Split / Splitless Injection for Capillary GC

Size: px
Start display at page:

Download "Split / Splitless Injection for Capillary GC"

Transcription

1 Split / Splitless Injection for Capillary GC

2 Speakers John V Hinshaw CHROMacademy GC Dean GC Connections editor Moderator Tony Taylor CHROMacademy Trainer Technical Director, Crawford Scientific Dave Walsh Editor In Chief LCGC Magazine

3

4 Aims & Objectives 1. Overview of Injectors for Capillary GC 2. Injector components 3. Split / Splitless Injection Overview 4. Gas flows in Split and Splitless Mode 5. Critical operating parameters in Split and Splitless mode 6. Optimising injector settings for maximum sensitivity and repeatability 7. Inlet liners the critical facts 8. Troubleshooting Inlet Hardware 9. Investigating Irreproducibility and Poor Peak Shape

5 Overview of Injectors for Capillary GC 1. Injectors and Inlets are used to introduce the sample to the GC column 2. Different classes of Injectors and Inlets available: a. Vaporising Injector (including Split / Splitless) b. Cool-on-Column Injector (Thermally Labile or Accurate Low Level Quant) c. Large Volume / Programmed Thermal Vaporising Injector d. Headspace Inlet e. Purge and Trap Inlet f. Thermal Desorption Inlet g. Pyrolysis Inlet

6 Characteristics of GC Injectors 1. Low / No Contribution to Band Broadening 2. Introduces representative & homogenous sample 3. No discrimination based on differences in analyte b.pt., polarity, concentration 4. Avoids thermal / catalytic degradation 5. Good accuracy & precision with a wide range of analyte concentrations

7 Injector Anatomy

8 01_overview.flv

9 Split Injection Mechanisms I 1. Sample syringe pierces septum which seals around needle 2. Sample rapidly introduced into heated inlet 3. Liquid sample volatilises and the gaseous plasma is contained within a quartz glass liner

10 Split Injection Mechanisms II 4. The sample gas is swept by the carrier gas through the liner and EITHER into the GC Column OR between the liner and inlet body and down the Split Line 5. % of sample reaching the column depends upon the relative flow rates in the column and split flow line

11 02_Split_Inj.flv

12 Setting the Split Ratio I 1. Split ratio is the ratio of gas flows through the column and split line 2. Represents the volume fraction of sample entering the column 3. Split Ratios from 1:1 to 500:1 are common 4. Higher split smaller amount of sample on column

13 Setting the Split Ratio II 5. Avoids column overload fronting peaks and poor area reproducibility 6. Column capacity depends upon film thickness, column i.d., polarity and retention 7. HIGHER split ratios give SHARPER (more EFFICIENT) PEAKS

14 03_Split_Ratio.flv

15 Effect of Split Flow on Peak Shape 1. As split flow increases Liner Flow increases 2. Liner flow is a combination of the column flow and the split flow 3. Gaseous sample is transferred more rapidly onto the column 4. Net result is a decrease is analyte band width (peak width) at the column head 5. The analyte band will disperse during elution but initial bandwidth has impact on peak effciency

16 05_peakShape.flv

17 Split Injection Discrimination Normalised response of n-alkanes in Hexane

18 06_discrimination.flv

19 Split Injection Set-Up Summary 1. Used as the Default Vaporising Injector 2. Primarily used for non-trace analysis of volatile samples 3. Need to consider gas flows (particularly split flow) carefully / Don t forget septum purge flow! 4. Increasing split flow: a. Improves peak shape b. Lowers column loading c. Lowers analytical sensitivity d. Decreases analyte inlet residence time therefore reduces the opportunity for thermal degradation 5. Need to consider Discrimination effects

20 Split Injection Default / Development Conditions Temperature: Mode: Septum Purge flow: Split Flow: Column Flow: Injection Volume: Liner: Injection Solvent: Column Temp.: 250 o C Split 1-3 ml/min (instrument dependant) 100ml/min ml/min (note: depends upon column) 1ml (check for backflash) Straight through (deactivated and packed if necessary) Match to column chemistry o C (note: analyte dependant)

21 Split Injection Advantages & Disadvantages Advantages 1. Simple to Use 2. Rugged Design 3. Narrow analyte band on column 4. Protects column from involatile sample components 5. Easy to Automate Disadvantages 1. Not suitable for ultra-trace analysis 2. Suffers from Discrimination 3. Liner geometry dictates injector settings 4. Analytes susceptible to thermal degradation

22 Splitless Injection Mechanism I 1. Same principle as Split Injection 2. DIFFERENCES INCLUDE 3. Initial injector state is SPLITLESS i.e. The split line flow is turned off 4. All sample reaches the column 5. Sample vapours trapped onto head of column (solvent and thermal effects)

23 Splitless Injection Mechanism II 6. Column temperature programmed to initiate elution 7. At some point after analyte transfer to the column the split line is turned on to empty the injector 8. Primarily used for trace and ultra-trace analysis

24 07_Splitless.flv

25 Splitless Injection Gas Flows I 1. Flow through the liner = Column Flow during Splitless phase 2. Analyte can take MINUTES to transfer to the column 3. If no action is taken, chromatographic peaks will be unacceptably wide

26 Splitless Injection Gas Flows II 4. The answer is to FOCUS the analytes onto the head of the column using Thermal and Solvent focussing effects 5. The SPLITLESS (Split-On or Purge) time needs to be carefully considered

27 08_Liner_Flow.flv

28 Splitless Analyte Thermal Focussing Rule of thumb: Initial Oven Temp 10 o C < Solvent B.Pt.

29 09_thermal_focussing.flv

30 Splitless Analyte Solvent Focussing N-alkanes in CS2

31 10_solvent_focussing.flv

32 Splitless Analyte Solvent Focussing II N-alkanes in CS2 with Wax Column Rule of thumb: Match solvent and column polarity

33 11_solvent_focus_2.flv

34 Splitless Injection - Optimising Purge Time Use the peak area of an early eluting peak to determine analyte transfer time

35 Splitless Injection Default / Development Conditions Temperature: Mode: Septum Purge flow: Purge Time: Split Flow: Column Flow: Injection Volume: Liner: Injection Solvent: Column Temp.: 250 o C Splitless 1-3 ml/min (instrument dependant) 30 seconds (NOTE: optimise empirically) 100ml/min ml/min (note: depends upon column) 1ml (check for backflash) Straight through (splitless liner) Match to column chemistry <10 o C below solvent B.Pt.

36 Split Injection Advantages & Disadvantages Advantages 1. Simple to Use 2. Rugged Design 3. Excellent for trace analysis 4. Less Risk of Analyte Discrimination than Split Mode 5. Easy to Automate Disadvantages 1. Need to carefully optimise conditions 2. Risk of backflash 3. Analytes susceptible to thermal degradation

37 Optimising Injection Volume 1. Solvent vapour volume should not exceed liner volume 2. Blank solvent injection following sample will give a mini-chromatogram

38 12_optimising_injection.flv

39 Calculating Optimum Injection Volume 1. Vapour expansion volume is dependant upon: a. Solvent type b. Injection Port Temp. c. Inlet Pressure 2. Backflash dependant upon: a. Liner Type b. Vapour expansion volume c. Injection Volume

40 13_backFlash_calc.flv

41 Liners for Split / Splitless Injection Straight Through Top Features Middle Features Bottom Features Liner i.d.

42 Troubleshooting Temperature / Activity Thermal Degradation If in doubt use 250 o C as an inlet starting point Peak Tailing 1. Consider Volumetric tailing (due to unswept dead volumes) 2. Chemically de-activate the Liner / inlet / column

43 Troubleshooting Septa Septum Bleed Septa Considerations 1. Temperature Limit 2. Size 3. Material 4. Sandwich 5. PTFE Faced 6. Pre-Drilled

44 e-learning Topics 1000 s of articles by your favourite LCGC Authors 1000 s of Applications 100 s of News Items every day Social Network LC / GC / MS / Sample Prep. Annual Subscription $

Gas Chromatography Liner Selection Guide

Gas Chromatography Liner Selection Guide Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing

More information

Mass Spectrometry. Fundamental GC-MS. GC Considerations

Mass Spectrometry. Fundamental GC-MS. GC Considerations Mass Spectrometry Fundamental GC-MS GC Considerations i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax. Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative

More information

Expectations for GC-MS Lab

Expectations for GC-MS Lab Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will

More information

THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles

THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles THERMAL DESORPTION Introduction and Principles Thermal Desorption Thermal desorption is a simple extension of the technique of gas chromatography (GC) It involves the use of heat and a flow of inert gas

More information

Analysis of Liquid Samples on the Agilent GC-MS

Analysis of Liquid Samples on the Agilent GC-MS Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >

More information

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory

More information

Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder

Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder Lecture Chromo-3: Gas Chromatography CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder Outline Introduction Instrument overview Carrier gas Sample

More information

GC-xt. The most flexible GC /GC-MS liquid sample injector

GC-xt. The most flexible GC /GC-MS liquid sample injector GC-xt Prep and Load Platform The most flexible GC /GC-MS liquid sample injector Environmental Foods/Beverages Consumer Products Forensics Petrochemicals/Polymers Pharmaceuticals GC-xt Liquid mode Prep

More information

The Theory of HPLC. Gradient HPLC

The Theory of HPLC. Gradient HPLC The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims

More information

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2

More information

Analyzing Small Molecules by EI and GC-MS. July 2014

Analyzing Small Molecules by EI and GC-MS. July 2014 Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,

More information

Gas Chromatograph GC Consumables. p/n: 220-94898-02

Gas Chromatograph GC Consumables. p/n: 220-94898-02 Gas Chromatograph GC Consumables p/n: 220-94898-02 Diag. # 1) Septum Nut 2) Needle Guide 3) Septum 4) Injection Port A Assy 5) O-ring for Glass Insert 6) Glass Insert 7) Gold Gasket 8) Capillary Adaptor

More information

How To Use Gc-Ms

How To Use Gc-Ms The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter

More information

23 The Thermal Conductivity Detector

23 The Thermal Conductivity Detector 23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing

More information

consultancy services

consultancy services consultancy services Method Troubleshooting / Method Improvement (On-site or at our UK Laboratories) Problematic & poorly performing analytical methods have a high cost in terms of wasted analysis time

More information

Background Information

Background Information 1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software

More information

Micro Cryo-Trap Model 971

Micro Cryo-Trap Model 971 Micro Cryo-Trap Model 971 for Liquid CO2 Cooling Scientific Instrument Services, Inc. 1027 Old York Road Ringoes, NJ 08551 Phone (908) 788-5550 Part #951001M revised 10/ /98 Notice The information in this

More information

A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water

A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water Sjaak de Koning 1, Mark van Lieshout 2, Hans-Gerd Janssen 2 and Wil van Egmond 1 1 ATAS, De Run 4334, PO Box

More information

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated

More information

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC The Essential CHROMacademy Guide Mobile Phase Optimization Strategies for Reversed Phase HPLC Webcast Notes Type your questions in the Submit Question box, located below the slide window You can enlarge

More information

Gas Chromatography. Gas Chromatograph

Gas Chromatography. Gas Chromatograph Gas Chromatography Gas Chromatograph Chromatography: Separate analytes in a mixture with a resolution 1.5 in the shortest amount of time and detection of separated components. 1 = m.p. 6 2 4 5 3 Gas supplies

More information

GC Troubleshooting 10. How to use this Troubleshooting Help system... 10. GC Troubleshooting - Top Level... 10 Generic problem descriptions...

GC Troubleshooting 10. How to use this Troubleshooting Help system... 10. GC Troubleshooting - Top Level... 10 Generic problem descriptions... Table of Contents GC Troubleshooting 10 How to use this Troubleshooting Help system... 10 GC Troubleshooting - Top Level... 10 Generic problem descriptions... 11 Baseline related problems... 11 Peak related

More information

Operating Hints for Using Split/Splitless Injectors

Operating Hints for Using Split/Splitless Injectors Technical Guide 1 Operating Hints for Using Split/Splitless Injectors Inside: Overviews of split and splitless injection tech niques Backpressure-regulated injection systems Headpressure-regulated injection

More information

Quantitative Evaluation of Perfume Fleuressence Samples using the znose

Quantitative Evaluation of Perfume Fleuressence Samples using the znose Quantitative Evaluation of Perfume Fleuressence Samples using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses Conventional electronic noses (enoses) produce a recognizable response

More information

ORGANIC SAMPLE PREPARATION

ORGANIC SAMPLE PREPARATION ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,

More information

AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY. Andrew Tipler

AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY. Andrew Tipler AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY Andrew Tipler Chromatography Research and Technology Manager PerkinElmer, Inc. Table of Contents Introduction 3 Fundamental

More information

Fractional Distillation and Gas Chromatography

Fractional Distillation and Gas Chromatography Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils

More information

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ)

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ) Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) Closing Date and Time: Wednesday, April 13, 2016 at 5:00 p.m. (Pacific Time) Contact: Aaron Galloway Fax: 541-888-3250

More information

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

TechNote 3/2002. The Use of Different PTV Inlet Liner Types for Trapping Alkanes, Aromatics and Oxygenated Compounds During Thermal Desorption

TechNote 3/2002. The Use of Different PTV Inlet Liner Types for Trapping Alkanes, Aromatics and Oxygenated Compounds During Thermal Desorption TechNote 3/22 The Use of Different PTV Inlet Liner Types for Trapping Alkanes, Aromatics and Oxygenated Compounds During Thermal Desorption Bita Kolahgar Gerstel GmbH & Co. KG, Eberhard-Gerstel-Platz 1,

More information

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at

More information

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and

More information

C184-E014D GC-2014. Shimadzu Gas Chromatograph System

C184-E014D GC-2014. Shimadzu Gas Chromatograph System C184-E014D GC-2014 Shimadzu Gas Chromatograph System Big Performance & Small Space GC-2014 GC-2014 Shimadzu Gas Chromatograph System High Performance and Expandability Merged at a higher level High Performance

More information

Diffusion and Fluid Flow

Diffusion and Fluid Flow Diffusion and Fluid Flow What determines the diffusion coefficient? What determines fluid flow? 1. Diffusion: Diffusion refers to the transport of substance against a concentration gradient. ΔS>0 Mass

More information

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

Automation of Solid Phase Extraction and Column Chromatographic Cleanup Automation of Solid Phase Extraction and Column Chromatographic Cleanup Haibin Wan Ltd. Problems with Manual Operation 1. Unstable flow rate affects reproducibility and throughput. 2. Not convenient for

More information

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note. EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer HT3 Application Note Introduction The analysis of dissolved gases in ground water is important in

More information

Chapter 28: High-Performance Liquid Chromatography (HPLC)

Chapter 28: High-Performance Liquid Chromatography (HPLC) Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion

More information

Identification of Arson Accelerants by Gas Chromatography

Identification of Arson Accelerants by Gas Chromatography Identification of Arson Accelerants by Gas Chromatography Purpose Arson is an insidious crime that annually claims the lives of hundreds of Americans and costs billions of dollars. 1 Scientists and criminal

More information

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas

More information

FOR RISK ASSESSMENT. Analysis of mineral oil: Manual pre-separation, followed by LV-GC/FID Off-line HPLC-GC/FID FEDERAL INSTITUTE.

FOR RISK ASSESSMENT. Analysis of mineral oil: Manual pre-separation, followed by LV-GC/FID Off-line HPLC-GC/FID FEDERAL INSTITUTE. FEDERAL INSTITUTE FOR RISK ASSESSMENT Analysis of mineral oil: Manual pre-separation, followed by LV-GC/FID Off-line HPLC-GC/FID Christoph Hutzler Topics Design of the manual methods Large Volume GC injections

More information

AMD Analysis & Technology AG

AMD Analysis & Technology AG AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds

More information

EnAlgae SOP: Biogas analysis

EnAlgae SOP: Biogas analysis Table of content 1. PURPOSE... 2 2. PRINCIPLE... 2 3. REQUIREMENTS... 3 3.1. EQUIPMENT AND MATERIALS... 3 3.2. REAGENTS... 3 3.3. PREPARATION OF MEASUREMENT... 3 4. PROCEDURE... 4 4.1. SAMPLE MEASUREMENT...

More information

Theory and Instrumentation of GC Introduction

Theory and Instrumentation of GC Introduction Theory and Instrumentation of GC Introduction i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Multi-Shot Pyrolyzer EGA/PY-3030D. Flexible Versatile Reproducible

Multi-Shot Pyrolyzer EGA/PY-3030D. Flexible Versatile Reproducible Multi-Shot Pyrolyzer EGA/PY-3030D Flexible Versatile Reproducible Why Evolved Gas Analysis? Why Pyrolysis? Today, analytical pyrolysis encompasses much more than simple flash pyrolysis of polymeric materials.

More information

Purification of reaction mixtures using flash chromatography.

Purification of reaction mixtures using flash chromatography. Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?

More information

SUCRALOSE. White to off-white, practically odourless crystalline powder

SUCRALOSE. White to off-white, practically odourless crystalline powder SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D.

Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D. Gas Chromatograph Mass Spectrometer No. GCMS-1603 Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS Shilpi Chopra, Ph.D. Introduction Organophosphorus (OP) pesticides are a class

More information

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas

More information

Clarus 580 Gas Chromatograph

Clarus 580 Gas Chromatograph S P E C I F I C A T I O N S Gas Chromatography Clarus 580 Gas Chromatograph The PerkinElmer Clarus 580 Gas Chromatograph (GC) is a fully automated gas chromatograph. The system offers the capabilities

More information

Series 2. Air Server. Options for UNITY 2. A robust, cryogen-free solution for round-the-clock on-line air monitoring

Series 2. Air Server. Options for UNITY 2. A robust, cryogen-free solution for round-the-clock on-line air monitoring Series 2 Air Server Options for UNITY 2 A robust, cryogen-free solution for round-the-clock on-line air monitoring Series 2 Air Server Three- and eight-channel options for UNITY offering round-the-clock

More information

The world's largest elearning website for analytical scientists

The world's largest elearning website for analytical scientists The world's largest elearning website for analytical scientists CHROMacademy will help increase your knowledge, efficiency and productivity in the lab HPLC GC MS SPE IR BLS BIO www.chromacademy.com HPLC

More information

General aspects of GC analysis of volatiles

General aspects of GC analysis of volatiles General aspects of GC analysis of volatiles Kaunas University of Technology Content Introduction: food volatiles, principles of GC separation Main parameters for the optimisation of volatile analysis columns

More information

Thermal Desorption Unit TDU

Thermal Desorption Unit TDU Thermal Desorption Unit TDU GERSTEL Thermal DesorptionUnit TDU Thermal Desorption Unit TDU The world s most flexible Thermal Desorption Instrument The Thermal Desorption Unit (TDU) is a flexible automated

More information

PLOT Columns. Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative

PLOT Columns. Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative PLOT Columns Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative PLOT Column Selection 93 Features & Benefits of Restek PLOT Columns Features

More information

ß-CYCLODEXTRIN SYNONYMS

ß-CYCLODEXTRIN SYNONYMS ß-CYCLODEXTRIN Prepared at the 44th JECFA (1995), published in FNP 52 Add 3 (1995) superseding specifications prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications

More information

Series 2. Continuous monitoring of trace chemical vapours

Series 2. Continuous monitoring of trace chemical vapours Series 2 TT24-7 Continuous monitoring of trace chemical vapours Series 2 TT24-7 TT24-7 Series 2 Continuous monitoring of trace chemical vapours The TT24-7 Series 2 is an advanced thermal desorption (TD)

More information

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic

More information

METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY

METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8020 is used to determine the concentration of various aromatic volatile organic compounds. The following

More information

Aromatic plants: from extraction to analysis

Aromatic plants: from extraction to analysis Aromatic plants: from extraction to analysis Thierry TALOU Agro-industrial Chemistry Laboratory National Polytechnic Institute of Toulouse ENSIACET 4 allée Emile Monso 31432 Toulouse Olfactory system Olfactory

More information

With proper calibration, the upper limit may be extended as needed.

With proper calibration, the upper limit may be extended as needed. While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the

More information

GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits

GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits Issue Implementation Page 1 of 7 GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits Scope and Application This method will determine the level of vanillin and ethyl vanillin in distilled

More information

ChromQuest 5.0 Chromatography Data System

ChromQuest 5.0 Chromatography Data System ChromQuest 5.0 Chromatography Data System User Guide for the Surveyor LC CHROM-97257 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. Surveyor and ChromQuest are trademarks

More information

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview. Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide

More information

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,

More information

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* Method 552.1 Revision 1.0 Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* UCT Products: EUQAX156 (quaternary amine with Cl - counter ion, 6 ml cartridge)** CLTTP050 (CLEAN-THRU

More information

Syringes 5. Contents. Autosampler 6. MEPS 18 Product Data 18 Part Numbers 19. Manual 20. Guided Plunger 24 Product Data 24 Part Numbers 25

Syringes 5. Contents. Autosampler 6. MEPS 18 Product Data 18 Part Numbers 19. Manual 20. Guided Plunger 24 Product Data 24 Part Numbers 25 Syringes Injection Port Contents Syringes 5 Autosampler 6 Diamond Headspace 6 Product Data 6 Part Numbers 7 Diamond MS 8 Product Data 8 Part Numbers 9 GC Autosampler 11 Product Data 11 Part Numbers 12

More information

SCION SQ Series GC-MS

SCION SQ Series GC-MS SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative

More information

Enhancing Utility and Throughput of SPME using Advanced Automation Strategies

Enhancing Utility and Throughput of SPME using Advanced Automation Strategies Supelco Innovation Seminars November 11 th, 2010 Atlanta, GA Enhancing Utility and Throughput of SPME using Advanced Automation Strategies Presented by Tom Tobien, Ph.D. Analytical Chemist, LEAP Technologies,

More information

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract

More information

Application. Gas Chromatography September 1997

Application. Gas Chromatography September 1997 Trace Level Pesticide Analysis by GC/MS Using Large-Volume Injection Application Gas Chromatography September 1997 Author Philip L. Wylie Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE

More information

High-Throughput 3-D Chromatography Through Ion Exchange SPE

High-Throughput 3-D Chromatography Through Ion Exchange SPE High-Throughput 3-D Chromatography Through Ion Exchange SPE Application Note 205 Luke Roenneburg and Alan Hamstra (Gilson, Inc.) Introduction 2-dimensional (2-D) separation is the separation of a sample

More information

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen

More information

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting

More information

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen. The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen. Mark Crawford Single Presenter Name Abstract With the recent decline in hive bee populations this year the investigation of acaricides

More information

Filter Vials. Product Knowledge. PTFE PVDF Nylon PES. Pat. US #7,790,117 Worldwide patents pending. Thomson Instrument Company Toll Free 800 541.

Filter Vials. Product Knowledge. PTFE PVDF Nylon PES. Pat. US #7,790,117 Worldwide patents pending. Thomson Instrument Company Toll Free 800 541. Product Knowledge Filter Vials PTFE PVDF Nylon PES.2µm.45µm.2µm.45µm.2µm.45µm.2µm.45µm Pat. US #7,790,117 Worldwide patents pending Thomson Instrument Company Toll Free 800 541.4792 Email folks@htslabs.com

More information

Setting up a Quantitative Analysis MS ChemStation

Setting up a Quantitative Analysis MS ChemStation Setting up a Quantitative Analysis MS ChemStation Getting Ready 1. Use the tutorial section "Quant Reports" on the MSD Reference Collection CD-ROM that came with your ChemStation software. 2. Know what

More information

Micromass LCT User s Guide

Micromass LCT User s Guide Micromass LCT User s Guide 1) Log on to MassLynx with your username & password. 2) After you have logged in, the MassLynx software will automatically run. 3) After MassLynx has come up, open your project

More information

CCME. Reference Method for the Canada- Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method

CCME. Reference Method for the Canada- Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method CCME Canadian Council of Ministers of the Environment Le Conseil canadien des ministres de l environnement Reference Method for the Canada- Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method

More information

Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps

Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps Application Note: ANTV10 Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps Abstract In order to regulate product safety and quality

More information

Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana*

Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana* Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana* a Department of Pharmaceutical Analysis, Andhra University, Visakhapatnam. b Department of Pharmaceutics,

More information

GCxGC COUPLED TO FAST SCANNING QUADRUPOLE MS FOR TRACE ANALYSIS OF POPs

GCxGC COUPLED TO FAST SCANNING QUADRUPOLE MS FOR TRACE ANALYSIS OF POPs GCxGC COUPLED TO FAST SCANNING QUADRUPOLE MS FOR TRACE ANALYSIS OF POPs Kinet C, De Pauw E, Focant JF CART, Mass Spectrometry Laboratory, Chemistry Department, University of Liège, Allée de la Chimie 3,

More information

Shimadzu Simulated Distillation Gas Chromatograph System C184-E030

Shimadzu Simulated Distillation Gas Chromatograph System C184-E030 Shimadzu Simulated Distillation Gas Chromatograph System C184-E030 Meets All Simulated Distillation Gas Chromatograph Standards - Provides Highly Accurate Analysis Results with Excellent Reproducibility-

More information

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Carmen T. Santasania and David S. Bell SUPELCO, 595 North Harrison Road, Bellefonte, PA 16823 T404045 GUS Abstract

More information

Capillary and Packed Gas Chromatograph

Capillary and Packed Gas Chromatograph Capillary and Packed Gas Chromatograph C184-E014G High Performance and Expandability Merged at a higher level Capillary and Packed Gas Chromatograph High Performance Superior Performance Improved design

More information

Process Instrumentation

Process Instrumentation Process Instrumentation Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Process Gas Chromatography Features of Process Chromatographs Agilent 7890A

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

Agilent G1701DA MSD ChemStation

Agilent G1701DA MSD ChemStation Agilent G1701DA MSD ChemStation 5973 MSD / ChemStation Quick Reference Agilent Technologies What s New in Revision D.01.00 A new toolbar has been added to the Top view and an expanded set of toolbars have

More information

Consumable Parts Catalogue Ver.3.06

Consumable Parts Catalogue Ver.3.06 Consumable Parts Catalogue Ver.3.06 February, 2012 This catalogue contains consumable parts for ; Pyrolyzer (EGA/PY-3030D, PY-3030S, PY-2020iD/iS) Selective Sampler (SS-1010E) MicroJet Cryo-Trap (MJT-1030Ex)

More information

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK. VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK. VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS PRINCIPLE This method confirms the identity of volatile compounds in biological

More information

Agilent 7697A Headspace Sampler Data Sheet

Agilent 7697A Headspace Sampler Data Sheet Agilent 7697A Headspace Sampler Data Sheet Overview The Agilent 7697A Headspace Sampler is part of Agilent s line of gas chromtography products. The 7697A based on the architecture of the market-leading

More information

CHROMOCTANE O N LINE ANALYZER

CHROMOCTANE O N LINE ANALYZER GECIL Process GT instruments CHROMOCTANE O N LINE ANALYZER ASTM D6733 DHA and RON and MON on line GC for Reformer unit Automatic Chromatographic system for analysis of naphta feeds and reforming gasolines

More information

the fastest injection-to-injection time

the fastest injection-to-injection time Clarus 600 Gas Chromatograph the fastest injection-to-injection time in conventional gas chromatography Clarus 600 Gas Chromatograph the innovation is simply its speed Clarus 600 Gas Chromatograph Every

More information

Analysis of Drugs by using the Ultra Inert Inlet Liners with Wool

Analysis of Drugs by using the Ultra Inert Inlet Liners with Wool nalysis of Drugs by using the Ultra Inert Inlet Liners with Wool pplication Note Forensic Toxicology uthors Limian Zhao, ruce Quimby gilent Technologies, Inc. 280 Centerville Road Wilmington, DE 19808

More information

IFC. A new perspective i l ifi ti. p p g p. Single softw FractPAL software plug-in for Agilent ChemStation. Medicinal Chemistry.

IFC. A new perspective i l ifi ti. p p g p. Single softw FractPAL software plug-in for Agilent ChemStation. Medicinal Chemistry. IFC A new perspective i l ifi ti Medicinal Chemistry p p CombiChem Compound Isolation Autopurification Drug Discovery y y q p p p p g p Single softw FractPAL software plug-in for Agilent ChemStation The

More information

Tuning & Mass Calibration

Tuning & Mass Calibration Tuning & Mass Calibration 1 1 The Sample List Sample List Name Project Name 2 The sample list is the top level screen in the TurboMass Gold Software. Data storage is set up in PROJECT files and within

More information