Fingertip Pulse Oximeter

Similar documents
ADC Diagnostix 2100 Fingertip Pulse Oximeter. Directions for use

INSTRUCTION MANUAL Contactless Clinical Infrared Thermometer GT-101

Critical Care Therapy and Respiratory Care Section

Instructions to User

Policy Name: Patient Monitoring via the Patient SafetyNet (PSN) Monitoring System

Automatic Wrist Blood Pressure Monitor with

Operating Manual FINGERSIM SET /5.11. for SECULIFE NIBP in Combination with SECULIFE OX1

digital fever alert ear thermometer suitable from birth instruction manual

SpO2 PULSE OXYMETER MD-300

EMMA Emergency Capnometer/ Finger Pulse Oximeter Version 2 Related Documents SOP-Medical and Equipment packs Stuart Daly; Phil Munro

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Forehead. Use and Care Manual. Thermometer IMPORTANT! READ AND SAVE THESE INSTRUCTIONS. Model V977

DSM155 Digital LED Light Meter User s Manual DSM155

INSTRUCTION MANUAL HEM-6052

FiberOptix TM. AP Sensor Zeroing and Calibration Guide INTRA-AORTIC BALLOON PRODUCT SUPPORT HOTLINE. U.S. and Canada: IABP (4227)

High Temperature InfraRed Thermometer with Laser Pointer

12 Volt 30 Amp Digital Solar Charge Controller

Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D / / 2012

DX-395. FM/AM/SW1/SW2 PLL Synthesized Receiver OWNER S MANUAL. Cat. No Please read before using this equipment.

Pulse Oximeter. Manual

User s manual FLIR VP50/VP52 Non-contact AC voltage detector

User's Guide. Pinless Moisture Meter. Model MO257

Instant Ear Thermometer

Ultrasonic Distance Meter. Operation Manual

Touchless Forehead Thermometer

Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS

e I N S T RUC TION M ANU A L

Water Tank Level Monitor

Milwaukee USER MANUAL

Mold Alert... Operation Manual P0270

User manual TAB-10C

MEMO ALARM CLOCK Operating Manual

Dual Laser InfraRed (IR) Thermometer with Color Alert

Hygro-Thermometer + InfraRed Thermometer Model RH101

Dual Laser InfraRed (IR) Thermometer

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W

TMS TANK MANAGEMENT SYSTEM

Instant Ear Thermometer

Wireless Home Security Alarm System AM 500

Soil Sensor Reader PRODUCT MANUAL

DegerPipes. Electronic Bagpipe Chanter. Operating Manual. Overview. Precautions. PHONES Jack MIDI-Jack. Mouthpiece

Solar Weather Station Model: BAR806 / BAR806A

Dual Laser InfraRed (IR) Thermometer

510 / 511 Cardiorespiratory Monitor Preventative Maintenance Test Procedure

How To Use A Power Supply Unit (Upu)

RomiApex A-15. APEX LOCATOR User Manual

The Weather Station. Hanging Hole. LCD Display Battery Cover. Function Keys. Detachable Stand

Atomic Alarm Clock Instruction Manual

DR 70 Portable Digital DAB+ and FM Radio Receiver User Manual English Please read carefully before using this product

Getting started with

REMOTE TERMOMETER TE 219 LED USER MANUAL

User manual. Videomatic VERDE MF

4.3-inch Back-Up Camera

LG Wireless Remote Controller

CFMRI Physiological Monitoring System. Operator s Manual

FUJI DIGITAL SOUND DETECTOR FSB-8D INSTRUCTION MANUAL

BODY FAT & HYDRATION % MONITOR SCALE

Reizen Atomic Talking Analog Alarm Clock Item # Instruction Manual

Periodic Verification System Model 775PVS

ENGLISH INSTRUCTION & INSTALLATION MANUAL DUCTLESS MINI SPLIT AIR CONDITIONING SYSTEMS

body fat scale + Instruction Manual complete body analysis scale HEALTH

Thermistor Thermometer MODEL NO

HDTV Anywhere USER MANUAL / HDTV Anywhere ALL RIGHTS RESERVED MARMITEK

EzFlex Combustible Gas Leak Detector Model EZ40

Lexmark E260, E260d, and E260dn Series. Maintenance Guide

Light Meter (Foot Candles)

Portable Air Conditioner

How To Use A Meter

INSTRUCTIONS FOR VOICE CONTROLLED TALKING ALARM CLOCK

PROFESSIONAL HOME ALARM SYSTEM. Important! Please read these instructions carefully.

QFP910. Water detector. Building Technologies. CE1N2732en Synco living

Operating Manual Color Changing Fountains

Operator s Manual. Model 2500A PalmSAT. Pulse Oximeter with Alarms English

Tina 2A/B Adaptor unit

Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer

User s Manual AURORA 1.2K/2.2K

Advantium 2 Plus Alarm

DAB+ / FM Tuner Model: TU-201

Microwave Meter. Instruction Manual

Electronic Brake Controller Hayes Brake Controller Company ENERGIZE III P/N # 81741B or ENERGIZE XPC P/N #81745 OPERATION MANUAL

Suggested BiliChek Usage Protocol

Laser Distance Meter Laser-Distanzmesser Laser distance-mètre Metro di distanza laser Medidor Láser de Distancia User Manual

Dual Laser InfraRed (IR) Thermometer

el-720 Carbon Dioxide Gas Leak Detector User s Guide

EXERGEN. English. TemporalScanner TM. Accurate Temperature with a Gentle Forehead Scan. Operators Manual for TAT-5000

RT-1261A Ear and Forehead Thermometer Owner s Manual

RADIO CONTROLLED DIGITAL CLOCK MODELS / 88906

NEULOG HEART RATE AND PULSE LOGGER SENSOR GUIDE

KNX TH65-AP. Thermo-Hygrometer. Technical specifications and installation instructions Item number 70184

Tympanic Thermometer

SHUTTER REMOTE USER MANUAL UNIVERSAL WIRED & IR RS-UWIR100 COMPATIBLE WITH MOST DSLR CAMERAS

INSTRUCTION MANUAL. Ultra Premium Blood Pressure Monitor featuring ComFit TM Cuff. Model HEM-775

Model: Manual DC: WIRELESS COLOR FORECAST STATION

WIRELESS ALARM KIT INSTRUCTION MANUAL COMPLETE ALARM SYSTEM FOR HOME OR BUSINESS. Model Number: IT214054

2-Way Wireless Indoor Siren Installation Guide

The EasySense unit can detect that the Smart Q Heart Rate Sensor is connected and the range it is set to.

Important Safety Instructions

IN-OUT Thermometer with Cable Free Sensor and Clock

LCD Display. Function keys. Stand. Adapter socket

Transcription:

Fingertip Pulse Oximeter Instruction manual GT-300C320 0118

TABLE OF CONTENTS General Description 14 Precautions 15 Range of Application 16 Battery Installation 16 Description of Front Panel 17 Instructions for Use 17 Evaluation of Pulse Oximetry 20 Specification 21 Possible Problems and Solutions 22 Cleaning Instructions 24 Limited Warranty 24 GENERAL DESCRIPTION Haemoglobin saturation is the percentage of oxygen-bearing haemoglobin (HbO 2 ) molecules in proportion to the amount of oxygen that the haemoglobin (Hb) is capable of carrying in the blood. It expresses the concentration of oxyhaemoglobin in the blood and is a very important parameter of the respiratory and circulation system. Many respiratory diseases can result in a decline in haemoglobin saturation levels. In clinical medicine it is very important to determine a patient s haemoglobin saturation level in good time so as to detect problems sufficiently early. Pulsoximetry is a non-invasive monitoring procedure for the determination of oxygen saturation in arterial (oxygenated) blood by means of a photosensor. To obtain a diagnosis, it is only necessary for the patient to lay one of his fingers on a photoelectric sensor. The display screen shows the measured value of haemoglobin saturation. Clinical experiments have demonstrated the high degree of precision and repeatability of this procedure. 14

PRECAUTIONS 1 Do not use the pulse oximeter in an MRI or CT environment. 2 Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere. 3 The pulse oximeter is intended only as an ancillary procedure in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms. 4 Prolonged use or the patient s condition may require that the sensor site be changed regularly. Change the sensor site and check the patient s skin integrity, circulatory status, and correct alignment at least every 4 hours. 5 Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or immersing the sensors in liquid, which will result in inaccurate readings. 6 Store the pulse oximeter in a dry environment. 7 The following factors may impair the SpO 2 measurements and/or lead to inaccurate readings: Significant levels of dysfunctional haemoglobins (such as carboxy-haemoglobin or methaemoglobin) Intravascular dyes such as indocyanine green or methylene blue High levels of ambient light (shield the sensor area against direct sunlight) Excessive patient movement Venous pulsations Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line The patient has hypotension, severe vasoconstriction, severe anaemia, or hypothermia The patient is in cardiac arrest or is in shock Fingernail polish or artificial fingernails Cold or dirty fingers 15

RANGE OF APPLICATION The Geratherm oxy control 4c can be used to measure human haemoglobin saturation levels and the heart rate via a finger. The product is suitable for use at home, in hospital (including clinical use in internal medicine/surgical wards, in anaesthesia, paediatrics, intensive care units etc.), in social medical institutions or in sports medicine (it can be used before and after sporting activity - use during sporting activities is not recommended). BATTERY INSTALLATION 1. Place the two AAA batteries inside the battery compartment ensuring correct polarity. 2. Slide the battery cover on horizontally as shown by the arrow. Notes: Batteries must be inserted with the correct polarities. Otherwise, damage might be caused to the device. Please insert or remove batteries in the correct order, or the device bracket may be damaged. Please remove the batteries if the oximeter will not be used for a long period. When the battery status indicator appears, replace the batteries in good time. 16

DESCRIPTION OF FRONT PANEL Perfusion Index (PI) Pulse frequency indicator SpO 2 / oxygen saturation Pulse rate Battery status indicator 97 7 4 Graphic representation of SpO 2 The pulse frequency indicator flashes on and off in accordance with the pulse rate displayed. INSTRUCTIONS FOR USE 1 Insert two AAA batteries into the battery compartment before replacing the cover. 2 Press the device at the rear end so as to open the front section for the insertion of a finger. 3 When your finger is in the oximeter, the nail surface must face upwards. 4 After inserting the finger let go of the rear end. 5 Press the power button on the front panel once. 6 While the oximeter is working, you should not move your finger and you should keep still and quiet. 7 Read the values shown on the display screen. 17

INSTRUCTIONS FOR USE After turning on the oximeter, each time you press the power button the oximeter will switch to another display mode. There are 6 display modes, as follows: 1. 2. 97 7 4 97 7 4 3. 4. 97 7 4 97 7 4 5. 6. 97 7 4 97 7 4 Press the power button for a fairly long time (longer than 1 second) in order to set the brightness level of the display panel. Holding the button down will cause the brightness of the oximeter to change gradually. There are 10 brightness levels. The default level is level four. 18

INSTRUCTIONS FOR USE Audible pulse rate indicator During the measuring procedure you will hear a tone indicating your pulse rate. Alarm: Range: SpO 2 : <90% Pulse rate: <60 bpm or >100 bpm Alarm modes: audible and visual alarm Visual alarm: The measured pulse and oxygen values flash if they exceed the previously set lower or upper limits. Audible alarm: This can be switched on or off. The alarm sounds briefly twice if the measured values exceed the previously set limits. Switching the alarm on and off: In order to switch the alarm off, press the on/off button when the audible alarm sounds. You will enter alarm mode for a period of 30 seconds, thus switching the alarm off. However, you only switch off the alarm tone, not the audible pulse rate indicator. The alarm mode is automatically switched off if a new alarm sounds or the measured values are within the normal range. Perfusion Index (PI) The perfusion index (PI) shows the signal strength of the arterial pulse and indicates the proportion of pulsatile signals to non-pulsatile signals. The bar is highest when the quality of the blood supply is at its best. PI is influenced by the strength of the pulse rate, the quality of the blood supply to the fingertip and the temperature of the finger. If, for example, the pulse rate is weak, the blood supply to the fingertip poor or the finger temperature low, the PI will also be low. 19

EVALUATION OF PULSE OXIMETRY In healthy patients, oxygen saturation levels lie between 96 and 98 %. Values below 95 % can indicate a health disorder and make it necessary to supply the patient with additional oxygen. If your oxygen saturation value is below 95 %, please consult your physician. Normal pulse values depend on the age and fitness level of the individual. Here are a few reference values indicating normal pulse rates: Infants under 1 year: 100-160 bpm Children between 1 and 10 years: 70-120 bpm People over 10 years: 60-100 bpm Highly trained individuals/athletes: 40-60 bpm 20

SPECIFICATION 1. Display type: 4-colour LCD-display 2. SpO 2 : Measurement range: 70 99 % Accuracy: ±2 % between 80% and 99%; ±3% between 70% and 80% 3. Pulse rate: Measurement range: 30-235 bpm Accuracy: ±2 bpm or ±2 % Pulse intensity: Pulse rate indicator 4. Power requirements: Two alkaline batteries, type AAA 1.5 V, 600mAh Power consumption: Less than 40mA Battery status indicator: Battery life: up to 30 hours in continual use 5. Dimensions: Length: 60mm; Width: 34mm; Height: 30mm Weight: 50g (including two AAA batteries) 6. Environmental requirements: Operating temperature: 5-40 C Storage temperature: -10-40 C Humidity: 15 % - 80 % in operation 10 % - 80 % in storage 7. Quality standards: Certification in compliance with EN ISO 13485, Conformity in compliance with EN 9919:2005 Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use, EN60601-1, EN60601-1-2, standard Council Directive 93/42/ EEC 21

POSSIBLE PROBLEMS AND SOLUTIONS Problems Possible reason Solution SpO 2 or PR cannot be shown normally. 1. Finger is not inserted correctly. 2. Patient s oxyhaemoglobin level is too low to be measured. 1. Try inserting the finger again. 2. Try several more times if you are sure that there is no problem with the device. Please go to a hospital in good time to obtain a precise diagnosis. SpO 2 or PR is shown unstably 1. Finger might not be inserted far enough. 2. Finger is trembling or the patient s body is moving. 1. Try inserting the finger again. 2. Try not to move. The oximeter cannot be switched on 1. Battery power may be inadequate or not be there at all. 2. Batteries may have been inserted incorrectly. 3. The oximeter might be damaged. 1. Please replace batteries. 2. Please reinsert the batteries. 3. Have the device tested by the manufacturer 22

POSSIBLE PROBLEMS AND SOLUTIONS Problems Possible reason Solution Indication lamps suddenly go off 1. The product is automatically switched off if no signal is detected for longer than 8 seconds. 2. Low power/ Batteries empty 1. Normal 2. Replace the batteries. Error3 or Error4 displayed on screen Error7 displayed on screen 1. Low power/ Batteries empty 2. Receiving sensor shielded or sensor damaged. 1. Low power/ Batteries empty 2. Infrared sensor damaged. 1. Insert new batteries. 2. Have the device tested by the manufacturer. 1 Change the batteries. 2 Have the device tested by the manufacturer. 23

CLEANING INSTRUCTIONS Please use medical alcohol to clean the synthetic material touching the finger inside the oximeter, and clean the test finger before and after using. Please clean the surface of the pulse oximeter before each use. LIMITED WARRANTY This pulse oximeter is guaranteed for 2 years from the date of purchase against manufacturer s defect under conditions of normal use. Following the instructions precisely will ensure that the device functions reliably for many years. If your pulse oximeter does not function properly due to defective parts or assembly we will repair it free of charge. All parts of the pulse oximeter are covered by this warranty. The warranty does not cover damage to your unit due to improper handling. The manufacturer recommends that the accuracy of the device be checked by an authorized laboratory every 2 years (in Germany this is a statutory requirement). You must adhere to the applicable regulations when disposing of the device. Manufacturer: Geratherm Medical AG Fahrenheitstraße 1 98716 Geschwenda Germany 24