RaySafe Solo. Specifications

Similar documents
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Barracuda & QABrowser

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

USER GUIDE. LED Light Meter. Model LT45

iva Cable & Antenna Analyzer

iva Cable & Antenna Analyzer

CLV440/442 Bar Code Scanner

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

DRTS 33. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters

DSM155 Digital LED Light Meter User s Manual DSM155

Precision Lightmeters Made in Germany

Nederland België / Belgique

User Manual.

Series 7500 Torque Sensor

Series 6000 Torque measured metal bellow coupling

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Technical Data. back-up time in minutes 100% load % load

How To Power A Schen

Multi-Function Ground

Material: Weight: Bearing Life: Shaft Speed: Starting Torque: Mass Moment of Inertia: Shaft Loads:

Description. Dimensions. Features. precision works better

Light Meter (Foot Candles)

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Digital input modules

SENSORS. Fiber Optic Sensors - S70. Advanced fiber optic amplifiers for high speed and low contrast applications

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Digital Breast Tomosynthesis QC Requirements

Bar Code Reader Series BCL 20 A flexible reader for a variety of applications

MS9500 SERIES VOYAGER PRODUCT GUIDE. Single Line, Hand-Held, 1D, High-Density 1D, PDF-417 2D and Wireless Laser Bar Code Scanners

SCS Directory Accreditation number: SCS 0115

SpectraTec II. Polarized Multi-Laser Source BLUE SKY RESEARCH WAVELENGTHS. The SpectraTec II

TX OPEN RS232/485 module

Quick Connect. quick - simple - efficient.

L-LAS-TB-CL serie. laser light curtains for inline measuring tasks

Wahl C50 Calibrator FEATURES

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

ABB i-bus EIB / KNX Analogue Input AE/S 4.2

LED red (-): Measuring value < lower tolerance threshold LED red (+): Measuring value > upper tolerance threshold. Page 1/6

Performance testing for Precision 500D Classical R/F System

MAVO-MONITOR / MAVO-SPOT Instrument Set for Contact or Distant Measurements of Luminances

Pocket Tach 99 (PT99) Non-Contact Tachometer

Agilent AEDB-9140 Series Three Channel Optical Incremental Encoder Modules with Codewheel, 100 CPR to 500 CPR Data Sheet

Three Channel Optical Incremental Encoder Modules Technical Data

Nardalert S3 Non-Ionizing Radiation Monitor

NI USB-5681 RF Power Meter Specifications

Telephone- and leased line modem for industrial applications TD-36

4-Channel Thermometer / Datalogger

R&S ZNC Vector Network Analyzer Specifications

USB-500/600 Series Low-Cost Data Loggers and Accessories

1 Technical Description Lokal-200PC

Measuring electric and magnetic fields

FEATURES. ! Sealing to IP68 (fully waterproof) Infra-red optical navigation technology

USER'S GUIDE NUSCAN 2100 Series

Models HP HDMI to VGA Adapter HP EliteFolio Ethernet & VGA Adapter New! HP HDMI to DVI Adapter New! HP DisplayPort to HDMI 1.4 Adapter New!

Schedule of Accreditation issued by United Kingdom Accreditation Service High Street, Feltham, Middlesex, TW13 4UN, UK

7.0, 4 Wire Touch Panel

VeHiCle diagnostics the autocom way

DS 600. A contact free flux gate based current measurement sensor 600A rms

RISH EM 3490 DS Dual Source Energy Meter RISH EM 3490 DS. Application : Product Features:

Barcode Reader Series BCL 31, BCL 32, BCL 34

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

HT580 User Reference Guide. Version 1.0

Operating Manual UMB Analogue/Digital Converter ANACON Order Number: 8160.UANA

Datalogger DL/N 70 Version:

PD30ETB20xxIS. Photoelectrics, Background Suppression reflective with IR light. Main features. Description

Waveguide Access Point WGA631. Product Guide

T Substation Maintenance and Commissioning Test Equipment

Technical Data. FLIR T (incl. Wi-Fi)

Models HP Dock Connector to Ethernet & VGA Adapter New! HP EliteFolio Ethernet & VGA Adapter HP HDMI to DVI Adapter

LM10. Micro Laser Displacement Sensor. The LM10 makes laser sensors super easy to use! New circuitry lowers costs

APSYN420A/B Specification GHz Low Phase Noise Synthesizer

Electromagnetic Compatibility Test Report Test results of Floww equipment

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Perle Gigabit Fiber to Fiber Media Converters

Multifunction HVAC and IAQ Meter

Hand-held thermometer Model CTH7000

SENSORS. Miniature Sensors - S3Z. Advanced line of miniature Asian style of photoelectric sensors mm background suppression

KS732 Family. Rev KS732. 1/3 CCD Color USB Camera

How To Use A Ds340 (Dsp)

R&S SMZ Frequency Multiplier Specifications

USB-500 Series Low-Cost Data Loggers and Accessories

Cross-beam scanning system to detect slim objects. 100 mm in

Product Data. Thermal Comfort Data Logger Type Including Application Software Type 7301

USB Smart Power Sensor

USB-500/600 Series Low-Cost Data Loggers and Accessories

Remote RadEye Product Family

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Features. Display. Measurements. Intelligent. Accuracy. Models. Installation DEIF A/S. Multi-instrument D

APC Matrix-UPS 5kVA 208V/240V In 120/208/240V Out

Dual Laser InfraRed (IR) Thermometer

The quadrature signals and the index pulse are accessed through five inch square pins located on 0.1 inch centers.

Perle Fast Fiber to Fiber. Media Converters

Laser diffuse reflection light scanner with background suppression. Dimensioned drawing

LED light engine / OLED LED linear / area. Umodule STARK QLE G3 lens CLASSIC umodule QLE

Transcription:

RaySafe Solo Specifications

RAYSAFE SOLO GENERAL EMC TESTED According to EN 61000-6-1:2007 and EN 61000-6-3:2007 EXPOSURE NEEDED One RESET Automatic TEMP. 15 35 C (59 95 F) DETECTOR CABLE LENGTH 2 and 10 m (6.5 and 33 ft) PATENT Germany DE69430268.6-08 UK 0758522 Japan 3449721 Sweden 9302909-8 France 075822 USA 5761270 SIZE BASE UNIT 28 x 74 x 142 mm (1.1 x 2.9 x 5.6 in) 250 g (9 oz) POWER OFF Never, or after 5, 20 or 60 min of inactivity POWER SOURCE Rechargeable 7.4 V Li-on battery BATTERY TIME 20 40 hours (depending on detector and if Bluetooth is used) READ OUT Three row alphanumerical backlit display with four digits numerical resolution RAYSAFE SOLO mas ma 0.2 2000 ma ma 1 % or ± 0.02 ma mas 0.05 9999 mas mas 1 % or ± 0.02 mas MAX LOAD < 200 ma continuously, 500 ma < 1 s, 1000 ma < 0.5 s REPRODUCIBILITY < 0.5 % OVER VOLTAGE 70 V PROTECTION PEAK RATE mas PER 0.5 % or 0.2 ms > 8 ma 1/6 120 pulses/s 0.001 2000 mas/pulse

RAYSAFE SOLO R/F 12 x 22 x 117 mm (0.5 x 0.9 x 4.6 in) DOSE (R/F LOW) 10 ngy 9999 Gy (1 µr 9999 R) 200 ngy/s (1.4 mr/min) 5 % or ± 10 ngy (1 µr) (40 150 kvp, HVL: 1.5 14 mm Al (1), DOSE (R/F HIGH) 10 µgy 9999 Gy (1 mr 9999 R) 100 µgy/s (0.7 R/min) 5 % or ± 10 µgy (1 mr) (40 150 kvp, HVL: 1.5 14 mm Al (1), DOSE RATE (R/F LOW) 20 ngy/s 1 mgy/s (140 µr/min 7 R/min) MIN. PEAK 200 ngy/s (1.4 mr/min) 5 % or ± 10 ngy/s (70 µr/min) (40 150 kvp, HVL: 1.5 14 mm Al (1), DOSE RATE (R/F HIGH) 20 µgy/s 1000 mgy/s (2) (140 mr/min 7000 R/min) MIN. PEAK 100 µgy/s (0.7 R/min) 5 % or ± 10 µgy/s (70 mr/min) (40 150 kvp, HVL: 1.5 14 mm Al (1), kv/kvp 45 150 kv/kvp 2 % (for total filtrations from 2.5 mm Al up to 15 mm Al or equivalent, SENSITIVITY 0.4 ma, 40 kv, 40 cm (15.7 in), (R/F LOW) SENSITIVITY 0.8 ma, 70 kv, 50 cm (19.7 in), (R/F HIGH) 0.5 % or 0.2 ms PEAK > 3 µgy/s (R/F LOW) PEAK > 1 mgy/s (R/F HIGH) HVL (3) TOTAL FILTRATION (4) 1.0 14.0 mm Al 10 % or ± 0.2 mm Al (at signal levels above 1.5 35 mm Al (60 120 kv) 10% or ± 0.3 mm Al (at signal levels above (1) 45 mm Al added filtration at 145 kvp gives a HVL of ~13 mm Al. (2) 1000 mgy/s up to 70 kvp, 400 mgy/s at 100 kvp, 250 mgy/s at 140 kvp. (3) HVL is an option to RaySafe Solo R/F. (4) Total filtration is included in the HVL option.

RAYSAFE SOLO DOSE 12 x 22 x 117 mm (0.5 x 0.9 x 4.6 in) DOSE (R/F LOW) 10 ngy 9999 Gy (1 µr 9999 R) 200 ngy/s (1.4 mr/min) 5 % or ± 10 ngy (1 µr) DOSE (R/F HIGH) 10 µgy 9999 Gy (1 mr 9999 R) 100 µgy/s (0.7 R/min) 5 % or ± 10 µgy (1 mr) DOSE RATE (R/F LOW) 10 ngy/s 1 mgy/s (70 µr/min 7 R/min) MIN. PEAK 200 ngy/s (1.4 mr/min) 5 % or ± 10 ngy/s (70 µr/min) DOSE RATE (R/F HIGH) 20 µgy/s 1000 mgy/s (2) (140 mr/min 7000 R/min) MIN. PEAK 100 µgy/s (0.7 R/min) 5 % or ± 10 µgy/s (70 mr/min) 0.5 % or 0.2 ms HVL (3) PEAK (R/F LOW) PEAK (R/F HIGH) 1.0 14.0 mm Al 10 % or ± 0.2 mm Al (at signal levels above > 3 µgy/s > 1 mgy/s RAYSAFE SOLO RAD DOSE 12 x 22 x 117 mm (0.5 x 0.9 x 4.6 in) 10 µgy 9999 Gy (1 mr 9999 R) 100 µgy/s (0.7 R/min) 5 % or ± 10 µgy (1 mr) DOSE RATE 20 µgy/s 1000 mgy/s (2) (140 mr/min 7000 R/min) MIN. PEAK 100 µgy/s (0.7 R/min) 5 % or ± 10 µgy/s (70 mr/min) kv/kvp 45 150 kv/kvp 2 % (for total filtrations from 2.5 mm Al up to 15 mm Al or equivalent, SENSITIVITY 0.8 ma, 70 kv, 50 cm (19.7 in), 0.5 % or 0.2 ms PEAK > 1 mgy/s HVL (3) TOTAL FILTRATION (4) 1.0 14.0 mm Al 10 % or ± 0.2 mm Al (at signal levels above 1.5 35 mm Al (60 120 kv) 10% or ± 0.3 mm Al (at signal levels above (1) 45 mm Al added filtration at 145 kvp gives a HVL of ~13 mm Al. (2) 1000 mgy/s up to 70 kvp, 400 mgy/s at 100 kvp, 250 mgy/s at 140 kvp. (3) HVL is an option to RaySafe Solo dose and RaySafe Solo RAD. (4) Total filtration is included in the HVL option for RaySafe Solo RAD.

RAYSAFE SOLO DENT 12 x 22 x 117 mm (0.5 x 0.9 x 4.6 in) DOSE (DENT LOW) 10 ngy 9999 Gy (1 μr 9999 R) 1 μgy/s (7 mr/min) 5 % or ± 10 ngy (1 μr) (40 150 kvp, HVL: 1.5 14 mm Al (1), DOSE (DENT HIGH) 10 μgy 9999 Gy (1 mr 9999 R) 100 μgy/s (0.7 R/min) 5 % or ± 10 μgy (1 mr) (40 150 kvp, HVL: 1.5 14 mm Al (1), Active Compen sation) DOSE RATE (DENT LOW) 10 ngy/s 1 mgy/s (70 μr/min 7 R/min) MIN. PEAK 1 μgy/s (7 mr/min) 5 % or ± 10 ngy/s (70 μr/min) Active Compensation) DOSE RATE (DENT HIGH) 20 μgy/s 1000 mgy/s (2) (140 mr/min 7000 R/min) MIN. PEAK 100 μgy/s (0.7 R/min) 5 % or ± 10 μgy/s (70 mr/min) kv/kvp 45 150 kv/kvp 2 % (for total filtrations from 2.5 mm Al up to 1 mm Cu or equivalent, SENSITIVITY (DENT LOW) 0.4 ma, 40 kv, 40 cm (15.7 in), SENSITIVITY (DENT HIGH) 0.8 ma, 70 kv, 50 cm (19.7 in), PEAK (DENT LOW) PEAK (DENT HIGH) HVL (3) TOTAL FILTRATION (4) 0.5 % or 0.2 ms > 3 μgy/s > 1 mgy/s 1.0 14.0 mm Al 10 % or ± 0.2 mm Al (at signal levels above 1.5 35 mm Al (60 120 kv) 10% or ± 0.3 mm Al (at signal levels above (1) 45 mm Al added filtration at 145 kvp gives a HVL of ~13 mm Al. (2) 1000 mgy/s up to 70 kvp, 400 mgy/s at 100 kvp, 250 mgy/s at 140 kvp. (3) HVL is an option to RaySafe Solo DENT. (4) Total filtration is included in the HVL option.

RAYSAFE SOLO MAM DOSE BEAM QUALITITES 117 x 22 x 12 mm Mo/Mo, Mo/Al, Mo/Rh, Rh/Rh, Rh/Al, W/Rh, W/Ag 5 µgy 9999 Gy (0.5 mr 9999R) 10 µgy/s (70 mr/min) 5 % or ± 5 µgy (0.5 mr) (20 40 kv: Mo/Mo 22 49 kv: Mo/Al, Rh/Rh, Rh/Al 22 40 kv: Mo/Rh, W/Rh, W/Ag, 0 2.5 mm Al added filtration, Active Compensation) DOSE RATE 10 µgy/s 100 mgy/s (70 mr/min 700R/min) MIN. PEAK 10 µgy/s (70 mr/min) 5 % or ± 5 µgy/s (35 mr/min) (20 40 kv: Mo/Mo 22 49 kv: Mo/Al, Rh/Rh, Rh/Al 22 40 kv: Mo/Rh, W/Rh, W/Ag, 0 2.5 mm Al added filtration, Active Compensation) kv (only RaySafe Solo MAM, not RaySafe Solo MAM Dose) BEAM QUALITIES Mo/Mo, W/Rh 20 40 kv 2 % or ± 0.5 kv (no paddle) 2 % or ± 0.7 kv (paddle) (Active Compensation for inherent Mo filtration of 25 35 µm and for inherent Rh filtration of 50 60 µm. User selectable paddle compansation.) (1) SENSITIVITY 10 ma, 28 kv, 65 cm (25.6 in), no added filtration HVL (2) BEAM QUALITIES Mo/Mo, Mo/Al, Mo/Rh, Rh/Rh, Rh/Al, W/Rh, W/Ag 0.2 1.2 mm Al, depending on beam quality 5 % (for up to 2.5 mm Al added filtration to each beam quality) RAYSAFE SOLO CT 0.5 % or 0.2 ms (1) Definition: Paddle = 0.1 mm Al Note! Variation in paddle thickness and homogeneity may affect kv results. To achieve the most accurate result, kv measurements without paddle is recommended. (2) HVL is an option to RaySafe Solo MAM. 200 x 20 x 12 mm (7.9 x 0.8 x 0.5 in) SIZE Ø DETECTOR 7.5 mm (0.30 in) SIZE Ø PHANTOM 12.5 mm (0.49 in) ADAPTER EFFECTIVE LENGTH 100 mm (3.94 in) 50 g (1.75 oz) 10 µgy 9999 Gy (1 mr 9999 R) 20 µgy/s 100 mgy/s (140 mr/min 680 R/min) 5 % (at reference point RQA9; 120 kv, 42.5 mm Al) ENERGY DEPENDENCE < 5 % (at 80 kv to 150 kv; RQR and RQA qualities) RADIAL UNIFORMITY ± 2 % AXIAL UNIFORMITY ± 3 %, within rated length INFLUENCE OF RELATIVE < 0.3 % (for RH < 80 %) HUMIDITY IN 2 % TEMP. AND PRESSURE CORRECTION PRESSURE 80.0 106.0 kpa INTERNATIONAL Fulfills requirements in IEC 61674 STANDARD RaySafe Solo CT detector comes with a phantom adapter to fit a standard head and/or body phantom.

RAYSAFE SOLO LIGHT Cosine Response 1 0.8 0.6 0.4 0.2 RELATIVE AIR HUMIDITY 170 g (6 oz) < 80 % 3 % ILLUMINANT A DETECTOR MEMORY 30 measurements per sensor CLASSIFICATION Class B (according to DIN 5032, part 7) MAX. DEVIATION FROM 4 % (see figure Photopic Response) THE CIE CURVE FOR THE HUMAN EYE (V(Λ)) SIZE LIGHT DETECTOR 30 x 104 x 21 mm (1.2 x 4.1 x 0.83 in) LUMINANCE TUBE Ø = 29 mm (1.1 in) L = 84 mm (3.3 in) SHADOW RING Ø = 50 mm (2 in) LUMINANCE (AUTO) 0.05 50 000 cd/m 2 RESOLUTION 0.01 cd/m 2 LUMINANCE DETECTOR OPTICS ILLUMINANCE (AUTO) RESOLUTION MAX. DEVIATION FROM COSINE ANGULAR RESPONSE Ø 10 mm (0.4 in) measuring field. Contact measure ment focusing lens 1:1 0.05 50 000 lux 0.01 lux 1.7 % (see figure Cosine Response) 0-90 -60-30 0 30 60 90 Relative Response 1 0.8 0.6 0.4 0.2 Photopic Response 0 400 RAYSAFE SOLO PC KIT COMPATIBLE WITH Windows 10, Windows 8, Windows 8.1, Windows 7, Windows Vista, Windows XP, Windows 2000 FILE FORMAT XML COMMUNICATION RS-232 /USB or Blue tooth BLUETOOTH OPTION CONNECTOR OPERATING DISTANCE WAVEFORM BANDWIDTH (R/F LOW, DENT LOW) BANDWIDTH (R/F HIGH, DENT HIGH & RAD) BANDWIDTH (MAM) BANDWIDTH (mas) MEMORY DEPTH 450 500 550 600 650 700 750 Wavelength (nm) 9-pin D-SUB, pre-configured for communication with Xi View 100 m nominal (actual performance depends on environment and receiving Bluetooth module) 0.1 khz 2.5 khz 2.4 khz 1 khz 1 200 ms CIE Response RaySafe Response

2015.11 Unfors RaySafe 5200047-2.2 RAYSAFE DEFINITION The expanded uncertainty is stated as the combined uncertainty of measurement multiplied by the coverage factor k=2, which assuming a normal distribution has a coverage probability of 95 % (complies with GUM by ISO (1995, ISBN 92-67-10188-9)). Instrument specifications are subject to purchased configuration. All specifications may change without notice. www.raysafe.com