Nasal Pressure Airflow Measurement

Size: px
Start display at page:

Download "Nasal Pressure Airflow Measurement"

Transcription

1 Nasal Pressure Airflow Measurement An Introduction Dr. David Rapoport, M.D. Robert Norman, M.S., R.R.T. Michael Nielson, R.R.T., R.PSG.T. Copyright 2001 by Pro-Tech Services, Inc. All rights reserved Pro-Tech Services, Inc Harbour Pointe Blvd. S.W. Mukilteo, WA (425) (800) Fax: (425) Printed in the USA

2 Preface The Polysomnogram, the Gold Standard test in the field of Sleep Medicine, has remained largely unchanged since the 1950 s. With the recognition of sleep disordered breathing as a common disease the focus has shifted beyond what was the original exclusive use of the EEG to characterize sleep, with an incidental interest in respiration, to a greater emphasis on respiratory monitoring. The tools to detect respiration, however, have only recently begun to evolve. Consequently, many laboratories still rely almost exclusively on nasal and oral thermistors (or thermocouples) to detect the presence and nature of breathing. Newer technologies for the measurement of airflow and detection of elevated upper airway resistance include the nasal cannula/pressure transducer system. The intensified interest in nasal cannula monitoring and the fact that many technologists are unfamiliar with its use have led us to conclude that a primer on this technique would be beneficial to provide information on the proper use and interpretation of the nasal pressure waveform. Dr. David Rapoport, M.D. Robert Norman, M.S., R.R.T. Michael Nielson, R.R.T., R.PSG.T.

3 Contents Preface... iii Chapter 1: A Brief History of Respiratory Monitoring on NPSG...1 Chapter 2: Introducing Nasal Pressure Airflow Measurement...3 Chapter 3: How to Record the Nasal Pressure Signal...5 Chapter 4: Extraction of the Snoring Signal from the Pressure Waveform...9 Chapter 5: Common Problems...11 Chapter 6: Nasal Pressure Airflow and NCPAP...15 Chapter 7: Scoring Atlas...21 Chapter 8: Frequently Asked Questions...33 Bibliography and Acknowledgements...43

4 Chapter 1 A Brief History of Respiratory Monitoring on NPSG Prior to the recognition of the syndrome of Obstructive Sleep Apnea (OSA), the main purpose of measuring respiration was to detect the timing of breaths. As the need to detect apnea increased during the 1970 s, the thermistor/thermocouple came into their own because they were ideal tools to identify the total absence of airflow. Their ease of use and comfort on patients has led to their being accepted as an integral tool in polysomnography. However, it is important to understand what is being detected and the limitations of this type of technology. By definition, thermistors and thermocouples are not airflow sensors but are simply detecting the temperature oscillation (between inhaled room air temperature and warmer exhaled air) as a qualitative indicator of airflow. Total cessation of airflow (apnea) is well detected because of the complete lack of temperature variation. In the mid 1980 s it became clear that it was clinically necessary to detect and count partial airway obstruction events, and the term hypopnea was introduced. Although the original description of hypopnea emphasized significant reductions in airflow it was natural to try to extract this information from thermistors and thermocouples that were in extensive use in most laboratories. However, in the case of incomplete airway obstruction, the relationship between temperature change and true flow is not straightforward. Thus the amplitude of the signal resulting from a partially obstructed breath may be reduced not at all (if the full effect of exhaled warm air is detected) or to a variable extent (depending on characteristics of the sensor and the way in which the signal is processed, e.g., filtered). An example of the lack of relationship between true flow amplitude of breaths and the signal obtained from a thermistor is shown on the left graph of Figure 1.1. PEAK INSPIRATORY FLOW (SIMULTANEOUS COLLECTION) Thermister Amplitude (% full scale) Pneumotach Flow (% full scale) Nasal Cannula Flow (% full scale) Pneumotach Flow (% full scale) Figure 1.1

5 Despite the limitations of the thermistor/thermocouple, their use has persisted until recently as the usual clinical practice in most laboratories, and scoring of hypopnea has evolved to the following practice: events of reduced airflow are first identified from the respiratory (thermistor) signal; they are then validated as true hypopneas by their coincidence with desaturation or arousal on EEG. Typical definitions in use today required a decrease in the thermistor signal (50 75%) associated with oxyhemoglobin desaturation of 2% 4% or EEG microarousal. However, if the thermal sensor does not detect a candidate respiratory event and there is no desaturation, arousals are left unexplained. As this pattern has been recognized in more symptomatic patients, this has led to the recognition of the Upper Airway Resistance Syndrome (UARS) - defined as the presence of repetitive arousals following periods of increased respiratory effort (detected by esophageal manometry) but undetected by thermal sensors. The degree to which thermistors under-detect respiratory events has recently been investigated. Up to 70% of respiratory events may be missed in patients with symptoms consistent with OSA but without frank apnea/hypopnea. In some cases, the number of events missed with simple thermal technology may be sufficient to alter the ultimate 4, 12 patient diagnosis. A recent position paper from the American Academy of Sleep Medicine (AASM) reemphasized that the detection of reduced airflow (the entry into defining a hypopnea) could not be performed with thermistors or thermocouples and that these devices should be reserved for the detection of apnea only. 1

6 Chapter 2 Introducing Nasal Pressure Airflow Measurement. What is Measuring Nasal Pressure and How is this Different from Thermistors and Thermocouples Data coming from direct measurement of respiratory airflow (using masks and pneumotachographs) began to accumulate in the 1980 s. These data showed that transient elevations in resistance in the upper airway could be detected as a plateau or flattening of the inspiratory waveform. Initially, though, the only technique available to accomplish this was full facial pneumotachography, which was not suited to general clinical practice. Others and we began to use a more non-invasive alternative to measure airflow during diagnostic studies: this consisted of a nasal cannula (standard for oxygen administration) connected to a sensitive pressure transducer. In a number of papers, 2, 12 it was shown that the relative amplitude of breaths and morphology of the respiratory waveform obtained from this pressure signal reproduces that obtained from full facial pneumotach in most situations. Nasal cannula pressure systems generate this respiratory waveform signal by detecting the fluctuations in pressure caused by inspiration and expiration. Unlike the inspiratory and expiratory fluctuations recorded via temperature from a thermocouple or thermistor, these signals are truly proportional to flow. There is a theoretical need to apply a linearization to the pressure signal by taking its square root, but this does little to affect the shape of the signal in most cases. Consequently, nasal pressure (NP) airflow appears to be far more sensitive and accurate in the capture of hypopnea and Respiratory Effort Related Arousals (RERA) events. This increased sensitivity should clearly enhance the diagnostic yield obtainable in the Polysomnographic evaluation of sleep patients. As published in the AASM guidelines, the only validated, and thus recommended, technique for the identification of RERAs was the use of an esophageal balloon manometer to measure intrathoracic pressure in order to identify a pattern suggesting increasing respiratory efforts ending in arousal. This technique, however, has not gained wide acceptance because of patient comfort issues (perceived or otherwise). Consequently, in many clinical laboratories, the diagnosis of UARS is made by inference. It is assumed that subtle changes in respiratory pattern (building respiratory effort and/or cascading snoring) terminated by EEG arousal represent a respiratory event (RERA). An alternative used by some clinical laboratories is to demonstrate that the repetitive arousals (or movements) are reduced in a patient when nasal CPAP is applied. The availability of a simple direct measure of nasal airflow (the nasal cannula/pressure system) provides an alternative to the thermal based flow devices and affords a more direct method to detect RERAs. Recent data shows that the nasal cannula technique contains information about respiratory airflow not found in the thermistor, and can be used to detect the subtle flow/resistance changes characteristic of UARS (now called RERA) with results similar to those obtained with an esophageal manometer. The use of the nasal cannula technique is also in accord with the AASM recommendations that state that thermistors/thermocouples are NOT appropriate to the detection of 2, 6, 9, 12 hypopnea and RERAs.

7 In summary, recent published data have shown that the use of the nasal cannula pressure based technique for assessing airflow and upper airway resistance changes, unlike temperature based probes, reliably detect apnea, hypopnea, and the subtle flow 2, 3, 4, 5, 6, 8, 11, 12, 13 limitation observed in the Upper Airway Resistance Syndrome.

8 Chapter 3 How to Record the Nasal Pressure Signal In our experience, the most frequent problem that laboratories have with implementing a recording of NP airflow relates to amplifier choice and filter settings. Attention to these issues is critical in recording a hi-fidelity representation of respiratory airflow. This section deals with appropriate use of AC and DC amplifiers and filter settings. Filter Settings There are three types of filters provided by most recorders. These are: Low frequency filter (LFF) (often referred to somewhat confusingly as the high pass filter ) High frequency filter (HFF) and 60 cycle filter (often referred to as the notch filter). On some polygraphic recorders the low frequency filter may be referred to as the Time Constant. This concept relates specifically to AC amplifiers. Low frequency filters are used to prevent the wandering baseline which occurs in some signals. The behavior of these amplifiers can be seen dramatically during the typical calibration that is done with a DC voltage (usually 50µV). When this voltage (a constant signal) is passed through the amplifier, a calibration signal, such as is shown in Figure 3.1 is generated. While the peak of the rise provides a value for calibration i.e., the size of a 50µV wave, the AC amplifier characteristic is to force any constant value back to baseline (even though the input value from the calibration source does not change). The time required for the pen deflection to fall back to 37% of its full amplitude is known as the fall time constant. There is an inverse relationship between the time constant and the low filter setting. Increasing the low frequency filter will decrease the time constant and vice versa. In general, a low frequency filter will attenuate or suppress any signal which changes at a frequency below the assigned frequency. If the time constant is relatively short, say 1 or 2 seconds, a respiratory airflow signal will begin to decay prematurely during a sustained breath, thus significantly affecting the morphology of the waveform. In the case of the NP airflow waveform, a short time constant can cause the plateau associated with inspiratory flow reduction to decay prematurely and become undetectable. In fact, with a short enough time constant (such as that used in thermistor recordings) ALL respiratory waveforms are changed to a near perfect sine wave appearance.

9 The high frequency filter attenuates or suppresses all signals whose frequency is above the set value. This is extremely crucial when the user is attempting to isolate information generated from a specific frequency range, and gets rid of high frequency signals like muscle artifact in the EEG. This is also illustrated in the case of the snoring signal captured from the nasal airflow signal. When the high frequency filter is set to 70Hz, all information up to 70Hz is captured; this includes the snoring information. When the high frequency filter is set to 5Hz, only the slow information below 5 Hz is captured which would in this case be respiration. Notch or 60 cycle filters are used to attenuate or suppress a single narrow range of frequencies (typically 60 cycle noise ). If chosen to eliminate a frequency which is not in the range of the desired signals, they do little to affect the physiological signal. Amplification: AC or DC In order to properly record the NP airflow signal, it is necessary to pay close attention to the decisions regarding the amplification (or system input type) of the airflow signal. This is true whether the airflow signal will be recorded on a polygraph or with a digital recording system. The first decision is whether you wish to record the data via AC or DC inputs. The main difference between AC and DC amplification is whether constant-voltage input will result in a constant-voltage output (this happens when you chose DC), or whether constant, or near-constant, voltage signals will be filtered out of the signal (this happens when you chose AC, and is described in the section on filtering, above). In the following descriptions, polygraph refers to either the classic analog polygraph or a computerized data acquisition system. Use of AC Input for the Nasal Cannula Pressure Signal During flow limitation, the inspiratory airflow can be nearly constant for over 2 seconds. A nasal pressure airflow transducer will output a nearly constant voltage throughout inspiration under this condition. In order for these events to be properly visualized in an AC configuration, the polygraph or other recording device must NOT be set to suppress such nearly constant signals. Thus it must have the capability of establishing a low frequency cutoff filter setting of 0.05Hz or less (or a time constant approximately 3 seconds). The optimal setting would be 0.01Hz (time constant approximately 5 seconds). This setting will allow the respiratory waveforms to appear essentially identical to a DC signal. An example of the effect of different recordings of the identical flow signal obtained from a nasal cannula pressure transducer output is shown in Figure 3.2. The bottom trace shows a flow limitation event recorded using a DC amplifier. During the flow limitation event, only a slight decrease in flow occurs during each inspiration. The top tracing shows the same event recorded through an AC amplifier with a low frequency filter that is set too high. During each inspiration flow falls back to zero flow (center line) well before inspiration ends. The middle tracing shows the same signal recorded with an AC amplifier with an appropriate low frequency filter set and is virtually identical to the DC recorded signal.

10 The greatest disadvantage of using the AC recording settings is that many older systems, designed to record EEG type signals, do not have appropriate low frequency filters for respiration signals. Despite this, AC configurations have the advantage of ease of use and familiarity for the typically trained EEG or sleep technician. The transducer connects directly into the AC headbox, which is already in the patient room. Most commonly, a 1.5 mm safety connector is used in this configuration. This is not true of all recorders, though, and it is best to verify the correct input connector before purchasing a transducer. DC signals, however, must be connected to a special DC input on the polygraph, which may not be on the headbox. Figure 3.2: Effect of differing time constants. Use of DC Input for Nasal Cannula Pressure Signal The best way to record a good NP airflow signal is to use a DC amplifier or input port. This bypasses the need to worry about low frequency filter settings. The placement of the DC ports varies from recorder to recorder. Some systems place the DC ports in the tech room while others position them at the patient bedside. If they are located far from the patient in the tech room, it will be necessary to obtain an extension cable. The most common DC input connector is a 1/8 mini-phone jack but this is by no means universal. If you are unsure of the correct DC configuration for your Polygraph, contact the host Polygraph manufacturer. Sampling Rate When signals are recorded on a digital system, i.e., a computer, the actual values are not stored continuously, but rather are sampled at a set frequency. Sampling rate is often referred to as the digitizing rate or sampling frequency. This is the number of discrete values per seconds that are stored to represent the signal. For example, a sampling rate of 10Hz means that the value of the recorded signal amplitudes will be measured at 10 equally spaced samples in each second. These can be displayed as a

11 continuous tracing by having the computer connect the points on the display to make a waveform. A higher sampling rate therefore provides a more refined expression of the data. Unfortunately, this comes at the cost of increased file size (more points to store). The assignment of the correct sampling rate is determined by weighing the complexity of the waveform on any given channel against the amount of space required to record the channel. For example, EEG channels (which are inherently more complex than respiratory channels because they vary rapidly) require more definition to be correctly interpreted and are therefore recorded with higher sampling rates, e.g., or more samples per second. Respiratory channels are typically recorded with a lower sampling rate (10-50 samples per second) because further sampling of the data may provide no additional usable information. When a signal is sampled, distortion of the data can occur if the sampling rate is too low relative to the natural frequencies occurring in the signal (aliasing). For this reason, a good general rule of thumb, based on theoretical considerations, is to set the sampling rate to no less than twice the highest frequency contained in the signal, which is often determined by the value chosen for the high filter setting (Nyquist limit). The following table gives some reasonable settings. Table 3.1 shows some typical recommended filter and sample rate settings for the various types of outputs available from NP airflow transducers. Table 3.1 Pressure Transducer Output Signal Recommended Filter Settings and Sampling Rates Sensitivity Time Constant (Low Filter) High Filter Sampling Frequency Unfiltered Output (Low Level AC) 30µV/mm (Variable * ) 3 6 seconds ( Hz) 70Hz or higher Hz Unfiltered Output (High Level DC) Varies ** NA 70Hz or higher Hz Filtered Output (Low Level AC) 30µV/mm (Variable*) 3 6 seconds ( Hz) 5Hz or higher 10Hz or higher Filtered Output (High Level DC) Varies ** NA 5Hz or higher 10Hz or higher Snoring Output (Low Level AC) 30µV/mm (Varies * ) 10Hz or lower (>0.016 seconds) 70Hz or higher Hz Snoring Output (High Level DC) Varies ** NA 70Hz or higher Hz DC Signals are referred to as high level signals because the signals are transduced at a higher output voltage than AC signals. A typical DC voltage might be 2V peak to peak while a typical AC signal ( low level ) might be 600µV peak to peak. The specific sensitivity settings may also vary for some transducer systems. * Varies from patient to patient. **Varies from patient to patient and for a specific polygraph.

12 Chapter 4 Extraction of the Snoring Signal from the Pressure Waveform Once properly collected, the raw (unfiltered) pressure waveform from a nasal cannula contains more than just the respiratory airflow. Breathing is a signal represented by the slow changes in pressure. In addition, fast vibrations made during snoring show up on the pressure signal, (Figure 4.1). As a result, by altering the filtering settings, the waveform may be conditioned to display respiration only (Figure 4.2) or snoring only (Figure 4.3). Figure 4.1: The unfiltered pressure waveform. In the above example, you will note both the flattening of the waveform (characteristic of flow reduction) with fast activity (snoring) superimposed on the third and fourth breaths. By adjusting the high frequency filter to 5-15Hz, the snoring is removed, leaving a respiratory waveform only. Figure 4.2: The pressure waveform filtered for respiration only. Conversely, if the raw waveform is conditioned by setting a low frequency filter to 10Hz and a high frequency filter to 70Hz or greater, only the snoring is captured. The reliability of the snoring information captured in this fashion has not been rigorously compared to that obtained with a microphone or other snoring sensor. However, it is our impression that this signal is as reliable as these other sensors and has the advantage of a lower background noise. In addition, the nasal cannula is not subject to the application issues associated with those devices (loosened tape, the influence of external physical characteristics of the patient, etc.). Figure 4.3: The pressure waveform filtered for snoring only.

13 Chapter 5 Common Problems Poor Nasal Pressure Flow Signal Patient Issues. In the example below (Figure 5.1), a loss of amplitude in the NP airflow signal is apparent. The loss of either inspiratory or expiratory flow can occur with poor positioning of the nasal prongs on the cannula (they must be inside the nares to function properly). It is possible to trim the prongs somewhat to improve the comfort of the patient but care must be taken to insure that a sufficient amount of tubing remains to fully capture the signal. A small piece of tape applied to either side of the cannula may insure that the prongs remain well positioned throughout the night. Signal loss may also result from a complete blockage of the cannula tubing due to secretions, but this is rare if the patient does not have clinically evident rhinorhea. Furthermore, the signal amplitude varies from patient to patient. This is inversely related to the size of the patient s nares. In other words, a large nostril may promote a low amplitude signal and a small nostril may promote a larger signal. Oral breathing may occur but a complete cessation of nasal breathing is rare during sleep. Recent data suggests that this occurs in only 1 2% of the overall recording time. 11 Partial oral breathing is more common, but when this occurs cyclical changes in the shape and amplitude of NP flow signal are still indicative of respiratory events albeit that a sensitivity adjustment may be required. Many pressure users utilize an oral thermistor or a nasal-oral cannula to address the issue of oral breathing. Figure 5.1: Poorly positioned cannula. During wakefulness, the respiratory signal may be chaotic and should not be overinterpreted. With practice this can be recognized by the presence of large breaths with irregular frequencies. The tracing below, (Figure 5.2), shows a 2-minute transition from wake to sleep (second half). Note that the irregular amplitudes and frequencies stabilize with the onset of sleep.

14 Figure 5.2: Sleep onset. Poor Nasal Pressure Flow Signal Technical Factors. The most common problem is an incorrect filter setting when the AC output of the PTAF is used. The AC time constant of the recording device must be longer than 3-5 sec (low frequency filter of Hz or lower). When the time constant is too short, a square signal (such as the calibration wave) will decay and may cause flow-limited inspirations to appear biphasic, (Figure 5.3). Strictly speaking, the NP airflow signal should be collected as a DC signal when this low frequency filter / time constant requirement cannot be met by the host recorder. Figure 5.3: Signal decay from short time constant. Another commonly made mistake is in establishing the correct polarity of the channel. To determine if the polarity is correct, check to insure that the negative (upward) deflections correspond to inhalation and that the downward deflections correspond to exhalation. In general, when the polarity is correct, amplitude and morphological changes are reflected on the top of the waveforms.

15 That is to say, flattening or rounding of the waveform should be apparent on the tops of the waveforms while the lower part of the waveform should be more characteristically sharp, as shown below in Figure 5.4 and Figure 5.5. Figure 5.4: Correctly oriented NP channel. Figure 5.5: Inverted NP signal.

16 Chapter 6 Nasal Pressure Airflow and NCPAP The same analysis of airflow characteristics that is used during diagnostic studies performed with a NP airflow signal (using a cannula) can be applied during the titration of nasal CPAP. Using this information about flow waveforms may improve CPAP titration by identifying sub-optimal pressures through the presence of flow limitation. While many laboratory CPAP units possess the ability to export an airflow signal themselves (or an estimated patient flow signal in which any constant leak flow is subtracted out of the electronic signal), most of the simple systems used in the home setting do not have this feature. The same hardware that is used to record NP airflow can also be used to monitor airflow during nasal CPAP. This also allows some of the special features of the processed signal to be utilized, e.g., snoring detection. There are several ways to hook up such a system with the CPAP circuit. The specific techniques that can be used depend upon the details of the transducer. In-Line Pneumotach - The best technique for measuring a true flow signal is to use an in-line pneumotach. This device provides a small in-line resistance in the CPAP tubing. The pressure drop across this resistance is proportional to flow, but pressure must be detected at two sites. See Figure 6.1 for additional detail. Dual Hose Connection - A variation on this takes advantage of the natural resistance of the CPAP circuit. Thus, the hose itself becomes a type of pneumotachograph and pressure drop from start to finish of the hose is proportional to flow. Again, pressure must be detected at two sites, but no separate pneumotachograph is needed. See Figure 6.2 for additional detail. The Single Hose Mask Connection - Pressure can be measured in the CPAP mask directly and assume this is proportional to flow. With this approach, only one pressure needs to be measured. See Figure 6.3 for additional detail. Cannula Under the Mask - The last option is to use an unmodified nasal cannula under the CPAP mask. While apparently the simplest solution, this has several disadvantages, which are discussed under Figure 6.4, and we do not generally recommend this approach. See Figure 6.4 for additional detail.

17 In-Line Pneumotach Figure 6.1 The highest fidelity airflow signal is obtained with the use of an in-line pneumotach, the gold standard device for measurement of airflow. This technique requires a measurement of the difference between two pressures (above and below the pneumotachograph), thus a dual port transducer. In this configuration of flow monitoring, a pneumotach is placed in the CPAP circuit just below the mask or at the CPAP generator output. The signal that is produced by the drop in pressure in this device is relatively easy to set up and can be used with any CPAP generator. The signal from a pneumotach can be calibrated to actual airflow if one so desires. It thus provides not only respiratory waveforms, but also information about leak flow from the CPAP system. However, this last advantage can be one of the problems with the use of a pneumotach at this location. Since any flow from the CPAP generator will be detected, the airleak between the pneumotach and the patient will also be included in the measured airflow and will create an offset. One solution to this is to use a transducer that has an AC coupled output (or an AC amplifier with a long time constant), which will remove this baseline offset.

18 Dual Hose Connection Figure 6.2 Simply using the CPAP hose in place of the pneumotach can produce an airflow signal of high quality, comparable to that from a pneumotach. This circuit involves sampling the pressure at each end of the CPAP tube. Like the in-line pneumotach, it requires the use of a dual port pressure transducer, as the difference between two pressures is the signal sought. This configuration provides an excellent signal, but does require some additional materials and setup time. The materials required would be two in line pressure taps (O 2 Tees work well) and an additional one-foot length of CPAP circuit. To correctly set up this circuit, place the Tee adapters on each end of the 6-8 foot long CPAP tubing. The one-foot extension of tubing (or adapters) should connect the CPAP unit with one of the Tee adapters. Although the Tee could be attached directly to the CPAP generator, the use of a short extra tube helps to minimize noise transmitted from the generator. The other Tee adapter is attached near the mask. The nipple of the Tee closest to the CPAP should be connected to the negative ( - ) port of the transducer with small diameter e.g., O 2 tubing. The Tee closest to the mask should be connected to the positive (+) port. With this circuit, a generally excellent flow signal can be obtained, although no actual pneumotachograph is used.

19 The Single Hose Mask Connection Figure 6.3 Pressure fluctuations in the CPAP mask itself are created by the patient s inspiratory and expiratory airflow. These pressure fluctuations are superimposed on the CPAP pressure signal and are roughly proportional to this patient s airflow, provided the pressure at the CPAP generator does not vary much (as is the goal of CPAP). In this circuit configuration, the pressure signal must be processed so only the fluctuations are used to measure airflow. The advantage of the single hose to mask connection is its simplicity. This technique requires no additional cabling by the lab staff and is clearly one of the few possibilities for those who possess a pressure transducer with only one port for tubing. Unfortunately, though, this circuit requires a transducer that measures pressure in the range typical of that of CPAP (as opposed to the other circuits which directly measure the differential pressure between two sites). Pressure fluctuations during breathing (for which the transducers used with the nasal cannula are generally optimized) are in the range of 2cm H 2 O. While on CPAP, these fluctuations occur on a background of the set CPAP pressure, which may be up to 20cm H 2 O, and a transducer optimized for the nasal cannula may be incapable of measuring these higher pressures. In addition, flow turbulence in the mask may affect the quality of the signal and signal dampening may occur as CPAP pressures are raised. Turbulence may obscure any derived snore signal by creating an excessively thick baseline. Finally, the assumption that pressure fluctuations in the mask are proportional solely to patient breathing may be incorrect. Specifically, some CPAP generators are deliberately designed to minimize these pressure swings (and may even have a sensor in the mask for pressure feedback). These devices intentionally alter output flow from the CPAP generator in response to mask pressure fluctuations, and may render the single hose signal unusable.

20 Cannula Under the Mask Figure 6.4 This technique is typically employed with a single port transducer. It is easily configured as it is identical to the use of the nasal cannula alone. As with the single hose mask connection, the transducer must be capable of measuring pressures in the CPAP pressure range. The signal obtained may be somewhat more stable than that obtained by simply connecting the pressure transducer directly to the CPAP mask. However, the snoring signal derived may still be contaminated with some background CPAP noise, but less than with the direct mask connection. There may be some signal dampening as well due to the CPAP blower feedback control of pressure, but again less than with a simple mask connection. The major objection to this approach is patient comfort and potential mask leaks. Some patients may find that the cannula and the mask combine to irritate the skin where the cannula passes under the mask.

21 Chapter 7 Scoring Atlas Figure 7.1: Apnea Apnea is detected as a flat signal on the NP airflow. There is a tendency for small breaths (severe hypopnea) to appear as apnea on the NP airflow signal. Note the minimal fluctuation present on the thermocouple signal despite flat NP airflow signal.

22 Figure 7.2: Hypopnea Hypopnea is detected as a clearly reduced amplitude of the NP airflow signal. This may be more a more obvious reduction than that seen on the thermocouple signal. Normal inspiration has a sinusoidal appearance on the NP airflow signal. During obstructive hypopnea there is frequently progressive flattening of the inspiratory signal. Snoring may appear simultaneously.

23 Figure 7.3: Flow limitation Elevated upper airway resistance events are detected when the NP airflow does not show a reduction in amplitude to the level of hypopnea but clearly demonstrates flattening (which indicates inspiratory flow limitation). Note that there may be no indication of abnormality on the thermocouple channel.

24 Figure 7.4: Better event definition Events on the NP airflow signal are often easier to identify than on other respiratory channels and the reduction in amplitude during the event may be more severe.

25 Figure 7.5: Severe flow limitation and apnea Some apnea events begin as episodes of severe flow reduction and transition to apnea. These events may be scored as hypopnea if only a thermistor is used.

26 Figure 7.6: Sustained flow limitation The flattened inspiratory airflow contour on the NP airflow signal may occur as a transient event (hypopneas or RERAs) or in sustained runs. These are particularly common during delta sleep and may indicated sustained elevated upper airway resistance without arousal. Sustained flattened inspiratory airflow contour may overlap with benign snoring.

27 Figure 7.7: Flow limitation without hypopnea Patients may experience periods of flow limitation that do not generate neuro-cortical arousal. This is particularly common during slow-wave sleep when the arousal threshold is depressed. In the absence of arousal, these periods probably do not represent clinical disturbance.

28 Figure 7.8: Flow limitation with snoring Although it is possible that a patient may experience non-disruptive snoring (so called primary snoring) most patients that snore do exhibit some degree of sleep fragmentation. The cause of this fragmentation may appear ambiguous when a thermocouple is used for airflow but is often clearly related to respiration when NP airflow is used.

29 Figure 7.9: Flow limitation without snoring One of the more insidious forms of mild sleep disordered breathing is disruptive flow limitation seen in the absence of snoring. This condition is not easily identified by significant changes in other channels and, as a result, may be under-represented in the lab environment. The absence of snoring may lead a trained technologist to misinterpret a respiratory event as a leg jerk or transient arousal when a thermocouple is used. These events are usually clear when NP airflow is employed.

30 Figure 7.10: Snoring without flow limitation It is possible for arousals to be associated with cascading snoring and not a characteristically flattened nasal pressure signal. These events occur in the absence of flow limitation, although there is cascading snoring. Such events may be scored as RERA events.

31 Figure 7.11: Variant flow signal during snoring Fluctuations in the NP airflow signal are occasionally seen while the patient is snoring. It is important to remember that snoring, by definition, represents some degree of airway compromise but in the absence of neuro-cortical arousal should not be scored.

32 Figure 7.12: Cannula and esophageal pressure The AASM definition of a RERA is based on increasing esophageal pressure prior to arousal. The change in shape on the NP airflow signal is usually present and allows RERAs to be detected non-invasively from the NP airflow signal alone.

33 Chapter 8 Frequently Asked Questions 2, 12 QUESTION 1: Should I use the AC or DC inputs? The decision of whether to use the DC or AC inputs is based on a number of factors such as the number of available DC ports or pin inputs on the polygraph headbox, the filtering capability or configuration of your polygraph and the lab design. In general, the DC (high level) ports are preferred as the nasal cannula signal is a slowly varying respiratory signal and AC amplifiers are more suited to rapidly varying signals e.g., EEG. If an AC amplifier is used, a low filter setting of (preferably 0.01Hz) or lower (time constant of 3 seconds or longer) is required in order to correctly record the unfiltered and filtered flow signals. Either DC or AC inputs may be used as long as this filtering requirement can be met. QUESTION 2: What polygraph settings should I use? Pressure Transducer Output Signal Unfiltered Output (Low Level AC) Unfiltered Output (High Level DC) Filtered Output (Low Level AC) Filtered Output (High Level DC) Snoring Output (Low Level AC) Snoring Output (High Level DC) Sensitivity 30µV/mm (Variable*) Time Constant (Low Filter) 3 6 seconds ( Hz) High Filter 70Hz or higher Sampling Frequency Hz Varies** NA 70Hz or higher Hz 30µV/mm (Variable*) 3 6 seconds ( Hz) 5Hz or higher 10Hz or higher Varies** NA 5Hz or higher 10Hz or higher 30µV/mm (Varies*) 10Hz or lower (>0.016 seconds) 70Hz or higher Hz Varies** NA 70Hz or higher Hz *Varies from patient to patient. **Varies from patient to patient and for a specific polygraph.

34 QUESTION 3: What is a sampling rate? (also referred to as the sampling frequency or digitizing rate) Computers record data as a series of points. The number of points per second to be recorded is the sampling rate. In other words, if a sampling rate of 10Hz is established, 10 equally spaced samples of the data are recorded as points, which are then connected by the computer to make a waveform. A higher sampling rate provides a more refined expression of the data, but increases the size of the file to be recorded. The assignment of the correct sampling rate is determined by weighing the complexity of the waveform on any given channel against the amount of space required to record the channel. For example, EEG channels are inherently more complex than respiratory channels and require more definition to be correctly interpreted and are therefore recorded with higher sampling rates. Respiratory channels are typically recorded with a lower sampling rate because further refining the data and increasing the file size provide no additional clinical yield. A general rule of thumb is to set the sampling rate to no less than twice the high frequency filter setting. QUESTION 4: What is the difference between a low frequency filter and a time constant? AC amplifiers are typically calibrated to a DC voltage (usually 50µV). When this voltage is passed through an AC amplifier a calibration signal, such as is shown to the left, is generated. The time required for the deflection to fall back to 37% of its full amplitude is known as the fall time constant. The low frequency filter causes this fall back to baseline. There is an inverse relationship between the time constant and the low filter setting. Increasing the low frequency filter will decrease the time constant and vice versa. In general, a low frequency filter attenuates the signals below the assigned frequency. QUESTION 5: What are the advantages of the nasal cannula/pressure 2, 12 transducer over a thermocouple or thermistor? Temperature sensing devices (thermocouples and thermistors) have traditionally been used to measure airflow in Polysomnography. These devices, which sense temperature as a surrogate of airflow, are extraordinarily accurate when measuring frank apnea but are quite inaccurate when measuring moderate events such as hypopnea or RERA events. A measure of nasal airflow using a pressure transducer yields a much more accurate representation of the inspiratory flow reduction seen in these events. The nasal cannula pressure system, nearly as simple, uses a simple attachment to the patient (the cannula) and is comparable in cost.

35 QUESTION 6: What about oral breathing? 11 It has been estimated that clearly identifiable oral breathing only occurs for 1 2% of actual diagnostic recording time (Montserrrat abstract). Complete oral breathing may occur in patients with nasal pathologies, but is readily identifiable. In most cases of oral breathing, sufficient nasal breathing usually remains to allow for a NP reading. If oral breathing is a concern, this issue may be addressed by using a separate measure of the oral airflow, such as an oral thermocouple in combination with a nasal cannula. QUESTION 7: What does normal breathing look like? When measured by a nasal cannula, normal breathing appears as a series of sinusoidal inspiratory waveforms of equal deflection as depicted below. Figure 8.1 QUESTION 8: What is flow limitation and what does it look like? 2 During inspiration, the collapsibility of the upper airway causes a Starling resistor-like behavior and results in a maximum flow, regardless of increases in effort. The normal sinusoidal inspiratory airflow becomes flattened (develops a plateau). The incomplete airway obstruction also causes inspiratory time to increase. Overall, this appears as a flattening of the waveform and is typically seen in breaths of decreased amplitude (Figure 8.2). These events are usually associated with increased respiratory effort and may represent either hypopneas or RERAs. To correctly see this flattening, a time constant setting of 3 6 seconds (low frequency filter Hz) should be established. Figure 8.2

36 QUESTION 9: What does an apnea look like? By definition, an apnea refers to a period of complete airflow reduction. In actual practice, apneas may be defined as periods of airflow reduction below a certain percentage of a normal or baseline breath. During apnea, the NP airflow signal tends to be either perfectly flat or have a lower amplitude than the thermistor. Definitions range from below 10% to 25% and are often subjectively determined. In clinical practice, most sleep centers would score events associated with oxyhemoglobin desaturations (usually 3-4%). A period of repetitive apneas is shown in Figure 8.3. Figure 8.3

37 QUESTION 10: What does a hypopnea look like? Hypopnea is defined as airflow decreased below a given percentage of the surrounding baseline and may also require the presence of some degree of oxyhemoglobin desaturation or arousal. Typically, a flow reduction of 25 50% should be seen in order to score an event as a hypopnea. Scoring for both apnea and hypopnea is often subjectively determined, but the event duration should typically be at least 10 seconds. In the scoring of hypopnea one might wish to keep in mind the nature of the relationship between the NP airflow signal and true flow. The nasal cannula actually records the square of airflow. Thus, a NP airflow signal that is 50% of the surrounding baseline is actually 70% of the surrounding flow. A NP airflow signal that is 25% of surrounding flow is actually 50% of the actual airflow. A typical hypopnea is shown in Figure 8.4. Note, in addition to an amplitude reduction there is a marked change in shape with prolongation of inspiratory time. Figure 8.4

38 QUESTION 11: Can nasal pressure airflow be used to measure increased upper airway resistance (UARS)? Upper airway events (also known as RERA events) are arousals preceded by subclinical respiratory events (events that do not meet the criterion for hypopnea or are not detected with a thermistor at all). Nasal airflow has been shown to be far more reliable than thermal elements in detecting the subtle flow limitation characteristic of these events (Rapoport et al). Figure 8.5 shows a typical event characterized by flow limitation with some reduction in amplitude that coincides with the increased respiratory effort and is followed by arousal. Figure 8.5

39 QUESTION 12: How is snoring identified with a nasal cannula? Snoring causes small, fast fluctuations in the NP signal. Through the use of a high pass filter (either in the transducer or in the recorder), these fluctuations can be captured and recorded. To do this the signal is filtered through an AC amplifier with a low frequency cut off filter set at about 10Hz and a high pass filter set at 70Hz or above. The morphology of the NP waveform is very similar to that recorded using snoring microphones and vibraphones. An example of this is shown in Figure 8.6. Figure 8.6 QUESTION 13: What is the difference between a filtered and unfiltered airflow? With an unfiltered flow output, snoring information is displayed along with the flow signal on a single channel. A filtered flow signal uses a low pass filter (such as 5 to 10Hz) to remove the high frequencies such as that of snoring; what remains is the respiratory waveform. This is shown in Figure 8.6 above. It is important not to over filter, e.g., 1Hz low pass filter, the signal as this can remove components present in the signal during flow limitation.

40 QUESTION 14: If I am already using a snore microphone or snore sensor, what benefits are derived from the unfiltered flow signal? An unfiltered flow signal allows snoring and flow to be recorded simultaneously and may be advantageous if you have a limited number of available channels. However, the signal is generally noisier and may be more difficult to interpret than if snoring and airflow are presented on separate channels. QUESTION 15: What if I do not want to use the snore channel? If your transducer has multiple outputs (such as filtered flow, unfiltered flow and snoring) you may use any or all of these based on your personal preference. It is not required that all outputs be recorded. If you have a limited number of available channels, an unfiltered flow signal will combine both snoring with airflow into a single signal. QUESTION 16: Can a nasal pressure airflow system be used with a capnograph? Since both etco 2 and NP are collected via nasal cannula, it would be necessary to use a dual lumen cannula. These are commercially available and used in oxygen saver devices. This also provides a method for measuring nasal airflow while simultaneously administering oxygen. QUESTION 17: Are there validation studies for nasal pressure technology? Numerous papers have been written validating the use of NP airflow for the measurement of airflow. Some useful references are noted in the bibliography. QUESTION 18: Can the nasal pressure airflow signal be calibrated? NP airflow over long periods of time is only a semi-quantitative measure and cannot be readily calibrated to tidal volume. However, as noted by Guyatt, 7 calibration is possible for use over short periods though this would not be generally useful for sleep laboratories. In general, the definitions used in identifying respiratory events during sleep involve changes in amplitude relative to the surrounding region rather than absolute cutoffs that would require calibration. QUESTION 19: Can a nasal pressure transducer be used to measure esophageal pressure? In general, the pressure range of most nasal cannula pressure transducers is in the +/- 2 ( or +/- 20) cm H 2 O range while most esophageal pressure transducers can measure up to 50 cm H 2 O. Thus, a NP channel is not likely to be able to measure esophageal pressures. QUESTION 20: Can I use any cannula? Any cannula may be used though many manufacturers recommend cannulas designed specifically for use with a NP transducer. Care must be taken to insure that the cannula used does not occupy too large a portion of the cross section of the nares (over 50%) so that it acts as an obstruction to breathing in. Special cannulas, such as the dual lumen cannula mentioned above, are also available.

41 QUESTION 21: Can children be studied using nasal cannula airflow? Pediatric cannulas are available in most respiratory therapy departments. Again, care should be taken with small children to insure that the cannula does not, itself, become an obstruction. QUESTION 22: Do the nasal prongs ever become clogged? It is possible for the nasal prongs to become clogged with secretions although in actual practice the occurrence is rare. In the laboratory this can be readily resolved by cleaning or changing the cannula. A signal can still be processed if one prong is clogged. QUESTION 23: Can the nasal prongs be trimmed if needed? Yes, but it is important to retain some of the prong for a good signal. To function properly the tip of the prongs must be inside the nose (to measure the pressure in the nares, not the pressure outside). QUESTION 24: Do I need to tape the cannula on the patient? Most cannulas are designed for simple but secure attachment to the patient. Whether using a strap around the head or tubes secured around the ears, the nasal cannula does not usually require any additional method to secure it to the patient. QUESTION 25: What methods should be used to connect the nasal pressure transducer to a CPAP unit? There are four possible methods recommended: 1. In-Line Pneumotach 2. Mask Connection 3. CPAP Mask over Cannula 4. Dual Hose Connection The specific methods, advantages and disadvantages of each are discussed in the main section of the primer. (Pages 15-19) QUESTION 26: Why do I get a distorted signal when connecting a tubing between the CPAP mask and a pressure transducer? Turbulence inside the mask and automatic flow adjustments by the CPAP generator (to maintain mask pressure) may distort the waveform. In addition, with some transducers the pressure offset produced by the CPAP may saturate the pressure transducer. See the main portion of the primer for suggestions as to how to overcome this. (Pages 15-19) QUESTION 27: Why do I get a heavy baseline on the snore channel when connected to a CPAP machine? Vibrations caused from the operation of the CPAP device may be detected and displayed as baseline noise.

Summary of AASM Clinical Guidelines for the Manual Titration of Positive Airway Pressure in Patients with Obstructive Sleep Apnea - Updated July 2012

Summary of AASM Clinical Guidelines for the Manual Titration of Positive Airway Pressure in Patients with Obstructive Sleep Apnea - Updated July 2012 Summary of AASM Clinical Guidelines for the Manual Titration of Positive Airway Pressure in Patients with Obstructive Sleep Apnea - Updated July 2012 SUMMARY: Sleep technologists are team members who work

More information

MODULE. POSITIVE AIRWAY PRESSURE (PAP) Titrations

MODULE. POSITIVE AIRWAY PRESSURE (PAP) Titrations MODULE POSITIVE AIRWAY PRESSURE (PAP) Titrations POSITIVE AIRWAY PRESSURE (PAP) TITRATIONS OBJECTIVES At the end of this module the student must be able to: Identify the standards of practice for administering

More information

HEALTH EVIDENCE REVIEW COMMISSION (HERC) COVERAGE GUIDANCE: DIAGNOSIS OF SLEEP APNEA IN ADULTS DATE: 5/9/2013 HERC COVERAGE GUIDANCE

HEALTH EVIDENCE REVIEW COMMISSION (HERC) COVERAGE GUIDANCE: DIAGNOSIS OF SLEEP APNEA IN ADULTS DATE: 5/9/2013 HERC COVERAGE GUIDANCE HEALTH EVIDENCE REVIEW COMMISSION (HERC) COVERAGE GUIDANCE: DIAGNOSIS OF SLEEP APNEA IN ADULTS DATE: 5/9/2013 HERC COVERAGE GUIDANCE The following diagnostic tests for Obstructive Sleep Apnea (OSA) should

More information

Lab 1: The Digital Oscilloscope

Lab 1: The Digital Oscilloscope PHYSICS 220 Physical Electronics Lab 1: The Digital Oscilloscope Object: To become familiar with the oscilloscope, a ubiquitous instrument for observing and measuring electronic signals. Apparatus: Tektronix

More information

ROLE OF ORAL APPLIANCES TO TREAT OBSTRUCTIVE SLEEP APNEA

ROLE OF ORAL APPLIANCES TO TREAT OBSTRUCTIVE SLEEP APNEA 1 ROLE OF ORAL APPLIANCES TO TREAT OBSTRUCTIVE SLEEP APNEA There are three documented ways to treat obstructive sleep apnea: 1. CPAP device 2. Oral Appliances 3. Surgical correction of nasal and oral obstructions

More information

SUMMARY. Additional Digital/Software filters are included in Chart and filter the data after it has been sampled and recorded by the PowerLab.

SUMMARY. Additional Digital/Software filters are included in Chart and filter the data after it has been sampled and recorded by the PowerLab. This technique note was compiled by ADInstruments Pty Ltd. It includes figures and tables from S.S. Young (2001): Computerized data acquisition and analysis for the life sciences. For further information

More information

Comparing the Performance and Efficacy of the 3B/BMC RESmart. Auto-CPAP with the ResMed S9 AutoSet

Comparing the Performance and Efficacy of the 3B/BMC RESmart. Auto-CPAP with the ResMed S9 AutoSet 3B MEDICAL, INC., 21301 US HIGHWAY 27, LAKE WALES, FL. 33859 Comparing the Performance and Efficacy of the 3B/BMC RESmart Auto-CPAP with the ResMed S9 AutoSet Zhi Zhuang, PhD, Research and Development,

More information

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

Central, mixed and obstructive sleep apnea patient

Central, mixed and obstructive sleep apnea patient Central, mixed and obstructive sleep apnea patient Clinical scenario: A 57-year-old male with a history of TMJ, bruxism and arthritis is not on any daily medications but takes a daily multivitamin and

More information

Diagnosis and Treatment

Diagnosis and Treatment Sleep Apnea: Diagnosis and Treatment Sleep Apnea Sleep Apnea is Common Dangerous Easily recognized Treatable Types of Sleep Disordered Breathing Apnea Cessation of airflow > 10 seconds Hypopnea Decreased

More information

DIAGNOSING SLEEP APNEA. Christie Goldsby Florida State University PHY 3109 04/09/14

DIAGNOSING SLEEP APNEA. Christie Goldsby Florida State University PHY 3109 04/09/14 DIAGNOSING SLEEP APNEA Christie Goldsby Florida State University PHY 3109 04/09/14 Outline of Talk Background information -what is sleep apnea? Diagnosing sleep apnea -polysomnography -respiratory airflow

More information

Healthcare Online. Keyhole Touchproof Phone RCA RJ10 RJ11 (dual 1mm) (1.5mm safety) (male) (male 1/8")

Healthcare Online. Keyhole Touchproof Phone RCA RJ10 RJ11 (dual 1mm) (1.5mm safety) (male) (male 1/8) ACCESSORI COMPATIBILI PER APPARECCHIATURE COMPUMEDICS Model: 0510 Adult OroNasal Thermistor with 1.5 mm safety pins. 7-month warranty. Model: 0510PED Pediatric (ages 1-10 yrs) OroNasal Thermistor with

More information

Operational Amplifier - IC 741

Operational Amplifier - IC 741 Operational Amplifier - IC 741 Tabish December 2005 Aim: To study the working of an 741 operational amplifier by conducting the following experiments: (a) Input bias current measurement (b) Input offset

More information

11: AUDIO AMPLIFIER I. INTRODUCTION

11: AUDIO AMPLIFIER I. INTRODUCTION 11: AUDIO AMPLIFIER I. INTRODUCTION The properties of an amplifying circuit using an op-amp depend primarily on the characteristics of the feedback network rather than on those of the op-amp itself. A

More information

Arlington Dental Associates Ira Stier DDS PC 876 Dutchess Tpk 2 Lafayette Ct. Poughkeepsie, NY 12603 Fishkill, NY 12524 845-454-7023 845-896-4977

Arlington Dental Associates Ira Stier DDS PC 876 Dutchess Tpk 2 Lafayette Ct. Poughkeepsie, NY 12603 Fishkill, NY 12524 845-454-7023 845-896-4977 Home Sleep Test Liability Form Study Equipment Due: @ I, accept responsibility for the sleep monitoring device while it is in rny possession. I understand that if I fail to return the device or I return

More information

Laboratory 4: Feedback and Compensation

Laboratory 4: Feedback and Compensation Laboratory 4: Feedback and Compensation To be performed during Week 9 (Oct. 20-24) and Week 10 (Oct. 27-31) Due Week 11 (Nov. 3-7) 1 Pre-Lab This Pre-Lab should be completed before attending your regular

More information

Fits in the palm of your hand. Small and mighty. Stardust II portable diagnostic system

Fits in the palm of your hand. Small and mighty. Stardust II portable diagnostic system Small and mighty Powerful enough to generate comprehensive reports and small enough to fit the palm of your hand, the Stardust II portable diagnostic system provides you with an easy-to-use portable testing

More information

Frequency Response of Filters

Frequency Response of Filters School of Engineering Department of Electrical and Computer Engineering 332:224 Principles of Electrical Engineering II Laboratory Experiment 2 Frequency Response of Filters 1 Introduction Objectives To

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

Integration of the new miniaturized pneumotachograph for SOMNOscreen plus (Item no.: SEN513)

Integration of the new miniaturized pneumotachograph for SOMNOscreen plus (Item no.: SEN513) Software-News DOMINO 2.7.0 July 2015 The latest DOMINO version 2.7.0 comes with a handful of new features, allowing a high quality, user-friendly collection and analysis of the ambulatory PG and PSG recordings

More information

SINGLE-SUPPLY OPERATION OF OPERATIONAL AMPLIFIERS

SINGLE-SUPPLY OPERATION OF OPERATIONAL AMPLIFIERS SINGLE-SUPPLY OPERATION OF OPERATIONAL AMPLIFIERS One of the most common applications questions on operational amplifiers concerns operation from a single supply voltage. Can the model OPAxyz be operated

More information

What is Obstructive Sleep Apnoea?

What is Obstructive Sleep Apnoea? Patient Information Leaflet: Obstructive Sleep Apnoea Greenlane Respiratory Services, Auckland City Hospital & Greenlane Clinical Centre Auckland District Health Board What is Obstructive Sleep Apnoea?

More information

The Calculation of G rms

The Calculation of G rms The Calculation of G rms QualMark Corp. Neill Doertenbach The metric of G rms is typically used to specify and compare the energy in repetitive shock vibration systems. However, the method of arriving

More information

Q1. The graph below shows how a sinusoidal alternating voltage varies with time when connected across a resistor, R.

Q1. The graph below shows how a sinusoidal alternating voltage varies with time when connected across a resistor, R. Q1. The graph below shows how a sinusoidal alternating voltage varies with time when connected across a resistor, R. (a) (i) State the peak-to-peak voltage. peak-to-peak voltage...v (1) (ii) State the

More information

How To Calculate The Power Gain Of An Opamp

How To Calculate The Power Gain Of An Opamp A. M. Niknejad University of California, Berkeley EE 100 / 42 Lecture 8 p. 1/23 EE 42/100 Lecture 8: Op-Amps ELECTRONICS Rev C 2/8/2012 (9:54 AM) Prof. Ali M. Niknejad University of California, Berkeley

More information

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. CHAPTER 3: OSCILLOSCOPE AND SIGNAL GENERATOR 3.1 Introduction to oscilloscope 1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. 2. The graph show signal change

More information

Polysomnography in Patients with Obstructive Sleep Apnea. OHTAC Recommendation. Polysomnography in Patients with Obstructive Sleep Apnea

Polysomnography in Patients with Obstructive Sleep Apnea. OHTAC Recommendation. Polysomnography in Patients with Obstructive Sleep Apnea OHTAC Recommendation Polysomnography in Patients with Obstructive Sleep Apnea June 16, 2006 1 The Ontario Health Technology Advisory Committee (OHTAC) met on June 16, 2006 and reviewed a health technology

More information

Understanding Sleep Apnea

Understanding Sleep Apnea Understanding Sleep Apnea www.sleepmangementsolutions.com What is Obstructive Sleep Apnea (OSA)? OSA afflicts 20 million adult men and women in the U.S. People who have OSA stop breathing repeatedly during

More information

Sleep Apnea Dr. Douglas Tapper

Sleep Apnea Dr. Douglas Tapper Sleep Apnea Dr. Douglas Tapper Dr. Hendler: Hello and welcome to KP Healthcast. I m your host today, Dr. Peter Hendler, and our guest is Dr. Douglas Tapper who is the chief of Pulmonary Medicine at Kaiser

More information

LM 358 Op Amp. If you have small signals and need a more useful reading we could amplify it using the op amp, this is commonly used in sensors.

LM 358 Op Amp. If you have small signals and need a more useful reading we could amplify it using the op amp, this is commonly used in sensors. LM 358 Op Amp S k i l l L e v e l : I n t e r m e d i a t e OVERVIEW The LM 358 is a duel single supply operational amplifier. As it is a single supply it eliminates the need for a duel power supply, thus

More information

MECHINICAL VENTILATION S. Kache, MD

MECHINICAL VENTILATION S. Kache, MD MECHINICAL VENTILATION S. Kache, MD Spontaneous respiration vs. Mechanical ventilation Natural spontaneous ventilation occurs when the respiratory muscles, diaphragm and intercostal muscles pull on the

More information

Precision Diode Rectifiers

Precision Diode Rectifiers by Kenneth A. Kuhn March 21, 2013 Precision half-wave rectifiers An operational amplifier can be used to linearize a non-linear function such as the transfer function of a semiconductor diode. The classic

More information

Hands On ECG. Sean Hubber and Crystal Lu

Hands On ECG. Sean Hubber and Crystal Lu Hands On ECG Sean Hubber and Crystal Lu The device. The black box contains the circuit and microcontroller, the mini tv is set on top, the bars on the sides are for holding it and reading hand voltage,

More information

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION 1 SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION By Lannes S. Purnell FLUKE CORPORATION 2 This paper shows how standard signal generators can be used as leveled sine wave sources for calibrating oscilloscopes.

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 2011 Interference - 1 INTERFERENCE OF SOUND WAVES The objectives of this experiment are: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves. To observe interference phenomena

More information

Scaling and Biasing Analog Signals

Scaling and Biasing Analog Signals Scaling and Biasing Analog Signals November 2007 Introduction Scaling and biasing the range and offset of analog signals is a useful skill for working with a variety of electronics. Not only can it interface

More information

Current Loop Tuning Procedure. Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) General Procedure AN-015

Current Loop Tuning Procedure. Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) General Procedure AN-015 Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) The standard tuning values used in ADVANCED Motion Controls drives are conservative and work well in over 90%

More information

Trigonometric functions and sound

Trigonometric functions and sound Trigonometric functions and sound The sounds we hear are caused by vibrations that send pressure waves through the air. Our ears respond to these pressure waves and signal the brain about their amplitude

More information

MAS.836 HOW TO BIAS AN OP-AMP

MAS.836 HOW TO BIAS AN OP-AMP MAS.836 HOW TO BIAS AN OP-AMP Op-Amp Circuits: Bias, in an electronic circuit, describes the steady state operating characteristics with no signal being applied. In an op-amp circuit, the operating characteristic

More information

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b DIODE CIRCUITS LABORATORY A solid state diode consists of a junction of either dissimilar semiconductors (pn junction diode) or a metal and a semiconductor (Schottky barrier diode). Regardless of the type,

More information

Home Sleep Testing Common Questions and Answers

Home Sleep Testing Common Questions and Answers Home Sleep Testing Common Questions and Answers Suite 1, 100 Schneider Road Kanata, Ontario, Canada K2K 1Y2 Tel: 613.831.6690 Fax: 613.831.6699 braebon.com D.MP8.80200.2 November 2013 I thought BRAEBON

More information

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS Objective In this experiment you will study the i-v characteristics of an MOS transistor. You will use the MOSFET as a variable resistor and as a switch. BACKGROUND

More information

Underwriting Sleep Apnea

Underwriting Sleep Apnea Underwriting Sleep Apnea Joel Weiner, MD, FLMI April 29, 2014 WAHLU The Northwestern Mutual Life Insurance Company Milwaukee, WI A Brief Survey Before We Get Started The Weiner Sleepiness Scale How likely

More information

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter

More information

Isolated AC Sine Wave Input 3B42 / 3B43 / 3B44 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated AC Sine Wave Input 3B42 / 3B43 / 3B44 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated AC Sine Wave Input 3B42 / 3B43 / 3B44 FEATURES AC averaging technique used to rectify, amplify, and filter 50 Hz to 400 Hz sine-wave signals. Accepts inputs of between 20 mv to 550 V rms to give

More information

General Information about Sleep Studies and What to Expect

General Information about Sleep Studies and What to Expect General Information about Sleep Studies and What to Expect Why do I need a sleep study? Your doctor has ordered a sleep study because your doctor is concerned you may have a sleep disorder that is impacting

More information

EEG IN CHILDREN: NORMAL AND ABNORMAL. Warren T. Blume, MD,FRCPC EEG Course FSNC/CNSF JUNE 2007

EEG IN CHILDREN: NORMAL AND ABNORMAL. Warren T. Blume, MD,FRCPC EEG Course FSNC/CNSF JUNE 2007 EEG IN CHILDREN: NORMAL AND ABNORMAL Warren T. Blume, MD,FRCPC EEG Course FSNC/CNSF JUNE 2007 OBJECTIVES Survey some abnormal and normal patterns Maturation characteristics Artefact recognition Patterns

More information

Technical Note Series

Technical Note Series Technical Note Series SKIN CONDUCTANCE SENSOR (SA9309M) S TN0 0 0 8-0 0 S k i n C o n d u c t a n c e S e n s o r Page 2 IMPORTANT OPERATION INFORMATION WARNING Type BF Equipment Internally powered equipment

More information

Building the AMP Amplifier

Building the AMP Amplifier Building the AMP Amplifier Introduction For about 80 years it has been possible to amplify voltage differences and to increase the associated power, first with vacuum tubes using electrons from a hot filament;

More information

Lecture - 4 Diode Rectifier Circuits

Lecture - 4 Diode Rectifier Circuits Basic Electronics (Module 1 Semiconductor Diodes) Dr. Chitralekha Mahanta Department of Electronics and Communication Engineering Indian Institute of Technology, Guwahati Lecture - 4 Diode Rectifier Circuits

More information

Reading: HH Sections 4.11 4.13, 4.19 4.20 (pgs. 189-212, 222 224)

Reading: HH Sections 4.11 4.13, 4.19 4.20 (pgs. 189-212, 222 224) 6 OP AMPS II 6 Op Amps II In the previous lab, you explored several applications of op amps. In this exercise, you will look at some of their limitations. You will also examine the op amp integrator and

More information

TESTS OF 1 MHZ SIGNAL SOURCE FOR SPECTRUM ANALYZER CALIBRATION 7/8/08 Sam Wetterlin

TESTS OF 1 MHZ SIGNAL SOURCE FOR SPECTRUM ANALYZER CALIBRATION 7/8/08 Sam Wetterlin TESTS OF 1 MHZ SIGNAL SOURCE FOR SPECTRUM ANALYZER CALIBRATION 7/8/08 Sam Wetterlin (Updated 7/19/08 to delete sine wave output) I constructed the 1 MHz square wave generator shown in the Appendix. This

More information

Op amp DC error characteristics and the effect on high-precision applications

Op amp DC error characteristics and the effect on high-precision applications Op amp DC error characteristics and the effect on high-precision applications Srudeep Patil, Member of Technical Staff, Maxim Integrated - January 01, 2014 This article discusses the DC limitations of

More information

Prepared by: Paul Lee ON Semiconductor http://onsemi.com

Prepared by: Paul Lee ON Semiconductor http://onsemi.com Introduction to Analog Video Prepared by: Paul Lee ON Semiconductor APPLICATION NOTE Introduction Eventually all video signals being broadcasted or transmitted will be digital, but until then analog video

More information

Snoring and Obstructive Sleep Apnea (updated 09/06)

Snoring and Obstructive Sleep Apnea (updated 09/06) Snoring and Obstructive Sleep Apnea (updated 09/06) 1. Define: apnea, hypopnea, RDI, obstructive sleep apnea, central sleep apnea and upper airway resistance syndrome. BG 2. What are the criteria for mild,

More information

Changes in the Evaluation and Treatment of Sleep Apnea

Changes in the Evaluation and Treatment of Sleep Apnea Changes in the Evaluation and Treatment of Sleep Apnea Joseph DellaValla, MD FACP Medical Director Center for Sleep Medicine At Androscoggin Valley Hospital Sleep Related Breathing Problems Obstructive

More information

Exploring the Necessity of the Hot Hipot Test

Exploring the Necessity of the Hot Hipot Test Exploring the Necessity of the Hot Hipot Test Introduction In an industry comprised of workers with varying electronics knowledge, a Hipot test can seem a daunting task for some. Indeed, many test operators

More information

SLEEP STUDIES AND THERAPY MANAGEMENT

SLEEP STUDIES AND THERAPY MANAGEMENT SLEEP STUDIES AND THERAPY MANAGEMENT Effective November 1 st, 2012 Policy NHP has partnered with SMS (Sleep Management Solutions) and their parent company, CareCentrix (CCX) to provide sleep study and

More information

Transistor Amplifiers

Transistor Amplifiers Physics 3330 Experiment #7 Fall 1999 Transistor Amplifiers Purpose The aim of this experiment is to develop a bipolar transistor amplifier with a voltage gain of minus 25. The amplifier must accept input

More information

SARASOTA MEMORIAL HOSPITAL DEPARTMENT POLICY

SARASOTA MEMORIAL HOSPITAL DEPARTMENT POLICY PS1006 SARASOTA MEMORIAL HOSPITAL DEPARTMENT POLICY TITLE: NON-INVASIVE POSITIVE PRESSURE (NPPV) VENTILATION (CPAP/BIPAP) Job Title of Reviewer: EFFECTIVE DATE: REVISED DATE: POLICY TYPE: Director, Respiratory

More information

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007 PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow

More information

application note Directional Microphone Applications Introduction Directional Hearing Aids

application note Directional Microphone Applications Introduction Directional Hearing Aids APPLICATION NOTE AN-4 Directional Microphone Applications Introduction The inability to understand speech in noisy environments is a significant problem for hearing impaired individuals. An omnidirectional

More information

Type-D EEG System for Regular EEG Clinic

Type-D EEG System for Regular EEG Clinic Type-D EEG System for Regular EEG Clinic Type-D EEG amplifier Specifications 1. For Type-D Amplifier Input channels: 12/24/36/48 Monopolar EEG + 12channels Bipolar EEG+12 channels PSG. Power supply: Internal

More information

INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS

INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS This tutorial provides an overview of signal processing and conditioning for use in instrumentation and automatic control systems.

More information

ELECTRON SPIN RESONANCE Last Revised: July 2007

ELECTRON SPIN RESONANCE Last Revised: July 2007 QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron

More information

FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER

FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER 2014 Amplifier - 1 FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER The objectives of this experiment are: To understand the concept of HI-FI audio equipment To generate a frequency response curve for an audio

More information

AC 2012-3923: MEASUREMENT OF OP-AMP PARAMETERS USING VEC- TOR SIGNAL ANALYZERS IN UNDERGRADUATE LINEAR CIRCUITS LABORATORY

AC 2012-3923: MEASUREMENT OF OP-AMP PARAMETERS USING VEC- TOR SIGNAL ANALYZERS IN UNDERGRADUATE LINEAR CIRCUITS LABORATORY AC 212-3923: MEASUREMENT OF OP-AMP PARAMETERS USING VEC- TOR SIGNAL ANALYZERS IN UNDERGRADUATE LINEAR CIRCUITS LABORATORY Dr. Tooran Emami, U.S. Coast Guard Academy Tooran Emami received her M.S. and Ph.D.

More information

Effective Treatment for Obstructive Sleep Apnoea

Effective Treatment for Obstructive Sleep Apnoea Effective Treatment for Obstructive Sleep Apnoea The Series of Positive Airway Pressure devices from DeVilbiss Healthcare is designed to meet the varied needs of people suffering from Obstructive Sleep

More information

Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies

Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies Soonwook Hong, Ph. D. Michael Zuercher Martinson Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies 1. Introduction PV inverters use semiconductor devices to transform the

More information

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON. WHY DIFFERENTIAL? Voltage, The Difference Whether aware of it or not, a person using an oscilloscope to make any voltage measurement is actually making a differential voltage measurement. By definition,

More information

Lock - in Amplifier and Applications

Lock - in Amplifier and Applications Lock - in Amplifier and Applications What is a Lock in Amplifier? In a nut shell, what a lock-in amplifier does is measure the amplitude V o of a sinusoidal voltage, V in (t) = V o cos(ω o t) where ω o

More information

PCM Encoding and Decoding:

PCM Encoding and Decoding: PCM Encoding and Decoding: Aim: Introduction to PCM encoding and decoding. Introduction: PCM Encoding: The input to the PCM ENCODER module is an analog message. This must be constrained to a defined bandwidth

More information

CPAP (Continuous Positive Airway Pressure) Titration Study

CPAP (Continuous Positive Airway Pressure) Titration Study Helene A. Emsellem, MD Medical Director John R. Ruddy, MD Linda Croom, ANPC Karen Murtagh, CRNP Richard Currey, PA-C CPAP (Continuous Positive Airway Pressure) Titration Study About the CPAP Titration

More information

Troubleshooting accelerometer installations

Troubleshooting accelerometer installations Troubleshooting accelerometer installations Accelerometer based monitoring systems can be tested to verify proper installation and operation. Testing ensures data integrity and can identify most problems.

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 1/2016 Sound 1/8 INTERFERENCE OF SOUND WAVES PURPOSE: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves and to observe interference phenomena with ultrasonic sound waves.

More information

THE STRAIN GAGE PRESSURE TRANSDUCER

THE STRAIN GAGE PRESSURE TRANSDUCER THE STRAIN GAGE PRESSURE TRANSDUCER Pressure transducers use a variety of sensing devices to provide an electrical output proportional to applied pressure. The sensing device employed in the transducers

More information

Chapter 17 Medical Policy

Chapter 17 Medical Policy RAD-1 LCD for Respiratory Assist Devices (L11482) Contractor Information Contractor Name Contractor Number 00635 Contractor Type LCD Information LCD Database ID Number L11482 AdminaStar Federal, Inc. DMERC

More information

Introduction to Digital Audio

Introduction to Digital Audio Introduction to Digital Audio Before the development of high-speed, low-cost digital computers and analog-to-digital conversion circuits, all recording and manipulation of sound was done using analog techniques.

More information

The Different Types of UPS Systems

The Different Types of UPS Systems Systems White Paper 1 Revision 6 by Neil Rasmussen Contents Click on a section to jump to it > Executive summary There is much confusion in the marketplace about the different types of UPS systems and

More information

CPAP titration: PSG technologist or at Home

CPAP titration: PSG technologist or at Home CPAP titration: PSG technologist or at Home Carolyn D Ambrosio, MD, MS Associate Professor of Medicine Director, The Center for Sleep Medicine Tufts Medical Center Conflict of Interest No financial interest

More information

These help quantify the quality of a design from different perspectives: Cost Functionality Robustness Performance Energy consumption

These help quantify the quality of a design from different perspectives: Cost Functionality Robustness Performance Energy consumption Basic Properties of a Digital Design These help quantify the quality of a design from different perspectives: Cost Functionality Robustness Performance Energy consumption Which of these criteria is important

More information

See Horenstein 4.3 and 4.4

See Horenstein 4.3 and 4.4 EE 462: Laboratory # 4 DC Power Supply Circuits Using Diodes by Drs. A.V. Radun and K.D. Donohue (2/14/07) Department of Electrical and Computer Engineering University of Kentucky Lexington, KY 40506 Updated

More information

Buffer Op Amp to ADC Circuit Collection

Buffer Op Amp to ADC Circuit Collection Application Report SLOA098 March 2002 Buffer Op Amp to ADC Circuit Collection Bruce Carter High Performance Linear Products ABSTRACT This document describes various techniques that interface buffer op

More information

LABORATORY 10 TIME AVERAGES, RMS VALUES AND THE BRIDGE RECTIFIER. Bridge Rectifier

LABORATORY 10 TIME AVERAGES, RMS VALUES AND THE BRIDGE RECTIFIER. Bridge Rectifier LABORATORY 10 TIME AVERAGES, RMS VALUES AND THE BRIDGE RECTIFIER Full-wave Rectification: Bridge Rectifier For many electronic circuits, DC supply voltages are required but only AC voltages are available.

More information

Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide)

Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide) Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide) OVERVIEW Students will measure a sound wave by placing the Ward s DataHub microphone near one tuning fork A440 (f=440hz). Then

More information

Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003

Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003 Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003 Knowing the functions of the anesthetic delivery system

More information

Centrifugal Fans and Pumps are sized to meet the maximum

Centrifugal Fans and Pumps are sized to meet the maximum Fans and Pumps are sized to meet the maximum flow rate required by the system. System conditions frequently require reducing the flow rate. Throttling and bypass devices dampers and valves are installed

More information

Manual Analysis Software AFD 1201

Manual Analysis Software AFD 1201 AFD 1200 - AcoustiTube Manual Analysis Software AFD 1201 Measurement of Transmission loss acc. to Song and Bolton 1 Table of Contents Introduction - Analysis Software AFD 1201... 3 AFD 1200 - AcoustiTube

More information

RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA

RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA ABSTRACT Random vibration is becoming increasingly recognized as the most realistic method of simulating the dynamic environment of military

More information

The Different Types of UPS Systems

The Different Types of UPS Systems White Paper 1 Revision 7 by Neil Rasmussen > Executive summary There is much confusion in the marketplace about the different types of UPS systems and their characteristics. Each of these UPS types is

More information

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display

More information

Common-Emitter Amplifier

Common-Emitter Amplifier Common-Emitter Amplifier A. Before We Start As the title of this lab says, this lab is about designing a Common-Emitter Amplifier, and this in this stage of the lab course is premature, in my opinion,

More information

The 2N3393 Bipolar Junction Transistor

The 2N3393 Bipolar Junction Transistor The 2N3393 Bipolar Junction Transistor Common-Emitter Amplifier Aaron Prust Abstract The bipolar junction transistor (BJT) is a non-linear electronic device which can be used for amplification and switching.

More information

Impact Uni-Vent 754 Portable Ventilator

Impact Uni-Vent 754 Portable Ventilator Impact Uni-Vent 754 Portable Ventilator Description - Indications- Contraindications- Side Effects- Special Considerations- The Uni-Vent 754 Portable Ventilator is a portable electronically controlled,

More information

Line Reactors and AC Drives

Line Reactors and AC Drives Line Reactors and AC Drives Rockwell Automation Mequon Wisconsin Quite often, line and load reactors are installed on AC drives without a solid understanding of why or what the positive and negative consequences

More information

Interference in the ECG and its elimination

Interference in the ECG and its elimination Interference in the ECG and its elimination General points The signal voltages in ECG recording are known to be very small and are in the millivolt range, i.e. they have amplitudes of only a few thousands

More information

Characterizing Digital Cameras with the Photon Transfer Curve

Characterizing Digital Cameras with the Photon Transfer Curve Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently

More information

A Low-Cost, Single Coupling Capacitor Configuration for Stereo Headphone Amplifiers

A Low-Cost, Single Coupling Capacitor Configuration for Stereo Headphone Amplifiers Application Report SLOA043 - December 1999 A Low-Cost, Single Coupling Capacitor Configuration for Stereo Headphone Amplifiers Shawn Workman AAP Precision Analog ABSTRACT This application report compares

More information

Basics of Digital Recording

Basics of Digital Recording Basics of Digital Recording CONVERTING SOUND INTO NUMBERS In a digital recording system, sound is stored and manipulated as a stream of discrete numbers, each number representing the air pressure at a

More information

Lab 5 Operational Amplifiers

Lab 5 Operational Amplifiers Lab 5 Operational Amplifiers By: Gary A. Ybarra Christopher E. Cramer Duke University Department of Electrical and Computer Engineering Durham, NC. Purpose The purpose of this lab is to examine the properties

More information