1 Tools for IMRT QA N. Dogan, Ph.D Department of Radiation Oncology Virginia Commonwealth University Medical College of Virginia Hospitals Richmond, VA, USA

2 Objectives To identify the QA tasks involving IMRT To describe the QA tools for all aspects of IMRT process To explain the limitations of the current IMRT QA tools To compare the IMRT QA tools and techniques

3 QA tasks for IMRT Machine QA- Acceptance and routine QA of the MLC for IMRT delivery - dosimetric and geometric characteristics Algorithm QA for IMRT - QA of planning system and data consistency with machine Patient Specific QA prove plan works 1D and 2D dosimetry of treatment components such IM beams and segments 3D dosimetry of entire treatment delivery Post Treatment QA Log-file analysis

4 Detectors Phantoms Scanners Dosimetric Analysis Tools

5 Detector Requirements for IMRT QA Geometric and dosimetric accuracy Volumetric simultaneously integrating dosimeter to faithfully quantify the dose delivered over the total time of treatment Good spatial resolution, tissue equivalent response Ability to provide 3-D information Portability to multiple phantoms Ease of use Sufficiently large dynamic range and be insensitive to photon energy spectrum and dose rate response which is independent of the energy spectrum

6 Detectors Many of them available for IMRT measurements Necessary to characterize the detector response for both static and dynamic fields for linearity Need to be calibrated for absolute measurements Need to determine stem and cable effects

7 Detectors IMRT QA Tools Need to determine energy dependence and angular response Small field detectors required for small field characterization Sensitive to position Detector should be smaller than homogeneous region of dose to be measured Assess electrometer response

8 Detectors, cont. Need to determine necessary resolution depends on the resolution of the beamlet grid that is used for planning and sequencing fields for delivery Chambers with the smaller volumes are more sensitive to position and will have a higher response when positioned at an opposing leaf pair junction and between adjacent leaves More stable measurement point Courtesy of Jean Moran, UofM Dose (cgy) Poor detector position

9 1-D and 2-D Detectors Ion chamber (1-D) TLDs and MOSFETs (1-D) Detector arrays (2-D) Film (2-D) Radiographic Radiochromic Gels (3-D)

10 Small 1-D Detectors Detector Volume (cm 3 ) Diameter (cm) Disadvantages Microchamber Poorer resolution than diodes Pinpoint chamber p-type Si diode Over-respond to low energy photons Martens et al Stereotactic diode NA 0.45 MOSFET NA NA Non-linear dose response for <30 cgy Diamond < resolution than diodes, dose rate dependence, expensive

11 Ion Chamber Advantages IMRT QA Tools Available in different shape and sizes Dosimetric response is well understood. Absolute dose measurements theory is well establish, they can be used as a benchmark standard Easy to calibrate

12 Ion Chamber, cont. Disadvantages Only one measurement point for each irradiation does not yield sufficient information to evaluate the dose throughout the target and/or critical structures Volume averaging the measurements are to be considered as an average throughout the chamber s active volume - does not yield significant errors if the ion chamber is placed in a low dose-gradient region even for relatively large chambers

13 Ion Chamber volume averaging, cont. Micro cham: 0.009cc PTW: 0.125cc Farmer:0.65cc D.A. Low et al. Ionization chamber volume averaging effects in dynamic intensity modulated radiation therapy beams, Med. Phys.30(7): (2003

14 TLDs Advantages Multiple measurement points in a single irradiation Reusable Easy to use in multiple phantoms Small size and versatility in placement Readily available readout equipment Achievable accuracy: 2-3%

15 TLDs Disadvantages Requires calibration to determine calibration factor for each TLD chip Requires calibration of subset of TLD chips for each measurement TLD reader response and oven temperature should me routinely monitored to maintain consistent TLD response Automatic reader recommended for IMRT field verification due to large number of TLDs required for verification in a plane (60 or more) inefficient for routine IMRT QA

16 TLDs, cont. IMRT QA Tools D.A. Low et al. Phantoms for IMRT Dose Distribution Measurement and Treatment Verification, Int J Radiat Oncol Biol Phys 40: (1998).

17 MOSFET systems Advantages Excellent spatial resolution small size (~0.04mm 2 ) Multiple detectors can be irradiated simultaneously Automatic and immediate readout Can be re-used immediately Linear dose response > 30 cgy Response independent of depth Commercially available phantoms to accommodate the small detectors

18 MOSFET systems Disadvantages Decrease linearity for < 30 cgy limited to high dose applications Over-response for the phantom scatter factor for small fields Specific application and measurement conditions should be carefully assessed and the detector should be used in the appropriate dose range

19 MOSFET systems IMRT QA Tools Bias Box Reader MOSFET TNRD50 system Courtesy of Cynthia Chuang, UCSF An axial image of MOSFET phantom

20 MOSFET systems IMRT QA Tools 1400 MOSFET Linearity Mosfet Consistency 1200 MOSFET1 MOSFET2 MOSFET3 MOSFET MOSFET1 MOSFET2 MOSFET Radiation (cgy) Number of Measurements Courtesy of Cynthia Chuang, UCSF

21 MOSFET systems IMRT QA Tools Percent Depth Dose Comparison Ion Chamber MOSFET Angular Dependence Percentage (%) Degrees Depth (cm) Courtesy of Cynthia Chuang, UCSF

22 Cal Gy Meas Gy Diff 4.6 % Cal Gy Meas Gy Diff % Courtesy of Cynthia Chuang, UCSF Calc Gy Meas Gy Diff 4.35% Calc Gy Meas Gy Diff 3.52% Calc Gy Meas Gy Diff 3.45%

23 Current IMRT QA Tools 2-D Detectors Film Radiographic Radiochromic Beam imaging system, CCD, SLIC, AMFPI 2-D Detector arrays Diode array (Mapcheck) Ion chamber Active matrix flat panel detector (AMFPD)

24 Radiographic Film Advantages IMRT QA Tools Readily available (XV, EDR2, ) Can be cut into any desired shape Excellent spatial resolution (<1mm) Less expensive than other 2-D systems

25 Radiographic Film, cont. Disadvantages Over-response to low energy x-rays high atomic number of the active material not good for absolute dosimetry Dependent on QA of film batch Dependent on processor and digitizer Sensitive to storage conditions Need to measure the response to dose for each experiment H&D curve each time Proper normalization is critical

26 Current IMRT QA Tools Radiographic Film, cont. Other issues Store in a cool and dry place Make sure that the temperature for the film processor is stable Film digitizer pixel spacing, integrity of OD, beware of artifacts Verify spatial and optical density accuracy

27 Rapid Film Calibration 120 MU 90 MU 60 MU 30 MU 240 MU 210 MU 180 MU 150 MU Multiple dose levels per film-3x3 cm 2 fields of different dose levels Step-and-shoot or SMLC delivery Different dose values required for XV and EDR2 film (15-120MU for XV and MU for EDR2) Saves both time and film Childress et al Med Phys 29(10), 2002.

28 XV vs. EDR Film Depth-corrected H&D Net Optical Density XV EDR XV2, 6 MV XV2, 15 MV EDR2, 6 MV EDR2, 15 MV Optical Density Co60-EDR2 10MV-EDR2 6MV-EDR2 18MV-EDR Dose (cgy) Chetty and Charland 2002 PMB 47: Dose (cgy) Dogan et al PMB 47:

29 XV Depth-corrected H&D curves MV-EDR2-Depth corrected 6MV-EDR2-Regular EDR MV-EDR2-Depth corrected 18MV-EDR2-Regular Optical Density Dose (cgy) Optical Density Dose (cgy) Dogan et al PMB 47:

30 Ion Chamber Film- depth corrected H&D Film regular H&D Dogan et al PMB 47: (a) (b) Ion chamber and EDR2 film depth-dose curves for a) 6 x 6 cm2, b) 14 x 14 cm2 films for 10 MV beam. Films were positioned parallel to the beam and OD to dose conversion was done using regular and depth-corrected H&D curves.

31 Childress et al. Med. Phys. 32(2) 2005

32 As compared to XV film, EDR2 film has less dependence on the processor, field size less response to low energy photons have better reproducibility and agreement with ion chamber measurements can be used to measure a complete fraction of an IMRT treatment

33 Radiographic Film: 2-D Dosimetric Measurements Intensity map from Opt System Calculated Leaf Sequencer Calc-Meas Courtesy of Jean Moran, UM Measured

34 Radiographic Film: Routine DMLC QA Using radiographic films Intensity-modulated pattern field Check leaf position, acceleration, motion stability Check for hot and cold density Visual check DMLC field 14x14 cm 2 at SSD =100 cm, 2 cm separated strips

35 Film Processor issues Should do routine maintenance and quality assurance verify spatial intensity, characteristic response, noise due to large changes in optical density ( Dempsey et al, Med Phys, 26; , 1999). Should be warmed up prior to use Should have appropriate amount of chemicals - Several films should be run in advance Should have stable temperature Should have a consistent rate of feeding into the processor

36 Film Other Issues Accurate positioning of the film in the phantom for the registration with treatment planning system Minimized errors by using a solid-water slab designed for film Have pins between slabs that puncture the film

37 Radiochromic Film (RCF) Advantages No significant energy dependence decreased sensitivity to low-energy photons Insensitivity to visible light Very high spatial resolution - well-suited for measurements in high-dose gradient fields Self-developing no developer or fixer is required Easy to handle Tissue equivalent

38 Radiochromic Film, cont. Disadvantages Takes a couple of hours for the color change to stabilize, and it may be necessary to wait up to two days before evaluating the film Sensitive to the air temperature and humidity Ultraviolet light may cause a color change without exposure to ionizing radiation Size, availability, and cost Non-uniform response to radiation double exposure technique minimizes this effect Issues with thermal history, wavelength dependence, and local sensitivity of the film

39 RCF (Gafchromic HS and MD55-2) vs radiographic films (XV and EDR2) O. Zeidan et al., Med. Phys., 31 (10): (2004)

40 RCF profiles vs. Ion chamber J. Dempsey et al., Med. Phys., 27 (10): (2000)

41 RCF digitizer issues Response of the digitizer Light source characteristics Design Gluckman et al, Med Phys, 29(8); , 2002.

42 Other 2-D systems IMRT QA Tools Beam imaging system, CCD, SLIC, Amorphous silicon flat panel detector (AMFPD) EPID systems attached to gantry Investigated more for pre-treatment QA currently 2-D Detector arrays Diode array (e.g; MapCheck) Ion chamber (e.g; LA48 linear array)

43 EPID Systems IMRT QA Tools Charged coupled device (CCD) camera systems Scanning liquid ion chambers (SLICs) Amorphous silicon flat panel detector (AMFPD) Active matrix flat panel imagers (AMFPIs) Transit Dosimetry Patient or Phantom Pre-Tx 2-D Measurements Film Replacement

44 EPID Systems as500 EPID 1 mm copper plate Phosphor scintillating layer (Kodak Lanex Fast B Gd2O2S:Tb, 70 mg/cm3) Array of photodiodes Amorphous Silicon panel each pixel consists of: IMRT QA Tools Light sensitive photodiode Thin film transistor 16-bit ADC Munro et. al, Med. Phys. 25, 1998

45 EPIDs Advantages Many centers have installed EPIDs and being primarily used for patient-specific pretreatment field verification and MLC QA Logical extension to investigate dosimetric applications Mounted to linear accelerator - known geometry with respect to the beam Detector sag must be accounted for at different gantry angles Positioning reproducibility important Real time digital evaluation No processor, data acquisition takes less time

46 EPIDs - Challenges EPIDs were primarily designed for patient localization High resolution, good contrast images Additional dose to the patient should be minimized The conversion of imager response to dose is complex Imaging system dependent Other problems Ghosting Lag

47 EPIDs Dose determination Imager response must be calibrated to a standard Absolute calibration to ion chamber at a point over a ROI E.g. ion chamber in a mini-phantom or slab at same SDD as EPID 2-D calibration to actual beam distribution at the imager plane Can be measured with film or a diode array

48 Factors for EPID Response Water-equivalent depth of the detector Field size dependence and scatter properties within the imager Short- and long-term reproducibility Dose rate Energy dependence Spatial integrity

49 EPID: DMLC measurements Overall: Good agreement + Predicted EPID Ion Chamber 10 MV 25 MV Discrepancies in the penumbra region (up to 10%) Pasma Med Phys 26: (2376) 1999

50 Linear Diode Array in water vs. CCD Without short range penumbra correction Courtesy of Jean Moran, UofM

51 Dose Determination using EPID (SLIC) Chang et al., Int J Radiat Oncol Phys 47: (p. 233)

52 Calculation vs. measured using AMFPD for DMLC Calculated (Calculated Measured) Agreement : Within +/- 2 cgy Courtesy of Jean Moran, UofM

53 EPIDs can provide a much-needed replacement for pre-tx QA film dosimetry Only if proper QA of the EPID is established Need better understanding of regions where EPIDs are inadequate for dosimetry Systems must be verified at more centers against accepted QA methods such as film and ion chamber Additional software is required before more facilities can do proper validation of the methods (Software must be commissioned) Can be part of a comprehensive QA program in conjunction with other methods such as computational checks (monitor programs, log file analysis, etc.)

54 Gel Dosimeters Advantages 3-D information in one irradiation Energy and dose-rate independent High sensitivity and linear response Cumulative Gel density can be changed - Ideal for anthropomorphic phantoms Near tissue equivalent Multiple readout techniques (MR, optical-ct) New gel formulations and readers commercially available

55 Gel Dosimeters Disadvantages Sensitive to time, preparation, temperature Cylindrical container required for optical readers - less accurate readout at gel/container interface MR time is often limited and expensive - long scan times for accurate readout, e.g. 5% accuracy over 10 hr scan time (Gum et al. 2002) Relative dosimeter -require cross-calibration technique batch to batch they are different Cost

56 Gel dosimetry IMRT QA Tools 8cm In-house optical CT scanner cost is less Oldham and Kim, Med. Phys. 31 (5), Upgraded motors, motion control, and user interface. (Pacific Scientific: step motors. National Instruments: motion control and Labview.)

57 Gels: Optical Density to Dose Calibration 6 Beam calibration irradiation BANG gel phantom diameter 17.4cm Courtesy of Mark Oldham, Duke University

58 Gel Dosimeters IMRT QA Tools Five Field Prostate IMRT Courtesy of Mark Oldham, Duke University Re-computed for a 3 L BANG gel dosimeter. Dmax scaled to 1.8 Gy to fit dynamic range of optical scanner BANGkit TM from MGS Research. 1x1x3mm, 5hours

59 Gel Dosimeters IMRT QA Tools Isodose comparison: Pinnacle (red), Gel-dosimetry (black) Courtesy of Mark Oldham, Duke University

60 Gel Dosimeters IMRT QA Tools Gum, et al. Preliminary study on the use of an inhomogeneous anthropomorphic Fricke gel phantom and 3D magnetic resonance dosimetry for verification of IMRT plans, Phys Med Biol 47; N

61 Phantoms for IMRT Measurements multiple phantoms for commissioning Fiducials for reproducible setup of phantom and detectors User-customized for different detectors allow special holders Simple vs. anthropomorphic Homogeneous or heterogeneous

62 Current IMRT QA Tools Simple Geometric Phantoms Water tank Accommodate different ion chambers Use for measurements of depth dose and profiles Output, flatness, symmetry, and linearity assessment Cylindrical mini-phantom Use with ion chamber to assess dependence of output on gantry angle Water-equivalent plastics: slab w/ custom chamber inserts 1-D and 2-D measurements Detector position can be varied with depth Cylindrical phantoms (plastic or water filled) Straightforward geometry Ion chamber at single position Plastic phantoms may hold films

63 Water-equivalent square IMRT Verification Phantom D.A. Low et al. Phantoms for IMRT Dose Distribution Measurement and Treatment Verification, Int J Radiat Oncol Biol Phys 40: (1998).

64 Current IMRT QA Tools A Cylindrical Phantom containing movable ion chamber L. Xing et al. Dosimetric verification of a commercial inverse treatment planning system, Phys. Med. Biol. 44: (1998).

65 A cylindrical Plastic Phantom Detector

66 Plastic Cylindrical Phantom with MOSFETs Calc Gy Meas Gy Diff 3.52% Calc Gy Meas Gy Diff 3.45% Courtesy of Cynthia Chuang, UCSF

67 Spiral Phantom IMRT QA Tools Paliwal et al A spiral phantom for IMRT and tomotherapy treatment delivery verification Med Phys (2000).

68 Anthropomorphic: RPC Head Phantom Target Volumes Water Critical Structure Removable Dry Insert Water TLDs in Target Volumes Courtesy of Jean Moran, UofM Radiochromic film through multiple plans Delivery is required by RTOG for participation in IMRT trials

69 Dosimetric Analysis Tools Provide a comprehensive and quantitative comparison between two dose distributions Different ones available Important to know the limitations

70 Dosimetric Analysis Tools Overlay of isodoses 2-D dose difference displays with colorwash Dose difference histograms Distance-to-agreement (DTA) Gamma evaluation Normalized agreement test (NAT)

71 Isodose lines and Dose Difference Display 70 cgy 60 cgy 50 cgy 20 cgy 10 cgy Calcs Film +/- 10% Courtesy of Jean Moran, UofM

72 Dose difference display Useful in shallow dose gradients Overly sensitive in steep dose gradients e.g.; a small spatial shift (due to experimental measurement errors) between two dose distributions yield large dose differences

73 Dose difference histogram and profiles

74 Distance to Agreement (DTA) Is the distance between a reference point and the nearest point in the compared dose distribution that exhibits the same dose Is not overly sensitive in steep dose gradients In shallow dose gradients, a large DTA value may be computed even for relatively small dose differences May be hard to interpret

75 Combination of dose difference and DTA Identify regions where the dose difference and DTA are simultaneously by greater than a pre-selected criteria points that fail both criteria are identified on a composite distribution The display of the dose difference may emphasize the impression of failure in high dose gradient region Provides no information on the magnitude of the failure

76 Gamma Analysis- Generalization of composite distribution Measures the closest distance between each reference point and evaluated dose distribution after scaling by D and d r ( r, r ) δ ( r, r ) Γ ( r, r ) = + = Γ r r r d D 2 2 e r e r e r γ (r 2 2 r ) m in ( e, r ) e { } { } rr (,): e r spatial distance between evaluated and reference dose points D : Dose difference criteria d : DTA The point with the smallest deviation from reference point is a quantitative measure of the accuracy of the correspondence -> the quality index, γ (r r ) of the reference point γ (r r ) : 1 ->correspondence is within the specified acceptance criteria Low et al, Med Phys 30(9) (2003).

77 Dose Difference and DTA Dose Difference and DTA Analysis Summary Dose Diff and DTA criteria : 2% of Dmax and 2mm Points Checked = 5348 Points Passed DTA = 5312 Points Passed DD = 4363 Points Passed Either = 5343 Points Passed Both = % of the points passed DTA % of the points passed DoseDiff % of the points passed either Either % of the points passed Both Dose Difference Statistics Summary Mean Dose Diff = DTA Summary Mean DTA =

78 Gamma Analysis Gamma Analysis Summary Dose Diff and DTA criteria : 2% of Dmax and 2mm Points Checked = 5348 Points Passed = % of the points passed Gamma Gamma Statistics Summary GammaBar = Dose Diff and DTA criteria : 3% of Dmax and 3mm Points Checked = 5348 Points Passed = % of the points passed Gamma Gamma Statistics Summary GammaBar =

79 Normalized Agreement Test (NAT) Is based on a 2D array of calculated image of NAT values derived from comparisons of measured and computed doses. Assumes that two dose distribution images are registered each other and NAT is calculated using NAT NAT index scale ( δ 1) = D = A ve( NAT ) Ave( D ) scale δ : lesser of Abs( D/ D m ) or d/ d m D scale : D i /D max NAT index represents the average deviation from the D m and d m criteria for every dose pixel, ignoring the ones less than the set criteria N. Childress et al, The design and testing of noval clinical parameters for dose comparison, Int J. rad. Oncol Biol Phys 56(5) (2003).

80 NAT Index N. Childress et al, The design and testing of noval clinical parameters for dose comparison, Int J. rad. Oncol Biol Phys 56(5) (2003).

81 Other Analysis Tools MU check software In-house dose calc Commercial packages (e.g; Radcalc) Monte Carlo (e.g; Peregrine, EGS4, ) Patient QA Software for Post-treatment QA Analysis of IMRT delivery log files (e.g; inhouse analysis software, Argus IMRT QA package)

82 MC verification IMRT QA Tools Superposition Monte Carlo =10%

83 Summary Multiple detectors and phantoms are typically required for IMRT QA Quantitative dose analysis tools are necessary for proper evaluation of delivery - identify the cause of discrepancies between delivery and measurements Treatment planning vendors are starting to provide dosimetric evaluation tools Aware of the limitations of each tool

84 Summary Verify that all equipment is functioning properly Film processor, digitizer Detectors, cables, electrometers (automatic leakage correction) TLD reader, ovens Input/output to treatment planning system Standardize measurement setup when possible Monitor software and hardware changes and QA

85 Summary Measurements may show dosimetric differences that planning systems may not model at this time curved leaf ends Need to know the limits of the mechanical systems and interactions with controller and accelerator software for delivery Continued need for improvements to software for delivery system, measurement devices, phantoms, and dose analysis tools

86 Acknowledgements Jean Moran U of Michigan Cynthia Chuang UCSF Mark Oldham Duke University

QA of intensity-modulated beams using dynamic MLC log files

QA of intensity-modulated beams using dynamic MLC log files 36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,

More information

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry

More information

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac

More information

Cyclotron Centre in Poland and 2D thermoluminescence dosimetry

Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of

More information

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How

More information

Quality Assurance of Radiotherapy Equipment

Quality Assurance of Radiotherapy Equipment Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria

More information

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,

More information

External dosimetry Dosimetry in new radiotherapeutic techniques

External dosimetry Dosimetry in new radiotherapeutic techniques External dosimetry Dosimetry in new radiotherapeutic techniques Albert Lisbona Medical Physics Department CLCC René Gauducheau 44805 Saint-Herblain France a-lisbona@nantes.fnclcc.fr Objective : To describe

More information

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical

More information

Challenges in small field MV photon dosimetry

Challenges in small field MV photon dosimetry Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland maria.aspradakis@ksl.ch Why are

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

Comparison of Epson scanner quality for radiochromic film evaluation

Comparison of Epson scanner quality for radiochromic film evaluation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Comparison of Epson scanner quality for radiochromic film evaluation Hani Alnawaf, 1,3 Peter K.N. Yu, 2 Martin Butson 1,2,3a South

More information

Clinical Photon Beams

Clinical Photon Beams Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips

More information

Manual for simulation of EB processing. Software ModeRTL

Manual for simulation of EB processing. Software ModeRTL 1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended

More information

Azam Niroomand-Rad, PhD Georgetown University Medical Center Washington D.C. AAPM 41st Annual Meeting Continuing Education Courses

Azam Niroomand-Rad, PhD Georgetown University Medical Center Washington D.C. AAPM 41st Annual Meeting Continuing Education Courses AAPM 41st Annual Meeting Continuing Education Courses Special Dosimetry Measurements-3 Two Dimensional Radiation Field Mapping Using Radiochromic Film Wednesday, July 28, 1999 7:30 AM - 8:30 AM Azam Niroomand-Rad,

More information

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film. INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

More information

Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014

Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014 Multi-Channel Radiochromic Film Dosimetry Adapted from A.Micke Spain, April 2014 Single Channel Film Dosimetry Calibration Curve X=RR ave = R ave (D) D R = D R (R ave ) Color channels X=RGB D X = D( X

More information

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage 3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter

More information

Overview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota

Overview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota Creation of 3D printed phantoms for clinical radiation therapy Eric Ehler, PhD Assistant Professor University of Minnesota ehler 046@umn.edu Overview Background of 3D Printing Practical Information Current

More information

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures

More information

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular

More information

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Performance evaluation and quality assurance of Varian enhanced dynamic wedges JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department

More information

Quality Assurance of accelerators; the technologists responsibility

Quality Assurance of accelerators; the technologists responsibility Quality Assurance of accelerators; the technologists responsibility Christa Timmermans Radiation Technologist, Erasmus MC- Daniel den Hoed, Rotterdam, The Netherlands EORTC-ROG RT technologists Section

More information

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative

More information

Worldwide Quality Assurance networks for radiotherapy dosimetry

Worldwide Quality Assurance networks for radiotherapy dosimetry Worldwide Quality Assurance networks for radiotherapy dosimetry Joanna Izewska Dosimetry and Medical Radiation Physics Section dosimetry@iaea.org Audit in radiotherapy dosimetry Objectives to enhance confidence

More information

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools 3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground

More information

Quality Assurance for Treatment Planning Systems

Quality Assurance for Treatment Planning Systems Quality Assurance for Treatment Planning Systems PTCOG Educational Session May 19 th 2010 Oliver Jäkel Heidelberg Ion Beam Therapy Center and German Cancer Research Center, Heidelberg Introduction Outline

More information

GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2

GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2 GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures

More information

Key words: treatment planning, quality assurance, 3D treatment planning

Key words: treatment planning, quality assurance, 3D treatment planning American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical

More information

Use of 3D Printers in Proton Therapy

Use of 3D Printers in Proton Therapy Use of 3D Printers in Proton Therapy Nicholas Remmes, PhD Mayo Clinic, Rochester, MN Department of Radiation Oncology Remmes.Nicholas@mayo.edu 2015 MFMER slide-1 Disclosures/Confessions My 3D Printing

More information

Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com

Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com Serial Tomotherapy IGRT Total Solutions Treatment Planning Brachytherapy Imaging Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA BMI.v.08.2010 Best NOMOS Your Single Source Oncology Solutions Provider

More information

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50

More information

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,

More information

Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer

Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer Evaluation of complexity and deliverability of IMRT- treatment plans Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer June 11 Abstract Background: Intensity modulated beams are used

More information

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D. 4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology jiang.steve@mgh.harvard.edu http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?

More information

A Thesis. entitled. Retrofitted Micro-MLC SRS System. George H. Hancock

A Thesis. entitled. Retrofitted Micro-MLC SRS System. George H. Hancock A Thesis entitled Development of a Comprehensive Linac-based Quality Assurance Program for a Retrofitted Micro-MLC SRS System. by George H. Hancock Submitted to the Graduate Faculty as partial fulfillment

More information

Secondary Neutrons in Proton and Ion Therapy

Secondary Neutrons in Proton and Ion Therapy Secondary Neutrons in Proton and Ion Therapy L. Stolarczyk Institute of Nuclear Physics PAN, Poland on behalf of WG9 EURADOS Acknowledgments EURADOS Workig Group 9 Roger Harrison Jean Marc Bordy Carles

More information

Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.

Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D. Efficacy of IMRT with dot decimal compensators for radiotherapy of malignant pleural mesothelioma post extrapleural pneumonectomy Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan

More information

Quality Assurance in Stereotactic. Radiotherapy

Quality Assurance in Stereotactic. Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

Acknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers

Acknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers H. Helen Liu, PhD Department of Radiation Physics, U.T. MD Anderson Cancer Center, Houston, TX AAPM, Seattle, 2005 Acknowledgement

More information

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu

More information

1. Provide clinical training in radiation oncology physics within a structured clinical environment.

1. Provide clinical training in radiation oncology physics within a structured clinical environment. Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible

More information

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually

More information

Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care

Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require

More information

ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS

ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS INFN-Laboratori Nazionali del Sud Via S. Sofia 44, Catania, Italy Patient positioned for treatment System under consideration (experimental

More information

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS JOHN L. HORTON Department of Radiation Physics University of Texas

More information

Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021

Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021 Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021 Introduction Prostate is one of the treatment sites that is well suited for IMRT. For

More information

Digital Mammography Update: Design and Characteristics of Current Systems

Digital Mammography Update: Design and Characteristics of Current Systems : Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

Toward Clinically Intuitive Quality Assurance. Hannah Jillian Norris. Graduate Program in Medical Physics Duke University.

Toward Clinically Intuitive Quality Assurance. Hannah Jillian Norris. Graduate Program in Medical Physics Duke University. Toward Clinically Intuitive Quality Assurance by Hannah Jillian Norris Graduate Program in Medical Physics Duke University Date: Approved: Mark Oldham, Supervisor Jackie Wu Robert Reiman Thesis submitted

More information

Transit Dosimetry for Patient Treatment Verification with an Electronic Portal Imaging Device. Sean L Berry

Transit Dosimetry for Patient Treatment Verification with an Electronic Portal Imaging Device. Sean L Berry Transit Dosimetry for Patient Treatment Verification with an Electronic Portal Imaging Device Sean L Berry Submitted in partial fulfillment of the requirements for the degree of Doctor of Philosophy in

More information

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015 INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design

More information

Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification

Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification Valeria Casanova Borca, Massimo Pasquino, a

More information

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.

More information

Scripps Proton Therapy Center: Configuration and Implementation

Scripps Proton Therapy Center: Configuration and Implementation Scripps Proton Therapy Center: Configuration and Implementation Anthony Mascia Proton Symposium 2013 Annual AAPM Meeting Indianapolis, Indiana USA Facility Configuration Scripps Proton Therapy Center ***All

More information

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity

More information

Analysis of Trajectory Log Files of TrueBeam Medical Electron Linear Accelerator for Patient Specific IMRT QA

Analysis of Trajectory Log Files of TrueBeam Medical Electron Linear Accelerator for Patient Specific IMRT QA Cancer Research and Oncology: Open Access Received: Nov 16, 2015, Accepted: Jan 12, 2016, Published: Jan 15, 2016 CancerResOncol, Volume 1, Issue 1 http://crescopublications.org/pdf/crooa/crooa-1-001.pdf

More information

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment. DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering

More information

Measuring Patient Exposure in Interventional Radiology

Measuring Patient Exposure in Interventional Radiology Measuring Patient Exposure in Interventional Radiology GAFCHROMIC XR-R, a Wide-Area Dosimetry Film to Assess Peak Skin Dose (PSD) Xiang Yu, Ph.D. Director, Advanced Materials International Specialty Products

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada

More information

THE DOSIMETRIC EFFECTS OF

THE DOSIMETRIC EFFECTS OF THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,

More information

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed

More information

QA & DOSIMETRY NEWS. Respiratory MotionSim TM. Introducing. Quality Reports [EMR] Plan Quality Management»

QA & DOSIMETRY NEWS. Respiratory MotionSim TM. Introducing. Quality Reports [EMR] Plan Quality Management» QA & DOSIMETRY NEWS Serving the Radiation Oncology Community Since 1984 Published by Sun Nuclear Corporation 2014 ESTRO Edition Quality Reports [EMR] Plan Quality Management» From the CEO The Metrics of

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Volumetric modulated arc therapy (VMAT) for prostate cancer

Volumetric modulated arc therapy (VMAT) for prostate cancer Planning study Volumetric modulated arc therapy (VMAT) for prostate cancer Institution: Purpose: Swedish Cancer Institute, Seattle, WA, USA In a community cancer center, treating prostate cancer comprises

More information

Digital vs. Analogue Control Systems

Digital vs. Analogue Control Systems Digital vs. Analogue Control Systems Presented at the 2011 Annual Meeting of the American College of Medical Physics, Chattanooga, TN, May 1, 2011 Ivan A. Brezovich, PhD, Dept. of Rad Onc, Univ of Alabama

More information

Page: 1 of 6 Page: 1 of 6

Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique

More information

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ. Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ., Raleigh, NC One vital step is to choose a transfer lens matched to your

More information

Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY

Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY A. Mrozik, B. Marczewska, P. Bilski, W. Gieszczyk Institute of Nuclear Physics PAN,

More information

Chapter 3: Radiation Dosimeters

Chapter 3: Radiation Dosimeters Chapter 3: Radiation Dosimeters Set of 113 slides based on the chapter authored by J. Izewska and G. Rajan of the IAEA publication (ISBN 92-0-107304-6): Review of Radiation Oncology Physics: A Handbook

More information

Proton tracking for medical imaging and dosimetry

Proton tracking for medical imaging and dosimetry Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4

More information

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der

More information

Accreditation Is Coming

Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt

More information

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules Abstract J.L. Crozier, E.E. van Dyk, F.J. Vorster Nelson Mandela Metropolitan University Electroluminescence (EL) is a useful

More information

Gated Radiotherapy for Lung Cancer

Gated Radiotherapy for Lung Cancer Gated Radiotherapy for Lung Cancer Steve B. Jiang, Ph.D. Depart Of Radiation Oncology University of California San Diego sbjiang@ucsd.edu radonc.ucsd.edu/research/cart Two Types of Gating Internal gating

More information

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim- QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound

More information

METHODS OF BOLUSING THE TRACHEOSTOMY STOMA

METHODS OF BOLUSING THE TRACHEOSTOMY STOMA PII S0360-3016(00)01550-9 Int. J. Radiation Oncology Biol. Phys., Vol. 50, No. 1, pp. 69 74, 2001 Copyright 2001 Elsevier Science Inc. Printed in the USA. All rights reserved 0360-3016/01/$ see front matter

More information

DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT

DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT BACKGROUND AND PRODUCT DESCRIPTIONS Fracture Diagnostics provides state-of-the-art test equipment, utilizing the Rising Step Load testing technique.

More information

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture

More information

Introduction to CCDs and CCD Data Calibration

Introduction to CCDs and CCD Data Calibration Introduction to CCDs and CCD Data Calibration Dr. William Welsh San Diego State University CCD: charge coupled devices integrated circuit silicon chips that can record optical (and X-ray) light pixel =

More information

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard

More information

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam J. Korean Soc Ther Radiol Oncol : Vol. 15, No. 2, June, 1997 Reduction of Electron Contamination Using a Filter for 6MV Photon Beam Choul Soo Lee, MPH., Myung Jin Yoo, Ph.D. and Ha Yong Yum, M.D. Department

More information

INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS

INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS HOW DOES RADIATION THERAPY WORK? Cancer is the unrestrained growth of abnormal cells within the body. Radiation therapy kills both

More information

MACHINE VISION FOR SMARTPHONES. Essential machine vision camera requirements to fulfill the needs of our society

MACHINE VISION FOR SMARTPHONES. Essential machine vision camera requirements to fulfill the needs of our society MACHINE VISION FOR SMARTPHONES Essential machine vision camera requirements to fulfill the needs of our society INTRODUCTION With changes in our society, there is an increased demand in stateof-the art

More information

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

Digital radiography conquers the veterinary world

Digital radiography conquers the veterinary world Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing

More information

Powder diffraction and synchrotron radiation

Powder diffraction and synchrotron radiation Powder diffraction and synchrotron radiation Gilberto Artioli Dip. Geoscienze UNIPD CIRCe Center for Cement Materials single xl diffraction powder diffraction Ideal powder Powder averaging Textured sample

More information

Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry

Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry N. Gopishankar, 1a

More information

SEAAPM Symposium The Practice of Quality Assurance in an Era of Change

SEAAPM Symposium The Practice of Quality Assurance in an Era of Change SEAAPM Symposium The Practice of Quality Assurance in an Era of Change Directors: Dan Bourland and Ingrid Marshall Vendors, breaks, and food next door Thank you to SEAAPM EXCOM, Commercial Partners, Faculty,

More information

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM

More information

Development of on line monitor detectors used for clinical routine in proton and ion therapy

Development of on line monitor detectors used for clinical routine in proton and ion therapy Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary

More information

Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis

Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis David Bernard & Steve Ainsworth Dage Precision Industries Abstract Non-destructive testing during the manufacture of

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

Dental Radiography Core Subject. Digital Radiography

Dental Radiography Core Subject. Digital Radiography Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

Fundamentals of Cone-Beam CT Imaging

Fundamentals of Cone-Beam CT Imaging Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation

More information

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology product overview family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology scmos knowledge base scmos General Information PCO scmos cameras are a breakthrough

More information