STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H)

Similar documents
Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter

Fundamentals of Mass Flow Control

DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS

SDT 2960 Simultaneous DSC-TGA including DTA capabilities

Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System

QGA Quantitative Gas Analyser

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

TA INSTRUMENTS DIFFERENTIAL SCANNING CALORIMETER (DSC) Insert Nickname Here. Operating Instructions

Table of content. L81/RITA high speed Thermo Balance. Quattro Dilatometer. L75/1250/B/S Macro Dilatometer. New air cooled furnace program

G8 GALILEO. Innovation with Integrity. High-End Melt-extraction Analyzer. Inert Gas Method

Calibration of Dallas sensors

FP628. Nitrogen / Protein Determination by Combustion. Rev 1-6/8/11

Hands on an infrared gas analyzer (IRGA) to measure photosynthesis

Physics 41, Winter 1998 Lab 1 - The Current Balance. Theory

Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm

1 DESCRIPTION OF THE APPLIANCE

23 The Thermal Conductivity Detector

User checks and maintenance of laboratory balances

Phase Transitions and Differential Scanning Calorimetry

MILESTONE H E L P I N G C H E M I S T S

Differential Scanning Calorimetry DSC

TGA/SDTA851 e Module. Thermal Analysis. Technical data Quality Service

DSC Differential Scanning Calorimeter

SECTION 5 COMMERCIAL REFRIGERATION UNIT 22 CONDENSERS

Technical Specification for Model RSK2503 (Z)-06A IR Belt Solar Cell Firing Furnace

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors

Small Animal Respiratory Exchange Ratio (RER)

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory *

The NEW Advanced Range of Portable Temperature Calibrators PROVISIONAL. Portable Dry Blocks, Liquid Baths and Multi-functional Heat Sources

DMA-1 Direct Mercury Analyzer

Procon Engineering. Technical Document PELR TERMS and DEFINITIONS

AUTOCLAVE CORROSION INHIBITOR EVALUATION

Mass Flow Meters & Controllers

Determination of the enthalpy of combustion using a bomb calorimeter TEC

Physical Properties of a Pure Substance, Water

Typical Aluminum GFM Mass Flow Meter

Effect of Magnesium Oxide Content on Final Slag Fluidity of Blast Furnace

Description of Thermal Oxidizers

HEATING AND COOLING SYSTEMS THAT FIT COMFORTABLY WITHIN YOUR BUDGET.

APPLICATION OF DENSITOMETERS TO LIQUID MEASUREMENT Class # 2010

Control Device Requirements Charts For Oil and Gas Handling and Production Facilities

T5 LongLast. GE Lighting DATA SHEET

Folding Bath Screen Instruction Manual

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Chapter 4: Newton s Laws: Explaining Motion

Effect of Magnesium and Aluminum Oxides on Fluidity of Final Blast Furnace Slag and Its Application

DIAMOND Retractable Rodding Robot Model SPRAYROD-R

Bourns Resistive Products

Infrared Thermometer Calibration A Complete Guide

COUNTERBALANCE VALVES

FMA 3100/3100ST/3300/3300ST Series Thermal Mass Flow Sensors and Meters

DENSITY MEASURING SYSTEMS

CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA ABSTRACT

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Operator Quick Guide EC SENSOR

Environment-Laboratory Ambient Conditions

ASPE Seminar Water Heater Piping Strategies. Presented by Joe Kalarickal, EIT Regional Sales Manager AERCO International

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article

Instructions Manual. Electromagnetic sensor Series FLOMID FX. instrumentation for fluids. R-MI-FlomidFX Rev.: 0 English version

Characterization of Electronic Materials Using Thermal Analysis

M&V of Hot Water Boiler Plant. Boban Ratkovich, P. Eng, CEM, BESA, LEED AP President CES Engineering Ltd

Determination of the enthalpy of combustion using a bomb calorimeter TEC. Safety precautions

Thermal diffusivity and conductivity - an introduction to theory and practice

G4 ICARUS HF. Innovation with Integrity. Simultaneous Determination of Carbon and Sulfur in Solids. Combustion Method

RGC-IR Remote Gas Calibrator for IR400

Multiple Choice For questions 1-10, circle only one answer.

Geomembrane Manufacturing Quality Assurance Manual

Estimating Weighing Uncertainty From Balance Data Sheet Specifications

Thermal Flow Sensor Die

Manual piston sampler...m 11

WIRE, TERMINAL AND CONNECTOR REPAIR CONDUCTORS

Load Cell Accuracy in Relation to the Conditions of Use

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

To measure the solubility of a salt in water over a range of temperatures and to construct a graph representing the salt solubility.

RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE

450TOC Analyzer. Portable, Real-time, Continuous Total Organic Carbon Measurement. 450TOC Analyzer

Curved Quadrant Instruction Manual

Influence of Belt Furnace on Glass-to-Metal Seal Process

Unit 3: States of Matter Practice Exam

Guidelines for Choosing and Installing Radar in Stilling Wells and Bypass Chambers

ME Heat Transfer Laboratory. Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS

Sensors / Modules / Monitors. Breath Gas Sensing and Monitoring. For life supporting systems in intensive, emergency and ambulant care

PREPARATION. INTERNAL CALIBRATION (No external weight needed.) SPAN CALIBRATION

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration

Reliability Accuracy Performance

Technique of Monitoring Dioxins in Flue Gas from MSW Incinerators using Dioxin Precursor Analyzer

General Purpose Thermal Mass Flowmeter Model 4140/4143

METHOD EXPLOSIVE ATMOSPHERE

An Overview of Calibration Methods and Procedures for Process and Inventory Weigh Systems

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

R=Required by Lab S=May be subcontracted IEC SYSTEM FOR CONFORMITY TESTING AND CERTIFICATION OF ELECTRICAL EQUIPMENT COMMITTEE OF TESTING LABORATORIES

Instruction Manual FL

Controller Automation, Model II+

Sensor Integration on a W-CVD Cluster Tool for Real-Time Process Monitoring and Control

EuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases

Temperature Calibration; Depths of Immersion

Transcription:

STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H) by Joseph F. Greenbeck, Jr. And Courtney W. Taylor INTRODUCTION The following procedure is to be used for the maintenance and calibration of the TGA-50(H) and TGA-51(H). Because the TGA measures the weight loss of materials by thermal separation, some of the evolved material can be deposited onto the active parts of the weighing assembly (e.g. quartz rod, hanging pan, sample cell). This can cause erroneous readings in subsequent analyses by showing excessive or exaggerated weight losses which are not indicative of those particular samples. The following materials and supplies will be needed for cleaning and calibration: 1. A supply of Oxygen (O 2 ) or air. 2. A supply of 90/10 mixture of N 2 /H 2 (Optional). 3. Certified Curie Point standards (SRM-761) 4. Calibrated mg weight set (P/N 201-52724-01) 5. TGA calibration magnet (P/N 220-96401-00) CLEANING Method 1 (Standard) This cleaning method requires O 2 or air to reduce the built up organic residues in the furnace and on the active parts [e.g. quartz rod, Platinum (Pt) cell holder, etc.] of the balance assembly. This simple procedure is as follows: 1. Connect oxidative purge gas to the rear of the instrument. 2. Set flow rate to 50 ml/min. 3. Set heating rate to 20 C/min. 4. Set hold temp. to 700 C/min. 5. Set hold time to 120 min.

6. Set PID's to 10, 10, 10. 7. Start the run (with empty Pt cell). 8. When the run is complete the inside of the furnace and the purge inlet tube should have a clean, white appearance. The quartz rod and Pt cell and holder should also be clean. Method 2 (Optional) This optional method requires a mixture of 90% N 2 and 10% H 2 and can be used to clean the TGA if the residue is mostly inorganic, or if the O 2 method does not sufficiently clean the furnace and active weighing parts. The same instrument parameters should be used as in Method 1. CALIBRATION Since the TGA measures the change in weight of a sample over a given temperature range, the instrument must be calibrated for both temperature and weight signals. Weight calibration is not necessary if the TGA analysis is to be performed in percent weight loss only. The temperature will be calibrated using magnetic transition or Curie point standards. These standard reference materials (SRM) are available from the National Institute of Standards and Technology (N.I.S.T.) as SRM-761. The weight signal will be calibrated using a mg weight kit with certifications traceable to N.I.S.T. If traceability is of no concern then the weight kit supplied with the TGA can be used. Temperature Calibration Temperature calibration of the TGA-50 is relatively easy to perform but sample preparation is a little bit more involved. Once the sample is made, however, it can be reused. The materials needed are a TGA calibration magnet (P/N 220-96401-00) and the SRM-761 Curie point standard kit available from N.I.S.T. This kit contains one metal and four alloy standards which are magnetically permeable. When subjected to a magnetic field the sample's apparent weight changes due to a alteration in the sample s structural order brought on by a thermal change. In this case the transition will appear as a weight loss as the temperature increases, and a weight gain as the temperature decreases. When the 1st derivative curve is displayed, the weight loss and weight gain will appear as negative and positive peaks respectively. This is where the Curie point result will be taken. Sample Configuration To prepare for the test, dismount the Pt cell and pan from the quartz rod. Mount the TGA calibration magnet on the thermocouple by slightly spreading apart the leads and "cradling" the magnet so that the flat side is horizontal to the

axis of the quartz rod. Now hang the Curie point standard from the quartz rod and position it as close to the magnet as possible without touching it (about 1mm). The magnet should appear to draw the standard down towards it. Close the furnace and allow the weight signal to settle out. Zero the weight signal and set the temperature calibration constants (in this case Permanorm 3 and Mumetal) to Null: 1. Temp Gain = 1.000 Temp Offset = 0.0 2. T1 Expected = 259.1 T1 Measured = 259.1 3. T2 Expected = 381.6 T2 Measured = 381.6 4. SIG Gain = 1.000 5. SG Expected = 10 SG Measured = 10 Now set the run parameters to the following: 1. Heating Rate: 10 C/min. 2. Hold Temp: 100 C above Curie pt. 3. Hold Time: 0 min. 4. PID's: 10, 10, 10. 5. ATM: N 2 at 30-40 ml/min. 6. Sampling Interval: 1.0 sec. When the run is over, display the 1st derivative curve (smooth it if necessary), scale it and determine the peak temperature of the derivative curve for the apparent weight loss. Once two or more Curie point standards have been run the data can be entered into the TGA calibration function (FUNC 7) using the standard method: 1. Enter Calibration Function (Press FUNC, go to 7, and press ENT). 2. Press ENT 3 times. 3. Enter Measured T1 temp (from sample run) using the or arrow keys and press ENT. 4. Enter Measured T2 temp (from sample run) using the or arrow keys and press ENT. 5. Re-analyze Permanorm 3 and Mumetal standards. Temp should be within ± 5.0 C. Weight Calibration Performing the weight calibration for the TGA-50(H) or TGA-51(H) is quick and very easy. Again, this calibration does not need to be performed if weight losses will only be analyzed in percent. Use the mg weight set supplied with the TGA or, if necessary, any mg weight set traceable to N.I.S.T. Although this calibration uses one weight measurement, it is really a two point measurement since the reference point (or first point) is

Zero. Keep in mind that the calibration weight range should be just above the actual sample weight. That is, if the calibration weight used is 50mg then the sample weight should not be more than 50mg but just under. This should be performed at ambient temperature using the following method: 1. Enter the Calibration Function and set the SIG Gain to 1.000. 2. Set the SG Expected and SG Measured to the desired mg weight. 3. Close the furnace (empty cell on hanging pan), allow to settle. 4. Zero out the weight signal. 5. Lower the furnace and place calibration weight on pan assembly. 6. Close furnace and allow the weight to settle (about 1 min.). 7. Record weight reading and enter it into the SG Measured under Calibration Function. 8. Weight calibration is complete. BASELINE Since Thermogravimetric Analyzers measure the weight loss of materials as a function of temperature and time, a good baseline is necessary for optimum weight loss sensitivity. The ideal TGA baseline will be flat and parallel to the x-axis with no buoyancy effect at the start. This is never the case with low sample mass-high sensitivity instruments, as there are many things which can affect the baseline. The main de-stabilizing effect is buoyancy. Buoyancy is seen as the false increase in the weight signal at the start of a run and is caused by atmospheric disturbance. Before a run is started the sample and the atmosphere inside the furnace are in a stable state. The temperature is steady, and the purge atmosphere is slowly introduced into the furnace chamber. When the run is started, however, the heat energy being pumped into the furnace begins to warm the stable internal atmosphere. Since hot air rises, so also does the purge gas which begins to circulate. This tends to push down slightly on the sample pan assembly which the instrument sees as a weight gain. De-stabilizing baseline effects can be minimized several ways. Using large sample masses will help reduce the buoyancy effect simply by being more massive and resistant to any atmospheric "push." Lower temperature gradients will also reduce atmospheric disturbance. Finally, slightly higher purge rates can also act as an opposing force to thermally induced atmospheric currents.

Finally, most of the unstable baseline effects can be eliminated from sample data by performing a baseline subtraction. This is done by running a blank or empty sample cell under the same program and atmospheric conditions as the sample and then subtracting the baseline file from the raw sample data file. This procedure is covered on pp. 3-97/98 in the TA-50WSI users manual. One important note about TGA operation is that the instrument must always be run with a purge gas flowing into the back of the instrument. The reason for this is that it eliminates the chance of any evolved materials drifting up into the detector housing, which can damage the detector. As the purge gas enters the rear of the instrument it is passed through the regulator and then into the detector housing. From there it is directed down the purge inlet tube and into the furnace area. This keeps positive pressure within the detector housing and eliminates back drafting of potentially damaging evolved materials. CONCLUSION The TGA-50(H) and TGA-51(H) are both highly sensitive instruments which require proper care and attention. The operation is simple and, with a little time and patience, can easily be temperature and weight calibrated. Obviously a clean instrument will provide the user with quality data for a longer time than one that is not properly maintained. The frequency of cleaning, however, will vary from user to user and depend on the type of samples tested. The general rule of thumb is to clean and calibrate the instrument once a month.