Calibration Myths, Misunderstandings & Mistakes. How Not to Worry about Calibration in Your Next ISO Audit

Similar documents
calibrate confidence with Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Abstract Introduction White Paper

Quality Assurance Manual January 2008 Revision I Revision date of 01/17/13. Pi Tape Corporation. 344 North Vinewood St. Escondido CA USA

Procedure for Equipment Calibration and Maintenance

Calibration Certificate

Calibration Service Selection Guide

Copies of this document shall be classified either as Controlled or Uncontrolled, and shall be identified as such on the cover sheet.

Measuring equipment in a laboratory.

Regulatory Requirements for Medical Device Calibration Programs

Quality Management System Policy Manual

Measuring instruments in industry.

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z ALLTITE, INC Murdock, Wichita, KS Bruce Demaree Phone:

calibrate confidence with Risk Management: The Critical Role of Measurement and Calibration WHITE PAPER SERVICES

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

CALIBRATION PRINCIPLES

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

Luminus Devices, Inc Luminus Testing Laboratory Quality Management Systems Manual

MEASURING MACHINES. Pratt & Whitney METROLOGY LABORATORY. Measurement Systems, Inc.

EVALUATION AND ACCREDITATION OF SUPPLIER TEST FACILITIES GP 10

Enhanced calibration High quality services from your global instrumentation partner

Calibrating DC Shunts: Techniques and Uncertainties

Testing Facility Laboratory Manual

Calibration of Electronic Test Equipment: An Introduction to ISO & Accreditation

Model 601A02. Low-cost Industrial ICP Accelerometer. Installation and Operating Manual

ENERGY STAR Program Requirements Product Specification for Data Center Storage. Draft 2 Test Method March 2013

OPERATIONAL STANDARD

Agilent P940xA/C Solid State PIN Diode Switches

MET/TEAM Test Equipment Asset Management Software

Understand, Maintain and Service Your Torque Wrench

SUPPLIER Form #: 007 AUDIT QUESTIONNAIRE

Supplier Quality Agreements

QUALITY MANUAL REVISION RECORD

ISO 9001: 2008 Boosting quality to differentiate yourself from the competition. xxxx November 2008

Calibration Solutions. be certain.

Beamex. Calibration White Paper. How often should instruments be calibrated

ISO 9001:2008 Audit Checklist

Lab Service, Asset Management and Extending the Life of Your Lab Equipment. Scott Greenwood Director, Services

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.

GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards

The Fulfillment of AS 9100 Rev C Requirements by EnterpriseIQ

DICKSON PW4 DICKSON. Pressure Chart Recorders. Applications & Specifications. Product. Getting Started, Charts & Accessories.

Keysight Technologies Specifications Guidelines. Technical Overview

DICKSON PR100/PR300/PR500 DICKSON. Pressure Logger Operation. Contents:

THERMAL INSULATION MATERIALS

Beamex. Calibration White Paper. How often should instruments be calibrated?

Quality Management System General

CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company

Environment-Laboratory Ambient Conditions

Non-Controlled Copy. Quality Manual Revision Log. Rev # Date Description

A.1 Sensor Calibration Considerations

RTP s NUCLEAR QUALITY ASSURANCE PROGRAM

1 For more information T: / E: DAP@ul.com / W: ul.com/dap

CALIBRATION DATA MANAGEMENT: MEETING THE REPORTING REQUIREMENTS OF ISO/IEC FDIS

NIST HANDBOOK 150 CHECKLIST

How DCMA Helps To Ensure Good Measurements

User s manual FLIR VP50/VP52 Non-contact AC voltage detector

Quality Management System Manual ISO9001:2008

Load Cell Troubleshooting

10EaZy SW White Paper Choosing the right hardware

TITLE: Quality System Manual DOCUMENT NUMBER: QMS

The Prophotonix (UK) Ltd Quality manual

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z

Automated Calibration Software Guardbanding. Author: Matt Nicholas Fluke Corporation. Speaker: Lars Andersson Fluke Europe B.V

American Association for Laboratory Accreditation

Tips for a Successful Diesel Retrofit Project

QUALITY MANAGEMENT SYSTEMS MANUAL

ISO17025 INTERPRETATION DOCUMENT FOR CEC TEST METHODS

Metrology, Quality and ISO 9000

Revision Date Author Description of change Jun13 Mark Benton Removed Admin. Manager from approval

ISO-9001:2000 Quality Management Systems

3M CORPORATE METROLOGY SERVICES St. Paul, MN

ISO 9001:2008 Quality Management System Requirements (Third Revision)

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

Introduction to EMC Antenna Calibration Methods:

Internet-Based Calibration of a Multifunction Calibrator

UNCONTROLLED COPY FOR REFERENCE ONLY

CUSTOMER SPECIFIC REQUIREMENTS

RUSKA PRODUCTS - PRICE LIST

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

American Association for Laboratory Accreditation

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Quality management systems

(Uncertainty) 2. How uncertain is your uncertainty budget?

Essential Electrical Concepts

RF & Microwave Contract Manufacturing

NORTH AMERICA OPERATIONS. (Fairmont and Montreal Facilities) QUALITY MANUAL. Prepared to comply with the requirements of ISO 9001:2008

R&S ZCxxx Millimeter-Wave Converters Specifications

FMC Technologies Measurement Solutions Inc.

Experience with taximeters with a look at the corresponding tables

ISO 9001:2008 Requirements Explained - An Adobe PDF File for Use on a Network System

a. Are the foundry s employees trained and familiar with portions of the system applicable to their position?

Selection and use of ISO 9000

Calibration. How can you get lower costs for the calibration? What is traceability? What is measurement uncertainty? What is calibration?

Integrating Sound Level Meter and Datalogger

Eagle Machining, Inc. Quality Management System

SUPPLIER QUESTIONNAIRE

Technical Management? We Don t Need No Technical Management!

Multimeter measurements on variable frequency drives using the new Fluke 289 DMM

Basic Steps to Establish ISO/IEC Compliant Testing Laboratories

Transcription:

Calibration Myths, Misunderstandings & Mistakes How Not to Worry about Calibration in Your Next ISO Audit ASQ Section 1508 St Petersburg, FL May 9, 2011

What is calibration? To verify by comparison It is part of the science of measurement known as metrology Calibrations are performed on instruments used to measure (IMTE inspection, measurement, test equipment) often referred to as gages Another instrument of known magnitude or correctness is used called a standard It is not the adjustment, cleaning, repair or other service performed on the gage

What is verified in a calibration? Gage parameters There are 5 independent units of measure temperature interval linear distance electrical current frequency mass All other measurement types are one or more of these independent units example torque is a combination of length and mass measured in foot pounds or Newton meters

What is verified in a calibration? Most common parameters in manufacturing Physical/dimensional Force, mass, temperature, humidity, torque, length Electronic High frequency Time, RF power, DC to 50 Ghz Low frequency AC/DC current and voltage, resistance

Traceability An unbroken chain of comparisons NIST (National Institute of Standards and Technology) is part of the US Dept of Commerce NIST traceable means that the standard used to perform a calibration is traceable back to NIST. NIST traceable does not mean a calibration is accredited

What does accreditation mean? In the US it usually means ISO/IEC 17025 In January 2009 17025 superseded 9001:2008 Prior to 1999 referred to as Guide 25 Multiple accrediting bodies (AB) A2LA, NVLAP, L-A-B, etc. Only an accredited lab can issue an accredited certificate of calibration A 17025 accredited lab no longer needs to maintain separate 9001 certification

Tiers or levels of calibration Pass/fail sometimes called short form With data sometimes called as found/as left or long form Accredited includes uncertainty data Only accredited cals have the AB s logo A lab can only perform an accredited cal if that parameter is included under their scope of accreditation

Sample first page of a Scope of Accreditation

Accreditation issues Can an accredited lab perform a non accredited calibration? Yes Lab locations are accredited not the company A company can advertise being accredited at a location, but check the scope of accreditation to determine what parameters are accredited Non accredited cals are typically performed std) to ANSI/NCSL Z540-1-1994 (the old mil

How does calibration fit into your quality management system? Quality management systems call for an effective metrology system which includes formal, periodic, and documented calibration of all measuring instruments. ISO 9000 and ISO 17025 sets of standards and require that these traceable actions are to a high level and set out how they can be quantified. Documentation can be divided into two types: One related to the task The other related the administrative program.

Example of what task documentation must address When should a gage be calibrated? With a new instrument When a specified time period is elapsed When a specified usage (operating hours) has elapsed When an instrument has had a shock or vibration which potentially may have put it out of calibration Sudden changes in weather or environment Whenever observations appear questionable

Examples of what administration of the program documentation must address Where should the gage be calibrated Who is responsible for ensuring the gages are calibrated What tracking tool is utilized ie database Where are the certificates of calibration maintained

ISO audits Auditors usually dig into calibration Easy to find errors There is a lot of misinformation/misunderstandings about calibration they like to find The administrative documentation is usually the weak link For outsourced cals the supplier s accreditation is usually sufficient to satisfy an auditor Gages without supporting documentation (stickers, certs) of current cals are common targets

Avoiding calibration issues in your ISO audits Establish on-going program Don t just clean up a few weeks/months before audit Assign a central point of control Distributed responsibility gives inconsistent results Person assigned as central POC must have quality manager s on-going support Be creative to meet your specific issues/problems and document those processes Acquire or develop an easy to use database Track and report past due gages Perform internal audits

Questions?

Question Some manufacturers recommend calibration for their equipment such as digital multi-meters but don t provide certificates of calibration or affix stickers when the units are purchased. Then there are some that do offer the certs and stickers at an additional fee. Does NIST actually have a standard that recommends calibration intervals based upon the device type? If so is it possible to gain access to this data? Answer NIST does not recommend cal cycles. The best place to start to decide what a cal interval should be is what the OEM recommends. Then in order to properly determine what the cal interval should be for a particular item look at a history of the recent calibration reports and determine if the history of the item is stable enough to lengthen the recommended cal cycle. This works the other way too if you have a unit that is failing cal every year then the cal cycle needs to be moved from maybe 1 year to every six months. And one other major factor to consider before changing any recommended cal cycle is what the calibration cost is in relation to the potential loss of revenue due to bad or faulty parts being shipped out

Question We have come to the conclusion that digital multi-meters need to be considered somewhat as disposable equipment. Many of the DMMs are well built and priced such that it is more economical to buy them, use them for a year and then replace them as opposed to recalibrating them. Have you seen any trends such as this? Answer The average cost of a decent multi meter is around $300. The average cal cost is around $35 and most meters can last several years if maintained correctly, so in short if a good meter such as a Fluke is purchased it can last a very long time and should only be replaced when repair charges exceed 50-75% of the replacement cost. Also many new Fluke meters have a life time warranty.

Question Can calibration labs can be a useful reference for test and measurement equipment based upon their observations of the as found condition of various equipment? In other words I have seen some quality standards written where frequency of calibration is based upon the calibration customer s standard as opposed to the manufacturer or NIST standards. Do calibration labs accumulate data to support this sort of recommendation or is totally based upon data accumulated by the calibration customer? I know from my prior dealings with TMI you have been able to make equipment purchase recommendations. Answer Yes, a good calibration lab stores all of this data and can filter thru it and make recommendations on their use. TMI calibrates well over a 100,000 items each year and have a lot of data in the parameters we test. However one thing that must always be considered is how the equipment is treated. Even the best and most expensive test equipment can fail very fast if mistreated or mishandled.