1 Tools for IMRT QA N. Dogan, Ph.D Department of Radiation Oncology Virginia Commonwealth University Medical College of Virginia Hospitals Richmond, VA, USA
2 Objectives To identify the QA tasks involving IMRT To describe the QA tools for all aspects of IMRT process To explain the limitations of the current IMRT QA tools To compare the IMRT QA tools and techniques
3 QA tasks for IMRT Machine QA- Acceptance and routine QA of the MLC for IMRT delivery - dosimetric and geometric characteristics Algorithm QA for IMRT - QA of planning system and data consistency with machine Patient Specific QA prove plan works 1D and 2D dosimetry of treatment components such IM beams and segments 3D dosimetry of entire treatment delivery Post Treatment QA Log-file analysis
4 Detectors Phantoms Scanners Dosimetric Analysis Tools
5 Detector Requirements for IMRT QA Geometric and dosimetric accuracy Volumetric simultaneously integrating dosimeter to faithfully quantify the dose delivered over the total time of treatment Good spatial resolution, tissue equivalent response Ability to provide 3-D information Portability to multiple phantoms Ease of use Sufficiently large dynamic range and be insensitive to photon energy spectrum and dose rate response which is independent of the energy spectrum
6 Detectors Many of them available for IMRT measurements Necessary to characterize the detector response for both static and dynamic fields for linearity Need to be calibrated for absolute measurements Need to determine stem and cable effects
7 Detectors IMRT QA Tools Need to determine energy dependence and angular response Small field detectors required for small field characterization Sensitive to position Detector should be smaller than homogeneous region of dose to be measured Assess electrometer response
8 Detectors, cont. Need to determine necessary resolution depends on the resolution of the beamlet grid that is used for planning and sequencing fields for delivery Chambers with the smaller volumes are more sensitive to position and will have a higher response when positioned at an opposing leaf pair junction and between adjacent leaves More stable measurement point Courtesy of Jean Moran, UofM Dose (cgy) Poor detector position
9 1-D and 2-D Detectors Ion chamber (1-D) TLDs and MOSFETs (1-D) Detector arrays (2-D) Film (2-D) Radiographic Radiochromic Gels (3-D)
10 Small 1-D Detectors Detector Volume (cm 3 ) Diameter (cm) Disadvantages Microchamber Poorer resolution than diodes Pinpoint chamber p-type Si diode Over-respond to low energy photons Martens et al Stereotactic diode NA 0.45 MOSFET NA NA Non-linear dose response for <30 cgy Diamond < resolution than diodes, dose rate dependence, expensive
11 Ion Chamber Advantages IMRT QA Tools Available in different shape and sizes Dosimetric response is well understood. Absolute dose measurements theory is well establish, they can be used as a benchmark standard Easy to calibrate
12 Ion Chamber, cont. Disadvantages Only one measurement point for each irradiation does not yield sufficient information to evaluate the dose throughout the target and/or critical structures Volume averaging the measurements are to be considered as an average throughout the chamber s active volume - does not yield significant errors if the ion chamber is placed in a low dose-gradient region even for relatively large chambers
13 Ion Chamber volume averaging, cont. Micro cham: 0.009cc PTW: 0.125cc Farmer:0.65cc D.A. Low et al. Ionization chamber volume averaging effects in dynamic intensity modulated radiation therapy beams, Med. Phys.30(7): (2003
14 TLDs Advantages Multiple measurement points in a single irradiation Reusable Easy to use in multiple phantoms Small size and versatility in placement Readily available readout equipment Achievable accuracy: 2-3%
15 TLDs Disadvantages Requires calibration to determine calibration factor for each TLD chip Requires calibration of subset of TLD chips for each measurement TLD reader response and oven temperature should me routinely monitored to maintain consistent TLD response Automatic reader recommended for IMRT field verification due to large number of TLDs required for verification in a plane (60 or more) inefficient for routine IMRT QA
16 TLDs, cont. IMRT QA Tools D.A. Low et al. Phantoms for IMRT Dose Distribution Measurement and Treatment Verification, Int J Radiat Oncol Biol Phys 40: (1998).
17 MOSFET systems Advantages Excellent spatial resolution small size (~0.04mm 2 ) Multiple detectors can be irradiated simultaneously Automatic and immediate readout Can be re-used immediately Linear dose response > 30 cgy Response independent of depth Commercially available phantoms to accommodate the small detectors
18 MOSFET systems Disadvantages Decrease linearity for < 30 cgy limited to high dose applications Over-response for the phantom scatter factor for small fields Specific application and measurement conditions should be carefully assessed and the detector should be used in the appropriate dose range
19 MOSFET systems IMRT QA Tools Bias Box Reader MOSFET TNRD50 system Courtesy of Cynthia Chuang, UCSF An axial image of MOSFET phantom
20 MOSFET systems IMRT QA Tools 1400 MOSFET Linearity Mosfet Consistency 1200 MOSFET1 MOSFET2 MOSFET3 MOSFET MOSFET1 MOSFET2 MOSFET Radiation (cgy) Number of Measurements Courtesy of Cynthia Chuang, UCSF
21 MOSFET systems IMRT QA Tools Percent Depth Dose Comparison Ion Chamber MOSFET Angular Dependence Percentage (%) Degrees Depth (cm) Courtesy of Cynthia Chuang, UCSF
22 Cal Gy Meas Gy Diff 4.6 % Cal Gy Meas Gy Diff % Courtesy of Cynthia Chuang, UCSF Calc Gy Meas Gy Diff 4.35% Calc Gy Meas Gy Diff 3.52% Calc Gy Meas Gy Diff 3.45%
23 Current IMRT QA Tools 2-D Detectors Film Radiographic Radiochromic Beam imaging system, CCD, SLIC, AMFPI 2-D Detector arrays Diode array (Mapcheck) Ion chamber Active matrix flat panel detector (AMFPD)
24 Radiographic Film Advantages IMRT QA Tools Readily available (XV, EDR2, ) Can be cut into any desired shape Excellent spatial resolution (<1mm) Less expensive than other 2-D systems
25 Radiographic Film, cont. Disadvantages Over-response to low energy x-rays high atomic number of the active material not good for absolute dosimetry Dependent on QA of film batch Dependent on processor and digitizer Sensitive to storage conditions Need to measure the response to dose for each experiment H&D curve each time Proper normalization is critical
26 Current IMRT QA Tools Radiographic Film, cont. Other issues Store in a cool and dry place Make sure that the temperature for the film processor is stable Film digitizer pixel spacing, integrity of OD, beware of artifacts Verify spatial and optical density accuracy
27 Rapid Film Calibration 120 MU 90 MU 60 MU 30 MU 240 MU 210 MU 180 MU 150 MU Multiple dose levels per film-3x3 cm 2 fields of different dose levels Step-and-shoot or SMLC delivery Different dose values required for XV and EDR2 film (15-120MU for XV and MU for EDR2) Saves both time and film Childress et al Med Phys 29(10), 2002.
28 XV vs. EDR Film Depth-corrected H&D Net Optical Density XV EDR XV2, 6 MV XV2, 15 MV EDR2, 6 MV EDR2, 15 MV Optical Density Co60-EDR2 10MV-EDR2 6MV-EDR2 18MV-EDR Dose (cgy) Chetty and Charland 2002 PMB 47: Dose (cgy) Dogan et al PMB 47:
29 XV Depth-corrected H&D curves MV-EDR2-Depth corrected 6MV-EDR2-Regular EDR MV-EDR2-Depth corrected 18MV-EDR2-Regular Optical Density Dose (cgy) Optical Density Dose (cgy) Dogan et al PMB 47:
30 Ion Chamber Film- depth corrected H&D Film regular H&D Dogan et al PMB 47: (a) (b) Ion chamber and EDR2 film depth-dose curves for a) 6 x 6 cm2, b) 14 x 14 cm2 films for 10 MV beam. Films were positioned parallel to the beam and OD to dose conversion was done using regular and depth-corrected H&D curves.
31 Childress et al. Med. Phys. 32(2) 2005
32 As compared to XV film, EDR2 film has less dependence on the processor, field size less response to low energy photons have better reproducibility and agreement with ion chamber measurements can be used to measure a complete fraction of an IMRT treatment
33 Radiographic Film: 2-D Dosimetric Measurements Intensity map from Opt System Calculated Leaf Sequencer Calc-Meas Courtesy of Jean Moran, UM Measured
34 Radiographic Film: Routine DMLC QA Using radiographic films Intensity-modulated pattern field Check leaf position, acceleration, motion stability Check for hot and cold density Visual check DMLC field 14x14 cm 2 at SSD =100 cm, 2 cm separated strips
35 Film Processor issues Should do routine maintenance and quality assurance verify spatial intensity, characteristic response, noise due to large changes in optical density ( Dempsey et al, Med Phys, 26; , 1999). Should be warmed up prior to use Should have appropriate amount of chemicals - Several films should be run in advance Should have stable temperature Should have a consistent rate of feeding into the processor
36 Film Other Issues Accurate positioning of the film in the phantom for the registration with treatment planning system Minimized errors by using a solid-water slab designed for film Have pins between slabs that puncture the film
37 Radiochromic Film (RCF) Advantages No significant energy dependence decreased sensitivity to low-energy photons Insensitivity to visible light Very high spatial resolution - well-suited for measurements in high-dose gradient fields Self-developing no developer or fixer is required Easy to handle Tissue equivalent
38 Radiochromic Film, cont. Disadvantages Takes a couple of hours for the color change to stabilize, and it may be necessary to wait up to two days before evaluating the film Sensitive to the air temperature and humidity Ultraviolet light may cause a color change without exposure to ionizing radiation Size, availability, and cost Non-uniform response to radiation double exposure technique minimizes this effect Issues with thermal history, wavelength dependence, and local sensitivity of the film
39 RCF (Gafchromic HS and MD55-2) vs radiographic films (XV and EDR2) O. Zeidan et al., Med. Phys., 31 (10): (2004)
40 RCF profiles vs. Ion chamber J. Dempsey et al., Med. Phys., 27 (10): (2000)
41 RCF digitizer issues Response of the digitizer Light source characteristics Design Gluckman et al, Med Phys, 29(8); , 2002.
42 Other 2-D systems IMRT QA Tools Beam imaging system, CCD, SLIC, Amorphous silicon flat panel detector (AMFPD) EPID systems attached to gantry Investigated more for pre-treatment QA currently 2-D Detector arrays Diode array (e.g; MapCheck) Ion chamber (e.g; LA48 linear array)
43 EPID Systems IMRT QA Tools Charged coupled device (CCD) camera systems Scanning liquid ion chambers (SLICs) Amorphous silicon flat panel detector (AMFPD) Active matrix flat panel imagers (AMFPIs) Transit Dosimetry Patient or Phantom Pre-Tx 2-D Measurements Film Replacement
44 EPID Systems as500 EPID 1 mm copper plate Phosphor scintillating layer (Kodak Lanex Fast B Gd2O2S:Tb, 70 mg/cm3) Array of photodiodes Amorphous Silicon panel each pixel consists of: IMRT QA Tools Light sensitive photodiode Thin film transistor 16-bit ADC Munro et. al, Med. Phys. 25, 1998
45 EPIDs Advantages Many centers have installed EPIDs and being primarily used for patient-specific pretreatment field verification and MLC QA Logical extension to investigate dosimetric applications Mounted to linear accelerator - known geometry with respect to the beam Detector sag must be accounted for at different gantry angles Positioning reproducibility important Real time digital evaluation No processor, data acquisition takes less time
46 EPIDs - Challenges EPIDs were primarily designed for patient localization High resolution, good contrast images Additional dose to the patient should be minimized The conversion of imager response to dose is complex Imaging system dependent Other problems Ghosting Lag
47 EPIDs Dose determination Imager response must be calibrated to a standard Absolute calibration to ion chamber at a point over a ROI E.g. ion chamber in a mini-phantom or slab at same SDD as EPID 2-D calibration to actual beam distribution at the imager plane Can be measured with film or a diode array
48 Factors for EPID Response Water-equivalent depth of the detector Field size dependence and scatter properties within the imager Short- and long-term reproducibility Dose rate Energy dependence Spatial integrity
49 EPID: DMLC measurements Overall: Good agreement + Predicted EPID Ion Chamber 10 MV 25 MV Discrepancies in the penumbra region (up to 10%) Pasma Med Phys 26: (2376) 1999
50 Linear Diode Array in water vs. CCD Without short range penumbra correction Courtesy of Jean Moran, UofM
51 Dose Determination using EPID (SLIC) Chang et al., Int J Radiat Oncol Phys 47: (p. 233)
52 Calculation vs. measured using AMFPD for DMLC Calculated (Calculated Measured) Agreement : Within +/- 2 cgy Courtesy of Jean Moran, UofM
53 EPIDs can provide a much-needed replacement for pre-tx QA film dosimetry Only if proper QA of the EPID is established Need better understanding of regions where EPIDs are inadequate for dosimetry Systems must be verified at more centers against accepted QA methods such as film and ion chamber Additional software is required before more facilities can do proper validation of the methods (Software must be commissioned) Can be part of a comprehensive QA program in conjunction with other methods such as computational checks (monitor programs, log file analysis, etc.)
54 Gel Dosimeters Advantages 3-D information in one irradiation Energy and dose-rate independent High sensitivity and linear response Cumulative Gel density can be changed - Ideal for anthropomorphic phantoms Near tissue equivalent Multiple readout techniques (MR, optical-ct) New gel formulations and readers commercially available
55 Gel Dosimeters Disadvantages Sensitive to time, preparation, temperature Cylindrical container required for optical readers - less accurate readout at gel/container interface MR time is often limited and expensive - long scan times for accurate readout, e.g. 5% accuracy over 10 hr scan time (Gum et al. 2002) Relative dosimeter -require cross-calibration technique batch to batch they are different Cost
56 Gel dosimetry IMRT QA Tools 8cm In-house optical CT scanner cost is less Oldham and Kim, Med. Phys. 31 (5), Upgraded motors, motion control, and user interface. (Pacific Scientific: step motors. National Instruments: motion control and Labview.)
57 Gels: Optical Density to Dose Calibration 6 Beam calibration irradiation BANG gel phantom diameter 17.4cm Courtesy of Mark Oldham, Duke University
58 Gel Dosimeters IMRT QA Tools Five Field Prostate IMRT Courtesy of Mark Oldham, Duke University Re-computed for a 3 L BANG gel dosimeter. Dmax scaled to 1.8 Gy to fit dynamic range of optical scanner BANGkit TM from MGS Research. 1x1x3mm, 5hours
59 Gel Dosimeters IMRT QA Tools Isodose comparison: Pinnacle (red), Gel-dosimetry (black) Courtesy of Mark Oldham, Duke University
60 Gel Dosimeters IMRT QA Tools Gum, et al. Preliminary study on the use of an inhomogeneous anthropomorphic Fricke gel phantom and 3D magnetic resonance dosimetry for verification of IMRT plans, Phys Med Biol 47; N
61 Phantoms for IMRT Measurements multiple phantoms for commissioning Fiducials for reproducible setup of phantom and detectors User-customized for different detectors allow special holders Simple vs. anthropomorphic Homogeneous or heterogeneous
62 Current IMRT QA Tools Simple Geometric Phantoms Water tank Accommodate different ion chambers Use for measurements of depth dose and profiles Output, flatness, symmetry, and linearity assessment Cylindrical mini-phantom Use with ion chamber to assess dependence of output on gantry angle Water-equivalent plastics: slab w/ custom chamber inserts 1-D and 2-D measurements Detector position can be varied with depth Cylindrical phantoms (plastic or water filled) Straightforward geometry Ion chamber at single position Plastic phantoms may hold films
63 Water-equivalent square IMRT Verification Phantom D.A. Low et al. Phantoms for IMRT Dose Distribution Measurement and Treatment Verification, Int J Radiat Oncol Biol Phys 40: (1998).
64 Current IMRT QA Tools A Cylindrical Phantom containing movable ion chamber L. Xing et al. Dosimetric verification of a commercial inverse treatment planning system, Phys. Med. Biol. 44: (1998).
65 A cylindrical Plastic Phantom Detector
66 Plastic Cylindrical Phantom with MOSFETs Calc Gy Meas Gy Diff 3.52% Calc Gy Meas Gy Diff 3.45% Courtesy of Cynthia Chuang, UCSF
67 Spiral Phantom IMRT QA Tools Paliwal et al A spiral phantom for IMRT and tomotherapy treatment delivery verification Med Phys (2000).
68 Anthropomorphic: RPC Head Phantom Target Volumes Water Critical Structure Removable Dry Insert Water TLDs in Target Volumes Courtesy of Jean Moran, UofM Radiochromic film through multiple plans Delivery is required by RTOG for participation in IMRT trials
69 Dosimetric Analysis Tools Provide a comprehensive and quantitative comparison between two dose distributions Different ones available Important to know the limitations
70 Dosimetric Analysis Tools Overlay of isodoses 2-D dose difference displays with colorwash Dose difference histograms Distance-to-agreement (DTA) Gamma evaluation Normalized agreement test (NAT)
71 Isodose lines and Dose Difference Display 70 cgy 60 cgy 50 cgy 20 cgy 10 cgy Calcs Film +/- 10% Courtesy of Jean Moran, UofM
72 Dose difference display Useful in shallow dose gradients Overly sensitive in steep dose gradients e.g.; a small spatial shift (due to experimental measurement errors) between two dose distributions yield large dose differences
73 Dose difference histogram and profiles
74 Distance to Agreement (DTA) Is the distance between a reference point and the nearest point in the compared dose distribution that exhibits the same dose Is not overly sensitive in steep dose gradients In shallow dose gradients, a large DTA value may be computed even for relatively small dose differences May be hard to interpret
75 Combination of dose difference and DTA Identify regions where the dose difference and DTA are simultaneously by greater than a pre-selected criteria points that fail both criteria are identified on a composite distribution The display of the dose difference may emphasize the impression of failure in high dose gradient region Provides no information on the magnitude of the failure
76 Gamma Analysis- Generalization of composite distribution Measures the closest distance between each reference point and evaluated dose distribution after scaling by D and d r ( r, r ) δ ( r, r ) Γ ( r, r ) = + = Γ r r r d D 2 2 e r e r e r γ (r 2 2 r ) m in ( e, r ) e { } { } rr (,): e r spatial distance between evaluated and reference dose points D : Dose difference criteria d : DTA The point with the smallest deviation from reference point is a quantitative measure of the accuracy of the correspondence -> the quality index, γ (r r ) of the reference point γ (r r ) : 1 ->correspondence is within the specified acceptance criteria Low et al, Med Phys 30(9) (2003).
77 Dose Difference and DTA Dose Difference and DTA Analysis Summary Dose Diff and DTA criteria : 2% of Dmax and 2mm Points Checked = 5348 Points Passed DTA = 5312 Points Passed DD = 4363 Points Passed Either = 5343 Points Passed Both = % of the points passed DTA % of the points passed DoseDiff % of the points passed either Either % of the points passed Both Dose Difference Statistics Summary Mean Dose Diff = DTA Summary Mean DTA =
78 Gamma Analysis Gamma Analysis Summary Dose Diff and DTA criteria : 2% of Dmax and 2mm Points Checked = 5348 Points Passed = % of the points passed Gamma Gamma Statistics Summary GammaBar = Dose Diff and DTA criteria : 3% of Dmax and 3mm Points Checked = 5348 Points Passed = % of the points passed Gamma Gamma Statistics Summary GammaBar =
79 Normalized Agreement Test (NAT) Is based on a 2D array of calculated image of NAT values derived from comparisons of measured and computed doses. Assumes that two dose distribution images are registered each other and NAT is calculated using NAT NAT index scale ( δ 1) = D = A ve( NAT ) Ave( D ) scale δ : lesser of Abs( D/ D m ) or d/ d m D scale : D i /D max NAT index represents the average deviation from the D m and d m criteria for every dose pixel, ignoring the ones less than the set criteria N. Childress et al, The design and testing of noval clinical parameters for dose comparison, Int J. rad. Oncol Biol Phys 56(5) (2003).
80 NAT Index N. Childress et al, The design and testing of noval clinical parameters for dose comparison, Int J. rad. Oncol Biol Phys 56(5) (2003).
81 Other Analysis Tools MU check software In-house dose calc Commercial packages (e.g; Radcalc) Monte Carlo (e.g; Peregrine, EGS4, ) Patient QA Software for Post-treatment QA Analysis of IMRT delivery log files (e.g; inhouse analysis software, Argus IMRT QA package)
82 MC verification IMRT QA Tools Superposition Monte Carlo =10%
83 Summary Multiple detectors and phantoms are typically required for IMRT QA Quantitative dose analysis tools are necessary for proper evaluation of delivery - identify the cause of discrepancies between delivery and measurements Treatment planning vendors are starting to provide dosimetric evaluation tools Aware of the limitations of each tool
84 Summary Verify that all equipment is functioning properly Film processor, digitizer Detectors, cables, electrometers (automatic leakage correction) TLD reader, ovens Input/output to treatment planning system Standardize measurement setup when possible Monitor software and hardware changes and QA
85 Summary Measurements may show dosimetric differences that planning systems may not model at this time curved leaf ends Need to know the limits of the mechanical systems and interactions with controller and accelerator software for delivery Continued need for improvements to software for delivery system, measurement devices, phantoms, and dose analysis tools
86 Acknowledgements Jean Moran U of Michigan Cynthia Chuang UCSF Mark Oldham Duke University
QA of intensity-modulated beams using dynamic MLC log files
36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,
More informationUse of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
More informationRapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
More informationCyclotron Centre in Poland and 2D thermoluminescence dosimetry
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of
More informationPost Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y
Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How
More informationQuality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
More informationkv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
More informationExternal dosimetry Dosimetry in new radiotherapeutic techniques
External dosimetry Dosimetry in new radiotherapeutic techniques Albert Lisbona Medical Physics Department CLCC René Gauducheau 44805 Saint-Herblain France a-lisbona@nantes.fnclcc.fr Objective : To describe
More informationM D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
More informationChallenges in small field MV photon dosimetry
Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland maria.aspradakis@ksl.ch Why are
More informationAn Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
More informationComparison of Epson scanner quality for radiochromic film evaluation
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Comparison of Epson scanner quality for radiochromic film evaluation Hani Alnawaf, 1,3 Peter K.N. Yu, 2 Martin Butson 1,2,3a South
More informationClinical Photon Beams
Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips
More informationManual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
More informationAzam Niroomand-Rad, PhD Georgetown University Medical Center Washington D.C. AAPM 41st Annual Meeting Continuing Education Courses
AAPM 41st Annual Meeting Continuing Education Courses Special Dosimetry Measurements-3 Two Dimensional Radiation Field Mapping Using Radiochromic Film Wednesday, July 28, 1999 7:30 AM - 8:30 AM Azam Niroomand-Rad,
More informationPerformance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
More informationMulti-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014
Multi-Channel Radiochromic Film Dosimetry Adapted from A.Micke Spain, April 2014 Single Channel Film Dosimetry Calibration Curve X=RR ave = R ave (D) D R = D R (R ave ) Color channels X=RGB D X = D( X
More information3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage
3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter
More informationOverview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota
Creation of 3D printed phantoms for clinical radiation therapy Eric Ehler, PhD Assistant Professor University of Minnesota ehler 046@umn.edu Overview Background of 3D Printing Practical Information Current
More informationGAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3
GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures
More informationRADIATION THERAPY. Dosimetry Pioneers since 1922 NEW
RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular
More informationPerformance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
More informationQuality Assurance of accelerators; the technologists responsibility
Quality Assurance of accelerators; the technologists responsibility Christa Timmermans Radiation Technologist, Erasmus MC- Daniel den Hoed, Rotterdam, The Netherlands EORTC-ROG RT technologists Section
More informationQUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
More informationWorldwide Quality Assurance networks for radiotherapy dosimetry
Worldwide Quality Assurance networks for radiotherapy dosimetry Joanna Izewska Dosimetry and Medical Radiation Physics Section dosimetry@iaea.org Audit in radiotherapy dosimetry Objectives to enhance confidence
More information3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
More informationQuality Assurance for Treatment Planning Systems
Quality Assurance for Treatment Planning Systems PTCOG Educational Session May 19 th 2010 Oliver Jäkel Heidelberg Ion Beam Therapy Center and German Cancer Research Center, Heidelberg Introduction Outline
More informationGAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2
GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures
More informationKey words: treatment planning, quality assurance, 3D treatment planning
American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical
More informationUse of 3D Printers in Proton Therapy
Use of 3D Printers in Proton Therapy Nicholas Remmes, PhD Mayo Clinic, Rochester, MN Department of Radiation Oncology Remmes.Nicholas@mayo.edu 2015 MFMER slide-1 Disclosures/Confessions My 3D Printing
More informationTotal Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com
Serial Tomotherapy IGRT Total Solutions Treatment Planning Brachytherapy Imaging Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA BMI.v.08.2010 Best NOMOS Your Single Source Oncology Solutions Provider
More informationMLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline
Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50
More informationNIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,
More informationEvaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer
Evaluation of complexity and deliverability of IMRT- treatment plans Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer June 11 Abstract Background: Intensity modulated beams are used
More information4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.
4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology jiang.steve@mgh.harvard.edu http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?
More informationA Thesis. entitled. Retrofitted Micro-MLC SRS System. George H. Hancock
A Thesis entitled Development of a Comprehensive Linac-based Quality Assurance Program for a Retrofitted Micro-MLC SRS System. by George H. Hancock Submitted to the Graduate Faculty as partial fulfillment
More informationSecondary Neutrons in Proton and Ion Therapy
Secondary Neutrons in Proton and Ion Therapy L. Stolarczyk Institute of Nuclear Physics PAN, Poland on behalf of WG9 EURADOS Acknowledgments EURADOS Workig Group 9 Roger Harrison Jean Marc Bordy Carles
More informationDepartment of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.
Efficacy of IMRT with dot decimal compensators for radiotherapy of malignant pleural mesothelioma post extrapleural pneumonectomy Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan
More informationQuality Assurance in Stereotactic. Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationAcknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers
Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers H. Helen Liu, PhD Department of Radiation Physics, U.T. MD Anderson Cancer Center, Houston, TX AAPM, Seattle, 2005 Acknowledgement
More informationFeasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
More information1. Provide clinical training in radiation oncology physics within a structured clinical environment.
Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible
More informationClinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
More informationPrecise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care
Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require
More informationON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS
ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS INFN-Laboratori Nazionali del Sud Via S. Sofia 44, Catania, Italy Patient positioned for treatment System under consideration (experimental
More informationCHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS JOHN L. HORTON Department of Radiation Physics University of Texas
More informationProstate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021
Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021 Introduction Prostate is one of the treatment sites that is well suited for IMRT. For
More informationDigital Mammography Update: Design and Characteristics of Current Systems
: Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,
More informationPhysics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
More informationToward Clinically Intuitive Quality Assurance. Hannah Jillian Norris. Graduate Program in Medical Physics Duke University.
Toward Clinically Intuitive Quality Assurance by Hannah Jillian Norris Graduate Program in Medical Physics Duke University Date: Approved: Mark Oldham, Supervisor Jackie Wu Robert Reiman Thesis submitted
More informationTransit Dosimetry for Patient Treatment Verification with an Electronic Portal Imaging Device. Sean L Berry
Transit Dosimetry for Patient Treatment Verification with an Electronic Portal Imaging Device Sean L Berry Submitted in partial fulfillment of the requirements for the degree of Doctor of Philosophy in
More informationTom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015
INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design
More informationDosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification Valeria Casanova Borca, Massimo Pasquino, a
More informationRecognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
More informationScripps Proton Therapy Center: Configuration and Implementation
Scripps Proton Therapy Center: Configuration and Implementation Anthony Mascia Proton Symposium 2013 Annual AAPM Meeting Indianapolis, Indiana USA Facility Configuration Scripps Proton Therapy Center ***All
More informationPET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
More informationAnalysis of Trajectory Log Files of TrueBeam Medical Electron Linear Accelerator for Patient Specific IMRT QA
Cancer Research and Oncology: Open Access Received: Nov 16, 2015, Accepted: Jan 12, 2016, Published: Jan 15, 2016 CancerResOncol, Volume 1, Issue 1 http://crescopublications.org/pdf/crooa/crooa-1-001.pdf
More informationIGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
More informationMeasuring Patient Exposure in Interventional Radiology
Measuring Patient Exposure in Interventional Radiology GAFCHROMIC XR-R, a Wide-Area Dosimetry Film to Assess Peak Skin Dose (PSD) Xiang Yu, Ph.D. Director, Advanced Materials International Specialty Products
More informationIndependent corroboration of monitor unit calculations performed by a 3D computerized planning system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and
More informationMODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada
More informationTHE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
More informationCorrecting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed
More informationQA & DOSIMETRY NEWS. Respiratory MotionSim TM. Introducing. Quality Reports [EMR] Plan Quality Management»
QA & DOSIMETRY NEWS Serving the Radiation Oncology Community Since 1984 Published by Sun Nuclear Corporation 2014 ESTRO Edition Quality Reports [EMR] Plan Quality Management» From the CEO The Metrics of
More informationINTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
More informationVolumetric modulated arc therapy (VMAT) for prostate cancer
Planning study Volumetric modulated arc therapy (VMAT) for prostate cancer Institution: Purpose: Swedish Cancer Institute, Seattle, WA, USA In a community cancer center, treating prostate cancer comprises
More informationDigital vs. Analogue Control Systems
Digital vs. Analogue Control Systems Presented at the 2011 Annual Meeting of the American College of Medical Physics, Chattanooga, TN, May 1, 2011 Ivan A. Brezovich, PhD, Dept. of Rad Onc, Univ of Alabama
More informationPage: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
More informationChoosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.
Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ., Raleigh, NC One vital step is to choose a transfer lens matched to your
More informationSubmitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY
Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY A. Mrozik, B. Marczewska, P. Bilski, W. Gieszczyk Institute of Nuclear Physics PAN,
More informationChapter 3: Radiation Dosimeters
Chapter 3: Radiation Dosimeters Set of 113 slides based on the chapter authored by J. Izewska and G. Rajan of the IAEA publication (ISBN 92-0-107304-6): Review of Radiation Oncology Physics: A Handbook
More informationProton tracking for medical imaging and dosimetry
Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4
More informationIn room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der
More informationAccreditation Is Coming
Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt
More informationHigh Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules
High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules Abstract J.L. Crozier, E.E. van Dyk, F.J. Vorster Nelson Mandela Metropolitan University Electroluminescence (EL) is a useful
More informationGated Radiotherapy for Lung Cancer
Gated Radiotherapy for Lung Cancer Steve B. Jiang, Ph.D. Depart Of Radiation Oncology University of California San Diego sbjiang@ucsd.edu radonc.ucsd.edu/research/cart Two Types of Gating Internal gating
More informationTISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
More informationMETHODS OF BOLUSING THE TRACHEOSTOMY STOMA
PII S0360-3016(00)01550-9 Int. J. Radiation Oncology Biol. Phys., Vol. 50, No. 1, pp. 69 74, 2001 Copyright 2001 Elsevier Science Inc. Printed in the USA. All rights reserved 0360-3016/01/$ see front matter
More informationDIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT
DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT BACKGROUND AND PRODUCT DESCRIPTIONS Fracture Diagnostics provides state-of-the-art test equipment, utilizing the Rising Step Load testing technique.
More informationPersonal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture
More informationIntroduction to CCDs and CCD Data Calibration
Introduction to CCDs and CCD Data Calibration Dr. William Welsh San Diego State University CCD: charge coupled devices integrated circuit silicon chips that can record optical (and X-ray) light pixel =
More informationHigh-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound
High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard
More informationReduction of Electron Contamination Using a Filter for 6MV Photon Beam
J. Korean Soc Ther Radiol Oncol : Vol. 15, No. 2, June, 1997 Reduction of Electron Contamination Using a Filter for 6MV Photon Beam Choul Soo Lee, MPH., Myung Jin Yoo, Ph.D. and Ha Yong Yum, M.D. Department
More informationINTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS
INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS HOW DOES RADIATION THERAPY WORK? Cancer is the unrestrained growth of abnormal cells within the body. Radiation therapy kills both
More informationMACHINE VISION FOR SMARTPHONES. Essential machine vision camera requirements to fulfill the needs of our society
MACHINE VISION FOR SMARTPHONES Essential machine vision camera requirements to fulfill the needs of our society INTRODUCTION With changes in our society, there is an increased demand in stateof-the art
More informationCoating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
More informationQuality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
More informationScan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
More informationDigital radiography conquers the veterinary world
Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing
More informationPowder diffraction and synchrotron radiation
Powder diffraction and synchrotron radiation Gilberto Artioli Dip. Geoscienze UNIPD CIRCe Center for Cement Materials single xl diffraction powder diffraction Ideal powder Powder averaging Textured sample
More informationIndigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry N. Gopishankar, 1a
More informationSEAAPM Symposium The Practice of Quality Assurance in an Era of Change
SEAAPM Symposium The Practice of Quality Assurance in an Era of Change Directors: Dan Bourland and Ingrid Marshall Vendors, breaks, and food next door Thank you to SEAAPM EXCOM, Commercial Partners, Faculty,
More informationMicro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM
More informationDevelopment of on line monitor detectors used for clinical routine in proton and ion therapy
Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary
More informationComparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis
Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis David Bernard & Steve Ainsworth Dage Precision Industries Abstract Non-destructive testing during the manufacture of
More informationDose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography
More informationDental Radiography Core Subject. Digital Radiography
Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of
More informationThinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
More informationFundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
More informationproduct overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology
product overview family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology scmos knowledge base scmos General Information PCO scmos cameras are a breakthrough
More information