Keywords: Torque, calibration, uncertainty, torque tools, torque testers. Contents



Similar documents
Guidelines on the Calibration of Static Torque Measuring Devices

Series 6000 Torque measured metal bellow coupling

OPERATOR S MANUAL SUPPORTED TORQUE CALIBRATION BEAM. Part Number Issue 4 Original Instructions (English)

SCHATZ - TEST Tool testing systems. SCHATZ -certest-w. Dynamic testing of torque tools

Operating instructions

Dampers. Inlet guide vane. Damper. Louvre damper. Handbook Radial Fans / Chapter 3 - Dampers REITZ List 2010 DO 1. Dampers

INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS. Published: April 1999

Series 7500 Torque Sensor

DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements

Reaction Torque Sensor

AUTOMATED, FULL LOAD MOTOR TESTING AT PRODUCTION SPEEDS

Gravimetric determination of pipette errors

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS

Surface roughness. Picture 1: Friction coefficient as a function of surface roughness Curve 1: Roughness R z 0,5µm Curve 2: Roughness R z 5 µ

Torque motors. direct drive technology

Torque and Rotation. Physics

PTB 311E Manual 3-in-1 Tablet Testing Instrument

Schedule of Accreditation issued by United Kingdom Accreditation Service High Street, Feltham, Middlesex, TW13 4UN, UK

LEVEL GAUGES WITH MAGNETIC JOINT L 14 - L 22 - L 34

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

Chapter 2 Lead Screws

Active Vibration Isolation of an Unbalanced Machine Spindle

Torque Transducers MEASURING ELECTRONIC

THE COMPOSITE DISC - A NEW JOINT FOR HIGH POWER DRIVESHAFTS

LOCATION DEPENDENCY OF POSITIONING ERROR IN A 3-AXES CNC MILLING MACHINE

The Effects of Wheelbase and Track on Vehicle Dynamics. Automotive vehicles move by delivering rotational forces from the engine to

Polarization of Light

User checks and maintenance of laboratory balances

AP Physics 1 and 2 Lab Investigations

Estimating Weighing Uncertainty From Balance Data Sheet Specifications

MINIMAT-EC-Servo Screwdriver Spindles

HPF Hollow Shaft Unit Type

Chapter 11 Equilibrium

FREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5

Design of a Universal Robot End-effector for Straight-line Pick-up Motion

Step 11 Static Load Testing

Procon Engineering. Technical Document PELR TERMS and DEFINITIONS

E X P E R I M E N T 8

TECHANGLE November 30, 2005 TECHNICAL PAGE

Machine Tool Inspection & Analyzer solutions

Standex-Meder Electronics. Custom Engineered Solutions for Tomorrow

Model Year: 2010 Model: Prius Doc ID: RM000001Y3B015X. Title: ALIGNMENT / HANDLING DIAGNOSIS: FRONT WHEEL ALIGNMENT: ADJUSTMENT (2010 Prius)

Experiment: Static and Kinetic Friction

E/ECE/324/Rev.2/Add.116/Rev.2/Amend.2 E/ECE/TRANS/505/Rev.2/Add.116/Rev.2/Amend.2

Mounting Instructions. Torque Transducer. A en

Wheeled Vehicle Design For Science Olympiad By Carey I. Fisher

Subminiature Load Cell Model 8417

Steering unit. Table of contents. Features RE 11872/ Type LAGL. Frame sizes 500 to 1000 Component series 1X Maximum flow 80 l/min

due to uncertainty. This, in turn, has a direct impact on equipment availability and maintenance costs. Unfortunately, due to misconceptions and

Chapter 3.8 & 6 Solutions

STATIC AND KINETIC FRICTION

Application Information

A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion

Questions: Does it always take the same amount of force to lift a load? Where should you press to lift a load with the least amount of force?

Description. Dimensions. Features. precision works better

Estimates of Uncertainty of the Calibration of Balances

F1 Fuel Tank Surging; Model Validation

Mobile field balancing reduces vibrations in energy and power plants. Published in VGB PowerTech 07/2012

POWERS CONTROL RC 195 Single Input Receiver-Controller

CALIBRATION PRINCIPLES

SPINDLE ERROR MOVEMENTS MEASUREMENT ALGORITHM AND A NEW METHOD OF RESULTS ANALYSIS 1. INTRODUCTION

DINAMIC AND STATIC CENTRE OF PRESSURE MEASUREMENT ON THE FORCEPLATE. F. R. Soha, I. A. Szabó, M. Budai. Abstract

Via Stelvio, Marnate (VA), Italy Tel Fax info@italcontrol.net - Website:

Ti-Pure TITANIUM DIOXIDE. DETERMINATION OF BROOKFIELD VISCOSITY, TiO 2 SLURRY METHOD: T WP

MONDOATHENS BASKETBALL SET (Reference PK110)

Eco Green Labs India Pvt. Ltd., Ground Floor, Plot No. 100, New Sidco Industrial Estate, Srinagar, Hosur, Tamil Nadu

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Longitudinal and lateral dynamics

NI USB-5681 RF Power Meter Specifications

HAZET Torque Wrenches

Harmonic oscillations of spiral springs Springs linked in parallel and in series

Understand, Maintain and Service Your Torque Wrench

Perthometer. Surface Texture Parameters New Standards DIN EN ISO / ASME

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM

Technical Datasheet Scalar Network Analyzer Model MHz to 40 GHz

Uncertainty of Force Measurements

TELESCOPE Motorized Lighting Suspension

n Measuring range 0... ± 0,02 Nm to 0... ± 1000 Nm n Low linearity deviation of ± 0.05 % F.S. n Intelligent operating state indicator

How To Test A Titrette And Digital Burette

EXPERIMENTAL ERROR AND DATA ANALYSIS

So You Want to Think Lean? Jürgen Hilbinger and Martin Schreiber

Rotational Motion: Moment of Inertia

ABB i-bus EIB / KNX Analogue Input AE/S 4.2

University Physics 226N/231N Old Dominion University. Getting Loopy and Friction

D.C. Motors. Products and specifications subject to change without notice.

BoWex SD, SD1 and SD2 Operating-/Assembly Instructions

Practice Exam Three Solutions

EFC Frequency converters

TENSIONMETER ** code 183_LNRSTW **

Calibration Certificate

NITP 6.1 to 6.4 National Instrument Test Procedures for Non-automatic Weighing Instruments

Instructions for Use Anti-theft alarm DWA 6

The simulation of machine tools can be divided into two stages. In the first stage the mechanical behavior of a machine tool is simulated with FEM

Traditional Drawing Tools

Clutch Systems. Innovative Solutions for All Roads

SCS Directory Accreditation number: SCS 0115

ROTEX SD shiftable jaw coupling

MInIMAT-ec. flexible data recording highly accurate

B.TECH. (AEROSPACE ENGINEERING) PROGRAMME (BTAE) Term-End Examination December, 2011 BAS-010 : MACHINE DESIGN

Transcription:

Measurement uncertainty in torque calibration Luca G. Bochese, ASQ CQE, CRE (Boch - Milano), info@boch.net, www.boch.net Dr. A. Bochese ( Feanor - Tallinn), info@feanor.com, www.feanor.com Keywords: Torque, calibration, uncertainty, torque tools, torque testers Contents 1. Introduction - Abstract 2. Standard 3. Typical torque curve 4. Actual calibration procedure 5. Considerations on traditional calibration procedure 6. New torque calibration device 7. Calibration of the device 8. Calibration of reference torque tools 9. Calibration certificate 10. Measurement uncertainty 11. Acknowledgements 12. References 13. Acronyms 14. Slides 1. Introduction - Abstract Calibration of torque measurement instruments (mainly torque wrenches and torque testers), even if widely used in the automotive, aviation and building industry, has received much less attention than other types of instruments (i.e. length measurement instruments). Very few laboratories offer calibrations services for torque measurement instruments, compared to length measurement, pressure, etc. All car manufacturers have developed internal measurement systems to measure torque (i.e. braking systems, frictions, etc), but out of these laboratories the specific situation for torque calibration is still unclear. Since torque is given in Nm, the usual way to calibrate such instruments is to use a torque balance, set the torque tool at the target value (i.e. 40 Nm) and apply reference weights (N) until the requested KP (click point) is reached. The difference between the nominal KP (on the wrench) and the true KP (measured by the sensor on the torque balance) should be less than 4% or 6%, depending from tool type. This presentation focus on calibration of torque measurement and the comparison between this traditional method (torque balance) and a new calibration device, also used at the PTB Braunschweig - Germany. Experience shows that the new system applicable to calibration of torque tools is more effective than traditional torque balance systems, since it tests them in the same operating condition (dynamic) and with very small almost continuous increments of forces while torque balance operates almost statically and need discrete mass

increments. The new systems is also much faster (requires about 15 minutes for calibration according to ISO 6789) than the traditional torque balance method. The new draft of ISO/CD 6789 (not an international standard yet) is also discussed and compared with the previous edition of ISO 6789. A formula for the determination of measurement uncertainty of a new concept of calibration device is given and briefly discussed. Some experience of a calibration laboratory in Milano (Italy) will give useful hints at the laboratory level calibration of torque tools (both setting and indicating torque tools) and testers. Slides show that torque tools from different manufacturers do not operate in a similar way, some of them shows a distinct KP (clicking point) while other have a smooth curve with local peaks, in this situation the experience of the laboratory personnel is still important to find out where the correct KP is and some calibration parameters accordingly. This situation is not optimal, since it implies an analysis, and therefore an influence, of the laboratory personnel and risks to affect the independence of the calibration laboratory. Anyway, tolerance at 4% is a close tolerance and will force manufacturers to improve the design and costruction of torque tools components (expecially springs) in order to satisfy these requirements. 2. Standard The standard documents which refer to calibration of these instruments are, respectively ISO 6789:1992 and DIN 51309, along with DKD guidelines R 3-7 and R 3-8. ISO 6789:1992 was prepared by ISO TC/29. This Committee is constituted by representatives of torque manufacturers and representatives of research and calibration institutes. ISO/CD 6789 is a Committee draft and cancels and replaces the second edition ISO 6789:1992, of which it constitutes a technical revision (slide 10). New clauses have been added: 3 Terms and definitions 4.1 Design conformance testing 4.2 Quality conformance testing 4.3 Recalibration Figures 1 and 2 have been revised, figures 3 have been added, the standard has been editorially revised. The document show a strong influence of DIN 51309. In particular following changes have been introduced with the ISO/CD 6789: 1. the change of tolerance (from 4% to 6% tolerance) also for instruments of type II and classes A and B up to 10 Nm (slide 15). This tolerance was, in the previous edition, accepted only for type C, D and E instruments. The internal friction has a greater influence with small torque values, and thus the original tolerance of 4% has been enlarged, as requested by torque wrench manufacturers.

2. A period of validity of a calibration certificate has been introduced (12 months as a maximum). This period should be reduced when the tool is frequently used. 3. Chapter nr. 3 with common terms and definitions has been added. A multilanguage section would be very useful in future editions. 3. Typical torque curve Slides 2, 3,4,7 and 8 show different shapes of torque vs. time curves for different brands of type II wrenches. The KP (click point) can sometimes be determined very easily (slide 3 and 4), but sometimes the curve has local peaks (slide 2), is too smooth (slide 8), or shows friction effects (slide 7). In this situation the experience of the laboratory personnel is still important to find out where the correct KP is and set certain parameters on the calibration device accordingly. This parameters are be the rotation speed, the trigger used to select a KP (the requested fall in % of the torque value after the peak, in order to classify it as a true peak, i.e. 80%, 85%, 96%, 99%), the angle of rotation, the point where speed reduces (usually at 80% of the torque target value. This situation is not optimal, since it implies an analysis, and therefore an influence, of the laboratory personnel and risks to affect the independence of the calibration laboratory. By looking at the torque manufacturer the laboratory personnel can set following parameters: rotation speed and KP sensitivity. Unfortunately, given the different behaviours of torque wrench brands, it cannot be avoided. All these curves represent a single measurement acquisition at a certain torque, i.e. 100 Nm. ISO 6789 requires a series of 5 measurement values for each step at 20, 60 and 100%. 4. Actual calibration procedure A) Torque balance Having torque dimensions of force X distance, the traditional calibrating device is constituted by a torque balance (slide 12) and a set of primary weight reference. With this system, the necessary force is obtained using a reference mass, keeping in consideration the local gravity constant g, in both directions (right and left). The torque is transmitted through a double arm, mounted on an aerostatic radial and axial bearing. This guarantees a transmission of radial and axial forces with a minimum level of friction. A combination of motor and a transmission with high sensitivity is used to compensate the rotation of the reference. The information for the correction is obtained through non-contact sensors positioned on the arm. Possible force increments with this system are: 20 Nm, 50 Nm, 100 Nm, 200 Nm, 500 Nm and 1000 Nm, each one can be divided in 10 discrete steps. With this system a typical measurement uncertainty of : 0.01% from 5 Nm to 1000 Nm 0.02% from 1 Nm to 5 Nm

can be reached. Torque sensors, torque testers, reference torque tools and torque tools are calibrated with this torque balance. Calibration with torque balance needs skilled laboratory personnel, requires time and, since we switch from mass to force, can be influenced by vibrations and by changes of the gravitation constant g. With this system the axis is horizontal and the results contains also the contribution of the weight of the wrench grip (lever effect). Systems with vertical axis are being also developed, to reduce this problem. B) Torque tester After calibrating the sensor or the reference torque with the promary torque balance, we can transfer this result to a torque tester (slide 16). These instruments are often used as primary reference in companies or calibration rooms. The torque tester should then be mounted on a stable support in Aluminium (slide 13) and used for calibration of torque tools. A transmission is needed to reduce the influence of the operator (position of the hand on the grip, local increase and decrease of speed, expecially close to the click point. Experience shows that measurement values change slightly from user to user when such a transmission is not used. C) Torque wrench First of all we should distinguish 2 types of wrenches: 1. setting torque tools (indicated as type II wrenches - slide 15) 2. indicating torque tools (indicated as type I torque tools slide 14). Type II -setting torque tools (Click torque tools) The release point must be determined at intervals of 20%, 60% and 100% of the maximal nominal torque. For each measurement we take 5 measurement, if any one of them (even a single one) is out of the tolerance band of ±4% the wrench is considered out of tolerance. Experience shows that points have an asymptotic distribution like this, with the first value higher than following points. This is mainly due to an adaptation of the spring in the guide and an hysteresis effect on the first clicks.

Figure 1: Typical distribution of sequential torque values This behaviour cannot be considered a random effect, it is due to internal adaptation of the torque wrench. This phenomenon is typical with almost all torque tools even if the tool has been, as requested by the ISO 6789 standard, operated for at least 5 times at full scale a few seconds before the start of the calibration procedure at 20% of full scale torque value. It is therefore not meaningful to calculate a mean value and to compare this value with the 4% tolerance limits. For this reason ISO 6789 requests to use at least 5 times the torque tool at the maximum nominal value just before starting the calibration procedure, so that first clicks can be considered outliers and would not affect the automatic calibration procedure. Type I - Indicating torque tools (Dial torque wrenches) The calibration intervals are the same as type II wrenches. The calibration cannot be automatically performed, since the operator has to look at the indicator, write manually the value and the systems checks the difference nominal-true torque value. Considerations on both types of wrenches As seen before, tolerance limits for small torque tools up to 25 Nm have recently been reduced to ±6% (the same level as torque screwdrivers (see types D, E, F in slide 16 and types D, E in slide14) in the previous edition of the ISO 6789).

5. Considerations on traditional calibration procedure This system works fine with dial wrenches, but it is not good for KP wrenches, because we can give only finite increments and cannot eventually decide exactly find out where the KP point is. That is the most important issue: KP wrenches should be calibrated dynamically, applying a force increasing with continuity, until the release point (KP) is reached and recognised by the sensor. This is the key factor who inspired the construction of a new torque calibration device. 6. New torque calibration device This device (slide 11) is manufactured by a company in Germany, who also manufactures equipment directly for the PTB. The device is equipped with 2 different sensors and motors, one for small torque values up to 30 Nm and another one for torque values up to 1000 Nm. Torque axis is vertical, so that any influence on wrench own weight is excluded. The wrench is rotated at higher speed until 80% of target torque value or until contact between grip and device, torque is applied until KP is found. 200 measurement are taken every second and can be optionally reported on a graph (slides 2,3,4). The torque wrench to be calibrated has to be stabilised at a temperature of 20% for at least 12 hours, air humidity should be also maintained at 50% ± 5%. It is also very important that the torque wrench spring is kept at the minimum level or less ( spring uncharged ) for all that period. In effect, it is always a good practice to bring the indicator at the minimum level after the wrench has been used and leave the spring uncharged when not in use. All other wrench components (i.e. adapters) must be available. The calibration of type II tools is fully automatic, while type I require laboratory personnel. 7. Calibration of the new device This device is calibrated using 2 reference torque tools of the PTB (slide 13). In our laboratory we actually use these reference wrenches to calibrate not only the device, but also to calibrate torque testers (both digital and analogic slide 16). The calibration is performed dividing the range in at least 5 equidistant step (i.e. up to 100 Nm we shall have steps at 20, 40, 60, 80 and 100 Nm) and performing 4 series of measurements. See slide 20 for a matrix with measurement values and a graphical view of these 4 series. For each series one measurement is taken at every step, in this sequence: 1. going up, maximum arm length 2. going down, maximum arm length 3. going up, maximum arm length 4. going up, minimum arm length The different contributions to the measurement uncertainty are determined as follows: difference between 1 and 2 = contribution of hysteresis (as with i.e. dial gauges)

difference between 1 and 3 = contribution of repeatability with same arm distance difference between 1 and 4 = contribution of repeatability with reduced arm distance Figure 2. Graphical overview of the calibration procedure 8. Calibration of reference torque tools The two references torque tools used in our laboratory (up to 100 Nm and from 10 to 1000 Nm) are calibrated every year by using a torque balance (slide 12) with reference weights (PTB primary system) directly at the PTB. The procedure is the same as the one described in the previous paragraph. 9. Calibration certificate Which information should be given in a calibration certificate? Standard calibration certificates have a standard format for the front page like this (slide 9). In the front page should always appear the traceability to national standard. The back page usually contains the measurement values and calculations of parameters like: measurement uncertainty, mean values and standard deviation, Cp and Cpk (or Cm and Cmk) derived from SPC techniques (slide 5). A particular consideration should be given to some calculation which can be found in some certificates like this (slide 18): 20 or 25 consecutive measurements are taken and parameters like mean value, standard deviation are calculated. In our opinion this information is useful only in calibration of torque testers, which are given with a tolerance of ±1%, and could be avoided when calibrating type II torque tools. Unfortunately norms do not indicate a standard format for the technical information of a torque calibration certificate, apart from the front page (slide 9). Some work still has to be done in order to reach a common standard like i.e. in calibration of dial gauges or block gauges. 10. Measurement uncertainty

The formula used to determine measurement uncertainty is shown in slide 19. This discussion refers to the calibration of a digital torque tester (DIN ISO 51309) (slide 16) using a PTB reference torque wrench (slide 13). The expanded uncertainty is given by U, where U MG is the measurement uncertainty of the measuring instrument and U TN is the measurement uncertainty of the reference torque wrench (given by PTB): U = ( U MG) + ( UTN) 2 2 1 2 1 2 1 2 U MG = f q + k f + 0 b + l b' + 12 8 8 1 12 r 100 Mi 2 + 1 12 h 2 Where: U TN uncertainty of the transfer normal (reference wrench) U MG uncertainty of the measurement instrument (calibration equipment) fq Contribution of the digital display (error is known) (7) DIN 51309 f0 Contribution of initial zero position (error is unknown) (4) DIN 51309 bl Contribution of the distance (error is unknown) DKD R 3-8 b Repeatability (error is unknown) DKD R 3-8 r Resolution of the display DKD R 3-8 Mi Nominal torque DKD R 3-8 h Contributions of hysteresis (6) DKD R 3-8 k Coverage factor, standard is k=2 (95,45%) Standard U TN values are as follows: Torque M Measurement uncertainty U TN M 5 Nm 0,2 % 5 Nm < M 150 Nm 0,1 % 150 Nm < M 1000 Nm 0,05 % In particular following contributions are taken in account for the determination of measurement uncertainty: hysteresis (h) (only with both right and left direction) display (fq) it can be corrected and therefore is does not enter in the square first set to zero (f 0 ) torque reference (U TN ) arm distance (bl) repeatability (b ) digital display resolution (r)

Remarks Environmental factors as temperature and air humidity are NOT taken into account, given their minimal influence on measurement results. All electrical equipment must have been switched on at least 15 minutes before starting the calibration procedure Distribution A rectangular distribution is assumed for all uncertainty contributes. This explains the 1/12 coefficient in the given formula (slide 19). The conventional coverage factor is k = 2. This value is reported on the calibration certificate, together with the indication that the given uncertainty refers to expanded uncertainty. 11. Acknowledgements Dieter Kenzler DIN Salvatore Merlo SCHATZ Remscheid Deutsches Institut für Normung e.v. IMGC CNR Torino 12. References 1. ISO 6789:1992 2. ISO/CD 6789 (Draft) Assembly tools for screws and nuts Requirements and test methods for design conformance testing, quality conformance testing and recalibration procedure 3. DIN 51309 4. DKD Richtlinien R 3-7 and R 3-8 5. ISO GUM - Guide for determination of measurement uncertainty - 1995 13. Acronyms DKD IMGC PTB SIT Deutscher Kalibrierdienst Istituto Metrologico Gustavo Colonnetti Physikalisch Technische Bundenanstalt Sistema Italiano di taratura

14. Slides 1. Example of 3 different curves torque/time 2. Example of a curve with local torque peaks 3. Example of a good torque curve (KP easily determined) 4. Example of a good torque curve 5. Example of a calibration certificate (back page) 6. Typical distribution graph on a single torque wrench 7. Example of a torque curve with friction 8. Example of torque curve with indistinct KP and local peak 9. Example of front page of a calibration certificate 10. ISO 6789 draft (english) 11. Calibrating device (new concept) 12. Typical torque balance 13. Reference torque wrench 14. Typical torque tools class I (dial) 15. Typical torque tools class II (click) 16. Typical torque tester 17. Calibration laboratory 18. Alternative certificate (measurement on 25 consecutive values) 19. Formula for torque measurement uncertainty 20. Calibration certificate (page nr.3) of a digital torque tester