TLDs measurements Standard deviation (%) Measured doses (ugy)
22 Evaluation of TLD measurements TLDs compared with Ionization Chamber (all data) y = 0,9683x R 2 = 0, TLD, ugy Toate masuratorile 2000 Linear (Toate masuratorile) IC, ugy
23 Evaluation of TLD measurements TLDs compared with Ionization Chamber (all data) TLD, ugy 6000 TELEDIAGNOST 4000 POLYMOBIL 2 PRACTIX 100 PLUS TOSHIBA TOSHIBA 80 DIAGNOMAX DENTAR 0 0, , , , , , ,00 IC, ugy
24 Evaluation of TLD measurements TLDs compared with Ionization Chamber (all data) Log Axis y = 0,9683x R 2 = 0, TLD, ugy 100 Toate masuratorile 10 Linear (Toate masuratorile) IC, ugy
25 Evaluation of TLD measurements TLDs compared with Ionization Chamber (all data) Log Axis TLD, ugy TELEDIAGNOST 100 POLYMOBIL 2 PRACTIX 100 PLUS TOSHIBA TOSHIBA 80 DIAGNOMAX DENTAR 1 1,00 10,00 100, , , ,00 IC, ugy
26 Plan a strategy for patient-dose recording 1. Have a Radiology Information System (RIS) Computers, Network Software for patient data (and compliant software-support) Integrated into the Hospital Information System (HIS) To easy get patient data Be sure that the information is given at the relevant levels 2. Have a collaboration with a dosimetry body
27 Plan a strategy for patient-dose recording 3. For every X-Ray unit, every type of radiography and every size of patient (that means for all exposures in your practice during a period of time) Input into the RIS the exposure parameters Use a TLD to measure the incident dose. Let your specialised partner to process the TLDs and find the doses.
28 Plan a strategy for patient-dose recording 4. Input the measured doses into the RIS, linked with the specific X-Ray units, size of patients, type of radiograph and exposure conditions. 5. Now-on, the doses will be directly assigned to each exposure by the RIS, and also transferred in the patient file and practice statistics/reports.
29 Plan a strategy for patient-dose recording 6. Schedule new TLD measurements from time to time and in case of new technology acquisitions (X-Ray units, Screens / Detectors, ) or other changes in the practice.
30 Plan a strategy for patient-dose recording 7. Collaborate with your dosimetry partner, in scientific projects, to determine the most correct estimations of the Effective Dose, for each Incident Dose already recorded into the RIS for each type of radiography.
31 Take-Home Messages For Plain X-Rays use TLDs DAP meters are good for fluoroscopic procedures, not for radiography The doses cannot be measured for every exposure Instead, assign the dose values to exposure parameters and let the RIS do the records
32
ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY
ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY B. Wall, (Co-ordinator) 1, D. Hart 1, H. Mol 2, A. Lecluyse 2, A. Aroua 3, P. Trueb 3, J. Griebel 4, E. Nekolla 4, P. Gron 5, H. Waltenburg 5, H. Beauvais-March
More informationX-Rays Benefits and Risks. Techniques that use x-rays
X-Rays Benefits and Risks X-rays are a form of electromagnetic radiation, just like light waves and radiowaves. Because X-rays have higher energy than light waves, they can pass through the body. X-rays
More informationDIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE
DIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE A web module produced by Committee 3 of the International Commission on Radiological Protection (ICRP Key Points Diagnostic
More informationPRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT
PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,
More informationNational reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK
The British Journal of Radiology, 82 (2009), 1 12 COMMENTARY National reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK D HART, PhD, M C HILLIER and B F WALL,
More informationThe Exposure Index and Its Standardization
The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany ulrich.neitzel@philips.com Content The Exposure Index (EI) What is it? What is it good
More informationImage Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
More informationRadiation Dose Measurements in Routine X ray Examinations
Tenth Radiation Physics & Protection Conference, 27-3 November 21, Nasr City - Cairo, Egypt Radiation Dose Measurements in Routine X ray Examinations H. Osman *1, A. Sulieman 1, I.I. Suliman 2 and A.K.
More informationPatient Exposure Doses During Diagnostic Radiography
Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University
More informationStudy the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
More informationMulti-slice Helical CT Scanning of the Chest
Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.
More informationSOURCES AND EFFECTS OF IONIZING RADIATION
SOURCES AND EFFECTS OF IONIZING RADIATION United Nations Scientific Committee on the Effects of Atomic Radiation UNSCEAR 2008 Report to the General Assembly with Scientific Annexes VOLUME I UNITED NATIONS
More informationQuality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
More informationSUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
More informationEvaluation of the radiation load of children in neonatal departments of Slovak hospitals Maruniaková A. 1), Nikodemová D. 2), Greschner J.
Evaluation of the radiation load of children in neonatal departments of Slovak hospitals Maruniaková A. 1), Nikodemová D. 2), Greschner J. 3) 1) Regional Authority of Public Health in Banská Bystrica,
More informationStaff Doses & Practical Radiation Protection in DEXA
Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /
More informationComputed Tomography Radiation Safety Issues in Ontario
Computed Tomography Radiation Safety Issues in Ontario Healthcare Human Factors Group Centre for Global ehealth Innovation University Health Network Toronto, ON, Canada June 16, 2006 Table of Contents
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
More informationThe Challenge of CT Dose Records
The Challenge of CT Dose Records Kimberly Applegate, MD, MS, FACR Professor of Radiology and Pediatrics Emory University *financial disclosures: -Springer Textbook contracts -AIM advisory board for patient
More information5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
More informationCHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
More informationQuality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric
More informationX-ray (Radiography) - Abdomen
Scan for mobile link. X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small dose of ionizing radiation to produce pictures of the inside of the abdominal cavity. It is used to evaluate the stomach,
More informationAssessing Radiation Dose: How to Do It Right
Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine
More informationRequired RS Training Info
C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness
More informationRadiation Exposure of the UK Population from Medical and Dental X-ray Examinations
NRPB-W4 Radiation Exposure of the UK Population from Medical and Dental X-ray Examinations D Hart and B F Wall ABSTRACT Knowledge of recent trends in the radiation doses from x-ray examinations and their
More informationPersonal Dosimetry in Dental Radiology
Personal Dosimetry in Dental Radiology Position Statement Personal Dosimetry in Dental Radiology Radiological Protection Institute of Ireland May 2011 3 Clonskeagh Square, Dublin 14 Background The RPII
More informationUnderstanding Digital Modalities: System Integration and Use
Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information
More informationDiagnostic Exposure Tracking in the Medical Record
Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand
More informationIrradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
More informationMEDICAL IMAGING. 90/30 Remote Controlled RF System
MEDICAL IMAGING 90/30 Remote Controlled RF System The Easy solution for your Remote Controlled room Apollo EZ represents the Easy approach for the radiographic/fluoroscopic room and it comes from the experience
More informationIONISING RADIATION. X-rays: benefit and risk
IONISING RADIATION X-rays: benefit and risk Impress Federal Office for Radiation Protection Postfach 10 01 49 D - 38201 Salzgitter Telephone: + 49 (0) 30 18333-0 Fax: + 49 (0) 30 18333-1885 Website: www.bfs.de
More informationDose Tracking System:
Dose Tracking System: A Paradigm Shift in Patient Dose Monitoring Andrew Kuhls-Gilcrist, Ph.D. XRVL Clinical Science Manager Toshiba America Medical Systems, Inc. 2 Dose Tracking System: A Paradigm Shift
More informationThe effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.
RADIATION SAFETY: HOW TO EDUCATE AND PROTECT YOURSELF AND YOUR STAFF John Farrelly, DVM, MS, ACVIM (Oncology), ACVR (Radiation Oncology) Cornell University Veterinary Specialists The Veterinary Cancer
More informationSUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES
SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act
More informationRadiation Safety Characteristics of the NOMAD Portable X-ray System
Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality
More informationX-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581
X-ray Radiation Safety Course James Kane & Rob Deters Office of Radiological Control 545-7581 About the Course X-ray Radiation Safety X-ray radiation safety training is mandatory for radiation workers
More informationProduction of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
More informationFluoroscopy and Radiation Safety Content for Radiologists Marta Hernanz Schulman MD, FAAP, FACR Keith Strauss, M.Sc Ishtiaq H. Bercha, M.
Fluoroscopy and Radiation Safety Content for Radiologists Marta Hernanz Schulman MD, FAAP, FACR Keith Strauss, M.Sc Ishtiaq H. Bercha, M.Sc Primary Objective: Review Pediatric fluoroscopic procedures,
More informationGestión global de la dosis en TC. Sistema de registro y gestión
Gestión global de la dosis en TC. Sistema de registro y gestión IV Jornada de Protección Radiológica Hospitalaria SARH FEA Radiofísica Hospital Virgen de las Nieves. Granada jalmansa.lopez@gmail.com 22
More informationINTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
More informationX-ray (Radiography) - Bone
Scan for mobile link. X-ray (Radiography) - Bone Bone x-ray uses a very small dose of ionizing radiation to produce pictures of any bone in the body. It is commonly used to diagnose fractured bones or
More informationIntroduction. Chapter 15 Radiation Protection. Regulatory bodies. Dose Equivalent. Regulatory bodies. Main Principles of Radiation Protection
Introduction Chapter 15 Radiation Protection Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. F.M. Khan, The Physics of Radiation Therapy, 4th ed., Chapter
More informationIAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)
IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization
More informationR/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
More informationSOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration
Implementation Page #: Page 1 of 10 Last Reviewed/Update 1. Purpose and Scope The purpose of this document is to describe the Medical Physics and Radiation Safety program at Boston University (BU) and
More informationRadiation Exposure in X-ray and CT Examinations
Patient Safety-Xray: Radiation Exposure in X-ray and CT Examinations What are x-rays and what do they do? X-rays are forms of radiant energy, like light or radio waves. Unlike light, x-rays can penetrate
More informationREGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
More informationC A M P E P. Standards for Accreditation of Graduate Educational Programs in Medical Physics
Standards for Accreditation of Graduate Educational Programs in Medical Physics Preamble Revised July 2015 Medical Physics is a branch of physics that applies the concepts and principles of physics to
More informationImplementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
More informationP R E S E N T S Dr. Mufa T. Ghadiali is skilled in all aspects of General Surgery. His General Surgery Services include: General Surgery Advanced Laparoscopic Surgery Surgical Oncology Gastrointestinal
More informationMeasuring Patient Exposure in Interventional Radiology
Measuring Patient Exposure in Interventional Radiology GAFCHROMIC XR-R, a Wide-Area Dosimetry Film to Assess Peak Skin Dose (PSD) Xiang Yu, Ph.D. Director, Advanced Materials International Specialty Products
More informationDOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES
DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES Ausra Urboniene, Birute Griciene Lithuania Introduction Medical staff during interventional radiology procedures
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
More informationThe University of Southern Maine. XRF X-Ray Radiation Protection Program
The University of Southern Maine XRF X-Ray Radiation Protection Program 1 University of Southern Maine Department of Environmental Sciences 105 Bailey Gorham Hand Held XRF Analyzer X-Ray Radiation Protection
More informationUse of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
More informationDesign and Implementation of an Institution-Wide Patient-Specific Radiation Dose
Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose Monitoring Program for Computed Tomography, Digital Radiography, and Nuclear Medicine by Olav Christianson Medical Physics
More informationfor Image Scaling Enabling Digital Orthopaedics
Calibration Devices, Use & Positioning for Image Scaling Enabling Digital Orthopaedics DIGITAL IMAGE SCALING As Radiology becomes filmless. Current planning practices used by surgeons with acetate sheets
More informationCT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
More informationEasy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology
Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment
More informationSubmitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY
Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY A. Mrozik, B. Marczewska, P. Bilski, W. Gieszczyk Institute of Nuclear Physics PAN,
More informationRESPONSE BY THE CEO OF ARPANSA TO THE ADVICE ON MEDICAL RADIATION ISSUES from the RADIATION HEALTH AND SAFETY ADVISORY COUNCIL
RESPONSE BY THE CEO OF ARPANSA TO THE ADVICE ON MEDICAL RADIATION ISSUES from the RADIATION HEALTH AND SAFETY ADVISORY COUNCIL INTRODUCTION The practice of radiology, nuclear medicine and radiotherapy
More informationClinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
More informationABSTRACT 1 LIST OF PUBLICATIONS 3 NOMENCLATURE 4 AIM OF THE STUDY 5 1 INTRODUCTION 6
UNIVERSITY OF HELSINKI REPORT SERIES IN PHYSICS HU-P-D82 CALCULATING PATIENT SPECIFIC DOSES IN X-RAY DIAGNOSTICS AND FROM RADIOPHARMACEUTICALS Juha Lampinen Department of Physics Faculty of Science University
More informationPage: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
More informationChemistry 1000 Lecture 2: Nuclear reactions and radiation. Marc R. Roussel
Chemistry 1000 Lecture 2: Nuclear reactions and radiation Marc R. Roussel Nuclear reactions Ordinary chemical reactions do not involve the nuclei, so we can balance these reactions by making sure that
More informationCT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of
More informationScan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
More informationEPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography
Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information
More informationQ.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
More informationMonitoring Patient Radiation Dose in VA. Charles M. Anderson MD, PhD Chief Consultant for Diagnostic Services Veterans Health Administration
Monitoring Patient Radiation Dose in VA Charles M. Anderson MD, PhD Chief Consultant for Diagnostic Services Veterans Health Administration TWENTY ONE/NOVEMBER 2011 Computed Tomography (CT), Magnetic Resonance
More informationRadiation Safety Issues for Radiologic Technologists
Radiation Safety Issues for Radiologic Technologists Greg Sackett, M.S., CHP Medical Physicist Radiation Worker Risks? 1 Patient Risks? Acute Effects? Delayed Effects? Patient Questions? Radiation Dose
More informationCyclotron Centre in Poland and 2D thermoluminescence dosimetry
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of
More informationRadiation Protection Update
Radiation Protection Update Pennsylvania Dental Association Revised December 2013 Reviewed by the Pennsylvania Department of Environmental Protection Radiation Protection Update Radiation in the Healing
More informationX-ray (Radiography), Upper GI Tract
Scan for mobile link. X-ray (Radiography), Upper GI Tract What is Upper Gastrointestinal (GI) Tract Radiography? Upper gastrointestinal tract radiography, also called an upper GI, is an x-ray examination
More informationChecklist for submission of application form for Registration
Checklist for submission of application form for Registration *Incomplete submission may cause delay in processing of the application. Name of the Hospital/Institution: City State Sr.No. Checkpoints Status
More informationIonizing Radiation, Czech Republic, CMI (Czech Metrology Institute)
Ionizing Radiation, Czech Republic, (Czech Metrology Institute) Calibration or Measurement RADIOACTIVITY 1.0E+00 1.0E+02 Bq cm -2 C-14 1.0E+01 1.0E+02 Bq cm -2 Co-60 1.0E+01 1.0E+02 Bq cm -2 Sr-90 1.0E+01
More informationThe Radiologic Technology Program at Trenholm University
H. Councill Trenholm State Technical College 243 Location: Trenholm Campus Program Information The Radiology Technology program at TrenholmState will provide students with the necessary training to gain
More informationI.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS
I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS A. Slavtchev 1, I. Manolov 2 1 National Centre of Radiobiology and Radiation Protection 2 Medrom Ltd. Abstract In recent years the number
More informationRevised Radiation Protection Guidance for Diagnostic and Interventional X-ray Procedures
Revised Radiation Protection Guidance for Diagnostic and Interventional X-ray Procedures Mike Boyd US EPA Office of Radiation and Indoor Air NRSB Meeting July 20, 2010 What is Federal Guidance? 2 What
More informationMoving Forward What does this mean for the Medical Physicist and the Imaging Community?
Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical
More informationVideo Fluoroscopic Swallowing Exam (VFSE)
Scan for mobile link. Video Fluoroscopic Swallowing Exam (VFSE) A video fluoroscopic swallowing exam (VFSE) uses a form of real-time x-ray called fluoroscopy to evaluate a patient s ability to swallow
More informationX-Mind. Instinct for perfection
X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in
More informationCT Scan Thorax and Upper Abdomen. Respiratory Unit Patient Information Leaflet
CT Scan Thorax and Upper Abdomen Respiratory Unit Patient Information Leaflet Introduction This leaflet gives you general information about your CT (computerised tomography) scan. It does not replace the
More informationMammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial
Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY rothenbl@mskcc.org
More informationRadiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
More informationDose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography
More informationRecognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
More informationAcknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):
Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 47 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL Refresher Course
More informationRecommendations for a technical quality control program for diagnostic X-ray equipment
Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer
More informationWorldwide Quality Assurance networks for radiotherapy dosimetry
Worldwide Quality Assurance networks for radiotherapy dosimetry Joanna Izewska Dosimetry and Medical Radiation Physics Section dosimetry@iaea.org Audit in radiotherapy dosimetry Objectives to enhance confidence
More informationRadiation dose from medical imaging has come under
CONCISE REVIEW FOR CLINICIANS RADIATION RISK FROM MEDICAL IMAGING Radiation Risk From Medical Imaging Eugene C. Lin, MD This review provides a practical overview of the excess cancer risks related to radiation
More informationRESIDENT MANUAL RADIOLOGY RESIDENT ROTATION PEDIATRIC RADIOLOGY
RESIDENT MANUAL RADIOLOGY RESIDENT ROTATION PEDIATRIC RADIOLOGY INTRODUCTION Welcome to Pediatric Radiology at Arkansas Children s Hospital. We hope you learn lots of Pediatric Radiology, and that you
More informationMeasure #145: Radiology: Exposure Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety
Measure #145: Radiology: Exposure Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety 2016 PQRS OPTIONS FOR INDIVIDUAL MEASURES: CLAIMS, REGISTRY DESCRIPTION:
More informationFrequently Asked Questions: The following regulations are provided to answer some of the many questions that are asked of this office.
Steven L. Beshear Governor CABINET FOR HEALTH AND FAMILY SERVICES DEPARTMENT FOR PUBLIC HEALTH Division of Public Health Protection and Safety Radiation Health Branch 275 East Main Street, HS 1C-A Frankfort,
More informationComparison of Epson scanner quality for radiochromic film evaluation
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Comparison of Epson scanner quality for radiochromic film evaluation Hani Alnawaf, 1,3 Peter K.N. Yu, 2 Martin Butson 1,2,3a South
More informationTA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1
TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT
More informationRadiation Therapy. 1. Introduction. 2. Documentation of Compliance. 3. Didactic Competency Requirements. 4. Clinical Competency Requirements
PRIMARY CERTIFICATION DIDACTIC AND CLINICAL COMPETENCY REQUIREMENTS EFFECTIVE JANUARY 2014 Radiation Therapy 1. Introduction Candidates for certification and registration are required to meet the Professional
More informationDiagnostic X-ray facilities as per quality control performances in Tanzania
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 4, FALL 2006 Diagnostic X-ray facilities as per quality control performances in Tanzania Yesaya Y. Sungita, 1 Simon S.L. Mdoe, 1 and Peter
More informationRADIOLOGY SERVICES. By Dr Lim Eng Kok 1
INTRODUCTION RADIOLOGY SERVICES By Dr Lim Eng Kok 1 Radiology is the branch of medicine that deals with the use of ionising (e.g. x- rays and radio-isotopes) and non-ionising radiation (e.g. ultrasound
More information