CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School

Purchasing a cardiac CT scanner: What the radiologist needs to know

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Assessing Radiation Dose: How to Do It Right

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

How To Improve Your Ct Image Quality

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Table of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

American College of Radiology CT Accreditation Program. Testing Instructions

Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives

GE Healthcare. Great treasures in small places. BrightSpeed Elite

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

GE Healthcare. Revolution EVO. More than just high tech. Higher purpose.

Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

Low-dose CT for Pulmonary Embolism

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

The Challenge of CT Dose Records

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

Practical exercise: Effective dose estimate in CT

State-of-the-Art Technology in Cardiac CT

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

IAEA HUMAN HEALTH SERIES

Percutaneous Abscess Drainage

Dose Modulation Technique in CT

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

Diagnostic Exposure Tracking in the Medical Record

NEMA XR Standard Attributes on CT Equipment Related to Dose Optimization and Management

Techniques and Applications of Automatic Tube Current Modulation for CT 1

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

Data. microcat +SPECT

The power of Multislice CT Philips MX 6000 Dual

How To Prepare For A Ct Scan

Abdomen X-Ray (AXR) Collimation is ideally from diaphragms to lower border of the symphysis pubis and the lateral skin margins.

Development of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35

A d v a n c e d 1 2 8

Mx8000 IDT 16 provides dose control methodologies like DoseRight ACS (Automatic Current Selection) and DoseRight DOM (dynamic dose modulation).

Cardiac CT for Calcium Scoring

Computed Tomography (CT) - Sinuses

Gestión global de la dosis en TC. Sistema de registro y gestión

Children's (Pediatric) CT (Computed Tomography)

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

Prepublication Requirements

Recommendations for cross-sectional imaging in cancer management, Second edition

X-ray (Radiography) - Chest

Computed Tomography Radiation Safety Issues in Ontario

The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement.

Pediatric Hospitals Bring Low-dose CT to the Middle East

Running head: OBESITY AND MEDICAL IMAGING 1. Obesity and Medical Imaging. November 13, 2012

INTRODUCTION. A. Purpose

The Exposure Index and Its Standardization

X-ray (Radiography) - Abdomen

Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.

CT Colonography. What is CT Colonography?

Brilliance CT 64-channel confi guration

Computed Tomography (CT) - Spine

CT scans and IV contrast (radiographic iodinated contrast) utilization in adults

WHERE IN THE WORLD JILL LIPOTI?

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Siemens Computed Tomography

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

AAPM Medical Physics Practice Guideline 1.a: CT Protocol Management and Review Practice Guideline

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

OPTIMIZING PATIENT EXPOSURE TO IONIZING RADIATION (OPEIR) MEASURES GROUP OVERVIEW 2015 PQRS OPTIONS FOR MEASURES GROUPS:

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen.

Development of pediatric CT protocols to obtain minimal doses in PET/CT studies a physicists approach.

3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays


Fundamentals of Cone-Beam CT Imaging

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

DENTAL Cone beam 3D X-RAY SYSTEM with

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Patient Prep Information

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad.

Multi-slice Helical CT Scanning of the Chest

Computed Tomography (CT) - Body

Patient sample criteria for the OPEIR Measures Group are all patients regardless of age, that have a specific CT procedure performed:

SWABIK Project. Software Tools for DICOM Media Exchange in Clinical Research: the. SWABIK Project

SITE IMAGING MANUAL ACRIN 6698

Acute abdominal pain in the elderly patient: Impact of early MDCT examination on diagnosis and management

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Mammography. What is Mammography?

Protocol Management and Review Strategies The MD Anderson Experience

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

Patient Exposure Doses During Diagnostic Radiography

Transcription:

DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION OF PHYSICS TO MEDICINE AND BIOLOGY, ENCOURAGE INTEREST AND TRAINING IN MEDICAL PHYSICS AND RELATED FIELDS ("AAPM"), 'AS IS' WITHOUT WARRANTIES OR CONDITIONS OF ANY KIND, WHETHER ORAL OR WRITTEN, EXPRESS OR IMPLIED. AAPM SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTIES OR CONDITIONS OF MERCHANTABILITY, SATISFACTORY QUALITY, NONINFRINGEMENT AND FITNESS FOR A PARTICULAR PURPOSE. SOME JURISDICTIONS DO NOT ALLOW EXCLUSIONS OF IMPLIED WARRANTIES OR CONDITIONS, SO THE ABOVE EXCLUSION MAY NOT APPLY TO YOU. YOU MAY HAVE OTHER RIGHTS THAT VARY ACCORDING TO LOCAL LAW. TO THE EXTENT ALLOWED BY LOCAL LAW, IN NO EVENT WILL AAPM OR ITS SUBSIDIARIES, AFFILIATES OR VENDORS BE LIABLE FOR DIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL OR OTHER DAMAGES (INCLUDING LOST PROFIT, LOST DATA, OR DOWNTIME COSTS), ARISING OUT OF THE USE, INABILITY TO USE, OR THE RESULTS OF USE OF THE PROVIDED INFORMATION, WHETHER BASED IN WARRANTY, CONTRACT, TORT OR OTHER LEGAL THEORY, AND WHETHER OR NOT ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. YOUR USE OF THE INFORMATION IS ENTIRELY AT YOUR OWN RISK. THIS INFORMATION IS NOT MEANT TO BE USED AS A SUBSTITUTE FOR THE REVIEW OF SCAN PROTOCOL PARAMETERS BY A QUALIFIED AND CERTIFIED PROFESSIONAL. USERS ARE CAUTIONED TO SEEK THE ADVICE OF A QUALIFIED AND CERTIFIED PROFESSIONAL BEFORE USING ANY PROTOCOL BASED ON THE PROVIDED INFORMATION. AAPM IS NOT RESPONSIBLE FOR A USER'S FAILURE TO VERIFY OR CONFIRM APPROPRIATE PERFORMANCE OF THE PROVIDED SCAN PARAMETERS. SOME JURISDICTIONS DO NOT ALLOW THE EXCLUSION OR LIMITATION OF LIABILITY FOR DAMAGES, SO THE ABOVE LIMITATION MAY NOT APPLY TO YOU. The disclaimer on page 1 is an integral part of this document. 1

ROUTINE ADULT ABDOMEN/PELVIS CT PROTOCOLS Indications (include but are not limited to) Evaluation of abdominal, flank, or pelvic pain Evaluation of known or suspected abdominal or pelvic masses or fluid collections Evaluation of primary or metastatic malignancies Evaluation of abdominal or pelvic inflammatory processes Evaluate infections of the abdomen and pelvis Evaluate patients with fever or suspected abscess To identify the cause of bowel obstruction Evaluation of weight loss Clarification of findings from other imaging studies or laboratory abnormalities. For reference, see ACR SPR Practice Guideline for the Performance of Computed Tomography (CT) of the Abdomen and Computed Tomography (CT) of the Pelvis Diagnostic Tasks (include but are not limited to) Detect soft tissue masses and abnormal fluid collections and determine sizes Identify abnormal collections of blood Identify air outside the intestinal tract Detect nodules or soft tissue masses adjacent to vascular structures Detect calcifications in abnormal locations or in abdominal and pelvic organs Characterize soft tissue edema around the organs of the abdomen and pelvis Key Elements Patient positioning is especially important when using tube current modulation. Scanning should be performed helically whenever possible. Automatic Exposure Control (AEC): Each manufacturer has unique nomenclature and operating characteristics for their AEC system(s). Users must become very familiar with how the AEC systems on their particular scanners operate. See Singh et al. Automatic Exposure Control in CT: Applications and Limitations. JACR 2011;8(6):446-449. Contrast Oral: Per radiologist Injected: Certain indications require administration of intravenous contrast media. Intravenous contrast enhancement should be performed as directed by the supervising radiologist using appropriate injection protocols and in accordance with the ACR-SPR Practice Guideline for the Use of Intravascular Contrast Media and the ACR Manual on Contrast Media. Patient Positioning Center the patient within the gantry; this is critical for proper functioning of AEC systems. Patient supine, typically feet first Scan Range Scan from top of liver to either iliac crest or pubic symphysis, depending on clinical indications. Suspension of Respiration Patient should be instructed to hold his/her breath at end of inspiration. 2

Additional Image Reconstructions Certain indications may require that images be reconstructed in coronal and/or sagittal planes. Very thin images (approximately 1 mm) may need to be reconstructed to serve as source images for the sagittal and/or coronal reformatted images. Creation, use, and archival of these additional images are at the discretion of the supervising radiologist and/or departmental policy. Very large datasets may result from these additional reconstructions. Radiation Dose Management AEC should be used whenever possible. Pay careful attention to the values selected to define the desired level of image quality (eg, Noise Index, Quality Reference mas, Standard Deviation). Each manufacturer will have recommendations unique to their systems and system features. Be sure to work with your CT equipment manufacturer and a qualified medical physicist to ensure safe and appropriate operation of AEC systems. If more than one CT localizer radiograph is acquired, AEC systems from different manufacturers can differ with respect to which one is used to determine ma and/or kv settings. Please refer to individual manufacturer protocol instructions. CTDI measurements and calculations Some manufacturers utilize a z-axis flying focal spot, in which two unique projections are acquired at the same z-axis table position. When this technique is used, we identify it with **. The CTDIvol on the console accurately accounts for use of this feature. 3

Approximate Volume CT Dose Index (CTDIvol) Values Adult Routine Abdomen/Pelvis CT Protocols Version 1.1 08/07/2015 Approximate values for CTDIvol are listed for three different patient sizes: Average Patient 70-90 155-200 15-25 The approximate CTDIvol values are for reference only and represent a dose to the CT Dose Index phantom under very specific conditions. The CTDIvol displayed on the scanner for a patient of a given size should be similar, but not necessarily an exact match, to those listed in the above table. The provided values are all based on the 32 cm diameter body CTDI phantom. It is essential that users recognize that the CTDIvol values reported on the user console prior to acquiring CT localizer radiographs on a particular patient do not represent the CTDIvol that will be delivered during that patient s scan. CT systems rely on the CT localizer radiograph to 1) estimate the patient s size, 2) determine the tube current settings for each tube angle and table position that will yield the requested level of image quality, and 3) calculate the average CTDIvol for the patient over the prescribed scan range. Until the CT localizer radiograph is acquired, the reported CTDIvol is not patient-specific, but is based on a generic patient size. The CTDIvol values provided here are approximate, and are intended only to provide reference ranges for the user to consider. They are for a routine CT of an adult s abdomen/pelvis for the general indictations given at the beginning of this document. Other indications or diagnostic tasks may have different image quality and dose requirements, and hence reasonable ranges of CTDIvol may differ according to those requirements. In this document, a small patient is considered to be approximately 50-70 kg (110-155 lbs), an average patient approximately 70-90 kg (155-200 lbs), and a large patient 90-120 kg (200-265 lbs). However, weight is not a perfect indication of patient size. A person s height, gender and distribution of weight across the body also must be taken into account. The thickness of the body over the area to be scanned is the best indication of patient size. Bodymass index (BMI) may also be considered: Underweight = BMI <18.5 Normal weight = BMI of 18.5 24.9 Overweight = BMI of 25 29.9 Obesity = BMI of 30 or greater It is recognized that the median (50 th percentile) patient size for adults in the USA is larger than 70 kg. However, the 70 kg patient represents the Reference Man, as defined by the International Commission on Radiation Protection (ICRP), upon which AEC systems and tissue weighting factors (used for effective dose estimation) are based. 4

INDEX OF ADULT ROUTINE ABDOMEN-PELVIS PROTOCOLS (by manufacturer) GE Hitachi Neusoft Philips Siemens Toshiba 5

ADULT ABDOMEN & PELVIS ROUTINE (Selected GE scanners) SCOUT: AP if automatic exposure control is used. PA if manual ma is used. Landmark at xyphoid process. Range: S60-I500. Lateral scout optional, but recommended to ensure accurate vertical (AP) centering for AEC operation. If two Scouts are obtained, the second one must cover the entire intended scan range, as it is used to determine ma settings. GE LightSpeed Ultra (8) BrightSpeed 16 Select LightSpeed 16 BrightSpeed 16 LightSpeed Pro 16 Scan Type Helical Helical Helical Helical Rotation Time (s) 0.6 0.8 0.5 0.5 Detector Configuration 8 x 1.25 mm (10 mm, 2i) 16 x 1.25 mm (20 mm, 4i) 16 x 1.25 mm (20 mm, 4i) 16 x 1.25 mm (20 mm, 4i) Pitch 1.375 0.938 1.375 1.375 Table Feed/Interval (mm) 13.5 18.75 27.5 27.5 kv 120 120 120 120 Auto-mA range 50-440 50-350 100-440 50-650 Noise Index (NI)* 11.57 12 11.57 11.57 SFOV Large Large Large Large Plane Axial Axial Axial Axial Algorithm Stnd Stnd Stnd Stnd Recon Mode Plus Plus Plus Plus Thickness (mm)* 5 5 5 5 Interval (mm) 5 5 5 5 Plane Axial Axial Axial Axial Algorithm Stnd Stnd Stnd Stnd Recon Mode Plus Plus Plus Plus Thickness (mm) 1.25 1.25 1.25 1.25 Interval (mm) 1.25 1.25 1.25 1.25 *The Noise Index value and the primary () image reconstruction thickness both strongly impact CTDIvol and patient dose. See: Kanal KM et al. Impact of Operator-Selected Image Noise Index and Reconstruction Slice Thickness on Patient Radiation Dose in 64-MDCT. AJR 2007; 189: 219-225. 6

ADULT ABDOMEN & PELVIS ROUTINE (selected GE scanners)(continued) SCOUT: AP if automatic exposure control is used. PA if manual ma is used. Landmark at xyphoid process. Range: S60-I500. Lateral scout optional, but recommended to ensure accurate vertical (AP) centering for AEC operation. If two Scouts are obtained, the second one must cover the entire intended scan range, as it is used to determine ma settings. GE LightSpeed VCT Discovery CT750 HD LightSpeed VCT (w/asir) Discovery CT750 HD (w/asir) Scan Type Helical Helical Helical Helical Rotation Time (s) 0.5 0.5 0.5 0.5 Detector Configuration 64 x 0.625 mm (40 mm, 8i) 64 x 0.625 mm (40 mm, 8i) 64 x 0.625 mm (40 mm, 64i) 64 x 0.625 mm (40 mm, 64i) Pitch 1.375 1.375 0.984 0.984 Table Feed/Interval (mm) 55 55 39.37 39.37 kv 120 120 120 120 Auto-mA 50-650 100-650 50-670 80-750 Noise Index (NI)* 11.57 11.57 18.0 (DR 50%) 25.2 SFOV Large Large Large Large ASiR None None SS50 SS50 Plane Axial Axial Axial DMPR create Sag/Cor reformat Axial DMPR create Sag/Cor reformat Algorithm Stnd Stnd Stnd Stnd Recon Mode Plus Plus Plus Plus Thickness (mm)* 5 5 5 5 Interval (mm) 5 5 5 5 ASiR None None SS50 SS50 Plane Axial DMPR create Sag/Cor reformat Axial DMPR create Sag/Cor reformat Algorithm Stnd Stnd Stnd Stnd Recon Mode Plus Plus Plus Plus Thickness (mm) 0.625 0.625 0.625 0.625 Interval (mm) 0.625 0.625 0.625 0.625 ASiR None None SS50 SS50 *The Noise Index value and the primary () image reconstruction thickness both strongly impact CTDIvol and patient dose. See: Kanal KM et al. Impact of Operator-Selected Image Noise Index and Reconstruction Slice Thickness on Patient Radiation Dose in 64-MDCT. AJR 2007; 189: 219-225. Axial Axial 7

ADULT ABDOMEN & PELVIS ROUTINE (selected HITACHI scanners) SCANOGRAM: PA, scan from above diaphragm to below ischium. HITACHI CXR4 ECLOS 16 SCENARIA 64 Scan Type Volume Volume Volume Rotation Time (s) 0.8 0.8 0.75 Detector Configuration 2.5 mm x 4 1.25 mm x 16 0.625 mm x 64 Pitch 1.25 1.0625 0.578 Speed (mm/rot) 6.25 21.25 23.125 kv 120 120 120 ma 225 300 IntelliEC Adaptive ma/intelliec YES YES SD-9.5 SFOV 500 500 500 Series Description Abdomen Pelvis Routine Abdomen Pelvis Routine Abdomen Pelvis Type Volume Volume Volume Algorithm Abd STD 4 Abd STD 32 Abd STD 30 Thickness (mm) 5 5 5 Interval (mm) 5 5 5 Series Description Thins Thins Thins for MPR Type Volume Volume Volume Algorithm Abd STD 4 Abd STD 32 Abd STD 30 Thickness (mm) 2.5 1.25 1.0 Interval (mm) 1.25 0.625 0.5 8

ADULT ABDOMEN & PELVIS ROUTINE (selected NEUSOFT scanners) SURVIEW: AP, 120 kvp, 40 ma. Scan from above diaphragm to below ischium. NEUSOFT NeuViz DUAL NeuViz 16 Scan Type Helical Helical Rotation Time (s) 1.0 0.75 Collimation 2 5 mm 16 1.5 mm kvp 120 120 mas 265 368 Pitch 1.5 0.8631 Type Axial Abdomen/Pelvis (Axial)* Filter Body Standard (B) SB Thickness (mm) 5.0 5.0 Increment (mm) 5.0 5.0 Type Small Recon (Axial)* Filter B/C N / A Thickness (mm) 2.0 Increment (mm) 2.0 RECON 3 Type Coronal* Filter SB N / A Thickness (mm) 5.0 Increment (mm) 5.0 *Adaptive Filter and Metal Artifact Reduction can be selected prospectively or retrospectively. 9

ADULT ABDOMEN & PELVIS ROUTINE (selected PHILIPS scanners) SURVIEW: AP, 500 mm length (or adjusted to patient size), starting at mid-sternum (feet first) or pubic symphysis (head first). If two Surviews are obtained, the first one must cover the entire intended scan range, as it is used to determine ma settings. PHILIPS Brilliance 16 Brilliance 64 Ingenuity CT Brilliance ict SP Brilliance ict Scan Type Helical Helical Helical Helical Helical Rotation Time (s) 0.75 0.5 0.75 0.5 0.5 Collimation 16 1.5 mm 64 0.625 mm 64 0.625 mm 64 0.625 mm 128 0.625 mm kvp 120 120 120 120 120 mas DoseRight DoseRight DoseRight DoseRight DoseRight DoseRight ACS ON ON ON ON ON Pitch 0.93 0.75 1.0 1.17 1.0 SP Filter Yes Yes Yes Yes Yes Adaptive Filter Yes Yes Yes Yes Yes Resolution Setting Standard Standard Standard Standard Standard Type Axial Axial Axial Axial Axial Filter B/C B/C B/C B/C B/C Thickness (mm) 5 5 5 5 5 Increment (mm) 5 5 5 5 5 Type Coronal Coronal Coronal Coronal Coronal Filter B/C B/C B/C B/C B/C Thickness (mm) 3 3 3 3 3 Increment (mm) 3 3 3 3 3 RECON 3 Type Axial Axial Axial Axial Axial Filter YA YC YC YC YC Thickness (mm) 2 0.9 0.9 0.9 0.9 Increment (mm) 1 0.45 0.45 0.45 0.45 RECON 4 Type Coronal Coronal Coronal Coronal Coronal Filter YA YC YC YC YC Thickness (mm) 1 1 1 1 1 Increment (mm) 1 1 1 1 1 10

ADULT ABDOMEN & PELVIS ROUTINE (selected SIEMENS scanners) TOPOGRAM: AP, 512 or 768 mm. Scan from above diaphragm to below ischium. If two Topograms are obtained, both will be used to determine ma settings. SIEMENS Sensation 16 Sensation 64 Definition (dual source, 64-slice) Definition AS (128-slice) Definition Flash (dual source, 128-slice) Software version VB30 VB30 VA34 VA44 VA44 Scan mode Spiral Spiral Spiral Spiral Spiral Tube voltage (kv)* 120 120 120 120 120 Qual. Ref. mas (QRM) 200 200 210 210/150*** 210/150*** Rotation time (s) 0.5 0.5 0.5 0.5 0.5 Acq. (Detector Configuration) 16 x 1.5 mm **64 0.6 mm (32 x 0.6 mm = 19.2 mm) **64 0.6 mm (32 x 0.6 mm = 19.2 mm) **128 0.6 mm (64 x 0.6 mm = 38.4 mm) **128 0.6 mm (64 x 0.6 mm = 38.4 mm) Pitch 0.8 1.4 0.6 0.6 0.6 Dose modulation CARE Dose4D CARE Dose4D CARE Dose4D CARE Dose4D CARE Dose4D Kernel B30f B30f B30f B30f/I30f*** B30f/I30f*** Slice (mm) 5 5.0 5.0 5.0 5.0 Slice increment (mm) 5 5.0 5.0 5.0 5.0 Kernel B20f B20f B20f B20f/I26f*** B20f/I26f*** Slice (mm) 1.0 1.0 1.0 1.0 1.0 Slice increment (mm) 0.7 0.7 0.7 0.7 0.7 * If scanner is equipped with automatic kv selection (CARE kv), this should be activatedby selecting On. For non-contrast-enhanced exams, a Reference kv of 120 and a Dose saving optimized for slider position of 3 is recommended. For contrast-enhanced exams, a Reference kv of 120 and a Dose saving optimized for slider positionof 7 is recommended. For exams requiring both a non-contrast-enhanced and a contrast-enhanced scan, where the change in mean CT number of a region may be relevant to the diagnosis, it is important that the kv be the same for both scans. Hence, the same slider position should be used for both scans. Typically, the slider position for the contrast-enhanced scan (e.g. 7) is used in this scenario. ** Indicates that a z-axis flying focal spot technique is used to obtain twice as many projections per rotation as detector rows. *** With SAFIRE and a strength setting of 3. 11

ADULT ABDOMEN & PELVIS ROUTINE (selected TOSHIBA scanners) SCANOGRAM: Dual Scano: AP and LAT, from 800 mm above diaphragm to below ischium. Both scanograms are used for tube current modulation. Toshiba Aq16 Aq32 Aq64 AqPremium AqONE Scan Type Helical Helical Helical Helical Helical Rotation Time (s) 0.5 0.5 0.5 0.5 0.5 Detector Configuration 16 x 1.0 mm 32 x 0.5 mm 64 x 0.5 mm 160 x 0.5 mm 160 x 0.5 mm Pitch Std (0.9375) Std (0.844) Std (0.828) Std (0.869) Std (0.869) Speed (mm/rot) 15 13.5 26.5 69.5 79.5 kv 120 120 120 120 120 ma SURE Exposure SURE Exposure SURE Exposure SURE Exposure SURE Exposure SURE Exposure Std (SD = 12.5) Std (SD = 12.5) Std (SD = 12.5) Std (SD = 12.5) Std (SD = 12.5) SFOV L L L L L Type Axial Axial Axial Axial Axial SURE IQ Setting* Body Std Axial Body Std Axial Body Std Axial Body Std Axial Body Std Axial Thickness (mm) 5 5 5 5 5 Interval (mm) 5 5 5 5 5 VOLUME RECON Type Axial Axial Axial Axial Axial SURE IQ Setting* Body Std Volume Body Std Volume Body Std Volume Body Std Volume Body Std Volume Thickness (mm) 1 0.5 0.5 0.5 0.5 Interval (mm) 5 5 5 5 5 Type Coronal Coronal Coronal Coronal Coronal SURE IQ Setting* Body Std Body Std Body Std Body Std Body Std Thickness (mm) 4 4 4 4 4 Interval (mm) 4 4 4 4 4 RECON 3 Type Sagittal Sagittal Sagittal Sagittal Sagittal SURE IQ Setting* Body Std Body Std Body Std Body Std Body Std Thickness (mm) 4 4 4 4 4 Interval (mm) 4 4 4 4 4 *The SURE IQ setting determines the reconstruction FC as well as other post-processing and reconstruction options, such as AIDR. The SURE IQ settings listed in this document refer to the manufacturer s default settings. 12