Agilent RF Preselector, 20 Hz to 2 GHz For the 8566B or 8568B Spectrum Analyzer Data Sheet

Similar documents
Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

Agilent N2717A Service Software Performance Verification and Adjustment Software for the Agilent ESA Spectrum Analyzers Product Overview

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy

Agilent E5100A Network Analyzer

81110A Pulse Pattern Generator Simulating Distorted Signals for Tolerance Testing

Agilent 8753ET/8753ES Network Analyzers

Agilent E6832A W-CDMA Calibration Application

Agilent 8904A Multifunction Synthesizer dc to 600 khz

Agilent 8762/3/4A,B,C Coaxial Switches

Agilent AN 1315 Optimizing RF and Microwave Spectrum Analyzer Dynamic Range. Application Note

Agilent Creating Multi-tone Signals With the N7509A Waveform Generation Toolbox. Application Note

Agilent E4401B, E4402B, E4404B, E4405B, and E4407B ESA-E Series Spectrum Analyzers

khz to 2.7 GHz Spectrum Analyzer

Agilent E3830 Series Wide-bandwidth Signal Analyzer Custom Systems 100 MHz Bandwidth Microwave Vector Signal Analysis

Agilent Electronic Calibration (ECal) Modules for Vector Network Analyzers

HP 8590 E-Series Portable Spectrum Analyzers Technical Specifications

Agilent Test Solutions for Multiport and Balanced Devices

Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics

Agilent 87421A/87422A Power Supply

R&S ZNBT8 Vector Network Analyzer Specifications

R&S ZNC Vector Network Analyzer Specifications

Selecting RJ Bandwidth in EZJIT Plus Software

Agilent 4338B Milliohm Meter

A Guide to Calibrating Your Spectrum Analyzer

RECOMMENDATION ITU-R SM Measuring sideband emissions of T-DAB and DVB-T transmitters for monitoring purposes

Agilent GSM/EDGE Base Station Test with the E4406A VSA and ESG-D Series RF Signal Generators Product Overview

DSA800 Series Spectrum Analyzer

R&S ZVL Vector Network Analyzer Specifications

AN Application Note: FCC Regulations for ISM Band Devices: MHz. FCC Regulations for ISM Band Devices: MHz

Agilent Split Post Dielectric Resonators for Dielectric Measurements of Substrates. Application Note

Loop Bandwidth and Clock Data Recovery (CDR) in Oscilloscope Measurements. Application Note

HF Receiver Testing. Issues & Advances. (also presented at APDXC 2014, Osaka, Japan, November 2014)

How To Use A Sound Card With A Subsonic Sound Card

Technical Datasheet Scalar Network Analyzer Model MHz to 40 GHz

HP 70950B OPTICAL SPECTRUM ANALYZER

MATRIX TECHNICAL NOTES

APSYN420A/B Specification GHz Low Phase Noise Synthesizer

Keysight N9320B RF Spectrum Analyzer

Agilent Technologies E8047B Analysis Probe System for the Intel Xeon Processor Family

Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics

Keysight Technologies N9320B RF Spectrum Analyzer

X-Series Signal Analysis. Future-ready instruments Consistent measurement framework Broadest set of applications and software

Application Note AN-00125

Network Analyzer Operation

Agilent 8590C/E Series Spectrum Analyzers and Accessories Configuration Guide

AM/FM/ϕM Measurement Demodulator FS-K7

Impedance 50 (75 connectors via adapters)

R&S ZCxxx Millimeter-Wave Converters Specifications

Agilent E4980A Precision LCR Meter 20 Hz to 2 MHz

Agilent 4339B/4349B High Resistance Meters

Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz

Agilent 8720 Family Microwave Vector Network Analyzers

Agilent Technologies 3000 Series Oscilloscopes

SPECTRUM MONITOR MC-277B

Raptor RXi Ultra-fast scanning Countersurveillance Receiver

Agilent U2000 Series USB Power Sensors

FREQUENCY RESPONSE ANALYZERS

R&S ZVA Vector Network Analyzer Specifications

Agilent U2000 Series USB Power Sensors

Agilent Evolution of Test Automation Using the Built-In VBA with the ENA Series RF Network Analyzers

DT3: RF On/Off Remote Control Technology. Rodney Singleton Joe Larsen Luis Garcia Rafael Ocampo Mike Moulton Eric Hatch

Low-cost EMI Pre-compliance Testing Using a Spectrum Analyzer

RECOMMENDATION ITU-R BS.704 *, ** Characteristics of FM sound broadcasting reference receivers for planning purposes

Keysight Technologies 8 Hints for Better Spectrum Analysis. Application Note

Agilent E363xA Series Programmable DC Power Supplies. Data Sheet

LT-82 Stationary IR Transmitter

Agilent Measuring Noninsertable Devices

Current Probes. User Manual

F = S i /N i S o /N o

MYRICOM, INC. TEST REPORT FOR THE SERVER RACK, 10G-PCIE2-8B2-2S & 10G-PCIE2-8B2-2QP FCC PART 15 SUBPART B SECTIONS AND CLASS A TESTING

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

Conquering Noise for Accurate RF and Microwave Signal Measurements. Presented by: Ernie Jackson

SR2000 FREQUENCY MONITOR

Measurement, analysis, and monitoring of RF signals

Agilent PN Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth

Agilent PN RF Component Measurements: Amplifier Measurements Using the Agilent 8753 Network Analyzer. Product Note

Some Measurements on DVB-T Dongles with E4000 and R820T Tuners:

DVT913 TV CHANNEL CONVERTER

RF Communication System. EE 172 Systems Group Presentation

HD Radio FM Transmission System Specifications Rev. F August 24, 2011

Measurement of Adjacent Channel Leakage Power on 3GPP W-CDMA Signals with the FSP

Agilent 8509C Lightwave Polarization Analyzer

RF and Microwave Accessories. CD-ROM Catalog. Find the right component for your Rohde & Schwarz test & measurement equipment

Keysight N9000A CXA Signal Analyzer

DDX 7000 & Digital Partial Discharge Detectors FEATURES APPLICATIONS

ETSI EN V1.2.1 ( )

Application Note Receiving HF Signals with a USRP Device Ettus Research

The Next Generation in Automated Oscilloscope Test

DRM compatible RF Tuner Unit DRT1

R&S ESL EMI Test Receiver Compact, cost-effective measuring receiver

802.11ac Power Measurement and Timing Analysis

HP 8590 C/E/L/Q and EM Series Spectrum Analyzers and Accessories Configuration Guide

PXI. GSM/EDGE Measurement Suite

Field Calibration Software

Electromagnetic Compatibility and International Regulatory Approvals

CX380X Advanced Spectrum and Burst QAM Analyzer

ETSI EN V1.2.2 ( )

Cond-IS RF Conducted Immunity System

High Power Amplifier Measurements Using Agilent s Nonlinear Vector Network Analyzer

Advanced Test Equipment Rentals ATEC (2832)

Transcription:

Agilent RF Preselector, 20 Hz to 2 GHz For the 8566B or 8568B Spectrum Analyzer Data Sheet Your High-Performance Spectrum Analyzer can also be a Test Receiver Add the 85685A RF preselector to the 8566B or 8568B general-purpose spectrum analyzer, and you have a test receiver for specialized applications. The RF preselector enhances these high-performance spectrum analyzers with tracking filters and preamplifiers that cover the 20 Hz to 2 GHz range. The spectrum analyzer/test receiver is sensitive to low-level signals while providing overload protection from out-of-band signals. The result, for tests in the presence of high-level interference, is a measurement range 30 db greater than that of the spectrum analyzer alone. (The improvement is even greater in the presence of broadband pulse-interference.) The test receiver is easy to operate; simply use the spectrum analyzer controls without concern for preselector settings. The preselector automatically adjusts input-filter tracking, and the spectrum analyzer reports preselector operating conditions on the CRT display. For remote operation with a computer, the receiver system is fully GPIB programmable (GPIB is hardware, software, documentation, and support for IEEE-488 and IEC-625, worldwide standards for interfacing instruments.) The 85685A preselector adds: Automatic filter tracking Input overload protection Low system noise to the 8566B and 8568B spectrum analyzers.

EMI Testing According to CSPR Publication 16 Recommendations Combining the 85685A RF preselector with the 8566B or 8568B spectrum analyzer and the 85650A quasi-peak adapter produces an EMI receiver with the characteristics recommended in CISPR publication 16 1. The RF preselector provides the measurement sensitivity and overload protection needed for FCC and VDE radiated emission testing at open sites. For commercial and MIL-STD conducted EMI tests, the low-frequency input tolerates large impulses and line impedance stabilization network (LISN) transients. A built-in calibrator ensures ±2.0 db absolute-amplitude accuracy as required by the FCC and VDE. For measurement confidence, a linearity check tests for system overload and distortion. The sample display shows results for VDE 0871 emission testing at an open site. To prevent overload from the highlevel ambient signals, a spectrum analyzer needs input attenuation. This however, increases displayed noise, which can mask the low-level emissions. Using the RF preselector in this test eliminates the need for adding attenuation. The noise of the spectrum analyzer, as shown, exceeds the VDE limit, but the spectrum analyzer/test receiver noise is more than 10 db below the lowest limit. 1. Comite International Special Des Perturbations Radio-Electriques (CISPR) publication 16 is the "CISPR specification for radio interference measuring apparatus and measurement methods." Signal Monitoring in the Presence of High-Level Interference The RF preselector gives the 8566B or the 8568B spectrum analyzer the sensitivity and selectivity of a receiver. The spectrum analyzer/test receiver measures signals within the preselector filter passband and rejects outof-band interference by 40 db. Even in the presence of interference with a total input power of +10 dbm, your test receiver can monitor signals down to -150 dbm. The RF preselector improves signal reception for the broadband signal environment of a mobile test station. In the 150 170 MHz business band, the RF preselector reduces interference from radio and TV broadcast stations and from amateur radios. For the 800 900 MHz cellular radio band, the spectrum analyzer/test receiver makes field strength measurements on cellular base-station transmitters while rejecting interference from cellular mobile transmitters. Transmitter Spurious Testing with Fast Wideband Measurements Reduce transmitter-spurious test time by adding the RF preselector to your 8566B or 8568B spectrum analyzer. The preselector decreases input overload from the out-of-band carrier and thereby increases the range for measuring lowlevel signals. Compared to the spectrum analyzer alone, the spectrum analyzer/ test receiver uses a wider resolution bandwidth for the same measurement range, resulting in a faster sweep time. For example, to measure spurious signals over a 1 GHz frequency range at 100 db relative to the transmitter carrier, a spectrum analyzer by itself takes 3000 seconds, but when configured as a test receiver it requires only 3 seconds. 2

s s describe the instrument's warranted performance over the temperature range of 0 C to +55 C unless otherwise noted. Typical values for the specified parameters represent probable but non-warranted performance. Nominal values provide useful but non-warranted information about functional performance. Test Receiver The following system specifications apply when the 85685A RF preselector (RFP) operates with the 8566B or 8568B spectrum analyzer (SA) and with or without the 85650A quasi-peak adapter (QPA). The test receiver system is fully GPIB programmable. Refer to the specific instrument data sheets for more detailed information. Frequency range RFP bypass 1 SA frequency resolution SA frequency accuracy SA frequency reference error SA Frequency selectivity 3 3 db resolution BW Test receiver frequency 20 Hz 1.5 GHz with 8568B 20 Hz 2.0 GHz with 8566B 20 Hz 18 GHz (nominal) with 8568B 1 Hz ±(tuned frequency x freq. reference error) Hz for SA zero span Settabilit y2 < 10-8, Drift < 2 x 10-7 /year 10 Hz 3 MHz in 1, 3, 10 sequence Measurement range Gain compression (0 db atten. for RFP and SA Inside RFP passband Outside RFP passband Displayed average noise level (0 db atten. for RFP and SA 10 Hz resolution BW) Broadband signal sensitivity 5,6 (0 db atten. for RFP and SA, 10 Hz resolution BW) Second harmonic distortion (0 db atten. for RFP and SA) For -35 dbm input signal: Within RFP passband Outside RFP passband Third order intermodulation (0 db atten. for RFP and SA) For two -35 dbm input signals: Within RFP passband Outside RFP passband Residual responses (RFP input terminated, 0 db atten. for RFP and SA Amplitude accuracy 6 (+10 to + 30 C, 10 db SA atten. 0 20 db RFP atten.) Test receiver amplitude -150 to +30 dbm (-43 to +137 dbµv) (nominal) < 1 db (nominal) for input -30 dbm < 1 db (nominal) for input +10 dbm < -132 dbm (-25 dbµv) for input 9 khz 1 MHz 4 < -150 dbm (-43 dbµv) for input 1 1500 MHz < -147 dbm (-40 dbµv) for input 2.0 GHz < 31 dbµv/mhz (nominal) for input < 1 MHz 4 < 13 dbµv/mhz (nominal) for input = 1 1500 MHz < 16 dbµv/mhz (nominal) for input 2.0 GHz < -55 dbc (nominal), Equiv. SOI = +20 dbm < -75 dbc (nominal), Equiv. SOI = +40 dbm < -40 dbc (nominal), Equiv. SOI = -15 dbm < -110 dbc (nominal), Equiv. SOI = +20 dbm < -90 db for input 2 khz 1 MHz < -112 db for input 1 MHz < -120 db (typical) for input 1 MHz ±2.0 db for input < 1.0 GHz ±3.0 db for input 1.0 GHz RFP bypass SA specification plus: ± 0.5 db (nominal) for input 2.0 GHz ± 1.0 db (nominal) for input > 2.0 GHz 3

CISPR EMI Receiver The following system specifications apply when the 85685A RF preselector (RFP) operates with the 8566B or 8568B spectrum analyzer (SA) and with the 85650A quasi-peak adapter (QPA). The EMI receiver system has the characteristics recommended in CISPR publication 16 and is fully GPIB programmable. Refer to the specific instrument data sheets for more information. Frequency range SA frequency resolution SA frequency accuracy SA frequency reference error SA Frequency selectivity 9 CISPR 6 db bandwidth (QPA) CISPR EMI receiver frequency 10 khz 1 GHz 1 Hz ±(tuned frequency x freq. reference error) Hz Settability 2 < 10-8, Drift < 2 x 10-7 /year 200 Hz for input = 10 150 khz (band A) 9 khz for input = 150 khz 30 MHz (band B) 120 khz for input = 300 1000 MHz (band C, D) Make dedicated receiver measurements with a general-purpose spectrum analyzer 4

Measurement range Amplitude accuracy 8 (+10 to +30 C, 0 20 db RFP atten.) Calibration 10 Linearity check Impulsive signal tolerance Detection 9 Audio output CISPR EMI Amplitude +137 dbµv to noise floor (refer to amplitude sensitivity table) ±2.0 db for 10 db SA atten. ± 2.5 db for 0 db SA atten. Automatic calibration for absolute amplitude with comb generator (RFP) 3 db input attenuator test for system overload and distortion (RFP) Input #1 withstands LISN transients (RFP) Input #2 withstands CISPR calib. pulses (RFP) Peak and average (SA), quasi-peak (QPA) Speaker with volume control (QPA) CISPR EMI Amplitude sensitivity EMI receiver Frequency band Resolution bandwidth (6 db) Average noise 6 (RFP/SA atten. = 0 db) Quasi-peak noise (RFP/SA atten. = 0 db) 10 150 khz (band A) 200 Hz -12 dbµv 4-17 dbµv typical -15 dbµv (nominal) 150 khz 1 MHz (band B) 9 khz +4 dbµv -1 dbµv typical 1 30 MHz (band B) 9 khz -13 dbµv -16 dbµv typical 30 1000 MHz (band C, D) 120 khz -1 dbµv -6 dbµv typical 0 dbµv (nominal) -11 dbµv (nominal) 0 dbµv (nominal) 5

85685A RF Preselector The following instrument specifications apply to the 85685A RF preselector (RFP). 85685A RF preselector input Input #1 Input #2 Frequency range 20 Hz 50 MHz 20 MHz 2 GHz RFP bypass 1 DC 18 GHz (nom.) Connector type Fuse blow time Maximum safe input power Average power Impulsive signals 10 db RFP atten. 20 db RFP atten. DC voltage Standing wave ratio (SWR) 12 10 db RFP atten. 0 db RFP atten. RFP bypass Attenuator range Linearity check Filter characteristics Fixed filter BW (3 db) Tracking filter BW (3 db) Stopband attenuation Preamplifier gain BNC, 50 Ω (nominal) < 0.1 sec for > + 35 dbm +30 dbm (1 Watt) 100 Watts peak for a 10 µs pulse 2,000 Watts peak for a 10 µs pulse 11 (nom.) 0 Volts < 1.5:1 < 1.5:1 (nominal) NA 0 50 db in 10 db steps (nominal) Add 3 db (nominal) Type N, 50 Ω (nom.) NA +30 dbm (1 Watt) NA BW range = 10 khz 1 MHz (nom.) for input < 2 MHz 10% (nom.) of tuned freq. for input = 2 55 MHz 5% (nom.) of tuned freq. for input = 55 500 MHz 20 MHz (nom.) for input = 500 2000 MHz > 40 db (nominal) 20 db (nominal) for 0 db RFP atten. < 1.5:1 < 2.0:1 (nominal) < 1.5:1 (nominal) Comb generator Line spacing Line amplitude SWR GPIB interface* Private BUS Public BUS interface functions (public BUS) 85685A RF preselector output 100 khz, 500 khz, 1 MHz, 5 MHz (nominal) -40 to -60 dbm (nominal) < 1.5:1 (nominal) Digital BUS and sweep control between RFP and SA IEEE 488-1978 BUS SH1, AH1, T6, L4, SR1, RL1, PPO, DC1, DT1, C0, E1 6

Temperature Warm-up time Temp. equilibrium EMI Power requirements Weight Size Firmware compatibility for RFP operation with: 8566B, firmware data 5.5.85 8566B, firmware data < 5.5.85 8566A, any firmware date 8568B, firmware data 14.1.85 8568B, firmware data < 14.1.85 8568A, any firmware date 0 to + 55 C operating -40 to +75 C storage 30 minutes after cold start 2 hours after warm-up General Conducted and radiated interference is in compliance with MIL-STD 461A methods in CE03 and RE02, CISPR publication 11 (1975), and Messempfaenger-Postverfuegung 526/527/79 (Kennzeichnung Mit F-Nummer/Funkschutzzeichen). 100, 120, 220, or 240 V RMS +5%, -10%; 50 400 Hz; 35 VA (nominal) 19 kg (42 lbs) net (nominal) 133 mm (5.25") H x 425 mm (16.75") W x 508 mm (20.0") D (nominal) Compatible 85685A Option 066 required 8566A + 01K retrofit kit required Compatible 85685A Option 068 required 8568A + 01K retrofit kit required Product Information 85685A RF preselector Option 010, Rack mount slide kit Option 908, Rack flange kit without handles Option 910, Extra manual Option 913, Rack flange kit with handles Firmware compatibility options (see compatibility specifications above) 85685A Option 066, Firmware upgrade kit* for 8566B 85685A Option 068, Firmware upgrade kit* for 8568B 8566A + 01K Retrofit kit 8568A + 01K Retrofit kit 8566B spectrum analyzer 8568B spectrum analyzer 85650A Quasi-peak adapter Accessories 85864B EMI measurement software Transit case, part number 1540-0663 EMI accessories 13 : Antennas, LISNs, current probes, and amplifiers 7

Footnotes: 1. The RFP bypass switch connects input #2 directly to the RF output. 2. Refer to spectrum analyzer test and adjustment manual for the correct frequency reference adjustment. 3. Refer to spectrum analyzer data sheet for filter BW accuracy and filter shape. 4. 17 db higher with 8566B for frequency < 50 khz. 5. Sensitivity is approximately equal to system peak noise level. 6. Peak noise is nominally 8 db higher than average noise. 7. Measurement sensitivity is normalized to 1 MHz resolution BW. 8. s apply after system calibration with the comb generator in the 85685A. Refer to 85685A operating manual for correct amplitude measurement technique. 9. The system meets CISPR publication 16 recommendations for resolution BW, detector time constant, and detector pulse response. 10. System calibration can be verified with an external CISPR standard pulse calibration generator. 11. Typical LISN transients. 12. SWR is measured at the tuned frequency of the spectrum analyzer. 13. Refer to the EMI measurement solutions guide (5954-2702) for more information. Agilent Email Updates www.agilent.com www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. Agilent Open www.agilent.com/find/open Agilent Open simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent offers open connectivity for a broad range of system-ready instruments, open industry software, PC-standard I/O and global support, which are combined to more easily integrate test system development. Agilent Direct www.agilent.com/find/agilentdirect Quickly choose and use your test equipment solutions with confidence. Agilent Technologies Test and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Two concepts underlie Agilent s overall support policy: Our Promise and Your Advantage. Our Promise Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you receive your new Agilent equipment, we can help verify that it works properly and help with initial product operation. Your Advantage Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and onsite education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products. Agilent T&M Software and Connectivity Agilent s Test and Measurement software and connectivity products, solutions and developer network allows you to take time out of connecting your instruments to your computer with tools based on PC standards, so you can focus on your tasks, not on your connections. Visit www.agilent.com/find/connectivity for more information. For more information on Agilent Technologies products, applications or services, please contact your local Agilent office. Phone or Fax United States: Korea: (tel) 800 829 4444 (tel) (080) 769 0800 (fax) 800 829 4433 (fax) (080)769 0900 Canada: Latin America: (tel) 877 894 4414 (tel) (305) 269 7500 (fax) 800 746 4866 Taiwan: China: (tel) 0800 047 866 (tel) 800 810 0189 (fax) 0800 286 331 (fax) 800 820 2816 Other Asia Pacific Europe: Countries: (tel) 31 20 547 2111 (tel) (65) 6375 8100 Japan: (fax) (65) 6755 0042 (tel) (81) 426 56 7832 Email: tm_ap@agilent.com (fax) (81) 426 56 7840 Contacts revised: 05/05/05 The complete list is available at: www.agilent.com/find/contactus Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2005 Printed in USA, October 5, 2005 5954-2708