Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Similar documents
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Production of X-rays and Interactions of X-rays with Matter

Assessing Radiation Dose: How to Do It Right

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Atomic and Nuclear Physics Laboratory (Physics 4780)

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

DIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE

Lecture 2 Macroscopic Interactions Neutron Interactions and Applications Spring 2010

Prepublication Requirements

ACCELERATORS AND MEDICAL PHYSICS 2

Conversion coefficients from air kerma to personal dose equivalent H p (3) for eye-lens dosimetry

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Patient Exposure Doses During Diagnostic Radiography

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Diagnostic Exposure Tracking in the Medical Record

Radiation Safety Characteristics of the NOMAD Portable X-ray System

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Practical exercise: Effective dose estimate in CT

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY

AAPM protocol for kv x-ray beam dosimetry in radiotherapy and radiobiology

The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement.

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Introduction to the Monte Carlo method

Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS

Computerized Tomography, X-rays, and the Radon Transform

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Purchasing a cardiac CT scanner: What the radiologist needs to know

Qualitätsmanagement-Handbuch

Appendix A. An Overview of Monte Carlo N-Particle Software

Variance reduction techniques used in BEAMnrc

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Multi-slice Helical CT Scanning of the Chest

Chemistry 1000 Lecture 2: Nuclear reactions and radiation. Marc R. Roussel

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.

Dose Tracking System:

Principles of dosimetry The ionization chamber

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Chapter NP-5. Nuclear Physics. Nuclear Reactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 NUCLEAR REACTIONS 2.0 NEUTRON INTERACTIONS

Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards

Monitoring Patient Radiation Dose in VA. Charles M. Anderson MD, PhD Chief Consultant for Diagnostic Services Veterans Health Administration

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

The Challenge of CT Dose Records

How To Calculate A Patient Dose In Ct

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Patient Doses for CT Examinations in Denmark

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Main properties of atoms and nucleus

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

Physics testing of image detectors

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays

Bonn Call-for-Action Joint Position Statement by the IAEA and WHO

The Phenomenon of Photoelectric Emission:

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

Lectures about XRF (X-Ray Fluorescence)

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS

Chapter XIII. PDD TAR TMR Dose Free Space

Human Exposure to Outdoor PLC System

PCXMC 2.0. User s Guide. Markku Tapiovaara, Teemu Siiskonen STUK-TR 7 / NOVEMBER 2008

APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS

Chapter 9 CALIBRATION OF PHOTON AND ELECTRON BEAMS

SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS

Training in Emergency Preparedness and Response

Physics 111: Lecture 4: Chapter 4 - Forces and Newton s Laws of Motion. Physics is about forces and how the world around us reacts to these forces.

Experimental study of beam hardening artefacts in photon counting breast computed tomography

Diagnostic x-ray imaging relies on the attenuation of

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

Elasticity Theory Basics

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver /9/07) by M. Jensen and J. E.

3rd International Symposium of the System of Radiological Protection Seoul, Korea, October 2015

1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: kv range

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector

Physical Quantities, Symbols and Units

Cargo Container X-ray Inspection Systems

Annals of the ICRP ICRP PUBLICATION 121. Radiological Protection in Paediatric Diagnostic and Interventional Radiology. Editor-in-Chief C.H.

Postgraduate Educational Course in Radiation Protection and the Safety of Radiation Sources Standard Syllabus

Radiation survey meters used for environmental monitoring

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

Physics 9e/Cutnell. correlated to the. College Board AP Physics 1 Course Objectives

Transcription:

2033-2 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetric Quantities and Units Introduction Donald McLean IAEA Vienna Austria

Joint ICTP-IAEA Advanced school on Dosimetry in Diagnostic Radiology: And its Clinical Implementation 11-15 May 2009; Miramare, Trieste, Italy Dosimetric quantities and units Introduction Donald McLean Dosimetry and Medical Radiation Physics Section Division of Human Health IAEA International Atomic Energy Agency

INTRODUCTION The connection to the medical profession is obvious. The term dose of radiation was initially used in a pharmacological sense: and is analogous to its meaning when used in prescribing a dose of medicine. Very soon it turned out that physical methods to describe a "dose of radiation" proved superior to any biological methods. IAEA 1. Slide 2

INTRODUCTION Radiation dosimetry is a now a pure physical science. Central are the methods for a quantitative determination of energy deposited in a given medium by directly or indirectly ionizing radiations. A number of physical quantities and units have been defined for describing a beam of radiation and the dose of radiation. This talk deals with the most commonly used dosimetric quantities and their units in diagnostic radiology. IAEA 1. Slide 3

DOSIMETRIC QUANTITIES AND UNITS Basic dosimetric quantities Application specific dosimetric quantities Quantities related to stochastic and deterministic effects Conversion coefficients for the assessment of organ and tissue doses IAEA 4

Basic dosimetric quantities Fluence Energy fluence Kerma and kerma rate Energy imparted Absorbed dose Based on ICRU Report 60 IAEA 5

RADIATION FIELD OR RADIOMETRIC QUANTITIES Radiation Field The term radiation field is a very general term that is used to characterize in a quantitative way the radiation in space consisting of particles. There are two very general quantities associated with a radiation field: the number, N of particles the energy, R transported by the particles (which is also denoted as the radiant energy) IAEA 2.1 Slide 2

RADIATION FIELD OR RADIOMETRIC QUANTITIES Radiation Field ICRU-Definition of particle number: The particle number, N, is the number of particles that are emitted, transferred, or received.unit: 1 ICRU-Definition of radiant energy: The radiant energy, R, is the energy (excluding rest energy) of particles that are emitted, transferred, or received. Unit: J For particles of energy E (excluding rest energy): R = E N IAEA 2.1 Slide 3

RADIATION FIELD OR RADIOMETRIC QUANTITIES Particle Fluence How can the number of particles be determined at a certain point in space? Consider a point P( ) in space within a field of radiation. Then use the following simple method: r In case of a parallel radiation beam, construct a small area da around the point P in such a way, that its plane is perpendicular to the direction of the beam. Determine the number of particles that intercept this area da. da P IAEA 2.2 Slide 1

RADIATION FIELD OR RADIOMETRIC QUANTITIES Particle Fluence In the general case of nonparallel particle directions it is evident that a fixed plane cannot be traversed by all particles perpendicularly. A somewhat modified concept is needed! The plane da is allowed to move freely around P, so as to intercept each incident ray perpendicularly. Practically this means: Generate a sphere by rotating da around P da Count the number of particles entering the sphere P IAEA 2.2 Slide 2

Fluence & Energy fluence The fluence, Φ, is the quotient dn by da, where dn is the number of particles incident on a sphere of cross-sectional area da, thus dn Φ = Unit: m da -2. The energy fluence, Ψ, is the quotient dr by da, where dr is the radiant energy incident on a sphere of cross-sectional area da, thus dr da Ψ = IAEA Unit: J m -2. 10

Example: Bremsstrahlung fluence RQR 5: 70 kv; HVL = 2.58 mm Al Target; W, 12 0, Total filtration; 2.6 mm Al At 75 cm for 1 mas Total fluence = 2030295 photons/sq.mm. = 2.03 x 10 12 photons m -2 70000 60000 50000 Photon per mm 2 40000 30000 20000 10000 0 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 IAEA Photon energy (kev) 11

Example: Bremsstrahlung energy fluence RQR 5: 70 kv; HVL = 2.58 mm Al Target; W, 12 0, Total filtration; 2.6 mm Al At 75 cm for 1 mas Total energy fluence = 81378447 kev/sq.mm. = 1.30 x 10-2 J.m -2 h=6.62 x 10-34 J/s h=4.13 x 10-18 kev/s Energy fluence (kev per mm 2 ) 3e+6 3e+6 2e+6 2e+6 1e+6 5e+5 0 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 IAEA Photon energy (kev) 12

Comparison of fluence and energy fluence 3.5 3.0 2.5 relative units 2.0 1.5 1.0 0.5 0.0 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 Photon energy (kev) IAEA 13

DOSIMETRIC QUANTITIES: FUNDAMENTALS General Introduction Common characteristics of Kerma and Absorbed Dose: They are generally defined as: radiation energy (transferred or absorbed) mass They can also be defined as: J kg J radiation field quantity mass interaction coefficient kg IAEA 3.1 Slide 2

Kerma and kerma rate The kerma, K, is the quotient de tr by dm, where de tr is the sum of the initial kinetic energies of all the charged particles liberated by uncharged particles in a mass dm of material, thus K K = = de t r dm Unit: J kg -1. The special name for the unit of kerma is gray (Gy). The kerma rate, dk dt Unit: J kg -1 s -1 or (Gy s -1 ). IAEA 15

DOSIMETRIC QUANTITIES Kerma Illustration of kerma: photons E k,3 E k,2 secondary electrons V E = E + E Collision energy transferred in the volume: tr k, 2 k, 3 where is the initial kinetic energy of the secondary electrons. E k E k,1 E k,1 Note: is transferred outside the volume and is therefore not taken into account in the definition of kerma! IAEA 4.1 Slide 3

Energy imparted The mean energy imparted, to the matter in a given volume equals the radiant energy, R in, of all those charged and uncharged ionizing particles which enter the volume minus the radiant energy, R out, of all those charged and uncharged ionizing particles which leave the volume, plus the sum, ΣQ, of all changes of the rest energy of nuclei and elementary particles which occur in the volume, thus ε ε = R in - R out + ΣQ Unit: J. For the photon energies used in diagnostic radiology, ΣQ is zero. IAEA 17

DOSIMETRIC QUANTITIES: FUNDAMENTALS Fundamentals of the Absorption of Radiation Energy Example for energy deposit ε i with Q = 0 (electron knock-on interaction): fluorescence photon, hν primary electron, E out ε in electron Auger electron 1 E A,1 Auger electron 2 E A,2 i in out A,1 A,2 δ electron, E δ ε = ε ( E + E + E + E + hν) δ IAEA 3.2 Slide 2

Absorbed dose The absorbed dose, D, is the quotient d ε by dm, where d D ε mass dm, thus = is the mean energy imparted to matter of dε d m Unit: J kg -1. gray (Gy). Where the production of Bremsstrahlung negligible, absorbed dose and kerma are numerically equal when secondary electron equilibrium is established. IAEA 19

Dose to air at PMMA/air interface 70 kv, 2.5 mm Al, 100 kv, 3.97 mm Al, 150 kv, 6.57 mm Al IAEA 20

ICRU versus IAEA basic dosimetry quantities Quantity ICRU IAEA Fluence Energy fluence Kerma Energy imparted Absorbed dose Exposure no Ψ K (K a for air) no D X Φ Ψ K (also for air) ε D no IAEA 21

Application specific dosimetric quantities Rationale for choice of quantities Incident air kerma Entrance surface air kerma X ray tube output Air kerma-area product Air kerma-length product Quantities for CT dosimetry Based on ICRU Report 74 IAEA 22

Rationale for choice of quantities Several application specific quantities have been found useful in the past for measurements in diagnostic radiology. However, there has been ambiguity in the names of the quantities and their (sometimes incorrect) use. ICRU 74 provides a consistent set of quantities, which is adopted in this Code of Practice IAEA 23

Incident air kerma & Entrance surface air kerma The incident air kerma, K i, is the kerma to air from an incident X ray beam measured on the central beam axis at the position of the patient or phantom surface. Only the radiation incident on the patient or phantom and not the backscattered radiation is included. The entrance surface air kerma, K e, is the kerma to air measured on the central beam axis at the position of the patient or phantom surface. The radiation incident on the patient or phantom and the backscattered radiation (B) are included. IAEA K 24 e = K i B

X ray tube output The X ray tube output, Y(d), is defined as the quotient of the air kerma at a specified distance, d, from the X ray tube focus by the tube currentexposure time product P It., thus Y(d) = K(d)/P It Unit: Gy (A s) -1 or Gy (ma s) -1 The tube current exposure time product, P It is also referred to in this Code of Practice as the tube loading. IAEA 25

Zero Dimensional quantities IAEA 26

Air kerma-area product The air kerma-area product, P KA, is the integral of the air kerma over the area of the X ray beam in a plane perpendicular to the beam axis, thus P KA = A K( x, y)dxdy Unit: Gy m 2 P KA has the useful property that it is approximately invariant with distance from the X ray tube focus (when interactions in air and extra-focal radiation can be neglected), as long as the planes of measurement do not include a significant contribution from backscattered radiation from the patient or phantom. IAEA 27

Air kerma-length product The air kerma-length product, P KL, is the integral of the air kerma over a line, L, thus P = KL K( z) L dz Unit: Gy m air kerma-length product is applied to the dosimetry of CT and to the dosimetry of dental panoramic examinations. In the literature for dental panoramic dosimetry, this quantity has been termed the "dose-width product" IAEA 28

Quantities for CT dosimetry (1) The computed tomography air kerma index, C a,100, measured free-in-air for a single rotation of a CT scanner is the quotient of the integral of the air kerma along a line parallel to the axis of rotation of the scanner over a length of 100 mm and the nominal slice thickness, T. The integration range is positioned symmetrically about the volume scanned, thus 1 50 C a +, 100 ( z) dz = K Unit: Gy T 50 Multi detector CT: 1 50 C a = K()dz z N simultaneously,100 IAEA N T 50 acquired slices 29

Quantities for CT dosimetry (2) The computed tomography air kerma index is also measured inside PMMA head and body phantoms 1 C W = PMMA, 100, c + 3 ( C 2 C ) PMMA, 100, The quantity C PMMA,100,c is measured at the centre of the standard CT dosimetry phantom and C PMMA,100,p is the average of values measured at four positions around the periphery of the same phantom. A weighted computed tomography dose index (CTDI) was first introduced by Leitz et al. and is used in IEC and the European CT Dosimetry protocol. p IAEA 30

Quantities for CT dosimetry (3) The subscript n is used to denote when the value of C a,100 or C W has been normalised to unit tube current-exposure time product, P It, thus n C C W W = : nca,100 = PIt A further quantity, C VOL, takes into account the helical pitch or axial scan spacing thus NT CW C CVOL = CW = ; n CVOL = l p P C a,100 where N is the number of simultaneously acquired tomographic slices, T the nominal slice thickness, l is the distance moved by the patient couch per helical rotation or between consecutive scans for a series of axial scans and P It is the tube loading for a single axial scan P It VOL It IAEA 31

Quantities for CT dosimetry (4) The quantity p = l NT is known as the CT pitch factor (or pitch) for helical scanning. The CT air kerma-length product determined for the standard CT dosimetry phantom and a complete CT examination, P KL,CT, is calculated using P KL, CT = where the index j represents each serial or helical scan sequence forming part of the examination, l j is the distance moved by the patient couch between or during consecutive scanner rotations and is the total tube loading for scan sequence j. This quantity is IAEA 32 analogous to the dose length product introduced in EC guidelines j n C VOL j l j P It j

ICRU versus IAEA application specific quantities Quantity ICRU IAEA Weighted CT air kerma index C K,PMMA,w C W Normalized weighted CT air kerma index n C K,PMMA,w n C W CT air kerma-length product P KL,CT P KL,CT (defined using n C VOL ) IAEA 33

Quantities related to stochastic and deterministic effects Organ and tissue dose Equivalent dose Effective dose IAEA 34

Organ and tissue dose The mean absorbed dose in a specified tissue or organ is given the symbol, D T, in ICRU Report 51. It is equal to the ratio of the energy imparted, ε T, to the tissue or organ to the mass, m T, of the tissue or organ, thus ε T DT = m T The mean dose to the glandular tissues within the breast has been referred to in the literature as both mean glandular dose and average glandular dose. The term mean glandular dose and the notation D G are used in this Code of Practice. IAEA 35

Equivalent dose The equivalent dose, H T, to an organ or tissue, T, is defined in ICRP Report 60. For a single type of radiation, R, it is the product of a radiation weighting factor, w R, for radiation R and the organ dose, D T, thus H T = w R D T Unit: sievert (Sv) The radiation weighting factor, w R, allows for differences in the relative biological effectiveness of the incident radiation in producing stochastic effects at low doses in tissue or organ, T. For X ray energies used in diagnostic radiology, w R is taken to be unity IAEA 36

Effective dose The effective dose, E, is defined in ICRP Report 60. It is the sum over all the organs and tissues of the body of the product of the equivalent dose, H T, to the organ or tissue and a tissue weighting factor, w T, for that organ or tissue, thus E = T w T H T Unit: sievert (Sv) The tissue weighting factor, w T, for organ T represents the relative contribution of that organ to the total detriment arising from stochastic effects for uniform irradiation of the whole body IAEA 37

Conversion coefficients for the assessment of organ and tissue doses A conversion coefficient, c, relates the dose to an organ or tissue to a readily measured or calculated dosimetric quantity, thus c = organ or tissue dose measured or calculatedquantity Suffices are added to c to indicate the two quantities that are related, for example the coefficient c D K = T i relates the organ dose, D T, to T, i D K the incident air kerma, K i IAEA 38

Thank you for your attention IAEA 39